SONOLINE G20

US

Configuration
System
USB Printer Configuration

08647831
08647815

© Siemens AG 2004
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
Siemens Medical
Ultrasound
ejt DivisionSolutions
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: US06-201.843.01.01.02 English


Replaces: n.a. Doc. Gen. Date: 10.05
n.a.

Part No.: n.a,


sd
2 Revision / Disclaimer
1Revision / Disclaimer

Copyright
Copyright © 2004 by Siemens Corporation. All rights reserved. No part of this publication
may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into
any language or computer language, in any form or by any means, electronic, mechanical,
magnetic, optical, chemical, manual, or otherwise, without the prior written permission of
Siemens Corporation. Siemens Corporation reserves the right to change its products and
services at any time. In addition, this manual is subject to change without notice. Siemens
Corporation welcomes customer input on corrections and suggestions for improvements
to this manual.Although Siemens Corporation has attempted to ensure accuracy through-
out this manual, Siemens Corporation assumes no liability for any errors or omissions, nor
for any damages resulting from the application or use of this information.
Trademarks
ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE,
Adara, Antares, G20, G40, G50, G60S, Omnia and KinetDx are trademarks of Siemens
Corporation registered in the U.S. Patent and Trademark Office. CWS3000, DS3000,
WS3000, and WebPro are trademarks of Siemens Corporation.
Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet
Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are
registered trademarks of Netscape Communications Corporation. Java and the JavaScript
name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered
trademark of Adobe Systems, Inc.
Other products and brand names are trademarks of their respective owners.
Disclaimer
The service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie-
mens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
License Agreement
All computer programs copyright 1990-2004 by Siemens Corporation or its suppliers. Pro-
grams are licensed under the following agreement:
Siemens or its suppliers retain(s) ownership of and title to any computer program supplied
with the equipment and to the trade secrets embodied in such computer programs. Sub-
ject to the Buyer’s acceptance and fulfillment of the obligations in this paragraph, Siemens
grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any
computer program supplied with the Equipment that is necessary to operate the Equip-
ment solely on the medium in which such program is delivered for the purpose of operat-
ing the equipment in accordance with the instructions set forth in the operator’s manuals
supplied with the Equipment and for no other purpose whatsoever. Buyer may not
reverse-assemble, reverse-compile or otherwise reverse-engineer such computer pro-
grams nor may Buyer make a copy of such program or apply any techniques to derive the
trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms

SONOLINE G20 US06-201.843.01.01.02 Page 2 of 12 Siemens


10.05 sd Medical Solutions USA, Inc.
Revision / Disclaimer 3

of this license, the license granted by this paragraph shall terminate. Further, because
unauthorized use of such computer programs will leave Siemens without an adequate
remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to
restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s
suppliers of software is a direct and intended beneficiary of this end-user sublicense and
may enforce it directly against Buyer with respect to software supplied by such supplier,
and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indi-
rect, consequential, incidental or other damages arising out of the sublicense of the com-
puter programs supplied with the equipment.

Siemens US06-201.843.01.01.02 Page 3 of 12 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
4 Table of Contents
1- 0Table of Contents

1 _______ USB Report Printers _____________________________________________ 5

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
USB Report Printers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Connecting and Configuring a USB Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Alternate USB Printer Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Confirming Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

SONOLINE G20 US06-201.843.01.01.02 Page 4 of 12 Siemens


10.05 sd Medical Solutions USA, Inc.
USB Report Printers 5
2-
Overview
1USB Report Printers

A USB printer for printing patient reports can be connected to the ultrasound system at
one of the USB ports located on the I/O panel of the system. A USB port has the capabil-
ity for use with a USB printer compatible with the HP PCL 3 protocol.
Observe all safety precautions for connecting a non-medical device to the ultrasound sys-
tem.

NOTE The SET key on the control panel functions as a point-and-select


device (similar to a computer mouse) when used with the trackball.
To select an on-screen object such as a button or a check box, roll
the trackball to position the pointer (cursor) on the object and then
press the SET key on the control panel.

USB Report Printers 0

Report Printers

WARNING Accessory equipment connected to the USB port must be certified


according to the respective EN and IEC standards (for example, EN
60950 and IEC 60950 for data processing equipment and EN
60601-1 and IEC 60601-1 for medical equipment). Furthermore, all
configurations shall comply with the system standard EN
60601-1-1 and IEC 60601-1-1. Anyone who connects additional
equipment to the signal input or signal output port configures a
medical system and is therefore responsible that the system com-
plies with the requirements of the system standard EN 60601-1-1
and IEC 60601-1-1.
¹ Siemens can only guarantee the performance and safety
of the devices listed in the Accessories and Options
chapter of the System Reference (shipped with the ultra-
sound imaging system). If in doubt, consult Siemens ser-
vice department or your local Siemens representative.

