than
More s of
ar
20 ye nce
rie
expe
Low Temperature
Logging Instruments
www.antares-geo.de
ANTARES Datensysteme GmbH
Rudolf-Diesel-Str. 6
28816 Stuhr
Germany
Phone: +49 (0) 421 87194-0
Fax: +49 (0) 421 87194-25
Email: info@antares-geo.de
Web: www.antares-geo.de
The specifications contained in this document can be modified without previous announce-
ment. No part of this document may be duplicated or transmitted without explicit written
permission of ANTARES Datensysteme GmbH.
Dear Prospective Client
Over the last 20 years ANTARES has grown from a single small office in Germany to an
international company with offices in Houston and Cairo serving customers in over
30 countries. Close collaboration with our customers has been key to design our product
lines. As a result, we developed low-maintenance and reliable equipment for low-
temperature operations:
Our customers see ANTARES as a partner not only for leading edge technology but also for
training, maintenance and Research & Development. They appreciate the individual and
comprehensive support provided by ANTARES and the reliability of the tools and surface
system.
Together with my team, I hope we will become partners in the future. Discover the 'German
Engineering' difference!
Conrad Blumhagen
ANTARES 2017 1
Low Temperature Instruments
TABLE CONTENT
Tool ID OD Tmax. Cable Page
requirements
Preface 1
Standalone Instruments
Radioactive Instruments
GR 1448 38 mm/43 mm 90C 1- or 4-cond. 5
GR Adapter 1495 43 mm 90C 1- or 4-cond. 6
runs with several other instruments
Spectrum GR 2901 52 mm 60C 1- or 4-cond. 7
Spectrum GR 29N7 70 mm 60C 1- or 4-cond. 8
Slim Dual Detector Neutron 1472 38 mm 90C 1- or 4-cond. 9
Dual Detector Neutron 1497 43 mm 70C 1- or 4-cond. 10
Slim Sidewall Density 29N9 38 mm 90C 1- or 4-cond. 11
Compensated Sidewall Density 1476 52 mm 70C 1- or 4-cond. 12
Combi Instruments
Multipurpose Instrument (MPI) 1143 43 mm 90C 4-cond. 13
GR/Resistivity/Temp/SP/SPR
Combined Flowmeter Tool (CFT) 1631 43 mm/90 mm 100C 1- or 4-cond. 14
SAL/Press/Temp/Flow
Combined Salinity Tool (CST) 1628 43 mm 70C 1- or 4-cond. 15
SAL/Temp/GR
Resistivity Instruments
GR/Focused Electrolog 1125 52 mm 90C 4-cond. 19
Formation Evaluation
4-Arm Dipmeter 12N7 43 mm 100C 4-cond. 20
Magnetic Susceptibility 1108 43 mm 70C 1- or 4-cond. 21
Water Quality
Salinity/Temperature 1623 43 mm 90C 1- or 4-cond. 22
Milieu Monitoring 1707 43 mm 70C/125C 4-cond. 23
Temp/Press/pH/Redox/Cond./O2/Chloride
Continuous Flowmeter 1627 45 mm / 90mm 90C 1- or 4-cond. 24
High Resolution Temperature 16N7 43 mm 100C 1- or 4-cond. 25
Fluid Sampler 0409 43 mm 70C 1- or 4-cond. 26
2 ANTARES 2017
Low Temperature Instruments
TABLE CONTENT
Cavern Logging 38
Furthermore available 45
ANTARES 2017 3
Standalone Tool Product Line
Several instruments combine various sensors in one instrument and allow for easy handling and short
logging time.
As a special feature, we offer a GR Adapter for sevaral instruments to add lithology and depth
determination measurements.
4 ANTARES 2017
Standalone Tool Product Line
Gamma Ray
The ANTARES 1448 Gamma Ray (GR) measures gamma rays emitted by the formation surrounding
the borehole. All detected gamma rays are counted for the total-gamma ray curve. The Gamma
Ray Log can be ran in open and cased holes, in all mud systems and in dry holes. The instrument is
available in two different tool diameters.
Specifications
Diameter: 42 mm/ (1.65") Detector:
38 mm (1.50") GR NaI scintillation crystal
Length: 1,185 mm (46.7)
Weight: 6 kg (13 lbs)
Measuring Parameters
Measuring Range: Accuracy
GR: 0 to 3000 gAPI GR: + 5 % of measured values*
Logging parameters
Recommended Recommended
Min. Hole Diameter: 70 mm (2.8") Logging Speed: 9 m/min (1800 ft/hr)
63 mm (2.5") Sample Rate: selectable
Max. Hole Diameter: NA
Combinability
none
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 5
Standalone Tool and Combinable Radioactive Tool Product Line
GR Adapter
The ANTARES 1495 GR Adapter can be added to several ANTARES Low Temperature instruments of
the Standalone Tools as well as on the Combinable Radioactive Tool Product line to add lithology and
depth determination information. The Gamma Ray Log can be ran in open and cased holes, in all mud
systems and in dry holes.
Specifications
Diameter: 42.4 mm (1.67") Sensors:
Length: 1,594 mm (62.8) NaI Scintillation Crystal
Weight: 8 kg (17 lbs)
Measure Points (from bottom):
Max. Temp: 90C (194F) 1,316 mm (51.8")
Max. Pressure: 20 MPa (2900 psi)
Measuring Parameters:
Measuring Range: Accuracy:
GR: 0 to 3000 gAPI GR: + 5 % of measured values
Resolution:
GR: 3 API
Logging Parameters
Recommended Recommended
Min. Hole Diameter: 65 mm (2.6) Logging Speed: 9 m/min
Max. Hole Diameter: NA (1800 ft/hr)
Sample Rate: selectable
Combinability
With other tools from the ANTARES Standalone and Combinable Radioactive Tool Families.
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
6 ANTARES 2017
Standalone Tool Product Line
The ANTARES 2901 Standalone Spectrum Gamma Ray (SGR) is an instrument logging the natural
gamma radiation. The tool measures the entire gamma ray spectrum from 0 to 3000 keV. It records
the total amount of gamma rays and provides quantitative measurements of the 40Potassium,
232
Thorium and 238Uranium contents of the formation. The Spectrum Gamma Ray Log can be run in
open and cased holes with all mud systems and also in dry boreholes.
