ms_11876325122V19.

PreciControl Anti-HBc
11876325 122 16 x 1.3mL

English Handling
Intended use The controls are supplied readyforuse in bottles compatible with the
system. The controls should only be left on the analyzer during
PreciControl AntiHBc is used for quality control of the Elecsys AntiHBc performance of quality control. After use, close the bottles as soon as
immunoassay on the Elecsys and cobase immunoassay analyzers. possible and store upright at 28C.
Summary Due to possible evaporation effects, not more than 7 quality control
PreciControl AntiHBc is a readyforuse control serum based on human procedures per bottle should be performed.
serum both in the negative and positive concentration range. The controls Please note: Both the vial labels, and the additional labels (if available)
are used for monitoring the accuracy of the Elecsys AntiHBc contain 2 different barcodes. The barcode between the yellow markers is
immunoassay. for cobas8000 systems only. If using a cobas8000 system, please turn
Reagents - working solutions the vial cap 180 into the correct position so the barcode can be read by the
system. Place the vial on the instrument as usual.
PC AHBC1: 8bottles, each containing 1.3mL of control serum
Storage and stability
Human serum, negative for antiHBc; preservative.
Store at 28C.
Target range for the cutoff index: 1.053.0
Store controls upright in order to prevent the control solution from adhering
PC AHBC2: 8bottles, each containing 1.3mL of control serum to the snapcap.
AntiHBc antibodies (human) approximately 1U/mL
(PaulEhrlichInstitute units) in human serum; preservative. Stability:
Target range for the cutoff index: 0.140.87 unopened at 28C up to the stated expiration date
The exact ranges, given in the form of a cutoff index (COI), are encoded in after opening at 28C 8weeks
the barcodes as well as printed on the enclosed (or electronically available)
value sheet. on Elecsys2010 and cobase411 up to 5hours
analyzers at 2025C
Target values and ranges
The target values and ranges were determined and evaluated by Roche. on MODULARANALYTICSE170, up to 4hours
They were obtained using the Elecsys AntiHBc assay reagents and cobase601 and cobase602
analyzers available at the time of testing. analyzers at 2025C
Traceability information is given in the Method Sheet of the relevant Elecsys Materials provided
assay.
Results must be within the specified ranges. In the event that increasing or PreciControl AntiHBc, 2barcode cards, control barcode sheet
decreasing trends, or any other suddenly occurring deviations beyond the Materials required (but not provided)
range limits are observed, all test steps must be checked. Elecsys2010, MODULAR ANALYTICSE170 or cobase immunoassay
When necessary, measurement of the patient sample tested should be analyzers and assay reagents
repeated. See the assay Method Sheet and the operators manual for additionally
Each laboratory should establish corrective measures to be taken if values required materials.
fall outside the defined limits.
Assay
Note: Treat the control serum in the system-compatible labeled bottles for
For technical reasons re-assigned target values valid only for a specific analysis in the same way as patient samples.
reagent and control lot combination, must be entered manually on all
analyzers (except for the cobase602 analyzer). Therefore always refer to Read the data into the analyzer.
the value sheet included in the reagent kit or PreciControl kit to make sure Ensure the controls are at 2025C prior to measurement.
that the correct target values are used. Run controls daily in parallel with patient samples, once per reagent kit, and
When a new reagent or control lot is used, the analyzer will use the original whenever a calibration is performed. The control intervals and limits should
values encoded in the control barcodes. be adapted to each laboratorys individual requirements.
Precautions and warnings Follow the applicable government regulations and local guidelines for
For in vitro diagnostic use. quality control.
Exercise the normal precautions required for handling all laboratory References
reagents. 1 Occupational Safety and Health Standards: bloodborne pathogens.
Disposal of all waste material should be in accordance with local guidelines. (29CFR Part 1910.1030). Fed. Register.
Safety data sheet available for professional user on request.
All human material should be considered potentially infectious. All products 2 Directive 2000/54/EC of the European Parliament and Council of
derived from human blood are prepared exclusively from the blood of 18September 2000 on the protection of workers from risks related to
donors tested individually and shown to be free from HBsAg (PCAHBC1 exposure to biological agents at work.
only) and antibodies to HCV and HIV. The testing methods applied were For further information, please refer to the appropriate operators manual for
FDA-approved or cleared in compliance with the European Directive the analyzer concerned, the respective application sheets, the product
98/79/EC, AnnexII, ListA. information and the Method Sheets of all necessary components (if
The serum containing antiHBc used for the positive control (PCAHBC2) available in your country).
was inactivated using propiolactone and UVradiation. A point (period/stop) is always used in this Method Sheet as the decimal
However, as no inactivation or testing method can rule out the potential risk separator to mark the border between the integral and the fractional parts of
of infection with absolute certainty, the material should be handled with the a decimal numeral. Separators for thousands are not used.
same level of care as a patient specimen. In the event of exposure, the Symbols
directives of the responsible health authorities should be followed.1,2 Roche Diagnostics uses the following symbols and signs in addition to
The controls may not be used after the expiration date. those listed in the ISO 152231 standard.
Avoid foam formation in all reagents and sample types (specimens,
calibrators and controls). Contents of kit
Analyzers/Instruments on which reagents can be used

2015-11, V 19.0 English 1/2


ms_11876325122V19.0

PreciControl Anti-HBc
Reagent
Calibrator
Volume after reconstitution or mixing
GTIN Global Trade Item Number

COBAS, COBASE, ELECSYS and PRECICONTROL are trademarks of Roche.


All other product names and trademarks are the property of their respective owners.
Additions, deletions or changes are indicated by a change bar in the margin.
2015, Roche Diagnostics

Roche Diagnostics GmbH, SandhoferStrasse116, D-68305 Mannheim


www.roche.com

2/2 2015-11, V 19.0 English