M-Mode 2-35
Ultrasound Module Questions 2-37
REVISION HISTORY
A3175 58212 Rev. 2 J. Madarasz S. Williams Dec. 2000 Incorporate reviewer comments
INTRODUCING ULTRASOUND
OVERVIEW The field of diagnostic ultrasound has developed rapidly over the
past decade, but the basic physical principles have remained
consistent. The individuals who perform routine maintenance and
work with ultrasound equipment must understand the underlying
physical principles. This module prepares the reader to
conceptualize basic ultrasound principles in order to communicate
with the clinical user and to troubleshoot Acuson ultrasound
equipment when required.
wavefronts
TYPICAL FREQUENCY The following table displays the typical frequency ranges of sound
RANGES waves.
ULTRASOUND APPLICATIONS
Diagnostic ultrasound is one of the fastest growing and most
exciting imaging technologies. High performance ultrasound is
now the primary diagnostic imaging modality for many
applications. Acuson is a leading manufacturer, international
marketer, and service provider of diagnostic ultrasound systems
and image management products in the following main
applications: cardiology, OB/GYN, radiology, and vascular
imaging.
VASCULAR Both 2-D and Doppler modes are utilized extensively in vascular
work. The 2-D image provides anatomical information about blood
vessels, while Doppler modes display and measure blood flow.
Clinical applications of ultrasound in vascular work include the
detection and evaluation of abnormalities such as aneurysm and
stenosis.
ULTRASOUND IMAGING
Transducer
Transmitted Energy
Transducer
Anatomy
Reflected Energy
TRANSDUCERS
DEFINITION A transducer is a device that converts one form of energy into
another.
In ultrasound, the transmission and reception of ultrasound waves
is achieved by using a transducer. An ultrasound transducer
converts voltage impulses into ultrasound waves, transmits those
waves into the body, and converts returning echoes into voltages.
The transducer is acoustically coupled to the patient by means of a
water-based gel that matches the transducer-to-patient acoustic
impedance. Transducers come in various shapes and sizes to meet
the needs of various medical diagnostic applications.
Linear Array
Linear Transducer
Format
Nearfield Scan
Lines
Farfield
Curved Linear The curved linear array format provides a wide field of view. The
Array Transducers piezoelectric elements are arranged along a curved arc in linear
fashion.
Transducer
Vector and Sector The Vector wideview array transducer format is designed
Transducers exclusively by Acuson and was introduced in 1991. The Vector
transducer increases the near field of view without increasing the
transducer footprint. This is achieved with Acusons proprietary
software control and solves the usual sector problem of an over-
sampled near field. Vector transducers obsolete conventional sector
technology.
Additional Nearfield
Ribs
Transducer
Footprint Size
Sector Apex
Ribs
Scan Lines
NewView Format
Other Transducers Besides the basic transducer formats, there are many other types of
transducer available for speciality applications, including
transesophageal, endovaginal, endorectal, interoperative, and
catheter.
Acoustic Lens
Summation produces
Electronic Focus spherical wavefronts con-
verging toward the focal point
Timed Transmit Pulses
Focal
Point
The following two examples show how the focal point is positioned
at the center of the image or off-axis by adding a linear delay across
the array of the piezoelectric elements.
Delay
Line t0
Delay
Line t1
Delay
Line t2
Focal Point
Delay
Line t1
Delay
Line t0 Piezoelectric
Start Pulse Elements
Delay
Line t0
Delay
Line t1
Delay
Line t2
Delay
Line t3
Delay
Line t4 Piezoelectric Focal Point
Start Pulse Elements
Figure 2-12 Beam Steering
Focal Point
Focal Point
Figure 2-13 Improved Focus
Delay
Dynamic focus is achieved by controlling the delay of each signal
arriving at each element through each channel, such that only
signals from the computed sliding focal point arrive at a final
summation point at the same time. Signals originating from other
points do not sum in phase at the final output, and therefore tend to
cancel each other out, providing insignificant addition.
