Estetoscopio o estafiloscopio? Potencial vector


en las infecciones asociadas a la atencin de la salud
Andrs Ziga, Jaime Maalich y Rosario Corts

Universidad San Sebastin

Stethoscope or staphyloscope? Potential vector in nosocomial infections. Facultad de Medicina (AZ).


Instituto de Polticas Pblicas en
Salud IPSUSS (JM, RC).
Healthcare-associated infections (HCAI) are a problem worldwide. In our country, the estimated incidence of
HCAI is 70,000 per year. This results in an increase in the average length of hospital stay by 10 days per patient, Los autores declaran no tener
conflictos de inters.
an estimated annual cost of US $ 70 million and an overstay of 700 thousand bed days a year. For over 30 years
stethoscopes have been considered as potential HCAI vectors, since pathogens like methicillin-resistant Staphylo- Recibido: 22 de abril de 2015

coccus aureus and vancomycin-resistant Enterococcus strains adhere and colonize them. These organisms can be Aceptado: 26 de noviembre de
2015
transmitted between patients if the instruments are not sanitized. Several studies conclude that disinfecting the
stethoscope with isopropyl alcohol eliminates up to 99% of bacteria. Simple, economic measures such as imple- Correspondencia a:

mentation of guidelines for stethoscope disinfection are a clear opportunity for preventing infections. Jaime Maalich Muxi
jaime.manalich@uss.cl
Key words: Stethoscope, equipment contamination, Healthcare Associated Infections, Methicillin-Resistant
Staphylococcus aureus, decontamination, HCAI
Palabras clave: Estetoscopio, contaminacin, infeccin asociada a la atencin de salud, infeccin de la herida
operatoria, Staphylococcus aureus resistente a meticilina (SARM).

Introduccin que gobiernan la auscultacin, dndole una validez a su

E
uso como herramienta con fines semiolgicos2.
senciales para la vida son los sistemas cardiovascu- Durante el siglo XX se hicieron muchas mejoras me-
lar y pulmonar por lo que no nos ha de extraar que nores a los estetoscopios, reduciendo el peso y mejorando
cuando el hombre comenz a estudiar la fisiologa la calidad acstica filtrando el ruido externo1. A pesar de
y fisiopatologa del cuerpo humano, buscara herramientas todas las mejoras y cambios, el principio bsico detrs
que le permitieran poder examinar de mejor manera dichos del estetoscopio sigue siendo el mismo: proporcionar a
sistemas. Dentro de stos, el estetoscopio del griego, los mdicos los medios para llevar a cabo la auscultacin,
stthos que significa pecho, y skop que significa identificar los sonidos especficos dentro del cuerpo hu-
examinar, tambin llamado fonendoscopio ha sido el mano y actuar en caso de identificar alguna anormalidad1.
instrumento de tipo acstico que ms se ha utilizado para La infeccin asociada a la atencin de salud (IAAS),
cumplir con dicho propsito durante los ltimos 200 aos. tambin denominada infeccin intrahospitalaria o
A principios del siglo XIX, y antes del desarrollo del nosocomial se define como una infeccin que tiene
estetoscopio, los mdicos realizaban el examen fsico lugar en un paciente durante su atencin en un hospital u
utilizando la percusin y auscultacin inmediata, la otro establecimiento de atencin sanitaria, que no estaba
que consista en colocar directamente el odo sobre el presente o no se estaba incubando al momento de la
paciente1. admisin. Incluye las infecciones adquiridas en el lugar
En el ao 1816 el mdico Francs Ren Lannec in- de atencin sanitaria que se presentan luego del alta, y
vent el primer estetoscopio que consisti en un tubo de las infecciones ocupacionales entre los trabajadores de
madera mono-auricular el cual dio a conocer en su obra la salud del establecimiento3.
De lauscultation mdiate ou Trait du Diagnostic des Los principales factores de riesgo para la adquisicin
Maladies des Poumon et du Coeur tres aos ms tarde1. de una IAAS son las condiciones propias del paciente,
Durante el ao 1851 el mdico Irlands Arthur Leared mo- los procesos asociados a la atencin de salud (PAAS)
dific el estetoscopio hacindolo bi-auricular1. Algunos y algunos aspectos ambientales. Los esfuerzos en todo
aos despus los estetoscopios bi-auriculares comenzaron el mundo han centrado su inters en los PAAS, ya que
a divulgarse y su calidad fue mejorando paulatinamente, intervenciones a este nivel han demostrado ser efectivas,
especialmente despus de las investigaciones de Rappa- disminuyendo los riesgos para adquirir una IAAS4.
port y Sprague (1915), quienes estudiaron las leyes fsicas Hace ms de 160 aos se instaur un protocolo de lava-

