Anda di halaman 1dari 5

ORIGINAL ARTICLE

Turn-of-the-Nut Method Is Not Appropriate for Use in


Cancellous Bone
Melissa K. Ryan, PhD,* Aaron A. Mohtar, PhD,* John J. Costi, PhD,* and Karen J. Reynolds, PhD*

INTRODUCTION
Objective: The level to which bone screws are tightened is Whether used alone or with plates, bone screws are the
determined subjectively by the operating surgeon. It is likely that most common implant device. Mechanically, the purpose of
the tactile feedback that surgeons rely on is based on localized tissue a screw is to transform rotational force into axial motion or
yielding, which may predispose the screwbone interface to failure. force.1 Lag screws, in particular, are used to provide compres-
A limited number of studies have investigated the ratio between sion across fracture fragments; the threaded portion of the
clinical tightening torque and stripping torque. The purpose of this
screw is placed distal to the fracture line, and compression
study was to measure, for the rst time, the ratio between yield
is achieved once head contact occurs and the screw is
torque (Tyield) and stripping torque (Tmax) during screw insertion
restricted in axial translation by the bone surface or plate that
into the cancellous bone and to compare these torques with clinical
it is positioned against. The screws are then tightened until
levels of tightening reported in the literature. Additionally, a rota-
adequate compression is achieved at the fracture site. Sur-
tional limit was investigated as a potential end point for screw inser-
geons perform this by manually tightening to what they sub-
tion in cancellous bone.
jectively perceive to be the optimal torque, depending on
Methods: A 6.5-mm outer diameter commercial cancellous bone the quality of the host material.2,3 A surgeons ability to
screw was inserted into human femoral head specimens (n = 89). accurately gauge the appropriate level of tightening torque
Screws were inserted to failure, while recording insertion torque, depends heavily on experience,4 because there is no quanti-
compression under the screw head, and rotation angle. cation as to what this torque should be. Average clinical
tightening torque levels lie within the range of 84%88% of
Results: The median, interquartile ranges, and coefcient of stripping torque (Tmax)5,6; where, Tmax of screws in the can-
variation were calculated for each of the following parameters: cellous bone lies in the range of 13 Nm, depending on screw
Tyield, Tmax, Tyield/Tmax, slope, Tplateau, and rotation angle. The geometry, bone strength, and anatomical location.711 An
median ratio of Tyield/Tmax and rotation angle was 85.45% and adequate tightening torque is necessary to achieve sufcient
96.5 degrees, respectively. The coefcient of variation was greatest compression and primary stability of the xation; however,
for the rotation angle compared with the ratio of Tyield/Tmax (0.37 tightening beyond this can result in microfailure of the peri-
vs. 0.12). implant bone that may lead to screw loosening. In osteopo-
Conclusions: The detection of yield may be a more precise method rotic bone, overtightening can lead to complete failure of the
than the rotation angle in cancellous bone; however, bonescrew surrounding material and immediate loss of xation because
constructs that exhibit a Tyield close to Tmax may be more susceptible of the weakened bone structure.12 In patients older than 50
to stripping during insertion. Future work can identify factors that years, the incidence of screw stripping during internal xation
inuence the ratio of Tyield/Tmax may help to reduce the incidence of of displaced lateral malleolar fractures was reported to be as
screw stripping. high as 38%,13 with another study reporting an incidence of
45.4% in synthetic cancellous bone specimens.14 These con-
Key Words: insertion torque, yield torque, stripping torque, screw sequences demonstrate the need for an improved method of
stripping, cancellous bone screw insertion into the cancellous bone to avoid overtighten-
(J Orthop Trauma 2015;29:e437e441) ing and inadvertent stripping.
In cortical bone, Thakkar et al15 suggested the use of the
Turn-of-the-nut method that uses a rotational limit and is
commonly used in building construction.16 However, their
results revealed that the strength of the bonescrew construct
is compromised at a lower rotational angle than hypothesized,
Accepted for publication June 22, 2015. and that the optimum rotation angle is likely between 90 and
From the *Medical Device Research Institute, and School of Computer Science
Engineering and Mathematics, Flinders University, Adelaide, Australia. 180 degrees. We sought to identify whether a rotational limit
Supported by the National Health and Medical Research Council, Grant ID is applicable for screw insertion into the cancellous bone.
595933. Furthermore, we wanted to address the overarching question
The authors report no conict of interest. what is the mechanism that signals to the surgeon that ade-
Reprints: Karen J Reynolds, PhD, School of Computer Science Engineering
and Mathematics, Flinders University, GPO Box 2100, Adelaide 5001,
quate tightening has been achieved? It seems likely that the
Australia (email: Karen.reynolds@inders.edu.au). tactile feedback the surgeons are detecting is the onset of
Copyright 2015 Wolters Kluwer Health, Inc. All rights reserved. tissue yielding in the peri-implant bone.17

