YOUER NAME
Experiment #10:
Determination of the Gas Law
Constant
Laboratory Partners:
Objectives: The sole learning objective of this laboratory experiment is to determine the value of
Materials:
a. Chemicals: magnesium metal (Mg), magnesium ribbon (Mg), 0.1M hydrochloric acid
tweezers, analytical balance, boiling tube, pipet, plunger of the syringe, rubber stopper,
Methods:
1. The equipment shown in the laboratory manual under the experimental procedure was set
up. Care was exercised to avoid breaking the glass syringe. The boiling tube and the
to the reaction. The coating was removed by dipping the magnesium ribbon in a 50 mL
beaker containing approximately 20 mL of 0.1M hydrochloric acid, HCl. The ribbon was
held with tweezers. The tweezers never came in contact with the acid. Once the coating
was removed, the ribbon was thoroughly rinsed with distilled water and then dried with a
paper towel.
3. Between 0.016-0.018 g of the cleaned magnesium ribbon was weighed to four (4)
decimal places on the analytical balance and its mass was recorded on the data sheet. This
was done because using more that 0.018 g of Mg would have resulted in more gas being
tube. It was noted that the concentration is different than that used in step 2. Splashing the
acid on the sides of the tube was avoided. Doing this made introducing the Mg into the
acid easier.
5. The boiling tube containing the acid was tilted and the weighed magnesium ribbon was
placed on the upper inside wall of the tube. The Mg ribbon did come in contact with the
acid at this point. With the tube still tilted, the plunger of the syringe was pushed all the
way in and the rubber stopper was then inserted into the tube. The reading was recorded
solution. The side of the tube was tapped if the ribbon stuck. The tube was clamped into
place on the ring stand and the reaction was allowed to proceed until no further bubbles
were seen and all the Mg ribbon had disappeared. The plunger was then carefully rotated
to ensure it was not sticking; the plunger was not pushed. After the reaction was
completed, the apparatus was allowed to sit for 5 minutes to allow the gas to equilibrate.
Then, the final volume of the gas in the syringe was recorded on the data sheet.
7. Using the barometer on the instructors table, the atmospheric pressure in mmHg was
measured. Also, the room temperature was measured using a thermometer. Both values
obtained. It was not necessary to clean the boiling tube between determinations, but a dry
Reading, mL
Final Syringe 15.8 17.4 17.0 16.73
Reading, mL
Volume H2 15.8 17.4 17.0 16.73
collected, mL
(Final reading
initial reading)
Volume H2 0.0158 0.0174 0.0170 0.01673
collected, L
Temperature C 20 20 20 20
Temperature K 293.15 293.15. 293.15 293.15
Barometric 788 788 788 788
Pressure, mmHg
Vapor pressure of 18.7 18.7 18.7 18.7
H2O, mmHg
Corrected 769.3 769.3 769.3 769.3
pressure, mmHg
Corrected 1.0122 1.0122 1.0122 1.0122
pressure, atm
Calculated R 6.96 x 10-6 6.32 x 10-6 6.47 x 10-6 6.58 x 10-6
value, L-
atm/mol-K
Average R value, 6.58 x 10-6 6.58 x 10-6 6.58 x 10-6 6.58 x 10-6
L-atm/mol-K
True value of R, 6.964 x 10-6 6.964 x 10-6 6.964 x 10-6 6.964 x 10-6
L-atm/mol-K
% error of 5.514 x 10-6 % 5.514 x 10-6 % 5.514 x 10-6 % 5.514 x 10-6 %
average R
Calculations:
3. Sample conversion of mL H2 to L H2
4. Sample calculation for corrected pressure and conversion of corrected pressure from
mmHg to atm.