Rev Chilena Infectol 2016; 33 (1): 19-25 www.sochinf.cl 19


Infecciones Asociadas a la Atencin de Salud
Artculo Original

do de manos y desinfeccin de instrumental mdico, pero Tan slo en E.U.A. se han reportado 2 millones de
en la actualidad muchos mdicos no se lavan las manos IAAS al ao, con una sobre-estada hospitalaria promedio
con regularidad, ni mucho menos limpian o desinfectan de 5 das por paciente. En el caso de las IHO se estima
adecuadamente su estetoscopio5. La falta de control estric- una sobre-estada promedio de 7,5 das, 7 a 21 das en
to se ve agravado por el aumento de bacterias resistentes bacteriemias y de 6,8 a 30 das en neumonas. Todo lo
a los antimicrobianos. Dentro de estos, Staphylococcus anterior se traduce para el pas en un gasto anual de US$
aureus resistente a meticilina (SARM), incluyendo 4.532 millones y una sobre-estada de 8.676 millones de
SARM que ha sido adquirido en la comunidad, genera un das cama al ao8.
especial inters, debido a que estos microorganismos son
responsables de infecciones que van desde lesiones graves Infecciones Asociadas a la Atencin
de piel hasta cuadros graves de sepsis, atribuyndoles A. de la Salud en Chile
Schroeder 18.650 muertes estimadas al ao6. Chile cuenta desde el ao 1983 con un Programa Na-
Segn el informe de vigilancia de IAAS del ao cional de Prevencin y Control de IAAS dependiente del
2012 del Ministerio de Salud (MINSAL), la IAAS ms MINSAL, el que a travs de un sistema de vigilancia epi-
frecuentes en nuestro pas corresponden a la infeccin demiolgica, difusin de normas y conductas basadas en la
del tracto urinario (21,5%), seguida por la infeccin de la evidencia mundial y local, fortalecimiento de la capacidad
herida operatoria (IHO) (14,2%), la neumona no asociada local para el diagnstico e intervencin de infecciones
a ventilacin mecnica (9%), bacteriemias/septicemias adems de una evaluacin externa de los programas loca-
(8,1%) y la neumona asociada a ventilacin mecnica les, tanto en su organizacin como en su funcionamiento,
(7,1%)4 (Tabla 1). busca disminuir la incidencia de las IAAS4.
En pases como Francia, se estima que el gasto prome- A pesar de estos esfuerzos, en Chile se estima una
dio por enfermo que cursa con una IAAS es de 1.800 a incidencia de 70 mil IAAS al ao, con un aumento de la
3.600 dlares con sobre-estadas que van desde 7 hasta 15 sobre-estada hospitalaria de 10 das por paciente, un costo
das. En pases desarrollados, el total de gastos asciende estimado en US$ 70 millones ao y una sobre-estada
a 5-10 mil millones de dlares anuales. Si se estima que anual de 700 mil das cama, segn un estudio publicado
la infeccin es la causa de muerte en 1 a 3% de los pa- el ao 20038 (Tabla 2).
cientes ingresados, se tendrn cifras como las reportadas El presente artculo revisa la literatura cientfica rela-
en Estados Unidos de Amrica (E.U.A.), de 25 a 100 mil tiva a la contaminacin microbiana de los estetoscopios
muertes anuales7 (Tabla 2). utilizados por el personal de salud y la efectividad de los
productos utilizados para su desinfeccin.

