Anda di halaman 1dari 12

This copy is for personal use only - distrib

Ann Transplant, 2009; 14(2): 61-72 Review Paper

Received: 2008.10.28
Review of the newest HPLC methods with mass
This copy is for personal use only - distribution prohibited.

Accepted: 2009.01.27
Published: 2009.05.08
spectrometry detection for determination of
immunosuppressive drugs in clinical practice

SE
Magdalena Korecka, Leslie M. Shaw
University of Pennsylvania, Department Pathology and Laboratory Medicine, Philadelphia, PA, U.S.A.
This work was presented at the International Conference Therapeutic Drug Monitoring in Optimizing the

U
Immunosuppressive Therapy, Warsaw, Poland, 26th27th September 2008

Summary

LY L
This copy is for personal use only - distribution prohibited.

High performance liquid chromatography coupled with electrospray mass spec-


N A
trometry is widely used for quantitative determination of immunosuppressive
drugs (sirolimus, tacrolimus, everolimus, CsA and MPA) in biological uids. The
growth in volume for testing these drugs and economic constraints in clinical lab-
O N
oratories has led to heightened demand for high throughput methods.
Fast-ow on-line extraction with switching valve technique and implementation of
automation accelerates sample preparation. For on-line purication the combina-
SO

tion of fast ow of washing solution and narrow-bore extraction column provides


a clean sample in a very short time without compromising precision and accura-
cy. The unique feature of multireactant monitoring tandem mass spectrometry
reduces signicantly the need for chromatographic separation, as long as matrix
effects are not detected, and permits simultaneous measurement of several drugs
R

in one run when they are present in the same specimen. Additionally, the same
method together with the identical sample preparation and HPLC-MS conditions
and setting can be used for measurement of all ve immunosuppressants, four
This copy is for personal use only - distribution prohibited.

PE

of them in blood, MPA in plasma. Thanks to the high sensitivity of LC-MS only a
small volume of biological sample is required for analysis. However for MPA quan-
titation, mass interference attributable to in-source fragmentation of its glucuro-
nide metabolite can be a problem if the latter is not chromatographically separat-
ed from the parent drug before introduction of the sample into the ion source.
Thus, chromatographic separation is extremely important for MPA analysis.
In conclusion, important features of LC-MS methodology for immunosuppres-
sive drugs include: shortened analysis time, increased throughput, selectivity and
lower cost of analysis.
Key words: immunosuppressive drugs therapeutic drug monitoring high-pressure liquid
chromatography with mass spectrometry

Full-text PDF: http://www.annalsoftransplantation.com/fulltxt.php?ICID=883850


py is for personal use only - distribution prohibited.

Word count: 4649


Tables: 3
Figures: 2
References: 73
Authors address: Leslie M. Shaw, University of Pennsylvania, Department Pathology and Laboratory Medicine, 3400 Spruce
Street, Philadelphia, PA 19104, U.S.A., e-mail: shawlmj@mail.med.upenn.edu

61

Electronic PDF security powered by ISL-science.com


This copy is for personal use only - distrib

Review Paper Ann Transplant, 2009; 14(2): 61-72

BACKGROUND tion is too high or rejection when the concen-


tration is too low. The variable pharmacokinet-
As of today 5 major maintenance immunosup- ics of immunosuppressants within and between
pressive drugs are used in modern transplan- patients as a result of variations in absorption,
tation: cyclosporine A (CsA Novartis), tac- distribution and/or elimination makes it impos-
This copy is for personal use only - distribution prohibited.

rolimus (TAC Astellas), sirolimus (SIR Wyeth), sible to reliably predict the best dose for each
everolimus (RAD Novartis) and mycophenol- patient [10]. The resulting variability in trough
ic acid (CellCept Roche, or Myfortic (sodium blood concentration can reach up to 50% [11].
salt) Novartis). For TAC, SIR and RAD trough samples are pri-
marily used for drug monitoring as an estimate

SE
CsA (a cyclic undecapeptide) and tacrolimus of drug exposure. However, for CsA and MPA ad-
(macrolide lactone) are the calcineurin inhibitors ditional samples collected at several additional
which by blocking interleukin-2 production lead- time points over the 12 hour dose interval can
ing to suppression of T lymphocyte proliferation provide a more accurate estimation of drug ex-
[1]. These two drugs are the basis of most main- posure than the trough sample alone and there-

U
tenance immunosuppressive protocols follow- by provide for better clinical outcome [10]. In
ing organ transplantation. Sirolimus (a 31-mem- busy transplant centers the number of samples
bered triene macrolide lactone) and everolimus sent to the laboratory for this essential test can
(semi-synthetic 40-O-(2-hydroxyethyl) derivative be considerable. Thus there is a need for robust,

LY L
of rapamycin) has immunosuppression mecha- sensitive and high throughput methods, with sim-
This copy is for personal use only - distribution prohibited.

nisms distinct from those of CsA and TAC. SIR ple sample preparation, that allows measuring a
N A
and RAD inhibit the functioning of the mamma- few drugs in one blood sample thereby saving
lian target of rapamycin (mTOR), a kinase that time and reagents.
induces intracellular signal transduction at the
O N
G1-S interface, leading to an inhibition of T-cell Mass spectrometry is already well established
cycle progression [2,3]. Mycophenolic acid is a as a quantitative tool for small molecules, and
potent, selective and uncompetitive reversible is based on producing, differentiating and de-
SO

inhibitor of inosine monophosphate dehydroge- tecting ions in the gas phase. Conversion of dis-
nase (IMPDH), which is the rate limiting enzyme solved analyte eluting from a separation system
in the de-novo synthesis of guanosine nucleotides into gas-phase ions occurs in the ion source and
and consequently DNA production [4]. is generally associated with evaporation, pres-
sure reduction and an ionization process. From
R

Due to their complementary mechanism of ac- the ve currently available ion sources, electro-
tion calcineurin inhibitors and inhibitors of spray (ESI) is the most widely used for mass spec-
the mTOR are often combined in the clinic. It trometric quantication of immunosuppressive
This copy is for personal use only - distribution prohibited.

