Anda di halaman 1dari 12

diograph.

The electrocardiograms (ECGs or EKG's)


it will produce can be analyzed by yourself or your
doctor. We are not suggesting that you practice medi-
cine using this device, but you should find it interest-
ing and educational in monitoring your health. YOLI f

will see some of the unlque techniques used in


medical electronics and you maybe surprised to see =,
how similar medical electronic equipment is to
most other types of electronic equipment.
The eiectr&rdiograph that &&will build pro-
duces ECG's that are essentially identical to those
p r o d u c e d by commercial m a c h i n e s c o s t i n g
$10,000 dollars or more. In order to keep our cost to
a minimum we use a standard PC as an operator
interface and output device. That way you can print
out a hard copy of your ECG or just display it on
your monitor.

Biological theory
In order to understand the electronic operation of
a n electrocardiograph, we need to understand
some basic biological principles. As shown in Fig.
1, the heart consists of four chambers which are
organized as two pumps-the so-called right and
left heart. The right heart collects the blood return-
ing from the body and pumps it to the lungs, while
the left heart collects blood from the lungs and
pumps it to the body.
Each pump has two parts: the upper chambei-
known a s the atrium a n d the lower chamber en an eye on your health and
known a s the ventricle. The atrium collects blood
between cycles and at the appropriate time con- 'learn about medical eleetrbnics
tracts. filling the lower ventricles. The ventricles
then contract and DumD blood to the luilgs or bodv. with the Radio-Electronics
The heart is cdntroiled by a pulse Kenerato;,
known a s the pacemaker, located in the right
atrium, which initiates cardiac action. It is analo- ' B
gous to the clock in a digital system. The pulse it " i I
t
aenerates is first sent to both atrium which causes
0
H. EDWARD ROBERTS, M.D.
them to contract, filling the ventricles. After a delay
of approximately 150 milliseconds, the ventricles
are then triggered by the same pulse. which causes
them to contract. As in a digital system, the timing
relationships are quite important and much of the i
disease associated with the heart is related to tim- -
6
t
L
ing defects.
Figure 2 shows a typical signal as seen on ail WARNING!/ Thls article deals wlth and involves subject k 'i.,'Fdze
matter and the use of materials and substances that
ECG. The first pulse, called the "P" wave, is gener-
y be hazardous to health and Ilfe. Do not attempt to
ated by the pacemaker. The next pulse, called the lement or use the lnformat~oncontalned hereln un-
"QRS complex." represents the electrical signal less you are experienced and sk~lledw~threspect to
generated by the ventricles contracting. The "T such subject matter, materials and substances.
wave" which follows the QRS complex is generated Furthermore, the lnformatlon contalned In thls artlcle
as the muscles of the ventricles relax, or repolarize. IS being prov~ded solely to readers for educational pur-
A standard ECG consist of 12 channels: each poses Nothlng conta~nedhereln suggests that the
channel "looks" at the heart from a different elec- monltorlng system described hereln can be orshould be
trical axis. The different "views" allow u s to inter- used by the assembler or anyone else In place of or as
pret the activity of different parts of the heart. The n adjunct to professional medical treatment or advlce. jz:;s.3L
timing relationships between different compo- Neither the publisher nor the author make any represen-
nents of the heart will identify defects in the con- tatlons as for the completeness or the accuracy of the
duction pathways. lnformatron conta~nedhereln and dlsclaim any liability
for damages or injuries, whether caused by or arlslng
from the lack of completeness, Inaccuracies of the Infor-
2
How ECG's are used 7
In patients with high blood pressure, the left matlon, mlslnterpretat~onsof the directions, mlsappllca-
w
ventricle will become quite large due to its ill- t ~ o nof the information or otherwise. Y
BODY

P-R SEGMENT S-T SEGMENT


\ I

P-R INTERVAL& I
I
QRS 'COVPLEX I
A O-T INTERVAL '

FIG. 2-A TYPICAL SIGNAL as seen on an


ECG. The first pulse, called the "P" wave,
is generated by the pacemaker. The next
pulse, called the "QRS complex," repre-
sents the electrical signal generated by
the ventricles contracting. The "T wave"
which follows the QRS complex is gener-
ated as the muscles of theventricles relax,
or repolarize.

