DISPOSITIVOS Y EQUIPOS:
Fuente D.C.
Voltmetro analgico
Miliampermetro
Micro ampermetro
Multmetro
Protoboard
Potencimetros: 1,5,10,25,100,1000(K)
Resistores: 0.1,1,2.2,3,10 (K)
PROCEDIMIENTO:
CALIBRACION:
VOLTIMETRO DC
TABLA1
1 2 3 4 5
P1 425.5 425.5 425.5 425.5 425.5
P2 1000 800 600 400 210
Volt. p 3.00 2.72 2.37 2.02 1.38
Volt. t 3.28 3.02 2.67 2.16 1.43
= Vt-Vp 0.28 0.30 0.30 0.14 0.05
AMPERIMETRO DC
TABLA2
1 2 3 4 5
R1() 100 100 100 100 100
P1() 429.5 429.5 429.5 429.5 429.5
P2() 100.6 300.7 500 700 904
Ip(mA) 8.5 6.4 5.2 4.4 3.8
It(mA) 7.94 6.02 4.86 4.07 3.49
= It-Ip -0.56 -0.38 -0.34 -0.33 -0.31
CURVA DE CALIBRACION:
VOLTIMETRO DC
R1
1.024K RP
10K
V1 +88.8
6V Volts
TABLA 3
R1 (K) 1K 1K 1K 1K 1K
RP() 200 520.5 1089 2440 9700
Vp(V) 1 2 3 4 5
Vt(V) 1 2.05 3.12 4.26 5.44
=Vt-Vp 0 0.05 0.12 0.26 0.44
0.5
0.4
0.3
0.2
0.1
0
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
Valores Y
AMPERIMETRO DC
R1
102.7 RP
5K
V1
6V
+88.8
Amps
Miliampermetro
Voltmetro
XMM1
V1
12 V
SENSIBILIDAD
RP
5K
V1 R1
5v 102.7
R2
+88.8
3.5K Amps
Fig. 7
A plena escala:
I=30 uA.
d=88mm
A media deflexin
I=15 uA.
d=46.9mm
E 5
It= = =1.39 mA
Rt 3.6 K
R1
I galvanmetro=It =28.83 uA
R 1+ Rp+ Rg
Ig 28.83uA
SI = = =0.32 uA/mm
d 88 mm
S I =0.32 uA/mm
MILIAMPERIMETRO DC:
I R
RP Mx. Deflexin 10mA 530
5K
V1 Media Deflexin 5mA 1.1K
5v
+88.8
Amps
Ra=1.1K-2*530 Ra=40
VOLTIMETRO DC:
SW
P1
100k
V1
4V
V
+88.8
Volts
SW CERRADO: V1 = 4V
SW ABIERTO: V1 = 2V
CONSUMO ESPECFICO:
I V Ce
mA 1.5mA 7mV 10.5V
uA 1000uA 0.87v 870V
LINEALIDAD DE ESCALA:
R1
4.33K RP
100K
V1
5V
+88.8 +88.8
A mA
0.7
0.6
0.5
0.4
I(mA)
0.3
0.2
0.1
0
0 100 200 300 400 500 600 700
I(A)
EFECTO DE CARGA:
VOLTIMETRO DC
10k
XMM1 20k
P1
V1 Key=A
3V 50%
Sensibilidad: SV 4000/V
Escala 6V 12V
Resistencia interna: RV 24K 48K
Vmedido 1.65V 1.06V
Terico 2.11V 2.48V
Error 21.8% 57.25%
AMPERIMETRO DC
R1
0.986K
V1
3v
+88.8
mA
It 3.0 mA
Ip 3.1 mA
e(%) 3.33 %
METOD.SUSTITUCION: AMPERIMETRO
R1 RX
500 Xk
V1
5V
+88.8
mA
Rx 999
I 2mA
Rp 1035
I 2mA
e -3.6%
METOD.SUSTITUCION: VOLTIMETRO
R1
500
Rx 1962
V 2.14v
Rp 1997
V1 RX
5V Xk +88.8 V 2.14v
Volts
e -1.78%
Terico Prctico V I
R1 1K 0.974K 1.015 V 0.8 mA
R2 2K 2.17K 2.001 V 0.8 mA
Rp 0.5K 0.5K 1.015V 0.8 mA
Rx 1K 0.9K 2.000 V 0.8 mA
MILIAMPERIMETRO
Ig
Sabemos: Rs= Rg
I Ig
E=3 Rsh1=452.5
E=30 Rsh2=31.25
E=75 Rsh3=12.2
E=150 Rsh4=6.07
VOLTIMETRO
Para 3V:
RSH1=VT/IT=3/10mA=300
Para 6V:
RSH2=VT/IT=6/10mA=600
Para 12V:
RSH3=VT/IT=12/10mA=1.2K
Para 30V:
RSH4=VT/IT=30/10mA=3K
VOLTIMETRO-AMPERIMETRO
Implemente un voltmetro miliampermetro con las siguientes escalas:
Voltaje: 5, 10, 30 V
Para 5V:
RSH1=VT/IT=5/10mA=500
Para 10V:
