16 slice CT scanners
CEP08025
March 2009
Contents 2
Introduction ............................................................................................. 3
Acknowledgements ............................................................................... 94
References............................................................................................ 95
GE BrightSpeed Elite
Philips Brilliance CT 16
Toshiba Activion 16
Toshiba Aquilion 16
The data given in this report are representative of the scanners as of August 2008,
and are liable to change. Features such as workstations and software packages may
be listed as cost options here, but such additional costs are generally negotiable.
Whilst every effort has been made to ensure the accuracy of this report, we have not
verified the data provided. Specific responses should be confirmed with the
manufacturers.
1.1 Scanner type 3rd generation 3rd generation 3rd generation 3rd generation 3rd generation
Focus-isocentre
1.3 541 570 535 600 600
distance [mm]
Focus-detector distance
1.4 949 1040 940 1072 1072
[mm]
Positioning lights -
accuracy
1.11 ±1 ± 0.5 ±1 ±1 ±1
(at centre of gantry)
[mm]
2.1 Type High frequency High frequency High frequency High frequency High frequency
Method of
2.3 Air cooled Air cooled Air Air Air
generator cooling
Power rating
2.4 53.2 60 50 42 60
[kW]
2.5 kV settings 80, 100, 120, 140 90, 120, 140 80, 110, 130 80, 100, 120, 135 80, 100, 120, 135
Moving (dynamic
3.3 / flying) focal No Yes Yes No No
spot, xy-plane
Moving (dynamic
3.4 / flying) focal No No No No No
spot, z-axis
80 kV : >24kW (300
mA)
80 kV: 300 mA 80 kV: 450 mA
100 kV : >31kW (310
Settings at which Small focal spot in
mA) 100 kV: 300 mA 100 kV: 350 mA
3.5 focal spot HR/UHR mode; Not power dependant
changes 120 kV : >40kW (335 120 kV: 250 mA 120 kV: 300 mA
Not kW limited
mA)
135 kV: 200 mA 135 kV: 260 mA
140 kV : >47kW (335
mA)
Total filtration at
max kV (inherent
+ beam shaping
2.5 (L wedge filter) 2.5 (L wedge filter)
filter) on central 5.90 mm (70 kV, Head)
3.6 axis 7.95 @ 80 kV 6.3 2.5 (M wedge filter) 4.0 (M wedge filter)
8.70 mm (70 kV, Body)
[mm Al 11.0 (DR wedge filter) 11.0 (DR wedge filter)
equivalent @120
kV]
Anode heat
3.8 6.3 8 5 4 7.5
capacity [MHU]
Maximum anode
3.9 cooling rate 840 1608 810 864 1386
[kHU / min]
Method of tube
3.10 Oil to Air Oil to air Oil Liquid Liquid
cooling
4.1 Detector type Solid state array Solid state array Solid state array Solid state array Solid state array
Decay constant
4.3 Info. not available Info. not available 3 3 3
[μs]
Maximum
number of
simultaneously
4.4 16 16 16 16 16
acquired data
sets (number of
'slices')
Sampling 11,050
4.6 frequency for 1968 2857 1200 1800
fastest scan [Hz] (Up to 4,640 samples/rev)
Data
4.7 transmission rate 850 1100 1250 4400 4400
[Mbps]
Number of
detector
elements per row
4.8 888 672 736 800 896
(excluding
reference
detectors)
Number of
reference n.a. (reference detector is
4.9 18 1 1 (total) 1 (total)
detector in tube collimator)
elements per row
Number of
detector
4.10 24 24 24 28 40
elements along
z-axis
Effective length
of detector 16 x 0.625 mm 16 x 0.75 mm 16 x 0.6 mm 16 x 0.5 16 x 0.5
4.11 elements in z-
axis (at 8 x 1.25 mm 8 x 1.5 mm 8 x 1.2 mm 12 x 1.0 24 x 1.0
isocentre) [mm]
Total effective
length of detector
4.12 20 24 19.2 20 32
array at isocentre
[mm]
Quarter detector
4.13 Yes Yes Yes Yes Yes
shift
Hardware
features to Spiral Groove Bearing
4.15 No No No No
improve z-axis and Smart Focal Spot
resolution
Table 5. Couch
Couch top
5.1 Carbon fibre Carbon fibre composite Carbon fibre Carbon fibre Carbon fibre
material
Elevation control
pedals on both sides of
Couch controls - the couch. Additional
5.5 Foot switch Yes Foot switch Foot switch
in room controls (joystick) at
the table if CT Fluro
option installed
Couch controls -
5.6 Cradle In/Out Yes Yes Yes Yes
on console
Accuracy/reprod
5.7 ucibility of table +/- 0.25 ± 0.25 ± 0.25 ± 0.25 ± 0.25
positioning [mm]
Scannable
horizontal range 175 (std) 175 (std)
170 (axial), 160 (helical
5.9 with table top 175 160
& scout) 145 (short) 145 (short)
extension(s)
[cm]
Vertical
movement range
5.10 51 - 99 57.8 - 102.8 45 - 83 31 - 95.4 31 - 95.4
out of gantry
[cm]
Vertical
5.11 movement range 78 - 99 85.0 - 102.8 65 - 84 77.9 - 95.4 77.9 - 95.4
in gantry [cm]
Maximum
5.12 clearance in 25 17.8 62 58 58
gantry [cm]
Minimum couch
top height
5.13 51 57.8 45.5 31 31
outside gantry
[cm]
Maximum weight
5.14 allowed on 205 204 (295 option) 200 205 205
couch [kg]
Maximum weight
[kg] on standard
couch which still
achieves stated 180 (± 0.25 mm)
5.15 204 (295 option) 200 205 205
performance 205 (± 1 mm)
specifications
eg.accuracy
(mm)
Maximum
6.1 continuous 120 100 100 100 100
scan time [s]
Maximum
number of
rotations in one 200
150 - 183 (270 - 250
6.2 helical run at Info. not available 113 133
mA) at 0.6 s (250 option)
standard
abdomen
parameters
Maximum helical
scan coverage
using a 1 mm 1750 (std) 1750 (std)
6.3 1600 1750 1500
imaged slice 1450 (short) 1450 (short)
thickness and a
pitch of 1.5 [mm]
Overscan
required for
6.4 sequential Not required 60 degree fan angle No Max 35 Max 35
scanning
[degrees].