PE

was shown that this co-administration provides drugs. This ion source has the ability to produce
synergistic immunosuppression that is a major singly or multiply charged gaseous ions directly
benet for transplant patients [3,5]. The exist- from an aqueous/organic solvent system by cre-
ing drugs have a few common features that are ating a ne spray of highly charged droplets in
very important for their simultaneous analysis. the presence of a strong electric eld [12,13].
They are soluble in organic solvents like alco- ESI offers many advantages over other ioniza-
hols, acetonitrile and they are practically insol- tion techniques, including the ability to analyze
uble in water [3]. They all, except for MPA, can low and high-mass compounds, excellent quanti-
be measured in whole blood due to high distri- tative capabilities and reproducibility, high sensi-
bution in erythrocytes, between 4060% for CsA, tivity, simple sample preparation, amenability to
about 95% for SIR, 9598% for TAC and 75% automation, soft ionization and minimal amount
for everolimus [68]. of in-source fragmentation. In selected circum-
stances ESI can be quite effective even without
Therapeutic drug monitoring (TDM) of immuno- separation, especially when combined with tan-
py is for personal use only - distribution prohibited.

suppressive drugs helps to achieve optimal thera- dem mass spectrometry [14].
peutic efcacy while minimizing toxicity. All the
drugs have narrow therapeutic ranges for exam- From the ion source the gas phase ions are di-
ple 515 ng/mL for SIR [9] and 13 g/L for rected into a mass analyzer, through the series
MPA [10]. Variation in concentration outside the of lenses. The mass analyzer differentiates these
narrow therapeutic index can result in adverse ions according to their mass-to-charge ratio (m/z).
clinical outcomes: toxicity when the concentra- The triple quadrupole mass analyzer is the most

62

Electronic PDF security powered by ISL-science.com


This copy is for personal use only - distrib

Ann Transplant, 2009; 14(2): 61-72 Korecka M et al Review of the newest HPLC methods with mass spectrometry

common mass analyzer for compounds with SAMPLE PREPARATION


m/z <1000, but also can operate at upper mass-
es above 4000 m/z [13]. The majority of methods For all immunosuppressive drugs the matrix is-
for quantication of immunosuppressive drugs sue has been already resolved. Cyclosporine, tac-
employ a triple quadrupole as the mass analyz- rolimus, sirolimus and everolimus are measured
This copy is for personal use only - distribution prohibited.

er. Quadrupoles consist of four precisely paral- in whole blood while for total MPA analysis plas-
lel rods (poles) equally spaced around a central ma or serum is the matrix of choice [10] and
axis. By applying precisely controlled voltages to the ultra-ltrate obtained from plasma is used
opposing sets of poles, we create what is known as for analysis of free MPA [18,19]. Dried blood
a mass lter. Only ions with a particular mass- spot (DBS) sampling could be an alternative

SE
to-charge ratio will pass through the lter to be and two methods were published that use DBS
detected at a particular applied voltage [13]. capillary blood obtained from nger prick for
analysis CsA and TAC [20,21]. The usefulness of
In the past few years, high-pressure liquid chro- dried blood spot sampling for therapeutic drug
matography with mass spectrometry (HPLC-MS) monitoring of tacrolimus was investigated in re-

U
has been popularly utilized in drug quantitation nal transplant patients. There was no signicant
and pharmacokinetics studies and now is consid- difference between the concentrations (ranging
ered to be the gold standard analytical method in 3.3353.9 g/L) of 34 samples from 26 stable re-
therapeutic drug monitoring [6]. Immunoassays nal transplant outpatients, measured both after

LY L
and HPLC methods with UV detection for meas- venous and DBS sampling. DBS seems promis-
This copy is for personal use only - distribution prohibited.

urement of immunosuppressive drugs reveal too ing for routine patient monitoring [22]. TAC can
N A
many disadvantages compared to HPLC-MS such be additionally measured in tissue biopsy sam-
as cross reactivity of metabolites with the antibod- ples [23], MPA in saliva [24] and CsA in T-lym-
ies for immunoassays [15] and long, laborious phocytes [1].
O N
sample preparation for HPLC-UV [16]. The main
attraction of HPLC-MS is high selectivity and sen- Mass spectrometry detection, since more sensi-
sitivity. This technique permits the quantication tive when compared with UV detection, allows
SO

of the main drug independently from its metab- decreased sample volume used for analysis of
olites. Since the immunosuppressive agents are immunosuppressive agents. For the majority of
used in combined regimens and their amenabil- HPLC-MS methods the volume of whole blood
ity to analysis by HPLC-ESI- mass spectrometry, used ranged from 0.01 [25] to 0.5 mL [26,27],
a very important advantage of HPLC-MS quanti- most often 0.1 mL [3,28,29], and for MPA anal-
R

tation methods is the possibility of simultaneous ysis 0.05 [30,31] 0.2 mL [18] of plasma.
analysis of several compounds in one short run.
For four out of ve major immunosuppressants The high specicity of mass spectrometry allows
This copy is for personal use only - distribution prohibited.

PE

(CsA, TAC, SIR and RAD) the desired specicity for a less intensive sample preparation compared
and sensitivity can be achieved without chromato- to the HPLC-UV technique [26]. The three most
graphic separation in time, which shortens anal- widely used methods of sample preparation for
ysis time, increases throughput and also makes the immunosuppressive drugs quantication
a big impact on the laboratory budget. One of are: 1) on-line sample clean-up with switching
the most recent studies has shown that the same valve and 2-dimentional chromatography and
chromatographic system used for analysis of 4 off line clean-up with 2) solid phase extraction
immunosuppressive drugs in blood can also be columns (SPE) or with 3) liquid-liquid extrac-
used for the analysis of MPA in plasma [17]. All tion (LLE).
HPLC-MS methods need less laborious sample
preparation when compared with HPLC with On line sample clean-up
UV detection. All these advantages of HPLC-MS
methods can shorten total cycle time and save Several HPLC-MS methods employed on-line
reagent usage. sample clean-up followed by protein precipita-
py is for personal use only - distribution prohibited.

tion generally with the mixture of ZnSO4 and


The purpose of this paper is to review the recent- methanol, acetonitrile or acetone with internal
ly published HPLC-MS methods for analysis of standard, as an approach for shortening sample
immunosuppressive drugs and to provide a use- preparation and increasing sample throughput
ful resource for those who are planning to use [3,25,28,29,3236]. Addition of water to blood
mass spectrometry in their laboratories for the before adding ZnSO4 and organic solvents pro-
analysis of immunosuppressants. tects the sample from clumping, gives cleaner

63

Electronic PDF security powered by ISL-science.com


This copy is for personal use only - distrib

Review Paper Ann Transplant, 2009; 14(2): 61-72

Figure 1. Example of setting the switching valve


for quantitation of immunosuppressive
drugs by HPLC with mass spectrometry
detection.
This copy is for personal use only - distribution prohibited.

