Anda di halaman 1dari 15

Pulse Oximetry in Pediatric Practice

AUTHORS: Sotirios Fouzas, MD,a Kostas N. Priftis, MD,b


abstract and Michael B. Anthracopoulos, MDa
aRespiratory Unit, Department of Pediatrics, University Hospital
The introduction of pulse oximetry in clinical practice has allowed for
of Patras, Patras, Greece; and bThird Department of Pediatrics,
simple, noninvasive, and reasonably accurate estimation of arterial Attikon Hospital, University of Athens School of Medicine,
oxygen saturation. Pulse oximetry is routinely used in the emergency Athens, Greece
department, the pediatric ward, and in pediatric intensive and periop- KEY WORDS
erative care. However, clinically relevant principles and inherent limi- pulse oximetry, children, hemoglobin oxygen saturation
tations of the method are not always well understood by health care ABBREVIATIONS
professionals caring for children. The calculation of the percentage of SaO2arterial blood oxygen saturation
SPO2arterial hemoglobin oxygen saturation by pulse oximetry
arterial oxyhemoglobin is based on the distinct characteristics of light ODCoxyhemoglobin dissociation curve
absorption in the red and infrared spectra by oxygenated versus deox- COHbcarboxyhemoglobin
ygenated hemoglobin and takes advantage of the variation in light www.pediatrics.org/cgi/doi/10.1542/peds.2011-0271
absorption caused by the pulsatility of arterial blood. Computation of doi:10.1542/peds.2011-0271
oxygen saturation is achieved with the use of calibration algorithms. Accepted for publication Jun 1, 2011
Safe use of pulse oximetry requires knowledge of its limitations , which Address correspondence to Sotirios Fouzas, MD, Respiratory
include motion artifacts, poor perfusion at the site of measurement, Unit, Department of Pediatrics, University Hospital of Patras, Rio,
irregular rhythms, ambient light or electromagnetic interference, skin 265 04 Patras, Greece. E-mail: sfouzas@gmail.com
pigmentation, nail polish, calibration assumptions, probe positioning, PEDIATRICS (ISSN Numbers: Print, 0031-4005; Online, 1098-4275).
time lag in detecting hypoxic events, venous pulsation, intravenous Copyright 2011 by the American Academy of Pediatrics
dyes, and presence of abnormal hemoglobin molecules. In this review FINANCIAL DISCLOSURE: The authors have indicated they have
we describe the physiologic principles and limitations of pulse oxime- no nancial relationships relevant to this article to disclose.
try, discuss normal values, and highlight its importance in common
pediatric diseases, in which the principle mechanism of hypoxemia is
ventilation/perfusion mismatch (eg, asthma exacerbation, acute bron-
chiolitis, pneumonia) versus hypoventilation (eg, laryngotracheitis, vo-
cal cord dysfunction, foreign-body aspiration in the larynx or trachea).
Additional technologic advancements in pulse oximetry and its incor-
poration into evidence-based clinical algorithms will improve the ef-
ciency of the method in daily pediatric practice. Pediatrics 2011;128:
740752

740 FOUZAS et al
Downloaded from by guest on September 4, 2016
REVIEW ARTICLES

The clinical assessment of hypoxemia


is notoriously unreliable because it
depends on many factors, including
ambient lighting, skin pigmentation,
tissue perfusion, and hemoglobin con-
centration.1 Even under optimal condi-
tions, arterial blood oxygen saturation
(SaO2) of 75% is required before
central cyanosis becomes clinically
detectable.1,2
The introduction of pulse oximetry in
clinical practice has led to a revolu-
tionary advancement in patient as-
sessment and monitoring, because it
allows for a simple, noninvasive, and FIGURE 1
Reference spectra depicting the absorption coefcients of oxygenated hemoglobin (OxyHb), deoxy-
reasonably accurate estimation of genated hemoglobin (DeoxyHb), methemoglobin (MetHb), and COHb as a function of wavelength. The
arterial oxygen saturation. Pulse oxi- vertical lines indicate the wavelengths (red and infrared) commonly used in pulse oximetry.
meters have become available for
widespread application in pediatric
care, and oxygen saturation has even pending tissue perfusion. However, all PRINCIPLES OF OPERATION
been proposed as the fth vital these early oximeters relied either
sign.3,4 However, clinically relevant The estimation of arterial hemoglobin
on compression and reperfusion of the oxygen saturation by pulse oximetry
principles and inherent limitations
measuring site or on the arterializa- (SPO2) is based on the specic charac-
of pulse oximetry are not always
tion of capillary blood through heat- teristics of oxygenated and deoxygen-
well understood by health care
ing; consequently, they were inconve- ated hemoglobin (oxyhemoglobin and
professionals.5,6
niently large, difcult to use, and, most deoxyhemoglobin, respectively) with
In this review we describe the physio- importantly, inaccurate.7,11 regard to light absorption in the red
logic principles, limitations, and com-
A true revolution in the development of and infrared spectra. Deoxyhemoglo-
mon applications of pulse oximetry in
noninvasive oximetry occurred after bin is characterized by greater red-
daily pediatric practice.
the work of the Japanese electrical en- light absorption (wavelength range:
HISTORY OF PULSE OXIMETRY gineer Aoyagi.12 In an experiment 600 750 nm) in comparison to oxyhe-
aimed to develop a dye-dilution tech- moglobin, whereas oxyhemoglobin ex-
The theoretical background for nonin-
nique to measure cardiac output, he hibits higher absorption in the infra-
vasive assessment of blood oxygen-
realized that the untoward changes in red spectrum (850 1000 nm)14,15 (Fig
ation was set in the early 1900s when it
tissue light absorption caused by the 1). By obtaining the ratio of light ab-
was observed that spectral changes of
pulsatile nature of the arterial blood sorption in the red and infrared spec-
light absorbance in vivo are related to
ow could be used to compute oxygen tra and then calculating the ratio of
tissue perfusion.7 Great advancements
saturation. Thus, the noise in his ex- these 2 ratios (ratio of absorption ra-
in the development of related instru-
periment became the signal for a dif- tios), the percentage of oxyhemoglobin
ments occurred during World War II in
ferent application, which led to the de- can be calculated.12,15
an effort to monitor oxygenation of mil-
itary pilots.7 In 1940, Squire8 reported velopment of the rst pulse oximeter Light absorption in vivo depends on the
on a blood-oxygen-meter for use on in late 1974.11,12 In the next 2 decades, characteristics of the tissues across
the hand, and in 1942, Millikan9 coined after the explosive development of the site of measurement.16,17 During
the word oximeter for a portable ear technologies in light emission and short time periods, the absorption
device that read energy absorption in signal processing, pulse oximeters by skin, subcutaneous fat, muscles,
the red and infrared light spectra. Im- underwent astonishing improve- bones, and capillary and venous blood
portant subsequent work was pre- ments and became available for remains practically constant (constant
sented by Wood,10 who managed to widespread application throughout absorbers). Therefore, any change in
measure oxygen saturation by sus- medical practice.11,13 light absorption should be attributed to

PEDIATRICS Volume 128, Number 4, October 2011 741


Downloaded from by guest on September 4, 2016
and the detector are exactly opposite
to each other with 5 to 10 mm of tissue
between them.15,30 Typical measuring
sites include the nger, the toe, the
pinna, and the lobe of the ear, whereas
for neonates and infants measure-
ments are commonly obtained from
the palm or the sole by using specially
designed probes.28,3032 Less commonly
used sites are the cheek and the
FIGURE 2
Principle of operation of pulse oximetry. Shown is a schematic representation of the layers of human
tongue.30
tissues that absorb light energy at the measuring site (left) and the components of light absorption
(constantDC and variableAC) by distinct tissue characteristics (right). A simplied version of the MISCONCEPTIONS
ratio of absorption ratios equation used to calculate SPO2 is shown also.
Safe use of pulse oximetry requires
comprehension of the information that
the variations of the arterial blood vol- and to the speed and quality of the mi- the method offers.33 SPO2 is, in fact, an
ume related to the cardiac cycle12,1719 croprocessor. There are numerous estimate of SaO2 as derived by arterial
(Fig 2; Supplemental Movie 1). studies of the accuracy and precision blood gas analysis, which in turn does
Currently available pulse oximeters of pulse oximeters in various adult2224 not accurately reect partial oxygen
are equipped with 2 light-emitting di- and pediatric2527 populations. Most tension of the arterial blood (PaO2). In-
odes (LEDs), 1 emitting at the red spec- manufacturers claim mean differ- deed, although SaO2 and PaO2 are re-
trum and the other at the infrared ences (bias) of 2% with SDs (preci- lated through the oxyhemoglobin dis-
spectrum, most commonly at wave- sion) of 4%.15,1820,28 It should be sociation curve (ODC), their relation is
lengths of 660 and 940 nm, respec- noted, however, that these results not linear. Moreover, a series of fac-
tively. Emission of these 2 wavelengths have been reported in subjects with tors can further inuence the shape of
alternates at frequencies of 0.6 to 1.0 SaO2 levels that exceed 80%15,1820,28; the the ODC (Fig 3). Hence, SPO2 (as well as
kHz,15,20,21 and the nonabsorbed energy performance of pulse oximeters dete- SaO2) does not necessarily provide re-
is detected by a semiconductor. A mi- riorates remarkably when SaO2 de- liable information regarding the oxy-
croprocessor subtracts the absorp- creases to 80%.17,24,29 genation status of tissues.30,34,35
tion by constant absorbers, thus ren- The probe of the device must be posi- SPO2 represents an estimate of func-
dering the nal signal, which is tioned in such manner that the emitter tional arterial hemoglobin saturation,
displayed electronically as a plethys-
mographic wave form. The SPO2 is cal-
culated from the conversion of the ra-
tio of absorption ratios by using
dedicated calibration algorithms stored
in the microprocessor of the device.
These algorithms are derived through
SaO2 measurements in healthy volun-
teers breathing mixtures of decreased
oxygen concentrations and are usually
unique for each manufacturer.15,1721
The displayed SPO2 represents the
mean of the measurements obtained
during the previous 3 to 6 seconds,
whereas the data are updated every FIGURE 3
0.5 to 1.0 second.15,1820 The perfor- ODC (continuous line) and factors that inuence its shape. The ODC is shifted to the right (lower dotted
mance of each device is strictly related line) by increased hydrogen ion (H) concentration (acidosis), increased 2,3-diphosphoglycerate
(DPG), increased temperature (To), increased partial pressure of carbon dioxide (PCO2), and the
to the reliability and complexity of the presence of hemoglobin S (HbS) (sickle cell disease). Decreased H (alkalosis), DPG, To, and PCO2 and
algorithms used in signal processing the presence of fetal hemoglobin (HbF) shift the curve to the left (upper dotted line).

