sterilization equipments
For health Be credible
Working medium: mixed gas of 30% ethylene oxide gas and 70%
units and bank, archeology, archive and carbon dioxide gas or mixed gas of 20% ethylene
culture relic departments, and appears oxide gas and 80% carbon dioxide gas
its special advantages in sterilization for Water source pressure: 0.15-0.3Mpa
those without resistance to heat, humidity Steam source pressure: 0.3-0.5Mpa
240L360L
and erosion. Compressed air pressure: 0.5-0.7Mpa
Post-teat range: 0 ~ 9 times
Sterilization time range: 0 ~ 99 hours
Relative humidity: 45 ~ 75%
Gas concentration: 450-1200mg/L
600L1500L
Safety features
The process can not be started when the door is unlocked. Chamber Loading trolley Transfer trolley
Model dimensions volume dimensions Overall dimensions(LWHmm) Weight(Kg) dimensions
Manual door can keep the safety status even under the condition of (LWHmm) (m3) (LWH)mm (LWH)mm
Leakage test is requisite before sterilization step. Compress air Water Power(Kw) Cycle time(h)
Model consumption consumption Control power Driving power
The door can not be opened if there is pressure in chamber for (m3/min) Kg/cycle 220V/50Hz 380V/50Hz
54 37 Recommend ventilation
ensuring the safety of both operators and machine. XG2.SHS-0.24B 0.036 250 05 15*+1.45# 6 10 10
The filtering precision of air filter can reach 0.2 micron, to ensure no
second pollution after sterilization.
Password is needed for parameters modification.
Emergency switch can stop any abnormal operation occurs.
Safety valves will release pressure of chamber and jacket if pressure
is out of range.
Various alarms can stop operation and indicate the trouble spot.