Anda di halaman 1dari 74

combi n line

Blood-Gas-Electrolyte-Analyser

For In-Vitro-Diagnostic Use

Service Manual

Eschweiler GmbH & Co. KG

Development - Manufacturing - Sales - Technical Service

Made in Germany
Service Manual
Manual Version
The latest Manual version is: 4.01
Order No.: 60 8 90 10

Copyright © 2010 by
Print: SMe COMBILINE 401.PDF
!
Eschweiler GmbH & Co. KG, Holzkoppelweg 35, 24118 Kiel

No part of this Manual may be reproduced in any form without the


written permission of Eschweiler GmbH & Co. KG.

Manual History
Manual ANALYSER-Software Revision
Version Version Date

4.01 2500 April 2010

Software Copyrights
The software for the combiline Analyser is the User and Instrument Con-
trol Software for the combiline. The combiline Software is the intellectual
property of Eschweiler GmbH & Co. KG hereafter Eschweiler.
Intellectual property rights shall remain with Eschweiler.

You are entitled to use the combiline Software and the printed accompa-
nying material at your place of work only.

Any violations of property rights or copyright or trademark or conditions


of use may be subject to legal action.

Manufacturer
Eschweiler GmbH & Co. KG
Holzkoppelweg 35
24118 Kiel, Germany

Phone: -49- 431-54 65 80


FAX: -49- 431-54 94 23

Internet: http://www.eschweiler-kiel.de
E-Mail: info@eschweiler-kiel.de

Service Manual
Consumable solutions and spare parts from other manufactur-
ers used in ESCHWEILER Analysers.

We point out specifically that fault-free operation of our analysers can be


guaranteed only when original ESCHWEILER solutions and spare parts are
used.

The name ESCHWEILER with its decades of experience in the field of blood
gas analysis also stands for the quality of the spare parts and consumable
solutions. ESCHWEILER sensors and solutions form a functional unit
whose components are optimally adapted to each other. Measuring accuracy
and long life of the sensors are the goal of this system.

Through intensive development work, the chemical compositions of our


solutions have been adjusted optimally to the ESCHWEILER sensors and
guarantee, besides precise measurements, the greatest possible protec-
tion of the sensitive sensor surfaces.

For example, special wetting solutions are used which have a decisive
influence on the unconstrained contact between sensor and solution in the
measuring capillary. At the same time suitable preservatives prevent the
growth of micro-organisms in the consumable solutions and keep them from
being introduced into the measuring capillaries of the ESCHWEILER
analysers.

The characteristics and concentrations of both the wetting- and the


preservative substances are selected according to the special require-
ments of our sensors.

Fault-free operation of our analysers with measuring accuracy and long


sensor life are the main goals of our quality requirements. Only if
ESCHWEILER original parts and solutions are used we can meet these
requirements.

Please support us in doing this!

Service Manual
Table of Contents
Safety Issues ................................................................................................................................................ 2-5
Technical Data. ................................................................................................................................................ 6

Menue Guide. .................................................................................................................................................. 7

System Description ...................................................................................................................................... 8-11


Tubing. ....................................................................................................................................................... 12-15

Specification of Valves. .................................................................................................................................... 16

Blockdiagram. ................................................................................................................................................. 17
Power Supply ..................................................................................................................................................18

Power Supply ESW-MVTC. ........................................................................................................................ 19-24

Adjustment of the operating voltage for Rollerpumps RP1 and RP2. ................................................................ 21
Adjustment of Thermostate Temperature ......................................................................................................... 23

Analog-Board ESW-ANALOG. .................................................................................................................... 25-35

Adjustment of O2 Polarization Voltage and O2 Zero Point. .............................................................................. 27


Circuit diagram for Sample-light-Barrier ...........................................................................................................28

Circuit Diagram for Hb-unit .............................................................................................................................. 28

Testing the Airpressure Indication. ................................................................................................................... 29


Adjustment of Thermostat Temperature-Indication. .......................................................................................... 30

Adjustment of the Sample Light-Barrier ...........................................................................................................31

Capillary exchange in Sample-Light-Barrier .....................................................................................................32


Hb-Measurement module ........................................................................................................................... 33-35

CPU / AD-Converter CL Z80ADC. .............................................................................................................. 36-40

Serial Interface with SD Card Reader RS232 SD INT1.4 ............................................................................. 41-47


Fill Level Control CL FÜLLSTAND. .............................................................................................................. 48-50

Keyboard. ....................................................................................................................................................... 51

Display / Contrast. .......................................................................................................................................... 52


Thermoprinter GPT-4352-60 .............................................................................................................................54

Liquid-Valves .............................................................................................................................................. 55-56

Delay Circuit for Ventilation Valve ....................................................................................................................57


Suction-Pump ............................................................................................................................................ 58-59

Sensor Spare Parts and Lifetime Information. ............................................................................................. 60-62

Consumeables ........................................................................................................................................... 63-65


Spare Parts ................................................................................................................................................ 66-68

TÜV Certificate ................................................................................................................................................ 69

Declaration of Conformity ................................................................................................................................70

combiline Service- Manual - Version 4.01 page 1


Safety Issues

Please read the Instruction Manual in its entirety prior to operating the
combiline
line. In order to ensure a high level of performance, all warnings and
line
references to technical safety in this Instruction Manual have to be followed.