+
BODY

FIG. 1-THE HEART CONSISTS OF FOUR CHAMBERS which are organized as two
pumps, known as the right and left heart. The right heart collects the blood returning from
the body and pumps it to the lungs, while the left heart collects blood from the lungs and
pumps it to the body.

TABLE 1-110 PORT ADDRESS FUNCTIONS

1 Port Address I Control Pulse Function 1


Generates Clock for Multiplexer Sequencer
Generates Clear for Multiplexer Sequencer
Latches ECG Control Signal Byte In IC16 FIG. 3-TEN LEADS ARE CONNECTED to
Latches Lead Offset Data in lC17 the patient in a 12-channel system: they
are right arm, left arm, left leg, and 6 chest
leads called the V leads. The right leg is
creased work load. That is seen duce an extra, or irregular heart used as a ground and as an input to re-
a s a significant increase in the beat every now and then, which duce system noise.
amplitude of the QRS complex. may occur in the top or bottom of
Treatment of the high blood pres- the heart. It is a condition known monitoring equipment is used in
sure will allow the left ventricle to a s arrythymias. The irregular ambulances and intensive care
return to normal size, which sig- beats can be quite dangerous if units for instantaneous analysis
nificantly decreases the chance they occur frequently or if they of irregular beats. Often t h i s
of a heart attack. occur during certain intervals in analysis is performed automat-
Since the amplitude of the elec- the normal cardiac cycle. Many ically by arrythymia detectors.
trical signals in the heart are a researchers believe that the most Another area of particular in-
function of chemicals i n the common cause of death in males terest in ECGs is the "ST seg-
body, it is possible to predict ab- is due to irregular beats occur- ment. That is the area between
"

normalities. For example, an ele- ring at a time such that they the QRS and the T wave. It is very
vated potassium level will pro- "scramble" the normal electrical predictive of obstructed arteries
duce a tall peaked T wave. timing in the heart-the situa- before any damage occurs to the
If a portion of the ventricle is tion is known as fibrillation. heart. Obstruction of an artery
damaged, a so-called "Q wave" is Special ECG systems, known will result in a depressed ST seg-
(I)
2 formed which is simply a nega- a s Holter monitors, can detect ment of the ECG-it will fall be-
* tive-going QRS complex. The lo- these irregular beats. They are low the base line in the affected
cation of the damage can be simply ECGs with one or more leads. The exercise cardiogram,
determined by noting which channels that store each of the or stress test, looks primarily at
1 leads contain the Q wave. That's 80,000 or so beats in one day. The the ST portion of the ECG to pre-
how a doctor can tell where you data from the Holter is then fed dict if any of the hearts arteries
E have had a heart attack. into a computer which analyzes are becoming clogged.
$ Most normal individuals pro- it for a r r y t h y m i a s . S i m i l a r It is possible to become quite
I

OPTICALLY ISOLATED 8

NOT ISOLATED

FIG. 4-BLOCK DIAGRAM OF THE COMPLETE ECG SYSTEM. The system logically
divides into the front-end e i e c t r ~ n i c sand the controller. Data communication between
the analog and digital portion of the ECG is accomplished through optical isolators,
which helps keep the patient isaiated.