The system computer runs the proprietary software necessary to
accomplish this in real time.
With Apodization
Reduced Side Lobes
Focus 40 mm Focus 80 mm
Aperture 20 mm Aperture 40 mm
Aperture 60 mm Aperture 80 mm
Figure 2-16 Dynamic Aperture
BEAMWIDTH Beamwidth is the size of the ultrasound beam at any location. It is
narrow at the near field and widens as depth increases. Focusing is
used to maintain beamwidth over depth. Since lateral resolution is
related to the ability of the ultrasound beam to detect small objects
across the width of the beam., decreasing the beamwidth improves
lateral resolution.
The amplified
reflectors are sent to
an ultrasound Scan
Converter where
Ultrasound they are converted to
Scanner brightness and
XMT RCV formatted to a specific
location on the CRT.
XDCR
The Body
Focused
Wave Front
Amplitude of
reflections at Anatomical
specific depths Structure
One Theoretical
Ultrasound (or Scan) Line
Transducer
Line1
Line2
Line3
Line4
Line5
Line6
Line7
Line8
SCAN CONVERSION
DEFINITION Scan conversion is the process of reformatting ultrasound data for
display on a raster. Ultrasound data must undergo a scan
conversion before it can be displayed on a monitor. Ultrasound data
acquired from the returning echoes is in the vertical format, but can
be displayed only in the horizontal raster format.
TV Raster Format
MEMORY MAPPING In scan conversion, incoming echo information has been digitized.
This digital information is then stored in a memory map and
manipulated. The (digital) output of the memory map is converted
into analog horizontal television raster lines and displayed on a
CRT.
Data In
Data Out
ULTRASOUND FRAME When the ultrasound system has completed transmitting and
RATE receiving across the 2-D image area, the information is stored and
displayed as a full frame.
The ultrasound frame rate is the number of frames captured per
second. The acoustic or ultrasound frame rate depends upon the
number of lines fired, depth, number of focal zones applied, and the
application used. The more ultrasound lines fired, the longer it
takes to produce a single frame.
Depth = 16 cm
Lines = 128
Setup time = 32s
Round triplets = 16 cm x 13
s/cm = 208 s
208 s + 32 s setup time =
160 mm 240 s
240 s x 128 lines = 30.7 ms/
frame
= 32.25 frames/second
~128 Ultrasound Lines
Axial
Lateral
DYNAMIC RANGE There are two definitions for the dynamic range. The signal dynamic
range is the measure of the ultrasound signal magnitudes that can
be processed by the ultrasound system. The display dynamic range is
the measure of the signal magnitudes that can be displayed by the
system. The amplitudes of the returning echoes may be up to 100dB
to 150dB below the original intensity transmitted. This wide range
of echo strength is also reflected in the induced signals at the
receiver section. Logarithmic amplification reduces the dynamic
range of the induced RF signal. This decreasing of the dynamic
range is called log compression. Further processing of the signal
reduces the dynamic range as displayed below.
Anatomical
Interfaces
Overall Processing Display
Transducer Gain &
DGC
SIGNAL STRENGTH The ultrasound signal is attenuated (reduced in signal strength) the
farther it goes into the body and depending on what it comes in
contact with. The attenuation reduces the intensity of the beam.
The transmit signal strength can be made stronger by using a lower
transmit frequency. This however will compromise resolution.
The received signal strength can be increased using overall 2-D gain
control or Depth Gain Compensation (DGCs). DGCs allows the user
to enhance specific areas of interest by increasing or decreasing the
amplification of echoes at different depths. It compensates for losses
in signal strength as the ultrasound wave passes through
anatomical structures. Normally, the field of view is divided into
eight DGC zones. A DGC slider controls the DGC for one zone. The
positions of the eight controls with respect to each other define the
DGC curve displayed with the image.