Specifications
Diameter: 52 mm (2.05") Detector Type: BGO scintillation crystal
Length: 1,080 mm (42.5")
Weight: 7.5 kg (17 lbs)
Spectrum: 0...3000 keV,
Max. Temp: 60C (140F) 256 Channels
Max. Pressure: 20 MPa (2,900 psi)
Measuring Parameters:
Measuring Range: Accuracy:
GR 0...3000 API GR + 3 % of measured values*
K 0...50 % K, U, Th + 4 % of measured values*
U 0...500 ppm
Th 0...500 ppm
Logging Parameters
Recommended Recommended
Min. Hole Diameter: 70 mm (3) Logging Speed: 3-6 m/min
Max Hole Diameter: NA 590 - 1200 ft/hr
Sample Rate: selectable
Combinability
none
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 7
Standalone Tool Product Line
Specifications
Diameter: 70 mm (2.76") Detector Type: BGO scintillation crystal
Length: 1,400 mm (55.1")
Weight: 24 kg (53 lbs)
Spectrum: 0...3000 keV,
Max. Temp: 60C (140F) 256 Channels
Max. Pressure: 20 MPa (2,900 psi)
Measuring Parameters:
Measuring Range: Accuracy:
GR 0...3000 API GR + 3 % of measured values
K 0...50 % K, U, Th + 4 % of measured values
U 0...500 ppm
Th 0...500 ppm
Logging Parameters
Recommended Recommended
Min. Hole Diameter: 95 mm (3.7) Logging Speed: 3-6 m/min
Max Hole Diameter: NA 590 - 1200 ft/hr
Sample Rate: selectable
Combinability
none
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
8 ANTARES 2017
Standalone Tool Product Line
Specifications
Diameter: 38 mm (1.50") Sensors:
Length: 2,038 mm (80.7) Neutron Detectors:
Weight: 10 kg (22 lbs) Type: Lithium-Glass
No: 2
Max. Temp.: 90C (194F)
241
Max. Pressure: 20 MPa (2900 psi) Source: 3,7 GBq (100 mCi) AmBe
Measuring Parameters:
Measuring Range: Accuracy:
Count rate LS: 0 to 65535 cps + 5 % of the reading
Count rate SS: 0 to 65535 cps
Logging Parameters
Recommended Recommended
Min. Hole Diameter: 63 mm (2.5) Logging Speed: 3-6 m/min
Max. Hole Diameter: 400 mm (15.8") (590 - 1200 ft/hr)
Sample Rate: selectable
Combinability
With 1495 GR Adapter on top.
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 9
Standalone Tool Product Line
Specifications
Diameter: 53 mm (2.09") Neutron Detectors:
Length: 2,529 mm (99.6) Type: Lithium Glas
Weight: 10 kg (22 lbs) No: 2
241
Max. Temp: 70C (158F) Source: 111 GBq (3 Ci) AmBe
Max. Pressure: 30 MPa (4350 psi) (optional 129.5 GBq, 3.5 Ci)
Max Cable Length: 2500 m (8200 ft) Measure Points (from source):
Cable Requirements: mono- or 4-conductor LS 760 mm (29.9")
Telemetry required: no SS 400 mm (15.8")
Data Transmission: digital
Measuring Parameters:
Measuring Range: Accuracy:
Count rate LS: 0 to 65535 cps Count rates: +5 % of the reading
Count rate SS: 0 to 65535 cps Caliper: + 2 mm (0.08")
Caliper: 457 mm (18")
Logging Parameters
Recommended Recommended
Min. Hole Diameter: 78 mm (3.1) Logging Speed:3-6 m/min (590 - 1200 ft/hr)
Max. Hole Diameter: 255 mm (10.0") Sample Rate: selectable
Combinability
With 1495 GR-Adapter on top
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
10 ANTARES 2017
Standalone Tool and Combinable Radioactive Tool Product Line
Specifications
Diameter: 38 mm (1.50") Detectors:
Length: 2,175 mm (85.6) Type: NaI Scintillation Crystal
Weight: 10 kg (22 lbs) No: 2
Max. Temp.: 90C (194F) Source: 3.7 GBq (100 mCi) 137Cs
Max. Pressure: 20 MPa (2900 psi)
Max Cable Length: 2,500 m (8,200 ft) Measure Points (from source):
Cable Requirements:4-conductor LS 622 mm (24.5")
(mono-conductor for SS 318 mm (15.0")
standalone version only)
Telemetry Required: no Accessories:
Data Transmission: digital Decentralizer
Measuring Parameters:
Measuring Range: Accuracy:
Count rate LS: 0 to 65535 cps +5 % of the reading
Count rate SS: 0 to 65535 cps
Logging Parameters
Recommended Recommended
Min. Hole Diameter: 63 mm (2.5) Logging Speed: 3-6 m/min
Max. Hole Diameter: 400 mm (15.8") (590 - 1200 ft/hr)
Sample Rate: selectable
Combinability
With ANTARES 1495 GR and Dual Detector Neutron tools from the ANTARES Combinable
Radioactive Tool Family.
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 11
Standalone Tool Product Line
Sidewall Density
The ANTARES 1476 Sidewall Density instrument includes a Cesium source emitting gamma rays and
two detectors. From the energy spectrum and count rates of the two counters, the borehole
compensated formation bulk density is calculated. A caliper measurement is provided, permitting
compensation for borehole effects. The tool can be run in any mud. A correction curve indicates the
influence of the borehole rugosity.
Specifications
137
Diameter (max): 53 mm (2.09") Source Type: Cs
Length: 2598 mm (102.3) Recommended Source Activity:
Weight: 30 kg (67.5 lbs) 3.7 to 11.1 GBq (0.1 to 0.3 Ci)
Detector Type: NaI Scintillation
Max. Temp: 70C (158F) Crystal
Max. Pressure: 20 MPa (2900 psi) No of Detectors: 2
Max Cable Length: 2500 m (8200 ft) Measure Points (from bottom):
Cable Requirements: 1- or 4-Conductor DEN: 258 mm (10,16)
Telemetry required: no (depending on source activity)
Data Transmission: digital Caliper 862 mm (33,94)
Measuring Parameters
Measuring Range: Accuracy:
DEN: 1-3 g/cm3 DEN +0.1 g/cm3
Caliper: 53-457 mm (18) (for 2 - 3 g/cm3 formation
density)
Logging parameters
Recommended Recommended
Min. Hole Diameter: 78 mm (3.1) Logging Speed: 6 m/min (1200 ft/hr)
Max. Hole Diameter: 457 mm (18)
Combinability
With 1495 GR-Adapter.
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
12 ANTARES 2017
Standalone Tool Product Line
The ANTARES 1143 Multipurpose Instrument measures various parameters (Gamma Ray/
Resistivity/Temperature) providing mud and surrounding formation parameters. As it logs
several measurements in one run it can save considerable rig time.
