Model: M2 - M3 - M4 - M5
(2)
RFID(1)reader LCD screen
User Multifunction
Keyboard
ok
Yellow LED
+ Enter -
Up Next Down Level 1 alarm
Visual: Flashing LED
Audible: Beeping
WSE134 10/2008
FPH.JML.RZ.CH
Level 2 alarm
Cancel Visual: Fixed LED
04/29/2009 05:01PM Audible: Beeping
Return
Access granted
(Simple user) Level 3 alarm
Visual: Flashing LED
Y o u c a n t u r n Mode
o nSecuret h e
Hood Audible: Continuous
Level 4 alarm
Visual: Fixed LED
Audible: Continuous
ok (3)
Red LED
A Global Solution For Green Buildings.
WSE130 09/2008
Authorizes access
ON
to the hood ON/OFF Lighting Ventilation ON/OFF Green LED
by RFID card
OFF
User A
Open Mode Function
User Operator User Card Fans OFF
This authorizes the holder to:
1 - Start and stop the fans (Ventilation ON/OFF key)
2 - Switch the lighting on and off (Lighting ON/OFF key)
3 - Deactivate the disengageable audible alarms if the right has bee granted by the
04/29/2009 05:01PM
administrator (Cancel key).
Access granted
(Simple user)
User X
B
04/29/2009 05:01PM 04/29/2009 05:01PM
User A
B
04/29/2009 05:01PM
(1)
RFID - Acronym (Radio Frequency Identification) referring to an identification Shut down in progress
Technology using radio frequency data communication. 00:00
(2)
LCD - Liquid Cristal Display Technologie where the image is produced by
liquid crystals excited by an electric current.
(3)
LED- Light-Emitting Diode.
Alarms
Causes and solutions
R
the alarm
Clearable
Level of
Solutions
Audible
Visual
alarm
alarm
alarm
Causes Alarm
Message *Authorized by the administrator
Keyboard used by an unauthorized user Fixed To use the hood,insert Place an authorized USER card
11A (or no authorized access card) 4 Red
Continuous
an authorizeduser card No in the RFID reader.
This alarm has priority even though it Flashing Reinsert card. Reinsert an authorized USER card
8D is level 3. Access card removed 3 Red
Continuous emergency stop in
00:XX
No in the RFID reader.
Fan out of service. Rotation speed less than 700rpm. Stop working
Faulty fan management electronic Fixed Fan X: out of order
8H Card or poor contact of the fan 4 Red
Continuous
stop working now No and
contact the Administrator
Connexion cable.
Acid alarm threshold permanently exceeded. Fixed Replace filters Contact the
8N Saturated filter modules. 4 Red
Continuous
(Sensor acid) No Administrator
Door open too wide: Opening greater than the Fixed Sensor acid Lower the door until the opening
8S configured alarm threshold. 2 Yellow
Beeping
out of order No complies with the threshold
*Press <ENTER> to
Fixed Check filters
8R Maintenance alarm 2 Yellow
Beeping
saturation YES mute current alarm sound
and call Administrator
The filter expiry date has passed. This date is Fixed Filters end of Partial or total replacement
8T saved when the filters are installed. 4 Red
Continuous
life(timer) No of the filter modules.
the established
the administrator
1 - Stop handling
According to
According to
Determined by
leve
Configuration does not comply with Fixed Hardware configuration Contact the
9A the factory configuration 4 Red
Continuous
Mismatch No Administrator
Anormal temperature in
40C temperature alarm threshold Flashing Check the heat sources.
8V exceeded in the working volume. 3 Red
Continuous working area.
call maintenance
No Contact the Administrator
Top priority alarm 60C Critical temperature. Stop your handling operation immediately do not
Fixed
8W temperature alarm threshold exceeded 3 Red
Continuous Fans stopped,
stop working now
No forget the heat sources. Completely close the
in the working volume. door. Call the Administrator
Faulty heat detection Flashing Sensor temperature Contact the
8V electronic card 1 Yellow
Beeping out of order No Administrator