Sample swatches of a specified amount (weight, size, somewhat inconsistent among laboratories national-
and surface area) are inoculated with a specified ly and internationally. There are no governing bodies
number of bacteria using a pipette. The inoculum enforcing strict adherence to published methods.
must be completely absorbed into the fabric and Modifications are often made and full customer
in intimate contact with the treated surface. If not, disclosure is not practiced by all laboratories.
microorganisms multiply in portions that pool apart
from the fabric, leading to misleading results for AATCC TM100
even properly treated samples. AATCC TM1001 uses a full nutrient broth for
dilution to achieve the required testing concentra-
The inoculated sample is incubated in a high tion of bacteria for inoculation. The nutrient level is
humidity environment. Surviving organisms are much higher than expected in most real-world situ-
recovered in a neutralizing broth containing an ations, allowing for aggressive bacterial growth and
agent that will stop the action of the antimicrobial. reproduction. Only a single replicate of the test is
Shaking the sample extracts surviving organisms. normally performed, as there is no specificity in the
The recovery broth is re-plated and re-incubated for method that requires more than one replicate. Doing
18-24 hours and the number of surviving bacteria the test in triplicate, as a minimum and as specified
counted as colony forming units (CFU). The techni- in JIS L 1902 and ISO 20743, is recommended due
cian calculates a reduction in bacteria versus either to the variability inherent in micro-testing.
the initial inoculum or the untreated control (if
available). Fig. 1 shows bacterial colonies, ready for AATCC TM100 is cited in military specifications
counting, for untreated and treated samples. These for evaluation of apparel since such apparel may
test methods do not specify an absolute standard become heavily soiled. Studies sponsored by the US
Congress and conducted jointly by Microban Inter-
national and the US Army Natick Soldier RD&E
Center, yielded correlation between positive results
in AATCC TM100 and odor control and comfort
in the field for army combat uniforms, T-shirts, and
socks.2,3 Natick has issued clarification to AATCC
TM 100 for improving consistency in testing and
adherence to protocol. These clarifications include
requiring clear documentation of test conditions and
Fig. 1. Final plates for counting. requiring use of the method regardless of
antimicrobial technology.4
for efficacy. Minimum requirements vary from
customer to customer and laboratory to laboratory.
JIS L 1902
JIS L 19025 was developed in Japan for testing
Quantitative bacterial testing can be used for all
silver-based antimicrobials. It primarily differs from
antimicrobials. Comparisons can be made between
AATCC TM100 in that the nutrient level in the
different antimicrobial treatments as well as various
inoculums broth is diluted to 1:20. JIS L 1902 also is
treatment levels on the same textile. These methods
explicit about calculating results for treated prod-
better simulate real-world conditions, such as a
ucts versus those for untreated controls and calls for
perspiration-soaked shirt or otherwise damp textile
testing in triplicate. The standard for a valid test is
thrown into a gym bag or locker, than
that there should be at least a 1.5 log increase on the
other methods.
untreated control.
There are also disadvantages to quantitative tests.
The low nutrient level for JIS L 1902 biases testing to
They are long, involved, and expensive, requiring a
provide more positive results for antimicrobials such
number of manipulations to the sample and organ-
as silver and cationic antimicrobials, which can be
isms. Current methods are highly dependent on
neutralized by proteins in the nutrient. The test can
operator interpretation and technique, and therefore
be used with other antimicrobials, such as triclosan,
but surprisingly triclosan seems to provide better From a practical commercial perspective, the Shake
results at the higher nutrient level used in AATCC Flask Method is only applicable to quat silanes and
TM100. AATCC TM100 is more aggressive since it cannot be used to compare results with or among
gives organisms far more resources for growth and other commercial antimicrobial treatments. It has
reproduction. However, actual nutrient levels on a also been implied that only the Shake Flask Method
fabric under normal use would be closer to those can be used to test grafted quat silane technolo-
used in JIS L 1902. The lower nutrient level increases gies, contrary to at least one report on grafted quat
the necessity for an operator skilled in running the silanes8 and experience in this lab.
test to get valid results.
There is little or no correlation between the Shake
ISO 20743 Flask Method and other quantitative tests. In this
There is now an International Standards laboratory, untreated cotton fabrics have yielded
Organization standard, ISO 20743, 6 modeled largely reductions as high as 95% (~ 1 log) after 1-hr con-
on the JIS L 1902, but allowing more flexibility in tact time, while showing 2-3 log increases over the
the conditions of the test. This is as much a curse as standard 24-h exposure time using AATCC TM100.
a blessing when trying to compare results Quick reductions for a quat silane treated sample
between laboratories. after 1-h exposure and then growth over the next 24
h have also been observed in this lab, indicating that
ASTM E2149 Shake Flask Method bacteria may be removed from the liquid buffer, but
The Shake Flask Method is a quantitative are still viable and capable of reproducing.
screening test, developed by Dow Corning for
The conditions of the Shake Flask Method, a
quaternary (quat) silane (poly-3-siloxy-propyldime-
small piece of fabric in a large amount of fluid with
thloctadecyl ammonium chloride) antimicrobial
samples being taken from the fluid and not from the
treatment, generally known as the “Dow Shake Flask
fabric surface, do not replicate real-life use condi-
Method.” 7 The Shake Flask Method can be faster
tions. As a result, the test is not widely accepted by
than AATCC TM100 or JIS L 1902 and requires
professionals as an indicator of efficacy.
considerably less technique to do consistently, but
differs radically from the other quantitative tests.