Overbeaming
required for Yes Yes (up to 3/4 rotation at
6.5 Up to 10% 2 (1 at each end) 2 (1 at each end)
helical scanning (Number not specified) either end)
[rotations]
Proprietary SPR
7.1 Scout Surview Topogram Scanogram Scanogram
name
Angular positions
of X-ray tube 90° (Lateral) 0, 90, 180, 270 0, 90, 180, 270
7.4 0 - 359 (1 deg step) a.p., p.a., lateral
available for 180° (AP) 5 degree steps 5 degree steps
SPR [degrees]
Maximum
number of SPR
7.5 2 images AP/PA & Lat 2 2 2 2
views available
for scan planning
Accuracy of slice
7.7 prescription from ± 0.25 ± 0.5 ± 0.5 ±2 ±2
SPR [mm]
Accuracy of
distance
7.8 measurements < 2 x image pixel size ± 0.25 ± 0.5 ±2 ±2
from SPR, at
isocentre [mm]
8a Key settings
8.1 kV steps 80, 100, 120, 140 90, 120, 140 80, 110, 130 80, 100, 120, 135 80, 100, 120, 135
Scan fields of
8.4 25 & 50 25, 50 5 - 50 18, 24, 32, 39, 50 18, 24, 32, 40, 50
view [cm]
8b Sequential mode
Rotation times
0.5, 0.6, 0.7, 0.8, 0.9, 0.4 (opt), 0.5, 0.75, 1.0, 0.4 (option), 0.5, 0.75,
8.5 for sequential 0.5 (option), 0.6, 1, 1.5 0.75, 1.0, 1.5, 2.0, 3.0
1, 2, 3, 4 1.5, 2.0 1.0, 1.5, 2.0, 3.0
scanning [s]
Partial rotation
8.6 0.25, 0.3 0.28 (option), 0.34 0.36 0.48 0.28 (option), 0.32
scan times [s]
Nominal z-axis
x-ray beam 2.4, 7.2, 9.6, 10.0,
14.4, 16.0, 19.2 1, 2, 4, 6, 8, 12, 16, 24,
8.7 collimations, 1.25, 10, 15, 20 1.2, 12, 18, 24, 2, 4, 8, 12, 16, 20
32
sequential
scanning [mm]
Maximum z-axis
x-ray beam
8.8 collimation with 10 12 9.6 8 8
sub-millimetre
slices [mm]
Thinnest
acquisition width
available at full
8.9 0.625 1.5 1.2 5 2
z-axis x-ray
beam collimation
[mm]
32: 4
24: 6
Sequential 20: 5
acquisition 16: 4
1.25: 0.625 19.2: 1.2 16: 4
options 24: 1.5, 3 12: 3
10: 0.625, 1.25 9.6: 0.6 12: 3
8.10 (beam 12: 0.75 8: 2, 8
collimation: 15: 3.75 4: 2 8: 2
available data 1.2: 0.6 6: 6
acquisition 20: 1.25, 2.5 1: 1 4: 1
4: 1, 4
widths) 2: 0.5
2: 0.5, 2
1: 1
8c Helical mode
Rotation times
0.5, 0.6, 0.7, 0.8, 0.9, 0.4 (option), 0.5, 0.75, 0.5 ( cardiac option ) 0.4 (option), 0.5, 0.75,
8.11 for helical 0.75, 1, 1.5
1, 2, 3, 4 1.0, 1.5, 2.0 0.6 ,1, 1.5 1, 1.5
scanning [s]
Nominal z-axis
x-ray beam
8.12 collimations, 10, 20 1.2, 12, 18, 24, 2.4, 9.6, 19.2 20, 16, 12, 8, 4, 2 32, 24, 16, 12, 8, 4, 2
helical scanning
[mm]
Maximum z-axis
x-ray beam
8.13 collimation with 10 12 9.6 8 8
sub-millimetre
slices [mm]
Thinnest
acquisition width
available at full
8.14 0.625 1.5 1.2 5 2
z-axis x-ray
beam collimation
[mm]
32: 2, 4
Automatic mA
control
9.1 (AEC / mA Yes DoseRight CARE Dose 4D Real-EC SUREExposure3D
modulation)
software
- mA adjustment
9.2 Yes Yes, ACS Yes Yes Yes
for patient size
- mA adjustment
9.3 Yes Yes, Z-DOM Yes Yes Yes
along the z-axis
- mA modulation
9.4 Yes Yes, D-DOM Yes No Yes
during rotation
Method for
Using single or dual Using single or dual
9.7 system control of Single scout Single or dual Surview SPR & online control
scanogram scanogram
mA
Yes,
Operator review
9.8 mA table display of min, mean, max No Yes Yes
of AEC
mAs
Siemens SOMATOM
GE BrightSpeed Elite Philips Brilliance CT 16 Toshiba Activion 16 Toshiba Aquilion 16
Emotion 16
10a Parameters
Minimum
reconstruction Std & high res: 5
10.1 9.6 5 5 5
field of view UHR: 2.5
[cm]
Siemens SOMATOM
GE BrightSpeed Elite Philips Brilliance CT 16 Toshiba Activion 16 Toshiba Aquilion 16
Emotion 16
Maximum
reconstruction Std & high res: 50
10.2 field of view at 50 50 50 50
full image UHR: 25
quality [cm]
Extended
10.3 scan field of No No Option 70 Not available Not available
view [cm]
Reconstructio 512²
10.4 512 x 512 512 x 512 512 x 512 512 x 512
n matrices 768², 1024² [optional]
32: 4
24: 6
20: 1.25, 2.5, 5, 10 19.2: 1.2, 2.4, 4.8, 9.6 20: 5
Sequential 16: 4
image slice 10: 0.625, 1.25, 2.5, 5, 16.0: 8, 16 16: 4
options 12: 3
10 0.6 - 12.0 14.4: 1.2, 3.6, 4.8 12: 3
10.5 (beam 8: 2, 8
15: 3.75, 7.5 9.6: 0.6, 1.2, 2.4, 4.8, 9.6 8: 2
collimation : 6: 6
slice widths) 5: 1.25, 2.5, 5 7.2: 0.6, 1.2, 2.4, 7.2 4: 1
[mm] 4: 1, 4
1.25: 0.625, 1.25 2.4: 0.6, 1.2, 2.4 2: 0.5
2: 0.5, 2
1: 1
Siemens SOMATOM
GE BrightSpeed Elite Philips Brilliance CT 16 Toshiba Activion 16 Toshiba Aquilion 16
Emotion 16
Minimum
reconstruction
interval of
10.7 images, 0.1 0.1 0.1 0.5 0.1
helical
scanning
[mm]
10 b Functionality
Number of
prospective Filter, Matrix Size, Slice
reconstruction Thickness, Slice
10.8 3 8 20 20
s that can be Increment, X/Y Offset,
set up in Zoom
protocol
Siemens SOMATOM
GE BrightSpeed Elite Philips Brilliance CT 16 Toshiba Activion 16 Toshiba Aquilion 16
Emotion 16
Any delay in
either
scanning or
10.9 reconstruction No No No No No
when
performed
concurrently
Simultaneous
10.1 scanning and Yes, with 2nd monitor
Yes Yes No Yes
0 routine option
analysis
Simultaneous
scanning and
10.1
archiving Yes Yes Yes No Yes
1
and/or hard
copying
Simultaneous
scanning and
10.1 transfer to
Yes Yes Yes No Yes
2 second
console/works
tation
Siemens SOMATOM
