General GC1000 Mark II Communication


Specifications Function and GCMT Gas Chromatograph
Maintenance Terminal
GS 11B03G03-01E

n OVERVIEW
Two types of communications are provided for GC1000 Mark II process gas chromatograph: DCS and personal
computer (PC) communications.
DCS communication transmits analytical and other data to a DCS systems, such as CENTUM, in digital form.
PC communication is provided for maintenance of GC1000 Mark II. By incorporating GCMT package software in
PC, the PC can keep chromatograms, check them by superimposition, and save other data. This makes it possible
to handle higher quality data and makes operation easier than conventional method that mainly use recorders.
In addition to the above-mentioned communication, Analyzer bus using Ethernet is also provided. About analyzer
bus, refer to GS 11B06A01-01E Ethernet Analyzer Bus System.

n SYSTEM CONFIGURATION
Field Panel/office Package software

GCMT
Communication converter
with explosion-proof
RS-422 maintenance function
(Up to 1 km) Note 1 Note 2
Note 3 PC Printer
K9404LA RS-232C
GC1000 Power-off signal
Mark II Communication converter
with explosion-proof
analyzer maintenance function
RS-422 Note 1
Note 3 DCS
K9404LD RS-232C

Power-off signal
F01.ai

Note 1: Cable recommended for RS-422 serial communication


Core: 0.75 to 1.5 mm2, twisted pair cable, 3 pairs
Cable OD: 10 to 15.9 mm
Cable length: 1 km max.
Sealing of analyzer side is required
Parts, cable and other materials are provided by customer.
Refer to TI 11B03A03-13E Process Gas Chromatograph GC1000 Mark II Installation Manual (Ethernet version) .
Note 2: Cable recommended for PC communication (wiring: Please refer to page 11.)

K9404LA side PC side


4

1 3

2 1: D-Sub 25 pins (male) 3: D-Sub 9 pins (female)


2: ISO M2.6-0.45 4: ISO #4-40 F02.ai

Note 3: The part number is changed when area classis FM-Y or CSA-Y.

Yokogawa Electric Corporation GS 11B03G03-01E


2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan Copyright Mar. 2009
Tel.: 81-422-52-5617 Fax.: 81-422-52-6792 1at Edition Jun. 01, 2009
<<Contents>> <<Index>> 
n DCS COMMUNICATION
1. Communication Specifications
Communication Standard:
RS-422 (between the GC1000 Mark II and the K9404LD communication converter)
RS-232C (between the K9404LD communication converter and the DCS)
Start-stop System: Start bit 1
Stop bit 1
Parity 1
Data Signal Rate: Select from 1200/2400/4800/9600/19200 bps
(switched from the keyboard the analyzer)
Error Detection: Parity check (even/odd/none)
Line Form: 4-wire type + 2 (power-off signal)
The power-off signal is used only between the GC1000 Mark II analyzer and the
communication converter.
Communication Data Format:
ASCII 7-bit code
(Select either non-procedure or handshake.)
Non-procedure: Output at any time when transmission data are generated
Handshake
Transmission: Full duplex
2. Operating Outline of GC1000 Mark II
(1) Status and Operation Mode
The GC1000 Mark II has the following statuses:
Process: Normal measurement, calibration and validation
Stream Sequence: Continuously measures streams in order specified in Stream Sequence
Stream (continuous): Continuously measures the specified stream
Stream (1 cycle): Measures the specified stream once
Calibration/Validation: Performs calibration or validation of the specified number
Manual: Manual operation
On/off of various valves
On/off of stream valves
On/off of several heaters
On/off of several types of detector (only On for FID and FPD)
Lab: Measurement with automatic peak detection, like a laboratory GC
The GC1000 Mark II has the following operation modes:
Stop: Mode in which the measurement stops
Run: Mode in which the measurement is running
Pause: Mode in which the measurement pauses
The figure of below shows the transitions of statuses and operation modes.

Lab Process Manual

Run command
Run mode Pause mode Run mode Run mode
Pause command
Pause
Stop Run command Run Stop Run
command command Stop command command Stop command command command

Change the status Change the status


Stop mode Stop mode Stop mode

Status and Operation Mode Transition Diagram


(2) Stream
31 streams can be used.
Specified parameters for streams
Stream valve number: 1 to 31
Method number: 1 to 4
Measurement stream: Used for measuring process samples.
Calibration stream: Comparing measured results before and after calibration, or checking whether the
GC1000 Mark II is operating properly can be performed using a standard sample.
Calibration stream: Used for measuring a standard sample at calibration.