Specification
Diameter 42 mm (1.65") Sensors:
Length: 7,940 mm (312.6") Gamma Ray
42 mm
w/o GR, Temp 7,500 mm (295.3") Spontaneous Potential
Weight: 26 kg (57 lbs) Resistivity: ES 16/64
Temperature
3670 mm
Max. Temperature: 100 C (212 F) Single Point Resistance
Max. Pressure: 20 MPa (2900 psi)
Measure points (from bottom):
Max Cable Length: 2500 m (8200 ft) GR 2,506 mm (98.7")
Cable Requirements:4-conductor Temperature 2,300 mm (90.6")
Telemetry required: no SP 1,780 mm (70.1")
Data Transmission: digital Long Normale 1,780 mm (70.1")
Short Normale 560 mm (22,0")
Top Connector: GO4 or GC4 SPR 135 mm (5.3")
Bottom Connector: none
Accessories:
Bridle (always included)
2130 mm
Measuring Parameters GR
Measuring range Accuracy
Temperature: -5 to 80 C Temperature: + 0.4 C
Gamma Ray: 0 to 3000 API GR: +3 % TEMP
SP: -10,000 to 10,000mV SP: +1 %
SPR 0.4 to 10,000 OHmm SPR: + 0.01 %
16/64 Resistivity (Short/Long Resistivity: + 0.1 OHmm +0.01.%
Normal):
0.2 to 10,000 Ohmm Resolution
Temperature 0.01 C
GR: 3 API Long
SP: 0.3 mV Normal,
SPR: 0.1 Ohmm SP
2085 mm
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 13
Standalone Tool Product Line
Accuracy
Conductivity +2 % PRES
Temperature + 0,2 C
Pressure + 0.1 %
Logging parameters
Recommended Recommended
Min. Hole Diameter: 70 mm (2.8) Logging Speed: 3-6 m/min
Max Hole Diameter: NA 590 - 1200 ft/hr
Sample Rate: selectable Centralizer
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
14 ANTARES 2017
Standalone Tool Product Line
SAL/TEMP/GR
The ANTARES 1628 Conductivity/Temperature/GR Instrument provides fluid temperature and con-
ductivity measurements. From fluid conductivity the fluid salinity can be calculated. The log can be
recorded upwards or downwards to minimize fluid disturbance. For correlation purpose the tool
includes a Gamma Ray scintillation detector, measuring the total gamma rays emitted by the
surrounding formation.
Specification
Diameter 43 mm (1.69 ) Sensors:
Length: 1,663 mm (65.47") Temperature
Weight: 7.1 kg (3.2 lbs) Fluid Conductivity
Gamma Ray
Max. Temperature: 70 C (158 F)
Max. Pressure: 20 MPa (2900 psi) Measure points (from bottom):
Temperature 63 mm (2.48")
Max. Cable Length: 2500 m (8200 ft) Fluid Conductivity 188.5 mm (7.42")
Cable Requirements: mono- or 4-conductor GR 560 mm (22.05")
Telemetry required: no
Data Transmission: digital
Measuring Parameters
Measuring range Accuracy
Temperature 5 to 80 C Conductivity +1 %
41 to 176 F Temperature + 0.1 C
Conductivity I 0 to 3 mS/cm GR + 3 % of measured
Conductivity II 0 to 30 mS/cm values
Conductivity III 0 to 300 mS/cm Resolution
GR 0 to 10,000 cps Conductivity I 0,001 mS/cm
0 to 9,000 gAPI Conductivity II 0.01 mS/cm
Conductivity III 0.1 mS/cm
Temperature 0.01 C
Logging parameters
Recommended: Recommended:
Min. Hole Diameter: 75 mm (3") Logging Speed GR: 6 m/min
Max Hole Diameter: NA (1200 ft/hour)
Sample Rate: selectable
Combinability
Application
With 1495 GR Adapter
ANTARES is continually updating and improving its high performance and quality logging system and instruments. For this reason the
specifications are subject to change without prior notice.
ANTARES 2017 15
Standalone Tool Product Line
3-Arm Caliper
The ANTARES 1014 Caliper tools is a motorized 3-Arm instruments. The measurements cover a
range between 43 mm and 700 mm (1.7" and 27.6").
Specifications
Diameter: 43 mm (1.69") Sensors: Potentiometric Linear
Length: 1,940 mm (76.4") Transducer
Weight: 12 kg (26 lbs)
Measure Points (from bottom):
Max. Temp: 100C (212F) 135 mm (5.3)
Max. Pressure: 20 MPa (2,900 psi)
Measuring Parameters:
Measuring Range: 43 - 700 mm Accuracy: + 2 mm (0.08")
(1.7"-27.6") Resolution: 0.1 mm (0.004")
Logging parameters
Recommended: Recommended:
Min. Hole Diameter: 68 mm (2.7") Logging Speed: 10 m/min (1950 ft/hr)
Max. Hole Diameter: 700 mm (27.6") Sample Rate: selectable
Combinability
With 1495 GR Adapter on top
ANTARES is continually updating and improving its high performance and quality logging system and instruments. For this reason the
specifications are subject to change without prior notice.
16 ANTARES 2017
Standalone Tool Product Line
4-Arm Caliper/Orientation
The ANTARES 10N7 Caliper tools is a motorized 4-Arm instruments. The X-Y caliper accurately
measures in two axes of the well bore for borehole diameters up to 1,000 mm (39.4 in). Together with
the orientation module the shape and direction of the borehole can be measured.
Specifications
Diameter: 52 mm (2.05") Sensors:
Length: 2,311 mm (91.0") Caliper: Potentiometric Linear
Weight: 18 kg (40 lbs) Transducer
Orientation : Magnetometer
Max. Temp: 70C (158F) Inclinometer
optional available: 100C (212F)
Max. Pressure: 20 MPa (2,900 psi) Measure Points (from bottom):
Caliper: 105 mm (4.1)
Max Cable Length: 2,500 m (8,200 ft) Orientation: 2,156 mm (84.9")
Cable Requirements: mono- or 4-
conductor
Telemetry required: no
Data Transmission: digital
Measuring Parameters:
Measuring Range: Accuracy:
Caliper: 55 to 700 mm Caliper: + 2 mm (0.08")
(2.2"-27.6") Deviation: + 0.5
w/extensions: to 1000 mm Azimuth: +3%
(39.4")
Deviation: 0 to 60 Resolution:
Azimuth: 0 to 360 Caliper: 0.1 mm (0.004")
Deviation: 0.1
Azimuth: 0.1
Logging parameters
Recommended: Recommended:
Min. Hole Diameter: 77 mm (3.0") Logging Speed: max. 10 m/min
Max Hole Diameter: 1000 mm (39.4") 1950 ft/hr
Sample Rate: selectable
Combinability
With 1495 GR-Adapter
ANTARES is continually updating and improving its high performance and quality logging system and instruments. For this reason the
specifications are subject to change without prior notice.
ANTARES 2017 17
Standalone Tool Product Line
Orientation
The ANTARES 1212 Orientation Instrument allows for detailed measurement of the borehole
direction. From the inclinometer and magnetometer measurements the azimuth and the deviation
are calculated. The tool can be run in open and cased holes (deviation only) and is independent from
the mud system.
