Abstract: This study evaluated the light intensity of various light curing units, the effect of distance of
the light guide, and the validity of a tapered light guide. Light curing units tested included (1) four blue
lightemitting diode curing units, Lux-O-Max, LEDemetron1, Ortholux LED, and The Cure; (2) two
tungsten-quartz halogen curing units, Optilux 501 and Co-bee; and (3) one plasma arc curing unit, Apol-
lo95E. The Optilux 501 was also evaluated for combinations of normal mode and boost mode and Standard
tip and Turbo tip light guide. The spectral output of each unit was measured from 300 to 600 nm with a
spectroradiometer. The light intensities at distances of zero, five, 10, 15, and 20 mm were determined with
the radiometer. The peak value of Ortholux LED and The Cure surpassed that of Apollo95E. The light
intensity significantly decreased with distance. Although The Cure showed a higher light intensity than
the LEDemetron1 at zero-mm distance, the light intensity of the LEDemetron1 was higher than that of
The Cure at five to 20 mm, resulting in no significant difference. The boost mode increased light intensity
at any distance. Although the Turbo tip enhanced light intensity at zero-mm distance, reduction of light
intensity by Turbo tip was demonstrated at five- to 20-mm distance. (Angle Orthod 2004;74:810815.)
Key Words: Light curing units, Blue lightemitting diode, Tungsten-quartz halogen, Plasma arc
of resins and resin-reinforced glass ionomer cements equal Co-bee, and Apollo95E (Table 1). All units were new and
to those produced with 40-second exposures of convention- unused. The cordless-type light curing units, Lux-O-Max,
al TQH.1518 However, PAC units have several disadvantag- LEDemetron1, Ortholux LED, and Co-bee, were supplied
es, eg, they are expensive and of a relatively large size and with fully charged batteries. Optilux501 was also evaluated
complex construction.19 using combinations of the normal mode and boost mode
Light curing units with gallium nitride blue lightemit- and two types of light guide tip (11-mm Standard tip and
ting diodes (LED) have also been developed.20 The spectral eight-mm Turbo tip).
output of LED falls within the absorptive region of cam- The spectral output of each light curing unit was mea-
phorquinone, so the LED requires no filters to produce blue sured from 350 to 550 nm with a spectroradiometer (USR-
light. The spectrum flux of LED is concentrated over a 40V, USHIO, Tokyo, Japan). A Broadband Power/Energy
much narrower bandwidth than that of TQH or PAC.2123 Meter (30 W, MELLES GRIOT, Carlsbad, Calif) was used
LEDs have many advantages compared with TQH such as to calibrate the value measured by the spectroradiometer.
a lifetime of more than 10,000 hours, invariable output en- A curing radiometer (Model100, Demetron Research
ergy over this term without degradation, and suitability for Corp, Danbury, Conn) was used to measure the light inten-
portable use because of low energy consumption and resis- sity of emissions between 400 and 525 nm. The diameter
tance against shock or vibration. They produce between a of all light guide tips was standardized to produce equiva-
410 and 500 nm. Although the LED shows 70% of the lent exposed areas. The measuring window of the radiom-
irradiance produced by the TQH, the depth of cure pro-
eter was covered with black paper with a five-mm hole.
duced by the LED was greater than that by the TQH.21
When the light intensity exceeded the range of the curing
Furthermore, advances in the power output of LEDs have
radiometer (1000 mW/cm2), measurement was performed
allowed LEDs to achieve a higher irradiance than TQH.
through a filter (ND0.3, FUJIFILM, Tokyo, Japan) and ad-
These high-intensity LEDs may decrease total light curing
justed. Each light curing unit was warmed up for 30 sec-
time.
