E-mail: gglasgo@lumc.edu
Abstract
We report new effective linear attenuation coefficients (eff) for refined gypsum
for Co-60 and 18 MV x-rays. 6 MV and 10 MV eff agree with published
data. For a 100 cm2 field size (FS), tissue defect x = 1.35 cm, target depth
d = 1.65 cm beneath the tissue defect x, eff is 0.123 cm1 (Co-60) and
0.0934 cm1 (6 MV). For 100 cm2, x = 1.35 cm, and d = 5.65 cm beneath x,
eff is 0.072 cm1 (10 MV), and 0.0614 cm1 (18 MV). For d, eff decreases
about 10% from 25 to 400 cm2. For a given FS, eff decreases with d by
3%5% for Co-60, and 3% for 6 MV, 10 MV and 18 MV, but depends on
(d x). For d, when x is large (8 cm), depending on energy and FS, eff
is 2%4% less than when x is small (2 cm). These data were used in a
treatment-planning computer to design compensator filters for a step phantom.
Compensation was to within 10% in the compensation plane (CP). Above and
below CP, computer-calculated ratios of doses with and without filters were
0.751.13. Chamber dose ratios with and without filters were 0.751.12.
1. Introduction
Gypsum Plates
Phantom
Figure 1. Schematic representation (not to scale) of the missing tissue compensator material
experiment. After baseline measurements with the complete phantom (left panel), polystyrene
plates are removed and gypsum attenuation plates inserted (right panel) to restore the charge
measured with the complete phantom.
coefficients (eff), measured in air, for refined gypsum, refined gypsum and steel and refined
gypsum and iron, for 6 MV and 15 MV photon beams. Arora and Weeks (1994) subsequently
reported eff coefficients, in tissue, for refined gypsum for 4 MV, 6 MV, 10 MV and 15 MV.
We extended the refined gypsum eff data to low and high energies by reporting new field-
size and depth-dependent eff for Co-60 and 18 MV x-rays, and 6 MV and 10 MV x-ray
eff coefficients that confirm previous work (Arora and Weeks 1994). We used these data to
implement the compensator filter module for a CMS (Computerized Medical Systems, Inc.,
St Louis, MO) FOCUS treatment-planning computer and confirmed the dosimetry for missing
tissue compensators. Additionally, the refined gypsum eff coefficients may be of use to those
desiring to deliver intensity-modulated radiotherapy (IMRT) using compensating filters.
The method of measurement was the classical compensation for the missing tissue
experiment, shown in figure 1. A polystyrene phantom, nominally 25 cm 25 cm 35 cm with
plates of 0.08, 0.16, 0.32, 0.64, 1.27, 2.54 and 5.08 cm thickness, was used for measurements.
A 0.99 scaling factor was used to correct the polystyrene thickness to water for Co-60 and
6 MV. A Capintec-192 (Capintec, Inc., Ramsey, NY) electrometer was used with a 0.6 cm3
PTW (Physikalisch-Technische Werkstatten-New York, New York, NY) Farmer-type chamber
for charge measurements. The chamber in the phantom was positioned at 100 cm source-
to-axis distance. Refined gypsum (sp gr 2.0) was chosen as the compensating material and
prepared following the recipe described by Weeks et al (1988) One hundred grams of gypsum
was thoroughly mixed with 20 ml of water. The refined gypsum mixture was poured into a
plate mould to form plates 20 cm 20 cm and nominally 0.635 cm thick. Each plate thickness
Gypsum effective attenuation coefficients 65
x
d
d-x
A1 A2 A3
Compensation Plane
B1 B2 B3 B4 B5
Figure 2. Schematic representation of the phantom for measuring the degree of compensation
provided by the gypsum missing tissue filters. The compensation plane, in which film
measurements were made, is at the top of the bottom slab of the phantom. One set of chamber
measurement points (A1A3) are 3.5 cm above the plane of compensation and the other set
of measurement points (B1B5) are 1.5 cm below the plane of compensation. All chamber
measurement points lie on the midline of the phantom.