!1 Hazards and Precautions


The cautions and safety regulations in this Instruction Manual meet interna-
tional classifications:

Warns of a risk of injury or of a risk to life (for example by electrical shock).

DANGER Warns of a risk of injury or of the analyser being severely damaged.

CAUTION Warns of a risk of biological material like blood or other human or animal
specimen. Risk of infection! Potential infectious area/material!

Biohazard
Introduces rules to be observed.

NOTE The following safety issues have to be observed at all


times:

Electrical safety
- Check that the operating voltage is set correctly before you connect the
device to the main power supply.

- To connect the analyser to the main power supply, use only sockets which
are grounded to avoid the risk of an electrical shock.

- Use only grounded extension power cords. The used power cord has to
correspond to the country specific regulations.
DANGER
- Never intentionally disconnect the grounding contacts.
There is the risk of electrical shock if the protective conductor is inter-
rupted within or outside the analyser, and/or the grounded contacts have
been disconnected from the line.

- Never remove protective guards or secured components since you could


expose electrically live parts in this way.

- Electrical connection contacts (plugs, sockets, etc.) can be electrically live.

page 2 Service- Manual - Version 4.01 combiline


Safety Issues

- Even after an analyser has been switched off, components (e.g. capaci-
tors) can be under voltage as the result of an electrical charge.

- All current carrying parts are sources of danger for an electrical


shock.

- Surfaces (floors, work tables) have not to be moist when you are working
with any electrical device.

- Carry out only the maintenance work and/or the replacement of parts
described in these Instruction Manual.

- Unauthorized work on the analyser can lead to the guarantee obligation


becoming null and void with necessary expensive service work to
correct it.

- All work, which requires the analyser to be opened, have to be carried


out by an authorised technician who is familiar with the risks related
thereto.

- Use only replacement fuses of the stated type and with the stated nomi-
nal current. Never use fuses, which have been “repaired”.

- There is a Lithium battery on MAIN BOARD that have to be replaced


by the authorised technical service every 5 years! In case of a battery
fault, all memorised parameter might be lost.

Mechanical Safety (Analyser is


operating)
- Never open or unscrew analyser's casing parts while it is switched ON.
There is a risk of injury due to moving parts as fans, pumps,
stepper-motors or any other mechanical movements.
CAUTION

Biological Material - Risk of Infection


- Wear gloves in all cases if there is a risk of infection.

- Avoid any direct contact with samples that are potentially infectious or
which may generate other risks to the human body (Aids, Hepatitis etc).
Biohazard In case of direct body contact go and wash the contaminated area
immediately. Use a suitable skin disinfection solution. Ask your doctor for
aid.

- If sample material (blood) or reagent is spilled onto the analyser, wipe it off
immediately and refer to the chapter ! 2 Maintenance and Hygiene.

- Do not open a waste bottle as long as the analyser is in process.

combiline Service- Manual - Version 4.01 page 3


Safety Issues

Reagents and Controls


- Observe the suggestions in the package inserts for an exact use of the
reagents and quality controls. Note that reagents and quality control mate-
rial can be biological material!
CAUTION
- Don't use Reagents, Controls and other Liquids after expiration date!
- Avoid any liquid penetration into the analyser.

Restrictions for Samples, Reagents


and Controls
- For consumables no guarantee can be provided for any resistance against
organic solvents. For this reason, do not use any organic solvents unless
such solvents are expressly indicated.
CAUTION
- Do not use any other rinsing or cleaning solutions as recommended by the
manufacturer or representative.

- Waste liquid has to be disposed in compliance with the legislation.

Accuracy and Precision of the


Measured Results
- In order to ensure a flawless operation of the analyser measure control
samples and watch the function of the analyser closely.
CAUTION - Faulty measurement results may result in an incorrect diagnosis or range
danger for patient.

Fire and Explosion Hazards


- Do not place any flammable or hazardous explosive material in the
proximity of the device. Electrical sparks could cause fire or explosions.
CAUTION

Operator Qualification
- The analyser should only be operated by trained personnel. Improper
operation of the analyser might cause inaccurate measuring results!

! 2 Maintenance and Hygiene


- No organic acid based cleaning substances should be applied. Instead
use cleaner designed for cleaning and disinfecting laboratory analysers.
Only use a dampened cloth to clean the analyser.

- Never spray or pour cleaning solution directly onto the analyser, that may
negatively impact the analyser’s functions significantly.
CAUTION - Keep the analyser clean and do not spill liquids onto it.

- In case liquids were spilled onto the analyser, immediately absorb liquid
with a suitable cloth.

- Contact your distributor if your control measurements do not produce the


expected results.

page 4 Service- Manual - Version 4.01 combiline


Safety Issues

Recommended Disinfectant
Solutions.
A suitable disinfectant is available by BODE Chemie, Hamburg:
Bacillol®, is recommended for areas at risk of infection, where rapid
contact times and drying times are necessary.

BODE Chemie GmbH & Co


Melachtonstr. 27 - 22525 Hamburg, Germany

Phone: +49- (0)40-54 00 60 - Internet: www. bode-chemie.de

Another suitable disinfectant is available by S&M Schülke & Mayr:


TERRALIN® Liquid

Schülke & Mayr GmbH


Robert-Koch-Str. 2 - 22851 Norderstedt, Germany

Phone: + 49- (0)40-521 00-0 - Internet: www.schuelke-mayr.com

See the website for your nearest representation. Ask for an EC Material
Safety Data Sheet according to EC 91/155.