742 FOUZAS et al
Downloaded from by guest on September 4, 2016
REVIEW ARTICLES

which refers only to the arterial hemo- of absorption (especially when due to black, blue, or green color and syn-
globin that is capable of transporting venous blood) will affect the signal-to- thetic nails might interfere with pulse
oxygen (functional hemoglobin oxy- noise ratio and drive SPO2 to lower oximetry and result in an underestima-
hemoglobin/[oxyhemoglobin deoxy- than true values.41,42 Fortunately, mo- tion of SaO2.54,55,59 This effect can be
hemoglobin]). Functional saturation tion artifacts can be recognized by avoided by mounting the probe on the
differs from fractional hemoglobin sat- motion alarms or distorted plethysmo- nger sideways.34 New-technology
uration (Fractional hemoglobin oxy- graphic waveforms. However, rhyth- pulse oximeters are less susceptible
hemoglobin/total hemoglobin), which mic motions or vibrations with a fre- to these limitations.21,34,5658
can be measured by most blood gas an- quency similar to heart rate (0.53.5 Bilirubin has no effect on pulse oxime-
alyzers with co-oximetry. The total hemo- Hz) can be particularly troublesome.41 try, because it presents a different
globin denominator in the calculation of Sophisticated read-through-motion spectrum of light absorption (at 450
fractional hemoglobin might include ab- and motion-tolerant technologies con- nm). Therefore, the method can be used
normal or variant hemoglobin mole- tinue to evolve and have improved the reliably for monitoring jaundiced pa-
cules with limited oxygen-carrying prop- performance of the new-generation tients, including neonates.13,15,20,21,28,30,34
erties.30,35,36 Therefore, the terms oximeters.30,4346 However, patients with severe hemo-
functional and fractional hemoglobin lytic jaundice might also have in-
saturation are not interchangeable.36 In Poor Perfusion
creased carboxyhemoglobin (COHb) lev-
situations such as dyshemoglobinemias, Adequate arterial pulsation at the site els, which could potentially lead to
pulse-oximetry readings do not ade- of measurement is essential for distin- erroneous pulse-oximetry readings.15 In
quately reect the oxygen-carrying prop- guishing true signal from background addition, falsely low SPO2 values have
erties of arterial blood.15,28,35,37 It should noise.41,42 Low-perfusion states, such been reported in bronze baby
be noted also that pulse oximetry does as low cardiac output, shock, hypo- syndrome.60
not provide information regarding venti- thermia, vasoconstriction, arterial oc-
lation or acid-base status.30,3840 clusion, or during blood pressure cuff Irregular Rhythms
ination, might impair the functioning
Inaccurate oximetry readings can be ob-
LIMITATIONS OF PULSE OXIMETRY of the device and result in lower SPO2
served with irregular heart rhythms, es-
readings or delayed recognition of
The limitations of pulse oximetry can pecially during tachyarrhythmias.21
acute hypoxemia.13,28,4650 For infants
be generally classied as safe or po- These artifacts can usually be recog-
with cold extremities, local rubbing or
tentially unsafe (Table 1). Safe limita- nized by observing the plethysmo-
heating before the application of the
tions refer to those circumstances in graphic wave form. Currently available
probe might temporarily improve perfu-
which the inaccuracy in the displayed devices possess signal-extraction tech-
sion; however, for hypothermic patients,
SPO2 can be suspected, and its cause is nologies that are capable of recognizing
monitoring by a forehead probe is an
recognizable. In this case the observer such events.20,21,34
alternative option.51 New-generation
is usually warned by the device
devices are equipped with signal- Electromagnetic Interference
(alarm) about the pitfall. A potentially
extraction algorithms and can perform
unsafe limitation is considered to be Electromagnetic energy from electro-
better in low-perfusion states.30,4649
any situation in which the inaccuracy surgical cauterization units and cellu-
is difcult to recognize; the displayed Skin Pigmentation, Nail Polish, and lar phones might interfere with pulse
SPO2 is erroneous, but the observer is Articial Nails oximeters and lead to erroneous SPO2
not warned about the pitfall. readings.61 Special devices with ber-
In theory, skin pigmentation presents
a constant level of absorption that is optic technology should be used dur-
Safe Limitations ing MRI to avoid both interference with
subtracted in the calculation of SPO2
Motion Artifacts and, therefore, should not inuence image quality and potentially danger-
the performance of the device.12,34 ous heating of the sensor with conse-
Motion artifact represents the most
However, dark skin pigmentation has quent thermal injury.61,62
common limitation of pulse oxime-
try.13,15,28 Because the normally pulsa- been incriminated for erroneous SPO2
Potentially Unsafe Limitations
tile (arterial) component of light ab- readings, especially at SaO2 values of
sorption represents no more than 5% 80%.21,52,53 Calibration Assumptions
of the total absorbed energy, any mo- Although data regarding the impact of As stated already, the displayed SPO2 is
tion that alters the remaining fraction nail polish are conicting,5458 polish of the result of the conversion of the ratio

PEDIATRICS Volume 128, Number 4, October 2011 743


Downloaded from by guest on September 4, 2016
744
TABLE 1 Limitations of Pulse Oximetry
Limitations Mechanism Bias Proposed Action
Safe limitations a

Motion Sensor movement Lower SPO2 readings Evaluate plethysmographic waveform

FOUZAS et al
Increased noise caused by changes in nonpulsatile False alarms Stabilize sensor
component of light absorption Change sensor position
Use new-generation pulse oximeters
Poor perfusion Decreased signal caused by decreased pulsatile (arterial) Lower SPO2 readings Evaluate plethysmographic waveform
component of light absorption Check and correct skin temperature and peripheral perfusion
Place sensor more centrally
Use new-generation pulse oximetersb
Skin pigmentation Probably caused by calibration assumptions for dark skin Lower or less reliable SPO2 readings at lower SaO2 Use new-generation pulse oximetersb
pigmentation values
Nail polish and articial nails Decreased signal because of decreased light absorption with Lower SPO2 readings Change sensor position
articial nails or nail polish of black, blue, or green color
Irregular rhythms Increased noise caused by tachyarrhythmias Lower or less reliable SPO2 readings Evaluate plethysmographic waveform
Use new-generation pulse oximetersb
Electromagnetic interference External electromagnetic energy interference caused by Lower SPO2 readings Evaluate plethysmographic waveform
electrosurgical cauterization units, cellular phones, or MRI False alarms Avoid external electromagnetic energy sources
devices Heating of the sensor and thermal injury (MRI) Use pulse oximeters with ber-optic technology (MRI)
Potentially unsafe limitationsa
Calibration Device-specic calibration algorithms derived by correlating SPO2 readings of 80%85% are less accurate Use new-generation pulse oximetersb
light absorption ratios over a SaO2 spectrum of 80%100% especially at the extremes of the age spectrum
in healthy young adults
Lower SPO2 values calculated by mathematical equations
Time lag Software-related delay between sudden changes in blood Delay in detecting clinically important desaturation, Use new-generation pulse oximetersb
oxygenation and SPO2 readings which may exceed 1520 s Do not use pulse oximetry as a substitute for
cardiorespiratory monitoring in critically ill patients
Probe positioning The emitted light energy is projected tangentially to the Lower SPO2 readings Place sensor with the emitter and the detector exactly
detector because of inappropriate sensor placement opposite to each other
(penumbra or optical shunting effect) Use probes of appropriate size in neonates and infants
Ambient light interference Intense external light energy (as in phototherapy) may Lower SPO2 readings Use new-generation pulse oximetersb
interfere with the photodetector (ooding effect) Cover the sensor