Tabla 1. Frecuencia de Infecciones Asociadas a la Atencin


de Salud segn tipo de infeccin en hospitales chilenos. Material y Mtodos
Ao 2012
Se efectu una revisin bibliogrfica de 11 trabajos
Tipo de infeccin % publicados durante los aos 1997-2015, cuyo objetivo
Infeccin del tracto urinario 21,5 era determinar el grado de contaminacin microbiana
Infeccin de herida operatoria 14,2 presente en los estetoscopios utilizados por mdicos, es-
Neumona no asociada a ventilacin mecnica 9 tudiantes de medicina, enfermeras(os) y otros trabajadores
Bacteremia/septicemia 8,1 de la salud y la efectividad de los productos utilizados
Neumona con ventilacin mecnica 7,1 para su desinfeccin (alcohol isoproplico, alcohol etlico
y jabn antispticos). La contaminacin bacteriana de
Adaptado de ref. 4 los estetoscopios y la eficacia de su desinfeccin fueron

Tabla 2. Incidencia de Infecciones Asociadas a la Atencin de Salud y costos asociados en Chile y Estados Unidos de Amrica. Ao 2003

Incidencia anual Sobre-estada hospitalaria Sobre-estada hospitalaria Costo pas ao US $ Mortalidad anual
anual x por paciente
IAAS Chile 70.000 casos 700.000 das cama 10 das 70.000.000 -
IAAS Estados Unidos 2.000.0000 casos 8.676.000 das cama 5 das 4.500-5.000 millones 25.000-100.000

Adaptado de ref. 8

20 www.sochinf.cl Rev Chilena Infectol 2016; 33 (1): 19-25


Infecciones Asociadas a la Atencin de Salud
Artculo Original

medidas, en unidades formadoras de colonias (ufc), de los considerar que es un instrumento de uso generalizado
cultivos procedentes de los diafragmas, antes y despus y con mucha frecuencia, especialmente por el mdico,
de haber sido desinfectados. durante el examen fsico.

