Anda di halaman 1dari 16

Simultaneous Quantitative and

Qualitative Analysis of Clozapine


and its Metabolites in Rat Plasma
Using the Agilent 6540 Q-TOF LC/MS
System
Application Note

Authors Abstract
Na Pi Parra and Shane E. Tichy The employment of a high resolution accurate mass (HRAM) LC/MS approach
Agilent Technologies, Inc.
Santa Clara, CA, USA for simultaneous quantitative and qualitative analysis was demonstrated for the
bioanalysis and metabolite identication of clozapine in rat plasma. Excellent
assay performance was achieved for the quantitation of clozapine and metabolites
based on the ultrahigh resolving power and mass accuracy of the Agilent6540
Quadrupole Time-of-Flight (Q-TOF) LC/MS system. Two clozapine phase I
metabolites were successfully identied in rat plasma. The HRAM MS and
MS/MS data acquired in this study can be retrospectively analyzed to search
for potential metabolites, biomarkers, and clinically interesting endogenous
components without sample re-injection.
Introduction and its metabolites, norclozapine Sample preparation
and clozapine-N-oxide (Figure 1),
Multiple reaction monitoring (MRM) by Calibration standards (110,000 ng/mL)
in rat plasma. Excellent sensitivity,
triple quadrupole mass spectrometry and quality controls (QCs) (50 ng/mL)
linearity, dynamic range, accuracy,
has been the most commonly used were prepared by spiking clozapine,
reproducibility, and precision were
method for the quantitation of drugs norclozapine, and clozapine-N-oxide
demonstrated in the quantitative
and their metabolites in complex at varied concentrations into rat
measurements. In addition, clozapine
biological matrices.1,2 With recent plasma. In the clozapine PK study,
metabolites were identied using
advances in Q-TOF HRAM mass rats were dosed at 1 mg/kg through
accurate mass MS and MS/MS data.
spectrometry, there is a growing an intravenous (IV) route, and blood
A combined targeted and untargeted
interest in the pharmaceutical samples were taken over a time course
workow is described for clozapine
industry and clinical laboratories of 5 minutes to 7 hours. PK samples,
metabolite ID using Agilent
using HRAM LC/MS and MS/MS calibration standards, and QCs were
MassHunter Qualitative analysis and
methods for quantitative bioanalysis.3-5 spiked with glyburide (internal standard)
Metabolite ID software tools.
Q-TOF HRAM LC/MS and MS/MS at 50 ng/mL and extracted with ice-cold
methods offer advantages over triple acetonitrile before LC/MS analysis.
quadrupole MRM methods by Blank plasma was used as a double
allowing rapid method development, blank and blank plasma with 50 ng/mL
and providing accurate mass and glyburide was used as a blank.
MS/MS fragmentation information
for further metabolite identication
(ID) and structural characterization.6
O
This application note presents HRAM N NH N N N N
LC/MS and MS/MS methods with Cl
N
Cl
N
Cl
N
Demethylation N-Oxidation
great selectivity and mass accuracy
N N
for the simultaneous quantitative H H
N
H
and qualitative analysis of clozapine Norclozapine Clozapine Clozapine-N-oxide
C17H17ClN4, 312.1142 C18H19ClN4, 326.8233 C18H19ClN4O, 342.1247

Figure 1. Clozapine and its metabolites, norclozapine and clozapine-N-oxide.