Specifications
Diameter: 43 mm (1.69") Sensors:
Length: 1,048 mm (41.3") Orientation : Magnetometer
Weight: 5 kg (11 lbs) Inclinometer
Orientation Instrument
Max. Pressure: 20 MPa (2,900 psi) Orientation: 261 mm (10.3")
Measuring Parameters:
Measuring Range: Accuracy:
Deviation: 0 to 180 Deviation: + Meas. Value x 0.001+
Azimuth: 0 to 360 0.7
Azimuth: + Meas. Value x
0.001+1.2
Resolution:
Deviation: 0.01
Azimuth: 0.1
Logging parameters
Recommended: Recommended:
Min. Hole Diameter: 68 mm (2.7") Logging Speed: max. 10 m/min
Max Hole Diameter: NA 1950 ft/hr
Sinker Bar
Combinability
With 1495 GR-Adapter
ANTARES is continually updating and improving its high performance and quality logging system and instruments. For this reason the
specifications are subject to change without prior notice.
18 ANTARES 2017
Standalone Tool Product Line
GR/Focused Electrolog
The ANTARES 1125 Gamma Ray Focused Electrolog measures gamma ray and resistivity parameters.
The focused electrolog works with one source electrode and two guard electrodes to focus the current
into the formation. It results in a high vertical resolution. The GR detector measures the natural
gamma radiation emitted from the surrounding formation.
Specification
Diameter 42.4 mm (1.67") Sensors:
Length: 5,327 mm (209.7") GR Scintilation Crystal
Transport Length: 2,210 mm (87.2") FEL Electrodes
Weight: 13.8 kg (30 lbs)
Measure points (from bottom):
3000 mm
Bridle
Max. Temperature: 90 C (194 F) GR 151 mm (5.9")
Max. Pressure: 20 MPa (2900 psi) FEL 996 mm (39.2")
Measuring Parameters
Measuring range Accuracy
Gamma Ray: 0 - 3000 API GR: + 3 % of measured
FEL1: 0 - 5,000 Ohmm values
5,327 mm
FEL2: 0 - 2,000 Ohmm FEL1: + 1%
FEL2: + 1%
Resolution
FEL1: 0.1 Ohmm
FEL2: 0.5 Ohmm
GR: 3 API
FEL
Logging parameters
Recommended Recommended
Min. Hole Diameter: 68 mm (2.7") Logging Speed: 10 m/min (2000 ft/hr)
Max Hole Diameter: NA
Combinability
Application
none
Application
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 19
Standalone Tool Product Line
4-Arm Dipmeter
The ANTARES 12N7 4-Arm Dipmeter is a pad-type device to detect changes in formation resistivity
which are obtained from short spaced focused electrodes pressed against the borehole wall.
Simultaneous measurements are recorded at each of the four pads. The curves are later correlated in
order to determine dip angles relative to the borehole. All curves including the caliper are used to
determine the true formation dip.
Specifications
Diameter: 52 mm (2.05") Sensors:
Length: 2,650 mm (104.3) Pad resistivity
Weight: 12 kg (26 lbs) Caliper
3-axis Accelerometer
Max. Temp: 70C (158F) 3-axis Magnetometer
Max. Pressure: 20 MPa (2900 psi) GR (optional)
Max. Cable Length: 2,500 m (8,200 ft) Measure Points (from bottom):
Cable Requirements: 4-conductor Resistivity: 580 mm (22.8")
Telemetry required: no
Data Transmission: digital
Logging Parameters
Recommended
Min. Hole Diameter: 65 mm (2.6) Recommended
Max. Hole Diameter: 250 mm Logging Speed: 6 m/min
(9.8") (900 ft/hr)
Combinability
none
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
20 ANTARES 2017
Standalone Tool Product Line
Magnetic Susceptibility
The ANTARES 1108 Magnetic Susceptibility instruments is a standalone logging tool. The
determination of magnetic susceptibility is widely used in mineral exploration and lithology
evaluation.
Specification
Diameter 43 mm (1.69") Sensors:
Length: 1,500 mm (59,1") Three coil systems
Weight: 7 kg (15 lbs)
Measure point (from bottom):
Max. Temperature: 70 C (158 F) 230 mm (9.1")
Max. Pressure: 20 MPa (2,900 psi)
Measuring Parameters
Measuring range : Resolution
1 x -10-5 SI Units to 1 x 10-1 SI-Unit -5
10 SI-Unit
Two Ranges Temperature Compensation
for 0 to 60C
Logging parameters
Recommended Recommended
Min. Hole Diameter: 78 mm (3.07") Logging Speed: 6 m/min
Max Hole Diameter: NA 1200 ft/hr
Combinability
Application
none
Application
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 21
Standalone Tool Product Line
SAL/TEMP
The ANTARES 1623 Fluid Conductivity/Temperature Instrument measures the electric resistivity of
water in the borehole. The salinity of the water can be determined from the conductivity readings.
Together with the temperature logs, The combination of the Temperature and salinity logs
provides an indication of vertical movement of the water.
Specification
Diameter: 42.4 mm (1.67") Sensors:
Length: 1,281 mm (50.4") Temperature: Pt 1000
Transport Length: 1,306 mm (51.4") Conductivity: 4 electrodes
Weight: 5.3 kg (12 lbs) conductivity
optional: GR NaI scintillation
Max. Temperature: 100 C (212 F) crystal
Max. Pressure: 20 MPa (2,900 psi)
Measure points (from bottom):
Max. Cable Length: 2,500 m (8,200 ft) Conductivity: 187 mm (7.4")
Cable Requirements: mono- or 4-Conductor Temperature: 50 mm (2.0")
Telemetry required: no
Data Transmission: digital
Measuring Parameters:
Measuring Range Resolution
Temperature 0 to 100 C Temperature 0.01 C0
Conductivity I 0 to 3 mS/cm Conductivity I 0.001 mS/cm
Conductivity II 0 to 30 mS/cm Conductivity II 0.01 mS/cm
Conductivity III 0 to 250 mS/cm Conductivity III 0.1 mS/cm
Accuracy
Conductivity +2 %
Temperature +0.2 C
COND
Logging parameters
Recommended Recommended
Min. Hole Diameter: 68 mm (2.7) Logging Speed: 3-6 m/min
Max Hole Diameter: NA 590 - 1200 ft/hr TEMP
Sample Rate: selectable
Combinability
With 1495 GR Adapter on top
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
22 ANTARES 2017
Standalone Tool Product Line
Milieu Monitoring
The 1707 ANTARES Milieu Monitoring Instrument measures parameters describing the borehole
fluid, such as pressure, temperature, fluid conductivity, oxygen and pH.