Downloaded from by guest on September 4, 2016


Abnormal hemoglobin COHb presents red-light absorption similar to oxyhemoglobin In carboxyhemoglobinemia pulse oximetry Check arterial SaO2 if abnormal hemoglobin molecules are
molecules overestimates blood oxygenation suspected (ie, carbon monoxide intoxication)
Methemoglobin absorbs equal amount of energy in the red In signicant methemoglobinemia, SPO2 tends Suspect abnormal hemoglobin molecules if the SaO2SPO2
and infrared spectra, which affects the ratio of absorption toward 85% difference exceeds 5%
Use pulse co-oximetryc
Pulsatile veins Increased noise because of pulsations of venous blood (ie, Lower or less reliable SPO2 readings Use new-generation pulse oximetersb
signicant tricuspid regurgitation, hyperdynamic
circulation states)
Intravenous dyes Intravenous dyes such as methylene blue, indocyanine green, Lower SPO2 readings Do not use pulse oximetry or interpret pulse-oximetry
and indigo carmine interfere with light absorption readings with caution
Check SaO2
a Safe limitations are circumstances in which a possible inaccuracy in the displayed SPO2 can be easily suspected; the observer is usually warned by the device (alarm) about the pitfall. Potentially unsafe limitations are those situations in which the
inaccuracy is difcult to recognize; the displayed SPO2 is erroneous but the observer is not warned about the pitfall.
b New-generation pulse oximeters are less susceptible to these limitations because of more sophisticated calibration and signal-extraction algorithms.

c Pulse co-oximeters are capable of detecting abnormal hemoglobin molecules by using multiwavelength technology.
REVIEW ARTICLES

of absorption ratios into percent satu- propriate size for neonates and equal amounts of energy in the red and
ration by using specic calibration al- infants.13,28,34 infrared spectrums14 (Fig 1). In signi-
gorithms. These algorithms are de- cant methemoglobinemia (MetHb
rived by correlating the ratio of the Ambient Light Interference 30%), the ratio of absorption ratios will
absorption ratios with arterial gas Intense white or infrared light might tend toward the unit (SPO2 85%),
SaO2 measurements in healthy young interfere with pulse oximetry and lead thus underestimating high saturation
volunteers over a range of desatura- to falsely low SPO2 readings. This phe- values and overestimating severe hy-
tion values. Because it is unethical to nomenon, known as the ooding ef- poxemia.71,73,74 If the difference be-
desaturate volunteers below SaO2 lev- fect, is caused by the excessive in- tween SaO2 and SPO2 (the SaO2SPO2
els of 80%, lower SPO2 values are de- crease of the light energy that literally gap) exceeds 5%, the presence of ab-
rived by extrapolation and, therefore, oods the photodetector and drives normal hemoglobin molecules should
are less accurate.15,17,24,29,34 Moreover, the ratio of absorption ratios toward be investigated by co-oximetry.72,73
because the subjects recruited for cal- the unit; this corresponds to an SPO2 Pulse co-oximeters, by taking advan-
ibration purposes are healthy young of 85%.16 Although new-generation tage of novel multiwavelength technol-
adults, the applicability of calibration devices can detect light interfer- ogies, have been shown to accurately
data to patients at the age extremes ence,16,21,34,68 health care professionals, measure both COHb and MetHb.71,7578
has been questioned.13,15,25,30,34 particularly those who handle neo-
nates exposed to phototherapy, must Fetal hemoglobin and hemoglobin S
Time Lag in the Detection of Hypoxic be aware of this potential limitation. present light-absorption characteris-
Events Ambient light interference can be tics similar to those of adult hemoglo-
avoided by simply covering the sensor bin and do not interfere with pulse
Most conventional pulse oximeters
with nontransparent material. oximetry.14,7981 However, physicians
present a clinically signicant delay
should remember that abnormal he-
between a sudden change in blood ox-
Abnormal Hemoglobin Molecules moglobin molecules affect ODC (Fig 3);
ygenation and the related change in
Abnormal or variant hemoglobin mole- thus, the displayed SPO2 value might
the displayed SPO2 values. This time lag
depends on the complexity of the algo- cules might interfere with pulse oxim- not reliably reect tissue oxygenation,
rithms used and might exceed 15 to etry and lead to inaccurate results particularly for children with sickle
20 seconds.34,6365 Although new- that might inuence clinical decision- cell disease.30,82
generation devices have improved re- making.69 Carboxyhemoglobinemia Anemia does not seem to affect the ac-
sponse times, and desaturation events represents the most dangerous limita- curacy of pulse oximetry, at least for
can be detected earlier if the probe is tion of pulse oximetry, because in the hemoglobin levels of 5 g/dL and
placed more centrally (eg, at the ear presence of COHb the method overes- if cardiovascular function is pre-
lobe),13,21,63 pulse oximetry should not timates arterial oxygenation. This served.34,80,83,84 Similarly, polycythemia
be used as a substitute for cardiore- effect is caused by the specic charac- does not seem to interfere with pulse
spiratory monitoring in critically ill teristics of COHb, which exhibits red- oximetry.80
patients.30,34 light absorption similar to that of oxy-
hemoglobin14 (Fig 1). Therefore, Venous Pulsation
Probe Positioning increased COHb levels affect the ratio
of absorption ratios and cause the In case of signicant tricuspid regurgi-
Lower SPO2 readings might occur when
pulse oximeter to overread by 1% for tation and in hyperdynamic circulation
the probe is inappropriately placed,
every 1% increase of circulating states, the pulsatile variation of ve-
especially on the small ngers of neo-
COHb.70,71 Therefore, SPO2 values nous blood might affect signal-to-noise
nates and infants.13,28 In this case, the
should be veried by SaO2 measure- ratio and result in erroneous SPO2
emitted light can be projected tangen-
ments using a co-oximetry method readings.85,86
tially to the detector, sometimes with-
out crossing an arterial bed, phenom- when the presence of COHb is sus-
pected (eg, carbon monoxide Intravenous Dyes
ena which have been described as the
penumbra and optical shunting ef- intoxication).21,69,72,73 Intravenous dyes such as methylene
fects, respectively.66,67 This pitfall can Methemoglobinemia also represents blue (actually used as a rst-line treat-
be avoided by positioning the emitter an important but less dangerous limi- ment for severe methemoglobinemia),
and the detector exactly opposite to tation of pulse oximetry.69 Methemo- indocyanine green, and indigo carmine
each other and by using probes of ap- globin (MetHb) absorbs approximately might cause lower SPO2 readings.15,34,87,88

PEDIATRICS Volume 128, Number 4, October 2011 745


Downloaded from by guest on September 4, 2016
APPLICATIONS OF PULSE OXIMETRY ing range from 95% to 100%, but nor- to 95%, and severe exacerbations with
IN PEDIATRIC PRACTICE mal saturation nadirs can be as low as values of 90%.110,111 Although SPO2
Pulse oximetry has become widely 84% to 86%.101103 However, although values of 92% at presentation have
available in various aspects of pediat- SPO2 values in the range of 90% to 93% been suggested to predict hospitaliza-
ric care. It is routinely found in the are not uncommon during sleep, they tion or return to the hospital,112 more
emergency department and the pedi- might be associated with poorer aca- recent studies have not conrmed this
atric ward, and it is regarded as an demic performance.104 nding.113116 Instead, a 1-hour post-
essential element of patient monitor- treatment SPO2 of 92% to 94% has
ing in pediatric intensive and perioper- Disease-Specic Applications been shown to be a better predictor of
ative care. Its use in the assessment of Respiratory Applications the need for hospitalization.113115
respiratory and hemodynamic param- In pediatric practice, pulse oximetry To date, there is no consensus on the
eters in advanced pediatric care set- must be readily available in any situa- SPO2 thresholds that should be used to
tings is beyond the intentions of this tion associated with hypoxemia. Oxy- admit, treat, and discharge infants
review. gen saturation is a particularly sensi- with acute bronchiolitis.117120 The
tive indicator of disease severity in American Academy of Pediatrics
Normal Values conditions associated with ventilation/ guideline recommends administration
Normal pediatric SPO2 values have not perfusion (V/Q) mismatch, such as ex- of supplemental oxygen if SPO2 values
yet been established. Pulse-oximetry acerbations of asthma or chronic lung fall to 90%.117 The Scottish Intercolle-
readings vary with age and altitude.89,90 disease of prematurity, acute bronchi- giate Guidelines Network (SIGN) rec-
The substantial variation of normal olitis, and pneumonia.3,4,21,26,105108 Con- ommends admission for all symptom-
SPO2 values among studies can be at- versely, SPO2 is not a reliable indicator atic infants with SPO2 values of 92%,
tributed to differences in sample size, of disease severity in proximal (laryn- whereas the decision to admit and/or
instruments used, health of partici- geal or tracheal) airway obstruction treat patients with an SPO2 value of
pants, probe positioning, and mea- such as acute laryngotracheitis, 93% to 94% should be made on an in-
surement protocols. Thus, in healthy foreign-body aspiration, and vocal dividual basis.118 Intermittent is pre-
infants and children, mean SPO2 values chord dysfunction.34 The principle ferred over continuous SPO2 monitor-
at sea level have been reported to be mechanism of hypoxemia in such ing in hospitalized infants, and
97% to 99% (2 SDs, 95%96%),9193 cases is hypoventilation, which pri- patients should be considered for dis-
and they might be lower in neonates marily leads to an increase in PaCO2. charge when the SPO2 is 94% in room
and young infants (range: 93% These patients might not present with air after an observation period of 8 to
100%).93 At moderate altitudes SPO2 particularly low SPO2 readings,3840 be- 12 hours.118 SPO2 values of 94% have
values are somewhat lower (mean: cause, per the alveolar gas equation,109 been shown to increase the likelihood
97%98%; 2 SDs, 93%96%)94,95 and an SPO2 of 90% requires a PaCO2 of of admission and to predict longer
decrease further at high altitudes 80 mm Hg. It should be noted that hospital stay121123; however, small dif-
(3000 m; mean: 86%91%; 2 SDs, coexistent diffuse peripheral airway ferences in SPO2 (92% vs 94%) might
74% 82%).89,90,9698 Authors of a recent obstruction (eg, laryngotracheobron- signicantly inuence the decision to
systematic review concluded that sup- chitis) might cause V/Q mismatch lead- admit or discharge.124 Therefore, it is
plemental oxygen should be adminis- ing to a lower SPO2 level. In the later evident that, on the basis of SPO2 val-
tered to children who reside at alti- scenario, however, hemoglobin de- ues alone, many infants with bronchi-
tudes of 3000 m if the SPO2 is saturation reects a secondary patho- olitis will be hospitalized and treated
85%.99 physiological process rather than the for prolonged periods of time while all
Most children also exhibit a progres- primary mechanism of the disorder. other problems have resolved.125,126
sive uctuation in SPO2 during a 24- Current guidelines state that oxygen Pulse oximetry is essential for prompt
hour cycle. Maximal values occur in saturation should be monitored by detection and management of pediat-
the late afternoon, whereas minimal pulse oximetry during asthma exacer- ric pneumonia, because infants and
values appear in the rst morning bations to assess severity of the dis- children might not appear cyanotic de-
hours. This pattern is evident regard- ease and response to treatment.110,111 spite signicant hypoxemia.127 The Brit-
less of whether children are asleep or Mild asthma exacerbations are associ- ish Thoracic Society guideline for the
awake.100 Basal SPO2 values reported ated with SPO2 values of 95%, moder- management of community-acquired
by polysomnography or home monitor- ate exacerbations with values of 90% pneumonia recommends that symp-