Qualitative Bacterial Tests
The fabric sample (~1 g), is immersed in 50 mL AATCC TM1479 is a qualitative, zone of inhibition
of an inoculated buffer solution in a flask, which is test adapted from the Kirby-Bauer test used in the
then agitated using a wrist action shaker. The bacte- medical field for decades. Both JIS L 1902 and ISO
rial concentration is ~105 mL-1, but with no nutrient 20743 have qualitative sections modeled on
beyond that transferred with the organisms from the AATCC TM147.
original culture. The time period for the exposure
is 1 h. After the specified time, the technician plates The bottom of a Petri dish is filled with nutrient
an aliquot of the buffer from the flask onto nutrient agar that is streaked with the organism of interest.
agar without neutralization, incubates for 18-24 h, The test sample, a strip of fabric, is then placed over
and then counts the number of colonies. A reduc- the streaks. Fig. 2 shows the results of an AATCC
tion is calculated using the known initial bacterial
concentration and the final count after exposure to
the test sample or is calculated versus an untreated
control. It is recommended that an untreated control
be tested in parallel to ensure a valid test and to run
the test in triplicate. The “inoculum only” flask, as
specified in the method, should be tracked alongside
all treated and untreated samples with each test, to
ensure there is no bactericidal effect resulting from
the non-nutritive test buffer or due to surfactants
added for wetting. Fig. 2. AATCC Test Method 147.
effective treatment completely prevents growth, continually create confusion in the antimicrobial
though no growth means that the sample will be marketplace. The comparative information avail-
highly resistant to fungal attack. The test is not very able on different antimicrobial technologies is often
discriminating with regards to performance; even misleading due to differences in test methodologies,
the minutest amount of microscopic or macroscopic inappropriate methodologies, sometimes hon-
growth will result in a rating other than 0. In this est confusion in running methods, and differing
laboratory, standard ratings are applied to the test, levels of expertise. Just because two laboratories cite
but comments related to percent coverage of the AATCC TM100 or JIS L 1902 does not mean that
fungus on the sample surface are added and photo- the laboratories are actually running those protocols
graphs included in exactly the same way. The quality of testing is
as documentation. distressingly poor; in blind round robin testing12 less
than half of the laboratories evaluated were capable
AATCC TM30 (Part IV) of running a dependable quantitative test protocol.
In AATCC TM3011 (Part IV), a dry, treated and The only way to be sure of comparisons is by run-
untreated 1 × 3 in. strip of nutrient saturated ning tests side-by-side in the same laboratory with
fabric, sprayed with a mixed-spore suspension of the same series of tests. Where differences occur
mildew-causing between laboratories, only detailed review of the test
organisms, is protocol will enable one to understand results.
I-suspended and
incubated in a References
closed jar with 1. AATCC Technical Manual, Vol. 85, 2010, pp142-144.
sterile water in 2. Swofford, H. W., et al., Antimicrobially Treated Products for
the bottom to Military Use: Final Report, US Dept. of Defense Contract
provide moist No. W911QY-04-C-0079, November 2006.
conditions 3. Swofford, H. W, et al., Advanced Antimicrobial & Comfort
Technologies for Military Applications: Final Report, US
(Fig. 4). After Dept. of Defense Contract No. W911QY-05-C-0087,
the incubation May 2008.
period, the tech- 4. AATCC TM 100 clarification, Amy Johnson (contact),
nician grades US Army Natick Soldier RD&E Center, Natick, MA, USA.
percent cover- 5. JIS L 1902:2002 (JAFET/JSA), Testing for Antibacterial
Activity and Efficacy on Textile Products, Japanese Indus-
age by fungal trial Standard.
Fig. 4. Test setup for AATCC Test Method growth. The 6. ISO 20743, Textiles–Determination of Antibacterial Activity
30, Part IV.
organism must of Antibacterial Finished Products, 1st edition, 2007.
germinate and 7. ASTM E2149-10, Standard Test Method for Determining
the Antimicrobial Activity of Immobilized Antimicrobial
establish itself on the treated fabric, whereas in the
Agents Under Dynamic Contact Conditions.
Part III test, the agar provides an antimicrobial-free 8. Gawish, S. M., et al., Textile Research Journal, Vol. 77, No. 2,
zone for the fungus to establish itself and then over- February 2007, pp92-104.
grow the sample. Thus, the Part IV method is less 9. AATCC Technical Manual, Vol. 85, 2010, pp251-252.
aggressive than the Part III method and allows for 10. Cutter, C. N., Journal of Food Protection, Vol. 62, No. 5,
1999, pp474-479.
somewhat better discrimination between treated and
11. AATCC Technical Manual, Vol. 85, 2010, pp76-79.
untreated samples. 12. Centola, D. T., internal memo, Microban International,
April 2007.
Testing Issues
Antimicrobial treatment of textiles provides Author
benefits in odor control and freshness, as well as
H. Wayne Swofford, Microban International,
protection against degradation due to bacteria and
1140 Vanstory Dr., Huntersville, NC 28078, USA;
mold. However, differences in test results between
phone +1 704 875 0806; fax +1 704 875 0810;
laboratories and claims of performance through the
wayne.swofford@microban.com.
use of unspecified and inappropriate test methods