SE
U
LY L
extracts with less ion suppression and improves (turbulent ow behavior) that allows for the rap-
This copy is for personal use only - distribution prohibited.

the extraction efciency [37]. Additionally, this id passage of large biomolecules of the biological
N A
extraction with water could be adapted to an au- sample (mainly phospholipids and carbohydrates)
todiluter for specimen delivery because water with the simultaneous retention of the small-mol-
allows the specimen to be ushed from the dis- ecule analyte(s) of interest [45]. This sample ex-
O N
penser with no precipitation in the line. All pi- traction stage lasts for less than 1.0 min. Second
petting during the sample preparation procedure dimension chromatography is performed using a
can be automated using a commercially availa- standard analytical column that will be discussed
SO

ble robotic pipetting system from Tecan Freedom later. After switching the valve from position A
EVO (Tecan, Switzerland) [38]. Supernatant ob- to position B the compound(s) of interest is/are
tained from the protein precipitation step is in- back-ushed from the extraction column into the
jected into the system (injection volume ranged analytical one and then to the mass spectrometer
from 5400 L [1,17,18,25,28,3234,3944] and using highly organic, acidied eluting solvent at
R

two-dimensional chromatography is performed a ow rate compatible with the analytical column


through a split arrangement of two pumps and and ion source of mass spectrometer. At the time
6- or 10-port switching valve plumbed for exam- of eluting the rst dimension chromatography is
This copy is for personal use only - distribution prohibited.

PE

ple as shown in Figure 1. In this system two col- in equilibration stage to prepare the extraction
umns (extraction and analytical) and two solvents column for next injection, so the equilibration
(extraction and eluting) are used. First dimen- time does not add to the run time that is usually
sion chromatography is accomplished through- 23 min. All of these unique features of on-line
out the extraction column. The pump that is ded- sample clean up increase the sample throughput
icated for extraction solution (highly aqueous, save on reagent costs, and make the two step sam-
for example water [32,34,41], ammonium acetate ple preparation a very simple and fast procedure
buffer with or without low (2040%) addition of and less likely to introduce errors. Based on a few
methanol [25,33,42] is connected to the sample years of experience with using this technique in
introduction unit. Sample is injected with the ex- our laboratory we can assert that the lack of con-
traction solution and cleaned on extraction col- tamination by samples is comparable to that ob-
umn, all debris and contaminants are ushed out tained using the SPE sample preparation proce-
to waste to protect the mass spectrometer from dure. This observation is additionally supported
contamination. Recently it is advised that as an by several published methods that directly use ex-
py is for personal use only - distribution prohibited.

extraction column a narrow-bore LC column (ID tract from protein precipitation without on-line
0.51.0 mm) packed with large particles of sta- cleaning [31,37,39].
tionary phase material (3050 m), with a very
high ow of extraction solution (35 mL/min for Solid phase extraction
column with ID 1mm) should be used. The com-
bination of the fast ow and large particle size Off line cleaning is another kind of sample prep-
provides the desired chromatographic behavior aration widely used in methods for quantitation

64

Electronic PDF security powered by ISL-science.com


This copy is for personal use only - distrib

Ann Transplant, 2009; 14(2): 61-72 Korecka M et al Review of the newest HPLC methods with mass spectrometry

of immunosuppressive drugs. Solid phase extrac- and the organic layers are then separated by cen-
tion (SPE) is a technique designed for rapid, se- trifugation. The robot is used to transfer the or-
lective sample preparation and purication prior ganic layers from the microtubes to a clean deep
to chromatographic analysis. Using liquid chro- 96-well collection plate. The extracts in the 96-well
matography principles to control selectivity, SPE plate are evaporated ofine and then the dried res-
This copy is for personal use only - distribution prohibited.

provides sample clean-up, recovery, and concen- idues in the 96-well plate are dissolved by adding
tration necessary for accurate quantitative anal- the reconstitution solution using the robotic sys-
ysis. The SPE procedure typically follows a pro- tem. The 96-well plate is sealed with a cover mat
tein precipitation step, similar to what is done and placed on the autosampler, ready for injec-
for on-line clean-up procedures. The superna- tion [45]. The Tomtec Quadra 96 station (Tomtec,

SE
tant is applied to a column previously equilibrat- Hamden, CT, USA) can be used for automation of
ed with methanol and then water. The most wide- LLE as described in the published method for si-
ly used columns are those with C18 as a matrix. multaneous analysis of CsA and everolimus [49].
The columns are washed sequentially with wa- The automation increases the sample through-
ter, 50% methanol in water, and heptane. Then put, shortens preparation time by approximately

U
compounds of interest are eluted from the ma- 3-fold when compared to manual LLE and gives
trix with a mixture of isopropyl alcohol and hep- the analyst more time for other tasks. However it
tane and dried under a stream of air. The dried has been found that there is no appreciable dif-
extract residue is usually dissolved in a small vol- ference in extraction efciency between manual

LY L
ume (50 L) of methanol [27] or of mobile phase and automated LLE [50].
This copy is for personal use only - distribution prohibited.

[26] and injected onto the column. Injection vol-


N A
ume ranged from 5 to 25 L [19,26,27,30,46]. When comparing these three methods present-
This whole procedure is more time, effort and ed here for sample preparation, protein precip-
reagent consuming compared to the on-line ex- itation with on line cleaning, SPE and LLE, the
O N
traction procedure. However, studies have shown latter provides cleaner extracts than SPE, and
that samples obtained from SPE columns are less SPE provides extracts with less matrix effect com-
likely to give rise to matrix effects [37]. pared to protein precipitation extract. However,
SO

the main attraction to the protein precipitation


Automated versions of the SPE procedure are with on-line cleaning technique, compared with
based on the use of the 96-well format for ex- manual LLE or SPE, is its speed, simplicity and
traction together with a robotic liquid handling universality in that the same basic procedure can
system. This approach automates most of the te- be applied to extract almost any analyte [45].
R

dious steps encountered in the traditional man-


ual extraction procedure and eliminates some of INTERNAL STANDARD SELECTION
the time consuming steps like tube labeling and
This copy is for personal use only - distribution prohibited.

PE

most capping/uncapping steps offering a desir- Selection of an appropriate internal standard


able short analysis time, which is important for is critical to the success of a quantitative meth-
high throughput in clinical applications [47]. od. Internal standards are used in high-per-
formance liquid chromatography mass spec-
Liquid-liquid extraction trometry methods to compensate for sample
to sample differences not only in extraction ef-
Liquid-liquid extraction (LLE) is an attractive al- ciency during sample preparation, as for clas-
ternative to SPE because of its relative simplicity sic HPLC with UV detection, but also for ioniza-
and ease of development. The extraction is per- tion variability during the transfer of the analyte
formed mainly with tert-butyl ether or 1-chlorobu- from the liquid phase to the gas phase and ma-
tane. LLE can be performed manually [48] or this trix effect [51]. Stable isotope labeled internal
process can be fully or semi-automated [49]. standards should be the rst choice of internal
standards for the measurement of immunosup-
In a typical automated 96-well LLE, biological sam- pressive drugs. An isotopically labeled internal
py is for personal use only - distribution prohibited.

ples are transferred, using a robotic liquid handling standard will have a similar extraction recovery,
system, into separate wells of a 96-well plate con- ionization response in ESI mass spectrometry,
sisting of a racked collection of microtubes. Then and a similar chromatographic retention time.
the internal standard solution and the extracting It will be affected by matrix effect in the same
organic solvent are added using the robotic sys- way like analyte of interest. Unfortunately, such
tem. The microtubes are capped with strip caps chemicals are not readily available for all immu-
and then the plate is shaken ofine. The aqueous nosuppressive agents and thus structural ana-

65

Electronic PDF security powered by ISL-science.com


This copy is for personal use only - distrib

Review Paper Ann Transplant, 2009; 14(2): 61-72

Table 1. Compounds used as internal standards for measurement


of immunosuppressive agent in published HPLC-MS
methods.