Specifications
Diameter: 43 mm (1.69") Sensors:
Length: 1.470 mm (57.9) Temperature
Weight: 6 kg (13 lbs) Pressure
Conductivity (Salinity)
Communication section
Max. Temp.: 50C (122F) O2
Max. Pressure: 15 MPa (2175 psi) pH
Redox
Max Cable Length: 2,500 m (8,200 ft)
Cable Requirements: 4-Conductor Optional sensors:
Telemetry Required: no Chloride
Data Transmission: digital GR
Measuring Parameters:
Measuring range Accuracy
Pressure 0 to 1000 dbar Oxygen + 0.1 ppm or
Temperature -1 to 50 C +1% saturation
Cond. Salt Water 0 to 64 mS/cm pH + 0.01 pH
Cond. Fresh Water 0 to 6400 S/cm Redox +1 mV
Oxygen 0 to 50 ppm
0 to 500 % sat. Resolution
pH 0 to 14 pH Pressure 0.0015 % full scale
Redox +/- 1000 mV Temperature 0.001 C
Cond.Salt Water 0.001 mS/cm
Accuracy Cond. Fresh Water 0.1 S/cm
Pressure + 0.05 % full scale Oxygen 0.01 ppm
Temperature + 0.005 C 0.1 % sat.
Sensor section
Cond. Salt Water + 0.007 mS/cm pH 0.001 pH
Cond. Fresh Water +5 S/cm Redox 0.1 mV
Logging Parameters
Recommended Recommended
Min. Hole Diameter: 63 mm (2.5) Logging Speed: 3-6 m/min
Max. Hole Diameter: NA (590 - 1200 ft/hr)
Sample Rate: selectable
Combinability
none
ANTARES
ANTARES is continually updating and improving its high performance and2017 23
quality logging system and for this reason the specifications
are subject to change without prior notice.
Standalone Tool Product Line
Continuous Flowmeter
The ANTARES 1627 Continuous Flowmeter is run to monitor downhole fluid flow rates. Exchangeable
heads allow to choose a right diameter to optimize the measurement according to the borehole
diameter. The tool can determine between upward and downward flow direction.
Specifications
Diameter Sensors: Spinner
Electronics section: 42 mm (1.65)
Spinner: 45 mm (1.77") Measure points (from bottom):
(exchangable) 90 mm (3.54") 45 mm spinner 92 mm (3.62")
Length: 1,748 mm (68.8") 90 mm spinner 111 mm (4.37")
Weight: 8.5 kg (19 lbs)
Measuring Parameters
Measuring range
Flow: 0.008 m sec to 5 m/sec
depending on cage size
Logging parameters
Recommended: Recommended:
Min. Hole Diameter: 70 mm (2.8") Logging Speed: max. 10 m/min
Max Hole Diameter: 250 mm (9.8") 1950 ft/hr
Sample Rate: selectable
Combinability
With 1495 GR Adapter on top
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
24 ANTARES 2017
Standalone Tool Product Line
The ANTARES 16N7 Temperature Instrument is used to determine temperature profiles with
resolutions of + 0.01 C and can detect even small changes in temperature.
Specification
Diameter 43 mm (1.69") Sensors:
Length: 1,005 mm (39.57") Temperature Pt 1000
Weight: 8 kg (18 lbs)
Measure points (from bottom)
Max. Temperature: 100 C (212 F) Temperature 20 mm (0.79")
Max. Pressure: 20 MPa (2900 psi)
Measuring Parameters:
Measuring Range: -5 to 100 C (212F)
Accuracy: +0.2 C
Resolution: 0.01C
Logging Parameters
Recommended Recommended
Min. Hole Diameter: 65 mm (2.6) Logging Speed: 6 m/min
Max. Hole Diameter: NA (900 ft/hr)
Sample Rate: selectable
Combinability
Application
none
Application
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 25
Standalone Tool Product Line
Fluid Sampler
ANTARES 0409 Fluid Sampler is designed to take fluid samples even under difficult borehole
conditions. Several tank sizes are availbale for the sample fluids and robust mechanism allows for
simple and trouble free operations.
Specification
Diameter Electronics: 42.4 mm (1.67") Sensors: Fluid Sensor
Fluid Sensor
Length Electronics: 800 mm (31.5")
Weight Electronics: 3.5 kg (8 lbs) Measure Points (from Top):
Fluid Sensor: 50 mm (1.97")
Max. Temperature: 70C (158F)
Max. Pressure: 20 MPa (2900 psi)
Electronic
Max. Cable Length: 2,500 m (8,200 ft) Section
Cable Requirements: mono- or
4-conductor
Telemetry required: no
Top Connector: GC1, GO1, GC4 or
GO4
Bottom Connector: none
Logging parameters
Recommended Recommended
Min. Hole Diameter: 70 mm (2.7) Logging Speed: stationary
Max Hole Diameter: NA Sample Rate: NA
Combinability
Application
none
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
26 ANTARES 2017
Combinable Low-Temperature Instruments
ANTARES offers a combinable string of nuclear instruments, which allows the measurement of GR,
neutron and density logs in just one run. The outer diameter of the string is only 43 mm and therefore,
it enables for measurements in very slim boreholes or casings. These tools can be run as standalone
or in combination.
The product line of ANTARES Telemetry low-temperature instruments comprises a set of tools, which
are run below a telemetry with Gamma Ray. These comprehensive low-temperature instruments
enable a high degree of flexibility. Several instruments can be combined in one string, reducing rig
time where every minute matters. These tools are designed according to oilfield standards and
measurements are accurate against the industry standards.
These tools have been field tested in offshore environments and proved their reliability and easy
handling ability. The pressure rating is up to 50 MPa (7,250 psi).
ANTARES 2017 27
Standalone Tool and Combinable Radioactive Tool Product Line
GR Adapter
The ANTARES 1495 GR Adapter can be added to several ANTARES Low Temperature instruments of
the Standalone Tools as well as on the Combinable Radioactive Tool Product line to add lithology and
depth determination information. The Gamma Ray Log can be ran in open and cased holes, in all mud
systems and in dry holes.
Specifications
Diameter: 42.4 mm (1.67") Sensors:
Length: 1,594 mm (62.8) NaI Scintillation Crystal
Weight: 8 kg (17 lbs)
Measure Points (from bottom):
Max. Temp: 90C (194F) 1,316 mm (51.8")
Max. Pressure: 20 MPa (2900 psi)
Measuring Parameters:
Measuring Range: Accuracy:
GR: 0 to 3000 gAPI GR: + 5 % of measured values
Resolution:
GR: 3 API
Logging Parameters
Recommended Recommended
Min. Hole Diameter: 65 mm (2.6) Logging Speed: 9 m/min
Max. Hole Diameter: NA (1800 ft/hr)
Sample Rate: selectable
Combinability
With other tools from the ANTARES Standalone and Combinable Radioactive Tool Families.
