Anda di halaman 1dari 5

Portable Coronary Active Perfusion System for

Off-Pump Coronary Artery Bypass Grafting


Yoshinao Koshida, MD, Go Watanabe, MD, Tamotsu Yasuda, MD,
Shigeyuki Tomita, MD, Shinichi Kadoya, MD, and Taro Kanamori, MD
Department of General and Cardiothoracic Surgery, Kanazawa University School of Medicine, Kanazawa, Ishikawa, Japan

Purpose. The present study was performed to develop a new perfusion system for
NEW TECHNOLOGY

off-pump coronary artery bypass grafting and to examine whether even a simple coronary
perfusion system can maintain adequate blood flow delivery during anastomosis.
Description. The experiment was performed in two stages. In procedure 1, 3 pigs with left
anterior descending artery occlusion were used to evaluate optimal perfusion flow rate and
coronary artery internal pressure, and to evaluate the safety area of perfusion. In procedure
2, 6 pigs were used to validate the new portable coronary perfusion system.
Evaluation. The optimal blood flow in the portable coronary active perfusion system was
less than approximately 40 mL/min. The small, easy to use pump system (ie, the portable
coronary active perfusion system) may prevent hemodynamic deterioration and ventricular
arrhythmia during coronary occlusion, resulting in better maintenance of left ventricular
function.
Conclusions. Even a simple pump system can achieve effective perfusion for safe
anastomosis. Further studies are required to allow the clinical use of this system.
(Ann Thorac Surg 2006;81:706 11)
2006 by The Society of Thoracic Surgeons

O ff-pump coronary artery bypass grafting (OPCABG)


is a promising technique for the treatment of
ischemic heart disease. However, one concern with
Technology and Technique
We recently developed the coronary active perfusion
system (CAPS) to avoid blood supply inadequacy in the
OPCABG is that temporary occlusion of the coronary
ischemic myocardium in OPCABG [6]. This is an active
artery, which is frequently required during anastomosis perfusion technique in which oxygenated blood is sup-
in OPCABG, can cause hemodynamic instability and plied from the femoral artery and pressurized by a pump
myocardial damage [1]. In addition, retraction and stabi- to optimize blood flow to the myocardium, and the
lization maneuvers during OPCABG often cause sys- amount of blood supply is not dependent on hemody-
temic hypotension, especially when the heart is displaced namic status during the procedure. However, CAPS
vertically for exposure of its lateral and posterior includes a special pump, controller, and power supply
branches [2]. system, and therefore is not easy to access, especially in
Several clinical studies have demonstrated the effec- emergencies. For that reason, use of this procedure has
tiveness of intracoronary shunts, external shunt circuits, not yet become widespread.
or ischemic preconditioning procedures [3, 4] to reduce We have developed a new perfusion system that may
ischemic injury. These methods are advantageous in enable active perfusion of arterial blood into the coronary
terms of both cost and time, and can be easily set up. artery during coronary artery occlusion, and thus avoid
However, these methods have shown little or no effect in ischemic injury during the procedure. Moreover, the
patients with severe proximal coronary artery stenosis or apparatus required is small and simple in comparison
severe ischemic heart disease with low coronary pressure with other coronary perfusion systems [6 8]. Herein we
and unstable flow. Moreover, it is unclear how the report details of our new method and describe experi-
adequacy of blood flow is affected by internal or external ence with the new coronary perfusion system.
shunts when systemic blood pressure deteriorates during
retraction or compression of the heart in OPCABG [5].
Clinical Experience
All animals received humane care in accordance with the
Accepted for publication June 22, 2005. Principle of Laboratory Animal Care formulated by the
Address correspondence to Dr Koshida, 13-1 Takaramachi, Kanazawa, National Society for Medical Research and the Guide
920-8641 Japan; e-mail: yoshinaoko2000@yahoo.co.jp. for Care and Use of Laboratory Animals, prepared by

2006 by The Society of Thoracic Surgeons 0003-4975/06/$32.00


Published by Elsevier Inc doi:10.1016/j.athoracsur.2005.06.064
Ann Thorac Surg NEW TECHNOLOGY KOSHIDA ET AL 707
2006;81:706 11 NEW PERFUSION SYSTEM FOR OPCABG

Fig 1. Measurement of coronary perfusion


flow and coronary arterial pressure during
coronary active perfusion. (A) Schematic pre-
sentation of perfusion circuit. (B) Efferent
pump and flow meter. (C) The CAPS can-
nula. (CAPS coronary active perfusion sys-
tem; LAD left anterior descending artery;
PCPS percutaneous cardiopulmonary
support.)