All Rights Reserved. Copyright 2009, Yokogawa Electric Corporation GS 11B03G03-01E Jun. 01, 2009-00
<<Contents>> <<Index>> 
(3) Method 4. Output Signal
Activation timing for various valves and parameters
with regard to the temperature of the isothermal/ (1) Types of Signals
programmed-temperature ovens should be specified 1 Analysis value data
for Method. Four Methods can be used with the Analyzer number
GC1000 Mark II. Stream number
Specified parameters for Method Peak number
Analysis cycle Component concentrations (including component
Warming up time concentration error alarm)
Peak detection stop time Unit (ppm or %)
Pause time Retention time (including retention time error
DO on time/DO off time alarm)
Stream valve on time/stream valve off time Computed value (component ration computation
V1 to V8 on/off time or linear polynominal computation)
Initial control temperature 2 Error alarm
Initial temperature time Analyzer number
(only for the GC1000D/GC1000T/GC1000W) Types of alarms
The first temperature gradient/first control Data and time of occurrence
temperature/first temperature time 3 Other information
The second temperature gradient/second control Calibration factor (including sensitivity error
temperature/second temperature time alarm)
The third temperature gradient/third control (2) Signal Lavel (for RS-232C)
temperature /third temperature time Output Voltage: +9 V 3 V at ON
(only when EPC is installed) -9 V 3 V at OFF
Initial control pressure Output impedance: 300 (for RS-232C)
Initial pressure time (3) Specifying Output Signal Transmission
First pressure gradient/first control pressure/first Analytical data and alarms are independently
control time specified with analyzer keys.
Second pressure gradient/second control
(4) Priority of Signal Transmission
pressure/second control time
Each signal is transmitted simultaneously upon
Third pressure gradient/third control pressure/
occurrence.
third control time
A error alarm occurring during an analytical data
3. Input Signal transmission of analytical data and calibration
factor are transmitted after the transmission is
(1) Types of Signals completed (for exclusive handshake).
1 Stream change command If two or more error alarms occur simulataneously,
2 Calibration command they are transmitted in the order of detection.
3 Range change command (5) Instruction in Handshake
4 Start/Stop command Output instruction
(Note) When a stop command is received during
1 Data (component concentration, computed value,
analysis, the operation is stopped after analysis is
completed. calibration factor, error alarm) receiving request
2 Data transmission completion
(2) Signal Level (for RS-232C) 3 Prompt of input command
Receiving threshold value (Note) This is made at the time of receiving data
+3 V and above at ON (component concentration, and computed
+0 V or less at OFF value).
Input impedance 3 to 7 k 4 Answerback (command execution)
Maximum permissible input voltage 30 V 5 Answerback (command not executed)
(3) Priority of Signal Transmission Input instruction
In the order of occurrence 1 Data transmission request
2 Data retransmission request
(Note) For signal types and communication control
procedure, see the next section.

All Rights Reserved. Copyright 2009, Yokogawa Electric Corporation GS 11B03G03-01E Jun. 01, 2009-00
<<Contents>> <<Index>> 
5. Data Format for DCS Communication
(1) GC6 Type Output Data Format (fixed to 45 characters)
1 Analysis value data (component concentration)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45

P P M A : C H L Dummy C L
D S T A : R T
, , % , , A : C L L , space R F

a b c d e f g CR: Carriage return


LF : Line feed
a: Stream number (01 to 24) [Stream number (01 to 31)] b: Component number (01 to 24) [Peak number (1 to 99)]
c: Component concentration (without zero suppression) d: Unit
e: Concentration error alarm (Upper limit = A:CHL, Lower limit = A:CLL) f: Retention time (seconds) g: Retention time error alarm
Note: The error alarm is output only if an error occurs. (Dummy space if no error occurs.) GC1000 data in bracket [ ].
F0504.ai

2 Analysis value data (component ratio operation, linear polynominal operation)


1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45

P P M C L
D S Dummy space R F
, , %

a b c d

a: Stream number (01 to 24) [Stream number (01 to 31)] b: Component number (01 to 24) [Peak number (1 to 99)]
c: Computed value (without zero suppression) d: Unit
Note: GC1000 data in bracket [ ].
F0505.ai