GE BrightSpeed Elite Philips Brilliance CT 16 Toshiba Activion 16 Toshiba Aquilion 16
Emotion 16
10 c Performance
Standard
reconstruction
rate for 512 x Approx. 6 std
10.1
512 images - Up to 6 16 6 4
3 Approx. 14 option
head [images
per second
(ips)]
Standard
reconstruction Approx. 6 std
10.1
rate for 512 x Up to 6 16 6 4
4 Approx. 15 option
512 images -
body[ips]
Reconstructio
10.1
n upgrade Up to 16 ips RapidView 6 is standard Not applicable No No
5
options
Fastest
10.1
reconstruction 6 Up to 20 (optional) 16 6 4
6
rate [ips]
Reconstruction
MUSCOT (multislice MUSCOT (multislice
algorithm -
11.1 Filtered Back Projection Filtered Back Projection Filtered Back Projection cone beam cone beam
sequential
tomography) tomography)
scanning
Reconstruction
approach - 2D 2D
11.2 2D 2D 2D
sequential Back projection Back projection
scanning
Cone beam
reconstruction -
11.3 CrossBeam Correction Not available No No No
sequential
scanning
MUSCOT MUSCOT
Reconstruction
Cone Beam
algorithm - GE Hyperplane & Adaptive Multiple Plane T-COT T-COT
11.4 Reconstruction Alogirthm
helical CrossBeam Reconstruction (AMPR)
(COBRA) (true cone beam (true cone beam
scanning
tomography) tomography)
Reconstruction
approach - 2D Filtered 2D Back projection 2D Back projection
11.5 2D Back Projection 3D Back Projection
helical Backprojection Volume Volume
scanning
Cone beam
Cone beam for slice ≤2
reconstruction -
11.6 CrossBeam Correction Yes mm @16x0.6 or ≤4 mm Yes Yes
helical
@16x1.2 mm
scanning
Low signal
11.7 Yes Yes, Adaptive Filtering Yes Yes Yes
correction
Motion
11.8 Yes Yes, with overscan Yes No No
correction
Beam
11.1 PFO is on in all soft-tissue
hardening All head protocols Yes, Protocol dependent Kernel Kernel
0 head modes
trigger
Noise
11.1 reduction
Yes, GE proprietary Yes, Adaptive Filtering 3D adaptive filter Yes Yes
1 techniques -
raw data based
Noise
11.1 reduction 2D advanced smoothing
Yes, GE proprietary Yes, Image Enhancement Yes Yes
2 techniques - algorithm
image based
Other artefact
11.1
reduction Yes, GE proprietary Info. not available No No No
3
methods
Power-on to
12.3 ready to warm- < 10 min Info. not available 4 2 2
up [mins]
Standy by to
12.4 ready to warm- <1 <1 0 0 0
up [mins]
Tube warm-up
12.5 1 min 3 0 5 5
time [mins]
Detector
12.7 calibration at Air cal recommended Yes, air cals Part of warm up Automatic Automatic
warm-up
Time to
perform
12.8 detector < 30 25 6 2 2
calibrations
[mins]
Recommended
frequency for
12.9 Once every 24hrs Weekly Daily Daily Daily
detector
calibration
12.1 Automated QC
No Std QA Protocols Yes No No
0 tests
12.1 Self-check
Not applicable Daily Daily Not applicable Not applicable
1 frequency
Any additional
12.1 calibration by
No Info. not provided Not required Not required Not required
2 the
radiographer
Time to
perform these
12.1
additional Not applicable Info. not provided Not required Not applicable Not applicable
3
calibrations
[mins]
MTF0: 24
In plane spatial MTF2: 21.4
MTF2: 23 MTF2: 21.4
resolution for MTF0: 15.4 MTF2: 15.6 (120 kV, 200mA, 2
sharpest MTF10:16 (120 kV, 300mA, 2
13.1 MTF10: 13.0 mm slice width,
clinical (130 kV, 60 mA, 2.4 mm, mm slice width, 0.5s,
MTF50: 8 0.75s, 240 mm FOV,
algorithm (10cm DFOV, Edge alg) 1.0 s, U90s) 240 mm FOV, FC90
FC90 alg, (small
[lp/cm] [0.75s, 2.5mm, UHR, alg, (small focus)
focus)
25cm FOV, E]
MTF0: 18.1
Longitudinal (z- MTF4: 14.2
axis) spatial
resolution for MTF10: 12.2
13.2 sharpest 15 lp/cm @ cut off MTF2: 15 Info not available Info not available
clinical MTF50: 7.3
algorithm (0.625mm, 0.562 pitch,
[lp/cm] 0.06mm interval, Detail
alg., small focus)
Contrast
resolution: 2mm @ 0.3 % contrast 5 mm @ 0.3% at 16 mGy 4 mm @ 0.3% @
5mm @ 0.3% w/o any
smallest rod noise filter (130 kV, 92 mAs, 0.6 s, 3 mm @ 0.3% @ 10.0 mGy: 120kV, 90
13.3 4mm @ 0.3% contrast
size 10mm slice, Sequence 16.4 mGy mAs, 10 mm, FC41
discernable (in 120kV, 100mAs, 10mm 5mm @ 0.3% contrast with adaptive filter
scan, B31s)
plane) [mm]
CT number
13.5 accuracy +3 ±4 2 3 3
(water) [+/-HU]
CT number
13.7 accuracy (air) -1000 +/- 10 Info. not available 10 3 3
[+/-HU]
CT number
HU of non-water materials
13.8 accuracy Info not available Acrylic (120 kV) = 140±15 Info not available Info not available
depends on kV
(bone) [+/-HU]
Scanner
settings for std 120 kV 120 kV
head scan, 120 kV, 200 mAs, 130 kV, 133 mA, 1.5 s, 100 mA 300 mA
13.9 used for 120 kV, 20 mm beam 16x1.5mm, 6mm ST, 250 4.8mm slice (coll. 12x1.2
quoting cm FOV, EB Filter mm), Large Focal Spot 1 s/rot 1 s/rot
CTDI100 16 mm 16 mm
values
CTDI100 at
13.1 isocentre in air
76.3 Info. not available 35.1 Info not available Info not available
0 [mGy /
100mAs]
CTDI100 at
13.1 centre of head 20.8 (Bow-tie filter) 18.4 (Bow-tie filter)
17.9 12.25 23.7
1 phantom [mGy 9.9 (Flat filter) 10.0 (Flat filter)
/ 100mAs]
CTDI100 at
periphery of 23.6 (Bow-tie filter) 21.5 (Bow-tie filter)
13.1
head phantom 18.7 13.3 24.7
