Error Codes
Functional Checks
Preventive Maintenance
Theory
Troubleshooting
Learning Solutions
This program and its contents are RESTRICTED to GE Medical Systems. For GE Medical How to Use this EPST
Systems personnel only. No rights licensed - do not use unless you are a GE Medical
Systems employee. If you are not a GE Medical Systems employee, you must close this
application immediately. Disclosure to third parties prohibited. GES-produced documents Self-Paced Training Modules
are not intended for use as documentation for field-based diagnostic imaging equipment.
Copyright GE Medical Systems 2000.
Reference Materials
Feedback to Author
Notes
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Detector Calibration
Low Voltage Power Supply Checks
Detector power supplies must be verified before
calibration or table loading.
1. Remove the detector cover.
2. Check the following power supplies:
• ±15V at the A/D board inductor
C41 = +15.05 ±0.05
C42 = -15.05 ±0.05
• 5V at the Digital I/F board inductor
1000/2000: U16 pin 20 to analog ground = 5V
±0.05V
3000: L1 to digital ground = 5V ±0.05V
• +10V reference voltage at the motherboard
Across TP37 and 36 = Mfg. Label ±200 mV
3. Replace cover.
4. Repeat for remaining detectors.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
• Source holder d. Click Next Step. The low tube DAC value is
automatically set.
• Detector tube template
6. Place the source holder over tube 2.
• Detector Data Sheet
7. Select Auto from the Calibration Auto/Manual box.
Procedure
8. Click Next Step. The system automatically prompts
1. Select Acquisition, Calibration, Detector, Tube you to move the source holder to the next tube and
Balance from the Workbench menu. will repeat for all tubes.
2. Place the tube template over the selected detector. Important! If a tube fails auto calibration, record the tube
3. Select the appropriate detector from the Head box. number and click OK to continue. When auto
calibration is complete, go to the Manual
4. Select the appropriate isotope from the Isotope box
Tube Balance procedure and manually bal-
(Tc99m).
ance any failed tubes.
5. Click in the Low Tube box:
9. When all tubes are balanced, click Done.
If no tube number is displayed perform the following
10. Click Save to save the calibration data.
steps, otherwise go to step 6.
11. Repeat for remaining detectors.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Table Loading
Energy Table 13. Repeat for remaining detectors.
1. Flood the detector with a 300-400 uCi point source
centered five FOV above the detector.
2. Select Acquisition, Calibration, Energy. The
Energy Calibration screen is displayed.
3. Select Detector 1 in the Head box.
4. Select Energy Refine in the Calibration box.
5. Click Next Step.
6. Select the flood name (intrinsic is the default).
7. Click Next Step. The PHA screen is displayed.
8. Click Refresh to display the peak. The PHA screen
is displayed.
9. Peak the detector, if necessary, by changing the
centerline (do not use High Voltage).
10. Click Done.
11. Click Next Step. The calibration begins (this
calibration takes about 30 minutes) and the table is
automatically saved.
12. Click Done.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
6. Click Next Step. The PHA screen is displayed. • Adjust the table or phantom to align the sources.
7. Verify the detectior is properly peaked. 14. Click OK when the sources are aligned. The gantry
will rotate to 90°.
8. Click Done.
15. Align all three point sources to the vertical lines.
9. Click Next Step.
• Do not adjust the detector radius.
Important! Do not enable the gantry. At this point the on-
screen message: Please enable the • Adjust the table or phantom to align the sources.
gantry with RESUME is displayed. 16. Click OK when the sources are aligned. The gantry
Ignore it. will rotate and collect data at two different radii. Then
10. Click on Pscope. the tables will be stored to disk.
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
RES (Resume) Push to list motions: Gantry Rotate, Gantry PAUSE Push to move to the next display
Linear, Table Head, Table Foot, Radius 1, during potentiometer/encoder
Radius 2, Tilt 1, Tilt 2. verification calibration.
Press simultaneously with BRAIN to store
calibration values.
MARK Push to list calibration for each motion: Pot BRAIN Press simultaneously with RES to
Calibration, Potentiometer/Encoder store calibration values.
Verification, Speed Calibration, Motor Control
Test.
Push to select positive (+) motion during
speed calibration.
TORSO Push to select negative (-) motion during WHOLE BODY Push to change the speed register
speed calibration. (DAC) value during speed
calibration.
START/STOP Push to go from the main menu to calibration UPRIGHT 1. Push to go to the test screen
display. during speed calibration.
2. Push to exit the current
calibration display and return to
the main menu.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
2. Press OVERRIDE/RESET once on the gantry panel. Balanced Load Speed Adjustment
3. Back the motion off the limit switch using ENABLE Caution Adjusting potentiometer RV1 (SG1) will
and direction (-) on the Rotate Limit/Manual cause the gantry to rotate. Ensure the
Movement board. rotation path is clear before adjusting.
4. When the limit switch is clear, press OVERRIDE/
RESET on the gantry power panel. This procedure adjusts the motor controller so the gantry
rotates at the same speed in both directions under a
5. Repeat and adjust the limit switch as necessary. balanced load.
6. Rotate the gantry in the (-) direction until the
minimum hardware limit is reached. The
potentiometer count on the gantry display should
read -11º ±5º.
7. Press OVERRIDE/RESET once on the gantry panel.
8. Back the motion off the limit switch using ENABLE
and direction (+) on the Rotate Limit/Manual
Movement board.
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
3. Select Tilt 1 using the RES. 13. Press RES and BRAIN simultaneously to save the
pot values.
4. Select Pot Calibration using MARK.
14. Press PAUSE to continue.
5. Press START/STOP to begin calibration.
6. Ensure Thompson rod is level.
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
2. Press START/STOP to calibrate. 5. Repeat and adjust the limit switch if necessary.
3. Move detector 1 to 0°. 6. Press SLOW INTERLOCK and TILT (-) to tilt the
detector to the minimum hard limit. The
4. Verify potentiometer and encoder value is within potentiometer count should be -5° ±1°.
±0.5°.
7. Press OVERRIDE/RESET once on the gantry power
5. Move detector 1 to 90°. panel.
6. Move detector 1 back to 0°. 8. Press ENABLE and direction (+) on the Tilt Limit/
7. Move detector 1 back to 90°. Manual Movement board to back the detector off the
limit switch.
8. Verify potentiometer and encoder value is within
±0.5°. 9. Press OVERRIDE/RESET once on the gantry power
panel.
Hard Limit Check
10. Repeat and adjust the limit switch if necessary.
1. Press SLOW INTERLOCK and TILT (+) to tilt the
detector to the maximum hard limit. The
potentiometer count should be 95° ±1°.
2. Press OVERRIDE/RESET once on the gantry power
panel.
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
The calculated DAC value is displayed. If the value is • 30 cm/minute ±5 cm for slow speed.
outside the ideal range (D00-EFF), it cannot be If the speeds are outside of specifications recalibrate
saved and the motor controller must be adjusted. the motor controller.
5. Press RES and BRAIN simultaneously to save the 10. Press UPRIGHT to return to the main menu and
calculated DAC value. The DAC value may also be continue calibration.
recalculated by pressing START.
