Anda di halaman 1dari 3

An ISO 9001:2008 Certified Company

Auto Clave

Kaizen make Autoclaves are designed and fabricated for steam sterilization process to Kill
micro-organism through the application of saturated steam under pressure to suit various
I. S. standards application in the growing field of Medical, Agricultural Institution, Bio-
Chemical Industrial, Research laboratories and Various Industries

B-1/6128, Dev Nagar, Karol Bagh, New Delhi 110005, INDIA


info@kaizenimperial.in, www.kaizenimperial.com
An ISO 9001:2008 Certified Company

Features a) Steam Temp. 121.4 oC at 15 psi.

As per BS 2646-5:1993. Autoclaves for sterilization


in laboratories Methods of Test for function and
performance.

Specification The Energy from the water is transferred into high


pressure steam inside the Stainless steel chamber
up to 15-psi pressure or 121.4oC temperature. The
Chamber (electrical) is fill with clean water for
sterilization. This heat process generated chemical
free steam for sterilization.

Temp Controller Microprocessor based LCD temp. controller ( GMP


model optional )

Pressure Controller Mechanically operated pressure/steam release


valve.

Pressure Gauge Analog Dial type Pressure Gauge

Temperature Sensor Standard coupling type PT 100 Sensor ( GMP


model)

Heat Exchanger The electrical heat exchanger transfer latent heat


to the clean fill water to generated steam under
pressure.

Water Level Control Low water level cut off electronic circuit protects
the electrical heat exchanger from low water level
inside the chamber, this Cut's off the power supply
to the heat exchanger.

Steam Outlet Steam generated from the unit is raised and exits
thought the electrically operated steam Purging
solonide valve from the outlet ( GMP model)

B-1/6128, Dev Nagar, Karol Bagh, New Delhi 110005, INDIA


info@kaizenimperial.in, www.kaizenimperial.com
An ISO 9001:2008 Certified Company

Models CS48a CS48b CS48c CS48d CS48e

Capacity 35 ltrs 52 ltrs 119 ltrs 178 ltrs 213 ltrs

Inner Size 30 dia X 50 h 35 dia X 55 h 40 dia X 75 h 55 dia X 75 h 55 dia X 90 h

Inner Chamber Stainless Steel 304 quality

External Stainless Steel 304 quality


Range Ambient to 15 psi pressure / 121.40C
Controller Mechanical Spring loaded value / based temp. controller (GMP)
Sensor PT 100 (GMP)
Accuracy +/-0.10C (GMP)
Uniformity +/-0.20C (GMP)
Max. Ramp up 45 min. upto 121.40C without load with respect to ambient (GMP)
time
Insulation Air Bank 50 mm
Top Lid With Dome Shape Die press Stainless Steel Plate

Optional

Printer Output Printer output port RS232

Validation Data logging facility for temperature validation report.


Facility

B-1/6128, Dev Nagar, Karol Bagh, New Delhi 110005, INDIA


info@kaizenimperial.in, www.kaizenimperial.com