2 10.6 (Flat filter) 11.1 (Flat filter)
[mGy /
100mAs]
Scanner
settings for std 120 kV 120 kV
body scan, 120 kV, 240 mAs, 110 kV, 150 mA, 1.0 s, 100 mA 300 mA
13.1
used for 120 kV, 20 mm beam 4x6mm, 6mm ST, 350 cm 4.8 mm slice (coll. 12x1.2
3 1 s/rot 1 s/rot
quoting FOV, B Filter mm), Large Focal Spot
CTDI100 16 mm 16 mm
values
CTDI100 at
13.1 isocentre in air
63.9 Info. not available 24 Info not available Info not available
4 [mGy /
100mAs]
CTDI100 at
13.1 centre of body 6.0 (Bow-tie filter) 6.6 (Bow-tie filter)
5.4 4.17 4.7
5 phantom 3.3 (Flat filter) 3.5 (Flat filter)
[mGy/100mAs]
CTDI100
13.1 periphery of 12.3 (Bow-tie filter) 14.8 (Bow-tie filter)
11.2 7.71 9.1
6 body phantom 6.9 (Flat filter) 7.3 (Flat filter)
[mGy/100mAs]
32 : 38 ± 9.5
Std. Body Collimation: 20 : 23 ± 5.0
24 : 28 ± 6.0
Nominal FWHM: 16.05 16 : 18 ± 5.0
Dose profile ±10% for all collimation in 16 : 20 ± 5.0
mm 12 : 16 ± 4.3
13.1 FWHM, air
Info. not available 12 : 17 ± 4.3
7 measured at Nominal Base Width (Film 8 : 12 ± 3.6
isocentre [mm] Measurement of 8 : 12 ± 3.6
Acceptance Test): 18.5 4 : 8 ± 2.4
mm 4 : 8 ± 2.4
2 : 6 ± 2.0
2 : 5 ± 2.0
Visual breath- Yes, with count down Yes, scan timing bar on
14.5 Yes Yes Yes
hold indicator timer console
Function of
each monitor Acquisiton control, patient Scan and Display Scan and Display
(state if any One for Operation, data base, image viewing,
All functionality can be console console
14.7 function is the other for review and clinical software
perfomed on 1 monitor (Reconstruction from (Reconstruction from
restricted to processing evaluation packages,
only one filming both) both)
monitor)
Display type
14.8 LCD LCD LCD LCD LCD
(LCD/CRT)
Diagonal
dimension of
14.9 50.8 (20 inches) 48 48 48.3 48.3
image screen
[cm]
Image area
14.1 512 x 512, 768 x 768,
matrix 1280 x 1024 1024 x 1024 1024x1024 1024x1024
0 1024 x 1024
dimensions
Usual range of
14.1 Dependent on application
CT Number -1024 - 3071 -1024 - +3071 -1024 - 8191 -1024 - 8191
1 and anatomy
displayed [HU]
Extended
14.1 range of CT
-30719 - 32767 -1024 - +3072 -10240 - +30710 -32768 - 32767 -32768 - 32767
2 number display
[HU]
Accuracy of
distance
14.1 < 2 times image pixel
measurements Info. not available 2 x pixel size 0.01 - 0.98 0.01 - 0.98
3 size
in xy-plane
[mm]
Geometric
14.1 Efficiency Not required - always >
Yes No Yes - if < 70% Yes - if < 70%
6 displayed on 70%
console
15 a Console computer
15.1 Location In console Info. not available In console / under desk Under desk Under desk
Make and
15.2 HP XW8200 Dell, Precision 690 Siemens Fujitsu TOSHIBA TOSHIBA
model
Operating
15.3 Linux Windows XP Windows XP Windows Windows
system
Number and
15.4 2 Xeon One Dual Core Dual Core Pentium Dual Dual
type of CPU
CPU speed
15.5 2 x 3.2 GHz 2 x 2.66 GHz 2 x 2.66 GHz 2 x 3.06 GHz 2 x 3.06 GHz
[GHz]
Standard
amount of 1.5 (scan)
15.6 4 2 2 3
computer RAM 3.0 (display)
[GB]
Maximum
amount of 1.5 (scan)
15.7 4 4 Not applicable 3
computer RAM 3.0 (display)
[GB]
15 b Reconstructor
15.8 Location In console Console Location dependant Under desk Under desk
Make and
15.9 GE custom made Info. not available Siemens TOSHIBA TOSHIBA
model
15.1 Operating
Info. not available Windows Windows XP Windows Windows
0 system
Standard
15.1 amount of 1.5 (scan)
Info. not available 3 3 3
3 computer RAM 3.0 (display)
[GB]
Maximum
15.1 amount of 1.5 (scan)
Info. not available 3 Not applicable 3
4 computer RAM 3.0 (display)
[GB]
Standard total
16.1 hard disc 291 146 965 297 450
capacity [GB]
Hard disc
capacity for
16.3 146 146 146 73 146
image storage
[GB]
Hard disc
capacity for
16.4 250000 250000 240000 100000 200000
image storage
[# images]
Capacity for
16.5 raw data 72 640 600 144 (3600 rotations) 144 (3600 rotations)
storage [GB]
Local
16.6 Un-compressed Info. not available No No No
compression
Image
compression
16.7 Not applicable Info. not available Not applicable Not applicable Not applicable
algorithms and
rates
16 b Image archive
Disc burning
16.9 Standard Yes, option Dicom part 10 standard Standard Standard
facility
17 a Standard workstation
Cerebral Blood
Perfusion
Cerebral Blood
Calcium Scoring
Perfusion
Cardiac Functional
Whole range of clinical Calcium Scoring
Analysis
options for CT in Dental
Workstation oncology, cardiology, Dental
software vascular imaging and Lung CAD (3D WS
17.2 All clincal applications Details on application SUREPlaque
upgrade neurology - Software only)
options upgrade path to further Lung CAD (3D WS
versions with a 18 Fusion (3D WS Only)
only)
months compatibility EP Electrophysiology
Fusion (3D WS Only)
(3D WS only)
EP Electrophysiology
Vessel Probe
(3D WS only)
Vessel Probe
Workstation
Comes by default with 4
hardware Additional RAM, Hard
17.3 GB RAM, expandable to Upgrade PC / RAM 3D WS 3D WS
upgrade Disk
16 GB
options
Operating
17.5 Linux Windows XP Windows XP Windows Windows
system
Amount of
computer RAM 3 (MC) 3 (MC)
