INTRODUCTION
Five aqueous metal solution samples were submitted for analysis by the QuanX EDXRF
system. The samples encompassed about 15 elements ranging in concentrations from
2.5 ppm up to 12.5 ppm, as illustrated below:
PROCEDURE
Instrumentation: The samples were analyzed using the QuanX system, which consists
of a 50kV Rhodium target X-ray tube, an eight-position filter wheel, and better-than 170
eV Si(Li) Peltier-cooled detector, coupled to an ADC. Samples were presented to the x-
ray beam using an automated 10-position sample tray with sample-spinning capability.
A PC controls setup, automation and data analysis through the Wintrace software.
Sample Preparation: No sample preparation was performed. A sample of each
solution was transferred into a standard 31-mm plastic cup and sealed by Prolene-
brand x-ray film.
Excitation Conditions: Each sample was analyzed using the following three excitation
conditions to optimize detection limits for the wide range of elements:
The total analysis time per sample was 82 minutes, of which 80 minutes were equally
split between the Mid Zc and High Za conditions.
As with most analytical techniques, the acquisition (measurement) time in EDXRF is the
main determinant of experimental precision, because it directly affects analyte peak size
and the counting statistics.
Peak Integration: Net peak intensities were extracted using the XML peak-fitting
method. The XML routine corrects for peak overlaps using a least-squares algorithm
that fits stored reference spectra of pure analyte elements to a sample spectrum.
Digital filtering of spectra (mathematical treatment) is used to correct for the background
before XML is performed. Together, these methods yield accurate x-ray intensities for
each element.
The analysis results exactly as reported by the software are shown in Appendix A
below. These results can also be exported and viewed in spreadsheet form, as in the
embedded Stillwater.xls spreadsheet in Appendix E.
The spreadsheet also contains the statistical analysis of 10 consecutive measurements
of Sample 3 to determine the experimental precision. For most elements, the precision
was under 10% rsd. The notable exceptions were Ir and Sn, as both were present
below the lower detection limit of the instrument. Se and Te were measured with 4.3%
and 13.4% rsd, respectively. The precision for Te, in particular, might be improved by
using a thicker Cu filter for the acquisition; this decision would be studied as part of final
method optimization.
Appendix C shows overlaid spectra of Sample 1, 3 and 5 under each of the conditions
used for the analysis. In addition, a comparison of the spectra with thick and thin Cu
filters is provided to illustrate the filters effect on peak definition.
The measured peak for Sample 5 was the same size as that measured for Samples 1
and 10, while its concentration was twice as high.
Additional examples of calibration curves for S, Fe, Co, Rh and Pb - are provided in
Appendix B.
CONCLUSION
The analysis results demonstrate the ability of the QuanX system to accurately and
precisely characterize a wide range of trace constituents of aqueous solutions.
Conditions
Low Za
Voltage 4 kV Current 1.8 mA
Livetime 100 seconds Preset counts 0
Filter No Filter Atmosphere Air
Maximum Energy 10 keV Count rate range Medium
Warmup time 0 seconds
Mid Zc
Voltage 24 kV Current 0.5 mA
Livetime 1000 seconds Preset counts 0
Filter Pd Thick Atmosphere Air
Maximum Energy 20 keV Count rate range Medium
Warmup time 0 seconds
High Za
Voltage 50 kV Current 0.08 mA
Livetime 100 seconds Preset counts 0
Filter Cu Thin Atmosphere Air
Maximum Energy 40 keV Count rate range Medium
Warmup time 0 seconds
Results
5
S 0.99117 % 6.6997e-003 % 244.66 28.836
Fe 0.15332 % 6.7763e-004 % 57.564 7.1621
Co 148.78 ppm 3.3187 ppm 6.8059 16.242
Ni 1.5423 % 1.4442e-003 % 1135.5 -4.843
Cu 0.23723 % 5.4435e-004 % 246.5 73.427
As 7.7911 ppm 0.44169 ppm 1.5044 5.7696
Se 4.7955 ppm 0.28564 ppm 1.4317 5.8403
Ir 8.7522 ppm 0.45008 ppm 1.8624 7.3106
Pt 7.6639 ppm 0.40591 ppm 1.4576 4.5024
Pb 4.7785 ppm 0.40142 ppm 1.0635 6.9185
Ru 4.9785 ppm 0.19703 ppm 2.4542 6.9798
Rh 4.9794 ppm 0.19959 ppm 2.4721 7.3469
Pd 4.4631 ppm 0.21028 ppm 2.3693 10.092
Sn 2.9826 ppm 0.35744 ppm 1.5113 31.292
Te 2.7133 ppm 0.55913 ppm 1.321 72.777
H2O 97.056 %
8
S 1.2672 % 7.7444e-003 % 295.33 30.434
Fe 7.498e-002 % 4.9733e-004 % 29.725 9.1462
Co 75.083 ppm 2.5001 ppm 3.7555 11.883
Ni 0.75624 % 1.0262e-003 % 548.4 5.3331
Cu 1.5088 % 1.236e-003 % 1525.1 35.623
As 13.156 ppm 0.56443 ppm 2.2795 7.2846
Se 12.898 ppm 0.36496 ppm 3.4507 6.0833
Ir 3.6955 ppm 0.90705 ppm 0.70818 29.506
Pt 6.8791 ppm 0.47179 ppm 1.1773 5.3417
Pb 13.023 ppm 0.54447 ppm 2.2529 6.6191
Ru 12.72 ppm 0.24889 ppm 5.6492 6.5688
Rh 12.731 ppm 0.25238 ppm 5.7037 7.0813
Pd 12.829 ppm 0.26034 ppm 6.1574 9.4546
Sn 2.0004 ppm 0.37905 ppm 0.92403 29.734
Te 13.113 ppm 0.6065 ppm 5.8518 67.4
H2O 96.375 %
10
S 1.5045 % 8.5574e-003 % 337.23 30.691
Fe 2.3621e-002 % 3.0465e-004 % 11.512 10.533
Co 26.335 ppm 1.7977 ppm 1.5162 9.1958
Ni 0.24646 % 6.0587e-004 % 175.73 10.884
Cu 2.4974 % 1.5958e-003 % 2453. 3.7384
As 0. ppm 0. ppm 0. 5.24
Se 3.0477 ppm 0.30102 ppm 0.74006 4.6029
Ir 0. ppm 0. ppm 0. 47.686
Pt 0. ppm 0. ppm 0. 5.288
Pb -0.44334 ppm -0.1938 ppm 0.17839 5.9026
Ru 2.7988 ppm 0.21176 ppm 1.1351 6.2409
Rh 2.1599 ppm 0.21707 ppm 0.88503 7.026
Pd 2.5496 ppm 0.22572 ppm 1.121 8.728
Sn 2.7106 ppm 0.39855 ppm 1.1575 27.807
Te 2.6847 ppm 0.6191 ppm 1.1126 64.715
H2O 95.724 %
Figure C-1 Spectra of Samples 1, 3 and 5 acquired under conditions optimized for S.
Sample S Fe Co Ni Cu As Se Ru R
% % ppm % % ppm ppm ppm pp
Notes :
1) Pb for Sample 10 came out negative (-0.4 ppm), because a non-zero intercept was used for the calibration
calibration
2) Ir and Sn were present below the lower detection limit of the instrument; Ir overlaps with Pt.
3) As in Samples 1 and 10 was present below the detection limit of the instrument, because of peak overlap w