PARTS LIST-CONTROLLER
All resistors are %-watt, 5%, unless IC14-74HC138 1-of-8 decoder maiics, assembly instructions, and
otherwise noted. IC15-82C52 UART checkout- and plot-software design
R1-10 ohms IC16, IC17-74HC573 octal latch specifications (ECG-DP): $27.00.
R2, R7-E10-10,000 ohms IC18-74HC74 dual D fl~p-flop e front-end PC board, controller
R3-R6, R15, R16-1000 ohms IC19-MC145406 RS232 transce~ver PC board, and design package from
Rll-7500 ohms IC20-AD0829 AID converter above (ECG-PC): $74.00.
R12-24,000 ohms IC21, lC22-DAC0830 DIA converter s Complete kit of parts, including
R13-30,000 ohms IC23, IC24-NE5532A op-amp both PC boards. IC's, sockets, pas-
R14-10,000 ohms > 8, SIP IC25-74HC14 hex Schm~ttinverter sive components, design package,
Capacitors IC2674HC00 quad NAND gate ECG software, and checkout soft-
C1-C22, C25, C26, C28. C29, C33. C34, IC27-PS2501A-2 opto~soiator ware (ECG-KIT): $289.00.
C42-0 47 FF, ceramlc d~sk lC28, IC31-not used Lead kit consisting of 50 feet of
C23, C24-22 pF, ceramlc d~sk IC29-ICL7660 DC-DC converter 29-gauge shielded cable, 10 alligator
C27, C30--0 001 p.F, metal fllm IC30--78L06AC voltage regulator clips, heat-shrink tubing, and in-
C31-220 pF, ceramlc disk lC32-7805 voltage regulator structions (ECG-LD): $53.00.
C32-10 p.F, 10 volts, electrolytic Dl-D4-IN914 dlode EPROM containing ECG soft-
C35, C38, C39,-10 p.F, 10 volts, D5-5 1-volt Zener diode ware, ECG resident portion of check-
tantalum D 6 6 - v o l t Zener dlode out software (ECG-PROG): $45.00.
C36, C37, C40, C41-1 p.F, 10 volts, QI-IRFZ10 N-channel MOSFET e Set of four programmed PAL'S
tantalum Other components (ECG-PAL): $67.00.
Semiconductors XTALI-2 4576 MHz crystal e Case as shown with mounting
ICI-Z80 CPU XTAL2-8 00 MHz osc~llator hardware (ECG-CASE): $29.00
IC2-IC5, IC12-74HC245 bus S1-SPDT momentary contact sw~tch @ Package of 100 self-adhesive elec-

transcelver 501-0825 connector trodes (ECG-EL): $20.00.


IC6, IC8-Altera EP320 PAL Please include $5.00 shipping and
IC7-Altera EP600 PAL Note: The following items are available handling for design package and L

169-27C256 EPROM
IC10, lC11-55257 stat~cRAM
from DataBlocks, Inc., Glenwood,
GA 30428, (912) 568-7101.
electrodes, $10.00 shipping and han-
dling for all other products. Georgia
2
IC13-74HC688 equallty comparator B Design package including sche- residents must add sales tax. (D

2
BlNO

FROM L E F T A 3 H

112
SO5

FF1OY L E F T LEC. /1

t-
I
I
2
i
1
W I
2 FIG. 5-THE FRONT-END ELECTRONICS takes in the signals from the 9 input leads
0 located on the patient. The small signal frcm each of the leads is fed into the quad op
$ amps iC7, IC2, and IC3-a.
BAT +SV 1
CLK

------
GTLA I

FIG. 6-PAL IC8 SEQUENCES THE MULTIPLEXER ADDRESS LINES so that each input
signal is sequentially passed to the multiplexer output for processing. This timing
diagram shows the relationships between the PAL'S input, output, and control signals.