!3 First-Aid Measures
If a person swallowed any chemical solution may be while servicing,
wash out mouth with water.
If provided person is unconscious, than call a physician.
If inhaled, remove to fresh air. If breathing becomes difficult, than call a phy-
sician.
In case of contact, immediately wash skin with soap and copious amount of
water.
In case of contact with eyes, flush with copious amounts of water for at least
15 minutes.
Assure adequate flushing by separating the eyelids with fingers, than call a
physician.

!4 Repairs
- Repairs to the analyser may only be carried out by trained personnel, and
replacement parts have to comply with the analyser specifications.

CAUTION - In case of analyser problems contact your representative.

Useful hint
In this Manual a bullet "•" draws your attention to an instruction.
Example:

NOTE • Press enter-key to confirm ...

combiline Service- Manual - Version 4.01 page 5


Technical Data

page 6 Service- Manual - Version 4.01 combiline


Menue-Guide

POWER UP COMBILINE
Rev. 2.500

WARM UP PHASE press 1 x


OPERATION key
OF 1 HOUR
to select Menu

AUTO-CALIBRATION
OF SENSORS OPERATION
1 = RINSE
2 = CALIBRATION *
STAND BY 3 = CALIBRATION BGA * **
ECONOMY
DIALOGUE 4 = CALIBRATION ISE * **
PROGRAM
5 = ACTIVATION HB *
6 = SENSOR PARAMETER *
7= ECONOMY MODE *
8 = DISPLAY MEASURED

DISPLAY MEASURING DIALOGUES press 2 x


MEASURED OPERATION key
Sample introduction by
to select Menu
selection of:
PRINT MEASURE
(SYRINGE or CAPILLARY)
specify:
SERVICE - TEST
- BGA and meta values only 1 = BGA 1 / 3 ***
- ISE and meta values only 2 = BGA 2 / 4 ***
3 = CAL 3 **
QC specify: 4 = CAL 4 **
- LEVEL 1 5 = GAS ***
- LEVEL 2 6 = AIR / TEMP
- LEVEL 3 7 = LIGHT BARRIER
- LEVEL 4 8 = PROTEIN *
9 = REAGENT TEST
ASPIRATION GAS
press 3 x
OPERATION key
Attention!
to select Menu
Sofware functions marked with *
are not accessible during Warming Up!

Software functions marked with ** STANDARD DATA


are not available in BGA types!
1 = CALIBRATION DATA ***
Software functions marked with *** 2 = DATE / TIME
are not available in ISE types! 3 = CORRELATION
4 = SETTINGS
5 = SENSOR SELECTION
6 = AB-PARAMETER ***

combiline Service- Manual - Version 4.01 page 7


System Description

Opened front cover

Order no. Order no.

1 - LCD-Display F100.4.2 11- Suction-Valve F200.25

2 - Function keys 50 8 20 70 (keyboard compl.) 12- Sampleport-Valve F200.28

3 - Numerical keys 50 8 20 70 (keyboard compl.) 13- Input-Valve F200.25

4 - Screw joint of Reagentwall 14- Sample Light-Barrier 40 7 10 20

5 - Display-Contrast 15- Thermostat with Sensors

6 - Peristaltic Pump meta 50 7 40 00 16- Hb-Sensor type II 50 2 10 01

7 - Peristaltic Pump 50 7 40 00 17- Sensorcatch 50 7 20 10

8 - WASH-Valve F200.25 18- Panel for therm. illumination 50 7 20 21

9 - Sampleport-Insert 1-8.106 19- Metabolite valve F200.25

10- Liquid-Valves for Cal.Solutions 50 8 30 20 20-Metabolite unit

page 8 Service- Manual - Version 4.01 combiline


System Description

Opened Front

Order no.

1 - Switching power supply 60 8 10 90

2 - Thermoprinter GPT-4352-60 60 8 20 90

3 - Intermediate Vessel for suction 40 7 10 50

4 - Dampness Absorber 40 7 10 50

5 - Interface RS232 with SD-Card Reader 60 8 20 40

6 - Pinch Valve VE1 and VE2 F200.25

7 - Filling Level Control board CL-FÜLLSTAND 40 8 20 00

8 - Suction Pump F10.19

combiline Service- Manual - Version 4.01 page 9


System Description

Opened Rear Wall

Order no.

11 - Switching Power Supply 60 8 10 90

12 - Intermediate Vessel 40 7 10 50

13 - Thermoprinter GPT-4352-60 60 8 20 90

14 - Analogboard ESW-ANALOG 60 8 10 xx / 60 8 11 xx

15 - Filling Level Control board CL-FÜLLSTAND 40 8 20 00

16 - Rollerpump Motor with gear 50 7 40 10

17 - Suction Pump F10.19

page 10 Service- Manual - Version 4.01 combiline


System Description

Opened Rearwall, inside


CPU Board Z80ADC 1.2 and Power Supply ESW_MVTC

ESW_MVTC 2.0 60 8 10 80

1 - Potentiometer Rollerpump 2

2 - Potentiometer Rollerpump 1

3 - Buzzer

4 - Voltage input from Switching Power Supply

CPU CL Z80ADC 1.2 60 8 10 70

5 - L4 Connector for Analogboard ESW ANALOG

6 - Connector for DC-Supply (+5V)