Siemens US06-201.843.01.01.02 Page 5 of 12 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
6 USB Report Printers

WARNING Equipment connected to the ultrasound system and in the patient


environment must be powered from a medically isolated power
source or must be a medically isolated device.
¹ Equipment powered from a non-isolated source can
result in chassis leakage currents exceeding safe levels.
Chassis leakage current created by an accessory or
device connected to a non-isolated outlet may add to the
chassis leakage current of the ultrasound system.

WARNING Non-medical grade report printers cannot be used within a patient


environment.
¹ During use of a non-medical grade report printer or when
a non-medical grade report printer is connected to the
ultrasound system, the ultrasound system cannot in any
way be in contact with a patient.

NOTE Siemens cannot support or service any external devices connect-


ed to the USB port. Siemens cannot assume responsibility for
functionality beyond the scope of this specification.

NOTE Siemens does not specifically recommend any particular commu-


nication or analysis software; please contact a local vendor for as-
sistance with interpretation and analysis of the data transmitted by
the ultrasound imaging system.

Relevant standards for some non-medical electrical equipment may have limits for enclo-
sure leakage currents higher than required by medical standards. These higher standards
are acceptable only outside the patient environment. It is essential to reduce enclosure
leakage currents when non-medical electrical equipment is to be used within the patient
environment. Measures for reducing leakage current include use of a medically-approved
isolation transformer.
The ultrasound system has printer drivers compatible with non-medical grade report print-
ers. Siemens does not guarantee the performance and safety of any non-medical grade
report printer. Non-medical grade report printers do not fulfill the following safety require-
ments:
• EN 60601-1-1 and IEC 60601-1-1 (Medical Electrical Equipment, Part 1: General Re-
quirements for Safety).
• EN 60601-1-2 and IEC 60601-1-2 (Electromagnetic Compatibility of Medical Devices).
If a non-medical grade report printer is to be used with the ultrasound system, then you
must ensure mitigation is provided to meet all safety requirements. It is the responsibility
of the user to ensure that the ultrasound system in combination with the non-medical
grade report printer complies with safety requirements.

SONOLINE G20 US06-201.843.01.01.02 Page 6 of 12 Siemens


10.05 sd Medical Solutions USA, Inc.
USB Report Printers 7

Always use a medically-approved isolation transformer with a non-medical grade report


printer.
Mitigate risk with regard to the Electromagnetic Compatibility of Medical Devices require-
ment. A non-medical grade device must be designed, manufactured, and certified to meet
the same EMC (electromagnetic compatibility) requirements as the ultrasound system, or
other means must ensure that the overall EMC requirements are met.
Tab. 1 Manufacturers of Medically Approved Isolation Transformers

Manufacturer Web Site


Tripp Lite www.tripplite.com
Toroid Corporation of Maryland www.toroid.com
Dale Technology www.daletech.com

Mitigating the Risk of Connecting a Report Printer


To fulfill EN 60601-1-1 and IEC 60601-1-1 (Medical Electrical Equipment, Part 1: General
Requirements for Safety) requirements for non-medical peripheral equipment:
The non-medical peripheral equipment must be approved according to any other EN or
IEC standard (EN XXXXX or IEC XXXXX, e.g., equipment complying with EN 60348 and
IEC 60348, EN 60950 and IEC 60950, etc.)
The connection of non-medical peripheral equipment to your ultrasound system must
adhere to the following conditions:

1. Connect the ultrasound system to an AC power outlet within a medically used room
within the patient environment. A patient environment is defined as an area in which
medical examination, monitoring, or treatment of the patient takes place. The patient
environment is located 1.5 meters (or, in Canada and the U.S.A., 1.8 meters [6 feet])
around the patient location.
2. Connect the peripheral equipment AC power cable to a medically approved isolation
transformer.
3. Connect the medically approved isolation transformer to a main AC outlet either inside
or outside the patient environment. The peripheral device and medically approved iso-
lation transformer combination can be located:
- Within the same room as the ultrasound system, or
- In a non-medically used room.
For additional information and other possible combinations, please refer to the Medical
Electrical Equipment Standard EN 60601-1-1 or IEC 60601-1-1, Annex BBB.7,
Scenario 3c.

Siemens US06-201.843.01.01.02 Page 7 of 12 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
8 USB Report Printers

NOTE The above information is based on current EN 60601-1-1 and IEC


60601-1-1 standards, dated 2000-12. If your country"s regulatory
standards for medical equipment do not correspond to EN 60601-1
and IEC 60601-1, as well as EN 60601-1-1 and IEC 60601-1-1, your
local requirements may differ.