ticated text written by Marriott is then, is to distinguish the car-


entitled Practical Electrocar- diac signal from the unwanted
diography. Sources for these sigiial. hat is accomplished in
texts are listed in the Sources biological instrumentation in
Box of this article. much the same way as industrial
instrumentatioi~:by the use of
Biological interface differential amplifiers. These cir-
The patient is connected to the c u i t s can attenuate the u n -
electrocardiogram via 3-12 leads wanted signal by 100 dB or more.
in a typical system. The 3-lead The electrode that connects the
systems are used when only the ECG to the patient must make a
cardiac rhythm is to be studied- low-impedance connection be-
in an ambulance, monitoring an tween the system and the pa-
athlete, in an intensive care u n ~ t . tient's skin. That is typically
etc. If a detailed analysis of the accomplished by the use of dis-
heart is required. a l2-channel posable silver electrodes, which
system is normally used. The sys- we will use in our system. The
tem that we will build will gener- electrodes are made -of a silver-
ate a full 12-channel read out. chloride gel that provides low im-
Ten leads are connected to the pedance a n d a minimum
patient in a 12-channel system: ~imountof electrical noise with
they are right arm, left arm, left the skin.
leg, and 6 chest leads called the V One of the fundamental princi-
leads (see Fig. 3). The right leg is ples that medicine is based on is
used as a ground and as an input from the Hypocrites Oath, "do no
to reduce system noise. harm." That means under no cir-
You might well ask how we get a cumstances should there ever be
12-channel system only using 9 a possibility of doing any harm
active leads. That is accom- using medical instrumentation.
plished by combining different In the case of an ECG, the main
leads together. For example, lead concern would be the shock po-
"aVR" is equal to the voltage at tential through the electrodes. In
the right arm minus the sum of our system we optically isolate
the voltages at the left leg and left the patient from the rest of the
arm. Table 1 shows how the sig- electronics. Therefore, even if all
nals are combined. Our system the safeguards in the computers
skilled at reading the ECG with- will collect the data in each lead, power supply were to fail, the pa-
cn digitize them, and then digitally tient would still be protected
2 out being a medical expert, and
there are a number of texts on the combine the signals within the from shock.
@ subject that you will find interest- host computer. More on this later. One other concern is to prevent
ing. In particular, try Duben's A typical QRS will have a peak the patient from doing harm to
-1 Rapid Interpretation o f EKGs- amplitude of 1 to 2 millivolts. our equipment. In our ECG we
That may mix with noise (60-Hz provide a resistor-diode network
Q with it, you can become fairly hum, for example) with much on each lead which prevents a
o knowledgeable of ECG's in a mat-
$ ter of a few hours. A more sophis- higher amplitudes. The problem. high voltage from entering the
leads are fed back to the patient
180 degrees out of phase with the
original noise. The signal from
the Wilson Electrode is again in-
verted i n o p - a m p IC3-c a n d
routed to the multiplexer to even-
tually form the reference against
which the nine input signals are
compared.
The multiplexer is made up of
two integrated circuits, IC6 and
IC7, in conjunction with the mul-
tiplexer-controller IC8. Analog-
switch IC6 has 8 inputs (XO-X7).
NOTE: Z = OUTPUT IN HIGH IMPEDANCE STATE One of the eight inputs is con-
FIG. 7-MULTIPLEXER TRUTH TABLE shows how the signals from the PAL are used to nected through a very-low-im-
control the sequencing of the input signals to the input of the instrumentation amplifier. pedance path to the output (X)
according to the 3-bit address ap-
front-end amplifiers. An example input-lead multiplexer. pearing on the control inputs
where this might be significant Notice that we have included a CO-C2. For example, XO is con-
would be in the case where a pa- personal computer and printer in nected to X when the control ad-
tient is "shocked," or defibril- the diagram. Although not a part dress is 000, X1 is connected to
lated, after a cardiac arrest. In of this construction article, they the output when the control ad-
that case up to 400 volts could are an integral part of the system dress is 001, and so on. The addi-
appear across t h e electrodes since they provide the display for tional control address C3, is a n
coming from the patient. the ECG traces. inhibit which, when high,
Now that we have a basic un- causes the output X to be a high
derstanding of the underlying bi- Front-end electronics impedance, effectively turning