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
28 ANTARES 2017
Standalone Tool and Combinable Radioactive Tool Product Line
Specifications
Diameter: 38 mm (1.50") Sensors:
Length: 2,038 mm (80.7) Neutron Detectors:
Weight: 10 kg (22 lbs) Type: Lithium-Glass
No: 2
Max. Temp.: 90C (194F)
241
Max. Pressure: 20 MPa (2900 psi) Source: 3,7 GBq (100 mCi) AmBe
sideloaded
Max Cable Length: 2,500 m (8,200 ft)
Cable Requirements: 4-conductor Measure Points (from source):
Telemetry Required: no LS 372 mm (14.7")
Data Transmission: digital SS 172 mm (6.8")
Measuring Parameters:
Measuring Range: Accuracy:
Count rate LS: 0 to 65535 cps + 5 % of the reading
Count rate SS: 0 to 65535 cps
Logging Parameters
Recommended Recommended
Min. Hole Diameter: 63 mm (2.5) Logging Speed: 3-6 m/min
Max. Hole Diameter: 400 mm (15.8") (590 - 1200 ft/hr)
Sample Rate: selectable
Combinability
With other tools of the Combinable Radioactive Tool Family
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 29
Standalone Tool and Combinable Radioactive Tool Product Line
Specifications
Diameter: 38 mm (1.50") Detectors:
Length: 2,175 mm (85.6) Type: NaI Scintillation Crystal
Weight: 10 kg (22 lbs) No: 2
Max. Temp.: 90C (194F) Source: 3.7 GBq (100 mCi) 137Cs
Max. Pressure: 20 MPa (2900 psi)
Max Cable Length: 2,500 m (8,200 ft) Measure Points (from source):
Cable Requirements:4-conductor LS 622 mm (24.5")
(mono-conductor for SS 318 mm (15.0")
standalone version only)
Telemetry Required: no Accessories:
Data Transmission: digital Decentralizer
Measuring Parameters:
Measuring Range: Accuracy:
Count rate LS: 0 to 65535 cps +5 % of the reading
Count rate SS: 0 to 65535 cps
Logging Parameters
Recommended Recommended
Min. Hole Diameter: 63 mm (2.5) Logging Speed: 3-6 m/min
Max. Hole Diameter: 400 mm (15.8") (590 - 1200 ft/hr)
Sample Rate: selectable
Combinability
With ANTARES 1495 GR and Dual Detector Neutron tools from the ANTARES Combinable
Radioactive Tool Family.
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
30 ANTARES 2017
Combinable Low-Temperature Tool Product Line
Telemetry / GR
The ANTARES 0921 Digital Telemetry instrument handles the communication between the ALADIN
surface system and ANTARES digital downhole instruments in both directions. It also monitors the
temperature of the telemetry electronics and the cable head voltage. The tool includes a Gamma Ray
scintillation detector, measuring the total gamma rays emitted by the surrounding formation. The
Gamma Ray Log (GR) can be ran in open and cased holes, in all mud systems and in dry holes.
Specifications
Measuring Parameters
Measuring Range: Accuracy:
GR: 0 to 3000 gAPI GR: + 5 % of measured values
Logging parameters
Recommended: Recommended:
Min. Hole Diameter: 70 mm (2.8") Logging Speed GR: 6 m/min
Max Hole Diameter: NA (2800 ft/hour)
Sample Rate: selectable
Combinability
With all tools from the ANTARES Combinable Low-Temperature Tool Product Line
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 31
Combinable Low-Temperature Tool Product Line
Specifications
Diameter: 52 mm (2.05") Detector Type: BGO Scintillation Crystal
Length: 1,110 mm (43.7)
Weight: 9.2 kg (21 lbs)
Measure Point (from bottom):
Max. Temp: 60C (140F) 257.5 mm (10.1")
Max. Pressure: 50 MPa (5,800 psi)
Spectrum: 60...3000 keV,
Telemetry required: yes, type 921 or 924 256 Channels
Top Connector: ANTARES Toolbus
Bottom Connector: ANTARES Toolbus
Measuring Parameters:
Measuring Range: Accuracy:
GR 0...3000 APi GR + 3 % of measured
K 0...50 % values
U 0...500 ppm K, U, Th + 4 % of measured
Th 0...500 ppm values
Logging Parameters
Recommended Recommended
Min. Hole Diameter: 78 mm (3) Logging Speed: 3 m/min
Max Hole Diameter: NA 590 ft/hr
Sample Rate: selectable
Combinability
With all tools from the ANTARES Combinable Low-Temperature Tool Product Line
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
32 ANTARES 2017
Combinable Low-Temperature Tool Product Line
Specifications
Diameter (max): 52 mm (2.05") Detectors:
Length: 1750 mm (68.9) Type: He-3
Weight: 30 kg (66 lbs) No: 2
Max. Temp: 70C (158F) Source: 130 GBq (3.5 Ci) 241 AmBe
Max. Pressure: 50 MPa (7,250 psi) side loadaed
Telemetry required: yes, type 921 or 924 Measure Points (from bottom):
Top Connector: ANTARES Toolbus CN: 547 mm (21.5")
Bottom Connector: ANTARES Toolbus
Measuring Parameters
Measuring Range: 0-100 % Porosity Accuracy: + 10 %
Logging parameters
Recommended Recommended
Min. Hole Diameter: 70 mm (2.76) Logging Speed: 6 m/min
Max Hole Diameter: 250 mm (18) (1200 ft/hr)
Combinability
With all tools from the ANTARES Combinable Low-Temperature Tool Product Line
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 33
Combinable Low-Temperature Tool Product Line
Sidewall Density
The ANTARES 1491 Sidewall Density instrument includes a Cesium source and two detectors. From
the energy spectrum and count rates of the two counters, the borehole compensated formation bulk
density is calculated. A caliper measurement is provided, permitting compensation for borehole
effects. The tool can be run in any mud except in barite mud, which absorbs the Gamma Rays and
therefore affects the density reading heavily. A Correction Curve indicates the influence of the
borehole rugosity.
Specifications
Diameter (max): 53 mm (2.09") Detectors
Length: 2598 mm (102.3) Type: NaI Scintillation Crystal
Weight: 30 kg (67.5 lbs) No: 2
Max. Temp: 70C (158F) Source: 3,7 GBq (0,1 Ci) 137Cs
Max. Pressure: 50 MPa (7250 psi)
Measure Points (from bottom):
Telemetry required: yes, type 921 or 924 DEN: 258mm (10,16)
Top Connector: ANTARES Toolbus (depending on source activity)
Bottom Connector: ANTARES Toolbus Caliper 862 mm (33,94)
Measuring Parameters
Measuring Range: Accuracy:
DEN: 1-3 g/cm3 DEN 0.02 g/cm3
Caliper: 53-457 mm (18) (for 2 - 3 g/cm3 formation
density)
Logging parameters
Recommended Recommended
Min. Hole Diameter: 70 mm (2.8) Logging Speed: 6 m/min
Max Hole Diameter: 457 mm (18) (1200 ft/hr)
Combinability
With most tools from the ANTARES Combinable Low-Temperature Tool Product Line
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
34 ANTARES 2017
Combinable Low-Temperature Tool Product Line
Acoustic Instrument
The ANTARES 1313 Borehole Acoustic instrument is designed for acquiring high-resolution, full-
wave acoustic data. One ceramic monopole transmitter emits the acoustic signal which is detected by
two ceramic receivers. During one logging run, the formation travelling time in the open-hole interval
and the cement bond log in the cased-hole section can be recorded.The log requires a fluid filled
borehole but is independent of the mud type.
