Anda di halaman 1dari 4

University of Santo Tomas

COLLEGE OF SCIENCE
Espaa Blvd., Manila, Philippines
Tel. No.: (63-2) 7315728, (63-2) 406-1611 loc. 8224
FAX: (63-2) 740-9730
Website: http://science.ust.edu.ph

Experiment 1(Part 2): Use of Micropipettor and Analytical Balance


Group6: ILAGAN, Ivylyn; GARCIA, Pamela; JAEN, Carmella Gloria; LIM, Leslie Cathleen T.

Keywords: Summary:
Accuracy, Analytical The micropipettor is a device used for measuring volumes as small
Balance, Micropipettor, as 1 . The accuracy and precision of the operators regarding the
Percent error, Precision, usage of this device can be determined by weighing measured
Standard deviation volumes on the analytical balance. Accuracy and precision can be
best measured using percent error and standard deviation
respectively. The closer the obtained value to the standard value,
the more accurate the operator. High levels of precision produce a
. smaller value for standard deviation, resulting to a more precise
operator. As the volume to be measured increases, percent error
increases as well, for the larger the calibration of the device, the
more prone the operator is to error.

Introduction bubble on the level indicator must be within


One of the main devices used in this the inscribed circle. The read out should
experiment was the analytical balance. An show 0.0000g if the weighing pan is empty.
analytical balance is a measuring device If not, the tare button must be pressed to
with a basic function or principle to simply recalibrate the balance. Lastly, the object to
determine the weight of a material. The unit be weighed must be in a container.
it measures is usually in grams (g) or The accuracy of the digital balance
milligrams (mg) depending on the settings. does not depend on the number of decimal
The parts of an analytical balance places it could read. Accuracy is inherent in
include a level indicator with an air bubble a balance based on its manufacturer, internal
positioned at its center for proper alignment, mechanism, model, design, and procedures
a balance pan where objects to be weighed followed. For higher accuracy, the surface of
are placed, a tare button that recalibrates the the object to be weighed must not be touch
device to 0.0000g, a display screen that by ones bare hands. Furthermore, the
reads the measurement, a power button that balance works well with a relatively
turns the device on, and a door handle that constant temperature.
open and close the glass doors. The other device used in this
Before using the analytical balance, experiment was the micropipettor. A
it is important to make sure that the micropipettor is a precision pump with a
weighing pan is clean, and that the doors of disposable tip that measures as little as 1
the analytical balance are closed. The level microliter () sample solution. A small
indicator should be checked wherein the volume micropipettor (colored red)
measures 0.5-10 ; a midrange water by the same student. The average
micropipettor (colored yellow) measures 10- weight and standard deviation was
100 ; and a large volume micropipettor computed.
(colored blue) measures 100-1000 of For the second part of the
sample solution. experiment, the precision of the operators
Two topics tested in this experiment was tested. The same procedures from the
was the accuracy and precision of the first part of the experiment was done.
measuring device. Accuracy is the closeness However, each member of the group
of a measured value to its true value. in the performed this task. Each member placed 50
experiment, only (1) member tested for of distilled water in the aluminum foil
accuracy. On the other hand, precision is the container. After which, the average weight,
closeness of measured values to one another. percent error, and standard deviation was
In the experiment, each member of the computed.
group tested for precision. Below were the formulas used in the
There were three objectives in this experiment:
experiment. First, the parts of the analytical
balance were identified; second, the
analytical balance was used accurately and =

precisely; and third the standard deviation Equation 1.1. Average
(SD), coefficient of variation (CV), standard
error of the mean (SE), and percent error
=
(%E) were determined.
Equation 1.2. Coefficient of Variation
Materials and Methods
The materials used in this experiment were ! 2
the three kinds of micropipettors (colored =
red, yellow, and blue), mircopipettor tips, ( 1)
analytical balance, distilled water, and Equation 1.3. Standard Deviation
aluminum foil.
The first part of the experiment
involved testing the accuracy of the =

operators. Only one member of the group Equation 1.4. Standard Error of Mean
performed this task. First, the analytical
balance was turned on. After which, a piece
of aluminum foil container was placed on % = l l100