J Orthop Trauma  Volume 29, Number 11, November 2015 www.jorthotrauma.com | e437

Copyright 2015 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.
Ryan et al J Orthop Trauma  Volume 29, Number 11, November 2015

Therefore, the goals of this study were 2-fold; rst, to Data Analysis
determine the yield/stripping torque ratio, with the hypothesis The torque and compression versus rotation angle
that this is coincident with current clinically reported curves were analyzed using a custom-written program
stopping/stripping torque ratios, and second, to determine (Matlab, MA). The point of screw head contact was dened
whether a rotational limit existed, which would reduce the once the slope of the compression trace exceeded a threshold
incidence of stripping, while maximizing compression, during of 10 N per degree. Plateau torque was dened as the average
screw insertion into the cancellous bone. torque calculated over the 60 degrees of rotation before head
contact, and stripping torque was dened as the maximum
torque (Tmax). Yield torque (Tyield) was determined from the
MATERIALS AND METHODS torque versus rotation plots as follows: a moving average
lter with a span of 5 samples was applied to the torque
Screws versus rotation angle curve to reduce signal noise. The lin-
A partially threaded, stainless steel, 6.5-mm outer ear region of the curve was dened as the region of the curve
diameter (OD) cancellous lag screw, with a 16-mm thread between the 10th and 50th percentiles of plateau torque to
length, 4.4-mm inner diameter and 2.7-mm pitch (Mathys, stripping torque. A line was constructed parallel to the slope
Australia) was used (Fig. 1). but offset by 0.2 degrees (Fig. 2). Tyield was dened as the
torque at which the constructed line intersected the smoothed
Bone Samples torque-rotation curve. The rotation angles between head con-
Twenty-four excised human femoral heads [mean (SD) tact, Tyield, and Tmax were also measured from the curve.
age = 72.8 (12.85) years, 17 women, 7 men] were used.18 ShapiroWilk tests for normality showed that the data
Specimens were retrieved from patients undergoing hemiar- were not normally distributed; consequently, nonparametric
throplasty or full-arthroplasty for osteoporosis or osteoarthritis. analyses were performed. The median and interquartile ranges
The excised heads were cut at the femoral neck, and bone (IQR) were calculated for each of the following parameters:
specimens were extracted from the central portion with parallel Tyield, Tmax, Tyield/Tmax, slope, Tplateau, and the rotation angle
cuts. Specimen slice thicknesses were either 20 mm or 25 mm between head contact and Tmax (ROTHCTmax) and head con-
in width. tact and Tyield (ROTHCTyield). The coefcient of variance
All specimens were individually wrapped in saline- (COV) is reported as both (SD/mean) and (IQR/median).
soaked gauze and stored at 2208C until the time of testing. All statistical analyses were performed in SPSS (version 20;
Specimen donors had given their consent for use in the SPSS, Inc, Chicago, IL) with P , 0.05 considered signicant.
research, and ethical approval was obtained from relevant
institutions for use in the project.
RESULTS
Screw Insertion Tests A total of 89 insertions were performed in the femoral
A table-top test rig was used for testing,18 which com- head specimens, with 9 of the tests excluded from the analysis
prised a torque transducer to monitor insertion torque, a load
cell to monitor compression under the screw head, and a rotary
encoder for monitoring screw rotation. All signals were dig-
itally recorded at a sample rate of 500 Hz. For each insertion,
bone specimens were secured in a self-centering 4-jaw chuck.
Consistent with surgical guidelines, 4.5-mm pilot holes were
drilled in each specimen before being transferred to the test
rig.19 Screws were continually inserted at a rate of 60 rpm
until failure occurred. Failure was dened as achieving the
maximum torque with the slope of the torque versus rotation
curve being negative.