RSH2=VT/IT=10/10mA=1K
Para 30V:
RSH3=VT/IT=30/10mA=3K
Clculo de Resistores para el miliampermetro:
Para 10 mA:
10 mA 4
Rsh 1= =Circuito Abierto
0
Para 50 mA:
10 mA 4
Rsh 2= =1
40 mA
Para 100mA:
10 mA 4
Rsh 3= =0.44
90 mA
V
O R1 R2 R3
500 1k 5k
L
T
I
m
M i
E l
i
T a
R m
p
O
+88.8
e
mA
r
i
R4 m
1 e
t
r
o
CUESTIONARIO FINAL:
0.28+0.30+0.30+ 0.14+0.05
= =0.21
5
b) Error porcentual:
3.283
1 caso: %error= 100 =8.53
3.28
3.022.72
2 caso: %error= 100 =9.93
3.02
2.672.37
3 caso: %error= 100 =11.23
2.67
2.162.02
4 caso: %error= 100 =6.48
2.16
1.431.38
5caso: %error= 100 =3.5
1.43
0.560.380.340.330.31
= =0.384
5
b) Error porcentual:
7.948.5
1 caso: %error= 100 =7.05
7.94
6.026.4
2 caso: %error= 100 =6.31
6.02
4.865.2
3 caso: %error= 100 =7.0
4.86
4.074.4
4 caso: %error= 100 =8.11
4.07
3.493.8
5caso: %error= 100 =8.88
3.49
Para aplicar la definicin de sensibilidad era necesario medir la corriente que deflacta
al instrumento, y el desplazamiento en mm que la aguja recorra debido a esa
deflexin. Cmo todos los parmetros se obtuvieron del circuito, se puede decir que si
corresponden al instrumento: La resistencia elevada de 905 corresponde al esperado
en un microampermetro, y la baja sensibilidad tambin.
R1 R potenciometro
k= =
R2 Rx
Se lleg al balance del puente y se verific con el micro ampermetro una corriente de
cero amperios. El voltaje se verific aproximadamente igual a la mitad del voltaje
aplicado al puente.
1 K 504.5
k= =
2K Rx
De donde se obtiene:
R x (teorico )=1009
10091000
e ( )= 100 =0.89
1009
Ampermetro
1,0 k
de presicin= x 100 = 96.62%
1,035 k
Rp =1,035K
Rx= 1,0k
1 k 1,035 k
e(%) = x 100 =3.5
1k
Voltmetro
1,962
de presicin= x 100 = 98,25%
1,997
Rp = 1,997K
Rx = 1,962k
1,9621,997
e(%) = x 100 =1.78
1,962
+
mA
3V 6V 12V 30V
10mA (fs)
1 Vpk
1kHz
0
V
O R1 R2 R3
500 1k 5k
L
T
I
m
M i
E l
i
T a
R m
p
O
+88.8
e
mA
r
i
R4 m
e
1
t
r
o
Si se puede medir AC con un instrumento DC solo que antes la seal AC debe pasar
por un circuito rectificador la cual la convertir en DC y de ah ira al instrumento DC.
En otras palabras un instrumento AC esa conformado por el medidor DC mas una
etapa de rectificacin.
CIRCUITO INSTRUMENTO DC
RECTIFICADOR Galvanmetro
Hay que tener en cuenta que la escala debe ser calibrada para que muestre valores
correspondientes a AC. Se sabe que el nivel de seal continua despus de pasar por
un rectificador est determinado por el tipo de rectificador que se use. As:
Rectificador de media onda: EDC=ES/
Rectificador de onda completa: EDC=2ES/
Esto es para el caso de seales sinusoidales, luego la escala puede ser calibrada para
que muestre el valor eficaz o pico de la seal AC.