17.8 4 4 6
supplied as 8 (3D WS) 8 (3D WS)
standard [GB]
Maximum
amount of 3 (MC) 3 (MC)
17.9 16 4 8
computer RAM 8 (3D WS) 8 (3D WS)
[GB]
Disc storage
17.1 capacity 297 (MC) 450 (MC)
300 292 220
0 supplied as 500 (3D WS) 500 (3D WS)
standard [GB]
Maximum total
17.1 hard disc 297 (MC) 450 (MC)
300 438 (optional) 220
1 storage 500 (3D WS) 500 (3D WS)
capacity [GB]
Speed of
17.1 connections to 1000 (MC) 1000 (MC)
1000 10 / 100 / 1000 1000
2 networks 1000 (3D WS) 1000 (3D WS)
[Mbits/s]
17 b Image display
17.2 Colour or
Colour Colour Colour Colour Colour
0 greyscale
17.2 Monitor
Landscape Landscape Landscape Landscape Landscape
1 orientation
Standard CT
17.2
number range -1024 - 3072 Dependent on protocol -1024 - +3071 -1024 - 8191 -1024 - 8191
6
[HU]
Extended CT
17.2
number range -32738 - 32767 -1024 - +3072 -10240 - +30710 standard -32768 - 32767 -32768 - 32767
7
[HU]
Accuracy of
distance
17.2
measurements 0.2 Info. not available 2 x pixel size 0.01 - 0.98 0.01 - 0.98
8
in xy-plane
[mm]
Ultrafast connection
(Directconnect) that Yes (3D WS) Yes (3D WS)
17.2 Shared image Yes, images reside on Optional : Satellite
allow several Functionality based Functionality based
9 database server console (CTWP)
workstations to share on configuration on configuration
hard disk
Unlimited licenses - up to
"Per seat" licence basis the capacity of the server
17.3 Multi-use
- special financial offer Per seat Concurrent Concurrent
4 licensing 10k, 30k, 60k slices
for multilicence
simultaneously
18 a Image transfer
Background function
Rate of image 12 ips 12 ips
during reconstruction from
transfer 6 std (up to 16 option)
scanner to Siemens Maximum 50 ips Maximum 50 ips
18.2 scanner to 512 x 512 images (GB Info. not available
workstation - otherwise 25 (enhanced DICOM, (enhanced DICOM,
workstation network)
ips with no compression optional) optional)
[ips]
applied
Background function
Rate of image during reconstruction from 12 ips 12 ips
6 std (up to 16 option)
transfer scanner to Siemens Maximum 50 ips Maximum 50 ips
18.4 512 x 512 images (GB Info. not available
scanner to PACS - otherwise 25 ips (enhanced DICOM, (enhanced DICOM,
network)
PACS [ips] with no compression optional) optional)
applied
Workstation to
18.5 Yes Yes - manual Operator triggered Manual Manual
PACS
Rate of image
6 std (up to 16 option)
transfer
18.6 512 x 512 images (GB Info. not available 25 ips uncompressed 12 ips 12 ips
workstation to
network)
PACS [ips]
18 b Interoperability
18.1 Workstation
Standard Standard Standard Standard Standard
2 with PACS
Structured
18.1 Clinical Software
reports No Not available Not available Not available
4 dependant
available
18 c IHE
1) Acquisition Modality +
options:
- Assisted Acquisition
Protocol Setting Acquisition Modality (
Patient Based Worklist
Broad Worklist - PPS Exception Query, Broad Worklist
Actors Management Query, Modality Group
18.1 Patient Based Worklist
implemented - Case), Image Acquisition Acquisition
6 Query - Patient Based Worklist
scanner Display,Evidence
Query
Modality Group Case Creator,Acquisition
- Broad Worklist Query Modality, Image
Display,Print Composer
2) Image Display
3) Portable Media Creator
and Reader
Image Display
Image Creator Image Display, Evidence
1) Image Display Creator, Print Composer,
Actors Portable Media Creator
18.1 Report Creator, Report
implemented - 2) Portable Media Creator Image Display Image Display
7 Client Authentication Reader, image Display,
workstation and Reader
Agent Portable Media Creator,
Report Reader
Personnel White Pages
Consumer
18 d DICOM objects
CT image
18.1
object (1993) Yes Yes Yes Yes Yes
8
supported
CT 'enhanced'
18.1 multiframe
No No Standard Not available Not available
9 object (2005)
supported
CT object
18.2
supported as Image Object (1993) Image Object (1993) CT image object (1993) 1993 1993
0
standard
18.2
Storage SCU, SCP Yes Yes SCU, SCP SCU, SCP
1
18.2
Print SCU Yes Yes SCU SCU
2
18.2
Query / retrieve SCU, SCP Yes Yes SCU, SCP SCU, SCP
3
18.2 Modality
SCU (Connect Pro) Yes Yes SCU SCU
4 worklist
18.2 Performed
SCU (Connect Pro) Yes Yes SCP SCP
5 procedure step
18.2 Storage
Yes Yes Yes SCU SCU
6 commitment
18.2
Storage SCU/SCP Yes Yes SCU/SCP SCU/SCP
7
18.2
Print SCU Yes Yes SCU SCU
8
18.2
Query / retrieve SCU/SCP Yes Yes SCU SCU
9
Modality
18.3
worklist No No No Not available Not available
0
management
18.3 Performed
No No MPPS service as SCU Not available Not available
1 procedure step
18.3 Storage
SCU Yes Yes SCU SCU
2 commitment
19.2 Insufflators Option Info. not available CO2 (option) Local Purchase Local Purchase
Stereotactic
19.4 No Info. not available Option Local Purchase Local Purchase
head restraint
Restraining
19.5 Standard Velcro Velcro (standard) Velcro (standard) Velcro (standard)
straps
Slicker mattress
19.6 Option Standard Standard Standard Standard
cover
Bariatric couch
19.7 No Yes (optional) No No No
option?