FIGURE 2. Typical output of the torque versus the rotation


FIGURE 1. Partially threaded, 6.5-mm OD, stainless steel angle recorded during insertion. The dotted line represents
cancellous lag screw (Mathys, Australia). the 0.2-degree offset that was used to determine Tyield.

e438 | www.jorthotrauma.com Copyright 2015 Wolters Kluwer Health, Inc. All rights reserved.

Copyright 2015 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.
J Orthop Trauma  Volume 29, Number 11, November 2015 Turn-of-the-Nut

because of errors associated with the torque recordings, tightening torque to Tmax; Cordey et al5 reported that on
resulting in the analysis of 80 tests. A typical torque versus average, surgeons tightened within a mean (SD) of 84%
rotation angle trace is shown in Figure 2. Visual inspection of (613%) of Tmax in the human cancellous tibial bone and
the screw upon removal, poststripping, revealed signicant 88% (618%) in the human cancellous femoral bone. These
bony debris deposits within the valleys of the screw threads. results are consistent with more recent ndings by Tsuji et al,6
No damage to the hardware was observed. who reported mean stopping/stripping torque ratios of 84.5%
The median (IQR) and coefcient of variation for Tyield, (69.7%) in the human cancellous femoral bone and Aziz
Tmax, ratio of Tyield/Tmax, slope, and ROTHCTmax and et al,23 who indicated stopping/stripping ratios between 80%
ROTHCTyield are listed in Table 1. The median ratio of and 85% in the human cancellous humeral bone. This study
Tyield/Tmax was 85.42%. The coefcient of variation was the has demonstrated that the median ratio of Tyield/Tmax is con-
greatest for the slope and the rotation angle between head sistent with the clinical ratios of tightening torque/Tmax re-
contact and Tyield (ROTHCTyield). Peak compression and tor- ported in the literature (Table 1). Furthermore, the COV of
que occurred at a rotation angle of 80 degrees past head this ratio is similar across all studies (0.15, 0.11, and 0.12, for
contact; however, by this point, just more than 33% of screws Cordey et al, Tsuji et al, and this study, respectively). This
had also stripped. The compression under the screw head and supports the theory that the tactile feedback the surgeons use
tightening torque were statistically signicantly larger than to detect that adequate tightening has been achieved is con-
the preceding rotation angle at each rotational increment from sistent with the onset of localized yielding of the peri-implant
10 degrees up to 70 degrees. After 70 degrees, statistically tissue.
signicant differences could not be identied until 120 de- Only 3 (3.8%) specimens exhibited a ratio of Tyield/Tmax
grees past head contact, at which point, the measurements less than or equal to 70%. Interestingly, 27/80 tests (33.8%)
became statistically smaller at each 10-degree increment exhibited a Tyield greater than or equal to 90% Tmax. However,
(Table 2). by 90% of Tmax, surgeons are very close to stripping torque and
A 1-tailed post hoc power analysis comparing the mean if they are waiting for the tactile feedback (that occurs past 90%
values and SDs between consecutive groups revealed that the of Tmax in 30% of cases), it is not surprising that stripping
number of specimens was sufcient to ensure that all occurs with a similar frequency (ie, around 30% of cases13).
statistical differences were actually present (ie, to avoid type Specimens with a high ratio of Tyield/Tmax are most likely at
2 statistical errors). a greater risk of stripping during insertion.
With the goal of reducing the incidence of screw
stripping during insertion in cortical bone, Thakkar et al15
DISCUSSION investigated the implementation of a rotational limit termed
Prevention of overtightening during screw insertion Turn-of-the-nut. Their results showed that the rotation
relies on the surgeons ability to accurately detect the onset of angle between head contact and stripping torque was much
the tightening phase, both visually and by the feel of the rapid lower than initially assumed, and that rotation past 180 de-
increase in torque.20 The stripping torque of a screw is deter- grees resulted in a minimal increase in screw tension, with
mined by the material and geometric properties of the sur- a large increase in the number of stripped screws. There did,
rounding bone,21 which can vary greatly within and between however, appear a little variation in the rotation angle to peak
patients.22 Consequently, this is difcult to ascertain before compression across specimens, suggesting that in cortical
surgery, and methods that have relied on torque-limiting bone, this may provide an alternative end point with reduction
devices have had little success in orthopaedics because the in the incidence of stripped screws.
quality of bone exhibits large individual and topographic We sought to establish whether this was also the case
variations.1 for screw insertion in cancellous bone. Our results indicate
The rst goal of this study was to quantify the ratio that in cancellous bone, the rotation angle between head
between Tyield and Tmax, with the hypothesis that this was contact and stripping is signicantly lower than that of the
consistent with tightening torque levels observed clinically. cortical bone (96.5 degrees vs. 286 degrees), and that there
Only 3 other studies have looked at the ratio of clinical was a large variation in the rotation angle to stripping (COV =