NEW TECHNOLOGY
the Institute of Laboratory Animal Resources, National resulting arterial blood passed through a modified per-
Research Council and published by the National Acad- cutaneous cardiopulmonary support circuit with a cen-
emy Press, revised in 1996. trifugal pump (Mix Flow [JMS, Tokyo, Japan]) was led to
Nine pigs were used in this study. All pigs were the installed inline electromagnetic flow probe (Nihon-
sedated by intramuscular injection of ketamine (20 Kohden, Tokyo, Japan) in the last part of the circuit to
mg/kg body weight). Anesthesia was maintained with measure flow of coronary active perfusion (Fig 1). Coro-
halothane (0.5% to 1.5%), and muscle relaxation was nary perfusion flow was gradually increased in a step-
induced with pancuronium (0.1 mg/kg), which was ad- wise manner to determine how much flow it raised to
ministered through the peripheral intravenous route. critical pressure and to evaluate the upper limit of
An arterial pressure line was inserted into the brachio- optimal perfusion flow.
cephalic artery through the common carotid artery. A
Swan-Ganz catheter was inserted into the pulmonary Procedure 2
artery through the right internal jugular vein for pressure In procedure 2, 6 pigs were used to validate the new
monitoring and continuous cardiac output measurement coronary perfusion system, including cardiac function
(IntelliCath CCO/VIP, SAT-2 [Deerfield, Baxter, IL]). A during perfusion. The portable CAPS is shown in Figure
4-French catheter was inserted into the right femoral 2. Arterial blood was passed through extension tubes and
artery to remove arterial blood for the perfusion system. was pumped out from the microdiaphragm pump (CM
The experiment consisted of two stages. 15W Enomoto Micro Pump [Enomoto Kogyo, Tokyo,
Japan]). This pump system is small, simple, battery-
Procedure 1 driven, and inexpensive. The pump and battery can be
In procedure 1, 3 left descending coronary artery oc- placed in a small waterproof case and can be operated by
cluded pigs were used to determine the optimal perfu- a surgeon alone without the assistance of a perfusionist.
sion flow rate. After systemic heparinization (200 u/kg), This small pump can transfer liquid at a maximum rate of
the left descending coronary artery was snared at a point 100 mL/min. The approximate flow rate can be set using
just distal from the first diagonal branch, and a coronary the coronary perfusion cannula. The coronary cannula
arteriotomy was performed. A pressure wire (WaveMap used in the present study has already been applied in our
[Endosonics, Rancho Cordova, CA]) and coronary perfu- unit as previously reported [6, 7]. The enlarged fixed
sion cannula were then inserted through the arteriotomy portion of the cannula can prevent back bleeding from
site. To prevent escape of pressure from the arteriotomy the arteriotomy site by selecting a size of enlarged
site, a site distal to that of arteriotomy was snared. The portion suitable for each coronary artery. In this proce-
708 NEW TECHNOLOGY KOSHIDA ET AL Ann Thorac Surg
NEW PERFUSION SYSTEM FOR OPCABG 2006;81:706 11

Fig 2. (A) Small diaphragm pump and motor


unit. (B) Portable coronary active perfusion
system (CAPS). (C) The CAPS cannula.
NEW TECHNOLOGY

dure, we used a cannula measuring 1.25 mm in external Results of Procedure 1


diameter and 0.8 mm in inner diameter. Coronary pressure increased proportionately with grad-
Coronary perfusion was performed for 30 minutes as ual increases in perfusion flow. Until a perfusion flow of 40
described in procedure 1, and hemodynamic and me- mL/min, the mean slope was expressed by Y 1.41X 22.1.
chanical data were recorded. Suitable portable CAPS However, with perfusion flow in excess of 40 mL/min, the
perfusion flow was set based on flow that did not surpass mean slope was greater than the values expected from
the upper limit set in procedure 1. A conductance cath- the perfusion flow until 40 mL (Y 6.66X 184; r 0.999;
eter (Conductance-PC software [Cardio Dynamics BV, p 0.01), and the increase in peripheral resistance
Zoetermeer, The Netherlands]) was used to evaluate left predicted endothelial injury (Fig 3). At a perfusion flow
ventricular function and mechanical data. Left ventricu- exceeding 60 mL/min, coronary pressure increased re-
lar contractility was quantified based on the slope of the gardless of changes in perfusion flow.
end systolic pressure-volume relation (end-systolic elas-
tance [Ees] [mm Hg/mL]). The endpoint of this study was
30 minutes of observation without ventricular arrhyth-
mia. At the end of the experiment, the pigs were given a
lethal intravenous injection of pentobarbital and potas-
sium chloride, and transmural samples of the left ven-
tricular anterior wall were taken for histologic
examination.
Statistical analyses were performed using SPSS for
Windows (version 10.1.3J [SPSS Inc, Chicago, IL]). Cu-
mulative data are expressed as means standard devi-
ation. Simple linear regression was used to analyze the
relationship between the perfusion flow rate and coro-
nary pressure. The paired t test was performed for
analysis of hemodynamics before and after the experi-
ments, and repeated analysis of variance was used for
analysis of changes in the hemodynamics, coronary ar- Fig 3. The relationship between perfusion flow and mean coronary
tery pressure, coronary perfusion flow, and Ees. artery pressure. Data are presented as means standard deviation.
Ann Thorac Surg NEW TECHNOLOGY KOSHIDA ET AL 709
2006;81:706 11 NEW PERFUSION SYSTEM FOR OPCABG