Drug Internal standard


This copy is for personal use only - distribution prohibited.

CsA CsD, CsA-d12


Tacrolimus Ascomycin, FR 900520
Sirolimus SIR-d3, Ascomycin, 32-O-desmethoxy-rapamycin,

SE
Everolimus Ascomycin, SDZ RAD 223-756
Mycophenolate Carboxybutoxy ether, indomethacine
mofetil
Figure 2. Amount of internal standard used in any given analysis
should be in the 1/3 lower end of calibration curve [52].

U
logues or another related molecule that close-
ly matches the chromatographic and ionization the CsD [55] and ascomycin [42] mass transition
characteristics of the analyte are preferred. The that come from patient samples, CsA metabolites
de-methylated (14) or hydroxylated (+16) ana- or CsD that increase falsely area under the inter-

LY L
logs should not be used as internal standards nal standard peak (ascomycin or CsD) leading to
This copy is for personal use only - distribution prohibited.

since these are the most common mass shifts ob- lower RAD or CsA results, respectively.
N A
served in naturally occurring metabolites of the
parent compound [52].The most widely used Another paper published in 2008 has shown that
internal standards for quantication of immu- the use of SIR-d3 as the internal standard in the
O N
nosuppressive drugs are presented in Table 1. assay for quantitation of SIR, provides results with
Deuterated everolimus can be obtained from lower inter-patient variation compared with the
Novartis Pharma but not for all scientist [53] results obtained from the assay that used 32-O-
SO

and as of today there is no report that this IS desmethoxyrapamycin [56].


was used for any method.
It was also reported that choice of appropriate
The amount of added internal standard should solvent can affect performance of internal stand-
be well above the limit of quantication but not ard and for example methanolic solutions of as-
R

so high as to suppress the ionization of the an- comycin should be avoided [57]. Another inter-
alyte. A good rule of thumb is to target the in- esting observation is that ascomycin degraded
ternal standard to the lower 1/3 of the working quickly in the poor quality acetonitrile [57]. Nine
This copy is for personal use only - distribution prohibited.

PE

standard curve [52] (Figure 2). commercial methanol used in HPLC eluting sol-
vents were evaluated for signal suppression of
For the method with simultaneous analysis of few sirolimus, tacrolimus, and MPA in MS/MS detec-
drugs one common or a few separate, one for each tion [58]. Product ion intensity was found to vary
immunosuppressant, internal standards can be by 10 folds among the methanol tested. Though
used. Ascomycin is the most popular internal stand- appropriate internal standards could compen-
ard for macrolides: TAC, SIR and RAD and CsD sate for the signal loss, performance of the as-
for cyclosporine A. The effectiveness of these two say (e.g., LLOQ) could be compromised. A rig-
internal standards was discussed in several papers orous assessment of an internal standard should
[42,51,54]. It was shown that the precision for CsA be a part of method validation.
quality control samples obtained from the assay that
used ascomycin as an internal standard is poorest CHROMATOGRAPHIC SEPARATION
when compared to the results obtained from the
assay that employed CsA-d12 or CsD as the internal The most commonly used type of stationary
py is for personal use only - distribution prohibited.

standard [54]. Also comparison of RAD results be- phase for all immunosuppressive drug analy-
tween the assays that used ascomycin or SDZ RAD ses is a reversed-phase C18 or C8 column, for ex-
223-756 internal standard and CsA results between ample Novopak-C18 [27] or Zorbax-C8 [3], with
the assays that used CsD or CsA-d12 internal stand- column heating that ranged from 35C to 75C.
ard showed the lower results for runs that used Column heating generally reduces peak broad-
ascomycin and CsD as internal standard [42,51]. ening and allows to use higher ow rate of mo-
Studies revealed the presence of interferences in bile phase by decreasing the pressure on the col-

66

Electronic PDF security powered by ISL-science.com


This copy is for personal use only - distrib

Ann Transplant, 2009; 14(2): 61-72 Korecka M et al Review of the newest HPLC methods with mass spectrometry

umn. Various analytical column dimensions are to such in-source transformation of drug conju-
used across the published method. The length gate metabolites to the respective target analytes.
ranged from 3.3 to 250 mm and the outside di- The specicity of LC/MS/MS analyses must be
ameter from 2 to 4.6 mm. A few reported meth- assessed critically whenever drug metabolites
ods employ only a short guard column (3 or 4 with a higher molecular mass than the target an-
This copy is for personal use only - distribution prohibited.

mm length) as an analytical one avoiding a chro- alytes must be assumed in patients samples. The
matographic separation step [25,33,37,39]. For in-source fragmentation of MPAG, AcMPAG and
this kind of method matrix effects must be as- MPAC to MPA can varied depend on the mobile
sessed very careful since chromatographic sepa- phase composition, instrument settings and de-
ration can be a key step for elimination of matrix sign and some mass spectrometer parameters like

SE
effects. The features of the extraction columns capillary or cone voltage or temperature in the
used for the methods with switching valve and 2 ion source [30,36]. Since MPA analysis has two
dimensional chromatography were already pre- requirements not needed for the others immuno-
sented in the section on Sample preparation. suppressive drugs (plasma as a matrix and neces-
sity of chromatographic separation) it is usually

U
The acidied mixture of methanol (usually more analyzed in a separate assay, not simultaneously
than 80%) and water with the addition of am- with the co administered immunosuppressants.
monium acetate [18,2530,39,4244,46,48,49], However one method that was published in 2007
and/or formic acid [1,44] or sodium/ammoni- [17] described the LC and MS conditions that can

LY L
um formate [3,17] is usually used as an eluting be used both for MPA and the macrolides and
This copy is for personal use only - distribution prohibited.

solvent. A few methods for MPA and CsA analysis CsA quantication without any changes. Since the
N A
introduced acetonitrile with or without methanol sample matrix for these two groups of drugs dif-
into the mobile phase [31,44,49,59]. The addi- fers there is a requirement to prepare two differ-
tives in mobile phase (for example ammonium ent samples for analysis, but the LC/MS setting is
O N
acetate and/or formic acid) increase the ioniza- the same and assures full separation of MPA from
tion of analyzed compounds and improve meth- the metabolites and internal standard.
od sensitivity. Adduct formation together with mo-
SO

bile phase pH adjustment has been commonly MASS SPECTROMETRY DETECTION


used to optimize detection conditions [60]. The
ow of eluting solvent is usually isocratic, howev- Ionization process and matrix effect
er a few methods use gradient [3,25,37,39]. Flow
rate of mobile phase varied from 0.1 mL/min Ionization procedure refers to the two processes
R

[61] up to 0.8 mL/min [28,38,42] for the meth- leading to the formation of gas-phase ions sampled
od with ESI and 1 mL/min for the method with by the mass spectrometer: (1) transfer of the an-
APCI [33] managing the total run time in the alyte into the gas phase and (2) the addition of a
This copy is for personal use only - distribution prohibited.

