METHOD DEVELOPMENT
OBJECTIVE
I. To identify the components of the mixture using the selected HPLC conditions.
ABSTRACT
The chromatography process begins by injecting a sample solution into the mobile
phase through the injector port. As a sample solution in the mobile phase flows through a
column, the compounds of the solution will migrate according to the interactions of the
components with the stationary phase. The chemical interactions of the mobile phase and
sample solution, as well as with the stationary phase, determine the degree of migration and
separation of components contained in the sample. For example, the components that have
stronger interactions with the mobile phase than with the stationary phase will elute faster
from the column and have shorter retention time compared to those components which have
stronger interactions with the stationary phase. The type and composition of the mobile phase
is one of the influence in separation of the components. Therefore, the mobile phase can be
altered in order to manipulate the interactions of the sample components and the stationary
phase. In isocratic elution, components are eluted using constant mobile phase composition.
In gradient elution, different components are eluted by changing the mobile phase
composition as the separations proceeds. By increasing the strength of the mobile phase will
subsequently results in elution of retained components.
INTRODUCTION
High performance liquid chromatography is the most widely used of all of the analytical
separation technique. Its suitable for separating nonvolatile species or thermally fragile ones.
Partition chromatography is the most widely used of all the four types of liquid
chromatography procedure. It divides into two; normal-phase chromatography and reverse-
phase chromatography.
For this analysis we used reversed phase chromatography. In reverse-phase chromatography,
the stationary phase is non polar and the mobile phase is relatively polar. The most polar
component will elute first, and increasing the mobile phase polarity increase the elution
time. Method development tends to be more complex in liquid chromatography because the
sample components interact with both the stationary phase and the mobile phase. Successful
chromatography with interactive mobile phase requires a proper balance of intermolecular
forces among the three active participants in the separation process- the solute, the mobile
phase, and the stationary phase. These intermolecular forces are described qualitatively in
term of the relative polarity of three reactants. The polarities of various analytes functional
groups in increasing order are: hydrocarbon <ether <ester < ketones < aldehyde < amides <
amines < alcohols. Water is more polar compounds than compounds containing any of the
preceding functional groups.
Often in choosing a column for a partition chromatographic separation, the polarity of the
stationary phase is matched roughly with that of the analytes; a mobile phase of considerably
different polarity is then used for elution. This procedure is generally more successful than
one in which the polarities of the solute and mobile phase are matched but different from that
of the stationary phase. Here, the stationary phase often cannot compete successfully for the
sample components; retention time becomes too short for practical application. At the other
extreme, of course, is the situation where the polarity of the solute and stationary phase are
too much alike and totally different from that of the mobile phase. Here, the retention times
becomes inordinately long.
In summary, polarities for solute, mobile phase and stationary phase must be carefully
blended if good partition chromatography separation are to be realized in a reasonable time.
Unfortunately, theories of mobile phase and stationary phase interaction with any given set of
sample component are impacted, and at best, we can only narrow the choice of stationary
phase to a general type. Having made this choice, we then perform a series of set trial and
error experiment in which chromatogram are obtained with various mobile phase until a
satisfactory separation is realized. If resolution of the entire component of a mixture proves to
be impossible, different types of column may have to be chosen.
ANALYTICAL PROCEDURE
RESULT
FOR ISOCRATIC ELUTION;
Response Factor Area Mobile Phase Ratio Retention Time
(minute)
79.3762 7937.625 1.128
28.1650 2816.501 1.273
26.7675 2676.751 2.591
25.3343 2533.430 50% H2O : 50% ACN 3.834
678.492 67849.20 10.392
Sample ID: Standard mixture (100 ppm)
Respond Factor (RF) = Peak Area
Sample Amount (ppm)
For Peak 1 = 7937.625 = 79.3762
100
For Peak 2 = 2816.501 = 28.1650
100
For Peak 3 = 2676.751 = 26.7675
100
For Peak 4 = 2533.430 = 25.3343
100
For Peak 5 = 67849.20 = 678.492
100
DISCUSSION
During this experiment, a High Performance Liquid Chromatography (HPLC) Agilent
G1314A equipped with UV detector, 5 m Reverse Phase C18 column and 20 l sample loop
was used. At flow rate 1.5 ml / min and detector wavelength at 254 nm, the mobile phase
ratio (v/v) was set at 50% water and 50% acetonitrile at the beginning in order to analyze and
observe the effect of mobile phase on LC separation. After all the standard samples which is
standard mixture, caffeine, acetone, phenanthrene, methyl benzoate, and phenatole were
injected, the ratio was changed to 70%:30% respectively on the same mobile phase. Based on
the actual procedure, from this experiment we need to identify the components contained in
the standard mixture by using the optimized LC conditions getting from the above ratio of the
mobile phase as well as we should perform a gradient elution separation to improve the
efficiency of the column. Meaning that, isocratic elution is performed with a single solvent or
constant solvent mixture. If one solvent does not provide sufficiently rapid elution of all
components, then gradient elution can be used. In this case, increasing amounts of water are
added to acetonitrile to create a continuous gradient.
But the result shows all the peaks from the injection process to the sample loop were
not separated well. In a reversed-phase separation, the strength of eluent decreases as the
solvent becomes more polar. Acetonitrile has high eluent strength, and all compounds are
eluted rapidly. All the peaks are observed overlapping. From the result of chromatogram and
area calculation, we can see that the Response Factor for all the standards injected is almost
same. It was so difficult to determine the resolution of the peaks since the peaks got overlap
because the mixture is in high concentration. As we know, the quantitative analysis in
separation method depends upon direct relationship between the area under a peak or peak
height in the chromatogram and the amount of the compound corresponding to that peak in
the analyzed sample. Therefore, each peak should be totally resolved from any neighboring
peaks. A co-elution or other anomalies such as tailing or fronting will distort or obscure the
beginning and ending points of the peak.
Another reason, there are some factors that contribute to all the problems stated
above. The sample must be degassing properly. Sometime when the pressure was not
consistent, there must be any air bubble in the mobile phase that fluctuant the instrument.
Therefore, the instrument should be purge to let the pressure stable. Mobile phase that is too
cooled also effect the pressure. The 254nm is the most suitable wavelength because give us
very nice and sharp peak. The flow rate or velocity of the mobile phase is very essential in
HPLC (according to the Van Deemter Equation).
CONCLUSION
The components is identified as well as their peak retention time.
REFERENCES
1. Skoog, Holler and Nierman, 5th Edition. Principles of Instrumental Analysis.
Thomson Learning 1998
2. Skoog, D.A., West, D.M, Holler, F.J. 7th Edition, Fundamental of Analytical
Chemistry
3. Saim, N., Tajuddin, R., & Saaid, M. (2014). Analytical separation methods
laboratory guide. Selangor: UiTM Press.
(EXPERIMENT 2)
HIGH PERFORMANCE LIQUID
CHROMATOGRAPHY (HPLC):
METHOD DEVELOPMENT
NAME: NURUL HAZIQAH BINTI HASAN
STUDENT ID: 2016666772
PARTNERS NAMES: 1. NUR AININA BINTI MOHAMED AINI
2. NURUL JUNAIDAH BINTI TERMIZI
3. RAHAYU BINTI ABDUL RAHMAN
DATE OF SUBMISSION: 09/06/2017