NEW TECHNOLOGY
Fig 4. Changes in mean arterial pressure (top). Changes in the slope Fig. 5. Changes in perfusion flow during the experiments (top).
of the end systolic pressure volume relation (Ees) during the experi- Changes in coronary pressure (bottom).
ments (bottom). Data are presented as means standard deviation.

indicated little flow with these methods. Muraki and


Results of Procedure 2 colleagues [5] reported that intracoronary shunts pro-
Ventricular arrhythmia was not observed in any of the vided only 10% 30% of baseline blood flow in dogs
pigs for 30 minutes. None of the hemodynamic variables even without stenosis of the coronary artery. Moreover, it
(ie, heart rate, aortic pressure, cardiac output) changed is unclear how the adequacy of blood flow is affected by
during the 30-minute observation period (Fig 4A). The internal or external shunts when systemic blood pressure
changes in Ees are shown in Figure 4B. The value of Ees deteriorates during retraction or compression of the
did not change significantly during the experiment. heart in OPCABG. Therefore, some active perfusion
Free flow was 35.8 1.16 mL/min. Perfusion flow was systems may be required to obtain stable flow.
30.5 3.50 mL/min, and the decrement in perfusion was
14.9%. Coronary pressure and perfusion flow rate during
the experiment are shown in Figure 5. The coronary
pressure was lower than that in procedure 1, and was not
greater than the physiologic range. Flow pattern was
slightly dominant in the diastric phase (Fig 6).
On pathologic examination none of the pigs showed
tissue edema or vessel injury. Endothelium cells were
well preserved at the coronary artery distal to the can-
nula tip.

Comment
Several methods have been developed for perfusion of
the coronary artery and to avoid myocardial ischemia
during OPCABG. The intracoronary shunt method and
external shunt circuit have been used extensively for this
purpose [3, 4]. These methods are advantageous with
regard to both cost and time because they are inexpen-
sive and can be easily set up as they do not require any
specialized apparatus. These two methods have the same Fig. 6. Waveform of coronary perfusion flow and electrocardiogram.
characteristics given that distal coronary bed perfusion The coronary active perfusion system pattern (upper). Electrocardio-
can be provided passively. However, several studies have gram (lower).
710 NEW TECHNOLOGY KOSHIDA ET AL Ann Thorac Surg
NEW PERFUSION SYSTEM FOR OPCABG 2006;81:706 11

To prevent myocardial ischemia during OPCABG, sev- fusion flow and pressure in humans are necessary to
eral active perfusion methods have been reported. Mu- allow clinical application of this system.
raki and colleagues [8] investigated the efficacy of active
coronary perfusion with a nonpulsatile pump (perfusion-
Disclosures and Freedom of Investigation
assisted direct coronary artery bypass). Recently we re-
ported the CAPS to avoid inadequate blood supply for The microdiaphragm pump and catheter were purchased
ischemic myocardium [6, 7]. These procedures are active by Kanazawa University. The authors have performed a
perfusion techniques and have the characteristic that the free and independent evaluation of this new technology.
amount of blood supply does not depend on hemody- The authors had full control of the design of the study,
namic status during the procedure. Vassiliades and col- methods used, outcome measurements, analysis of data,
leagues [9] reported that active coronary perfusion using and production of the written report.
an in-line pump resulted in superior myocardial protec-
NEW TECHNOLOGY