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
6. Select Motor Ctl Test using MARK. 2. Set the DAC speed value to 3FF using WHOLE
BODY.
7. Set the DAC speed value to DFF using WHOLE
BODY. 3. Move the table end stand up using FAST
INTERLOCK and TABLE (+). Note the speed value.
8. Press and hold FAST INTERLOCK and the motion This is the heavy load check.
direction button.
4. Move the table end stand down using FAST
9. Turn the potentiometer RV3 (Tach Gain) INTERLOCK and RADIUS 1 (-). Note the speed
counterclockwise. Adjust RV3 to approximately 60 value. This is the light load check.
cm ±6 cm/min.
5. Slowly adjust potentiometer RV2 (Sig Gain 1) until
Caution Do not exceed hardware limits. System dam- the speed under the heavy load is nearly equal to the
age will result.
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
1 Enable Step
2 Enable Six Point Orbit Setup
Motions
3 Not Used
• Gantry Rotate
4 Disable Safety Utilities
• Detector Radius
5 Enable Calibration Mode • Table Height
6 Enable PScope • Table Translate
7 Not Used 4. Press Operator Control Panel ROTATE 0 to toggle
8 Not Used through DAC values.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Alignment and Calibration
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
COMPONENT LOCATIONS
Data Acquisition and Processing Table Translate
Detector Motherboard Assembly Table Height
Detector Tube Maps Collimator Server/Storage 1000/2000
High Voltage Power Supply
Collimator Server/Storage 3000
CAP
PScope Collimator Installation
Odyssey Computer
Gantry Power Distribution 1000/2000
Gantry Control 1000/2000
Gantry Motion 1000/2000
Gantry Rotate
Gantry Linear
Detector Tilt
Detector Radius
Table Height
Gantry Power Distribution 3000
Fuse and Breaker Locations
Gantry Control 3000
Gantry Motion 3000
Detector Radius
Gantry Rotate
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Main Amps
F1
(125V 4A) TP3 (Sum X)
F2 TP1 (Sum Ze)
(125V 4A)
J1
JP6 (Top) TP4 (Sum Y)
JP5 (Bottom) A/D Board
To CAP
TP2 (Sum Zp)
TP5 (Convert)
TP38 (AGND)
Digital Interface TP15 (Ze)
Board J13 (Y)
J14 (Zp)
Echelon Network J12 (X)
Interface Board 3000 XP (Cover Removed)
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
1 2 3 4 5
6 7 8 9 10 11
12 13 14 15 16
17 18 19 20 21 22
23 24 25 26 27
28 29 30 31 32 33
34 35 36 37 38
39 40 41 42 43 44
45 46 47 48 49
XP 1000/2000 XP 3000
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Detector
Motherboard
High Voltage
Distribution
Board Board
High Voltage
Power Supply
High Voltage
Distribution
Board Board
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
CAP
1000/2000 To
ECG J18
Rear View Echelon
To
Gantry
PB +12V
MPU J17 J16
RED
LED -12V
J15 TAXI
BNC
DET 1
TAXI to
J8 PScope
R
I/O
G
Y
J7
TAXI
J10 BNC
G
GND ISO
Troubleshooting)
+12V ISO
J13
Y
To Gantry MPU -12V ISO
Serial I/O (RS232) J6
TAXI GND
BNC
LED Note
VCC
J11 DET 3 Red: Violation Condition
To Yellow: Pulse when data
R
Ethernet loaded into FIFO
J14 G
J12 Green: Pulse when data
Y
extracted from FIFO
Multi-Head Correction Correction Correction Acquisition PScope
Front Panel Digital Divider Processor Processor Processor Interface Interface
Assembly Board Board (Det 1) Board (Det 2) Board (Det 3) Board Board
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
PScope
PScope Display
Interface Board
PScope Display
PScope Display
Control Board
CPU Board
Rear View
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Odyssey Computer
Power
Video Card Supply
RAM
Odyssey Computer
Typical Configuration
SCSI Card
Motherboard
Floppy Drive
Odyssey Computer
Cover Removed
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Fan
EStop Switch Board
Gantry
Motor Power Transformer (Front View)
-5V, 12A PS
DC Distribution Board
+24V, +5V PS
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Gantry
(Front View)
Gantry MPU
Board
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Gear Box
Pot & Encoder
Assembly
Gantry
(Rear View)
Rotate Belt
Potentiometer
Limit/Manual
Movement Board
Limit Switch Assy
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Encoder
Gantry
(Front View)
Potentiometer
Limit/manual
Movement Board
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Gantry
Tilt Motor (Front View)
Limit Switch
Encoder Limit/Manual
Movement Board
Potentiometer
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Limit Switches
Motor Control Board
Limit/Manual
Movement Board
Potentiometer
Encoder
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Limit Switch
Note: The layout and part
numbers for the foot
Encoder Assy pedistal are the same as
the head pedistal.
Potentiometer Assy
Ball Screw
Limit/Manual Movement
Board
Motor Control
Board
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Relays
Callout Type
1 Main
Gantry Relays 2 Rotate
3 DUZ
Qty Type
4 Isolation
2 DPDT 3A
6 3PDT 24V
Gantry Fans
1 3DPT
1 Time Delay (K9) Qty Type
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
F1 7A MDL Fan
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
F1
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Gantry
(Front View)
Gantry
(Rear View)
Encoder
Potentiometer
Radius Motor
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Gantry
(Rear View)
Encoder
Gantry
Rotate Motor Gear Box
Gantry Rotate Motor Gantry Rotate Motor Controller
Controller Adjustment (Detail) Potentiometer
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Gantry
(Front View)
Gantry Rotate Max
Limit Switches
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Transverse Bearing
Block w/Bearings Spider Cable from SOS board
Potentiometer to Pot and Encoder
Encoder 5k/10 turn (not Shown)
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Coupling
Coupling
Sprocket #35 Chain
Connector Link
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Potentiometer
Eccentric
Cams (qty 16)
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Transverse
Carriage
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Component Locations
Collimator Installation
1000/2000 3000
1 1
2 2
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Error Codes
ERROR CODES
Gantry Calibration Messages & Error Codes
Gantry Normal Messages & Error Codes
Safety Error Codes
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Error Codes
< PREPROGRAM> Preprogrammed function activated. This Information only. No action required
message appears on line 4 on the right side of
the screen.
COLLIMATOR LOAD, This message is displayed for 5 seconds after The server must be removed from the detector
VERIFY SERVER IS COLLIMATOR LOAD 1 or COLLIMATOR LOAD 2 before pressing INTERLOCK to initiate the
REMOVED is pushed. motion.
COLLIMATOR The collimators mounted are not of a matched Replace with matched collimators.
MISMATCH set.
COLLISION <1> or Collimator collision occurred. Clear by pressing OVERRIDE/RESET once and
<2> or <3> moving the affected heads away from the
collision.
Restore normal operation by pressing the
OVERRIDE/RESET again.
E_STOP Emergency stop. An emergency stop button has Clear by pressing the emergency stop button
been pressed. All gantry motions are disabled. again.