2
Instrumentation Table 1. Liquid chromatography, Agilent Q-TOF MS source conditions, and acquisition method parameters.

Liquid chromatography (LC) was LC conditions


performed on the Agilent 1290 Innity Column Agilent Zorbax Eclipse Plus Rapid Resolution HT column,
LC System, which consisted of a 2.1 50 mm, 1.8 m
(p/n: 959757-902)
binary pump, a vacuum degasser,
a high performance thermostatted Column temperature 40 C
autosampler, and a thermostatted Injection volume 1 L in MS mode and 2 L in MS/MS modes
column compartment. Full acquisition Autosampler temperature 6 C
MS, auto MS/MS, and targeted Needle wash 10 seconds in wash port
MS/MS were performed on the 6540 Mobile phase A = 0.1 % formic acid in water
Q-TOF Mass Spectrometer equipped B = 0.1 % formic acid in acetonitrile
with an Agilent Jet Stream source in Flow rate 0.75 mL/min
positive ionization mode, and using the Gradient program Time (min) A (%) B (%)
mass resolving power (MRP) of20K.
Initial 90 10
LC method, ion source conditions,
0.70 90 10
and acquisition method parameters
were optimized for clozapine and its 4.00 10 90
metabolites (Table 1). 4.50 90 10
5.00 90 10
Post time 1 minute
Agilent QTOF MS source conditions
Drying gas temperature 300 C
Drying gas ow 7 L/min
Sheath gas temperature 400 C
Sheath gas ow 11 L/min
Nebulizer pressure 35 psi
Capillary voltage 3,750 V
Nozzle voltage 0V
Fragmentor voltage 200 V
Reference delivery Agilent 1200 Isocratic pump with 100:1 splitter (p/n: G1607-60000)
Reference pump ow 0.5 mL/min for 5 L/min to nebulizer
Reference ions 121.050873 and 922.009798
Instrument mass range 1,700 Da
Instrument mode Extended dynamic range
Data storage Centroid and prole

3
The preferred precursor ions in auto Table 1. Liquid chromatography, Agilent Q-TOF MS source conditions and acquisition method parameters.
MS/MS mode and the targeted (continued)
precursor ions in targeted MS/MS
Agilent Q-TOF MS acquisition method parameters
mode are listed in Table 2.
Mass range 100-1,000 m/z
Acquisition rate 2.5 Hz, 400 ms/scan
Agilent Q-TOF Auto MS/MS acquisition method parameters
Mass range (MS) 100-1,000 m/z
Acquisition rate (MS) 5 Hz, 200 ms/scan
Mass range (MS/MS) 50-1,000 m/z
Acquisition rate (MS/MS) 3 Hz, 333.3 ms/scan
Quadrupole isolation Width Medium
Collision energy 20 V
Maximum precursor ions/cycle 4
Precursor ion static exclusion 100-200 m/z and 500-1,000 m/z
Precursor ion active charge state 1 and Unknown
Agilent Q-TOF Targeted MS/MS acquisition method parameters
Mass range (MS) 50-1,000 m/z
Acquisition rate (MS) 5 Hz, 200 ms/scan
Mass range (MS/MS) 25-1,000 m/z
Acquisition rate (MS/MS) 2.5 Hz, 400 ms/scan
Maximum time between MS1 spectra 3 seconds

Table 2. Auto MS/MS preferred precursor ion list and targeted MS/MS targeted precursor ion list.

Precursor Dm/z Charge RT DRT Isolation CE Product


Compound name ion (ppm) state (min) (min) width (V) ion
Norclozapine 313.1215 20 1 1.64 0.4 Medium 30 270.0793
Clozapine 327.1371 20 1 1.75 0.4 Medium 20 270.0793
Clozapine-N-oxide 343.1320 20 1 1.88 0.4 Medium 15 256.0633
Glyburide (IS) 494.1511 20 1 3.05 0.4 Medium 20 369.0663

4
Data acquisition and analysis QC at 50 ng/mL

A MassHunter Workstation (version 104


B.03.01) was used for data acquisition. 4.5

MassHunter Quantitative (Quan) Norclozapine


4 ECI 327.1371
Analysis Software (version B.04.00)
EIC 313.1215
was used for quantitation. Extracted ion 3.5 Clozapine
EIC 343.1320
chromatograms (EICs) of m/z 327.1371,
313.1215, and 343.1320 in MS mode, 3 EIC 494.1511
and product ion EICs of m/z 327.1371 2.5
> 270.0793, 313.1215 > 270.0793, Counts
and 343.1320 > 256.0633 in targeted 2
MS/MS mode, were employed for
1.5
quantitation of clozapine, norclozapine, Clozapine-N-oxide
and clozapine-N-oxide, respectively. EIC 1
of m/z 494.1511 and product ion EIC of
Glyburide (IS)
m/z 494.1511 > 369.0663 were used for 0.5
glyburide. The mass extraction window
0
(MEW) was 10 ppm. MassHunter
Qualitative (Qual) Analysis Software 1.2 1.4 1.6 1.8 2 2.2 2.4 2.6 2.8 3 3.2 3.4
(version B.03.01) was used to nd and Acquisition time (min)
conrm clozapine metabolites.