3 Calibration factor
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45

C C Dummy space C L
, , , C O E R F

a b c d

a: Standard sample (STD1=1, STD2=2) [Stream number (1 to 9)] b: Component number (01 to 24) [Peak number (1 to 99)]
c: Calibration factor (without zero suppression) d: Calibration factor error (only if an error occurs)
Note: GC1000 data in bracket [ ].
F0506.ai

4 Error alarm
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45

A S / : Dummy space C L
, , R F

a b c d e

a: Month b: Day c: Hour d: Minute e: Error alarm types MEM: Memory pattern check error EXT1: External contact input 1
WDT: Watchdog timer EXT2: External contact input 2
8 items in total
AD1: Detector 1 calibration error
AD2: Detector 2 calibration error EXT8: External contact input 8
TMPH: Isothermal oven temperature error FLM1: FID1 extinguished
CAR1: Pressure switch 1 OFF FLM2: FID2 extinguished
CAR2: Pressure switch 2 OFF RPT: Calibration repeatability error
CAL: Calibration out of range
Note: For 3 characters, the 18th character is a dummy space. POF: Power off
NSD: Communication error (GC1000 GCCU)
F0507.ai

All Rights Reserved. Copyright 2009, Yokogawa Electric Corporation GS 11B03G03-01E Jun. 01, 2009-00
<<Contents>> <<Index>> 
(2) GC6 Type Input Data Format <Retention Time Data Format>
(a) Stream change command (d) Start/stop command XXXX.X
1 2 3 4 5 6 Start command <Calibration Factor Data Format>
1 2 3 4
X.XXX
S C C L
R F
B E C L
R F
Output Signal Transmission Timing of Non-procedure
a
a: Stream number
Stop command Send Data (SD)
(b) Calibration command 1 2 3 4 Analysis data
+9V
1 2 3 4 5
F I C L
R F
C A C L
R F Analysis completion point
Note: GC1000 data in bracket [ ]. (Transmission
a Calibration factor
+9V completion)
a: Standard sample (STD1=1, STD2=2) [Stream number]

(c) Range change command


During calibration
1 2 3 4 5 6 7 8 9 10 11 12

R A C L Error alarm
R F +9V

a b c
When an alarm occurs (after transmission
completion if the component concentration
a: Stream number value or calibration factor is being transmitted)
b: Component number [Peak number]
c: Component list number [Range number] Request to Send (RS)
F0508.ai

Terminal Equipment Ready (ER)


<Data Format for Component Concentration>
The component concentration takes the following (Always ON)
+9V
format by designating measurement ranges in the
GC1000 peak setting table: Processor power ON
Measurement range Output data format
0.001 to 9.999 X.XXX F0509.EPS

10.000 to 99.999 XX.XX


100.000 to 999.999 XXX.X
1,000.000 and above XXXXX
I/O Instruction Format and Communication Control Procedure for Handshake
I/O instruction format Communication control procedure
E: Data receiving Normal communication
request
# E C L
R F Supervisory computer A A A
E Data 1 Data n F
F: Data transmission Output upon analysis completion or occurrence of an error alarm
completion (after transmission completion if the component concentration value or
# F C L
R F calibration factor is being transmitted)
Retransmission
T: Input command Supervisory computer A R R
Output instruction

prompt
# T C L
R F
E Data 1 Data 1 Data 1

Note: If retransmission is requested, data can be transmitted any number of


times. If the total communication time exceeds one analysis cycle, the
B: Answerback data is changed into new data. (The limiter should be considered for the
(command executed)
# B C L
R F
supervisory computer side.)

Input command
W: Answerback Command Command Up to 6 commands including those
(command not Supervisory computer 1 2 not executed
executed) # W C L
5 seconds
R F T B or W B or W 5 seconds
Output after Re-prompted if there is no If there is no
If the command is completion of
in an incorrect command. If there is no command command during
data transmission upon re-prompting, it is assumed this period, it is
format, answerback # W # or every 8 to 10 that there is no command. assumed that
is in the format sec periodically there is no
shown at right.) Note: If the sequence has stopped, input commands can be command.
received continuously until the start signal is received.
A: Data transmission 3.2 seconds
request Error handling (for reference)
Input instruction