Fuentes de informacin Bacterias patgenas aisladas de estetoscopios


Se realiz una bsqueda de artculos en las principales Segn diversas publicaciones, los estetoscopios, tanto
bases de datos bibliogrficas disponibles en internet con- el diafragma, como la campana y las olivas, son contami-
cretamente en: PubMed/MEDLINE y Scientific electronic nados por microorganismos patgenos con una frecuencia
library online (Scielo). Para la bsqueda ms compleja, de 85% (tasa media segn OFlaherty y cols.), incluyendo
que fue aquella realizada en PubMed/MEDLINE, se cepas de SARM, Enterococcus resistentes a vancomicina
utilizaron los descriptores MeSH Stethoscope, conta- (ERV) y de Acinetobacter baumannii panresistente10,12,13,
mination bacterial, nosocomial infection. que causan morbilidad y mortalidad significativa.
La buena tolerancia a las condiciones de desecacin y
Criterios de inclusin carencia de nutrientes de Staphylococcus spp y Entero-
Del total de artculos encontrados con los criterios coccus spp permite que sean importantes contaminantes
de bsqueda (n: 64) se consideraron los estudios que de superficies10. La evidencia seala que S. aureus puede
evaluaron los estetoscopios pertenecientes al equipo de permanecer en superficies por perodos de siete das a siete
salud (mdicos, estudiantes de medicina, enfermeras, meses en superficies, Acinetobacter spp entre tres das
tcnicos paramdicos y otros que tuvieran contacto con y cinco meses, Enterococcus spp de cinco das a cuatro
los pacientes en algn servicio de un recinto hospitalario) meses y Clostridium difficile hasta por cinco meses16.
y que adems realizaran dos o ms de las siguientes Otros estudios sealan que adems de Staphylococcus
intervenciones: spp y Enterococcus spp, los principales agentes aislados
Cultivo de la muestra proveniente del diafragma del en los estetoscopios5,6,9-15,17,18, se ha aislado a ms de 10
estetoscopio en placas de agar sangre durante 24-48 h especies diferentes, como Corynebacterium spp, Bacillus
entre 36-38C. spp, Micrococcus luteus, Neisseria spp, Candida spp
Tipificacin bacteriolgica de los cultivos mediante y varias especies de bacilos gramnegativos, los que
tinciones y pruebas bioqumicas especficas. provienen en 80% de estetoscopios de mdicos5 (Tablas
Pruebas de susceptibilidad in vitro a las bacterias 3 y 4). Tambin fueron aisladas cepas de C. difficile y
aisladas de los cultivos procedentes de los diafragmas Pseudomonas aeruginosa en cultivos procedentes de los
de los estetoscopios. estetoscopios5,6,9-15,17,18.
Desinfeccin del diafragma del estetoscopio con algn Durante el ao 2010, Uneke Chigzoi J. y cols., reali-
desinfectante (alcohol isoproplico, alcohol etlico zaron una investigacin examinando un total de 107 este-
o jabn antisptico) y medicin de la eficacia de la toscopios; de stos, 73 pertenecan a mdicos, mientras 34
accin cultivando una muestra post desinfeccin del provenan de enfermeras y otros trabajadores de la salud.
diafragma para medir la contaminacin bacteriana De los 107 estetoscopios estudiados, 84 (78,5%) tenan
medida en ufc. contaminacin bacteriana, 59 de estos procedan de los
Comparacin de la eficacia de un desinfectante versus estetoscopios de los mdicos y 25 de otros trabajadores
otro mediante la cuantificacin de las ufc procedentes de salud. Las bacterias aisladas fueron S. aureus (53,6%),
de los cultivos post desinfeccin. P. aeruginosa (19,0%), Enterococcus faecalis (14,3%) y
Encuestas a personal de salud respecto a la frecuencia Escherichia coli (13,1%)8 (Tabla 3).
de la desinfeccin del estetoscopio y del producto Mndez y cols., cultivaron durante el ao 2012 mues-
utilizado. tras provenientes de 155 estetoscopios de estudiantes de
medicina, informando su contaminacin por diferentes
Resultados microorganismos patgenos; dentro de stos, Staphylo-
coccus spp. fue el principal agente aislado, incluyendo
Evidencia del estetoscopio como potencial vector de cepas de SARM10. Este estudio adems report que las
infecciones asociadas a la atencin de salud. cocceas grampositivas presentaron resistencia a oxacilina
Se identific al estetoscopio como potencial vector entre 22 y 37%. En los bacilos gramnegativos cultivados
de IAAS hace ms de 30 aos. La evidencia seala que la resistencia a cotrimoxazol alcanz a 67%10.
microorganismos, entre ellos bacterias de importancia De los artculos que fueron considerados para esta
mdica, se adhieren y contaminan los diafragmas, revisin, todos coinciden y concluyen que los principales
campanas y olivas de los estetoscopios, pudiendo ser contaminantes de los estetoscopios son SARM, ERV y A.
transmitidas a los pacientes6,9-15 si stos no son desin- baumannii panresistente, los que pueden ser transmitidos
fectados (Tabla 3). Lo anterior resulta preocupante al a los pacientes si no se practica una desinfeccin. Los

Rev Chilena Infectol 2016; 33 (1): 19-25 www.sochinf.cl 21


Infecciones Asociadas a la Atencin de Salud
Artculo Original

Tabla 3. Microorganismos aislados de estetoscopios de mdicos, enfermeras(os) u otro personal de la salud, segn diferentes estudios

Estudio/microorganismo Mndez I. Uneke C. Whittington Zuliani M. Nez S. Marinella M. Schroeder A.


201210 20109 A. 200913 200214 199912 199711 20096*
Staphylococcus aureus 63 (40,7%) 36 (49,3%) 22 (100%) 176 (59%) 6 (5%) 15 (38%) 2 ufc
Staphylococcus coagulasa negativa 153 (51%) 40 (100%) 100 ufc
Staphylococcus epidermidis 7 (4,51%) 118 (97%)
Staphylococcus aureus resistente a meticilina 7 (32%) 3 ufc
Bacillus spp 14 (12%) 51 ufc
Stenotrophomonas maltophilia 22 (100%)
Micrococcus spp 48 (40%) 24 ufc
Pseudomonas aeruginosa 11 (15,1%) 2 ufc
Escherichia coli 5 (3,2%) 5 (6,85%)
Listeria spp 21 (13,6%)
Corynebacterium spp 31 (26%)
Acinetobacter baumannii 12 (55%)
Enterococcus faecalis 7 (9,6%)
Enterobacter cloacae 7 (32%)
Enterococcus spp 3 (1,93%)
EC cultivados (n)/ 155/ a 63 73/59 22/22 300/261 122/118 40/40 92/ 92
EC contaminados (%) (40,65%) (80,8%) (100%) (87%) (97%) (100%) (100%)

*Medido en ufc. Adaptado de refs. 6, 9, 10, 11, 12, 13 y 14. EC: estetoscopio.