Specifications
Diameter: 43 mm (1.69") Transmitter: Ceramic, monopole
Length: 3,865 mm (156) Sensors: Ceramic transducer
Transport Length: 2 sections array
max. 2,300 mm
Weight: 15 kg (23 lbs) Transmitter-Receiver Spacing:
T1-R1 : 932 mm (3 ft)
Max. Temp: 70C (158F) R1-R2: 305 mm (1 ft)
Max. Pressure: 50 MPa (7,250 psi)
Measure Points (from bottom):
Telemetry required: yes, type 921 or 924 Delta T 917 mm (36.1")
Top Connector: ANTARES Runtime Far 765 mm (30.1")
Bottom Connector: ANTARES Runtime Near 1,070 mm (42.1")
Measuring Parameters
Measuring Range: Accuracy
DT: 100-800 s/m DT*: + 3 s/m (+1s/ft)
(35-250 s/ft) * centralized tool in a 100 mm diameter
borehole
Logging parameters
Recommended Recommended
Min. Hole Diameter:68 mm (2.7") Logging Speed: 6-15 m/min
Max Hole Diameter: 400 mm (15.4") 1200- 3000 ft/hr
Combinability
With all tools from the ANTARES Combinable Low-Temperature Tool Product Line
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 35
Combinable Low-Temperature Tool Product Line
Dual Laterolog
Cable Head
GO4
measuring formation resistivities in boreholes with conductive borehole fluids.
The Dual Laterolog measures both the Deep (RLD) and the Shallow Laterolog
resistivity (RLS). A SP measurement is also provided.
Specifications
523 IsoSub
Diameter: 43 mm (1.69") Type: LL3
Length: 2,343 mm (92.2)
SP Electrode
Weight: 16 kg (35 lbs) Measure Points
(from bottom of 1113 DLL)):
Max. Temp: 70C (158F) Deep: 910 mm (35.8")
524 IsoSub
Max. Pressure: 50 MPa (7,250 psi) Shallow 190 mm (7.5")
Telemetry required: yes, type 921 or 924 (from bottom of 524 IsoSub)
Top Connector: ANTARES Toolbus SP 1,990 mm (78.4")
Bottom Connector: ANTARES Toolbus
0921 Telemetry
Measuring Parameters
Measuring Range: Accuracy:
DLL, SLL 0.2 to 40,000 Ohmm DLL 0.2 to 10,000 Ohmm: 2 %
SP -1000 mV to 1000 mV SP 1 to 40,0000 Ohmm: 10 %
Intergrated IsoSub
Logging parameters
Recommended Recommended
Min. Hole Diameter: 70 mm (2.76) Logging Speed: 25 m/min
Max. Hole Diameter: 250 mm (9.84) (1200 ft/hr)
IsoSubs required
0523 IsoSub w/ SP Length: 2,002 mm (78.82")
590 IsoSub
Combinability
With most tools from the ANTARES Combinable Low-Temperature Tool Product Line
Instrument
Further
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
36 ANTARES 2017
Combinable Low-Temperature Tool Product Line
4-Arm Caliper
The ANTARES 1005 4-Arm Caliper tool is a motorized instrument. All four arms provide individual
measurements. The four caliper signals are generated by using high-resolution precision magnetic
sensors and cover a range between 52 and 800 mm (2.0 and 31.5 inch). From the four independent radii
measurements an average borehole diameter and X - Y average diameters can be calculated.
Specifications
Diameter: 52 mm (2.05") Sensors: Magneto Resistivity
Length: 2,177 mm (85.7) Angle Sensors
Weight: 15 kg (123 lbs)
Measure Points (from bottom):
Max. Temp: 80C (176F) 77 mm (3.0")
Max. Pressure: 50 MPa (7250 psi)
Measuring Parameters
Measuring Range: Accuracy:
Cal: 52-800 mm + 2% (0.08")
(2.1-31.5")
Resolution:
0.1 mm (0.004")
Logging parameters
Recommended Recommended
Min. Hole Diameter:75 mm (3.0") Logging Speed: 6-15 m/min
Max Hole Diameter: 800 mm (31.5") 1200- 3000 ft/hr
Combinability
With most tools from the ANTARES Combinable Low-Temperature
Tool Product Line
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 37
Cavern and Low-Temperature Cased-Hole Logging
Three-dimensional measure-
ment and shape determination
of an underground cavern. The
ANTARES DSC instrument has
been accepted by international
mining companies and Euro-
pean mining authorities.
Further cased hole instruments for applications in all temperature ranges you will find in our catolog
Cased hole instruments.
38 ANTARES 2017
Cavern and Low-Temperature Cased Hole Tools
Specifications
Diameter: 72 mm (2.83") Top Connector: GC4
Length: 3,600mm (142) Bottom Connector: none
Transport Length: max. 2000 mm (79")
Weight: 60 kg (133 lbs) Sensors:
Transducers 2 piezo ceramic
temperature sensor
Max. Temp: 70 C (160F) Magnetometer
Gyro 50C (120F) Inclinometer
velocimeter and
Max. Pressure: 30 MPa (4,350 psi) Velocimeter
Gyro
Max Cable Length: 2,500 m (8,200ft) Dual string collar locator (DSCL)
Cable requirements: 4-conductor Temperature
Telemetry required: no Pressure
Data transmission: digital
Measuring Parameters
electronics
Measuring range: Accuracy:
main
Caliper (open cavern): 2 to 200 m Caliper: 0.018 m + 0.09 %
Caliper (1 pipe): 2 to 67 m of the distance
Caliper (2 pipe): 2 to 60 m Beam Width : 1.0 (-3 db)
(depending on the concentricity of Side-Echo: q-20 db of the main
transportation
breakpoint for
pipes and physical conditions) echo
Acoustic frequencies: 250...750 kHz Pressure: 0.1 MPa
Fluid sound velocity: 1300-1900m/s Temperature: 0.1C
Azimuth: 0-360 Fluid velocity: <2m/s
Inclination: 0-30
Pressure: 0-40 Mpa Resolution:
Temperature: 0-85 C Caliper: 0.1 m
DSCL: 2 strings of Azimuth: 1
casing Inclination: 0.5
Pressure: 0.01 MPa
fluxgate
Temperature: 0.01 C
gyro
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 39
Directional Sonar Caliper
Application
The main application of the Directional Sonar Caliper log is the determination of the shape of the
cavern. The log data can be exported to 3rd-party software to visualize the position and shape of it.