TABLE 1. Results for Analysis of Torque Versus Rotation Data Curves for Screw Insertion Into the Femoral Head Cancellous Bone
(n = 80)
Median (IQR) COV (SD/Mean) COV (IQR/Median)
Tyield, Nm 1.50 (0.981.97) 0.50 0.66
Tmax, Nm 1.84 (1.252.42) 0.45 0.63
Tyield/Tmax, % 85.42 (76.9993.95) 0.12 0.20
Slope, Nm per degree 7.59 (5.799.93) 0.79 0.55
ROTHCTmax, degrees 96.48 (67.32123.12) 0.37 0.60
ROTHCTyield, degrees 38.88 (25.7454.18) 0.54 0.75
HC = head contact; ROTHCTmax = rotations between head contact and stripping torque; ROTHCTyield = rotations between head contact and yield torque; Tmax = stripping torque;
Tyield = yield torque.

Copyright 2015 Wolters Kluwer Health, Inc. All rights reserved. www.jorthotrauma.com | e439

Copyright 2015 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.
Ryan et al J Orthop Trauma  Volume 29, Number 11, November 2015

30 degrees would be necessary; setting a rotational limit


TABLE 2. Compression and Insertion Torque at 20-degree
of 40 degrees would yield approximately a 1% incidence of
Increments of the Rotation Angle Past Head Contact
screw stripping, with a higher median compression achieved.
Rotation However, because some specimens exhibited a rotation angle
Angle Screws
Past HC, Median Compression, Median Torque, Tripped, at Cmax as high as 200 degrees, a rotational limit so low may
degrees (IQR), N (IQR), Nm % result in less than optimal compression; while stripping may
0 19.33 (12.5828.65) 0.67 (0.460.96) 0 be reduced, the incidence of nonunion may increase because
20 131.96* (113.39162.25) 0.99* (0.771.36) 0 of inadequate compression achieved at the fracture site. If the
40 250.53* (170.52331.66) 1.43* (1.031.82) 1.25 rotation angle is to be used, there is a trade-off between
60 290.95* (202.95449.24) 1.73* (1.162.15) 16.25 having a higher incidence of screw stripping or an increased
80 315.64 (217.57452.73) 1.78 (1.232.30) 33.75 incidence of inadequate compression. The screw considered
90 304.23 (217.86441.72) 1.74 (1.232.30) 42.5 here, however, was over twice the OD of the screw analyzed
100 296.07 (212.97432.00) 1.69 (1.222.24) 51.25 by Thakkar et al15 and 2 times the thread pitch. Ricci et al
120 292.93* (208.43424.66) 1.61* (1.202.18) 70 found that although insertion torque increased with increasing
140 265.09* (198.411378.77) 1.48* (1.202.15) 90 thread pitch, no statistical difference in pullout strength was
160 242.03* (167.84349.88) 1.41* (1.041.94) 97.5 observed by varying the pitch for 5.0 mm OD stainless steel
180 221.30* (153.51308.36) 1.31* (0.951.71) 98.75 fully threaded screws in the cancellous bone. Therefore, the
200 194.51* (129.05263.87) 1.25* (0.861.54) 98.75 difference in the rotation angle to failure observed between
220 162.71 (101.273227.18) 1.11 (0.731.37) 100 this study and that of Thakkar et al is most likely because of
the highly porous and heterogeneous structure present in the
The percentage of screws stripped for each angular increment is also listed. cancellous bone compared with homogeneous cortical bone,
A rotation angle of 90 degrees was included for comparison with the literature.
*The current measurement is signicantly different from the preceding measure- rather than the difference in screw pitch.
ment (P # 0.05). Although a large variation in the rotation angle between
head contact and Tyield (ie, the current point of clinical tight-
ening) was observed (COV = 0.54), a much smaller variance
0.37). Because cortical bone is stiffer than cancellous bone, was seen in the ratio of Tyield/Tmax (COV = 0.12). This suggests
a larger rotation angle to stripping may seem surprising. How- that the current method of screw tightening (which attempts to