7 - Driver ICs for Valve and Motor Control

8 - Printer Interface 9600 Baud

9 - Connector for RS232 Interface, 9600 Baud

10- Lithium Battery 3V Type BR2325

11- L11 LCD-Display

12- L10 Keyboard

13- Eprom with Analyser Software

14- L12 Filling Level board CL FÜLLSTAND

15 -RAM

16 - SD-Card Reader with RS232 Interface

combiline Service- Manual - Version 4.01 page 11


page 12
SV20.6

S204-06
VE2
SR20.17
F200.25
combiline BGA

normally open
stromlos offen
SR20.24 suction-pump
Entlüftungsventil
Absaugpumpe
ventilation valve
F10.19

SR20.16

stromlos offen SR20.8


normally opened

SP SR20.18
S104-08

R20.28
Rollerpumpe
F200.25

Feuchtefilter
Rollerpump
Reagent-Transport

Zwischengefäß
(peristaltic-pump) SV20.7

Intermediate-Vessel
Dampness Absorber
SR20.19
F200.25
S204-06 SR20.13
RP SR20.23 SR20.20
stromlos offen VE1
SR20.3
Light barrier for
normally open level control
(CL Füllstand)
Entlüftungsventil
R20.8 ventilation valve

R20.8

SR20.12
SR20.12
Ventilblock
Valve-block SR30.14
stromlos offen Kalibrierlösung
35-8.101 (set) normally open 40-8.101 (set) F200.26
SR20.19 Calibration-Solution
ABS
SR20.22
S204-06
F200.25

Service- Manual - Version 4.01


LV BGA 1 BGA 2
sample-port
Probenport BGA 1
1-8.106
BGA 2

R20.8

stromlos offen PV
EV stromlos offen normally open
normally open
F200.25
F200.25 S204-06
S204-06

LS
Ref pH HB
Rm100.21

pCO2 pO2
waste bottle
Abfallflasche WASH

VERSCHLAUCHUNG / TUBING REV 1.1 ESCHWEILER GmbH & CO. KG KIEL GERMANY
COMBILINE BGA DATUM: 28.07.2008 BEZ.: SP-COMBILINE BGA+HB.TCD VERSION: 1.1

combiline
SV20.6

combiline
S204-06
VE2
SR20.17
combiline ISE
F200.25

normally open
stromlos offen
SR20.24 suction-pump
Entlüftungsventil
Absaugpumpe
ventilation valve
F10.19

SR20.16

stromlos offen SR20.8


normally opened

SP SR20.18
S104-08

R20.28
Rollerpumpe
F200.25

Feuchtefilter
Rollerpump

Zwischengefäß
Reagent-Transport

(peristaltic-pump) SV20.7

Intermediate-Vessel
Dampness Absorber
SR20.19
F200.25
S204-06 SR20.13
RP SR20.23 SR20.20
stromlos offen VE1
SR20.3
Light barrier for
normally open level control
(CL Füllstand)
Entlüftungsventil
R20.8 ventilation valve

R20.8

SR20.12
SR20.12
Ventilblock
Valve-block SR30.14
stromlos offen
35-8.101 (set) normally open 35-8.101 (set) F200.26
SR20.19
ABS LV
SR20.22
S204-06 CAL 3
SR20.9
F200.25 CAL 4
sample-port
Probenport

Service- Manual - Version 4.01


1-8.106

R20.8

stromlos offen PV
EV stromlos offen normally open
normally open
F200.25
F200.25 S204-06
S204-06

LS Ref Cl pH Na Ca Li K E
Rm100.21
Rm100.21
Rm100.21

waste bottle
Abfallflasche WASH 2 CAL 3 CAL 4

Kalibrierlösung
Calibration-Solution

VERSCHLAUCHUNG / TUBING REV 1 ESCHWEILER GmbH & CO. KG KIEL GERMANY


COMBILINE ISE DATUM: 06.10.2006 BEZ.: SP-COMBILINE ISE.TCD VERSION: 1.1

page 13
page 14
SV20.6

S204-06
VE2
SR20.17
F200.25

normally open
stromlos offen
SR20.24 suction-pump
Entlüftungsventil
Absaugpumpe
ventilation valve
F10.19
combiline BGA+E

SR20.16

stromlos offen SR20.8


normally opened

SP SR20.18
S104-08

R20.28
Rollerpumpe
F200.25

Feuchtefilter
Rollerpump

Zwischengefäß
Reagent-Transport

(peristaltic-pump) SV20.7

Intermediate-Vessel
Dampness Absorber
SR20.19
F200.25
S204-06 SR20.13
SR20.23 SR20.20
RP VE1
stromlos offen
SR20.3
Light barrier for
normally open level control
(CL Füllstand)
Entlüftungsventil
R20.8 ventilation valve

R20.8

SR20.12
SR20.12
Ventilblock
Valve-block SR30.14
stromlos offen
35-8.101 (set) normally open 40-8.101 (set) F200.26
SR20.19
ABS LV Calibration-Solution
SR20.22
S204-06 CAL 3 Kalibrierlösung
SR20.9

Service- Manual - Version 4.01


F200.25 CAL 4 BGA 3 BGA 4
sample-port
Probenport BGA 3
1-8.106
BGA 4

R20.8

stromlos offen PV
EV stromlos offen normally open
normally open
F200.25
F200.25 S204-06
S204-06