746 FOUZAS et al
Downloaded from by guest on September 4, 2016
REVIEW ARTICLES

tomatic infants and children with an tions of peripheral perfusion, SPO2 val- titration of inspired oxygen concentra-
SPO2 of 92% should be treated with ues can be reliably measured 2 min- tion, it cannot reliably prevent hyper-
oxygen and admitted to the hospital.128 utes after birth.137,139,140 Use of new- oxic events.13,19,30,34 SPO2 values of
However, despite its very good positive generation devices and sensors of 92% do not accurately correlate with
predictive value, the method cannot re- appropriate size, as well as probe at- PaO2, as is clearly depicted by the
liably exclude the disease in emer- tachment to a preductal location (ie, shape of the ODC (Fig 3). At such high
gency settings.127,129 Pulse oximetry is right upper extremity), preferably be- SPO2 values, small variations of SPO2
mandatory for monitoring hospitalized fore connecting the probe to the de- might relate to disproportionally
patients with pneumonia to guide man- vice, might result in more accurate wider variations of PaO2.145 Therefore,
agement and to assess response to and timely readings.133,134 However, caution is required when interpreting
treatment. It is recommended that the health care professionals should be pulse-oximetry readings in situations
SPO2 be maintained at 92% with a aware that, even in uncompromised in which hyperoxia is to be avoided, es-
fraction of inspired oxygen of 0.6; neonates, an increase in SPO2 at levels pecially in case of preterm and low
otherwise, transfer to intensive care of 90% might take 10 minutes to birth weight neonates for whom exces-
should be considered.128 achieve.135140 Therefore, pulse oxime- sive oxygen administration can be par-
try should be used in conjunction with, ticularly harmful.146151 Although a
Cardiovascular Applications but not as a substitute for, clinical as- single best range has not been estab-
Pulse oximetry can be used for heart sessment during the transitional pe- lished yet, there is convincing evidence
rate monitoring or might serve more riod after birth.133,134 that SPO2 values between 85% and 93%
specialized applications, such as the are sufcient to maintain normox-
assessment of peripheral perfusion Neonatal Screening for Congenital
emia152 and to decrease the incidence
and hemodynamic status.130,131 The ple- Heart Disease
of retinopathy of prematurity in in-
thysmographic waveform has been Pulse oximetry has been proposed as a fants receiving supplemental oxy-
shown to be useful in the estimation of reasonable screening tool for the early gen.148151 In extremely preterm neo-
blood pressure when manometry detection of asymptomatic newborns nates, however, lower SPO2 targets (ie,
fails.131 It can also offer a semi- with critical congenital heart disease 85% 89%) have been associated with
quantitative evaluation of pulsus (CCHD).141,142 Single lower-extremity an increased risk of mortality com-
paradoxus by identifying an exagger- SPO2 values obtained after 24 postnatal pared with higher SPO2 levels (ie, 91%
ated decrease of pulse-wave ampli- hours seem to be convenient for large- 95%).153 Further ongoing trials on this
tude during inspiration.132 scale screening.142 An SPO2 threshold issue are expected to resolve the un-
of 95% at low altitudes seems to be certainties surrounding optimum SPO2
Neonatal Resuscitation appropriate.142 Although the method range in premature neonates receiv-
Assessment of skin color is not a reli- has been shown to have excellent spec- ing supplemental oxygen.154
able indicator of oxygenation status icity and negative predictive value,
during the immediate postnatal pe- its sensitivity and false-positive rate NOVEL TECHNOLOGIES AND FUTURE
riod.133 Moreover, the optimal manage- might vary substantially.142144 The DIRECTIONS
ment of oxygenation during neonatal cost/benet balance of routine univer-
resuscitation is critical, because there sal screening has not been well quan- Pulse oximetry has been proven to be
is strong evidence that both hypoxia tied; however, important cost savings an extremely useful tool in patient as-
and hyperoxia can be harmful.134 The could emerge because of early diagno- sessment and monitoring in pediatric
feasibility and reliability of pulse oxim- sis and treatment of infants with practice. However, its widespread use
etry during neonatal resuscitation CCHD.141 Future studies, designed to over the last 3 decades has also re-
have been proven in several stud- assess the impact of routine neonatal vealed its inherent limitations.
ies.135138 Thus, SPO2 monitoring in the screening by pulse oximetry on mor- The theoretical model of conventional
delivery room is currently recom- bidity, mortality, and hospital costs re- pulse oximetry assumes that the arte-
mended for neonates with persistent lated to CCHD, are expected to clarify rial blood is the only light-absorbing
cyanosis, when assisted ventilation this issue.144 pulsatile component. However, this as-
and supplementary oxygen adminis- sumption has been challenged by SPO2
tration are required, or when neonatal Prevention of Hyperoxia readings during motion that fall to
resuscitation is anticipated (high-risk Although for ventilator-dependent pa- 85% (which corresponds to a ratio
deliveries).133 Under acceptable condi- tients pulse oximetry can assist in the of absorption ratios equal to 1); this