104 Blank
Results and Discussion 4.5

Quantitative analysis 4
Norclozapine, clozapine, clozapine-N-
3.5
oxide, and glyburide were separated ECI 327.1371
using UHPLC at retention times (RT) 3 EIC 313.1215
of 1.64, 1.75, and 1.88, 3.05 minutes, EIC 343.1320
respectively (Figure 2 and Figure 3). 2.5
Counts

EIC 494.1511
The high MRP and narrow MEW
2
employed in the HRAM LC/MS and
MS/MS methods greatly decreased 1.5
the endogenous interference from rat
plasma, thus signicantly improving the 1

quantitation performance. 0.5


Glyburide (IS)

1.2 1.4 1.6 1.8 2 2.2 2.4 2.6 2.8 3 3.2 3.4
Acquisition time (min)

Figure 2. EICs of clozapine, norclozapine, clozapine-N-oxide, and glyburide in MS mode.

5
Sensitivity QC at 50 ng/mL

The limit of quantitation (LOQ) is 104

1 ng/mL or 1 pg on-column for 4


clozapine and its metabolites in rat 3.5 Clozapine
PI 327.1371 & 270.0793
plasma using the MS method. Using 3 PI 313.1215 & 270.0793
the targeted MS/MS method, the
2.5 PI 343.1320 & 256.0633
LOQ is 1, 1, 5 ng/mL or 2, 2, and 10 pg

Counts
Norclozapine PI 494.151 & 369.0663
2
on-column for clozapine, norclozapine,
and clozapine-N-oxide, respectively. 1.5
The MS method is slightly more 1 Glyburide (IS)
sensitive than the targeted MS/MS 0.5 Clozapine-N-oxide
method. Based on our observations, the
0
relative sensitivity of MS method versus
targeted MS/MS method depends on 1.2 1.4 1.6 1.8 2 2.2 2.4 2.6 2.8 3 3.2 3.4
Acquisition time (min)
the nature of the interference from
the complex biological matrix. The Blank
104
MS method is highly selective based
4
on the ultra-high MRP of Q-TOF.
Alternatively, the targeted MS/MS 3.5
PI 327.1371 & 270.0793
method gains further selectivity based 3
on fragmentation in the collision cell PI 313.1215 & 270.0793
2.5
and selection of a specic product ion PI 343.1320 & 256.0633
Counts

2
for quantitation; however, the ultimate PI 494.151 & 369.0663
1.5
signal intensity is sacriced. Glyburide (IS)
1

0.5

1.2 1.4 1.6 1.8 2 2.2 2.4 2.6 2.8 3 3.2 3.4
Acquisition time (min)

Figure 3. Product ion EICs of clozapine, norclozapine, clozapine-N-oxide, and glyburide in targeted
MS/MS mode.

6
Calibration curve linearity and
range
The calibration curves for clozapine,
norclozapine, and clozapine-N-oxide in
the MS and targeted MS/MS methods
(Figure 4 and Figure 5) showed
excellent linearity (R2 > 0.998) and a
wide dynamic range ( 3 orders).

Clozapine, 1-10,000 ng/mL


A
103
101
2 1
Relative responses

0.8
1.8
0.6
1.6 0.4
0.2
1.4
0
1.2
Relative responses

0 5 10 15 20 25 30 35 40 45 50
Concentration (ng/mL)
1
0.8
0.6
0.4
0.2
0 R 2 = 0.9984
-0.2
0 1000 2000 3000 4000 5000 6000 7000 8000 9000 10000
Concentration (ng/mL)

B Norclozapine, -5,000 ng/mL C Clozapine-N-oxide, 1-5,000 ng/mL

103
102
0.9 3.5
Relative responses

Relative responses

3
0.7
2.5
0.5 2.0
1.5
0.3
1
0.1 0.5
R 2 = 0.9996 R 2 = 0.9982
-0.1
0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000 0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000
Concentration (ng/mL) Concentration (ng/mL)

Figure 4. Calibration curves of clozapine (A), norclozapine (B), and clozapine-N-oxide (C) in rat plasma using the MS method. Inset graph (A) demonstrates the
low concentration range for clozapine.