onds before commencement of each test. The end of the
The light intensity is affected by various clinical expo-
light guide was placed in contact with the center of the
sure uses. It is recommended to place the light guide as
close as possible to the surface of the light-cured adhesive, measuring window of the curing radiometer at right angles,
which is rarely performed in clinical orthodontic practice. and each reading of the maximum power output was ob-
The distance between the light guide and orthodontic brack- tained within five seconds after light passed through the
et base corresponds to the diameter of the light guide when measuring window. Similarly, measurements were per-
the light guide is placed against the tooth surface at an formed at five-, 10-, 15-, and 20-mm distance from the light
angle of 458. This distance of approximately 10 mm may guide. Each distance was standardized by using acrylic
adversely affect the light intensity received by the light- tubes, eight mm in diameter and five, 10, 15, and 20 mm
cured adhesives inside the bracket base.24 long. To obtain the consistency of the results in power out-
The purpose of this study was to evaluate the light in- put, five measurements were repeated after an interval of
tensity of conventional TQH, PAC, and LED and to deter- one minute. One operator performed all measurements to
mine the effect of distance while moving the light guide eliminate any bias.
away. The validity of a tapered light guide was also as- Means and standard deviations of the light intensity at
sessed. each distance were calculated for each group. The data were
subjected to a repeated-measures analysis of variance (AN-
MATERIALS AND METHODS OVA) to determine whether any significant difference ex-
Seven light curing units were used in this study: Lux-O- isted between the light intensity at each measurement (P ,
Max, LEDemetron1, Ortholux LED, The Cure, Optilux501, .05). The Scheffe test was carried out to identify statistical
TABLE 2. Total Sample Means and Standard Deviations (SD) of Light Intensity (mW/cm2) Measured with the Curing Radiometer
Distance From the Light Guide (mm)
Zero Five 10 15 20
Light Units Mean SD Mean SD Mean SD Mean SD Mean SD
Lux-O-Max 120 0 93 2.7 72 2.7 50 0 38 2.7
LEDemetron1 508 8.4 334 5.5 250 0 192 4.5 158 2.7
Ortholux LED 924 19.5 368 8.4 212 4.5 164 5.5 118 4.5
The Cure 708 17.9 314 8.9 188 4.5 139 6.5 110 6.1
Optilux501
Normal/Standard tip 404 8.9 278 4.5 180 0 132 2.7 100 0
Normal/Turbo tip 508 21.7 196 5.5 120 7.1 94 2.2 74 5.5
Boost/Standard tip 482 8.4 313 4.5 200 0 152 4.5 112 4.5
Boost/Turbo tip 600 7.1 216 5.5 131 2.2 100 0 84 2.2
Co-bee 548 14.8 326 5.5 202 4.5 154 8.9 123 4.5
Apollo95E 1004 8.9 798 14.8 602 14.8 494 8.9 406 5.5
minished severely at distances more than two mm and was be needed to evaluate the bond strength of adhesives cured
just by 25% of the maximum value at four-mm distance. by high-intensity LED with reduced curing time.
Clinically realistic distances of five to 10 mm would dra-
matically decrease the light intensity. In this study, a dis- CONCLUSIONS
tance of five mm with the Optilux501 in normal mode
showed decreases of 31% for Standard tip and 61% for The spectral output and light intensity were measured
Turbo tip, and at 10 mm, decreases of 55% for Standard with four LEDs, Lux-O-Max, LEDemetron1, Ortholux
tip and 76% for Turbo tip. LED, and The Cure; two TQH, Optilux501 and Co-bee;
Tapered light guides such as Turbo tip were developed and a PAC, Apollo95E. The effects of distance while mov-
to condense the light energy and might be expected to in- ing the light guide away were determined. In the measure-
crease the light intensity. Turbo tip showed its maximum ment of spectral output, Ortholux LED showed the highest
ability in contact with the surface of curing radiometer. peak value at 458 nm. The light intensity significantly de-
Moving away from the surface, however, a greater drop in creased with distance. Although The Cure showed a higher
power density occurred, and the Standard tip delivered light intensity than LEDemetron1 at zero-mm distance, the
greater light intensity than Turbo tip at distances greater light intensity of LEDemetron1 was higher than that of The
than five mm. In this study, Turbo tip increased the light Cure at five to 20 mm. As a result, there was no significant
intensity by 26% in comparison with the Standard parallel- difference between The Cure and LEDemetron1. The boost
sided light guide. However, Turbo tip decreased the light mode increased the light intensity at any distance. Although
intensity by 29% when the light guide and the radiometer Turbo tip enhanced light intensity at zero-mm distance, re-
were separated by five mm. These findings were in agree- duction of light intensity by Turbo tip was demonstrated at
ment with previous studies that Turbo tip resulted in a sig- five- to 20-mm distance.