80 mm 40 mm
40 mm
64 Real
Ultrasound
Lines 80
mm
160
mm 400
80 mm TV Lines
400
TV Lines 200 TV Lines
2N Ultrasound Lines
N Ultrasound Lines
Figure 2-21 Regional Expansion Selection (RES)
CINE In Cine mode,
f1
multiple f2
frames are f3 Time
f4
captured and f5
then viewed in
looping
playback, one
frame at a time.
The review of
captured
frames is
controlled by
increasing or
decreasing the
review speed.
DOPPLER PRINCIPLES
DOPPLER EFFECT
SPECTRAL DOPPLER
SPECTRAL DOPPLER In spectral Doppler mode, blood flow can be imaged or visualized,
MODES to depict changes in blood flow velocities over time. There are three
spectral Doppler modes:
PULSED WAVE In pulsed wave (PW) Doppler, a Doppler cursor is positioned over
DOPPLER the vessel of interest in the 2-D image. The vessel is further focussed
by using a sample volume or range gate to obtain flow information
from a selected depth.
In PW Doppler, one element or group of elements transmits and
receives. The elements transmit and the system waits to receive the
returning echoes. The time delay between transmit and receive
relates directly to the depth of the targeted tissue (reflectors.)
A single ultrasound line is fired. Echoes reflected from moving
blood undergo a shift in frequency. Using the Doppler equation,
echo information obtained within the range gate is analyzed for
frequency shift and amplitude. The frequency data is converted to
velocity and displayed in a scrolling strip format on the system
monitor.
Weak Signal
(Few cells moving at this velocity)
Peak Systolic Velocity
Strong Signal
(Many cells moving at this velocity
Dicrotic Notch Envelope
(Outer Edge)
TV, MV close
Window
(Indicating
normal flow)
Time
Systole Diastole Systole Diastole
PULSE REPETITION The number of times the transducer piezoelectric element is pulsed
FREQUENCY per second is called the pulse repetition frequency (PRF.) This
frequency depends on a physical limitation that a transducer cannot
transmit and receive at the same time. The maximum PRF is
constrained by the transmit depth and the propagation velocity. The
time required for an ultrasound beam to transmit and receive once
is called the pulse repetition period (PRP). It is equal to the inverse
of the PRF. PRP = 1 / PRF
NYQUIST LIMIT If the sampling rate is not adequate for high-frequency Doppler
shifts, artifactual Doppler shifts are displayed. The highest Doppler
shift velocity measurable in PW Doppler is limited to one-half of the
sampling rate defined by the PRF. PRF is determined mainly by the
sampling depth. This maximum measurable velocity for a given
transducer and depth is known as the Nyquist limit, and can be
calculated using the following equation: Nyquist Limit = PRF / 2
TX TX TX TX
Amplitude
RX RX RX RX
Ultrasound Line Ultrasound Line Ultrasound Line Ultrasound Line
Time
Pulse Repetition Period
HIGH PRF The Nyquist limit decreases with increasing depth of ultrasound
wave penetration. As this limit reaches 1/2 PRF, the system
automatically increases the sample rate by increasing the PRF and
the number of samples. As a result, multiple wavefronts propagate
through the tissue simultaneously. Therefore, information obtained
may be from more than one location. These other locations are
indicated by ghost gates. The ghost gates are real, and are
positioned by the user such that they are not in the region of flow.
PLUG FLOW In plug flow, most of the blood moves at the same velocity and in
the same direction. Plug flow occurs in a narrow range of velocities
and primarily in large vessels such as the aorta.
LAMINAR FLOW In laminar flow, the blood flows more slowly along the vessel walls
and somewhat faster at the center of the vessel. Flow is essentially
in one direction. Laminar flow occurs in a narrow range of
velocities and in vessels that are straight, smooth, and
unobstructed.
