Use a CT dataset of the pelvis. Create a CTV by contouring the rectum (start
at the anus and stop at the turn where it meets the sigmoid colon). Expand
this structure by 1 cm and label it PTV.
Create a PA field with the top border at the bottom of L5 and the bottom
border 2 cm below the PTV. The lateral borders of the PA field should extend
1-2 cm beyond the pelvic inlet to include primary surrounding lymph nodes.
Place the beam isocenter in the center of the PTV and use the lowest beam
energy available (note: calculation point will be at isocenter).
Contour all critical structures (organs at risk) in the treatment area. List all
organs at risk (OR) and desired objectives/dose limitations, in the table
below:
Figure 1-1. Hot spot locations and DVH comparing CTV and PTV
isodose lines.
b. Change to a higher energy and calculate the beam. How did your
isodose distribution change?
The penetration of the PA beam increases which shifted the isodose
lines further anterior. This also brought the overall maximum hotspot
dose from 67.3 Gy to 61.3 Gy. This reflects the skin sparing properties
of a higher beam energy. Compare with Figure 1-2 below.
Figure 1-3
d. Change the 2 lateral fields to a higher energy and calculate. How did
this change the dose distribution? The hotspots were eliminated from
the sides of the pelvis anatomy. The overall hotspot was also reduced.
See Figure 1-4 below:
Figure 2-2. Top to bottom, left to right: Axial plane, PA field BEV,
coronal, and sagittal planes displaying dose distributions covering CTV
and PTV volumes.
Figure 2-3. DVH with CTV, PTV, and critical structures.
4 field pelvis
Using the final 3 field rectum plan, copy and oppose the PA field to create an
AP field. Keep the lateral field arrangement. Remove any wedges that may
have been used. Calculate the four fields and weight them equally. How does
this change the isodose distribution? What do you see as possible
advantages or potential disadvantages of adding the fourth field? The
isodose distribution was relatively similar considering there was no beam
weighting or wedge compensation to optimize the coverage. The four field
plans dose distribution did present very evenly and coverage was almost
identical but on comparing the DVHs for both plans, higher bowel and
bladder dose were given with the four field plan. The 3 field plan also had a
steeper dose fall off in comparison. See Figure 3-1 below:
Figure 3-1. DVH lines tracked with squares reflect 3 field plan, lines tracked
with triangles were 4 field plan.