nificant reduction in light intensity with distance.4042 Bis-
hara et al11 evaluated the effect of the difference between REFERENCES
a four-mm mini Turbo tip and a 11-mm Standard tip and
1. Bassiouny MA, Grant AA. A visible light-cured composite re-
concluded that there was no significant difference in bond storative. Br Dent J. 1978;145:327330.
strength. Although the use of Turbo tip increased light in- 2. Newman SM, Murray GA, Yates JL. Visible lights and visible
tensity, severe reduction was noted with distance. light-activated composite resins. J Prosthet Dent. 1983;50:3135.
Some light curing units have an enhanced output mode 3. ReadMJF. The bonding of orthodontic attachments using a visible
light cured adhesive. Br J Orthod. 1984;11:1620.
such as boost mode for increasing light intensity. Boost 4. Nomoto R. Effect of light wavelength on polymerization of light-
mode showed a higher light intensity by approximately cured resins. Dent Mater J. 1997;16:6073.
20% than conventional output mode at zero-mm distance. 5. Cook WD. Spectral distributions of dental photopolymerization
However, the enhancement of light intensity by boost mode sources. J Dent Res. 1982;61:14361438.
6. Althoff O, Hartung M. Advances in light curing. Am J Dent.
was also diminished with increasing distance from the end
2000;13:77D81D.
of the light guide. At distances more than five mm, boost 7. Fujibayashi K, Ishimaru K, Takahashi N, Kohno A. Newly de-
mode increased light intensity by approximately 10%. This veloped curing unit using blue light-emitting diodes. Dent Jap.
increase was too low to compensate the decreased light in- 1998;34:4953.
tensity because Optilux501 in normal mode decreased 31% 8. Martin FE. A survey of the efficiency of visible light curing units.
J Dent. 1998;26:239243.
for Standard tip and 61% for Turbo tip at five-mm distance. 9. Rueggeberg FA, Caughman WF, Curtis JW Jr. Effect of light in-
Boost mode might not increase the useful light intensity tensity and exposure duration on cure of resin composite. Oper
when the light guide is positioned away from the bracket. Dent. 1994;19:2632.
The reduction of light intensity depends on various clin- 10. Swartz ML, Phillips RW, Rhodes B. Visible light-activated res-
ical exposure terms such as the direction of light guide, insdepth of cure. JADA. 1983;106:634637.
11. Bishara SE, VonWald L, Zamtua J. Effects of different types of
shade of enamel, and distance between the end of light light guides on shear bond strength. Am J Orthod Dentofacial
guide and the surface of adhesives. Distance is the most Orthop. 1998;114:447451.
important factor in decreased light intensity. Although Tur- 12. Evans LJ, Peters C, Flickinger C, Taloumis L, Dunn W. A com-
bo tip and boost mode are effective for restorative dentistry, parison of shear bond strengths of orthodontic brackets using var-
ious light sources, light guides, and cure times. Am J Orthod
the enhanced light exposure may not produce the expected
Dentofacial Orthop. 2002;121:510515.
light intensity because it is impossible to position the light 13. Curtis JW Jr, Rueggeberg FA, Lee AJ. Curing efficiency of the
guide at zero-mm distance from the adhesive under the Turbo Tip. Gen Dent. 1995;43:428433.
bracket base. Increased curing time and light intensity can 14. Oesterle LJ, Newman SM, Shellhart WC. Rapid curing of bond-
compensate the loss of light intensity caused by the clinical ing composite with a xenon plasma arc light. Am J Orthod Den-
tofacial Orthop. 2001;119:610616.
factors. PAC and high-intensity LEDs with parallel-sided 15. Peutzfeldt A, Sahafi A, Asmussen E. Characterization of resin
light guides may be advantageous alternatives for light ex- composites polymerized with plasma arc curing units. Dent Ma-
posure in orthodontic clinical practice. Further study will ter. 2000;16:330336.
16. Loney RW, PriceRBT. Temperature transmission of high-output 30. Nomoto R, Uchida K, Hirasawa T. Effect of light intensity on
light-curing units through dentin. Oper Dent. 2001;26:516520. polymerization of light-cured composite resins. Dent Mater J.