LS
Ref pH Ca Na K pCO2 pO2 HB
Rm100.30
Rm100.31
Rm100.31

waste bottle
Abfallflasche WASH 2 CAL 3 CAL 4
calibration solution ISE

VERSCHLAUCHUNG / TUBING REV 1.1 ESCHWEILER GmbH & CO. KG KIEL GERMANY
COMBILINE BGA+E DATUM: 28.07.08 BEZ.: SP-COMBILINE BGA+E+HB.TCD VERSION: 1.1

combiline
SV20.6

combiline
S204-06
VE2
SR20.17
F200.25

normally open
stromlos offen
SR20.24 suction-pump
Entlüftungsventil
Absaugpumpe
combiline meta

ventilation valve
F10.19

SR20.16

stromlos offen SR20.8


normally opened

SP SR20.18
S104-08

R20.28
Rollerpumpe Rollerpumpe
F200.25

Feuchtefilter
Rollerpump Rollerpump

Zwischengefäß
(peristaltic-pump) (peristaltic-pump) SV20.7

Intermediate-Vessel
Dampness Absorber
Reagent-Transport

SR20.19
F200.25
S204-06 SR20.13
SR20.23 SR20.20
RP 2 RP 1 VE1
stromlos offen
SR20.3
Light barrier for
normally open level control
(CL Füllstand)
Entlüftungsventil
R20.8 ventilation valve

R20.8
40-8.102 (set)
SR20.12
SR20.12
Ventilblock
Valve-block SR30.14
F200.26
40-8.102 (set) SR20.19
ABS stromlos offen LV Calibration-Solution
normally open sample-port SR20.22
Probenport CAL 3 Kalibrierlösung
S204-06 1-8.106 SR20.9
F200.25 CAL 4+M BGA 3 BGA 4
BGA 3
BGA 4

Service- Manual - Version 4.01


stromlos offen PV
R20.8
normally open
GLU F200.25

EV S204-06
LAC stromlos offen
normally open
F200.25
S204-06

LS
Ref pH Ca Na K pCO2 pO2 HB
Rm100.30
Rm100.31
Rm100.31

META
waste bottle
stromlos offen WASH
normally open Abfallflasche CAL 3 CAL 4+M
2+M
F200.25
S204-06 calibration solution ISE

VERSCHLAUCHUNG / TUBING REV 1.1 ESCHWEILER GmbH & CO. KG KIEL GERMANY
COMBILINE meta BGA+E DATUM: 01.09.2008 BEZ.: SP-COMBILINE II BGA+E HBII.TCD VERSION: 1.1

page 15
Valves

Determination of Valves

EV Input-valve, module holder (one stage pinch valve, normally open)

BGA1 Liquid-valve for calibration solution BGA 1 (normally closed)1

BGA2 Liquid-valve for calibration solution BGA 2 (normally closed)1

BGA 3 Liquid-valve for calibration solution BGA 3 (normally closed)2

BGA 4 Liquid-valve for calibration solution BGA 4 (normally closed)2

CAL 3 Liquid-valve for calibration solution CAL 3 (normally closed)2

CAL 4 Liquid-valve for calibration solution CAL 4 (normally closed)2

LV Liquid-valve for ventilation (normally closed)

ABS Suction-valve (one stage pinch valve, normally open)

META Valve for triggering of the metabolite sensors 3

VE1 Ventilation-valve, intermediate vessel (one stage pinch valve, norm. open)

VE2 Ventilation-valve, suction pump (one stage pinch valve, normally open)

PV Sample port valve (one stage pinch valve, normally open)

SP Rinsing solution valve (one stage pinch valve, normally closed)

1
not available in combiline with electrolytes or metabolites
2
not available in combiline BGA
3
available in combiline meta only

page 16 Service- Manual - Version 4.01 combiline


-15V
+5V M1
Rollerpumpe 2 LCD

combiline
CONVERTER Kontrast
Absaugpumpe Ventile Rollerpumpe 1 Roller-Pump 2
Contrast
Suction-Pump Valves Roller-Pump 1 for metabolites
Beleuchtung
Drucker Füllstands- TASTATUR illumination
Printer kontrolle
+24V
GPT-4352 Level-Control KEYBOARD DISPLAY
100 - 250 V~ CL FÜLLSTAND
50 - 60HZ LMBHT014GC
L1
2 x 250V / 2.5A T CL Tastatur DISP INT SD-CARD
Blockdiagram

M4B M4B RP RP-META


L12 L10 L11
L3

M1 M1 IC25

BUZZER
L2 SER-TTL
9600Baud 8N1
M2 M2 IC26
Lüfter 24V DC SD CARD INT

drivers
Treiber
Fan 1.0W
SD Card Reader
V1 M3 M3 IC27 CPU Z80 ADC RS232
+24V +24V
+5V LCD-Converter +5V