7
Clozapine, 1-5,000 ng/mL
A
102
7
6.5
Relative responses 6
6 5
5.5 4
5 3
2
4.5
Relative responses

1
4 0
3.5 0 5 10 15 20 25 30 35 40 45 50
Concentration (ng/mL)
3
2.5
2
1.5
1
0.5
0 R 2 = 0.9982
-0.5
250 500 1000 1500 2000 5000 3000 3500 4000 4500 5000
Concentration (ng/mL)

B Norclozapine, 1-1,000 ng/mL C Clozapine-N-oxide, 5-500 ng/mL

101 101
4
2.5
Relative responses
Relative responses

3.5
3 2
2.5
1.5
2
1.5 1
1
0.5
0.5
0 R 2 = 0.9988 0 R 2 = 0.9985

0 100 200 300 400 500 600 700 800 900 1000 0 500 1000 1500 2000 2500 3000 3500 4000 4500 5000
Concentration (ng/mL) Concentration (ng/mL)

Figure 5. Calibration curves of clozapine (A), norclozapine (B), and clozapine-N-oxide (C) in rat plasma using the targeted MS/MS method. Inset graph (A)
demonstrates the low concentration range for clozapine.

8
Accuracy, reproducibility, and PK analysis and metabolite MS/MS methods. Figure 6 illustrates
precision proles of clozapine the concentration-time proles
of clozapine and its metabolites,
The accuracy, reproducibility, and The concentrations of clozapine, norclozapine and clozapine-N-oxide,
precision were evaluated at up to norclozapine, and clozapine-N-oxide in from which the clearance, area-under-
nine standard concentrations, and rat plasma samples were successfully curve (AUC), and half-life of clozapine
the QC level for clozapine and its measured with excellent reproducibility in the rat PK study were determined to
metabolites. The accuracy, precision and (% RSD < 5 in triplicate) (Table 4). be 55 mL/min/kg, 300 ng/mL*h and
reproducibility all met the bioanalytical Consistent quantitation results were 1 hour, respectively.
accepted criteria (Table 3). observed using the MS and targeted

Table 3. Accuracy, reproducibility , and precision results using the MS and targeted MS/MS methods.

MS method Targeted MS/MS method


Accuracy Reproducibility Precision Accuracy Reproducibility Precision
Compound name (%) (% RSD, n = 2) (% RSD, n = 9) (%) (% RSD, n = 2) (% RSD, n =8)
Clozapine 89.6-109.5 0.37-9.03 7.24 87.2-106.1 0.31-6.05 6.51
Norclozapine 91.4-107.5 N/A 6.47 88.1-109.9 N/A 7.01
Clozapine-N-oxide 86.9-115.9 N/A 10.13 87.1-108.4 N/A 6.49

Table 4. Measured concentrations in rat plasma PK samples using the MS and targeted MS/MS methods.

Conc. (ng/mL) Clozapine (ng/mL) Norclozapine (ng/mL) Clozapine-N-oxide (ng/mL)


Time MS MS/MS MS MS/MS MS MS/MS
5 minutes 303 304 6.10 6.35 3.11 BLOQ
15 minutes 206 195 6.51 6.39 3.33 BLOQ
30 minutes 147 137 5.94 5.76 2.84 BLOQ
1 hour 78.1 81.5 4.50 4.88 1.61 BLOQ
2 hours 38.8 35.3 2.62 2.89 1.17 BLOQ
4 hours 9.74 10.6 BLOQ* BLOQ BLOQ BLOQ
7 hours 1.81 1.51 BLOQ BLOQ BLOQ BLOQ
*BLOQ = below limit of quantitation.