EEPROM WRITE Calibrated value not written to the EEPROM. The Possible MPU Fault.
ERROR EEPROM will automatically continue to try to
read the value and clear the message. If the
message remains, an EEPROM error condition
exists. Press the PAUSE to quit.
ERR Unrequested motion error. This message refers Refer to Gantry Troubleshooting (1000/2000, or
to the affected motion. All gantry motions are 3000)
disabled. (See UNREQUESTED MOTION).
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Error Codes
HARDWARE LIMIT Hardware motion limit reached. All system Press manual overide and back off the limit.
motions are halted.
HEAD 1 INACTIVE Head 1 is inactive. SW7 of switchpack 1 is ON. Information only.
HEADS INACTIVE Both heads inactive. Both SW6 and SW7 of Information only.
switchpack 1 are ON. Collision sensors will not
function.
LIMIT Software limit. This message is displayed under Enable manual override and carefully back off
the affected motion on the display screen. Only software limit.
that motion is disabled.
LOST HAND The hand controller is unplugged or is no longer Reconnect/fix hand controller.
CONTROL MSG communicating with the MPU. The system
defaults to the Motion Halt mode.
MOTION MOVED, Motion with no encoder change. The message is Refer to Gantry Troubleshooting (1000/2000, or
BUT ENCODER HAD displayed for two seconds on line 5 of the gantry 3000)
NO CHANGE display, and the calibration software returns to the
main menu.
NO COLLIMATOR No collimator mounted. All gantry motions are Install a collimatoror the collimator test blocks to
disabled. clear.
NO MOTION A valid motion command was issued but no Refer to Gantry Troubleshooting (1000/2000, or
motion occurred. 3000)
NO TABLE The table is not mounted. Mount table.
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Error Codes
OVERRIDE Operator motion override. Signaled by the Clear by moving away from the obstruction and
override light on the gantry control panel and pressing OVERRIDE/RESET a second time.
continuous beeping.
P<>E M:X P:XX.XX Potentiometer/encoder mismatch. The motion Refer to Gantry Troubleshooting (1000/2000, or
E:XX.XX number pot value and encoder values are 3000)
displayed.
PAUSE (Preprogram mode only) System motion is Press RESUME to clear.
manually paused.
PCLOAD/TST SW1, SW2, or SW3 of switchpack 1 is ON. Information only.
POT INSTALLED This message appears in Potentiometer Check pot for correct installation.
INCORRECTLY Calibration if the potentiometer is installed
incorrectly or if the mounting hardware has
become loose. The message is displayed for 5
seconds on line 3 of the gantry display.
PRESS "RESET" The table pedestals were not balanced when the PressOVERRIDE/RESET button twice to clear.
TWICE TO BALANCE system was last turned off.
THE TABLE
SERVICE The gantry is in Service mode. SW5 on Information only
switchpack #1 ON. All software safety devices
are disabled.
TABLE FOOT AT (At power up.) The table foot pedestal is at
HARDWARE LIMIT, hardware limit. See HARDWARE LIMIT
SYSTEM HALT message for more information.
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Error Codes
TABLE FOOT AT (At power up.) The table foot pedestal at software Information only.
SOFTWARE LIMIT, limit. The foot pedestal will level itself to the head
SYSTEM HALT pedestal automatically and the message will
disappear.
TABLE FOOT DOES The table foot pedestal not balanced to the table Press OVERRIDE/RESET twice.
NOT MOVE, SYSTEM head pedestal.
HALT
TABLE MISALIGNED The table is mounted incorrectly. Manually shift head table mount position (the
table mount will slide and the table does not have
to be mounted).
TABLE NOT LEVEL Table head and foot pedestals not vertically Align pedestals.
aligned.
THE CAL MOTION This message appears in Potentiometer See to Gantry Troubleshooting (1000/2000, or
HAS NOT CHANGED Calibration if the RES and BRAIN buttons are 3000)
POSITION pressed to save the pot values and the motion
did not occur. Scrolling through the calibration
menu and attempting to save a pot value also
displays this message.
UNREQUESTED Unrequested gantry motion. ERR is also See to Gantry Troubleshooting (1000/2000, or
MOTION displayed under the affected motion on the gantry 3000)
display The system defaults to the Motion Halt.
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Error Codes
Message Definition
EN : Converted encoder value in degrees or centimeters from the absolute encoder count value.
P: Converted potentiometer value in degrees or centimeters from the absolute potentiometer value.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Error Codes
E. STOP Emergency stop. An Emergency Stop button has Pressing the Estop again and then press
been pressed. All gantry motions are disabled. OVERRIDE\RESET twice.
OVERRIDE OVERRIDE/RESET has been pressed once. The Pressing OVERRIDE/RESET a second time.
override light on the gantry control panel will
illuminate. A motion is started override will cause
a continuous beeping sound.
COLLISION 1/ Collimator collision. Pressing OVERRIDE/RESET once and move the
COLLISION 2 affected head away from the collision. Then press
OVERRIDE/RESET again.
PAUSED A preprogrammed motion has been manually Press RESUME.
paused.
NO COLLIMATOR No collimator is mounted. The gantry will not Mount a collimator or test block then press
HEAD (1/2) move. OVERRIDE/RESET twice.
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Error Codes
GANTRY LINEAR A pinhole collimator is installed and the Move gantry linear to less than 5 cm.
POSITION MUST BE Collimator Load preprogrammed function is
LESS THAN 5 cm TO enabled. The Collimator Load preprogram
UNLOAD PINHOLE function may only be initiated when gantry linear
COLLIMATOR is less than 5 cm.
TABLE: Patient pallet is mounted incorrectly. This is Ensure the table head end pedestal is in the
MISMATCHED usually a result of the table head pedestal in the correct position.
opposite position of the table foot pedestal.
TABLE: CARDIAC Patient pallet is mounted in the offset cardiac Information only.
imaging position.
TABLE: NOT Head pedestal is not pushed completely to the Clear by pushing the head pedestal completely to
LOCATED center or off-center position. No CAP or gantry the correct position.
preprogrammed motion.
PRESS "RESET" The table pedestals are not balanced. Press OVERRIDE/RESET twice. The table
TWICE TO BALANCE pedestals will balance automatically.
THE TABLE
HEAD (1/2): NO The gantry software does not detect a collimator. Clear by pressing OVERRIDE twice.
COLLIMATOR, BOTH
HEADS: NO
COLLIMATORS
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Error Codes
E.Stop Emergency stop pressed Press emergency stop again and Override/Reset twice to clear
Paused Preprogrammed motion paused by Press Resume on hand control once to clear
hand control
No No collimator mounted Mount collimator
Collimator
Collimator More than one type of collimator Check collimators. Mount matched set.
Mismatch mounted
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Functional Checks and Procedures
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Functional Checks and Procedures
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Functional Checks and Procedures
Energy Resolution
1. Flood the detector using an 800 µci point source 5. Repeat for remaining detectors.
centered in the field of view and at least:
• 127cm/50in (for the 3000)
• 254cm/100in (for the 1000/2000)
from the detector.