Tabla 4. Frecuencia de limpieza de estetoscopio por mdicos y enfermeros

EC Profesional Frecuencia de limpieza


(n)
Diariamente Semanalmente Mensualmente/ Anualmente Nunca
2-4 semanas atrs* 5 o ms semanas atrs**
90 Mdicos (n = 43) 5,6% (n = 5) 8% 10% 26% 30% (n = 13)
Enfermeros (n = 47) 12% 22% 15%
EC (n)/EC (n) (%) contaminado segn frecuencia de limpieza
107 Mdicos (n = 73) 30/18 (60%) 12/11 (91,7%)* 5/4 (80%)** 26/26 (100%)
Enfermeros (n = 34) 22/13 (59,1%) 0/0 (0%)* 1/1 (100%)** 11/11 (100%)

Adaptado de refs. 9 y 12. EC: estetoscopio.

estudios contribuyen adems, al dejar en evidencia que depende de la concentracin; el rango comprendido entre
el personal de salud no realiza con frecuencia la desinfec- 60 y 90% de solucin en agua (volumen/volumen) es el
cin de su estetoscopio -los mdicos son quienes menos indicado para la accin bactericida. An as no destruyen
realizan dicha accin- lo que se agrava al considerar la esporas bacterianas por lo que su uso no es til para
tasa media de contaminacin de los estetoscopios (85%) prevenir por ejemplo brotes por C. difficile.
referida por OFlaherty y cols., para 28 estudios revisa- En un estudio realizado el ao 1997 por Marinella M. y
dos15 (Tabla 4). cols., se obtuvieron a partir de cultivos procedentes de 40
estetoscopios seleccionados aleatoriamente, bacterias de
Desinfeccin del estetoscopio 11 distintos gneros y especie. Hubo crecimiento de Sta-
El alcohol etlico y el alcohol isoproplico son sustan- phylococcus coagulasa negativa en 100% de las muestras
cias bactericidas para toda forma vegetativa de bacterias, y 38% de S. aureus; la media del nmero de ufc fue de 158
tuberculicidas, fungicidas y virucidas. Su actividad 33 por diafragma11. Los estetoscopios de los mdicos