PEDIATRICS Volume 128, Number 4, October 2011 747


Downloaded from by guest on September 4, 2016
should not be the case if these desatu- uations of low signal-to-noise ratio such vices.157 Reectance pulse oximeters
rations were merely the result of un- as low-perfusion states. Thus, new- that are based on absorption analysis
characterized noise. New theoretical generation devices use improved algo- of reected rather than transmitted light
models assume that nonarterial ab- rithms of signal extraction, which ulti- have been also introduced into clinical
sorbers also generate a pulsatile signal mately result in more accurate SPO2 practice.158 In light of these ongoing tech-
when motion occurs and that the ratio of readings, especially under critical con- nologic advancements, clinical trials on
absorption ratios should be considered ditions.155,156 In addition, new theories how to incorporate pulse oximetry into
a composite of arterial and nonarterial of multiwavelength pulse oximetry are evidence-based diagnostic and manage-
pulsatile components. These novel con- expected to further improve the per- ment algorithms in daily pediatric prac-
ceptual models are also applicable to sit- formance and applicability of these de- tice are urgently required.
REFERENCES
1. Martin L, Khalil H. How much reduced he- 13. Poets CF, Southall DP. Noninvasive moni- 25. Fanconi S. Reliability of pulse oximetry in
moglobin is necessary to generate central toring of oxygenation in infants and hypoxic infants. J Pediatr. 1988;112(3):
cyanosis? Chest. 1990;97(1):182185 children: practical considerations and ar- 424 427
2. ODonnell CP, Kamlin CO, Davis PG, Carlin eas of concern. Pediatrics. 1994;93(5): 26. Ramanathan R, Durand M, Larrazabal C.
JB, Morley CJ. Clinical assessment of in- 737746 Pulse oximetry in very low birth weight in-
fant colour at delivery. Arch Dis Child Fetal 14. Zijlstra WG, Buursma A, Meeuwsen-van der fants with acute and chronic lung disease.
Neonatal Ed. 2007;92(6):F465F467 Roest WP. Absorption spectra of human fe- Pediatrics. 1987;79(4):612 617
3. Mower WR, Sachs C, Nicklin EL, Safa P, Bar- tal and adult oxyhemoglobin, de- 27. Carter BG, Wiwezaruk D, Hochmann M, Os-
aff LJ. Effect of routine emergency depart- oxyhemoglobin, carboxyhemoglobin and borne A, Henning R. Performance of trans-
ment triage pulse oximetry screening on methemoglobin. Clin Chem. 1991;37(9): cutaneous PCO2 and pulse oximetry moni-
16331638 tors in newborns and infants after cardiac
medical management. Chest. 1995;108(5):
12971302 15. Sinex JE. Pulse oximetry: principles and surgery. Anaesth Intensive Care. 2001;
limitations. Am J Emerg Med. 1999;17(1): 29(3):260 265
4. Mower WR, Sachs C, Nicklin EL, Baraff LJ.
59 67 28. Salyer JW. Neonatal and pediatric pulse
Pulse oximetry as a fth pediatric vital
sign. Pediatrics. 1997;99(5):681 686 16. Mannheimer PD. The light-tissue interac- oximetry. Respir Care. 2003;48(4):386 396
tion of pulse oximetry. Anesth Analg. 2007; 29. Severinghaus JW, Kelleher JF. Recent de-
5. Elliott M, Tate R, Page K. Do clinicians know
105(6 suppl):S10 S17 velopments in pulse oximetry. Anesthesiol-
how to use pulse oximetry? A literature
review and clinical implications. Aust Crit 17. Zonios G, Shankar U, Iyer VK. Pulse oxime- ogy. 1992;76(6):1018 1038
try theory and calibration for low satura- 30. Marr J, Abramo TJ. Monitoring in critically
Care. 2006;19(4):139 144
tions. IEEE Trans Biomed Eng. 2004;51(5): ill children. In: Baren JM, Rothrock SG,
6. Fouzas S, Politis P, Skylogianni E, et al. 818 822 Brennan JA, Brown L, eds. Pediatric Emer-
Knowledge on pulse oximetry among pedi-
18. Hanning CD, Alexander-Williams JM. Pulse gency Medicine. Philadelphia, PA: Saun-
atric health care professionals: a multi-
oximetry: a practical review. BMJ. 1995; ders Elsevier; 2008:50 52
center survey. Pediatrics. 2010;126(3).
311(7001):367370 31. Bell C, Luther MA, Nicholson JJ, Fox CJ,
Available at: www.pediatrics.org/cgi/
content/full/126/3/e657 19. Hess DR, Branson RD. Noninvasive respira- Hirsh JL. Effect of probe design on accu-
tory monitoring equipment. In: Branson racy and reliability of pulse oximetry in pe-
7. Severinghaus JW, Astrup PB. History of
RD, Hess DR, Chatburn RL, eds. Respiratory diatric patients. J Clin Anesth. 1999;11(4):
blood gas analysis. VI. Oximetry. J Clin
Care Equipment. Philadelphia, PA: J. P. 323327
Monit. 1986;2(4):270 288 Lippincott; 1995:193 32. Sedaghat-Yazdi F, Torres A, Fortuna R,
8. Squire JR. Instrument for measuring 20. Jubran A. Pulse oximetry. Intensive Care Geiss DM. Pulse oximeter accuracy and
quantity of blood and its degree of oxygen- Med. 2004;30(11):20172020 precision affected by sensor location in cy-
ation in web of the hand. Clin Sci(Lond). anotic children. Pediatr Crit Care Med.
21. Callahan JM. Pulse oximetry in emergency
1940;4:331339 2008;9(4):393397
medicine. Emerg Med Clin North Am. 2008;
9. Millikan GA. The oximeter: an instrument 26(4):869 879, vii 33. Schnapp LM, Cohen NH. Pulse oximetry:
for measuring continuously oxygen satu- uses and abuses. Chest. 1990;98(5):
22. Hannhart B, Haberer JP, Saunier C, Lax-
ration of arterial blood in man. Rev Sci 1244 1250
enaire MC. Accuracy and precision of four-
Instrum. 1942;13:434 444 teen pulse oximeters. Eur Respir J. 1991; 34. Mechem CC. Pulse oximetry. Available at:
10. Wood EH. Oximetry. In: Glasser O, ed. Med- 4(1):115119 www.uptodateonline.com/online/content/
ical Physics. Vol 2. Chicago, IL: Year Book; 23. Wouters PF, Gehring H, Meyfroidt G, et al. topic.do?topicKeycc_medi/16589. Ac-
1950:664 680 Accuracy of pulse oximeters: the European cessed January 7, 2011
11. Severinghaus JW, Aoyagi T. Discovery of multi-center trial. Anesth Analg. 2002;94(1 35. Perkins GD, McAuley DF, Giles S, Routledge
pulse oximetry. Anesth Analg. 2007;105(6 suppl):S13S16 H, Gao F. Do changes in pulse oximeter ox-
suppl):S1S4 24. Nickerson BG, Sarkisian C, Tremper KK. ygen saturation predict equivalent
12. Aoyagi T. Pulse oximetry: its invention, the- Bias and precision of pulse oximeters and changes in arterial oxygen saturation?
ory, and future. J Anesth. 2003;17(4): arterial oximeters. Chest. 1988;93(3): Crit Care. 2003;7(4):R67
259 266 515517 36. Toffaletti J, Zijlstra WG. Misconceptions in