ological principles associated As previously discussed, the off the eight input signals to the
with the ECG, let's look at some of 12-trace ECG is derived from 9 multiplexer chip. The output of
the technical details of the ma- input leads located on the pa- IC6 is routed to one of the inputs
chine which we will construct. tient. The small (approximately l to IC7.
Then let's build one! millivolt) signal from each of the Another analog switch, IC7,
leads is fed into the quad op- has 2 outputs and 4 X-Y input
System Theory of Operation amps IC1, IC2, a n d IC3-a, a s pairs (XO-X3 and YO-Y3). The XO
A block diagram of the com- shown in the schematic of Fig. 5. input is the output from IC6. The
plete ECG system is shown in The op-amps are configured as corresponding input YO comes
Fig. 4. The system can be divided non-inverting, unity-gain ampli- from the Wilson Electrode, IC3-c.
into the front-end electronics fiers. They provide a very high in- The signal from IC3-a (input 9 ) is
a n d the controller. The analog put impedance that prevents the the input to X1. The Wilson Elec-
portion attaches to the patient signals from the body electrodes trode signal is also paired with
with 10 lead wires; 9 input leads, from being loaded down. Notice input 9 on Y1. In addition to the
and 1 output, or reference lead. also that the input to each ampli- the nine signal inputs from the
The analog portion of the ECG is fier circuit is shunted to ground patient, a 1-mV test signal and a
powered by two 9-volt batteries to by a 220-pF capacitor (Cl-C10) ground input are routed to the X2
isolate the patient from any po- and two diodes (Dl-D20) in par- a n d X3 i n p u t s respectively.
tentially dangerous power cir- allel. Those components are used Ground is the Y2 input for the
cuitry. In addition, d a t a com- to protect the input of the ampli- corresponding 1-mV signal pair
munication between the analog fier from the high voltages pres- as well as for the X3 ground input
and digital portion of the ECG is ent during cardiac defibrillation, on Y3.
accomplished through optical and to provide patient protection 'ItYo address lines, CTLA and
isolators, which also helps keep in the unlikely event that high. CTLB control which input pair is
the patient isolated. voltage should feed back through switched to the outputs. That is,
The controller section of the the amplifier. when the control address is 0, in-
system contains a 2-80 based The output of the three limb puts XO and YO are switched to
computer with 32K of RAM and leads from IC1-a, IC1-b and IC1-c outputs X a n d Y respectively.
32K of EPROM. The controller are summed into op-amp IC3-b, These control lines as well as the
section of the ECG also contains inverted, and fed back to the pa- control signals for IC6 are derived
the analog-to-digital (AID) con- tient through the 10th lead from the outputs of IC8.
version circuitry to convert the which is attached to the patients IC8 is a programmable array
patient's analog ECG signals to right leg. The composite signal logic (PAL) IC which sequences
the digital data required for com- from the three limb leads is called the multiplexer address lines so
puter processing. In addition, the Wilson Electrode. The Wilson that each input signal is sequen-
t h i s section generates control Electrode signal significantly re- tially passed to the multiplexer
signals to sequence the AID con- duces the common-mode noise output for processing. The PAL is 7
version, compensate for input in the system, since unwanted programmed to advance the ad-
channel offset, a n d control the signals common to the three limb dress on control lines CO through 2
FIG. 8-THE CONTROLLER CONTAINS A Z8O-BASED COMPUTER and a digital control
section which combine to provide all the sequencing, timing, and control signals for the
ECG.
C3 one count each time a pulse is TABLE 2-CONTROL SIGNAL DEFINITION
received on the clock input. Addi-
tionally, a decode function is pro-
grammed in the PAL to control
the state of the control lines
CTLA a n d CTLB a n d , hence,
which signal pairs from the mul-
tiplexer are fed to the differential
inputs of the instrumentation
amplifier. The PAL timing di-
agram in Fig. 6 shows the timing
relationships of the input, out-
put, and control signals from the
PAL. The multiplexer truth table troller board for AID conversion. vided by a 9-volt battery (BI) lo-
in Fig. 7 shows how the signals In addition to its gain function, cated on the front-end board. To
from the PAL are used to control IC1-d is also an active low-pass extend battery life, the system is
the sequencing of the input sig- filter, with a cutoff frequency of powered only during the time re-
nals to the input of the instru- about 100 Hz. quired for a single ECG, with