TURBULENT FLOW In turbulent flow, the blood flows at various speeds and in various
directions, even including opposing directions. Turbulent flow
usually occurs past an area of obstruction or restriction, such as
sharp vessel bends, bifurcation, narrowing, obstruction, or leaks.
m/s
COLOR DOPPLER
In color Doppler,
blood flow in real-
time is displayed
as a color overlay
across the 2-D
image, showing
blood flow
through vessels
and organs. Color Color Box
Doppler can
display either the
velocity of blood
flow, the energy,
or a combination of the two.
When Color Doppler mode is invoked, a color box appears in the 2-
D image. This box is used to select the area of interest for the CD
information. The size and position of the color box are adjustable.
SAMPLE VOLUME Multiple range gates are taken along multiple Color Doppler lines
MAPPING to compute the frequency shift of blood flow. This is converted into
velocity information and assigned a color to represent a certain
velocity and direction, and then combined with the 2-D image at the
location of origin.
Transducer Transducer
Multiple
Gates
CD Box
d1 d2d3d4 .....
Scanning Doppler Lines
Image
VELOCITY DISPLAY Hue indicates direction
(towards or away from
Yellow transducer or ultrasound
Flow towards Increasing line) and intensity.
the transducer Velocities Typically red is blood
flow towards the
transducer and yellow is
Red high flow towards.
0 Typically blue is blood
Blue flow away from the
transducer and green is
high flow away.
Flow away from Increasing The scale must be
the transducer Velocities adjusted higher than
Green maximum velocity of
blood flow or aliasing
(wrap around) will occur.
Figure 2-30 Velocity Color Display Key
VELOCITY Color flow imaging requires location as well velocity of the moving
DETECTION object. To assess the motion multiple echoes from the same object
must be collected using a series of transmitted pulses. The rapid
acquisition and analysis of flow data are achieved with multiple
gate PW Doppler signals. Doppler signal processing is initiated
with a quadrature detection or QD. This results in two output
signals:
Inphase or I Signal
Quardrature or Q Signal (Out of phase)
Just like the 2-D image data, the data acquired from the returning
echoes is analog format. These I&Q signals are then digitized, scan
converted, and color mapped to display the color Doppler
information.
Color velocity estimation does not depict spectral analysis of each
sample volume. An average Doppler shift is estimated, resulting in
a time-averaged velocity estimation.
COLOR SCAN Digitized color Doppler data undergoes a scan conversion similar
CONVERSION to that in 2-D mode. The main purpose of the color scan conversion
is to display flow Doppler lines in the raster format. The resultant
flow information is overlaid on top of the 2-D image and displayed
on the monitor. The following figure depicts how flow information
is collected from multiple range gates, flow velocity estimated, and
scan converted for display.
DOPPLER TISSUE The DTI or Doppler Tissue Imaging is exact opposite of color
IMAGING Doppler Velocity. CDV employs tissue motion discrimination to
filter out the Doppler signals being reflected by the moving tissues
in the anatomy in order to display just the blood flow motion.
Unlike CDV, the DTI automatically detects and displays moving
tissues and ignores the Doppler signals from blood flow.
DTI capability introduces new methodology to study myocardial
motion, conduction abnormalities, diastole function, and stress
echo. It is available in three options: Velocity mode, Acceleration
mode, and Energy mode.
M-MODE
DEFINITION M-mode, also called motion mode, is used to display a graphic
representation of a line of interest within the 2-D display, such as
the movement of a heart valve. M-mode displays a graph that
shows how the line changes over time.
Heart Valves
Figure 2-32 M-Mode Data Acquisition
DATA DISPLAY The lines are displayed side by side on the system monitor. A
motion-time map is created, which can be used to verify valve or
muscle operation.
RECORDING M-mode strip can be recorded on a page printer, but the most
common archival method is a strip chart recorder, which provides a
continuous reproduction of the M-mode trace. With the advent of
digital image management systems like AEGIS, M-mode data can
be stored digitally on various media, including hard disk and
magneto-optical disk.