Clozapine PK and Metabolite Prole Clozapine PK and Metabolite Prole

Clozapine 10 Clozapine
300 Norclozapine
Norclozapine
8
Concentration (ng/mL)

Clozapine-NO Clozapine-NO
Concentration (ng/mL)

200 6

4
100
2

0 0
0 1 2 3 4 5 6 7 8 0 1 2 3 4 5 6 7 8
Time (hour) Time (hour)

Figure 6. Pharmacokinetics time curve of clozapine and its metabolite proles in rat plasma PK samples. Plot on the right is the zoom-in of the plot on the left at
lower concentration range.

9
Qualitative analysis Metabolite ID of clozapine using through untargeted MFE algorithm
MS data in the plasma sample was further
Metabolite ID of clozapine was identied using database (DB) search
performed in the 5 minute rat plasma In FbF, the MS data were searched against the PCD of clozapine or MFG
sample using MS, auto MS/MS, and against a personal compound database to generate formulas using accurate
targeted MS/MS data. In MassHunter (PCD) for clozapine (Figure 7) to nd mass and isotope patterns. The triple
Qualitative Analysis Software, matching peaks using accurate mass criteria MFG score was based on
metabolite ID of clozapine was information. The database was created accurate mass of the monoisotopic
achieved using both targeted data by entering the formulas of known peak, isotope spacing, and isotope
mining algorithms, nd by formula (FbF) clozapine metabolites previously abundance pattern. Clozapine and the
and nd by targeted MS/MS (FbTMS2), published in literature. In MFE, multiple two metabolites were found in the
and untargeted or nave data mining related ion clusters detected from plasma sample using FbF and MFE plus
algorithms, molecular feature extraction the raw MS data are grouped into a DB search and MFG with mass errors
(MFE), and nd by auto MS/MS list of qualied molecular features (MS) < 1 ppm and match scores > 90.
(FbAMS2). FbF and MFE were used to or compounds based on ion species,
process MS data, while FbTMS2 and charge states, and dimer/trimer
FbAMS2 were used to process targeted formation. The list of compounds found
and auto MS/MS data, respectively.

Figure 7. Clozapine personal compound database (PCD).

10
Figure 8 and Figure 9 illustrate the MS
spectra, isotope patterns, and MFG
results for clozapine and norclozapine.
Notably, excellent mass accuracy with
average mass errors < 1 ppm was
observed for the isotopes (M+1, M+2,
and M+3) of clozapine and the two
metabolites, demonstrating the high
sensitivity of the 6540 Q-TOF LC/MS
System.

5-Minute Rat Plasma PK Sample

327.1369
10 5 (M+H) +
6
329.1344 922.0098
(M+H) +
5
328.1399
4 (M+H) +
Counts

330.1373
3
(M+H) + 331.1415
2 (M+H) +
327.1369
121.0509
(M+H) +
1

0
100 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 950
Mass-to-change (m/z)

Figure 8. Clozapine MFG results using MS data in Agilent MassHunter Qualitative Analysis Software. The red boxes in the gure insert represent the theoretical
isotope abundance and spacing of clozapine.

11
5-Minute Rat Plasma PK Sample
922.0098
10 4 121.0509
2
313.1211 315.1189
1.8
(M+H) + (M+H) +
1.6
1.4
312.1241
Counts

1.2 316.1199
(M+H) +
1 (M+H) +
0.8
0.6
313.1211
0.4 (M+H) +
0.2
0
100 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 950
Mass-to-change (m/z)

Figure 9. Norclozapine MFG results using MS data in Agilent MassHunter Qualitative Analysis Software. The red boxes in the gure insert represent the
theoretical isotope abundance and spacing of norclozapine.

12
Metabolite ID of clozapine using for the compounds and their fragment scores > 90. Figure 10 and Figure 11
MS/MS data ions. The MS/MS MFG scores were illustrate the auto MS/MS spectra
based on the coverage and mass errors and MFG results for clozapine and
The MS/MS data were searched using of the fragment ions. Clozapine and norclozapine. Clozapine-N-oxide was
FbAMS2 and FbTMS2 algorithms. norclozapine were found in the sample not found in rat plasma samples using
The list of compounds found was using FbTMS2 and FbAMS2 plus DB FbTMS2 and FbAMS2 with either auto
subsequently identied using DB search and MFG with MS/MS mass or targeted MS/MS data. This could
search or MFG to generate formulas errors < 2 ppm and MS/MS MFG be due to the low level (< LOQ) of
clozapine-N-oxide in the PK samples.