ever, one contributing factor to this could be the denition of detect yield) is likely a more reliable method than the rotational
head contact; Thakkar et al15 dened head contact as the angle in the cancellous bone. The issue with this, however, is in
rotation angle at which insertion torque increased beyond specimens that exhibit Tyield very close to Tmax.
the baseline, with baseline torque dened as the average of Theoretically, an insertion torque closer to Tmax will
the peak torques measured while the self-tapping screws were result in a more stable construct because axial compression
cutting threads into the bone. Because in this study compres- increases with increasing torque. However, numerous studies
sion under the screw head was measured as a representation have found only a moderate relationship between stopping
of the resultant axial force and this does not occur until after torque and holding strength as measured by the pullout
head contact, a threshold on the slope of the compression force.9,24 A previous study found that in ovine tibial cortical
trace was used to dene head contact. This is a more robust bone, peak pullout strength occurred at 70% of Tmax.17 How-
method, as it is independent of any noise present in the torque ever, a study in human humeral cortical bone reported no
trace because of the heterogeneous nature of the material the signicant difference in pullout strength of screws tightened
screw is being inserted into, particularly in the cancellous to 50%, 70%, or 90% Tmax.25 Both of these studies suggest
bone. Additionally, the high porosity of the cancellous bone that little is gained in tightening past 70% of Tmax; however,
may mean that microdamage to only a small number of sur- both were performed in diaphyseal bone, which is primarily
rounding trabeculae will result in overall failure of the bone cortical. The effects on pullout strength by tightening to
screw construct. various levels of Tmax in cancellous bone have not been
Thakkar et al15 concluded that in cortical bone, a rota- reported. It is possible that tightening to a lower ratio of
tion angle of 180 degrees past screw seating would achieve Tinsert/Tyield will still provide adequate compression and sta-
adequate xation and reduce the incidence of screw stripping bility while minimizing the incidence of screw stripping;
for a self-tapping 3.5 mm cortical screw. In this study of however, this has yet to be investigated in the cancellous
screws in the cancellous bone, however, by applying a rota- bone.
tion angle of 90 degrees past head contact, just more than It is important to note the limitations of the study. First,
20% of specimens had stripped, and by 180 degrees, all but only 1 anatomic location was considered. Because bone
one of the screws had stripped. If the rotation angle was to be volume fraction, elastic modulus, and apparent strength are
used as an end point, a signicantly lower rotation angle known to vary with the anatomic location,22 the relationship
would be necessary in cancellous bone compared with corti- between Tyield and Tmax may also differ. Second, we did not
cal bone and for the larger diameter screw considered in this measure clinical tightening torque as such but compared our
study. In this cohort of data, attempting to achieve maximum data with the clinical data reported in the literature. However,
compression (ie, 80 degrees past head contact) would have both studies considered also used the human cancellous fem-
resulted in the stripping of the screw in 34% of cases. To oral head bone, and the reported ranges of measured stripping
eliminate screw stripping entirely, a rotational limit of torques are comparable with that seen in this study