Patient Environment

Fig. 1: Patient Environment (Sheet 1)

SONOLINE G20 US06-201.843.01.01.02 Page 8 of 12 Siemens


10.05 sd Medical Solutions USA, Inc.
USB Report Printers 9

Fig. 2: Patient Environment (Sheet 2)

Connecting and Configuring a USB Printer 0

Before beginning this procedure, ensure that the USB printer is disconnected from the
ultrasound system.
To connect a USB printer:

1. Press the Standby button to power on the ultrasound system.


When the power on sequence is complete, the system displays the active image
screen.
2. Power on the USB printer and ensure that there is paper in the feeder tray.
3. Connect the USB cable from the printer to a USB port on the input/output panel on the
left side of the ultrasound system.
After a few seconds, the system displays the Found New Hardware Wizard dialog box.

NOTE If the system does not display the new hardware dialog box, then
it displays an error dialog box; use the “Alternate USB Printer con-
figuration” procedure.

4. Press the MENU key on the control panel to activate the trackball pointer.
5. Select Install from a list or specific location (Advanced) and then select Next.
The system displays a new dialog box.

Siemens US06-201.843.01.01.02 Page 9 of 12 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
10 USB Report Printers

6. Select “Don"t search. I will choose the driver to install.” and then select the Next but-
ton.
The system displays a new dialog box.
7. Ensure that the “Show compatible hardware” check box is not selected.
The system displays the Update Driver Warning box.
8. Select HP in the Manufacturer column, HP DeskJet in the Printers column, and then
select the Next button.
The system displays the Update Driver Warning dialog box.
9. Select Yes.
The system displays the Completing the Found New Hardware Wizard dialog box.
10. Select Finish.
The system displays the active image screen.
11. Select a USB printer and then print a report to confirm the installation.

Alternate USB Printer Configuration 0

Use this procedure the Found New Hardware Wizard dialog box does not automatically
display on the ultrasound system during an attempt to configure a USB printer.

1. Press the MENU key to activate the trackball pointer.


2. Select the Cancel button until the system displays the following message: Do you want
to continue Setup without copying this file.
3. Select No.
The system displays the active image screen.
4. Press the F6 key.
The system displays the Preset Main Menu.
5. Select DIMAQ Utility on the left pane of the screen.
The system displays the DIMAQ Utility screen.
6. Select Install Printer.
The system displays the Add Printer Wizard dialog box.
7. Select Next.

SONOLINE G20 US06-201.843.01.01.02 Page 10 of 12 Siemens


10.05 sd Medical Solutions USA, Inc.
USB Report Printers 11

8. Follow these instructions for each wizard screen presented by the Add Printer Wizard.

Wizard Screen User Action


Local or Network Printer 1. Select Local printer attached to this computer.
2. Clear (disable) the Automatically detect and in-
stall my Plug and Play printer check box.
3. Select Next.
Select a Printer Port 1. Select Use the following port:
2. Select the down arrow on the right.
3. Select an available USB port (for example, USB001
(Virtual printer port for USB)).
4. Press the SET key.
5. Select Next.
Install Printer Software 1. Select HP in the Manufacturer column.
2. Select HP DeskJet in the Printers column.
3. Select Next.
Use Existing Driver 1. Select Keep existing driver (recommended).
2. Select Next.
Name Your Printer 1. Select Yes.
2. Select Next.
Printer Sharing 1. Select Do not share this printer.
2. Select Next.
Print Test Page 1. Select Yes.
2. Select Next.
Completing the Add Printer 1. Select Finish.
Wizard 2. Select the OK button after the test page prints.

The system displays the DIMAQ Utility screen.


9. Select OK.
The system displays the active image screen.
10. Select a USB printer and then print a report to confirm the installation.

Confirming Configuration 0

To confirm configuration, select a USB printer and then print a report.

Siemens US06-201.843.01.01.02 Page 11 of 12 SONOLINE G20


Medical Solutions USA, Inc. 10.05 sd
12 USB Report Printers

Selecting the USB Printer


Use the system presets to select a USB printer as the system"s report printing device.
To select a USB printer:

1. Press the F6 key.


The system displays the Preset Main Menu.
2. Select Peripheral from the left pane of the screen.
3. Select the USB Printer check box.
4. Select the Select button to the right of the USB Printer check box.
The system displays a list of the installed USB printers.
5. Select the required printer and then select the OK button.
6. Select the Save button.
The system displays the active image screen.

Testing the USB Printer Configuration


Test the installation of the printer by printing a report.

NOTE It is possible to perform this test without actual data in a report.

To test the USB printer installation:

1. Press the F5 key.


The system displays the Exam and QuickSet screen.
2. Select OB.
3. Press the F2 key.
The system displays an obstetrical report.
4. Select the Send Report button on the lower right of the screen.
The system sends the report to the selected USB printer and changes the Send Report
button to Cancel.
- To interrupt immediately, select the Cancel button.
- To stop after the current page, press the Escape key.

SONOLINE G20 US06-201.843.01.01.02 Page 12 of 12 Siemens