was measured with callipers and the exact plate thicknesses were used in data analysis. For
square field sizes (25, 64, 144, 225 and 400 cm2) and numerous depths (from 1.65 cm to 14 cm),
measurements were first made with all polystyrene plates present. As polystyrene plates were
removed the gypsum plates were placed in the beam to restore the charge to a value less than
that measured for the full phantom. The plates were placed on a solid 0.635 cm Plexiglas
plate in the block tray slot at 65 cm on the Co-60 unit and each linear accelerator. The exact
thickness of gypsum required to restore the measured charge to that measured for the full
phantom was determined by interpolation. Values for eff were obtained using the tissueair
ratio methodology (Boyer 1983), modified to tissue-maximum ratios
exp(eff t) = [TMR(A, d)/TMR(A, d x)]. (1)
Here eff is the effective linear attenuation coefficient per centimetre, t is the thickness of
gypsum in centimetres, TMR is the tissue-maximum ratio, A is the field size, d is the depth of
the plane of compensation beneath the high point and x is the missing tissue defect. Linear
least-squares fits were performed on eff as a function of field size area and eff extracted
for the intermediate square field sizes of 36, 49, 81, 100, 121, 169, 196, 256, 289, 324 and
361 cm2. Because of space limitations the results of only the even field sizes are presented
in tables 14. Complete tables for all 16 field sizes are available from the author on
request.
To confirm the accuracy of the measured gypsum effective attenuation coefficients
confirmation measurements were performed. A test phantom (figure 2) was designed using
the same polystyrene phantom described. The plane of compensation was selected at a depth
of 12 cm below the high point of the phantom. On the central axis the tissue defect was 3 cm
and the target depth was 9 cm. The phantom was irradiated with an 8 cm 17 cm field at
66 G P Glasgow et al
Target
depth Total depth Field size (cm cm)
Tissue below high point to
defect (cm) defect (cm) target (cm) 6 6 88 10 10 12 12 14 14 16 16 18 18 20 20
36 64 100 144 196 256 324 400
1.35 1.65 3 0.1249 0.1239 0.1227 0.1211 0.1193 0.1171 0.1147 0.1120
2 1 3 0.1236 0.1227 0.1216 0.1202 0.1186 0.1167 0.1146 0.1123
2.6 3.4 6 0.1210 0.1204 0.1196 0.1187 0.1176 0.1163 0.1149 0.1134
2.6 6.4 9 0.1223 0.1215 0.1204 0.1192 0.1177 0.1159 0.1140 0.1119
2.6 11.4 14 0.1204 0.1196 0.1185 0.1172 0.1157 0.1140 0.1122 0.1099
3.95 2.05 6 0.1216 0.1208 0.1198 0.1186 0.1172 0.1155 0.1137 0.1116
5.1 0.9 6 0.1212 0.1204 0.1199 0.1183 0.1169 0.1152 0.1135 0.1115
5.2 3.8 9 0.1211 0.1203 0.1194 0.1181 0.1167 0.1151 0.1133 0.1113
5.2 8.8 14 0.1202 0.1194 0.1184 0.1171 0.1156 0.1139 0.1120 0.1097
6.55 2.45 9 0.1215 0.1206 0.1195 0.1181 0.1165 0.1146 0.1126 0.1102
8 1 9 0.1207 0.1198 0.1188 0.1174 0.1158 0.1140 0.1121 0.1098
8 6 14 0.1194 0.1186 0.1176 0.1164 0.1149 0.1134 0.1115 0.1094
10.4 3.6 14 0.1202 0.1193 0.1183 0.1170 0.1154 0.1139 0.1120 0.1098
Target
depth Total depth Field size (cm cm)
Tissue below high point to
defect (cm) defect (cm) target (cm) 6 6 88 10 10 12 12 14 14 16 16 18 18 20 20
36 64 100 144 196 256 324 400
1.35 1.65 3 0.0945 0.0940 0.0934 0.0926 0.0917 0.0906 0.0894 0.0881
2.6 3.4 6 0.0946 0.0940 0.0932 0.0923 0.0912 0.0900 0.0885 0.0869
2.6 6.4 9 0.0944 0.0937 0.0928 0.0916 0.0903 0.0888 0.0870 0.0850
2.6 11.4 14 0.9125 0.0908 0.0902 0.0894 0.0896 0.0892 0.0864 0.0851
3.95 2.05 6 0.0950 0.0944 0.0936 0.0926 0.0915 0.0902 0.0889 0.0873
5.2 3.8 9 0.0935 0.0928 0.0922 0.0910 0.0899 0.0885 0.0875 0.0853
5.2 8.8 14 0.0934 0.0920 0.0918 0.0900 0.0887 0.0873 0.0862 0.0837