Bariatric couch
19.8 Not applicable 295 Not applicable Not applicable Not applicable
weight limit [kg]
Monitor Monitor
Hand, footswitch, table
CT fluoroscopy Monitor, Exposure
19.9 float, ceiling suspended Option Exposure Control Exposure Control
hardware Control, Couch Control
monitor, 12fps option
Couch Movement Couch Movement
Hard-copy
19.10 Option Info. not available Option Local Purchase Local Purchase
imaging device
Gantry mounted
19.11 Option ECG Monitor Option No No
ECG
20 a Operator console
Maximum and
minimum
20.1 intensity Option Yes Yes Yes Yes
projections (MIP,
MINIP)
Multi-planar
20.2 reconstruction Standard Yes Yes Yes Yes
(MPR)
Planes available
20.3 Freely user defined Any Yes Yes Yes
in MPR
Shaded surface
20.4 Yes Yes Yes Yes Yes
display (SSD)
Volume
20.5 rendering Option Yes Yes Yes Yes
software
20 b Standard workstation
Maximum and
minimum
20.7 intensity Yes Yes Yes Yes Yes
projections (MIP,
MINIP)
Multi-planar
20.8 reconstruction Yes Yes Yes Yes Yes
(MPR)
Planes available
20.9 Yes Any Yes Yes Yes
in MPR
Shaded surface
20.10 Yes Yes Yes Yes Yes
display (SSD)
Volume
20.11 rendering Yes Yes Yes Yes Yes
software
Virtual
colonoscopy: Syngo Colonography CT Colon CAR (3D WS Colon CAR (3D WS
21.4 OC: Option (CTC) Yes - Virtual Colnoscopy
Advanced (OC & WS) Option) Option)
package
Brain perfusion Yes - Advanced Brain Syngo Neuro Perfusion CBP (MC Option) CBP (MC Option)
21.5 OC: Option (Perfusion)
software Perfusion CT (OC, & WS) CBP (3D WS Option) CBP (3D WS Option)
Computer
OC: n/a, WS: Option Yes - Lung Nodule Syngo Lung CAD (OC & Lung CAD (3D WS Lung CAD (3D WS
21.7 assisted
(LungVCAR) Assessement WS) Option) Option)
reading: Lung
Computer
OC: n/a, WS: Option Syngo Colonography PEV Colon CAR (3D WS Colon CAR (3D WS
21.8 assisted Yes - Polyp CAD
(ColonVCAR) (OC & WS) Option) Option)
reading: Colon
21.1 OC: not available, WS: Syngo Image Fusion (OC Fusion 7D (3D WS Fusion 7D (3D WS
Image fusion Yes - Master Match
1 Option & WS) Option) Option)
Prospective
ECG-triggered
Smartscore 4/ VCAR
22.1 cardiac scanning Yes, Heartbeat-CS Pro HeartView option Yes Yes
Option
for calcium
scoring
Yes, Heartbeat-CS,
Calcium scoring Smartscore 4/ VCAR
22.2 Agatston, Mass, Syngo Calcium Scoring Option Option
software Option
Average
Prospective
ECG-triggered
22.3 cardiac scanning No Not available HeartView No No
for coronary
angiography
Retrospective
CardIQ Pro,SnapShot
ECG-gated
Segment; SnapShot Yes, Rate Responsive
22.4 cardiac scanning HeartView option No Yes
Burst; SnapShot Burst CV Toolkit
for coronary
Plus
angiography
Coronary
angiography Yes, Comprehensive Syngo Circulation
22.5 Option No Option (3D WS)
vessel analysis Cardiac Analysis option
software
Temporal
resolution for (200 option)
22.6 single sector 250 210 250 375
reconstruction 250
[ms]
Automatic
adapted Yes according to heart
22.8 Always use 0.4 No Not applicable Yes
selection of rate
rotation time?
Maximum mA
22.9 available for 440 500 345 Not applicable 500
cardiac scanning
ECG-gated mA
modulation (max
22.10 80 80 80 Not applicable Not available
reduction from
peak mA) [%]
Pitches available
22.11 for cardiac 0.24 - 0.325 0.2 - 0.3 0.27, 0.31 Not applicable >0.125
scanning
Automatic
adapted Yes, according to heart Yes, based on heart
22.12 No Not applicable Yes
selection of rate rate
pitch?
Multi-sector
22.13 Yes, Up to 4 sector Yes, up to 5 sectors Yes - 2 Not applicable 5
reconstruction
kV = 120 kV
For 70 kg
patient, 60 bpm Rotation Time = 0.4s
give typical scan 120 kV, 300 mA, 0.4
mAs = 800 mAs
parameters and s/rot, Pitch 0.2
Pitch = 0.2
(a) effective 16 * 1 mm, scan length
22.16 Info not available Info not available Not applicable
dose without Scan Length = 120 mm 120 mm
ECG modulation
Effective Dose = 14 a) Effective dose 12.2
(b) effective mSv mSv
dose with ECG
modulation Effective Dose = 8 mSv
(with modulation)
Radiotherapy
23.1 planning table Yes (option) Info. not available Option Option Option
top
Carbon fibre
23.2 Yes (option) Info. not available Option Not available Not available
breast board
Dorado: 1 Sagittal
(moving); 2 Lateral
(fixed or moving); [1
External transverse (fixed or
23.4 positioning laser moving)] Info. not available Option Not available Not available
options A2J: 1 Sagittal
(moving); 2 Lateral
(fixed or moving); 2
frontal (moving)
Respiratory GE Advantage 4D
23.5 Info. not available Option Not available Not available
gating software Option
Radiotherapy CT
23.7 simulation AW Sim MD (option) Info. not available Option Option on WS Option on WS
software
Dual-energy /
24.1 multi-energy No Info. not available No No No
capabilities
Dual-energy
24.2 No Info. not available No Not applicable Not applicable
method used
Dual-energy
24.3 analysis No Info. not available No Not applicable Not applicable
software
Contrast media
Region of interest
bolus tracking;
24.6 Manual Info. not available threshold set by the HU HU
trigger
user
mechanism
4 images:
Top left - plan slice
Top right - enhancing
Contrast media image
Region of interest Image and graph Image and graph
24.7 bolus tracking; Info. not available
Bottom left - real time graph display display
user monitoring
enhancement graph
Bottom right - real time
scan times, elapsed
time clock
Contrast media
Yes, to stop & start
24.8 bolus tracking; Info. not available Yes, to force the scan Abort or Start Abort or Start
scan
manual override
32
16 24
CT fluoroscopy; 1 x 2.4, 3 x 2.4, 3 x 4.8, 12 16
beam 3 x 7.2,
24.12 1.25, 2.5, 5, 7.5 Info. not available 8 12
collimations
[mm] 1 x 7.2 and 1 x 10 4 8
2 4
2
(32) 8,16, 8
(16) 4, 8, 4 (24) 6, 12, 6 or 8, 8, 8
CT fluoroscopy;
24.14 other viewing Available Info. not available no No No
options
CT fluoroscopy;
Extension Operating Extension Operating
24.15 in-room scan Footswitch Info. not available Foot switch
Panel Panel
controls
Maximum scan
coverage for
24.21 perfusion scans 10 (@0.625) 48 (with Jog Scan) 19.2 (Head & Body) 20 32
- sequential
scanning [mm]
Maximum scan
coverage for
24.22 perfusion scans Not available Info. not available 19.2 (Head & Body) Not applicable Not applicable