22 www.sochinf.cl Rev Chilena Infectol 2016; 33 (1): 19-25


Infecciones Asociadas a la Atencin de Salud
Artculo Original

presentaron mayor cantidad de ufc de Staphylococcus


Tabla 5. Efectividad del alcohol isoproplico en la reduccin de unidades formadoras de
spp en comparacin con los de enfermeras. El agente de colonias de los estetoscopios
limpieza ms efectivo fue el alcohol isoproplico: despus
de ser utilizado para limpiar la superficie de la membrana Alcohol isoproplico 70% Jabn antisptico
de los diafragmas de los estetoscopios logr disminuir Estudio ufc antes de ufc despus ufc antes de ufc despus
desde 158 33 ufc a 0,2 0,2 ufc y en los bordes desde la limpieza de la limpieza la limpieza de la limpieza
289 54 ufc a 2,2 1,5 ufc (P = 0,01)11 (Tabla 2). del EC del EC del EC del EC
Otro estudio realizado en el Servicio de Urgencia del Marinella M. 1997 158 33 0,2 2 - -
Hospital La Candelaria, Espaa, durante el ao 1999 por Nez S. 1999
Nez S. y cols., compar el crecimiento bacteriano antes Staphylococcus aureus > 1.000 (-) > 100 270
y despus de la desinfeccin de 49 estetoscopios del perso- Acinetobacter spp > 1.000 (-) > 1.000 100
nal mdico. La limpieza de los estetoscopios con alcohol Enterobacter spp > 1.000 (-) > 1.000 40
isoproplico elimino el 99% de las colonias (p < 0,01),
mientras que el jabn antisptico fue inefectivo (Tabla EC: estetoscopio. Adaptado de refs. 11 y 12.
5). Adicionalmente, se realiz una encuesta a 90 profe-
sionales del servicio (mdicos y enfermeros) consultando
la frecuencia con la que desinfectaban sus estetoscopios, Segn lo referido por Whittington A. en el ao 2009,
destacando que slo 5 (5,6%) de stos sealaron limpiarlo en su publicacin Contaminacin bacteriana en este-
a diario; en contraparte, 13 mdicos (30%) respondieron toscopios de la unidad de cuidados intensivos (Bacterial
no haberlo desinfectado nunca12 (Tabla 4). contamination of stethoscopes on the intesive care unit)
Schroeder A y cols., el ao 2009, analizaron un total la desinfeccin de los estetoscopios con alcohol isopropli-
de 184 cultivos provenientes de 92 estetoscopios (Tabla co reduce drsticamente el nmero de colonias bacterianas
3). Se compar el recuento de colonizacin bacteriana en el diafragma de los estetoscopios, entre 94 y 100%.
pre limpieza y post limpieza de stos. En la pre limpieza An as, en ese estudio hubo dos cepas de SARM que no
todos los participantes imprimieron el diafragma de su es- pudieron ser eliminadas con slo una limpieza. Adems,
tetoscopio en una placa de agar chocolate, inmediatamente la contaminacin aument considerablemente de 0 a 69%
despus desinfectaron con una espuma a base de alcohol despus de 24 h de haber efectuado la desinfeccin12 por lo
etlico al 62,5% sus manos frotando de forma simultnea que la desinfeccin de los estetoscopios debiera realizarse
el diafragma del estetoscopio entre stas. Luego de un antes y despus del uso con cada paciente11-14,17.
breve tiempo de secado, los participantes volvieron a
imprimir el diafragma del estetoscopio sobre otra placa
de agar chocolate. Qued en evidencia la disminucin de Discusin
la colonizacin posterior a la limpieza. Tres colonias de
SARM se identificaron en el perodo de prelavado, siendo Las IAAS son un problema de salud pblica a nivel
totalmente eliminadas con alcohol etlico al 62,5%6. mundial. stas impactan negativamente en la calidad de
Para la desinfeccin de los diafragmas y campanas vida del paciente, ya que aumentan la estada hospitalaria,
de los estetoscopios las soluciones en base a alcohol han y repercuten en el pronstico y evolucin de la patologa
demostrado ser las ms eficaces logrando eliminar hasta de base.
en 100% la contaminacin bacteriana, lo que ha quedado En trminos econmicos, se traducen para al pas en
evidenciado al no observar crecimiento bacteriano en costos cercanos a los $70 millones de dlares con los
los cultivos provenientes de muestras post desinfeccin, que podran construirse al menos 12 Centros de Salud
ya fuese con alcohol isoproplico al 70% o con alcohol Familiar (CESFAM) si se considera el presupuesto del
etlico al 62,5%. Otras soluciones desinfectantes tambin MINSAL del ao 2014.
fueron evaluadas pero no demostraron ser efectivas como La evidencia seala que los estetoscopios son poten-
los alcoholes; an as, cualquier conducta destinada a la ciales vectores de IAAS, al ser contaminados por diversos
desinfeccin de los estetoscopios contribuye a disminuir microorganismos en 70 a 100% de los casos, aislndose
la posibilidad de transmitir potenciales patgenos a los microorganismos de importancia mdica como SARM y
pacientes5,6,9-15,17,18. otros potenciales patgenos, con especial trascendencia
En los artculos revisados por los autores, se compar para pacientes con algn estado de inmunocompromiso,
la eficacia del alcohol isoproplico al 70% versus alco- heridas, quemaduras, portadores de catteres u otros
hol etlico al 62,5% y soluciones en base a etanol, sin dispositivos invasores12.
observarse una diferencia estadsticamente significativa La limpieza del estetoscopio se realiza de manera
en su eficacia, ya que con ambos alcoholes se obtuvieron inconsistente, especialmente entre el personal mdico;
cultivos negativos post limpieza (Tabla 5). el hbito de la desinfeccin estetoscopio con diferentes