LCD
V1
+12V Data-Output 9600Baud 8N1
+5V +5V +24V Printer MUX ADC

PRINT
-12V
+5V
+12V +12V +12V CPUZ80ADC L4

Z80
-12V

NTC

Schaltnetzteil
switching power-supply
PM110-45-2A
-12V -12V

COOL
Peltier-Element

PELTIER
ESW-MVCT LAC GLU
-12V Sensor Sensor
+12V

ANALOG
HB-Enable

Service- Manual - Version 4.01


M4B VE HEAT TEMP

V1 ADC LAC GLU


ESW ANALOG
NTC Analog-Verstärker / Analog-Amplifiers
ASV

Thermostat- Ventil VE2 Heizer


illumination Valve VE2 Heater
Luftdruck
24V (LED) Air-
Thermostat pressure

LS HB TEMP K Na Ca Cl Li pH pCO pO
2 2

Sensoren / Sensors

Blockschaltbild / Blockdiagram 4.0 ESCHWEILER GmbH & CO. KG KIEL GERMANY


COMBILINE II DATUM: 26.01.2010 BEZ.: Block CL 2 BGA+E+M+HBII 40.tcd VERSION: 4.0

page 17
Switching Power-Supply

Type: PM110-45-2A

+5V Supply for Logic (with Protection Diode1N5630A): Z80 ADC, Level Control board CL Füllstand,
LCD Converter

±12V Analogboard ESW- ANALOG, CL Z80 ADC Multiplexer

+24V Printer, Valves, Suction-Pump, Thermostat-Heating

Power Supply für Pumps, Valves and Heating

Type: ESW-MVTC

Rollerpump 1

+6,2V (E slow) Rollerpump 1 - slow speed (Testpoint MOTOR)

+11,5V (E fast) Rollerpump 1 - fast speed (Testpoint MOTOR)

Rollerpump 2 (available in meta version only)

+6,0V (E slow) Rollerpump 2 - slow speed (Testpoint MOTOR_META)

+8,0V (E fast) Rollerpump 2 - fast speed (Testpoint MOTOR_META)

page 18 Service- Manual - Version 4.01 combiline


Power-Supply ESW-MVTC

Voltage Supply for Pumps and Valves

combiline Service- Manual - Version 4.01 page 19


Power-Supply ESW-MVTC

Voltage Supply for Pumps and Valves

page 20 Service- Manual - Version 4.01 combiline


Power-Supply ESW-MVTC

Adjustment of the operating voltage for Roller pump RP

Preparations:

1. Switch off the analyser.

2. Open the rear wall.

3. Have a voltmeter ready

4. The board ESW-MVTC is located at the inner side of the rear wall

Adjustment of the roller pump 1:

1. Switch the equipment on again.

2. Press Program key and Select menu SERVICE-TEST.

3. Activate the program 9 REAGENT TEST.

4. Connect the voltmeter to the test-points MOTOR (+) and GND1 (ground).

5. Press key „3“ (calibration solution) to run the roller pump 1 with slow speed.

6. Set the voltage at test-point MOTOR to 6.2V using potentiometer TR_SLOW.

7. Press key „2“ (WASH solution) to run the roller pump 1 with high speed.

8. Set the voltage at test-point MOTOR to 11.5V using potentiometer TR_FAST.

Continue with point 14 if second roller-pump is not available.

Adjustment of the roller pump 2:

9. Connect the voltmeter to the test-points MOTOR_META (+) and GND1 (ground).

10. Press key „8“ (Meta) to run the roller pump 2 with slow speed.

11. Set the voltage at test-point MOTOR_META to 6.2V ±0.3V using potentiometer TRM_SLOW.

12. Press key „7“ (Gas) to run the roller pump 2 with high speed.

13. Set the voltage at test-point MOTOR_META to 8.0V ±0.3V using potentiometer TRM_FAST.

14. Switch off the instrument, disconnect Voltmeter

15. Close the rear wall

16. Switch analyser on again

Remark:

The operating voltage for the roller pump has to be checked and adjusted only when an intact, good-as-new
roller pump tube is used.

combiline Service- Manual - Version 4.01 page 21


Power-Supply ESW-MVTC

Circuit Diagram Temperature-Regulation

page 22 Service- Manual - Version 4.01 combiline


Power-Supply ESW-MVTC

Temperature-Regulation

Description:

The Thermostate-Temperature Regulation is located on the power supply board ESW-MVTC

Via the plug connection V1 the low voltage 24 V~/4.5A is supplied. Furthermore the external elements for
heavy-duty heating resistors as well as the temperature sensor (both in thermostats) are connected.

Exchange of Temperature regulation board (ESW-MVTC) or the NTC

After exchange of one of these components, the temperature, indicated on the display, is adjusted to 37°C
with potentiometer T-CAL. The PIC IC must not be exchanged. Otherwise a re-adjustment of the circuit is
required. For this adjustment special equipment is required.

Thermostat temperature correction on-site:

There should be made no adjustments without the required auxiliaries. Adjustments should not be done
without contacting Eschweiler Service department!

LED for Temperature control indication:

A yellow LED for Temperature indication is located at the analyser’s front, below the reagent-wall. This LED is
an indicator for temperature regulation status.

The more the actual thermostat temperature reaches to the nominal value (37°C), the longer the LED is
“ON”.

The LED is permanently “ON” when the temperature is in a range of ±0.5°C from 37°C.

The LED is blinking when the temperature deviation from 37°C is more than ±0.5°C.

The higher the blinking frequency, the higher is the temperature difference (e.g. Warm-up Phase).

In case of defective NTC-sensor or released temperature safety switch, the LED just blinks very short.