- helical
scanning [mm]
Power supply
380 - 440 V , 3Phase,
25.2 requirements Refer to above Info. not available Not applicable Not applicable
N & E 50Hz
(other)
Plumbing
25.3 Not applicable None No Not applicable Not applicable
requirements
Minimum floor
1290 (area as whole
25.4 load-bearing Info. not available 428.2 ± 4% 670 732
area of gantry)
[kg/m2]
Minimum floor
area required for
25.5 20.4 19.95 14 Info not available Info not available
gantry & couch
[m²]
Minimum floor
area required for
25.6 Not applicable Info. not available 0.75 Info not available Info not available
other in-room
system [m²]
Recommended
floor area 22 (std) 27 (std)
25.7 22.6 35.69 Site dependant ≥14
required for 21(short) 25 (short)
scanner [m²]
Minimum floor
area required for Site Dependant - PC
25.8 12.1 5 Info not available Info not available
the control Tower 0.14m²
console [m²]
Gantry
dimensions
1930 x 2040 x 1018; 2030 x 2390 x 940; 1820 x 2300 x 690; 1910 X 2070 X 900;
25.9 (H x W x D 1950x2330x960; 1750
1770 1896 1200 1280
[mm]) and
weight [kg]
Control console
dimensions 680-885 x 1240 x Site Dependant - PC 700 x 450 x 815; 85 700 x 450 x 815; 100
25.11 760 x 1190 x 910; 56
(H x W x L [mm]) 1135-1235; 195 tower 450 x 620 x 220 700 x 450 x 815; 100 700 x 450 x 815; 140
and weight [kg]
Power unit
dimensions
25.12 (H x W x D 1062 x 700 x 550; 350 510 x 380 x 560; 34 820 x 750 x 350 870 x 669 x 800 870 x 669 x 800
[mm]) and
weight [kg]
Cooling unit
dimensions
25.13 (H x W x D Not required None Not applicable Not applicable Not applicable
[mm]) and
weight [kg]
Supplementary
25.14 Not required None Not applicable Not applicable Not applicable
plant
Supplementary
plant unit
dimensions
25.15 Not required None Not applicable Not applicable Not applicable
(H x W x D
[mm]) and
weight [kg]
Supplementary
25.16 Not applicable None Not applicable Not applicable Not applicable
plant locations
System cooling
25.18 Output to air in room Air cooled Air Air Air
method
Environmental
requirements
(max/min 15 - 26° C
20 - 26 ºC, 40 - 80% 20 - 26 ºC, 40 - 80%
25.20 temperature, 30 - 60% relative 18 - 24 C, 35 - 70% 18 - 30 °C, 20 - 85 %
relative humidity relative humidity
humidity) in humidity
scanner control
room
Environmental
requirements 15 - 26° C
(max/min 20 - 26 ºC, 40 - 80% 20 - 26 ºC, 40 - 80%
25.21 30 - 60% relative 15 - 24 C, 35 - 70% 18 - 30 °C, 20 - 85 %
temperature, relative humidity relative humidity
humidity) for humidity
control computer
Environmental
requirements
(max/min 18 ~ 26° C, 30 ~ 60% 20 - 26 ºC, 40 - 80% 20 - 26 ºC, 40 - 80%
25.22 15 - 24 C, 35 - 70% 18 - 30 °C, 20 - 85 %
temperature, relative humidity relative humidity relative humidity
humidity) for
reconstructor
Environmental
requirements
(max/min 18 ~ 24° C, 30 ~ 60% 20 - 26 ºC, 40 - 80% 20 - 26 ºC, 40 - 80%
25.23 15 - 24 C, 35 - 70% 18 - 30 °C, 20 - 85 %
temperature, relative humidity relative humidity relative humidity
humidity) for
workstation
Connection to
1000 MPS / 1000 Base
25.24 networks 100 1000 1000 100
T
[Mbits/s]
RJ45 x 4
Network RJ45 RJ45
25.25 connections (for hospital network, 10 MPS / 10 Base T 1 per PC
required Injector, EKG monitor, CAT5E or CAT6 CAT5E or CAT6
workstation)
Network
25.26 TCP/IP LAN LAN LAN LAN
requirements
Static network
25.27 addresses Recommended, 1 Yes Yes - one per PC Yes Yes
required
Standard hours
26.1 8:30 - 17:00 Info. not available 09:00 - 17:00 Info not available Info not available
engineer call-out
Out of hours
Available/ Not on Available as a costed
26.4 planned Info. not available Info not available Info not available
standard contract option
servicing
Included in a
Detector Available on standard
26.7 Info. not available Performance TOP Info not available Info not available
replacement contract
contract
Included in a
Available on standard
26.8 Main computers Info. not available Performance TOP Info not available Info not available
contract
contract
Included in a
Monitors and Available on standard
26.9 Info. not available Performance TOP Info not available Info not available
display devices contract
contract
Included in a
Separate contract for
26.10 Workstations Info. not available Performance TOP Info not available Info not available
workstation
contract
User / Included in a
Available on standard
26.11 application Info. not available Performance TOP Info not available Info not available
contract
support contract
Included in a
Available on standard
26.12 Parts Info. not available Performance TOP Info not available Info not available
contract
contract
Frequency of
26.16 Quarterly Info. not available 2 per annum Info not available Info not available
service
Duration of
26.17 service [+/- 0.5 1 Day Info. not available 0.5 day Info not available Info not available
day]
What elements
of scanner /
plant are not As listed (see Notes 6
26.20 Air conditioning Info. not available Info not available Info not available
covered in and 7)
standard
support?
26 c Reliability
What is the
expected life
span of the Not less than 7 years
26.21 10 yrs Info. not available Info not available Info not available
product ? (for (see Note 8)
five 8 hour days
per week)
Detectors should
What is the
typically have a life
26.24 expected Life of system Info. not available Info not available Info not available
span of 7 years (see
detector life ?
Note 10)
When first
27.1 Nov 2006 2003 Sept 2006 Info not available Info not available
installed in UK ?
Any adverse
incident reports
27.3 None Info. not available No No No
issued on this
model ?
What is the
audible noise
output of the
27.4 product under Meets IEC 60601 Info. not available 66dB(A) @ 1m 60 60
normal operating
conditions?
[dB(A)]
What is the
audible noise
output of the
27.5 product under Meets IEC 60601 Info. not available 63db(A) @ 1m 60 60
standby or idle
conditions?
[dB(A)]
Is vibration
generated by the
27.6 product under None Info. not available No No No
normal operating
conditions?
Do the waste
outputs from
product under
normal operating
conditions
27.8 Not applicable Info. not available No No No
require any
regulatory
permits,
consents or
licences?