Rev Chilena Infectol 2016; 33 (1): 19-25 www.sochinf.cl 23


Infecciones Asociadas a la Atencin de Salud
Artculo Original

tipos de alcohol, como el isoproplico al 70% es una hasta 99% de los microorganismos de importancia mdica
buena oportunidad para controlar la diseminacin de en las IAAS16.
bacterias intrahospitalarias ya que elimina 99% de los Si bien no se ha definido cul es el mtodo ptimo
microorganismos presentes en stos. Esta limpieza que para la desinfeccin de los estetoscopios, la evidencia
debe realizarse antes y despus del uso con cada paciente. coincide al sealar que los desinfectantes en base de
Es importante considerar que el lavado de manos alcohol son efectivos al reducir significativamente la
previo al contacto con el paciente podra resultar menos contaminacin bacteriana presente en los diafragmas
efectivo si el personal de la salud manipula y utiliza un y campanas. Sugerimos realizar la desinfeccin de los
estetoscopio sucio, ya que existe la posibilidad de con- estetoscopios con dichas soluciones, antes y despus de su
taminarse con microorganismos patgenos y viceversa. uso con cada paciente, con posterioridad a la desinfeccin
Resulta prioritario buscar medidas que disminuyan de las manos ya que bacterias de importancia mdica en
cualquier fuente de transmisin de IAAS, debido al las IAAS pueden ser transferidas desde los estetoscopios
aumento progresivo de pacientes en estado de inmuno- hacia los pacientes15.
compromiso (pacientes en tratamiento con quimioterapia,
con terapia inmunosupresora, diabetes mellitus, infeccin Agradecimientos. A M. Jos Monsalves, Directora de In-
por VIH en etapa SIDA, receptores de trasplantes u otros), vestigacin IPSUSS, por su colaboracin durante el proceso
aquellos sometidos a procedimientos quirrgicos o con de revisin y correccin del artculo.
alguna patologa crnica, que aumenta el riesgo para
adquirir una IAAS.
Lamentablemente no existe una norma en Chile Resumen
ni protocolos de limpieza especficamente diseados
para la desinfeccin de los estetoscopios; tampoco se Las infecciones asociadas a la atencin de la salud
ha establecido cul es la frecuencia con la que se deben (IAAS) son un problema a mundial. Slo en nuestro
limpiar. La pregunta es relevante considerando que se ha pas se estima una incidencia de 70.000 al ao, lo que
demostrado que existe una correlacin directa entre el se traduce en un aumento de la estada hospitalaria en
grado de contaminacin, la baja frecuencia con la que se 10 das promedio por paciente, un costo anual estimado
limpia el estetoscopio, y su potencial rol como vector de en 70 millones de dlares y una sobre-estada de 700 mil
microorganismos patgenos de importancia en las IAAS. das cama al ao.
No existe bibliografa que relacione la colonizacin La evidencia seala que diferentes instrumentos de uso
del estetoscopio con la incidencia de las IAAS, por lo que hospitalario resultan contaminados por patgenos; entre
la realizacin de estudios en recintos hospitalarios que ellos, el estetoscopio se ha identificado como potencial
permitan establecer esta relacin, normando por ejemplo vector de IAAS hace ms de 30 aos y adquiere gran
la limpieza del estetoscopio, deberan ser foco de nuestro relevancia al ser de uso generalizado. Microorganis-
inters. El prximo paso de los autores es, mediante un mos patgenos, incluyendo cepas de Staphylococcus
ensayo clnico, medir el impacto de la desinfeccin de aureus resistente a meticilina y cepas de Enterococcus
los diafragmas y campanas con alcohol isopropilico al resistente a vancomicina se adhieren y contaminan los
70% como medida para disminuir la incidencia de IAAS estetoscopios, pudiendo transmitirse a otros pacientes si
en hospitales de mediana y alta complejidad de Santiago. no son desinfectados. Diversos estudios concluyen que
Tampoco existe bibliografa que seale si es el material la desinfeccin del estetoscopio con alcohol isoproplico
del que estn compuestos los estetoscopios son los que elimina hasta 99% de estas y otras bacterias, por lo que
permiten la contaminacin y/o colonizacin. An as el existe una clara oportunidad para aportar a la prevencin
uso del cobre y su capacidad antimicrobiana podra ser de las IAAS, interviniendo a travs de la implementacin
una opcin a evaluar en la fabricacin de estetoscopios ya de medidas sencillas, econmicas y operativas al corto
que superficies de este material han demostrado eliminar plazo, normando por ejemplo la limpieza del estetoscopio.