e440 | www.jorthotrauma.com Copyright 2015 Wolters Kluwer Health, Inc. All rights reserved.

Copyright 2015 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.
J Orthop Trauma  Volume 29, Number 11, November 2015 Turn-of-the-Nut

(0.55.5 Nm and 15.9 Nm, Tsuji et al and Cordey et al, 6. Tsuji M, Crookshank M, Olsen M, et al. The biomechanical effect of
respectively). Finally, the femoral heads were retrieved from articial and human bone density on stopping and stripping torque during
screw insertion. J Mech Behav Biomed Mater. 2013;22:146156.
a combination of patients with osteoarthritis and osteoporosis, 7. Brown G, McCarthy T, Bourgeault C, et al. Mechanical performance of
and these results may not be reective of normal bone. standard and cannulated 4.0 mm cancellous bone screws. J Orthop Res.
Despite these limitations, this is the rst study that 2000;18:307312.
experimentally conrms that current reported clinical levels of 8. Dinah A, Mears S, Knight T, et al. Inadvertent screw stripping during
ankle fracture xation in elderly bone. Geriatr Orthop Surg Rehabil.
tightening torque are coincident with the onset of tissue yield.
2011;2:8689.
Perhaps, the most important nding clinically is that in some 9. Reitman CA, Nguyen L, Fogel GR. Biomechanical evaluation of rela-
specimens, the onset of tissue yielding, which seems to provide tionship of screw pullout strength, insertional torque, and bone mineral
the sensory signal to stop, occurs very close to Tmax. Although density in the cervical spine. J Spinal Disord Tech. 2004;17:306311.
it is necessary to provide adequate tightening to achieve pri- 10. Ricci W, Tornetta P, Petteys T, et al. A comparison of screw insertion
torque and pullout strength. J Orthop Trauma. 2010;24:374378.
mary stability of the xation, specimens with a high ratio of 11. Terrier A, Kochbeck S, Merlini F, et al. Tightening force and torque of
Tyield/Tmax may be at an increased risk of screw stripping. nonlocking screws in a reverse shoulder prosthesis. Clin Biomech
However, undertightening may have equally deleterious re- (Bristol Avon). 2010;25:517522.
sults; if sufcient torque is not achieved, inadequate reduction 12. Drew T, Allcock P. A new method of xation in osteoporotic bone. A
may occur, leading to fragment misalignment, xation failure, preliminary report. Injury. 2002;33:685689.
13. Andreassen G, Hoiness P, Skraamm I, et al. Use of a synthetic bone void
and possible nonunion.26 A priori knowledge of the relation- ller to augment screws in osteopenic ankle fracture xation. Arch
ship between Tyield and Tmax would be highly benecial for Orthop Trauma Surg. 2004;124:161165.
surgeons, specically in light of recent literature that demon- 14. Stoesz M, Gustafson P, Bipinchandra V, et al. Surgeon perception of
strates Tmax can be predicted from the plateau torque before cancellous screw xation. J Orthop Trauma. 2014;28:e1e7.
15. Thakkar S, Langdale E, Mears S, et al. Can the turn-of-the-nut method
head contact.18 Further investigations should look to identify improve cortical screw xation? J Orthop Trauma. 2014;28:195199.