The LED is permanently on when the heating system is functioning.

combiline Service- Manual - Version 4.01 page 23


Power-Supply ESW-MVTC

Metabolite Temperature Regulation, Circuit Diagram

page 24 Service- Manual - Version 4.01 combiline


Analog-Board ESW ANALOG 1_01

Layout

combiline Service- Manual - Version 4.01 page 25


Analog-Board ESW ANALOG 1_01

Circuit-Diagram for Sensor Amplifiers

C
A

page 26 Service- Manual - Version 4.01 combiline


Analog-Board ESW ANALOG 1_01

Adjustment of polarization voltage

Preparations:

1. Switch off the equipment

2. Open the rear cover of the instrument

3. The board ESW ANALOG is located at the rear-side of the reagent wall

Adjustment of polarization voltage:

1. Switch the equipment on again

2. Connect a voltmeter to the testpoint UPOL ant pos.B and Ground

3. Adjust the indicated voltage to 0.700V with the potentiometer POL at pos. A

4. Switch the equipment off

5. Close the analogue box, the reagent-wall and the frot cover again

Remark:

The polarization voltage is connected to the platinum cathode of the pO2 sensor. It is set to 0.700V at the
factory

Adjustment of O2 zero point

1. Activate Testprogram 5, GAS, in the SERVICE menue and remove the plug from the pO2-sensor

2. Adjust the voltage, indicated on the screen, to -1.700V with potentimeter ZERO at pos. C

3. Connect sensor cable to pO2 sensor, stop the Testprogram and start a calibration

4. Repeat this adjustment after approx. 30 min. operation time!

combiline Service- Manual - Version 4.01 page 27


Analog-Board ESW ANALOG 1_01

Circuit-Diagram for BP, TEMP, Sample-Light-Barrier, Hb


E

I
D

K
F

G
D

page 28 Service- Manual - Version 4.01 combiline


Analog-Board ESW ANALOG 1_01

Adjustment and Testing of the Barometric Pressure indication

1. Find out the actual atmospheric pressure (Barometric Pressure)

2. Press PROGRAM key and select the menu SERVICE-TEST

3. Select Program 6 AIR / TEMP

4. The atmospheric pressure measured by the internal air-pressure sensor is indicated as follows:

AIRPRESSURE 760 MMHG (example!)

5. Adjust the indicated value to the actual Barometric Pressure with the potentiometer T PS at pos. D

Remark:

A testpoint for measurement of the output voltage of the Barometric Pressure sensor is located at pos. E

combiline Service- Manual - Version 4.01 page 29


Analog-Board ESW ANALOG 1_01

Adjustment of thermostate-temperature indication-circuit


Preparation:

1. Switch OFF the analyser

2. Open the front cover of the instrument

3. The board ESW ANALOG is located at the rear side of the reagentwall

The adjustment can be made with a suitable PT100-Transmitter or alternatively with 2 calibrated resistors. The
adjustment is made for the value +20°C with a resistance of 107.80 Ohm and for +40ºC with a resistance of
115.54 Ohm. At the transmitter and the resistors the same lengths of cable must be used as on the built-in
temperature sensor.

Adjustment procedure:

1. Switch the analyser ON again

2. Press PROGRAM key, and Select menu SERVICE-TEST.

3. Then activate the program 6 - AIR / TEMP

4. Disconnect the temperature sensor at plug PT100 at pos. H

5. Set +20ºC at the PT100-Transmitter and connect it at pin1 and pin3 of plug M2. Alternatively, connect a
precision resistance* of 107.8 Ohm.

6. Set the temperature value in the display to 20.0°C with potentiometer T_ZERO at pos. G

7. Set +40ºC at the transmitter or connect a precision resistance* of 115.54 Ohm.

8. Set the temperature value in the display to 40.0ºC with Potentiometer T_GAIN at pos F

* Corresponding resistors with suitable plugs are available at ESCHWEILER

It may be necessary to repeat this procedure several times until both values are reached.

page 30 Service- Manual - Version 4.01 combiline


Analog-Board ESW ANALOG 1_01

Adjustment of the Sample Light-Barrier

The Sample Light Barrier can be adjusted at potentiometer TR_LB at pos. K on the analog board ESW
ANALOG

For Adjustment:

1. Open the Reagentwall and move it to the front

2. Press OPERATION key, and select menue SERVICE-TEST

3. Then activate the program 7, LIGHTBARRIER

4. Adjust the voltage, indicated for air, with potentiometer TR_LB at pos. K to a value between 2,000 V -
2.300V

4. Press key TEST (WASH solution is filled into the measuring cell). The indicated voltage must be between
0.300V and 0.500V!

Should it not be posssible to reach the a.m. voltage values, an exchange of the Sample Light Barrier or its
capillary might be required. For instructions for LB-exchange please note next page.

combiline Service- Manual - Version 4.01 page 31


Sample-Light-Barrier LB

Capillary exchange
in Sample Light-Barrier LB

By the time, the sensitivity of the sample-light-barrier can


decrease. That means samples as well as calibration
a
solutions are not recognized by the system. This leads to
filling level indication for calibration solutions or even to
permanent wash-cycles. Samples can be washed out
without beeing measured.

The loss of sensitivity is mainly caused by deposits on


figure 1
the inner surface of the light-barrier-capillary. If an
adjustment as described on previous page is not possible,
the capillary needs to be replaced. For replacement
proceed as follows:

- Activate Reagenttransport Program in menu SERVICE

- Remove sensors on left side (figure 1).

- Remove tube from LB-capillary (a) figure 2

- Put the exchange tool on the LB-capillary as shown in


figure 2.