PE

range from 2.5 [29,39] to 20 min [34]. charge to form an ion [63]. In the presence of oth-
er sample components the ionization process can
Mass spectrometry detection in the analysis of be changed and will be mostly observed as a loss in
macrolides and CsA allows to achieve the de- response and referred to as matrix effect that can
sire selectivity and sensitivity without chromato- inuence the accuracy, precision and lower lim-
graphic separation. All compounds can be elut- it of quantication of the assay. The matrix effect
ed with the same retention time with no effect on can occur with any biological matrix (plasma, se-
selectivity and sensitivity as long as no matrix ef- rum, blood or urine), and it may vary depending
fect is observed for any of the analytes involved. on the source of matrix [45]. It is also compound
However for MPA quantication adequate sepa- dependent [63] and it was shown that the meth-
ration on an analytical column is a crucial step of ods based on electrospray ionization, mainly used
the analysis. Two metabolites of MPA, MPA-glu- for quantication of immunosuppressive drugs, are
curonide (MPAG) and the acyl glucuronide of more likely to experience a matrix effect than that
MPA (AcMPAG) and the generally used internal based on heated nebulizer [6466]. Assessment of
py is for personal use only - distribution prohibited.

standard, carboxybutoxy ether of MPA (MPAC), matrix effect is a very important step during meth-
undergo in-source fragmentation and produce od development/validation and can not be omit-
additional peaks of the ammonium or protonat- ted. There are two commonly used tests for evalu-
ed adduct ions of MPA and can falsely increase ation of the matrix effect: post-extraction addition
the area under the MPA peak when chromato- and post-column infusion, and they have been de-
graphic separation is omitted [18,31,62]. In one scribed in numerous of publications [64,66,67].
paper published in 2000 [45] attention was drawn The probability of the occurrence of the matrix

67

Electronic PDF security powered by ISL-science.com


This copy is for personal use only - distrib

Review Paper Ann Transplant, 2009; 14(2): 61-72

Table 2. MRM transitions (quantifier and qualifier pairs of ions) dem mass spectrometry for detection of ions.
employed for tandem mass spectrometry detection of Single quadrupole mass spectrometry detects
immunosuppressive drugs. mainly sodium adducts which gives the strong-
est signal when compared with proton or potassi-
MRM-transition MRM-transition um adduct ions [3]. However only one published
This copy is for personal use only - distribution prohibited.

quantifier qualifier method with single quadrupole mass spectrom-


Sirolimus [M+NH4]+ 931/864 931/882 etry detection uses an additive in the mobile
phase (sodium formate) [3], the rest of them
Tacrolimus [M+NH4]+ 821/768 975/908 do not use any additives [32,34,41] and forma-
Everolimus [M+NH4] +
975/908 975/858 tion of sodium adduct is connected with com-

SE
mon presence of sodium ions in water. The pos-
MPA [M+NH4]+ 338/207 Not reported itive mode is generally more sensitive than the
MPA [M+H]+ 321/207 Not reported negative mode for single quadrupole mass spec-

trometry detection [32].
MPA [M-H] 318/191 Not reported

U
CsA [M+NH4]+ 1219/1203 1219/1185 The majority of methods that use tandem mass
spectrometry detection operate in the positive
CsA [M+H]+ 1202/425 Not reported
mode while detecting primarily ammonium ad-
duct [18,2729,33,3739,48] and its fragment.

LY L
effect can be reduced or eliminated by utilizing a The ammonium adduct is obtained by addition
This copy is for personal use only - distribution prohibited.

more thorough clean-up procedure of the biolog- of ammonium acetate or ammonium formate
N A
ical samples, modifying the chromatographic con- to the mobile phase. It was shown that ammoni-
ditions, or allowing longer run times in order to um adduct ions are less stable than sodium ad-
enhance the chromatographic separation between ducts, so more appropriate for fragmentation.
O N
the matrix-effect causing component and the an- Additionally in positive ionisation mode proton
alyte [45]. The choice of internal standard is also adduct ions and in negative ionisation mode de-
very important for minimizing the inuence of ma- protonated ions are used for detection.
SO

trix effect on accuracy of the assay. An isotopically


labeled compound used as internal standard will The precursor ions give usually more than one
be affected by any matrix effect in the same way product. The most abundant product ion is cho-
like analyzed drug so accuracy of the quantitation sen as a quantier ion, and one from the less
should not be impaired provided that adequate sig- abundant ions can be used as a qualier ion.
R

nal from the drug and internal standard occurs so In order to obtain an increase in compound re-
that signal to noise ratio is adequate for both. sponse, the peak area for these two mass transi-
tions can be monitored and summed [27,29].
This copy is for personal use only - distribution prohibited.

PE

Detection The most widely used quantier and qualier


transitions are listed in Table 2.
The analytical methods used for quantitation of
immunosuppressive agents utilize single or tan-

Table 3. Characteristics of HPLC methods with single and tandem mass spectrometry detection for quantification of immunosuppressive
drugs. There is no recently reported method for MPA analysis by HPLC with single mass spectrometry detection.

MPA g/L CsA ng/mL Sirolimus ng/mL Tacrolimus ng/mL Everolimus ng/mL
MS/MS MS MS/MS MS MS/MS MS MS/MS MS MS/MS
Free: 0.5-2.5
LLOQ 1.525.0 5.24-10 0.251.0 0.21.66 0.21.0 0.2-1.66 0.15-0.25 0.252.5
Total: 50-100
Free: 2.5-8.8
Precision* (%) 0.9-10.2 2.5-7.3 1.29.8 0.96-10.8 0.75.8 0.7515 0.99.1 1.9-14.0
py is for personal use only - distribution prohibited.

Total: 1.8-8.5
Free: 95109
Accuracy** (%) 82105 91.3-119 96108 92105 79109 92104 84107 90109
Total: 91110
18, 19, 3, 32, 34, 3, 32, 25, 26, 28, 29, 3, 34, 41, 2729, 33, 28, 29, 42, 43,
References 20, 49 3, 34, 41
30, 44 41, 59, 71 34, 59 33, 37, 39, 48 59, 72 39, 40 46, 49, 73
* Together in- and between day reported for quality control samples; ** reported for quality control samples.