748 FOUZAS et al
Downloaded from by guest on September 4, 2016
REVIEW ARTICLES

reporting oxygen saturation. Anesth 51. Schallom L, Sona C, McSweeney M, Ma- 65. Stoneham MD. Uses and limitations of
Analg. 2007;105(6 suppl):S5S9 zuski J. Comparison of forehead and digit pulse oximetry. Br J Hosp Med. 1995;54(1):
37. Zijlstra WG. Clinical assessment of oxygen oximetry in surgical/trauma patients at 35 41
transport-related quantities. Clin Chem. risk for decreased peripheral perfusion. 66. Kelleher JF, Ruff RH. The penumbra effect:
2005;51(2):291292 Heart Lung. 2007;36(3):188 194 vasomotion-dependent pulse oximeter ar-
38. Fu ES, Downs JB, Schweiger JW, Miguel RV, 52. Feiner JR, Severinghaus JW, Bickler PE. tifact due to probe malposition. Anesthesi-
Smith RA. Supplemental oxygen impairs Dark skin decreases the accuracy of pulse ology. 1989;71(5):787791
detection of hypoventilation by pulse oxim- oximeters at low oxygen saturation: the ef- 67. Guan Z, Baker K, Sandberg WS. Misalign-
etry. Chest. 2004;126(5):15521558 fects of oximeter probe type and gender. ment of disposable pulse oximeter probes
39. Witting MD, Hsu S, Granja CA. The sensitiv- Anesth Analg. 2007;105(6 suppl):S18 S23 results in false saturation readings that
ity of room-air pulse oximetry in the detec- 53. Bickler PE, Feiner JR, Severinghaus JW. Ef- inuence anesthetic management. Anesth
tion of hypercapnia. Am J Emerg Med. fects of skin pigmentation on pulse oxime- Analg. 2009;109(5):1530 1533
2005;23(4):497500 ter accuracy at low saturation. Anesthesi- 68. Fluck RR Jr, Schroeder C, Frani G, Kropf B,
40. Muoz X, Torres F, Sampol G, Rios J, Mart ology. 2005;102(4):715719 Engbretson B. Does ambient light affect
S, Escrich E. Accuracy and reliability of 54. Cot CJ, Goldstein EA, Fuchsman WH, Hoa- the accuracy of pulse oximetry? Respir
pulse oximetry at different arterial carbon glin DC. The effect of nail polish on pulse Care. 2003;48(7):677 680
dioxide pressure levels. Eur Respir J. 2008; oximetry. Anesth Analg. 1988;67(7): 69. Verhovsek M, Henderson MP, Cox G, Luo
32(4):10531059 683 686 HY, Steinberg MH, Chui DH. Unexpectedly
41. Petterson MT, Begnoche VL, Graybeal JM. 55. St iek H, Gms S, Deniz O, et al. Ef- low pulse oximetry measurements associ-
The effect of motion on pulse oximetry and fect of nail polish and henna on oxygen ated with variant hemoglobins: a system-
its clinical signicance. Anesth Analg. saturation determined by pulse oximetry atic review. Am J Hematol. 2010;85(11):
2007;105(6 suppl):S78 S84 in healthy young adult females. Emerg 882 885
42. Goldman JM, Petterson MT, Kopotic RJ, Med J. 2010; In press 70. Hampson NB. Pulse oximetry in severe car-
Barker SJ. Masimo signal extraction pulse 56. Yamamoto LG, Yamamoto JA, Yamamoto bon monoxide poisoning. Chest. 1998;
oximetry. J Clin Monit Comput. 2000;16(7): JB, Yamamoto BE, Yamamoto PP. Nail pol- 114(4):1036 1041
475 483 ish does not signicantly affect pulse oxi- 71. Barker SJ, Curry J, Redford D, Morgan S.
43. Hay WW, Rodden DJ, Collins SM, Melara DL, metry measurements in mildly hypoxic Measurement of carboxyhemoglobin and
Hale KA, Fashaw LM. Reliability of conven- subjects. Respir Care. 2008;53(11): methemoglobin by pulse oximetry: a hu-
tional and new pulse oximetry in neonatal 1470 1474 man volunteer study. Anesthesiology.
patients. J Perinatol. 2002;22(5):360 366 57. Hinkelbein J, Genzwuerker HV. Fingernail 2006;105(5):892 897
44. Sahni R, Gupta A, Ohira-Kist K, Rosen TS. polish does not inuence pulse oximetry 72. Lee WW, Mayberry K, Crapo R, Jensen RL.
Motion resistant pulse oximetry in neo- to a clinically relevant dimension. Inten- The accuracy of pulse oximetry in the
nates. Arch Dis Child Fetal Neonatal Ed. sive Crit Care Nurs. 2008;24(1):4 5 emergency department. Am J Emerg Med.
2003;88(6):F505F508 58. Brand TM, Brand ME, Jay GD. Enamel nail 2000;18(4):427 431
45. Barker SJ. Motion-resistant pulse polish does not interfere with pulse oxim- 73. Akhtar J, Johnston BD, Krenzelok. Mind the
oximetry: a comparison of new and old etry among normoxic volunteers. J Clin gap. J Emerg Med. 2007;33(2):131132
models. Anesth Analg. 2002;95(4):967972 Monit Comput. 2002;17(2):9396 74. Barker SJ, Tremper KK, Hyatt J. Effects of
46. Gehring H, Hornberger C, Matz H, Konecny 59. Hinkelbein J, Koehler H, Genzwuerker HV, methemoglobinemia on pulse oximetry
E, Schmucker P. The effects of motion arti- Fiedler F. Articial acrylic nger nails may and mixed venous oximetry. Anesthesiol-
fact and low perfusion on the perfor- alter pulse oximetry measurement. Resus- ogy. 1989;70(1):112117
mance of a new generation of pulse oxime- citation. 2007;74(1):75 82 75. Feiner JR, Bickler PE. Improved accuracy
ters in volunteers undergoing hypoxemia. 60. Hussain SA. Pulse oximetry interference in of methemoglobin detection by pulse co-
Respir Care. 2002;47(1):48 60 bronze baby syndrome. J Perinatol. 2009; oximetry during hypoxia. Anesth Analg.
47. Talke P, Stapelfeldt C. Effect of peripheral 29(12):828 829 2010;111(5):1160 1167
vasoconstriction on pulse oximetry. J Clin 61. Rajkumar A, Karmarkar A, Knott J. Pulse 76. Touger M, Birnbaum A, Wong J. Perfor-
Monit Comput. 2006;20(5):305309 oximetry: an overview. J Perioper Pract. mance of the Rad-57 pulse co-oximeter
48. Macleod DB, Cortinez LI, Keifer JC, et al. The 2006;16(10):502504 compared with standard laboratory car-
desaturation response time of nger 62. Dempsey MF, Condon B. Thermal injuries boxyhemoglobin measurement. Ann
pulse oximeters during mild hypothermia. associated with MRI. Clin Radiol. 2001; Emerg Med. 2010;56(4):382388
Anaesthesia. 2005;60(1):657 56(6):457 465 77. Coulange M, Barthelemy A, Hug F, Thierry
49. Hummler HD, Engelmann A, Pohlandt F, H- 63. Young D, Jewkes C, Spittal M, Blogg C, AL, De Haro L. Reliability of new pulse co-
gel J, Franz AR. Decreased accuracy of Weissman J, Gradwell D. Response time of oximeter in victims of carbon monoxide
pulse oximetry measurements during low pulse oximeters assessed using acute de- poisoning. Undersea Hyperb Med. 2008;
perfusion caused by sepsis: is the perfu- compensation. Anesth Analg. 1992;74(2): 35(2):107111
sion index of any value? Intensive Care 189 195 78. Suner S, Partridge R, Sucov A, et al. Non-
Med. 2006;32(9):1428 1431 64. Pinnamaneni R, Kieran EA, ODonnell CPF. invasive pulse co-oximetry screening in
50. Hinkelbein J, Genzwuerker HV, Fiedler F. Speed of data display by pulse oximeters the emergency department identies oc-
Detection of a systolic pressure threshold in newborns: a randomised crossover cult carbon monoxide toxicity. J Emerg
for reliable readings in pulse oximetry. Re- study. Arch Dis Child fetal Neonatal Ed. Med. 2008;34(4):441 450
suscitation. 2005;64(3):315319 2010;95(5):F384 F385 79. Harris AP, Sendak MJ, Donham RT, Thomas

PEDIATRICS Volume 128, Number 4, October 2011 749


Downloaded from by guest on September 4, 2016
M, Duncan D. Absorption characteristics Reference values for pulse oximetry in wheezing less than 24 months of age. Pe-
of human fetal hemoglobin at wavelengths healthy children in coastal Papua New diatr Pulmonol. 1999;27(6):423 427
used in pulse oximetry. J Clin Monit. 1988; Guinea. P N G Med J. 2009;52(12):8 12 106. Choi J, Claudius I. Decrease in emergency
4(3):175177 94. Balasubramanian S, Suresh N, Raeshmi R, department length of stay as a result of
80. Dawson JA, Davis PG, ODonnell CP, Kamlin Kaarthigeyan K. Comparison of oxygen sat- triage pulse oximetry. Pediatr Emerg care.
CO, Morley CJ. Pulse oximetry for monitor- uration levels by pulse oximetry in healthy 2006;22(6):412 414
ing infants in the delivery room: a review. children aged 1 month to 5 years residing 107. Brown L, Dannenberg B. Pulse oximetry in
Arch Dis Child Fetal Neonatal Ed. 2007; at an altitude of 1500 metres and at sea discharge decision-making: a survey of
92(1):F4 F7 level. Ann Trop Paediatr. 2008;28(4): emergency physicians. CJEM. 2002;4(6):
81. Ortiz FO, Aldrich TK, Nagel RL, Benjamin LJ. 267273 388 393
Accuracy of pulse oximetry in sickle cell 95. Beebe SA, Heery LB, Magarian S, Culberson 108. Duke T, Subhi R, Peel D, Frey B. Pulse oxim-
disease. Am J Resp Crit Care Med. 1999; J. Pulse oximetry at moderate altitude: etry: technology to reduce child mortality
159(2):447 451 healthy children and children with upper in developing countries. Ann Trop Paedi-
82. Blaisdell CJ, Goodman S, Clark K, Casella respiratory infection. Clin Pediatr (Phila) atr. 2009;29(3):165175
JF, Loughlin GM. Pulse oximetry is a poor 1994;33(6):329 332 109. West JB, ed. Respiratory Physiology: The
predictor of hypoxemia in stable children 96. Lozano JM, Duque OR, Buitrago T, Behaine Essentials. 8th ed. Philadelphia, PA: Lippin-
with sickle cell disease. Arch Pediatr Ado- S. Pulse oximetry reference values at high cott, Williams & Wilkins; 2008
lesc Med. 2000;154(9):900 903 altitude. Arch Dis Child. 1992;67(3): 110. National Asthma Education and Prevention
83. Jay GD, HughesL, Renzi FP. Pulse oximetry 299 301 Program. Expert Panel Report 3 (EPR-3):
is accurate in acute anemia from hemor- 97. Gamponia MJ, Babaali H, Yugar F, Gilman guidelines for the diagnosis and manage-
rhage. Ann Emerg Med. 1994;24(1):3235 RH. Reference values for pulse oximetry at ment of asthmasummary report 2007
84. McEwen MP, Bull GP, Reynolds KJ. Vessel high altitude. Arch Dis Child. 1998;78(5): [published correction appears in J Allergy
calibre and haemoglobin effects on pulse 461 465 Clin Immunol. 2008;121(6):1330]. J Allergy
oximetry. Physiol Meas. 2009;30(9): 98. Schult S, Canelo-Aybar C. Oxygen satura- Clin Immunol. 2007;120(5 suppl):S94 S138
869 883 tion in healthy children aged 5 to 16 years 111. Global Initiative for Asthma. Global strat-
85. Bucher HU, Keel M, Wolf M, von Siebenthal residing in Huayllay, Peru at 4340 m. High egy for asthma management and preven-
K, Duc G. Artifactual pulse oximetry estima- Alt Med Biol. 2011;12(1):89 92 tion. Available at: http://www.ginasthma.
tion in neonates. Lancet. 1994;343(8906): 99. Subhi R, Smith K, Duke T. When should ox- org/guidelines-gina-report-global-
11351136 ygen be given to children at high altitude? strategy-for-asthma-html. Accessed
86. Stewart KG, Rowbottom SJ. Inaccuracy of A systematic review to dene altitude- January 7, 2011
pulse oximetry in patients with severe tri- specic hypoxaemia. Arch Dis Child. 2009; 112. Geelhoed GC, Landau LI, Le Souf PN. Eval-
cuspid regurgitation. Anaesthesia. 1991; 94(1):6 10 uation of SaO2 as a predictor of outcome in
46(8):668 670 100. Vargas MH, Heyaime-Lalane J, Prez- 280 children with acute asthma. Ann
87. Saito S, Fukura H, Shimada H, Fujita T. Pro- Rodriguez L, Ziga-Vzquez G, Furuya ME. Emerg Med. 1994;23(6):1236 1241
longed interference of blue dye patent Day-night uctuation of pulse oximetry: an 113. Wright RO, Santucci KA, Jay GD, Steele DW.
blue with pulse oximetry readings. Acta exploratory study in pediatric inpatients. Evaluation of pre- and posttreatment pulse
Anaesthesiol Scand. 1995;39(2):268 269 Rev Invest Clin. 2008;60(4):303310 oximetry in acute childhood asthma. Acad
88. Momeni R, Ariyan S. Pulse oximetry de- 101. American Thoracic Society. Cardiorespira- Emerg Med. 1997;4(2):114 117
clines due to intradermal isosulfan blue tory sleep studies in children: establish- 114. Sol D, Komatsu MK, Carvalho KV, Naspitz
dye: a controlled prospective study. Ann ment of normative data and polysomno- CK. Pulse oximetry in the evaluation of the
Surg Oncol. 2004;11(4):434 437 graphic predictors of morbidity. Am J severity of acute asthma and/or wheezing
89. Huicho L, Pawson IG, Len-Velarde F, et al. Respir Crit Care Med. 1999;160(4): in children. J Asthma. 1999;36(4):327333
Oxygen saturation and heart rate in 13811387 115. Kelly AM, Kerr D, Powell C. Is severity as-
healthy school children and adolescents 102. Muzumdar H, Arens R. Diagnostic issues in sessment after one hour of treatment bet-
living at high altitude. Am J Hum Biol. 2001; pediatric obstructive sleep apnea. Proc ter for predicting the need for admission
13(6):761770 Am Thorac Soc. 2008;5(2):263273 in acute asthma? Respir Med. 2004;98(8):
90. Weitz CA, Garrutto RM. A comparative anal- 103. Urschitz MS, Wolf J, Von Einem V, Urschitz- 777781
ysis of arterial oxygen saturation among Duprat PM, Schlaud M, Poets CF. Reference 116. Keahey L, Bulloch B, Becker AB, Pollack CV
Tibetans and Han born and raised at high values for nocturnal home pulse oximetry Jr, Clark S, Camargo CA Jr; Multicenter
altitude. High Alt Med Biol. 2007;8(1):1326 during sleep in primary school children. Asthma Research Collaboration (MARC) In-
91. Mau MK, Yamasato KS, Yamamoto LG. Nor- Chest. 2003;123(1):96 101 vestigators. Initial oxygen saturation as a
mal oxygen saturation values in pediatric 104. Urschitz MS, Wolff J, Sokollik C, et al. Noc- predictor of admission in children pre-
patients. Hawaii Med J. 2005;64(2):42, turnal arterial oxygen saturation and aca- senting to the emergency department with
44 45 demic performance in a community sam- acute asthma. Ann Emerg Med. 2002;40(3):
92. Balasubramanian S, Suresh N, Ravichan- ple of children. Pediatrics. 2005;115(2). 300 307
dran C, Dinesh Chand GH. Reference values Available at: www.pediatrics.org/cgi/ 117. American Academy of Pediatrics, Subcom-
for oxygen saturation by pulse oximetry in content/full/115/2/e204 mittee on Diagnosis and Management of
healthy children at sea level in Chennai. 105. Pavn D, Castro-Rodrguez JA, Rubilar L, Gi- Bronchiolitis. Diagnosis and management
Ann Trop Paediatr. 2006;26(2):9599 rardi G. Relation between pulse oximetry of bronchiolitis. Pediatrics. 2006;118(4):
93. Laman M, Ripa P, Vince JD, Tefuarani N. and clinical score in children with acute 1774 1793