17. Klocke A, Korbmacher HM, Huck LG, Kahl-Nieke B. Plasma arc 1994;13:198205.
curing lights for orthodontic bonding. Am J Orthod Dentofacial 31. Strydom C. Curing lightsthe effects of clinical factors on in-
Orthop. 2002;122:643648. tensity and polymerisation. SADJ. 2002;57:181186.
18. Ishikawa H, Komori A, Kojima I, Ando F. Orthodontic bracket 32. Myers ML, Caughman WF, Rueggeberg FA. Effect of restoration
bonding with a plasma-arc light and resin-reinforced glass ionom- composition, shade and thickness on the cure of a photoactivated
er cement. Am J Orthod Dentofacial Orthop. 2001;120:5863. resin cement. J Prosthodont. 1994;3:149157.
19. Clinical Research Associates. Resin curing lights. Newsletter. 33. Shortall AC, Wilson HJ, Harrington E. Depth of cure of radiation-
2000;24:14. activated composite restorativesinfluence of shade and opacity.
20. Mills RW. Blue light emitting diodesanother method of light J Oral Rehabil. 1995;22:337342.
curing? Br Dent J. 1995;178:169. 34. Williams PT, Johnson LN. Composite resin restoratives revisited.
21. Mills RW, Jandt KD, Ashworth SH. Dental composite depth of J Can Dent Assoc. 1993;59:538543.
cure with halogen and blue light emitting diode technology. Br 35. Tavas MA, Watts DC. Bonding of orthodontic brackets by trans-
Dent J. 1999;186:388391. illumination of a light activated composite: an in vitro study. Br
22. Jandt KD, Mills RW, Blackwell GB, Ashworth SH. Depth of cure J Orthod. 1979;6:207208.
and compressive strength of dental composites cured with blue 36. Behrents RG, Wendt SL, Fox DM, Smith RT, King L. A trans-
light emitting diodes (LEDs). Dent Mater. 2000;16:4147. illumination technique for lingual bonding. J Clin Orthod. 1987;
23. Stahl F, Ashworth SH, Jandt KD, Mills RW. Light-emitting diode 21:324325.
(LED) polymerization of dental composites: flexural properties 37. Ash MM Jr. Wheelers Dental Anatomy, Physiology and Occlu-
and polymerization potential. Biomaterials. 2000;21:13791385. sion. 7th ed. Philadelphia, Pa: WB Saunders Co; 1992:128169.
24. Prati C, Chersoni S, Montebugnoli L, Montanari G. Effect of air,
38. Moseley H, Strang R, Stephen KW. An assessment of visible-
dentin and resin-based composite thickness on light intensity re-
light polymerizing sources. J Oral Rehabil. 1986;13:215224.
duction. Am J Dent. 1999;12:231234.
39. Sakaguchi RL, Douglas WH, Peters MCRB. Curing light perfor-
25. Yearn JA. Factors affecting cure of visible light activated com-
posites. Int Dent J. 1985;35:218225. mance and polymerization of composite restorative materials. J
26. Park YJ, Chae KH, Rawls HR. Development of a new photoini- Dent. 1992;20:183188.
tiation system for dental light-cure composite resins. Dent Mater. 40. Price RB, Derand T, Sedarous M, Andreou P, Loney RW. Effect
1999;15:120127. of distance on the power density from two light guides. J Esthet
27. Stansbury JW. Curing dental resins and composites by photopoly- Dent. 2000;12:320327.
merization. J Esthet Dent. 2000;12:300308. 41. Hansen EK, Asmussen E. Visible-light curing units: correlation
28. Dunn WJ, Taloumis LJ. Polymerization of orthodontic resin ce- between depth of cure and distance between exit window and
ment with light-emitting diode curing units. Am J Orthod Den- resin surface. Acta Odontol Scand. 1997;55:162166.
tofacial Orthop. 2002;122:236241. 42. Meyer GR, Ernst CP, Willershausen B. Decrease in power output
29. Shortall AC, Harrington E. Effectiveness of battery powered light of new light-emitting diode (LED) curing devices with increasing
activation units. Br Dent J. 1997;183:95100. distance to filling surface. J Adhes Dent. 2002;4:197204.