# A C L
R F The same
command is
Supervisory computer (No response from the computer) A A repeated.

R: Data E 3 E 3 E E (No response from GC1000)


retransmission seconds seconds
request # R C L
R F

F0510.EPS

All Rights Reserved. Copyright 2009, Yokogawa Electric Corporation GS 11B03G03-01E Jun. 01, 2009-00
<<Contents>> <<Index>> 
(3) GC8/GC1000 Type Output Data Format (fixed to 45 characters)
1 Analysis value data (concentration)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45
S P P M A : C H L
D 1 T A : R T C L
% R F
2 , , , , A : C L L ,

a b c d e f g h
CR: Carriage return
LF : Line feed
a: Stream number (01 to 31) b: Peak number (1 to 99) when the head is DS c: Component concentration (without zero suppression)
Peak number 0 to 99 (100 to 199) when the head is D1 d: Unit
Peak number 0 to 55 (200 to 255) when the head is D2
e: Concentration error alarm (Upper limit = A:CHL, Lower limit = A:CLL) f: Retention time (seconds) g: Retention time error alarm
h: Analyzer number 1 to 250 (zero suppression provided)
Note: The error alarm is output only if an error occurs. (Dummy space if no error occurs) F0511.ai

2 Analysis value data (component ratio calculation and linear polynominal calculation)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45
S P P M
D 1 Dummy space C L
% R F
2 , ,

a b c d e

a: Stream number (01 to 31) b: Peak number (1 to 99) when the head is DS c: Computed value (without zero suppression)
Peak number 0 to 99 (100 to 199) when the head is D1 d: Unit
Peak number 0 to 55 (200 to 255) when the head is D2
e: Concentration error alarm (Upper limit = A:CHL, Lower limit = A:CLL)
F0512.ai

3 Calibration factor
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45

C C Dummy space C L
, , , C O E R F

a b c d e

a: Stream number (1 to 9) b: Peak number (1 to 99) c: Calibration factor (without zero suppression)
d: Sensitivity error (if an error occurs) e: Analyzer number 1 to 250 (zero suppression provided) F0513.ai

4 Error alarm
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45

A S / : Dummy space C L
, , R F

a b c d e f

a: Month b: Day c: Hour d: Minute e: Error alarm types MEM: Memory pattern check error EXT1: External contact input 1
f: Analyzer No.1 to 250 WDT: Watchdog timer EXT2: External contact input 2
8 items in total
AD1: Detector 1 calibration error
AD2: Detector 2 calibration error EXT8: External contact input 8
TMPH: Isothermal oven temperature error FLM1: FID1 extinguished
CAR1: Pressure switch 1 OFF FLM2: FID2 extinguished
CAR2: Pressure switch 2 OFF RPT: Calibration repeatability error
CAL: Calibration out of range
Note: For 3 characters, the 18th character is a dummy space. POF: Power off
NSD: Communication error (GC1000 GCCU)
F0514.ai

All Rights Reserved. Copyright 2009, Yokogawa Electric Corporation GS 11B03G03-01E Jun. 01, 2009-00
<<Contents>> <<Index>> 
(4) GC8/GC1000 Type Input Data Format
(a) Stream change command (b) Calibration command
1 2 3 4 5 6 7 8 1 2 3 4 5 6 7

S M C L C M C L
R F R F

a b a b
1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9

S E C L C E C L
R F R F

a b a b
a: Stream number a: Stream number
b: Analyzer number (1 to 6) or (001 to 255) b: Analyzer number (1 to 6) or (001 to 255)

(c) Range change command


1 2 3 4 5 6 7 8 9 10 11 12 13 14

R M C L
R F

a b c d
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

R E C L
R F

a b c d
a: Stream number
b: Peak number
c: Range number
d: Analyzer number (1 to 6) or (001 to 255)

(d) Start/stop command


Start command Stop command
1 2 3 4 5 1 2 3 4 5

B M C L F M C L
R F R F

a a
1 2 3 4 5 6 7 1 2 3 4 5 6 7

B E C L F E C L
R F R F

a a

a: Analyzer number a: Analyzer number


(1 to 6) or (001 to 255) (1 to 6) or (001 to 255) F0515.ai

<Data Format for Component Concentration>


The component concentration takes the following format by designating measurement ranges in the GC1000 peak
setting table:
Measurement range Output data format
0.001 to 9.999 X.XXX
10.000 to 99.999 XX.XX
100.000 to 999.999 XXX.X
1,000.000 and above XXXXX
<Retention Time Data Format>
XXXX.X
<Calibration Factor Data Format>
X.XXX