68

Electronic PDF security powered by ISL-science.com


This copy is for personal use only - distrib

Ann Transplant, 2009; 14(2): 61-72 Korecka M et al Review of the newest HPLC methods with mass spectrometry

METHODS CHARACTERISTICS 3. Christians U, Jacobsen W, Serkova N et al:


Automated, fast and sensitive quantication of
The lower limit of quantication (LLOQ), preci- drugs in blood by liquid chromatography-mass
sion and accuracy for all methods that utilize single spectrometry with on-line extraction: immuno-
or tandem mass spectrometry are shown in Table 3. suppressants. J Chromatogr B Biomed Sci Appl,
2000; 748: 4153
This copy is for personal use only - distribution prohibited.

The lowest standard concentration for all present-


ed methods, which dened LLOQ, is lower than 4. Ransom JT: Mechanism of action of mycopheno-
the lowest concentration of each drug expected in late mofetil. Ther Drug Monit, 1995; 17: 68184
treated patients. It is often the practice that LLOQ 5. Stepkowski SM, Tian L, Napoli KL et al: Synergistic
is not investigated at concentrations lower than the mechanisms by which sirolimus and cyclosporin in-

SE
lowest concentration of the observed linear range hibit rat heart and kidney allograft rejection. Clin
Exp Immunol, 1997; 108: 6368
[39]. Based on the recommendations of Shah et
al. [68] and the FDA guideline for analytical per- 6. Ansermot N, Fathi M, Veuthey JL et al: Simultaneous
formance validation [69] the published assays for quantication of cyclosporine, tacrolimus, sirolimus
and everolimus in whole blood by liquid chroma-
analysis of immunosuppressive drugs have excellent

U
tography-electrospray mass spectrometry. Clin
between- and within-day analytical recovery and pre- Biochem, 2008; 41: 72835
cision and can be used for TDM test methods.
7. Yatscoff RW, Wang P, Chan K et al: Rapamycin: dis-
tribution, pharmacokinetics, and therapeutic range
CONCLUSIONS

LY L
investigations. Ther Drug Monit, 1995; 17: 66671
This copy is for personal use only - distribution prohibited.

8. Nagase K, Iwasaki K, Nozaki K, Noda K: Distribution


Liquid chromatography with mass spectrometry and protein binding of FK506, a potent immuno-
N A
detection is a major breakthrough in therapeutic suppressive macrolide lactone, in human blood
drug monitoring of immunosuppressive agents and and its uptake by erythrocytes. J Pharm Pharmacol,
is considered as the method of choice in TDM of 1994; 46: 11317
O N
immunosuppressants. It provides high specicity, 9. Burton MES, Schentag JJ, Evans WE: Applied
excellent sensitivity and robustness for drug meas- Pharmacokinetics & Pharmacodynamics
urement however the choice of internal standard,
SO

Principles of therapeutic drug monitoring. In:


the organic solvents and matrix effect should be Wilkins LW, ed, 2004
evaluated carefully during method validation. It 10. Holt DW, Armstrong VW, Griesmacher A et al:
has a potential for simultaneous determination of International Federation of Clinical Chemistry/
4 immunosuppressants in blood and additionally International Association of Therapeutic Drug
allows analysis of MPA in plasma using the same Monitoring and Clinical Toxicology working group
R

conditions. Despite the initial high cost for the in- on immunosuppressive drug monitoring. Ther
strumentation, HPLC-MS is more cost effective than Drug Monit, 2002; 24: 5967
microparticle enzyme immunoassay, as was shown 11. Kahan BD, Napoli KL, Kelly PA et al: Therapeutic
This copy is for personal use only - distribution prohibited.

PE

for TAC [70]. The reagent cost for immunoassay is drug monitoring of sirolimus: correlations with efca-
signicantly higher compared with the basic rea- cy and toxicity. Clin Transplant, 2000; 14: 97109
gents and solvents used for HPLC-MS. Since prices 12. Siuzdak G: The emergence of mass spectrometry
for mass spectrometers are decreasing more cent- in biochemical research. Proc Natl Acad Sci USA,
ers should consider purchase of this equipment 1994; 91: 1129097
thereby saving money on reagents, potentially im- 13. Willoughby S, Mitrovich: A global view of LC/MS:
proving the quality of results and achieving an in- Global View Publishing
crease in the number of reported results. 14. Want EJ, Cravatt BF, Siuzdak G: The expanding
role of mass spectrometry in metabolite prol-
REFERENCES: ing and characterization. Chembiochem, 2005;
6: 194151
1. Ansermot N, Fathi M, Veuthey JL et al: 15. Wilson D, Johnston F, Holt D et al: Multi-center
Quantication of cyclosporine A in peripheral evaluation of analytical performance of the micro-
blood mononuclear cells by liquid chromatogra- particle enzyme immunoassay for sirolimus. Clin
phy-electrospray mass spectrometry using a col- Biochem, 2006; 39: 37886
py is for personal use only - distribution prohibited.

umn-switching approach. J Chromatogr B Analyt 16. Napoli KL, Kahan BD: Sample clean-up and high-
Technol Biomed Life Sci, 2007; 857: 9299 performance liquid chromatographic techniques
2. Shaw LM, Kaplan B, Brayman KL: Advances in for measurement of whole blood rapamycin con-
therapeutic drug monitoring for immunosuppres- centrations. J Chromatogr B Biomed Appl, 1994;
sants: a review of sirolimus. Introduction and over- 654: 11120
view. Clin Ther, 2000; 22(Suppl.B): B113

69

Electronic PDF security powered by ISL-science.com


This copy is for personal use only - distrib

Review Paper Ann Transplant, 2009; 14(2): 61-72

17. Bogusz MJ, Enazi EA, Hassan H et al: Simultaneous 29. Koal T, Deters M, Casetta B, Kaever V: Simultaneous
LC-MS-MS determination of cyclosporine A, tac- determination of four immunosuppressants by
rolimus, and sirolimus in whole blood as well as means of high speed and robust on-line solid phase
mycophenolic acid in plasma using common pre- extraction-high performance liquid chromatography-
treatment procedure. J Chromatogr B Analyt tandem mass spectrometry. J Chromatogr B Analyt
Technol Biomed Life Sci, 2007; 850: 47180 Technol Biomed Life Sci, 2004; 805: 21522
This copy is for personal use only - distribution prohibited.