750 FOUZAS et al
Downloaded from by guest on September 4, 2016
REVIEW ARTICLES

118. Scottish Intercollegiate Guidelines Net- of peripheral perfusion. Intensive Care Cardiovascular Disease in the Young, Council
work. Bronchiolitis in children: a national Med. 2005;31(10):1316 1326 on Cardiovascular Nursing, and Interdisci-
clinical guideline. Available at: www.sign. 131. Bendjelid K. The pulse oximetry plethysmo- plinary Council on Quality of Care and Out-
ac.uk. Accessed January 7, 2011 graphic curve revisited. Curr Opin Crit comes Research; American Academy of Pedi-
119. Everard ML. Respiratory syncytial virus- Care. 2008;14(3):348 353 atrics, Section on Cardiology and Cardiac
associated lower respiratory tract dis- Surgery, Committee on Fetus and Newborn.
132. Amoozgar H, Ghodsi H, Borzoee M, Amir-
ease. In: Taussig LM, Landau LI, LeSouf PN, Role of pulse oximetry in examining new-
ghofran AA, Ajami G, Serati Z. Detection of
Martinez FD, Morgan WJ, Sly PD, eds. Pedi- borns for congenital heart disease: a scien-
pulsus paradoxus by pulse oximetry in pe-
atric Respiratory Medicine. Philadelphia, tic statement from the AHA and AAP. Pediat-
diatric patients after cardiac surgery. Pe-
PA: Mosby-Elsevier; 2008:491 499 rics. 2009;124(2):823 836
diatr Cardiol. 2009;30(1):41 45
120. Zorc JJ, Hall CB. Bronchiolitis: recent evi- 143. Thangaratinam S, Daniels J, Ewer AK,
133. Kattwinkel J, Perlman JM, Aziz K, et al;
dence on diagnosis and management. Pe- Zamora J, Khan KS. Accuracy of pulse oxi-
American Heart Association. Neonatal
diatrics. 2010;125(2):342349 metry in screening for congenital heart
resuscitation: 2010 American Heart Asso-
disease in asymptomatic newborns: a sys-
121. Maneker AJ, Petrack EM, Krug SE. Contri- ciation Guidelines for cardiopulmonary re-
tematic review. Arch Dis Child Fetal Neona-
bution of routine pulse oximetry to evalua- suscitation and emergency cardiovascu-
tal Ed. 2007;92(3):F176 F180
tion and management of patients with re- lar care. Pediatrics. 2010;126(5). Available
spiratory illness in a pediatric emergency at: www.pediatrics.org/cgi/content/full/ 144. Meberg A, Brgmann-Pieper S, Due R Jr, et
department. Ann Emerg Med. 1995;25(1): 126/5/e1400 al. First day of life pulse oximetry screen-
36 40 ing to detect congenital heart defects
134. Perlman JM, Wyllie J, Kattwinkel J, et al;
[published correction appears in J Pedi-
122. Mansbach JM, Clark S, Christopher NV, et Neonatal Resuscitation Chapter Collabora-
atr. 2009;154(4):629]. J Pediatr. 2008;
al. Prospective multicenter study of tors. Neonatal resuscitation: 2010 Interna-
152(6):761765
bronchiolitis: predicting safe discharges tional Consensus on Cardiopulmonary
145. Bucher HU, Fanconi S, Baeckert P, Duc G. Hy-
from the emergency department. Pediat- Resuscitation and Emergency Cardiovas-
peroxemia in newborn infants: detection by
rics. 2008;121(4):680 688 cular Care Science With Treatment Recom-
pulse oximetry. Pediatrics. 1989;84(2):
123. Wang EE, Law BJ, Stephens D. Pediatric In- mendations. Pediatrics. 2010;126(5). Avail-
226 230
vestigators Collaborative Network on In- able at: www.pediatrics.org/cgi/content/
full/126/5/e1319 146. Claure N. Automated regulation of inspired
fections in Canada (PICNIC) prospective
oxygen in preterm infants: oxygenation
study of risk factors and outcomes in pa- 135. Kamlin CO, ODonnell CP, Davis PG, Morley
stability and clinical workload. Anesth
tients hospitalized with respiratory syncy- CJ. Oxygen saturation in healthy infants
Analg. 2007;105(6 suppl):S37S41
tial virus lower respiratory tract infection. immediately after birth. J Pediatr. 2006;
J Pediatr. 1995;126(2):212219 148(5):585589 147. Sola A, Saldeo YP, Favareto V. Clinical
practices in neonatal oxygenation: where
124. Mallory MD, Shay DK, Garrett J, Bordley WC. 136. Rabi Y, Yee W, Chen SY, Singhal N. Oxygen
have we failed? What can we do? J Perina-
Bronchiolitis management preferences saturation trends immediately after birth.
tol. 2008;8(suppl 1):S28 S34
and the inuence of pulse oximetry and J Pediatr. 2006;148(5):590 594
respiratory rate on the decision to admit. 148. Castillo A, Deulofeut R, Critz A, Sola A. Pre-
137. Altuncu E, Ozek E, Bilgen H, Topuzoglu A,
Pediatrics. 2003;111(1). Available at: www. vention of retinopathy of prematurity in
Kavuncuoglu S. Percentiles of oxygen sat-
pediatrics.org/cgi/content/full/111/1/e45 preterm infants through changes in clini-
uration in healthy term newborns in the
cal practice and SPO2 technology. Acta Pae-
125. Schroeder AR, Marmor AK, Pantell RH, rst minutes of life. Eur J Pediatr. 2008;
diatr. 2011;100(2):188 192
Newman TB. Impact of pulse oximetry and 167(6):687 688
oxygen therapy on length of stay in bron- 149. Deulofeut R, Critz A, Adams-Chapman I,
138. Mariani G, Dik PB, Ezquer A, et al. Pre-
Sola A. Avoiding hyperoxia in infants
chiolitis hospitalizations. Arch Pediatr ductal and post-ductal O2 saturation in 1250 g is associated with improved
Adolesc Med. 2004;158(6):527530 healthy term neonates after birth. J Pedi- short- and long-term outcomes. J Perina-
126. Unger S, Cunningham S. Effect of oxygen atr. 2007;150(4):418 421 tol. 2006;26(11):700 705
supplementation on length of stay for in- 139. ODonnell CP, Kamlin CO, Davis PG. Obtain- 150. Tokuhiro Y, Yoshida T, Nakabayashi Y, et al.
fants hospitalized with acute viral bronchi- ing pulse oximetry data in neonates: a ran- Reduced oxygen protocol decreases the
olitis. Pediatrics. 2008;121(3):470 475 domised crossover study of sensor appli- incidence of threshold retinopathy of pre-
127. Tanen DA, Trocinski DR. The use of pulse cation techniques. Arch Dis Child Fetal maturity in infants of 33 weeks gesta-
oximetry to exclude pneumonia in chil- Neonatal Ed. 2005;90(1):F84 F85 tion. Pediatr Int. 2009;51(6):804 806
dren. Am J Emerg Med. 2002;20(6): 140. Dawson JA, Kamlin CO, Wong C, et al. Oxy- 151. Askie LM, Henderson-Smart DJ, Ko H. Re-
521523 gen saturation and heart rate during deliv- stricted versus liberal oxygen exposure
128. British Thoracic Society Standards of Care ery room resuscitation of infants 30 for preventing morbidity and mortality in
Committee. British Thoracic Society guide- weeks gestation with air or 100% oxygen. preterm or low birth weight infants. Co-
lines for the management of community Arch Dis Child Fetal Neonatal Ed. 2009; chrane Database Syst Rev. 2009;(1):
acquired pneumonia in childhood. Thorax. 94(2):F87F91 CD001077
2002;57(suppl 1):i1i24 141. Hoffman JIE. It is time for routine neonatal 152. Castillo A, Sola A, Baquero H, et al. Pulse
129. Shah S, Mathews B, Neuman MI, Bachur R. screening by pulse oximetry. Neonatology. oxygen saturation levels and arterial oxy-
Detection of occult pneumonia in a pediat- 2011;99(1):19 gen tension values in newborns receiving
ric emergency department. Pediatr Emerg 142. Mahle WT, Newburger JW, Matherne GP, et al; oxygen therapy in the neonatal intensive
Care. 2010;26(9):615 621 American Heart Association, Congenital care unit: is 85% to 93% an acceptable
130. Lima A, Bakker J. Noninvasive monitoring Heart Defects Committee of the Council on range? Pediatrics. 2008;121(5):882 889