Referencias bibliogrficas Disponible en http://www.who.int/gpsc/ fuente potencial de infeccin nosocomial. Acta


country_work/burden_hcai/es/ (Consultado el 4 Pediatr. Costarric 2005; 19: 8-12.
1.- Roguin A. Rene Theophile Hyacinthe Lannec de noviembre de 2014). 6.- Schroeder A. Whats growing on your
(1781-1826). The man behind the stethoscope. 4.- Otaza F, Bustamante R, Snchez P. Informe stethoscope? And what you can do about it.
Clin Med Res 2006; 4 (3): 230-5. de Vigilancia de IAAS, ao 2012. Ministerio J Fam Pract 2009; 58 (8): 404-9.
2.- Goic A. Semiologa Mdica. Ed. Mediterrneo de Salud (MINSAL), Gobierno de Chile, 7.- Rivero L, lvarez A, Balleste I, Villarreal A,
2a Edicin. 1999 p. 333-334. 2012. Galbn O. Tendencias y pronsticos de las
3.- OMS, Organizacin Mundial de la Salud. 5.- lvarez T, Herrera J, vila M. Estetoscopios: infecciones hospitalarias y sus gastos asociados.

24 www.sochinf.cl Rev Chilena Infectol 2016; 33 (1): 19-25


Infecciones Asociadas a la Atencin de Salud
Artculo Original

Rev Cubana Obstet Ginecol 2009-2; 35 (4): 11.- Marinella M, Pierson C, Chenoweth C. healthcare-associated infection: a snake in the
150-61. The stethoscope, a potential source of grass or innocent bystander? J Hosp Infect
8.- Brenner P, Nercelles P, Pohlenz M, Otaza nosocomial infection? Intern Med 1997; 157 2015; 91 (1): 1-7.
F. Costo de las infecciones intrahospitalarias (7): 786-90. 16.- Prado V, Vidal R, Durn C. Aplicacin de
en hospitales chilenos de alta y mediana 12.- Nez S, Moreno A, Rodrguez I, Garca P, la capacidad bactericida del cobre en la
complejidad. Rev Chilena Infectol 2003; 20 (4): Hernndez J, Izquierdo C. El estetoscopio prctica mdica. Rev Med Chile 2012; 140:
285-90. como vector de la infeccin nosocomial. Rev de 1325-32.
9.- Uneke Chigzoi J, Ogbonna A, Oyibo P, Onu C. Emergencias 1999; 11: 281-5. 17.- Parmar R, Valvi C, Sira P. A prospective,
Bacterial contamination of stethoscopes used 13.- Whittington A, Whitlow G, Hewson D, randomised, double-blind study of comparative
by health workers: public health implications. J Thomas C, Brett S A. Bacterial contamination efficacy of immediate versus daily cleaning of
Infect Dev Ctries 2010; 4 (7): 436-41. of stethoscopes on the intensive care unit. stethoscope using 66% ethyl alcohol. Indian J
10.- Mndez I, Calisto O, Becerra W, Vsquez J, Anaesthesia 2009; 64: 620-4. Med Sci 2004; 58 (10): 423-30.
Bravo J, Pachn D. Microorganismos presentes 14.- Zuliani M, Fogli A, Bercial M, Ayres S. 18.- Lecat P, Cropp E,McCord G,Haller N A.
en fonendoscopios, manos, cavidad oral y nasal Stethoscope: a friend or an enemy? Sao Paulo Ethanol-based cleanser versus isopropyl alcohol
de estudiantes de una facultad de medicina. Rev Med J/Rev Paul Med 2002; 120 (1): 13-5. to descontaminate stethoscopes. Am J Infect
Fac Med 2012; 20 (1): 90-100. 15.- OFlaherty N, Fenelon L. The stethoscope and Control 2009; 37 (3): 241-3.

Rev Chilena Infectol 2016; 33 (1): 19-25 www.sochinf.cl 25

Anda mungkin juga menyukai