5-Minute Rat Plasma PK Sample


10 5
270.0791
2
C15 H13 CI N3
1.8
1.6
1.4
Counts

1.2
1
0.8 84.0807 327.1365
0.6 C5 H10 N 296.0944
0.4 192.0677 227.0366
C17 H15 CI N3
0.2 C13 H8 N2 C13 H8 CI N2
0
70 80 90 100 110 120 130 140 150 160 170 180 190 200 210 220 230 240 250 260 270 280 290 300 310 320 330
Mass-to-change (m/z)

Figure 10. Clozapine MFG results using auto MS/MS data in Agilent MassHunter Qualitative Analysis Software.

13
5 minute rat plasma PK sample
10 2
270.0786
8
C15 H13 CI N3
7
6
5 192.0682
Counts

C13 H8 N2
4
3 227.0368
C13 H8 CI N2 296.0948 313.1209
2 73.0277 253.0517 C17 H15 CI N3
1 C15 H10 CI N2
0
70 80 90 100 110 120 130 140 150 160 170 180 190 200 210 220 230 240 250 260 270 280 290 300 310
Mass-to-change (m/z)

Figure 11. Norclozapine MFG results using auto MS/MS data in Agilent MassHunter Qualitative Analysis Software.

14
Conclusions References
This application note describes high 1. Pharmaceutical LC/MS Solutions
resolution accurate mass LC/MS from Agilent Technologies. Agilent
and MS/MS methods with excellent publication 5990-5854EN.
sensitivity and mass accuracy for
the simultaneous quantitative and 2. Metabolic Stability Study using
qualitative analysis of clozapine and its Cassette Analysis and Polarity
metabolites in rat plasma samples. Switching in an Ultra High Performance
Liquid Chromatography (UHPLC)-Triple
Excellent sensitivity with LOQ of Quadrupole LC/MS System. Agilent
1 ng/mL or 1 pg on-column in rat publication 5990-4469EN.
plasma.
3. van der Heeft, E. et al. Full-scan
Plasma calibration curves show accurate mass selectivity of ultra-
the excellent linearity (> 0.995) performance liquid chromatography
with > 3 orders of dynamic range. combined with time-of-ight and
orbitrap mass spectrometry in hormone
Accuracy (87-116 %), precision and veterinary drug residue analysis.
(%RSD < 11 %), and reproducibility J. Am. Soc. Mass Spectrom., 2009,
(% RSD < 10 %) of the assay were 20:451-463.
well within accepted bioanalytical
criteria. 4. Zhang, N.R. et al. Quantitation of
small molecules using high-resolution
Clozapine and its metabolites were accurate mass spectrometers a
identied in rat plasma samples different approach for analysis of
with high scores (> 90) and biological samples. Rapid Commun.
average mass errors of < 1 ppm Mass Spectrom., 2009, 23:1085-1094.
(MS) and < 2 ppm (MS/MS).
5. Rousu, T., Herttuainen, J., and
Powerful software processing tools Tolonen, A. Comparison of triple
with sophisticated data mining and quadrupole, hybrid linear ion trap triple
feature identication algorithms quadrupole, time-of-ight and LTQ-
(FbF, MFE, FbTMS2, FbAMS2, Orbitrap mass spectrometers in drug
DB search, and MFG) greatly discovery phase metabolite screening
facilitated metabolite identication. and identication in vitro-amitriptyline
and verapamil as model compounds.
Acknowledgements Rapid Commun. Mass Spectrom., 2010,
24:939-957.
We wish to acknowledge Dr. Yuqin Dai,
Agilent Technologies, for review and 6. An LC/MS Metabolomics Discovery
comments. Workow for Malaria-Infected Red
Blood Cells using Mass Proler
Professional Software and LC-Triple
Quadrupole MRM Conrmation. Agilent
publication 5990-6790EN.

15
www.agilent.com/chem/QTOF
This information is subject to change without notice.

Agilent Technologies, Inc., 2012


Published in the USA, November 28, 2012
5991-1573EN

Anda mungkin juga menyukai