R=(PV)/(nT)
5. Sample conversion of C to K.
1C = 273.15
20 + 273.15 = 293.15 K
6. Sample Calculation of R.
R=(PV)/(nT)
R = [1.0122(0.0158)]/[7.8381(293.15)] = 0.01599276/2,297.73902 = 0.00000696 L-
Discussion:
To understand the concept of the ideal gas equation and what the term ideal gas really
means and the behavior of gases, it is better to assume the gases are in an ideal state where it is
unaffected by conditions of the real world. The behavior of a gas must follow Kinetic-Molecular
Theory for it to be ideal. An equation was created to deal with solving gas problems known as
the Ideal Gas Equation. Even though it has been implanted in people to believe that nothing ideal
exists, two assumptions should have been made beforehand. These two assumptions are that the
particles have no forces acting among them and that these particles do not take up any space,
There are four gas variables and one constant in the equation that should be mentioned
before stating the Ideal Gas Equation. These include pressure (P), volume (V), number of mole
of gas (n), and temperature (T). Lastly, the constant in the equation shown is R, known as the gas
constant: PV = nRT. And another way to describe an ideal gas is to write it in mathematical
terms: (PV)/(nRT) = 1. When an ideal gas is plugged into the aforementioned equation, it will
always equal one and the more it deviates from one, the more it behaves like an ideal gas.
opposite side of the equation to n and T. Pressure is indirectly proportional to volume because P
is on the same side of the equation with V. An ideal gas is defined as one that obeys the ideal gas
equation. Real gases obey the equation reasonably well at lower pressures and higher
Ideal Gas Laws are Boyles Law, Charles Law, and Avogadros Law. Boyles Law
describes the inverse proportional relationship between pressure and volume at a constant
temperature and a fixed amount of gas. This law came from a manipulation of the Ideal Gas
Law: P1V1. (UC Davis). Charles's Law describes the directly proportional relationship
between the volume and temperature (in Kelvin) of a fixed amount of gas, when the pressure is
held constant: (V1/T1) = (V2/T2), (UC Davis). Lastly, Volume of a gas is directly proportional to
the amount of gas at a constant temperature and pressure: (V1/n1) = (V2n2), (UC Davis).
With the data recorded during the procedure of the laboratory experiment and upon the
completion of the experiment, the value of R was successfully determined using the equation R =
(PV)/(nT). This was the sample calculation for the first set of data: R=(PV)/(nT)
6.96 x 10-6 L-atm/mol-K. However, the average of all the calculated R values was 6.58 x 10-6 L-
atm/mol-K. This R value was fairly accurate considering that the percent error was calculated
using the following formula for percent data calculation: % error = [(|Accepted
Measured|)/Accepted] x 100. Therefore, substituting in all values using the data table included in
percent error was well below zero percent suggesting a fair amount of accuracy in the
However, precautions were taken for all the aforementioned possible errors. To avoid
having the magnesium ribbon slide into the tube to the HCl before the system was closed, the
tube was tilted enough for the magnesium ribbon to remain still until the system in closed. Then,
the tube was straightened and the ribbon was allowed to fall into the HCl. Next, to avoid having
excess coating, the MgO was cleaned in the 0.1M HCl for about five to seven seconds before
washing it with distilled water and dried with a dry paper towel. The other errors were avoided
by having more than one member calculate the values, weighing more than once, and adjusting
equipment if necessary and using other materials if the first ones obtained seemed incompetent
Conclusion:
There was one sole objective of this laboratory experiment and that was to determine the
value of the universal gas constant, R. As aforesaid, with the data recorded during the procedure
of the laboratory experiment and upon the completion of the experiment, the value of R was
successfully determined using the equation R = (PV)/(nT). This was the sample calculation for
6.96 x 10-6 L-atm/mol-K. However, the average of all the calculated R values was 6.58 x 10-6
L-atm/mol-K. This R value was fairly accurate considering that the percent error was calculated
using the following formula for percent data calculation: % error = [(|Accepted
Measured|)/Accepted] x 100. Therefore, substituting in all values using the data table included in
% error = [(|6.964 x 10-6 - 6.58 x 10-6|)/6.964 x 10-6] x 100 % error = 5.514 x 10-6 %. The
percent error was well below zero percent suggesting a fair amount of accuracy in the
References
Miller, Dionne. 2013. General Chemistry I SCC201 Lab Manual Revised Printing [Laboratory
UC Davis. "The Ideal Gas Law." UC Davis ChemWiki. University of California, Davis, n.d.
<http://chemwiki.ucdavis.edu/Physical_Chemistry/Physical_Properties_of_Matter/Phases
_of_Matter/Gases/The_Ideal_Gas_Law>.