6.55 2.45 9 0.0938 0.0931 0.0922 0.0911 0.0898 0.0883 0.0866 0.0847
7.8 6.2 14 0.0910 0.0905 0.0898 0.0890 0.0881 0.0870 0.0857 0.0843
10.4 3.6 14 0.0926 0.0915 0.0907 0.0897 0.0885 0.0871 0.0856 0.0838
100 cm SAD, nominally at the lower chamber measurement plane in figure 2. Kodak (Eastman
Kodak Co., Rochester, NY) XV-2 film was placed in the plane of compensation to measure
compensation in that plane. The test phantom design allowed the 0.6 cm3 chamber to be
placed at eight locations in the phantom so that the degree of compensation could be measured
at points 3.5 cm above and 1.5 cm below the plane of compensation. Compensator filters for
the test phantom for each energy were designed using the FOCUS (Computerized Medical
Systems, Inc 1998) compensating filter module, using the Clarkson algorithm to compensate
for irregular dose pattern produced when a radiation beam was incident perpendicular to the
irregular surface of this phantom. The software uses the same mathematical model previously
described, but solved for eff. Specifically
eff = ln[TMR(A, d)/TMR(A, d x)]/t. (2)
Gypsum effective attenuation coefficients 67
Target
depth Total depth Field size (cm cm)
Tissue below high point to
defect (cm) defect (cm) target (cm) 6 6 88 10 10 12 12 14 14 16 16 18 18 20 20
36 64 100 144 196 256 324 400
1.35 5.65 7 0.0728 0.0725 0.0720 0.0714 0.0707 0.0699 0.0690 0.0679
1.35 8.35 9.7 0.0739 0.0734 0.0730 0.0720 0.0711 0.0701 0.0689 0.0676
1.35 11.1 12.4 0.0727 0.0722 0.0716 0.0709 0.0700 0.0691 0.0680 0.0668
1.35 13.9 15.2 0.0708 0.0706 0.0702 0.0698 0.0693 0.0688 0.0682 0.0675
2.6 4.4 7 0.0740 0.0735 0.0730 0.0723 0.0715 0.0705 0.0695 0.0683
2.6 7.1 9.7 0.0732 0.0727 0.0721 0.0714 0.0705 0.0696 0.0684 0.0672
2.6 9.8 12.4 0.0722 0.0717 0.0710 0.0703 0.0693 0.0682 0.0670 0.0657
2.6 12.6 15.2 0.0711 0.0708 0.0703 0.0698 0.0692 0.0685 0.0677 0.0667
5.2 4.5 9.7 0.0731 0.0726 0.0719 0.0711 0.0702 0.0691 0.0679 0.0665
5.2 7.2 12.4 0.0725 0.0720 0.0714 0.0707 0.0698 0.0688 0.0676 0.0664
5.2 10 15.2 0.0713 0.0798 0.0703 0.0697 0.0689 0.0680 0.0670 0.0659
7.8 4.6 12.4 0.0718 0.0714 0.0708 0.0700 0.0692 0.0682 0.0671 0.0659
7.8 7.4 15.2 0.0712 0.0708 0.0702 0.0695 0.0686 0.0676 0.0665 0.0653
10.4 4.8 15.2 0.0718 0.0713 0.0706 0.0699 0.0689 0.0679 0.0667 0.0653
Target
depth Total depth
Field size (cm cm)
Tissue below high point to
defect (cm) defect (cm) target (cm) 6 6 88 10 10 12 12 14 14 16 16 18 18 20 20
36 64 100 144 196 256 324 400
1.35 5.65 7 0.0627 0.0621 0.0614 0.0605 0.0594 0.0581 0.0567 0.0551
1.35 8.35 9.7 0.0606 0.0602 0.0598 0.0593 0.0586 0.0579 0.0571 0.0562
1.35 11.1 12.4 0.0603 0.0599 0.0593 0.0586 0.0577 0.0567 0.0556 0.0543
1.35 13.9 15.2 0.0596 0.0593 0.0589 0.0584 0.0578 0.0570 0.0562 0.0553
2.6 4.4 7 0.0634 0.0629 0.0623 0.0615 0.0605 0.0595 0.0582 0.0569
2.6 7.1 9.7 0.0609 0.0604 0.0598 0.0591 0.0582 0.0572 0.0561 0.0549
2.6 9.8 12.4 0.0600 0.0596 0.0591 0.0585 0.0577 0.0569 0.0559 0.0549
2.6 12.6 15.2 0.0611 0.0606 0.0599 0.0590 0.0580 0.0569 0.0555 0.0541
5.2 4.5 9.7 0.0617 0.0612 0.0606 0.0598 0.0589 0.0578 0.0566 0.0553
5.2 7.2 12.4 0.0604 0.0597 0.0590 0.0587 0.0578 0.0569 0.0558 0.0546
5.2 10 15.2 0.0594 0.0590 0.0584 0.0577 0.0569 0.0560 0.0550 0.0538
7.8 4.6 12.4 0.0609 0.0604 0.0598 0.0590 0.0581 0.0570 0.0558 0.0544
7.8 7.4 15.2 0.0597 0.0593 0.0587 0.0580 0.0572 0.0563 0.0552 0.0540
10.4 4.8 15.2 0.0605 0.0600 0.0594 0.0587 0.0578 0.0568 0.0556 0.0543
This module requires the use of a eff that is a function of field size and the depth of the
selected plane of compensation beneath the highest point in the treatment field.