2. Select Acquisition, NEMA Testing, Energy
Resolution, Head # from the workbench menu on the
Odyssey computer. Follow on-screen instructions.
3. After the required number of counts, the following
information is displayed:
Peak Channel
Peak Centroid
FWHM in Channels
FWHM in %
4. The Full-Width Half-Mask (FWHM) in % should be:
≤10.2% (3000)
≤10.8% (1000/2000)
Note This percentage is greater than the NEMA speci-
fication for Energy Resolution. This is due to the
lack of a flood mask (when open crystal) and test
method differences.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Functional Checks and Procedures
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Functional Checks and Procedures
Flood Uniformity
This test verifies that camera field uniformity is sufficient 8. 2000 only: Move head 1 under the source using
to prevent the introduction of artifacts during normal GANTRY LINEAR.
operation.
9. Select Aquisition, Do Patient, Patient
Perform this test with the collimator attached to simulate Info. from the workbench menu.
normal clinical imaging.
10. Fill out patient information form.
Important Adjust the table to maximum height when 11. Click Done.
storing flood tables using medium or high
12. Select Aquisition, Do Patient, Protocol
energy collimators
Select from the workbench menu.
1. Have qualified personnel prepare a sheet source. 13. Select Flood Protocol.
Note Mix the sheet source thoroughly. Improper mixing 14. Click Done.
will cause inconsistencies in flood uniformity.
15. Select Aquisition, Do Patient,
2. Install collimators to be tested. Ensure the correct Acquisition Info. from the workbench menu.
flood tables are loaded for the collimators used.
16. Select the detector to be checked in the Heads field.
3. Mount sheet source support to foot-end pedestal.
17. Select the following inputs:
4. 1000/2000: Adjust table height to minimum.
Termination Type: COUNTS
5. 3000: Adjust table height to 10.6 cm.
Value: 10000
6. Place sheet source on support. Ensure it is level.
Image Size: 64 X 64
7. Move head(s) to maximum radius.
18. Check PHA settings.
19. Click Done.
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Functional Checks and Procedures
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Functional Checks and Procedures
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Functional Checks and Procedures
LEUHR 30 mci Note 2000 only: Ensure detectors are at the same
radius.
LEHR-Fan 15 mci
10. 3000 only: Select Heads Same to set other
LEUHR-Fan 20 mci
detectors to the same radius as detector 1.
ME 20 mci
11. Manually rotate the gantry to ensure all detectors
HE 20 mci clear the phantom.
3. Install low energy collimators and ensure 12. Select Aquisition, Do Patient, Patient
corresponding flood tables are also loaded. Info. from the workbench menu.
4. Set detector head radius to maximum for all 13. Fill out patient information form.
detectors. 14. Click Done.
5. Rotate system to 0º.
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Functional Checks and Procedures
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Functional Checks and Procedures
20. Reconstruct three slices using the following Slice Thickness Att.
parameters. For each section, adjust the first slice Collimator Filter
(plx) Cor.
(and last slice) so the slice contains the section to be
Rods
viewed (rods, spheres, uniform). Increment or
deincrement the first slice setting to get the best LEDY 28 (1k/2k) 35 (3k) BUT/5.25 .11
image quality LEGAP 28 (1k/2k) 35 (3k) BUT/5.25 .11
LEHR 28 (1k/2k) 35 (3k) BUT/5.25 .11
21. Compare images with standard images provided by
manufacturer. LEUHR 28 (1k/2k) 35 (3k) BUT/5.25 .11
LEHR-Fan 18 (1k/2k) 25 (3k) BUT/5.25 .11
.
LEUHR-Fan 18 (1k/2k) 25 (3k) BUT/5.25 .11
Spheres
LEDY 4 (1k/2k) 5 (3k) BUT/5.25 .11
LEGAP 4 (1k/2k) 5 (3k) BUT/5.25 .11
LEHR 4 (1k/2k) 5 (3k) BUT/5.25 .11
LEUHR 4 (1k/2k) 5 (3k) BUT/5.25 .11
LEHR-Fan 2 (1k/2k) 3 (3k) BUT/5.25 .11
LEUHR-Fan 2 (1k/2k) 3 (3k) BUT/5.25 .11
Uniform
LEDY 8 (1k/2k) 10 (3k) BUT/5.25 .11
LEGAP 8 (1k/2k) 10 (3k) BUT/5.25 .11
LEHR 8 (1k/2k) 10 (3k) BUT/5.25 .11
LEUHR 8 (1k/2k) 10 (3k) BUT/5.25 .11
LEHR-Fan 5 (1k/2k) 6 (3k) BUT/5.25 .11
LEUHR-Fan 5 (1k/2k) 6 (3k) BUT/5.25 .11
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Functional Checks and Procedures
7. Display the whole body image at the Odyssey Whole Body Uniformity
computer. The bar resolution images should be 1. Secure the flood source against the collimator.
uniform and free from artifacts. See the chart below:
2. Set up the following Whole Body Acquisition
Collimator Bars Bar Resolution Parameters at the CAP:
LEGAP-PAR 1/4" Good Scan Mode: Planar
Acq. Type Whole Body
LEHR-PAR 1/4" Excellent
Image Size: 256 x 1024
3/16" Fair Magnification: 1.0
LEUHR-PAR 1/4" Excellent Scan Time: 40 min
3/16" Good Scan Length: 214 cm or maximum
5/32" Fair
Scan Direction: L to R
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Logic and Block Diagrams
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Logic and Block Diagrams
Prism XP 1000/2000
Prism XP 3000
Back
Overall Signal
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Logic and Block Diagrams
Detector Signal
Prism XP 1000/2000
Prism XP 3000
Event X, Y Data
Flat Panel Serial Com. Motorola SCSI
640 X 480 525 MB
68040
3.5" Disk
CPU
Floppy
Serial Data
PScope PScope Gantry
Flat 386
Display Display MPU
Panel CPU
I/F Control 128 MB
Driver
Disk VME System Bus
PScope Chassis
Position Data
PScope
Head 1
Head 2 Multihead Correction Correction Correction CAP
Head 3 Acq.
Digital Processor Processor Processor PScope
I/F
Local Network Driver Det. 1 Det. 2 Det. 3 Interface
Back
Detector Signal
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Logic and Block Diagrams
Prism XP 1000/2000
Prism XP 3000
Gantry
HCR MPU
Detector
X,Y, Ze, Xp Data Serial I/O
Note:
1000/2000: RS232 Parallel
XP CAP 3000: RS422 Port
(Position Info.)