All Rights Reserved. Copyright 2009, Yokogawa Electric Corporation GS 11B03G03-01E Jun. 01, 2009-00
<<Contents>> <<Index>> 
Output Transmission Timing in the Case of NO Control

Send Data (SD)


Analysis data
+9V (RS-232C)
(Transmission completion)
AUTO mode analysis
completion point

Calibration factor
+9V (RS-232C)
(Transmission completion)

During AUTO mode calibration

Error alarm
+9V (RS-232C)

When an alarm occurs (after transmission


completion if the component concentration
value or calibration factor is being transmitted)

Request to Send (RS)


Terminal Equipment Ready (ER)
(Always ON)
+9V (RS-232C)

Processor power ON

F0516.EPS

I/O Instruction Format and Communication Control Procedure for Handshake


I/O instruction format Communication control procedure
E: Data receiving request Normal communication t (0.4 to 0.5 second) t
# E C L
R F Supervisory computer A A A
E Data 1 Data n F
F: Data transmission Output upon analysis completion or occurrence of an error alarm
completion (after transmission completion if the component concentration value or
# F C L
R F calibration factor is being transmitted)
Retransmission t t t

T: Input command prompt Supervisory computer A R R


Output instruction

# T C L
R F
E Data 1 Data 1 Data 1

Note: If retransmission is requested, data can be transmitted any number of


B: Answerback times. If the total communication time exceeds one analysis cycle, the
(command executed) data is changed into new data. (The limiter should be considered for the
# B C L
R F supervisory computer side.)

Input command
t t
W: Answerback
(command not executed) Command Command
# W C L
R F
Supervisory computer 1 2 Up to 6 commands including those not executed
T 5 seconds B or W B or W 5 seconds
If the command is in Output after Re-prompted if there is no command. If there is no command
an incorrect format, completion of If there is no command upon
# W # during this period, it is
answerback is in the data transmission re-prompting, it is assumed that there assumed that there is no
format shown at right.) or every 8 to 10 is no command. command.
sec periodically
Note: If the sequence has stopped, input commands can be received
A: Data transmission
continuously until the start signal is received.
request
Input instruction

# A C L
R F Error handling 3.2 seconds (for reference)
The same
command is
Supervisory computer (No response from the computer) A A repeated.
R: Data retransmission
request E 3 E 3 E E (No response from GC1000)
# R C L
R F seconds seconds

F0517.EPS

(Note) If #ACRLF or #RCRLF is not received in 20 seconds or more after the data are transmitted, the transmission of analysis
data of that time is aborted.

All Rights Reserved. Copyright 2009, Yokogawa Electric Corporation GS 11B03G03-01E Jun. 01, 2009-00
<<Contents>> <<Index>> 
Reference: An Examples of Error Control by Distributed Control System (DCS)
1. Parity Check
Even parity is used for GC1000 Mark II communication outputs.
When an odd parity is received, the DCS regards the data as invalid.
2. Check of Header
Each communication packet begins with two alphabet characters.
The DCS regards any packet beginning with unidentified characters as invalid data.
3. Check of Value
Analysis results and other data in a communication packet are numeric values.
If data contain a non-numeric character where a number is supposed to be set, the DCS regards the corresponding
data as invalid.
The decimal point position can also be checked.
4. Check of Number and Order of Components
The number and order of the analysis result data are fixed.
If the order of received result data is wrong or too many data are received, the DCS regards them as invalid data.
5. Check of Character Length
The length of a communication output is fixed to 45 characters.
The DCS regards any communication output having a length other than 45 characters as invalid data.

n GCMT GAS CHROMATOGRAPH MAINTENANCE TERMINAL


The GCMT gas chromatograph maintenance terminal is package software for the PC to maintain the GC1000 Mark
II.