factors that may inuence the ratio of Tyield/Tmax and any 16. Shigley J, Mitchell L. The Design of Screws, Fasteners, and Connec-
differences between healthy and diseased bone. tions. New York, NY: McGraw-Hill; 1983.
17. Cleek T, Reynolds K, Hearn T. Effect of screw torque level on cortical
bone pullout strength. J Orthop Trauma. 2007;21:117123.
18. Reynolds K, Cleek T, Mohtar A, et al. Predicting cancellous bone failure
ACKNOWLEDGMENTS during screw insertion. J Biomech. 2013;46:12071210.
The authors thank Richard Stanley and Roshida 19. Schatzker J. Screws and plates and their application. In: Muller M,
Ab-Lazid for their valuable contributions in the biomechan- Allgower M, Schneider R, et al, eds. Manual of Internal Fixation. Berlin,
NY: Springer-Verlag; 1992.
ical testing of this study. 20. Thomas R, Bouazza-Marouf K, Taylor G. Automated surgical screwdriver:
automated screw placement. Proc Ins Mech Eng. 2008;22:451454.
REFERENCES 21. Thiele O, Eckhardt C, Linke B, et al. Factors affecting the stability of
1. Messmer P, Perren S, Suhm N. Screws. In: Ruedi T, Buckley R, screws in human cortical osteoporotic bone: a cadaver study. J Bone
Moran C, eds. AO Principles of Fracture Management. Davos, Switzer- Joint Surg Br. 2007;89:701705.
land: AO Foundation publishing; 2007:213225. 22. Goldstein S. The mechanical properties of trabecular bone: dependence
2. Carter DR, Shimaoka EE, Harris WH, et al. Changes in long-bones on anatomic location and function. J Biomech. 1987;20:10551061.
structural properties during the rst 8 weeks of plate implantation. 23. Aziz MS, Tsuji MR, Nicayenzi B, et al. Biomechanical measurements of
J Orthop Res. 1984;2:8089. stopping and stripping torques during screw insertion in ve types of
3. Jacobs R, Rahn B, Perren S. Effect of plates on corticl bone perfusion. human and articial humeri. Proc Ins Mech Eng H. 2014;228:446455.
J Trauma. 1981;21:9195. 24. Ryken T, Clausen J, Traynelis V, et al. Biomechanical analysis of
4. McGuire R, St John K, Agnew S. Analysis of the torque applied to bone bone mineral density, insertion technique, screw torque, and holding
screws by trauma surgeons. Comparisons based on years of experience strength of anterior cervical plate screws. J Neurosurg. 1995;83:324329.
and material of implant construction. Am J Orthop (Belle Mead NJ). 25. Tankard S, Mears S, Marsland D, et al. Does maximum torque mean
1995;24:254256. optimal pullout strength of screws. J Orthop Trauma. 2013;27:232235.
5. Cordey J, Rahn B, Perren S. Human torque control in the use of bone 26. Jaeger M, Leung F, Li W. 11A3 open reduction. In: Colton C, Krikler S,
screws. In: Uhthoff H, ed. Current Concepts of Internal Fixation of Schatzker J, et al, eds. AO Foundation Surgery Reference. Davos,
Fractures. Berlin, NY: Springer-Verlag; 1980:235243. Switzerland: AO publishing; 2011.

Copyright 2015 Wolters Kluwer Health, Inc. All rights reserved. www.jorthotrauma.com | e441

Copyright 2015 Wolters Kluwer Health, Inc. Unauthorized reproduction of this article is prohibited.

Anda mungkin juga menyukai