Fiber Optic
Processor
Data Transfer
Event X,Y Data
Odyssey Ethernet
PScope
Computer Controls CAP
Operations
Gantry
Back
Detector/CAP
Interconnect
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Logic and Block Diagrams
Prism XP 1000/2000
Prism XP 3000
Image
ACQ I/O
SCSI Processing
I/F Interconnect
Workstation
Tape
Printer
Drive
Hard Laser
Drive Imager
LAN
Ethernet
Keyboard
CAP
Image Processing
Optical
Computer
Drive
Back
Image
Processing
Alignment &
Calibration
Component
Locations
Error
Codes
Functional Logic & Block
Checks Diagrams
Preventive
Maintenance
Software &
Diagnostics
Theory Trouble-
shooting
Reference
Materials
Notes
Computer Signal
GE Medical Systems Picker Prism XP Logic and Block Diagrams
Prism XP 1000/2000
Collimator ID Hand Controller PScope Motion Port Numbers
DC
& Collision J111: Gantry Rotate
Distribution EStop
(Detectors J112: Gantry Linear
PCB
1 & 2) Interface PCB Processor J113: Table/Head
J114: Table/Foot
J115: Head 1 Radius
J116: Head 1 Tilt
J117: Head 2 Radius
Turbo MPU J118: Head 2 Tilt
Motion
Ports J113 J114 J115 J117 J116 J118 J111 J112 MPU Motion Port
Pin Out
1: +5 VDC
Limit Limit Limit Limit Limit Limit Limit Limit 2: DIG. GND
Manual Manual Manual Manual Manual Manual Manual Manual 3: Encoder A+
Motion Motion Motion Motion Motion Motion Motion Motion 4: Encoder A-
5: Encoder B+
6: Encoder B-
Motor Motor Motor Motor Motor Motor Motor Motor 7: Pot. Ref. Voltage
Controller Controller Controller Controller Controller Controller Controller Controller 8: Pot. Wiper
PCB PCB PCB PCB PCB PCB PCB PCB 9: Analog GND
10: Motor Velocity
11: Analog GND
12: Motor Enable
Motor Motor Motor Motor Motor Motor Motor Motor 13: Limit Switch
14: Table Sensor 1
Table Table Radius Radius Tilt Tilt Gantry Gantry 15: Table Sensor 2
Head Foot Detector 1 Detector 2 Detector 1 Detector 2 Rotate Linear 16 Not Used
Back
Gantry Motion
Signal
Alignment &
Calibration
Component
Locations
Error
Codes
Functional Logic & Block
Checks Diagrams
Preventive
Maintenance
Software &
Diagnostics
Theory Trouble-
shooting
Reference
Materials
Notes
(1000/2000)
GE Medical Systems Picker Prism XP Logic and Block Diagrams
Prism XP 3000
Trans Height
Flat Panel Table Table
Display Translate Height
Motor Motor
(640 X 480) Linear Dual-Speed
& Display Controller Controller
Operator Driver Pot & Pot
Control Encoder Feedback
Panels Feedback
Left & Right PScope
Processor Table SOS
Table
Gantry MPU
Pot &
Encoder
Feedback J3 Detector
Pot &
Gantry Head Encoder Gantry
Rotate Radius Feedback SOS II Hand ACQ
Motor
Motor Board I/F
Motor Motor (3) Optional Control
Control Controller Connection
Gantry Detector
Rotate Radius Serial Connection
CAP
Gantry
Back
Gantry Motion
Signal (3000)
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Logic and Block Diagrams
Prism X P 1000/2000
Line In
F1
(3A Slo) DC PS
208/220/ +24/+5
240VAC F1, F2, CB1 Step Down 120 VAC VDC
Line Filter
(F1 & F2 30A Slow Blow) Transformer
F2 Outlet Box
( 7ASB) & Gantry
120 VAC Cooling Fans +24/+5
from G antry To Motor VDC
CB5
CB2 Motor Controller
(12A) F3
Controller Detector 1
CB1-5 on (3ASB) PS
(5A) AC To ±15
Back of CAP VDC
CAP
Line Filter F4 Detector 2
CB3 Relay (3ASB) PS ±15
CAP (12A) K2 VDC
F5 Det.1, 2
(2ASB) PS -5
Digital BP CB1 VDC
CB1 Relay
& Floppies Gantry
(12.5A) K3
PS ±12/+5 (12A) F6 MPU
VDC (4ASB) PS ±15/+5
Breaker VDC
F1- 6 on Gantry
Panel
Analog AC Distribution PCB
CB2
Backplane
(1A) +5 VDC
Gantry
PS
Analog &
CB3
Digital BP
(5A)
PS ±15 VDC
CAP
Back
Overall Power
Distribution
Alignment &
Calibration
Component
Locations
Error
Codes
Functional Logic & Block
Checks Diagrams
Preventive
Maintenance
Software &
Diagnostics
Theory Trouble-
shooting
Reference
Materials
Notes (1000/2000)
GE Medical Systems Picker Prism XP Logic and Block Diagrams
CB1 - 5 on Operator
Motor Back of CAP Terminal
Prism X P 3000
120VAC
CB1 ±12/+5 VDC Disk
T2 Rotate Gantry (12.5A) PS Drives
CB1 (25A)
Isolation Motor
Rotate Gantry
CB1 - 8 on Transformer Controller
Back of Gantry CB2 ±15VDC
(1A) PS Digital/
CB8 CB5
(12A) (20A) Analog
CB7 Backplane
Table TRANS
CB2 (3A) CB3 ±15VDC
Motor
Table (5A) PS
Controller
CAP
T1 Table Height
Input Motor
Controller
120VAC
Transformer
Gantry Fuses
CB5 (3A) -5VDC Main 220VAC
Gantry MPU PS F1 & F2 (25A)
AC Outlet
±15VDC Detector 1 Detector 2 Detector 3 F3 & F4 (5A)
PS
Laser & Fan
F1 (7A)
Back
Overall Power
Distribution
Alignment &
Calibration
Component
Locations
Error
Codes
Functional Logic & Block
Checks Diagrams
Preventive
Maintenance
Software &
Diagnostics
Theory Trouble-
shooting
Reference
Materials
Notes
(3000)
GE Medical Systems Picker Prism XP Logic and Block Diagrams
EStop Signal
Prism XP 1000/2000
TBL HP Motor
Control
GLIN Motor
24V Left Right Gantry MCDC Control
24V 24V 24V 24V
Power J208 AC J210 EStop EStop J110 Turbo J302 Power
Supply Dist. Board Board MPU RAD 1 Motor
Supply
Control
J208 J210 J302
1: 24V 2: 24V 1: 24V TILT 1 Motor
2: COM 3: COM 2: COM Control
RAD 2 Motor
Control
TILT 2 Motor
Control
Back
Estop
(1000/2000)
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Logic and Block Diagrams
K6 Manual Open
Overide
Back
Estop (3000)
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Logic and Block Diagrams
Prism S 1000/2000
Prism S 3000
Hit Collimator
Detect ID Sense
5VDC
Collimator
P 3 2 1 0
Ground Shield
Collimator
Upper Grid Layer Latch Switch
Back
Collimator ID
(1000/2000)
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Preventive Maintenance
PREVENTIVE MAINTENANCE
Prism XP 1000/2000
Required Materials
Required Sources
Schedule
Prism XP 3000 Schedule
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Preventive Maintenance
Prism XP 1000/2000
Required Materials Required Sources
N100722 Lead Pig Balance, Energy Acquisition, 200 µci point source
HV/KeV, Energy Resolution,
393598 DVM Spatial Resolution
NRTEK2230 Oscilloscope Pixel and Composite Five 200 µci vials
N Tube Template (1000/2000) Registration
N Calibration Kit (1000/2000) Flood Acquisition, Flood QC 15-30 mci flood tank
Deluxe Data Spectrum SPECT performance test 15-30 mci source for Jaszczak
Phantom phantom
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Preventive Maintenance
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Preventive Maintenance
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Preventive Maintenance
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Preventive Maintenance
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Preventive Maintenance
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Preventive Maintenance
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Preventive Maintenance
Patient Table Floor Lock Blocks Quarterly Table component Check for proper engagement into floor tracks and
locations check ramp is actuating locking pin.