1. Operating Conditions
Model: IBM PC/AT compatible machine that can run Microsoft Windows
OS Microsoft Windows Vista Microsoft Windows 2000 SP4,
Bussiness Edition Microsoft Windows XP SP2
CPU 32 bit - 1 GHz or higher Pentium II 350 MHz or higher
Memory 1 GB or more 256 MB or more
Hard disk 80 MB or more (for program and data)
VDT resolution VGA (640 x 480) or higher
Others Printer
CD-ROM drive

2. Communication Specifications
Communication Standard: RS-232C (between PC and communication converter K9404LA)
Communication Rate: 19200/38400 bps
Parameter: 8 data + 1 stop bit, without parity, XON / XOFF control

3. Functional Configuration
The GC maintenance terminal basically has the next two functions.
1 The number of analyzers that can receive service simultaneously is one.
2 Only the data that the analyzer currently has are handled.
(1) Connection Link to Analyzer
The database is always updated. The operating commands from PC are transmitted to the analyzer on a real-time
basis.
(2) Operation on the Display
The data in the database is indicated on the display and the display is updated with the messages for updating.
The operation mode can be changed and valves can be opened / closed on the display.
(3) Data Storage and Display
The database and chromatograms can be stored in files and redisplayed using the file display function.
(4) Data Computation Processing
The PC maintenance terminal computes and processes the analytical data and chromatogram data stored in the
files. Also, sophisticated computations and processing data for the analyzer can be easily carried out using general-
purpose software.
(5) Printing
The display of the PC maintenance terminal can be printed out (hard copy) with the printer which is provided with a
Windows printer driver.

All Rights Reserved. Copyright 2009, Yokogawa Electric Corporation GS 11B03G03-01E Jun. 01, 2009-00
<<Contents>> <<Index>> 10
4. Display Panel and Operation
The display panel which the system indicates are roughly composed of the following five types: These can be
simultaneously displayed except for the LCD emulator window.
(1) Analyzer operation window
(2) Analysis results window
(3) Chromatogram window
(4) Alarm window
(5) LCD emulator window

5. GC Maintenance Terminal Model Codes

Model Basic Code Optional Code Description


GCMT Software of GC Maintenance Terminal
Function -A01 Standard
Language E English
J Japanese

n COMMUNICATION CONVERTER K9404LA, K9404LD, RS-232C


COMMUNICATION CABLE, FIGURE, WIRING
This communication converter converts communications from RS-422 to RS-232C for PC or DCS communication
and shuts off the communication signals when a power-off signal is received from the analyzer. This maintains
explosion-protection on the analyzer side.

1. Signal Terminals on the GC1000 Mark II


M4 terminals
1 RS-422 SDA, SDB, RDA, RDB
2 D/I 12 V +, -

2. Signal Terminals on the Communication Equipment


M4 terminals
RS-232C D-Sub25 female terminals

3. Power Supply
100 to 240 V AC 10 %, 50/60 Hz 5 % , 15W Max., M4 terminals

4. Terminal Arrangement (common for K9404LA, K9404LD)


D-SUB female 25 pins Terminal: M4
CN1
CN3 CN2

SDA SDB RDA RDB + N L


RS-232C
RS-422 D/I 12V AC power supply

from the GC1000 analyzer F10.EPS

All Rights Reserved. Copyright 2009, Yokogawa Electric Corporation GS 11B03G03-01E Jun. 01, 2009-00
<<Contents>> <<Index>> 11
5. PC Communication Cable (Straight type)
Internal wiring
GND GND
FG 1
CD 8 1 CD
RD 3 2 RD
SD 2 3 SD
ER 20 4 ER
SG 7 5 SG
DR 6 6 DR
RS 4 7 RS
CS 5 8 CS
CI 22 9 CI
K9404LA Personal computer
F12.ai

6. External Dimension (K9404LA)


Unit: mm
203
126
35

Weight: approximately 2 kg

RS-232C
RS-422, D/I 12 V
D-SUB 25P Socket Power supply
F11.ai

All Rights Reserved. Copyright 2009, Yokogawa Electric Corporation GS 11B03G03-01E Jun. 01, 2009-00
<<Contents>> <<Index>> 12
7. External Dimension (K9404LD)
Unit: mm
233

218

203
126

110
96
7

5
4-4.5
Mounting holes
40

1.6
RS-232C RS-422, D/I 12 V
Power supply Weight: approximately 2 kg
D-SUB 25P Socket
K9404LD.ai

8. Terminal RS-232C: Wiring in the inside of converter (K9404LA, K9404LD)

RS-232C Wiring in the inside of converter

FG
1
SD
2
RD
3
RS
4
CS Outside
5
DR
6
SG
7
CD
8
ER
20

Other pins are open.


RS-232C.ai

All Rights Reserved. Copyright 2009, Yokogawa Electric Corporation GS 11B03G03-01E Jun. 01, 2009-00