The dosimetrist performs an initial isodose computation with a field size and an
approximate value of eff. This yields the depth of the plane of compensation beneath the
high point, and, on the central axis, the missing tissue defect and the target depth below
the missing tissue defect. Using these parameters and tables 14, the dosimetrist selects an
appropriate eff and repeats the computation. For the confirmation dosimetry, the eff values
were 0.1193, 0.0906, 0.0716 and 0.0599 for Co-60, 6, 10 and 18 MV, respectively.
68 G P Glasgow et al
Table 5. Confirmation ion-chamber dosimetry. Table entries are ratios of the measured absorbed
dose in water to the FOCUS calculated absorbed dose in water without (w/o) and with the filter
present.
The FOCUS compensator data files were transferred electronically to a Par Scientific
DIGIMILLTM (S & S Par Scientific, Inc., Brooklyn, NY) automated milling machine, where
the moulds for the compensator filters were milled. The mould material machined was regular
density (sp gr 0.0337) StyrofoamTM blocks 8 8 2.165 thick. The milling machine has
a milling accuracy of 1 mm. Quality assurance depth measurements were performed on the
milled moulds and confirmed that the accuracy of the milled depths was within 1 mm.
3. Results
Figure 3. 6 MV coronal per cent isodose scans in the compensation plane of the test phantom.
Without compensation (left panel) and with compensation (right panel). Per cent isodose curves
are normalized to the maximum absorbed dose in each scan.
other energies. Figure 3, for 6 MV, shows in the plane of compensation, the per cent isodose
curves normalized to the maximum dose in the compensation plane. A dose uniformity of
about 10% is achieved in the plane of compensation. A similar degree of dose uniformity was
obtained in the plane of compensation for the other three energies. Numerous factors could
contribute to the observed non-uniformity. Further careful studies will be required to identify
these factors and their relative contributions to the non-uniformity in dose.
Table 6 compares coefficients from this study to that reported by Arora and Weeks (1994).
The new coefficients for Co-60 and 18 MV x-rays have values expected for these energies.
The eff of 0.1227 cm1 (Co-60) is greater than the eff of 0.0998 cm1 (4 MV); the eff of
70 G P Glasgow et al
0.0614 cm1 (18 MV) is less than the 0.063 cm1 eff (15 MV). For 6 MV and 10 MV the
coefficients agree to within 1% to about 3%. This study included measurements for five square
field sizes (25, 64, 144, 225, 400 cm2); Arora and Weeks (1994) studied four square field sizes
(25, 100, 400 and 625 cm2). For the two energies (6 MV, 10 MV) this study shows a somewhat
stronger decrease in eff with increases in field size than the previous study (Arora and Weeks
1994). They also noted a weak decrease in eff as depth increased. The depth dependence
is a complex function, and depends on the tissue defect, or filter thickness, as noted later.
For example, for a 100 cm2 field, for 6 MV, for a target depth 6.4 cm below a 2.6 cm tissue
defect, eff is 0.0928 cm1; for essentially the same depth 6.2 cm below a 7.8 cm tissue defect,
eff is 0.0898 cm1, about 3% less. For the same field size, for 10 MV, for a target depth of
8.35 cm beneath a 1.35 cm tissue defect, eff is 0.0730 cm1; for a similar target depth of
7.4 cm beneath a 7.8 cm tissue defect, eff is 0.0702 cm1, about 4% less. For a given target
depth below the defect, when the tissue defect is large (about 8 cm), depending on energy and
field size, eff is 2%4% less than the corresponding eff when the tissue defect is small (about
2 cm). Arora and Weeks (1994) noted this dependence with tissue defect as a dependence
associated with the thickness of the attenuating filter. For a 100 cm2 field, to 10 cm depth,
for a 7.9 cm attenuator, Arora and Weeks (1994) reported a 1.7% decrease in eff (4 MV) and
0.6% decrease in eff (15 MV). In this study, the corresponding decreases in eff for a tissue
defect of 10.4 cm were 3.6% (Co-60), 2.9% (6 MV), 2.0% (10 MV), and 3.2% (18 MV).