Patient Table Floor Rail Quarterly Table component Check floor track groove. Clean with Fantastik or
locations equivalent
Patient Table Transverse Rails Quarterly Table component Clean transverse rails with Fantastic or equivalent.
locations
Gantry Drive Limit Switch Semi-Annual Gantry component Manually activate and check switch alignment.
Assembly locations
Radial Drive Roller Bearing Semi-Annual Gantry component Lubricate with Lubriplate #OI. #305005. Check
locations chain tension and adjust if necessary.
Caution
Do not overtighten chain.
Radius Limit Switches Semi-Annual Gantry component Manually activate switch and check alignment.
locations
Radius Drive IKO Bearing Rails Semi-Annual Gantry component Lubricate with Lithium Grease #2 #305005.
locations
Patient Table Drive Chain Semi-Annual Table component Check chain tension and adjust if necessary.
Tension locations
Energy Calibration Semi-Annual Energy Corrections
Flood Calibration Semi-Annual Flood corrections
IRC Calibration Semi-Annual IRC corrections
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Preventive Maintenance
Gantry Drive Gear Box Annual Gantry component Visually inspect for leaks around shafts.
locations
Gantry Drive Motor Annual Gantry component Check the two electrical motor connections.
locations
Patient Table Annual Table component Check table movement along floor tracks.
locations Movement should be smooth and steady with no
rocking. Adjust cam followers if necessary.
PMT Balance (All three Annual Detector Calibration
detectors)
Gantry Drive Gears Annual Gantry component Lubricate with Mollycoat spray OI #305001
locations
Cable Trough Annual Inspect and check for any rub marks on cable or
trough. Ensure area is clear and nothing has fallen
near the gears from previous PM. Ensure cable is
clamped and inline with trough.
Radius Drive Gear Box Annual Gantry component Inspect for leaks around shaft.
locations
Radius Drive Motor and Annual Gantry component Inspect for correct alignment.
Coupling locations
Radius Drive Acme Screw Annual Gantry component Manually activate switch and check alignment.
locations
Patient Table Drive Chain Annual Table component Lubricate lightly with LPS #3. OI #305002
Tension locations
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Preventive Maintenance
Patient Table Pallet Movement Annual Table component Inspect patient pallet for nicks or cracks in
locations detectorrest area and around bolt holes. Check for
any side-to-side pallet movement. Adjust side
camrollers if necessary.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Software and Diagnostics
Diagnostics
Introduction
On-Site Laptop Access
Running Diagnostics
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Software and Diagnostics
Diagnostics
Introduction 3. From the main Windows screen, select
Accessories, Terminal. Settings,
Diagnostics files reside on the Odyssey computer hard
Communications, COM1 (under Connector).
drive and in memory at the XP CAP. Diagnostics may be
Select 9600 baud (under Baud Rate).
run from the Odyssey computer or laptop (use the laptop
if the Odyssey computer is down or in the event of 4. With communication established with the Odyssey
ethernet problems). computer, login as prism.
Diagnostic requests are sent to the XP CAP diagnostic 5. Enter the following command to establish
files on the hard drive and results are sent back to the communication:
Odyssey computer for display.
/prism telnet cap2
Odyssey Computer Access The > prompt indicates the command was
1. Open the UNIX shell. successful.
2. Enter: 6. Type the command to load the diagnostics. Start the
\prism rlogin cap2 diagnostics as if using the Odyssey computer.
> ld</SVC/diagcap2
> diagcap2
On-Site Laptop Access
1. Connect the serial interface cable to J13 (Corn. port
1) on the 68040 XP CAP CPU and the serial port on
the laptop.
2. Boot the laptop and start Windows.
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Software and Diagnostics
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Software and Diagnostics
DSP Stage 1: OK, DSP Stage 2: OK, DSP 10.Energy Corrections/Cal Test
Stage 3: OK, DSP Stage 4: OK 11./Y Hermite Linearity Test
Head 2 Correction Processor Board is 12.Flood Table Correction Test
available
13.Utility/Option Menu
DSP Stage 1: OK, DSP Stage 2: OK, DSP
Stage 3: OK, DSP Stage 4: OK 14.Select Correction Processor Board
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Software and Diagnostics
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Software and Diagnostics
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
THEORY
Safety Power Distribution (3000)
Overall Safety Precautions Overall
Radiation Safety Relays
Electrical Safety Control
Mechanical Safety
Motion (1000/2000)
Fire Safety
Motion (3000)
Data Processing and Acquisition
Potentiometers and Encoders
Detector Optics
Motor Controllers
Detector Processor
Echelon Board Patient Table
Signal Processing 1000/2000
Persistence Scope Gantry Safety (3000)
Gantry Interface Overview
Hand Control System Initialization
Operator Control Panel Estop
Limit Switch Halt
Power Distribution (1000/2000)
Collimator Collision Halt
Overall
Gantry Power Distribution Isotope Information
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Safety
Always follow all safety procedures when operating or Overall Safety Precautions
working on all Picker Prism S Series equipment. This Always comply with the following general safety
equipment contains high voltages and heavy moving precautions:
parts that are hazardous if proper safety precautions are
not followed. For maximum safety, always operate and • Procedures on this system should be performed only
service this system according to procedures specified in by properly authorized and qualified personnel.
this EPST and the manufacturers Operators Guide.
• Do not operate the system unless you fully
Before operating or servicing this equipment you must understand the safety considerations and procedures
have a thorough knowledge of all operating controls, presented in this EPST and manufacturers operator
electro-mechanical functions, and a basic knowledge of guides.
nuclear physics.
• Do not defeat any interlock or safety device.
WARNING! Some procedures in this EPST require the use • Make sure remote circuit breakers are off and tagged
of radioactive material. Always use proper before working inside the equipment.
safety procedures when handling radioactive
material. You cannot feel, see, or hear ionizing • Always know actions to be taken in an emergency
radiation. Long-term exposure to low radiation situation.
levels may cause health problems.
In an emergency, such as an equipment malfunction or
Equipment safety procedures and potential hazards are fire, press the Power switch on the gantry side panels to
emphasized by warnings and cautions in this EPST and stop all operation.
the Operators Guide. It is absolutely essential that you
Note The PMT high voltage and the primary AC power
are familiar with both documents before attempting to
will remain on.
service the system.
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Safety Continued
Radiation Safety Mechanical Safety
All service personnel must wear film badges or thermo- The mechanical drives of the Prism systems can cause
luminescent dosimeters (TLDs) to monitor radiation serious injury.
exposure. This film badge or TLD must be worn at all
times when working near a radiation hazard area. DANGER! Always press an emergency stop button on the
Exchange the replaceable insert or TLD every month. gantry before working on the rear of the gantry.