mentation amplifier. A high pass filter is also imple- each ECG sequence initiated by
The instrumentation amplifier mented in the final stage by feed- depressing the reset switch. That
(IC4 and IC9-b back in Fig. 5 ) is ing the ECG signal from the is accomplished by powering the
one of the key signal-processing emitter of ICl6-c through an ac- s t a r t - u p c i r c u i t r y from t h e
elements in the ECG design. It is tive low-pass filter consisting of RS-232 port on the PC. V, for
a differential amplifier so its out- IC10-a and its associated compo- the power-on latch, IC26, is pro-
put is equal to the difference be- nents. The cutoff frequency for vided by the 5-volt Zener-diode
tween its two inputs multiplied the low-pass filter is about 0.1 Hz, regulator, D5.
by a gain. ln this application, the a n d t h e o u t p u t is fed back When the reset switch is de-
gain is approximately 1000. through IC16-d to the positive in- p r e s s e d , t h e power-on l a t c h
The amplifier also has a n input put of IC9-a where it it is used to changes state, turning 9 1on and
offset adjustment, R12, to com- cancel frequency components of completing the return path for
pensate for minute differences in the original signal below 0.1 Hz. B1. Power for the controller board
the input voltages, as well as a n As a result, the frequency compo- is provided by the 5-volt reg-
output offset capability at pin 7. nents of the ECG signal are lim- +
ulator, IC32. A 6-volt supply for
In this design, the output offset ited to a band between 0.1 Hz and the analog circuits on the con-
at pin 7 is biased by IC5, a preci- 100 Hz. troller board is provided by IC30.
sion voltage reference, to a con- The - 6-volt supply for the analog
s t a n t voltage n e a r 2 . 5 volts. Controller operation circuitry is provided by IC29, a
When the difference between the The controller contains a 280- DC-DC converter. The two 6-volt
inputs of the instrumentation based computer a n d a digital supplies are also used to power
amplifier is zero, the output will control section which combine to portions of the front-end that
be 2.5 volts. As the differential provide all the sequencing, tim- don't require patient isolation.
input voltage changes from zero, ing, and control signals for the The remaining IC's on the con-
the output of the amplifier will ECG (see Fig. 8). The 280 micro- troller board are used to generate
change from 2 . 5 volts by a n processor (IC1)is clocked by an 8- the control signals and to digitize
amount equal to the input dif- MHz oscillator, XTAL2. Octal bus the ECG data from the nine pa-
ference times the gain. transceivers IC2-IC5 buffer the tient leads. Components IC13
The final stage in the analog microprocessor control, address, and IC14 decode I10 instructions
signal path is the isolation cir- and data buses, and, in the case from the controller to produce
cuitry to the AID converter. It con- of IC5, provide bi-directional ca- control pulses that sequence the
sists of IC9-a, IC16-c, and IC10-d. pability on the data bus. Pro- acquisition of the ECG data. Ta-
As t h e v o l t a g e i n t o I C 9 - a grammable logic devices (PLD's) ble 1 lists the control pulse gener-
changes, the current through IC6, IC7, and IC8 generate vari- ated by each I10 port address.
the LED portion of optoisolator ous bus-control and chip-select Each time I10 port 52 is ad-
IC16-c changes, which modu- signals to select the appropriate dressed, the resulting pulse
lates the base of the light-sen- memory and I10 chips. The 32K clocks IC18-a, a D-type flip-flop
sitive t r a n s i s t o r i n t h e op- x 8 EPROM (IC9)stores the CPU wired to divide by two. W o out-
toisolator. That causes the collec- operating system and the ECG puts to the I10 port produce a sin-
tor current to change, developing control program. ?iNo 32K X 8 gle pulse at the output of IC18-a.
a a voltage change across the emit- static RAM'S, IClO and IC11, pro- The pulse is passed to the clock
s ter resistor. That voltage follows vide the CPU with 64K of RAM. input of IC8 on the front-end
the original voltage signal at the Communication with the outside board through IC14, a n op-
input of IC9-a with a 180-degree world is provided through a uni- toisolator also on the front-end
0 phase difference. To correct the versal asynchronous receiver1 board. In a similar manner, a
'
Q
phase reversal a n d to compen-
sate for bias and gain errors, the
transmitter (UART),IC15, and its
associated line transceiver IC19.
CLEAR pulse is developed at the
CLEAR input to IC8 on the front-
o signal is fed through amplifier Power for most of the circuitry end board when the controller
a
cc IC10-d prior to going to the con- on the controller board is pro- continued on page 46
ECC
continued from page 40