PEDIATRICS Volume 128, Number 4, October 2011 751


Downloaded from by guest on September 4, 2016
153. SUPPORT Study Group of the Eunice Ken- 155. Graybeal JM, Petterson MT. Adaptive lter- 157. Aoyagi T, Fuse M, Kobayashi N, Machida K,
nedy Shriver NICHD Neonatal Research ing and alternative calculations revolu- Miyasaka K. Multiwavelength pulse
Network; Carlo WA, Finer NN, Walsh MC, et tionizes pulse oximetry sensitivity and oximetry: theory for the future. Anesth
al. Target ranges of oxygen saturation in specicity during motion and low perfu- Analg. 2007;105(6 suppl):S53S58
extremely preterm infants. N Engl J Med. sion. Conf Proc IEEE Eng Med Biol Soc. 2004; 158. Agashe GS, Coakley J, Mannheimer PD.
2010;362(21):1959 1969 7:53635366 Forehead pulse oximetry: headband use
154. Tin W, Gupta S. Optimum oxygen therapy in 156. Cannesson M, Talke P. Recent advances in helps alleviate false low readings likely re-
preterm babies. Arch Dis Child Fetal Neo- pulse oximetry. F1000 Med Rep. 2009; lated to venous pulsation artifact. Anesthe-
natal Ed. 2007;92(2):F143F147 1(pii):66 siology. 2006;105(6):11111116

HOW MUCH IS ENOUGH?: Many of my friends exercise all the time, whereas
others hardly ever do. When I ask those not exercising why they dont, most say
they dont have enough time, that it is too hard to start, or that exercising just a
few minutes a day is unlikely to be benecial. Exercise physiologists and others
have long wondered just how much aerobic exercise each day or each week is
necessary to produce a health benet in adults. As reported in USA Today
(Fitness & Food: August 2, 2011), it turns out that it doesnt take much at all.
Federal guidelines suggest that adults should engage in 150 minutes of
moderate-intensity activity each week; this is still a reasonable goal. However,
new data suggest that almost any amount of exercise may be benecial. Adults
engaging in as little as 10 to 15 minutes/day of moderate-intensity exercise
accrue some benet in the prevention of heart disease. In studies evaluating the
risk of heart disease in sedentary and exercising adults, the most dramatic
health benets were seen in those who went from not exercising at all to
exercising a little bit. The data also show that there is an indirect relationship
between the amount of exercise and the risk of heart disease. Compared to
sedentary people, those who engaged in 150 minutes of moderate-intensity
exercise each week had a 14% reduced risk of heart disease. Those who exer-
cised 300 minutes/week had a 20% risk reduction, and a 25% risk reduction if
they exercised 750 minutes/week. Women, for unknown reasons, derive a
greater benet from exercise than men. Bursts of activity followed by long
periods of inactivity, however, were not benecial. This suggests that for better
health, one needs to keep moving. Although researchers have not been able to
quantify the exact health benet to 75 minutes of weekly moderate-intensity
exercise, the American College of Sports Medicine recently revised its guide-
lines. Although the guidelines still recommend that adults engage in at least 150
minutes of moderate-intensity exercise each week to achieve weight reduction
and help maximize the health benets of exercise, just a little exercise, such as
75 minutes/week, is likely to be benecial. The data are fairly clear. To borrow a
marketing phrase from Nike: just do it.
Noted by WVR, MD

752 FOUZAS et al
Downloaded from by guest on September 4, 2016
Pulse Oximetry in Pediatric Practice
Sotirios Fouzas, Kostas N. Priftis and Michael B. Anthracopoulos
Pediatrics 2011;128;740; originally published online September 19, 2011;
DOI: 10.1542/peds.2011-0271
Updated Information & including high resolution figures, can be found at:
Services /content/128/4/740.full.html
Supplementary Material Supplementary material can be found at:
/content/suppl/2011/09/12/peds.2011-0271.DC1.html
References This article cites 144 articles, 26 of which can be accessed
free at:
/content/128/4/740.full.html#ref-list-1
Citations This article has been cited by 12 HighWire-hosted articles:
/content/128/4/740.full.html#related-urls
Subspecialty Collections This article, along with others on similar topics, appears in
the following collection(s):
Pulmonology
/cgi/collection/pulmonology_sub
Respiratory Tract
/cgi/collection/respiratory_tract_sub
Permissions & Licensing Information about reproducing this article in parts (figures,
tables) or in its entirety can be found online at:
/site/misc/Permissions.xhtml
Reprints Information about ordering reprints can be found online:
/site/misc/reprints.xhtml

PEDIATRICS is the official journal of the American Academy of Pediatrics. A monthly


publication, it has been published continuously since 1948. PEDIATRICS is owned, published,
and trademarked by the American Academy of Pediatrics, 141 Northwest Point Boulevard, Elk
Grove Village, Illinois, 60007. Copyright 2011 by the American Academy of Pediatrics. All
rights reserved. Print ISSN: 0031-4005. Online ISSN: 1098-4275.

Downloaded from by guest on September 4, 2016


Pulse Oximetry in Pediatric Practice
Sotirios Fouzas, Kostas N. Priftis and Michael B. Anthracopoulos
Pediatrics 2011;128;740; originally published online September 19, 2011;
DOI: 10.1542/peds.2011-0271

The online version of this article, along with updated information and services, is
located on the World Wide Web at:
/content/128/4/740.full.html

PEDIATRICS is the official journal of the American Academy of Pediatrics. A monthly


publication, it has been published continuously since 1948. PEDIATRICS is owned,
published, and trademarked by the American Academy of Pediatrics, 141 Northwest Point
Boulevard, Elk Grove Village, Illinois, 60007. Copyright 2011 by the American Academy
of Pediatrics. All rights reserved. Print ISSN: 0031-4005. Online ISSN: 1098-4275.

Downloaded from by guest on September 4, 2016

Anda mungkin juga menyukai