This is done to prevent accidental gantry
The purpose of the badge/TLD is to alert the Radiation movement initiated from the CAP console or
Safety Officer to excessive radiation exposure. This hand controller.
monitoring device is furnished in compliance with the
Nuclear Regulatory Commission. Fire Safety
Electrical Safety Observe all fire safety precautions. Flammable liquids
Always use caution when working with the covers may be stored or used in the same room as the Prism S.
removed. 208 - 240 VAC line voltages are used as Familiarize yourself with the location of the fire
primary power. Primary wiring requires grounding extinguishers in the room and observe all No Smoking
conductors for all power feeds and neutrals with posted areas. Use care when using equipment that could
grounded conduits. generate a spark or excessive heat, such as a soldering
iron.
Rooms may also be equipped with oxygen. Observe all
fire and safety regulations at each designated location.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Gantry Interface
Note Some interface configurations will vary between The Motion Halted LED lights whenever the Gantry
machines. Functionally, however, they all operate MPU inhibits gantry motion. The activating signal comes
in the same way. from K7 on the SDI board.
Hand Control The Manual Override LED flashes when the Manual
Override switch is activated, from relay K1 at J1-20. The
The hand control consists of:
Blink signal from the SDI board causes it to flash when
• An Estop button activated.
• Control buttons to manually control gantry motions
• 3000 only The HEADS SAME button to position The Alarm sounds:
detectors 2 and 3 to the same radius as detector 1.
• At collimator collision
Operator Control Panel • When pre-programmed buttons are pressed
The manual override switch allows the operator to
override collimator collision or the Gantry MPU halts • At software motion limits
motion. U2 latches in the override condition. • When manual override is activated and INTERLOCK
Pre-programmed switches are used to control on the hand control is pressed.
automatic positioning of the gantry and detectors for The alarm is activated by the Gantry MPU, through the
certain types of studies or patient access. The pre- SDI board.
programmed switch signals route directly from the
operator control panels through the SDI board to the
Gantry MPU board.
The MPU periodically checks for active switches.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Six low-voltage and one high-voltage DC power supplies • Fuse F1 to a pilot, dual-voltage power supply through
J206
are located on the gantry:
• Fuse F2 to a Waber box through J207
• Four provide power to operate the detectors and
detector processing electronics • Gantry Electronics circuit breaker C82, gantry power
• One provides power to operate the Turbo MPU board, relay K3, individual line fuses F3-F6, and individual
Preprogram board, and gantry display connectors J208-J211, to the remaining low-voltage
power supplies
• One is a pilot power supply used throughout the
• CAP circuit breaker C84, CAP power relay K2, and
gantry
connector J212 in the CAP.
• One 100 VDC power supply for the eight gantry
motors. The CAP supplies an AC power signal through J205 to
energize relay K2 (normally open) when the CAP power
Gantry Power Distribution switch is on.
The gantry has an AC Power Distribution board, a DC The front control panel supplies a switched 24V signal
Power Distribution board, and an MPU Power through J200 to energize relay K3 (normally open) when
Distribution board. the gantry power switch is on. This signal originates on
the DC Power Distribution board and is routed to the
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Control
The gantry MPU board controls and monitors the
movements of the gantry. It also provides information
about the current position of the gantry, detectors, and
patient table on the gantry display.
The gantry MPU communicates with the CAP through a
serial line. On initialization, gantry operation software is
downloaded from the CAP.
All interface to and from the Gantry MPU board is made
through the System Distribution Interface (SDI) board
except for the Gantry Display, which plugs directly into
the Gantry MPU.
The Gantry MPU board receives commands from the
CAP, and signals from the hand control and the operator
control panels. The Gantry MPU board controls all
motion using onboard software. No direct link exists
between the hand control, operator control panels, CAP,
or drive components.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Motion (1000/2000)
The motion subsystem consists of eight identical motion • Provides manual motor control. Each Limit/Manual
control channels: Movement board connects to a Motor Controller
board.
• Detector Radius (1 and 2)
• Detector Tilt (1 and 2)
• Gantry Rotate
• Gantry Linear
• Table Height (head and foot)
All motion subsystems use identical encoders,
potentiometers, limit switches, and wiring. The table
head and foot subsystems also use the same table
sensors.
The Turbo MPU board has eight motion ports. Each
motion port is associated with a gantry movement and
connects to a Limit/Manual Movement board. The Limit/
Manual Movement board performs three major functions:
• Passes motor enable and velocity signals to the Motor
Controller board as long as the accompanying limit
switch is not tripped
• Provides interface between the position sensors,
(encoder, potentiometer, limit switch, table sensors)
and the motion port
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Motion (3000)
Motion commands originate from the CAP, the hand Potentiometers and Encoders
control, or the operator control panel. The CAP is linked Feedback components consist of potentiometers and
through a serial line to the Gantry MPU board. The hand encoders (except for table height which only uses a
control and operator control panels are linked by
potentiometer). The potentiometers provide absolute
dedicated wiring. The Gantry MPU board is responsible position feedback.
for initiating, monitoring, and displaying requested
motions. During calibration, each motion is positioned at
maximum and minimum reference points, and the
Gantry drive components include motor controllers,
digitized potentiometer values at these points are stored
motors, and associated mechanics. in EEPROM. All other position locations are derived from
Each motion is monitored to halt motion when the these reference points.
software limit of that motion is reached, to check for Encoder pulses clock an associated position counter to
unrequested motions, to check for commanded motions
provide position feedback. Because counter values are
which are not moving, and to ensure that the requested not permanently saved, each encoder counter is
motion is moving in the right direction. calibrated to its potentiometer value at system
initialization.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
The controller receives Velocity and Enable control The dual-speed table height controller is the same as
signals from the Gantry MPU. The Velocity signal is a bi- the table translate controller except for the interface to
the Gantry MPU.
polar value of ±10 VDC generated from the velocity
DAC. Instead of a Velocity signal, the MPU provides a Fast
Enable is generated from the Gantry MPU through relay signal, which pulls in relay K1 on the interface board.
K5 on the SDI board and is dependent on the This bypasses speed resistor K1 causing the table to
move at approximately double the slow speed.
engagement of EStop relay K8.
When the Fast signal is not present, the motor moves at
The detector radius controllers receive Velocity and
Enable control signals from the Gantry MPU. The the slower speed. Like the linear controller, the remaining
velocity signal is a bi-polar value of ±10 VDC generated signals from the gantry MPU are Enable and Direction
and their function is identical to the linear controller.
from the velocity DAC.
The Enable signal energizes relays on the radius
controllers to supply power to the controller. These relays
control base drive and anti-creep circuitry.
For added safety, the Enable signal controls relay K1 in
the controller to supply 60 VAC to the power transistors
for rectified bus voltage.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Patient Table
1000/2000 3000
The patient table has one motion: table height. The table The patient table has two motions: translate and table
may also be moved manually from side to side. height. The table may also be moved manually along the
floor rails for service and the installation or removal of
collimators.