addresses I10 port 53. These two


pulses control the operation of
the ECG signal multiplexer. The
other two pulses produced a t
IC14 when the controller ad-
dresses I10 ports 5 4 and 5 5 are
used to latch data into octal
latches IC16 and IC17.
Data latch IC17 stores data for
t h e DIA converters IC21 a n d
IC22. Latch IC16 is used to derive
additional control signals.
Table 2 shows that to turn both
batteries on a n d to place the
100 MHz Switchable lx-lox .... other control signals in an inac-
tive state, all output bits of IC16
must be a 1. To achieve that, the
Any Other Probe Ever Made CPU must output 255 to 110 port
* Economical 54. To subsequently activate, or
For Tektronix. Hewlett Packard, Substantial savings compared to lower, any of the control bits
Philips, Leader, B&K, Kikusui, OEM probes without disturbing the other bits
10 day return policy or turning off the battery power,
Guaranteed performance and quality
all bits must be high except for
the one corresponding to the ac-
tivated control signal. After the
Toll Free 1-800-368-5719 appropriate bits in IC24 have
been activated, the CPU must re-
turn them to their inactive state
C I R C L E 123 O N FRE i INFORMATION C A R D by sending 2 5 5 to I10 port 54.
0 a The same logic applies to the
other signals in Table 2.
To see how this all works to-
gether in the circuit, lets assume
we need to t u r n on both the
positive a n d negative battery-
powered supplies on the front-
end board, a n d write a 127 to
IC21 o n the controller board.
First, the CPU outputs a 2 5 5 to
110 port 54. That places a "1" at all
outputs of IC16. The "1" at Q 8 of
IC16 produces a current flow in
the LED of optoisolator IC16-a,
causing current to flow through
its associated transistor. The en-
suing voltage drop across R60
OF THE 1990's
IF you are able to work with common small hand turns FET Q1 on, completing the
tools, and are familiar with basic electronics (i.e, able battery input circuit to the - 6-
to use voltmeter, understand DC electronics). . . . volt regulator, IC11. The negative
IF you possess average mechanical ability, and have a
VCR on which to practice and learn. . . .then we can supply is now on. The positive
teach YOU VCR maintenance and repair! supplies are turned on by the "1"
FACT: up to 90% of ALL VCR malfunctions are due to
simple MECHANICAL or ELECTRO-MECHANICAL on Q7 of IC16.
breakdowns! Now the CPU must place our
FACT: over 77 million VCRs in use today nationwide! arbitrarily chosen value, 1 2 7 ,
Averaa~VCR needs service or reoair everv 12 to 18
rnonttis! into the octal latch, IC17, on the
* Viejo's 400 PAGE TRAINING MANUAL (over 500 pho- controller board. As shown in Ta-
tos and illustrations) and AWARD-WINNING VIDEO
TRAINING TAPE reveals the SECRETS of VCR rnainte- ble 1, that is accomplished by
nance and repair-"real world" information that is writing 127 to I10 port 55. Once
NOT available elsewhere! the correct number is in the
Also includes all the info you'll need regarding the