Any water
27.9 No Info. not available No No No
consumption?
To what extent
is the product
designed for use Not specifically Not specifically
27.12 Info. not available Info. not available Complies
in developing designed designed
health
economies?
What colours
27.13 No color selection Info. not available White
are available
What is the
typical power
consumption 10 10
28.2 when not 20 kVA Info. not available ≤3 kVA
scanning? [kVA] (PF > 0.9) (PF > 0.9)
What is the
power
consumption 10 10
28.3 15 kVA Info. not available ≤3 kVA
when in 'Stand- (PF > 0.9) (PF > 0.9)
By' mode ?
[kVA]
What
percentage of
the packaging
As per European Union
28.6 used is derived Info. not available Info. not available 0.03 0.03
regulations
from recyclate
as opposed to
virgin materials?
What
percentage of
the packaging
28.7 Info. not available Info. not available 1 1 1
can be recycled
? (under normal
circumstances)
What is the
percentage of
28.11 recyclate utilised Info. not available Info. not available Info. not available Info. not available
in the product
manufacture?
What is the
environmental
Energy consumption /
28.12 impact of Info. not available Info. not available Use of electricity Use of electricity
heat output
operating this
scanner ?
What is the
carbon footprint
of the
manufacturing
28.13 Info. not available Info. not available Unknown 0.04 0.04
processes ?
[Equivalent
tonnes of CO2
emissions]
What is the
carbon footprint
of operating the
scanner?
28.14 Info. not available Info. not available Unknown 14.2 15.2
[Equivalent
tonnes of CO2
emissions per
annum]
Additional notes
GE
The following statement was received in response to the questionnaire in relation to the sustainability questions
(table 28).
GE sustainability statement
The GE Healthcare Renewable Resources team in Oak Creek, Wisconsin ensures that nearly 15 million pounds of old
medical equipment and parts are reused, redeployed or recycled each year.
GE Healthcare Renewable Resources in Oak Creek, Wisconsin earned the Governor's Award for Excellence in
Environmental Performance for reprocessing nearly 15 million pounds of old medical equipment and parts each year.
As part of Global Parts & Repair Solutions in the service organization of the Diagnostic Imaging business, Renewable
Resources is an example of our ecomagination commitment to improve the energy efficiency of our business
operations while also increasing company revenue.
The Oak Creek facility receives a wide variety of devices and parts, including:
The Renewable Resources team sees that 94 percent of that material is reused, redeployed or recycled. It's a major
achievement in environmental responsibility that also produces more than $3 million in annual revenue by reusing and
recycling parts.
"We receive more then 4,000 truckloads of material per year," says Geoff Hunt, global repair manager. "That includes
up to 500 CT and X-ray systems and 800 diagnostic imaging system cabinets.
"We send some imaging systems to repair suppliers for use as test beds for repair parts. Others we take apart. We
harvest the usable components to be placed into inventory or added to our repair pools. This is especially valuable for
older components that no longer have a supply chain. The remainder we separate into material streams for recycling."
"The proper handling of used electronic equipment has become a major issue worldwide," Hunt observes. "We've
created a way to deal with retired medical devices that demonstrates both sound business practice and excellent
corporate citizenship."
Siemens
The following notes are additional comments that were received in response to the questionnaire, that have been
placed here in order to keep the responses in the comparison tables concise.
Note1: It is not possible to accurately predict "The lifespan" of some components such as X-ray tubes as there are too many variables
governing their usage, for example - the number, type and frequency of examination procedures carried out, the number of exposures
with reference to made, the parameters selected for those exposures, the intervals between exposures, etc. all factors which are clearly unknown at this
question 3.11 time. However based on average usage and historical data X-ray tube should typically have a life span of 3 years.
Note 2: Siemens Remote Services utilises a secure broadband internet connection, such as N3, such that systems can benefit from real-time,
pro-active monitoring which detects and reports parameter deviations and potential system malfunctions to our specialists in the Uptime
Q26.5 Support Centre. In addition your systems can receive software updates, can be protected against viruses and can even support you as
a user with clinical information and e-training.
Note 3: When a call is received at the UPTIME Support Centre, there is immediate call clarification with all resources to assist problem
resolution available within the team: immediate access to remote clarification information provided via Siemens Remote Services, the
Q26.13 call history on SAP, technical as well as application experts, and spare parts logistics. However, if required, we would aim to have an
engineer to site within 4 working hours.
Note 4: Our logistics operation is as effective as it can be; parts ordered by 6.00 pm can be delivered to site before 8.00 am the following day.
In some cases it may even be possible to get spare parts delivered on the same day. Our network of spare-part drop boxes,
Q26.15 strategically placed at hospital sites throughout the country makes parts delivery to site as transparent and efficient as possible. This is
complemented by our ‘Assisted-Fit’ programme where a larger part can be delivered by appointment with all the necessary manpower
to install it.
Note 5: The STRATON X-ray tube is delivered with one year full warranty. There are no restrictions in terms of scan seconds or patient limits.
This offer is not restricted to a pro rata bases, the warranty is delivered for one full year. If the tube should fail during the warranty
Q26.19 and Q26.23 period, Siemens will replace every tube during these 12 months free of charge. e.g. If a tube fails in month 10 a new tube will be
supplied free of charge, which will come with a 2 month warranty thus fulfilling a total warranty period of 12 months.
In addition to the items listed under “All systems” the following modality specific items, are not included within the Maintenance Contract
(MC) and w
Note 8: This is very new technology which is being developed and refined continuously. Siemens Medical Solutions systems are supported in
line with the provisions in CRACOE - The contractor shall make available for purchase by the authority, or its agents, any necessary
Q26.21 spare parts and components to ensure that the goods can be maintained in safe working operation for a period of not less than 7 years
from acceptance and, thereafter, shall use all reasonable endeavors to provide the said spare parts and components for a further period
of 3 years.
Note 9: Data to date shows that the average life expectancy of Straton x-ray tubes life is between 2-3 years. Please note that it is difficult to
determine life expectancy for these items since it is heavily dependent on the number, type and frequency of examination procedures
Q26.22 carried out, the number of exposures made, the parameters selected for these exposures, the intervals between exposures, as well as
whether servicing has been requested and carried out at appropriate intervals - factors which are clearly unknown at this time.
Note 10: It is not possible to accurately predict "The lifespan" of some components such as detectors as there are too many variables governing
their usage, for example - the number, type and frequency of examination procedures carried out, the number of exposures made, the
Q26.24 parameters selected for those exposures, the intervals between exposures, etc. all factors which are clearly unknown at this time.
Detectors technology is very new and therefore usage data is limited, however our expectations are that detectors should typically have
a life span of 7 years. Detector cover can be included within a Performance TOP maintenance contract.