In an emergency, the patient pallet may be manually
pulled out of the gantry head area even during translate
motion.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Back Continued
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Theory
Isotope Information
Isotope Photopeaks Half-life Uses Notes
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Troubleshooting
TROUBLESHOOTING
Location Problem
Detector Hole in Flood
CAP/Detector Flood Quality
CAP (Go to Software and Diagnostics) No Counts from One Head
Gantry No Power
Gantry (1000/2000) Collimator Errors
Gantry (3000) Intermittent Halt Condition
Gantry Motion Stops (1000/2000)
Gantry Motion Stops (3000)
Gantry Motion Speed Incorrect (3000)
No Detector Radius Motion (3000)
No Gantry Rotate Motion (3000)
No Table Height/Translate Motion (3000)
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Troubleshooting
Detector
Detector Troubleshooting
Pre-checks
Hole in Flood 1. Turn off corrections to ensure software corrections are not causing the problem.
2. Check outputs for other heads.
Prism XP 1000/2000, 3000 3. Clean and reseat Main Amp cards.
Loose or Hole
Broken Stays Note: It is highly recommended that a tube be on hand
before the detector is dismantled to replace it. Check TP 7
on the High Voltage distribution board to determine tube
type.
Bad Tube.
Fix and Retest
Replace •RCA 7.4%
•Amperex 1.48%.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Troubleshooting
Detector Continued
Detector Troubleshooting
Pre-checks
Flood Quality 1. Turn of corrections to ensure the uncorrected flood is good.
2. Check outputs for other heads.
Prism XP 1000/2000, 3000
Artifact
Degraded Image
Swap Boards Fix or Replace
Image problem?
Non-uniformity
Rerun
Correction
Table
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Troubleshooting
CAP/Detector
Pre-checks
CAP/Detector Troubleshooting 1. Power. Check fuses and breakers for the detector power supply.
2. Check outputs for other heads.
No Counts from One Head
Prism XP 1000/2000, 3000 Introduction
No counts from only one head can be caused by problems in one of three areas:
• Detector
• Ring Cable
• CAP
Check Check
No Counts Switch
Detector Detector Swap Detector
to One Processor
Power Good Signal PCBs
Head Boards No Counts No Output
Supplies Output
If the Ze signal is
Counts Bad Output not present, swap the
• Quad Linear Gate
• Quad Delay
• Ze Summer
Swap CAP Fix or Ring Cable
PCBs Replace or Wiring If the Gate signal is not
present swap the Pileup
Inspector Card.
Swap the following CAP Ensure all electrical connections
boards one at a time from are clean and tight. Measure
a good detector: resistance across the Ze wire on
• Signal Processor the ring cable.
• A/D Converter
• Energy
• Uniformity
• Linearity
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Troubleshooting
Gantry
Gantry Troubleshooting
No Power
Prism XP 1000/2000, 3000
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Troubleshooting
Gantry
Gantry Troubleshooting
Collimator Errors
Prism S 1000/2000, 3000
Change Contact
Fix or Replace
Sensor Cover on OK OK
and Test
collimator
Recheck Periodically
to Ensure Problem is
Fixed
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Troubleshooting
Gantry
Gantry Troubleshooting
Intermittent Halt Condition
Prism S 1000/2000, 3000
Pre-check
Ensure AC Input 208, 220, or 240 available.
Intermittent Check
System Have Replace CPU
Halt Check Input AC Low Voltage
UPS? No Board
Condition Good Supplies Good
Low or
Fluctuates Yes Bad
Check UPS or
Contact Remove UPS Replace or Fix
Facilities Mgmt
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Troubleshooting
Gantry (1000/2000)
Gantry Troubleshooting
Gantry Motion Stops
Prism XP 1000/2000
Pre-checks
1. Main power. Gantry plasma display should be illuminated. Check main fuses and breakers.
2. Estop or Collision. Check display for Estop condition. Check for flashing Estop buttons.
3. Limits. Movement will stop at soft limits. Check for movement in opposite direction.
Potentiometer
Gantry Select Motion Check Problem
Check gantry Enter Cal Check
Motion with Hand Gantry
display Mode Potentiometer
Stops Controller Display
Encoder
Problem
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Troubleshooting
Gantry (3000)
Gantry Troubleshooting
Gantry Motion Stops
Prism XP 3000
Pre-checks
1. Main power. Gantry plasma display should be illuminated. Check main fuses and breakers.
2. Estop or Collision. Check display for Estop condition. Check for flashing Estop buttons.
3. Limits. Movement will stop at soft limits. Check for movement in opposite direction.
Note: During table movements the potentiometers and
encoders track on a linear gear. Be sure to check for proper
mechanical adjustment to the gear track.
Potentiometer
Gantry Select Motion Check Problem
Check gantry Enter Cal Check
Motion with Hand Gantry
display Mode Potentiometer
Stops Controller Display
Encoder
Note: The table height movement
Problem
does not use an encoder. Check for
the pot and associated wiring only.
Encoder failure is less common.
Check Perform the following checks:
Encoder • Visual inspection
• Wiring to/from encoder
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Troubleshooting
Gantry Troubleshooting
Gantry Motion Speed Incorrect
Prism XP 3000
Gantry Motions:
• Gantry Rotate
• Detector Radius
• Table Height Motion causing fault
• Table Translate on gantry display.
OK
End
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Troubleshooting
Good
Movement
Swap
Replace
Controller
Assembly
Assys.
No Movement
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Troubleshooting
Gantry Troubleshooting
No Gantry Rotate Motion
Prism XP 3000
Pre-checks
1. Main power. Gantry plasma display should be illuminated. Check main fuses and breakers.
2. Estop or Collision. Check display for Estop condition. Check for flashing Estop buttons.
3. Limits. Movement will stop at soft limits. Check for movement in opposite direction.
Power LED
Gantry Motion Illuminated
• Gantry Rotate No DC In
No LED
No Check Fault Check DC In
Check
Gantry Rotate Motor DC In to Motor
Motion The Gantry Rotate Controller Motor
Motor Controller has
two status LEDs: Check the motor and
• Power Fault LED transmission for
• Fault Illuminated obstructions, binding,
or other mechanical
Problems.
Replace
Controller
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Troubleshooting
Good
Movement
Replace
Controller Good
Board
No Movement
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP How to Use this EPST
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP How to Use this EPST
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP How to Use this EPST
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP How to Use this EPST
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP How to Use this EPST
Printing
1. On the Acrobat toolbar, select File, Print.
2. Select whether to print all pages, only the current
page, or a range of pages. (Default is Print All
Pages.)
3. Enable the "Shrink to Fit" option when printing the
11" x 17" block diagrams on 8.5" x 11" paper.
4. Click on OK to begin printing.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP How to Use this EPST
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Self-Paced Training Modules
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Reference Materials
REFERENCE MATERIALS
This EPST has no reference materials.
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials
GE Medical Systems Picker Prism XP Feedback to Author
FEEDBACK TO AUTHOR
If you have comments or suggestions, please email them
to:
EPST.feedback@med.ge.com
Back
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Reference Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Materials