18. Streit F, Shipkova M, Armstrong VW, Oellerich 30. Annesley TM, Clayton LT: Quantication of myc-
M: Validation of a rapid and sensitive liquid chro- ophenolic acid and glucuronide metabolite in hu-
matography-tandem mass spectrometry method man serum by HPLC-tandem mass spectrometry.
for free and total mycophenolic acid. Clin Chem, Clin Chem, 2005; 51: 87277

SE
2004; 50: 15259 31. Vogeser M, Zachoval R, Spohrer U, Jacob K:
19. Willis C, Taylor PJ, Salm P et al: Quantication of Potential lack of specicity using electrospray
free mycophenolic acid by high-performance liquid tandem-mass spectrometry for the analysis of myc-
chromatography-atmospheric pressure chemical ophenolic acid in serum. Ther Drug Monit, 2001;
ionisation tandem mass spectrometry. J Chromatogr 23: 72224
B Biomed Sci Appl, 2000; 748: 15156

U
32. Kirchner GI, Vidal C, Jacobsen W et al: Simultaneous
20. Keevil BG, Tierney DP, Cooper DP et al: on-line extraction and analysis of sirolimus (rapamy-
Simultaneous and rapid analysis of cyclosporin A cin) and ciclosporin in blood by liquid chromatog-
and creatinine in nger prick blood samples using raphy-electrospray mass spectrometry. J Chromatogr
liquid chromatography tandem mass spectrometry B Biomed Sci Appl, 1999; 721: 28594

LY L
and its application in C2 monitoring. Ther Drug 33. Volosov A, Napoli KL, Soldin SJ: Simultaneous sim-
This copy is for personal use only - distribution prohibited.

Monit, 2002; 24: 75767 ple and fast quantication of three major immu-
N A
21. Hoogtanders K, van der Heijden J, Christiaans M nosuppressants by liquid chromatography tan-
et al: Therapeutic drug monitoring of tacrolimus dem mass-spectrometry. Clin Biochem, 2001; 34:
with the dried blood spot method. J Pharm Biomed 28590
O N
Anal, 2007; 44: 65864 34. Deters M, Kirchner G, Resch K, Kaever V:
22. Hoogtanders K, van der Heijden J, Christiaans M et Simultaneous quantification of sirolimus,
al: Dried blood spot measurement of tacrolimus is everolimus, tacrolimus and cyclosporine by liq-
SO

promising for patient monitoring. Transplantation, uid chromatography-mass spectrometry (LC-MS).


2007; 83: 23738 Clin Chem Lab Med, 2002; 40: 28592
23. Capron A, Lerut J, Verbaandert C et al: Validation of 35. Ceglarek U, Lembcke J, Fiedler GM et al: Rapid si-
a liquid chromatography-mass spectrometric assay multaneous quantication of immunosuppressants
for tacrolimus in liver biopsies after hepatic trans- in transplant patients by turbulent ow chroma-
R

plantation: correlation with histopathologic staging tography combined with tandem mass spectrom-
of rejection. Ther Drug Monit, 2007; 29: 34048 etry. Clin Chim Acta, 2004; 346: 18190
24. Mendonza AE, Gohh RY, Akhlaghi F: Analysis of 36. Vogeser M, Fleischer C, Meiser B et al: Quantication
This copy is for personal use only - distribution prohibited.

PE

mycophenolic acid in saliva using liquid chroma- of sirolimus by liquid chromatography-tandem mass
tography tandem mass spectrometry. Ther Drug spectrometry using on-line solid-phase extraction.
Monit, 2006; 28: 4026 Clin Chem Lab Med, 2002; 40: 4045
25. Wallemacq PE, Vanbinst R, Asta S, Cooper DP: 37. Annesley TM, Clayton L: Simple extraction pro-
High-throughput liquid chromatography-tandem tocol for analysis of immunosuppressant drugs in
mass spectrometric analysis of sirolimus in whole whole blood. Clin Chem, 2004; 50: 184548
blood. Clin Chem Lab Med, 2003; 41: 92125 38. Vogeser M, Spohrer U: Automated processing of
26. Taylor PJ, Johnson AG: Quantitative analysis of whole blood samples for the determination of im-
sirolimus (Rapamycin) in blood by high-perform- munosuppressants by liquid chromatography tan-
ance liquid chromatography-electrospray tandem dem-mass spectrometry. Clin Chem Lab Med, 2006;
mass spectrometry. J Chromatogr B Biomed Sci 44: 112630
Appl, 1998; 718: 25157 39. Wang S, Magill JE, Vicente FB: A fast and simple
27. Taylor PJ, Salm P, Lynch SV, Pillans PI: Simultaneous high-performance liquid chromatography/mass
quantication of tacrolimus and sirolimus, in hu- spectrometry method for simultaneous measure-
man blood, by high-performance liquid chroma- ment of whole blood tacrolimus and sirolimus.
py is for personal use only - distribution prohibited.

tography-tandem mass spectrometry. Ther Drug Arch Pathol Lab Med, 2005; 129: 66165
Monit, 2000; 22: 60812 40. Salm P, Taylor PJ, Clark A et al: High-performance
28. Streit F, Armstrong VW, Oellerich M: Rapid liquid liquid chromatography-tandem mass spectrometry
chromatography-tandem mass spectrometry rou- as a reference for analysis of tacrolimus to assess
tine method for simultaneous determination of two immunoassays in patients with liver and renal
sirolimus, everolimus, tacrolimus, and cyclosporin transplants. Ther Drug Monit, 1997; 19: 694700
A in whole blood. Clin Chem, 2002; 48: 95558

70

Electronic PDF security powered by ISL-science.com


This copy is for personal use only - distrib

Ann Transplant, 2009; 14(2): 61-72 Korecka M et al Review of the newest HPLC methods with mass spectrometry

41. Vidal C, Kirchner GI, Wunsch G, Sewing KF: 53. Holt D: Deuterated internal standards donated
Automated simultaneous quantication of the im- by Novartis. Available from: http://www.iatdmct.
munosuppressants 40-O-(2-hydroxyethyl) rapamy- org/index.php/publisher/articleview/action/view/
cin and cyclosporine in blood with electrospray- frmArticleID/58/, 2008
mass spectrometric detection. Clin Chem, 1998; 54. Taylor PJ, Brown SR, Cooper DP et al: Evaluation
44: 127582
This copy is for personal use only - distribution prohibited.

of 3 internal standards for the measurement of cy-


42. Korecka M, Solari SG, Shaw LM: Sensitive, high closporin by HPLC-mass spectrometry. Clin Chem,
throughput HPLC-MS/MS method with on-line 2005; 51: 189093
sample clean-up for everolimus measurement. 55. Taylor PJ, Brown SR, Cooper DP et al: Data sup-
Ther Drug Monit, 2006; 28: 48490 plemented to on line version of article: Evaluation

SE
43. Taylor PJ, Franklin ME, Graham KS, Pillans PI: of 3 internal standards for the measurement of cy-
A HPLC-mass spectrometric method suitable for closporin by HPLC-mass spectrometry. Available
the therapeutic drug monitoring of everolimus. J from: http://www.clinchem.org/content/vol51/issue10/,
Chromatogr B Analyt Technol Biomed Life Sci, 2005
2007; 848: 20814 56. OHalloran S, Ilett KF: Evaluation of a deuterium-