'
-I

0
n
BUSINESS-SIDE of running a successful service op-
eration! FREE INFORMATION
CALL TOLL-FREE 1-800-537-0589
latch, the CPU must write it to
the DIA converter. Notice that the
outputs of the octal latch, IC17,
2
ISI
Or write to: Viejo Publ~cationslnc.
5329 Fountain Ave continued on page 88
Los Angeles, C 4 90029 Dept. RE C I R C L E 108 O N F R E E INFORMATION C A R D
AC
9U
C I R C L E 187 O N F R E E INFORMATION C A R D I
c con tin ued from page 46

go to both D/A converters, IC21


and IC22. To write the infoi-ma-
tion to IC21 without disturbing
the contents of the other D/A con-
EDUCATION & INSTRUCTION verter, the CPU must output a
F.C.C. Commercial General Radiotelephone li- 252 to 110 port 5 4 to 1ower Q 1and
cense. Electronics home study. Fast, inexpensive! Q2 of IC16, followed by a 2 5 5 to
"Free" details. COMMAND, D-176, Box 2824, San
Francisco, CA 94126. 110 port 5 4 to return the control
LEARN IBM PC assembly language. 80 pro- their inactive state.
grams. Disk $5.00. Book $18.00. ZIPFAST, Box The function of the D/A con-
12238, Lexington, KY 40581-2238.
verters in the circuit are to pro-
COMPLETEcourse in electronicengineering. Eight
MAKE THE CONNECTION volumes. Includes all necessary math and physics. vide a DC offset to the analog
Free brochure. BANNER TECHNICAL BOOKS, input circuitry to compensate for
1203 Grant Avenue, Rockford, IL 61103.
the DC offset produced by each
ECG electrode on the patient's
skin. Since the offset, in general,
is different for each input lead
each time the system is con-
nected to a a means to
CABLE TV Secrets - the outlaw publication the measure the offset must be pro-
cablecompanies tried to ban. HBO, Mov~eChannel,
Showtime, descramblers, converters, etc. Sup- vided. That is
pliers list included. $9.95. CABLE FACTS, BOX 711- performing a series of calibration
R, Pataskala, OH 43062.
measurements just prior to mak-
DO i t yourself, systems, upgrades, m a j o r
brands discounted, save 30% - 60%. Lowest INVENTORS ing the ECG measurements.
prices anywhere. L.J.H. INC, call Larry at (609) Next time we will continue with
596-0656. INVENTORS! Can you patent and profit from your
AD WITH A idea? Call AMERICAN INVENTORS CORPORA- the construction and Operation
IS A REGULAR
TINT BACKGROUND, TO have your appear like TlON for free information. Overadecadeof service of the ECG device. R-E
1 (800) 338-5656. In Canada call (413) 568-3753.
this one, the cost is $3.90 per word.

Professional driver
for use In aud~tor~um
and studio sound
systems. Giant 100
oz. magnet. 250
watts RMS. 350

VAS= 10.5 cu f t

#RJ-260-495 $5g9$ch #RJ-290-145 $3ga0 $3680


(1-3) (4.~~1
Subwoofer Crossover Network
Super duty, ZOO watt RMS
Orlglnal Sanyo hlgh power handling capability.
end system m~drange. Des~gnedspecially for dual
Large 5" paper cone voice coil subwoofer
with gold look dust cap. systems. 12 dBper
Heavy 12 oz magnet. octave roll-off at
1" ferro fluid cooled 150 Hz lnto 8 ohms.

for use with your exlsting speaker system. Network is


totally passlve, requiring no power source of ~ t sown.
power handling capability.

80 #RJ-281-100 $5" $395 #RJ-260-220 $28" $24'=


(1-9) (10-up)
(1-5) (6-UP)

-
-15 day money back guarantee $15.00 minimum order -

88 CIRCLE 56 ON FREE INFORMATION CARD

Anda mungkin juga menyukai