tion and performance during OPCABG as compared with


References
either no coronary perfusion or passive coronary perfu-
sion. However, these active perfusion systems have a 1. Vassiliades TA Jr, Nielsen JL, Lonquist JL. Hemodynamic
collapse during off-pump coronary artery bypass grafting.
number of disadvantages in that they require large,
Ann Thorac Surg 2002;73:1874 9.
expensive, and complicated settings, and therefore these 2. Grundeman PF, Borst C, van Herwaarden JA, Verlaan CW,
systems have not yet gained widespread acceptance Jansen EW. Vertical displacement of the beating heart by the
among cardiac surgeons. octopus tissue stabilizer: influence on coronary flow. Ann
The diaphragm pump used in the present study was Thorac Surg 1998;65:1348 52.
3. Lucchetti V, Capasso F, Caputo M, et al. Intracoronary shunt
able to supply a stable flow. In addition, this pump prevents left ventricular function impairment during beating
system was small, inexpensive, and consumed little heart coronary revascularization. Eur J Cardiothorac Surg
power, and therefore could be driven by a small battery. 1999;15:2559.
The flow rate controller was omitted to simplify the 4. Arai H, Yoshida T, Izumi H, Sunamori M. External shunt for
off-pump coronary artery bypass grafting: distal coronary
system, but some flow rate adjustment was possible by perfusion catheter. Ann Thorac Surg 2000;70:6812.
adjusting the inner diameter of the circuit. The flow 5. Muraki S, Morris CD, Budde JM, et al. Preserved myocardial
rate was set at approximately 30 mL/min, the value blood flow and oxygen supply-demand balance with active
determined by procedure 1, so that the coronary artery coronary perfusion during simulated off-pump coronary
artery bypass grafting. J Thorac Cardiovasc Surg 2002;123:
pressure was maintained within the physiologic limits.
53 62.
Simplification of the system prevents measurement of 6. Kamiya H, Watanabe G, Doi T, et al. Coronary active
coronary pressure, but if the coronary artery pressure perfusion system can maintain myocardial blood flow and
increases too much, excess blood over the lesional myo- tissue oxygenation. Eur J Cardiothorac Surg 2002;22:410 4.
cardial demand may be washed away from the arteriot- 7. Watanabe G, Kamiya H, Nagamine H, et al. Off-pump
CABG with Synchronized Arterial Flow Ensuring System
omy site. Therefore, it is important to select an appropri- (SAFE-System) Ann Thorac Surg (in press).
ate cannula size. Thus, critical high pressure does not 8. Muraki S, Tsukamoto M, Komatsu K, et al. Minimally
represent a significant concern. ischemic off-pump coronary artery bypass grafting: active
The portable CAPS developed in the present study can perfusion-assist with nitroglycerin-supplemented blood.
Ann Thorac Surg 2003;76:298 300.
achieve adequate perfusion with a constant volume, 9. Vassiliades TA Jr, Nielsen JL, Lonquist JL. Coronary perfu-
nonpulsatile pump. There have been few reports regard- sion methods during off-pump coronary artery bypass: re-
ing small pumps for coronary perfusion. A miniature sults of a randomized clinical trial. Ann Thorac Surg 2002;
vibrating flow pump was reported for external shunt 74:S13839.
10. Kawano S, Isoyama T, Kobayashi S, et al. Miniature vibrat-
catheter [10]. However, its clinical application has not ing flow blood pump using a cross-slider mechanism for
been reported yet. In future studies it will be necessary to external shunt catheter. Artif Organs 2003;27:737.
perform comparisons between our system and such a
small pump.
In conclusion, hemodynamic deterioration and ventric-
Disclaimer
ular arrhythmias during coronary occlusion were pre- The Society of Thoracic Surgeons, the Southern Thoracic
vented by the portable CAPS. Even a simple pump Surgical Association, and The Annals of Thoracic Surgery
system can achieve effective perfusion. Further studies neither endorse nor discourage use of the new technol-
including examination of adequate coronary active per- ogy described in this article.

INVITED COMMENTARY
This article [1] describes a novel active coronary perfu- in myocardium perfused by their pump, and there was
sion system for use in off-pump coronary artery bypass no evidence of acute vascular injury from the intracoro-
(OPCAB) procedures. The authors tested this system in nary cannula.
normal in-situ porcine hearts. They demonstrated stable Other investigators have shown the potential for active
electrophysiologic and contractile function for 30 minutes coronary perfusion systems to diminish ischemic injury

2006 by The Society of Thoracic Surgeons 0003-4975/06/$32.00


Published by Elsevier Inc doi:10.1016/j.athoracsur.2005.07.093

Anda mungkin juga menyukai