U
44. Brandhorst G, Streit F, Goetze S et al: Quantication labeled internal standard for the measurement of
by liquid chromatography tandem mass spectrom- sirolimus by high-throughput HPLC electrospray
etry of mycophenolic acid and its phenol and acyl ionization tandem mass spectrometry. Clin Chem,
glucuronide metabolites. Clin Chem, 2006; 52: 2008; 54: 138689
196264

LY L
57. Napoli KL: Organic solvents compromise perform-
45. Jemal M: High-throughput quantitative bioanaly- ance of internal standard (ascomycin) in pro-
This copy is for personal use only - distribution prohibited.

sis by LC/MS/MS. Biomed Chromatogr, 2000; 14: ciency testing of mass spectrometry-based assays
42229
N A for tacrolimus. Clin Chem, 2006; 52: 76566
46. Salm P, Taylor PJ, Lynch SV, Pillans PI: 58. Annesley TM: Methanol-associated matrix effects
Quantication and stability of everolimus (SDZ in electrospray ionization tandem mass spectrom-
O N
RAD) in human blood by high-performance liquid etry. Clin Chem, 2007; 53: 182734
chromatography-electrospray tandem mass spec- 59. Poquette MA, Lensmeyer GL, Doran TC: Effective
trometry. J Chromatogr B Analyt Technol Biomed
SO

use of liquid chromatography-mass spectrometry


Life Sci, 2002; 772: 28390 (LC/MS) in the routine clinical laboratory for
47. Harrison AC, Walker DK: Automated 96-well sol- monitoring sirolimus, tacrolimus, and cyclosporine.
id phase extraction for the determination of do- Ther Drug Monit, 2005; 27: 14450
ramectin in cattle plasma. J Pharm Biomed Anal, 60. Gao S, Zhang ZP, Karnes HT: Sensitivity enhancement
1998; 16: 77783 in liquid chromatography/atmospheric pressure ion-
R

48. Holt DW, Lee T, Jones K, Johnston A: Validation ization mass spectrometry using derivatization and
of an assay for routine monitoring of sirolimus us- mobile phase additives. J Chromatogr B Analyt
ing HPLC with mass spectrometric detection. Clin Technol Biomed Life Sci, 2005; 825: 98110
This copy is for personal use only - distribution prohibited.

PE

Chem, 2000; 46: 117983 61. Taylor PJ, Jones A, Balderson GA et al: Sensitive,
49. Brignol N, McMahon LM, Luo S, Tse FL: High- specic quantitative analysis of tacrolimus (FK506)
throughput semi-automated 96-well liquid/liq- in blood by liquid chromatography-electrospray
uid extraction and liquid chromatography/mass tandem mass spectrometry. Clin Chem, 1996; 42:
spectrometric analysis of everolimus (RAD 001) 27985
and cyclosporin a (CsA) in whole blood. Rapid 62. Korecka M, Nikolic D, van Breemen RB, Shaw LM:
Commun Mass Spectrom, 2001; 15: 898907 The apparent inhibition of inosine monophos-
50. Jemal M, Teitz D, Ouyang Z, Khan S: Comparison phate dehydrogenase by mycophenolic acid glu-
of plasma sample purication by manual liquid- curonide is attributable to the presence of trace
liquid extraction, automated 96-well liquid-liquid quantities of mycophenolic acid. Clin Chem, 1999;
extraction and automated 96-well solid-phase ex- 45: 104750
traction for analysis by high-performance liquid 63. Bonglio R, King RC, Olah TV, Merkle K: The ef-
chromatography with tandem mass spectrome- fects of sample preparation methods on the vari-
try. J Chromatogr B Biomed Sci Appl, 1999; 732: ability of the electrospray ionization response for
5018 model drug compounds. Rapid Commun Mass
py is for personal use only - distribution prohibited.

51. Taylor PJ: Internal standard selection for immu- Spectrom, 1999; 13: 117585
nosuppressant drugs measured by high-perform- 64. Matuszewski BK, Constanzer ML, Chavez-Eng CM:
ance liquid chromatography tandem mass spec- Matrix effect in quantitative LC/MS/MS analyses
trometry. Ther Drug Monit, 2007; 29: 13132 of biological uids: a method for determination
52. The Internal Standard. Available from: http://www. of nasteride in human plasma at picogram per
ionsource.com/tutorial/msquan/is.htm milliliter concentrations. Anal Chem, 1998; 70:
88289

71

Electronic PDF security powered by ISL-science.com


This copy is for personal use only - distrib

Review Paper Ann Transplant, 2009; 14(2): 61-72

65. King R, Bonglio R, Fernandez-Metzler C et al: 70. Lensmeyer GL, Poquette MA: Therapeutic moni-
Mechanistic investigation of ionization suppres- toring of tacrolimus concentrations in blood: semi-
sion in electrospray ionization. J Am Soc Mass automated extraction and liquid chromatography-
Spectrom, 2000; 11: 94250 electrospray ionization mass spectrometry. Ther
66. Annesley TM: Ion suppression in mass spectrom- Drug Monit, 2001; 23: 23949
This copy is for personal use only - distribution prohibited.

etry. Clin Chem, 2003; 49: 104144 71. Zhou L, Tan D, Theng J et al: Optimized analyt-
67. Taylor PJ: Matrix effects: the Achilles heel of quan- ical method for cyclosporin A by high-perform-
titative high-performance liquid chromatogra- ance liquid chromatography-electrospray ioniza-
phy-electrospray-tandem mass spectrometry. Clin tion mass spectrometry. J Chromatogr B Biomed
Biochem, 2005; 38: 32834 Sci Appl, 2001; 754: 2017

SE
68. Shah VP, Midha KK, Findlay JW et al: Bioanalytical 72. Alak AM: Measurement of tacrolimus (FK506) and
method validation a revisit with a decade of its metabolites: a review of assay development and
progress. Pharm Res, 2000; 17: 155157 application in therapeutic drug monitoring and
pharmacokinetic studies. Ther Drug Monit, 1997;
69. FDA. Guidance for industry-bioanalytical method
19: 33851
validation. Available from: http://www.fda.gov/cder/

U
guidance/index.htm, 2001 73. Khoschsorur G: Simultaneous measurement of
sirolimus and everolimus in whole blood by HPLC
with ultraviolet detection. Clin Chem, 2005; 51:
172124

LY L
This copy is for personal use only - distribution prohibited.

N A
O N
SO
R
This copy is for personal use only - distribution prohibited.

PE
py is for personal use only - distribution prohibited.

72

Electronic PDF security powered by ISL-science.com

Anda mungkin juga menyukai