Multiple factors contribute to the effect. For the same depth below different tissue defects,
beam geometry and scatter is different. Moreover, a thicker filterrequired for a greater tissue
deficitprobably hardens the beam more than a thinner filter. Consequently, the eff for the
slightly higher energy beam is less than the eff for a thinner filter in the same nominal energy
beam.
4. Conclusions
For a 100 cm2 field size, the new refined gypsum eff of 0.1227 cm1 (Co-60) and the eff
of 0.0614 cm1 (18 MV) are values expected for these energies. Other eff coefficients
(6 MV, 10 MV) agree to within 1% to about 3% with published data. For all energies, for a
given depth, eff coefficients decrease by about 10% for field sizes from 5 cm 5 cm to 20 cm
20 cm. For a given field size, eff generally decreases with target depth from 3% to 5% for
Co-60, and 3% for 6, 10 and 18 MV, but the depth dependence is complex and a function
of the tissue defect. Compensator filters constructed using these data and a compensator
filter-planning module produced ratios of doses with filters to doses without filters that were
confirmed to within 1% by chamber measurements. In a test phantom designed to test
compensation using film, compensation to within about 10% was achieved across the plane of
compensation.
References
Arora V A and Weeks K J 1994 Characterization of gypsum attenuators for radiotherapy dose modification Med.
Phys. 21 7781
Boge R J, Edland R W and Matthes D C 1974 Tissue compensators for megavoltage radiotherapy fabricated from
hollowed Styrofoam filled with wax Radiology 111 1937
Boyer A L 1983 Design of compensating filters Advances in Radiation Therapy Treatment Planning ed A E Wright
and A L Boyer (New York: AIP) pp 185220
Clark B G and Evans M D C 1988 Standard compensators for ENT therapy fields Med. Dosim. 13 1737
Gypsum effective attenuation coefficients 71
Computerized Medical Systems, Inc 1998 FOCUS Dose Calculations and Clinical Applications Manual (chap 13,
Compensating FiltersTechnical Considerations) Release 2.1.0 (St Louis, MO: Computerized Medical
Systems)
Dmytrow B W, Kuchnir F T and Jamshidi A 1986 Verification of the efficacy of a compensator to produce a uniform
dose in depth over the entire field of irradiation Am. Assoc. Med. Dosim. J. 11 379
Ellis F, Hall E J and Oliver R 1959 A compensator for variations in tissue thickness for high energy beams Br. J.
Radiol. 32 4212
Khan F M, Moore V C and Burns D J 1970 The construction of compensators for cobalt teletherapy Radiology 96
18792
Hall E J and Oliver R 1961 The use of standard isodose distributions with high energy radiation beamsthe accuracy
of a compensator technique in correcting for body contours Br. J. Radiol. 34 4352
Harari M, Sharda N N, Brock L K and Paliwal B R 1998 Improving dose homogeneity in routine head and neck
radiotherapy with custom 3-D compensation Radiother. Oncol. 49 6771
Hendee W R and Garciga C E 1967 Tissue compensating filters for cobalt 60 teletherapy Am. J. Roentgenol. 99
93943
Jackson W 1970 Wax retraction as a technique for compensating the effect of surface irregularities in high-energy
radiotherapy Br. J. Radiol. 43 85967
Reinstein L F 1992 New approaches to tissue compensation in radiation oncology Advances in Radiation Oncology
Physics ed J A Purdy (New York: AIP) pp 53572
Robinson D M and Scrimger J W 1987 Megavoltage photon beam dose reduction with retracted tissue compensators
Phys. Med. Biol. 32 10317
Robinson D M and Robinson J W 1989 Limitations of retracted missing tissue compensators: an experimental
analysis Med. Dosim. 14 4954
Sundbom L 1964 Individually designed filters in cobalt 60 teletherapy Acta Radiol., Ther. Phys. Biol. 2 189206
Van de Geijn J 1965 The construction of individualized intensity modifying filters in cobalt 60 teletherapy Br. J.
Radiol. 38 86570
Weeks K L, Fraass B A and Hutchins K M 1988 Gypsum mixtures for compensator construction Med. Phys. 15 4104