the weighing balance. The tare button was Equation 1.5. Percent Error
pressed to recalibrate the balance to
0.0000g. 50 of distilled water was placed where N is the number of trials, CV
in the aluminum foil container, and the is the coefficient of variation, SD is the
measured weight was recorded in the lab standard deviation, SE is the standard error
manual. The tare button was pressed again of the mean, and %E is the percent error.
before the same procedures were repeated
five times using 100 and 500 distilled
Results and Discussion volume, indicate a high level of
The values obtained from the experiment precision, and thus will produce a
were inserted in the formulas listed smaller value for standard deviation. The
above (see Materials and Methods). smaller the value for standard deviation
However, to obtain percent error, the more precise the operator.
standard value must be obtained as well. Table 1.3. (See appendix) shows
To determine the standard value of the the compiled results for all of the
distilled water at a specific room groups. This table especially depicts the
temperature as is the case for this percent error [Equation 1.5.] for each of
experiment, the standard density (g/mL) the groups, a value that may be used to
of the sample (water) in room determine accuracy. The smaller the
temperature must be known. Then, this percent error, the more accurate the
will be converted to (g/L) since operator. For each of the respective
measurements in class was done in L. volumes 50L, 100L, and 500L,
The grams (g) of the sample will be Groups 2, 9, and 10 obtained the
obtained by multiplying the standard smallest percent error, and thus are the
density of water in L with the volume most accurate among the class.
of water placed in the aluminum foil. However, the correct trend of percent
The computed values for the standard error is seen in the results of Group 4.
values are 0.0499g for the 50L volume, This displays an increase in percent error
0.0977g for the 100L, and 0.4986g as volume to be measured increases.
500L volume. Finally, the percent error This is because the larger the calibration
can be determined with the given of the device, the more prone the
formula above (see Materials and operator is to error.
Methods). Possible sources of error for this
Table 1.1. (See Appendix) shows experiment include incorrect setting for
the weight of different water samples by the readout of the micropipettor,
1 member of the group. By comparing presence of bubbles at the end of the tip
the average weight [Equation 1.1] from when pipetting, when the plunger is not
the standard values computed earlier, depressed properly to the first and
one can determine the accuracy of the second stop, when the level indicator in
operator. The closer the obtained value the analytical balance is not in the
to the standard value, the more accurate middle, when the analytical balance is
the operator. A more quantitative not in a stable room temperature, and
measure of accuracy is depicted by when the wrong values are inputted in
percent error, which will be elaborated in the given formulas.
the proceeding discussion in Table 1.3.
Table 1.2. (See Appendix) shows Conclusion
the weight of different water samples by
the members of the group. This table In conclusion, the analytical balance is a
allows a measure of precision within measuring device that determines the
members of the group. This is indicated weight of an object in grams (g), or
by the closeness of values and can milligrams (mg) depending on the
quantified via standard deviation chosen settings. Accuracy is how close a
[Equation 1.3.] values. Highly close value is to the true value, and precision
values, such as those seen in the 100L is how close values are with each other.
Determination of SD, CD, SE, and %E https://www.sartorius.com/sartori
includes using the formulas stated in us/en/EU/products/laboratory/lab
Materials and Methods. As seen in the oratorybalances/analytical-
experiment, the closer the obtained value balances.
to the standard value, the more accurate Sigma Aldrich (2017). Uses of the
the operator. High levels of precision Micropipettor Retrieved October
produce a smaller value for standard 3,
deviation. The smaller the value for 2017fromhttp://www.sigmaaldric
standard deviation the more precise the h.com/catalog/product/aldrich/z1
operator. As the volume to be measured 14820?lang=en&region=P
increases, percent error increases as well.
This is because the larger the calibration
of the device, the more prone the
operator is to error.

Acknowledgements

The authors are grateful to the


University of Santo Tomas College of
Science for providing the equipment
needed for the experiment, to Professor
Abrantes and Professor Lumanlan for
giving a detailed prelab and postlab
explanation, and to all the parents who
gave their daughters a chance for
experiencing a great education.

References
Brooke, J. (2012). Biology. Accuracy
and precision. McGraw Hill:
USA.
Christian, G. (2011). Analytical
Chemistry. Accuracy and
precision. Wiley and Sons:
Washington.
Hart, C.S. (2015). The Analytical
Balance and Its Use. Use and
care for the analytical balance.
Popular Astronomy: New York.
Ohaus. (2016). Accuracy and Precision
Retrieved October 3, 2017
fromhttps://www.ohaus.com/enU
S/Products/Balances-
Scales/Analytical-Balances.
Sartorius. (2017). Analytical Balances
Retrieved October 3, 2017 from

Anda mungkin juga menyukai