ALADIN aquisitions and
Processing
procession system
at Customer
(3D-Processing
Software
Package)
Function Overview
Ground
Instrument function
3
The configuration of the tool with a tilting arm and
2
two transducers scanning different directions
1 Measurement
allow the client to produce 3-dimentional figures
principle
of caverns.
ANTARES is continually updating and improving its high performance and quality logging system and for this reason the specifications
are subject to change without prior notice.
40 ANTARES 2017
Cavern and Low-Temperature Cased Hole Tools
1509 DSCL/GR/TEMP
Specification
Diameter 73 mm (2.87) Sensors:
Length: 2900 mm (114.17") Temperature
Weight: 30 kg (66 lbs) DSCL
Gamma Ray
Max. Temperature: 70 C (158 F) Inclination
Max. Pressure: 60 MPa (8,700 psi)
Measuring Parameters
Measuring range Accuracy
Temperature 5 to 70C Temperature + 0.1 C
(41 to 158F) GR + 3 % of
GR 0 to 10,000 cps measured values*
0 to 3,000 gAPI DSCL NA
DSCL two strings of Deviation: + 1
casing
Deviation: 0...20 deg Resolution
Temperature 0.01 C
Logging parameters
Recommended: Recommended:
Min. Casing Diameter: 100 mm (4") Logging Speed: 10 m/min
Max. Casing Diameter: NA
Sample Rate: selectable
Combinability
Application
none
ANTARES is continually updating and improving its high performance and quality logging system and instruments. For this reason the
specifications are subject to change without prior notice.
ANTARES 2017 41
Cavern and Low-Temperature Cased Hole Tools
The 1463 Pulsed Neutron Blanket Instrument is designed to detect the location of the brine-
hydrocarbon-interface in salt caverns. The instrument contains a neutron generator and a gamma ray
detector. The number of detected Gamma Rays depends on the neutron capture cross-section of the
medium. As brine has a much higher capture cross-section than hydrocarbons, the interface is easily
detected. The tool is also equiped with a Casing Collar Locator and a temperature sensor.
Specifications
Measuring Parameters
Measuring Range: Accuracy
Blanket counter: 0-50000 cts/sec Blanket: + 5% of measured
Temperature: 0 - 80C values
CCL: one strings of Temperature: + 1C
casing CCL NA
Logging parameters
Recommended Recommended
Min. Hole Diameter: 70 mm (2.8") Logging Speed: 9 m/min (1800 ft/hr )
Max Hole Diameter: 300 mm (15.7")
Combinability
none
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
42 ANTARES 2017
Cavern and Low-Temperature Cased Hole Tools
The ANTARES 1464 Pulsed Neutron Blanket Instrument is designed to detect the location of the
brine-hydrocarbon or brine-air interface in salt-caverns. The instrument contains a neutron
generator and a gamma ray detector. The number of detected Gamma Rays depends on the
neutron capture cross-section of the medium. As brine has a much higher capture cross-section
than hydrocarbons, the interface is easily detected. The tool is also equiped with a Dual String
Casing Collar Locator (DSCL) and a temperature sensor.
Specifications
Diameter: 43/44 mm (1 11/16") Source:
Length: 3,858 mm (156) Type: Neutron Deuterium
Transport Length: 2 sections with Tritium target
Weight: 20 kg (123 lbs) Firing frequency: 0-20 Hz
Neutron output: 1x108n/s
Max. Temp: 70C (158F) Neutron length: 10-6 s
Max. Pressure: 20 MPa (2900 psi) Average Lifetime 100 Hrs @
optional 30 MPa (4300 psi) 10 Hz firing rate
Top Connector: GC1, GO1, GC4 or GO4 Measure Points (from bottom):
Bottom Connector: none DSCL 3,510 mm
Blanket Counter: 2,108 mm
Borehole Temp.: 50 mm
Measuring Parameters
Measuring Range: Accuracy
Blanket counter: 0-50000 cts/sec Blanket: + 5% of measured
Temperature: 0 - 80C values
DSCL: two strings of Temperature: + 1C
casing DSCL NA
Logging parameters
Recommended Recommended
Min. Hole Diameter: 70 mm (2.8") Logging Speed: 9 m/min
Max Hole Diameter: 300 mm (15.7") 1800 ft/hr
Combinability
none
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
ANTARES 2017 43
Cavern and Low-Temperature Cased Hole Tools
DSCL/Temperature/Gamma Blanket
The ANTARES 1466 Gamma-DSCL-Temperature is Blanket instrument designed to detect the fluid
interfaces in salt caverns and simultaneous performing high-precision temperature measurements.
The Dual String Collar Locator (DSCL) sensor can distinguish between collars of the outer and inner
casing strings.
Specification
Diameter 44.5 mm (1.75") Source:
Length: 2,464 mm (118.3") Type: 3.7 to 11.1 GBq 137Cs
Weight: 15 kg (28 lbs) (0.1 to 0.3 Ci)
Measuring Parameters
Measuring range Accuracy
Temperature: -5 to 80 C Temperature: + 0.5 C
Gamma Ray: 0 to 65535 cps GR: +/- 5 % of measured
DSCL two strings of values
casing
Resolution
Temperature: 0.05 C
Logging parameters
Recommended Recommended
Min. Hole Diameter: 69 mm (2.5") Logging Speed: 25 m/min
Max Hole Diameter: NA 5000 ft/hr
Combinability
Application
TEMP
none
Source
Holder
ANTARES is continually updating and improving its high performance and quality Logging system and for this reason the specifications
are subject to change without prior notice.
44 ANTARES 2017
Low Temperature Instruments
Furthermore available
The following instruments are available at ANTARES as well. Data sheets will be published soon.
Research and Development is the key feature for ANTARES. Therefore, we are able to enlarge our
product line continuously and will come up with several new tools each year.
ANTARES 2017 45
Surface System
The ANTARES Log Acquisition and Data Interpretation System (ALADIN) is very user friendly.It
comprises a modular 'plug-and-play' panel system and a modular software called GEOBASE. ALADIN
is exceptionally versatile and provides the user with a true multi-tasking logging environment.
SURFACE PANELS
GEOBASE SOFTWARE
46 ANTARES 2017
Contacts
Contacts
ANTARES 2017 47
One System - One Supplier - One Contact
www.antares-geo.de
ANTARES 2017
ANTARES Datensysteme GmbH
Rudolf-Diesel-Str. 6
28816 Stuhr
Germany
Phone: +49 (0) 421 87194-0
Fax: +49 (0) 421 87194-25
Email: info@antares-geo.de
Version1/2017 Web: www.antares-geo.de