- Push the capillary out of the light-barrier as shown in


figure 3.

- Take a new LB-capillary, place it on the tip of the


exchange tool and push it left into the light-barrier as figure 3
shown in figure 4.

- Reinstall sensors, connect tube to new LB-capillary


and Quit Reagenttransport Program.

After replacement, check and adjust light-barrier voltage


as described on previous page.

figure 4

page 32 Service- Manual - Version 4.01 combiline


Analog-Board ESW ANALOG 1_01 Hb

Measure method

The Hb-sensor is based on the combination of reflexion (scattered) and Transmission measurement of infrared
light.

The data processing of Hb-values in the range of 6 - 30g/dl is done by the Reflexion-measurement method.
The data processing of Hb-values in the range of 3 - 6g/dl is done by the Transmission-measurement method.

The Hb-measure unit is located on the analog amplifier board ESW-ANALOG 1_01.

The following has to be noticed:

- The use of Hb-calibration is not required, only the input of two parameters forrefexion and Transmission
(HB-R and HB-T) is necessary.

- After setting the Combiline into operation, sensor related parameters have to be entered by keyboard. After
that the Hb-sensor is active for measurement. These parameters (HB-R and HB-T) are printed on a white
flag at the Hb-sensor cable.

- The sample volume required for a fully equipped Combiline BGA+E+HB is min. 210 - 240µl. The
sample volume required for a combiline BGA+HB is approx. 160µl. So it is of particular
relevance to use capillaries with adequate volume.

- Please note that this measure method is suitable for measurement of whole-blood only! Quality controls
are not usable!

- The measure range of the Hb-sensor is 3 - 30g/dl

- Hb will be measured if the sample light-barrier detects a whole-blood sample. Otherwise the programmed
standard Hb-value is displayed (e.g. 15g/dl) and used for the calculation of the acid-base parameters.

- In case of too less sample volume, a “Transmission-voltage” of -4.000V is detected. The displayed Hb-
value will be 0.0g/dl then. This can be observed on a sensor-parameter printout which can be made after a
sample measurement (see attached example of printout)

combiline Service- Manual - Version 4.01 page 33


Analog-Board ESW ANALOG 1_01 Hb

Activation of Hb-Sensor type II

Carry out of an Hb-Sensor activation in STAND-BY mode

In order to maintain the measuring accuracy for determination of Hb and to activate the Hb-sensor, an input of
two factors HB-R and HB-T has to be carried out in STAND-BY mode

In case the COMBILINE was switched off for less than 24 hours the Hb-factors are retained in the memory
and the sensor remains ACTIVE.

In case the COMBILINE was switched off for more than 24 hours the Hb-factors have to be entered as
described below, the Hb-sensor is marked with OFF.

The mentioned Hb-factors are printed on a white flag at the Hb-sensor cable (see attached photo)

Carry out Hb-factor input to activate the Hb-sensor:

• Press the 5-key in OPERATION MENU to perform the Hb-activation process, the following dialogue
appears:

HB-SENSOR

HB R : 0.000

HB T :

• Enter the HB-R value given on the white ensign at the Hb-sensor by keypad e.g. 0.970 and press the
Enter-key to confirm.

The following dialogue appears:

HB-SENSOR

HB R : 0.970

HB T : 0.000

• Enter the HB-T value given on the white ensign at the Hb-sensor by keypad e.g. 0.517 and press the
Enter-key to confirm.

• The input range for HB-R and HB-T is 0.300 to 1.800

For quality control of the Hb-Sensor, whole blood specimens whose Hb-concentration have already been
determined with the help of suitable measuring methods might be used (e.g. oximeter).

page 34 Service- Manual - Version 4.01 combiline


Analog-Board ESW ANALOG 1_01 Hb

HB - SENSOR

HBR / HBT 0.970 0.517

MESS. [mV] -1182 -320

printout of Hb-sensor parameter

HB/R = Reflexion factor


HB/T = Transmission factor
MESS = Hb-sensor voltages for Reflexion and Transmissi-
on measurement of the last sample measurement (in
millivolt)

Hb-factors HB-R and HB-T

Hb-sensor type II

Hb-sensor type II installed in combiline (optional)

combiline Service- Manual - Version 4.01 page 35


CPU Board CL Z80 ADC12

Layout

page 36 Service- Manual - Version 4.01 combiline


+5V

+5V
+5V +5V
41 90 16 64
1 Vcc Vss
81 89 D0 1 2 11 10 11 10

combiline
2 CLK/TRG 0 D0 D0 A0 D0 A0
80 88 D1 /M1 3 4 /WR 12 9 12 9
3 CLK/TRG 1 D1 D1 A1 D1 A1
79 87 D2 A7 5 6 /WAIT 13 8 13 8
4 CLK/TRG 2 D2 D2 A2 D2 A2
78 86 D3 A6 7 8 CLK 15 7 15 7
5 CLK/TRG 3 D3 D3 A3 D3 A3
L1 77 85 D4 A5 9 10 /C0 16 6 16 6
6 ZC/TO 3 D4 D4 A4 D4 A4
76 84 D5 A4 11 12 /RD 17 5 17 5
7 ZC/TO 2 D5 D5 A5 D5 A5

HB INTERFACE
75 83 D6 A3 13 14 /C1 18 4 18 4
8 ZC/TO 1 D6 D6 A6 D6 A6
74 82 D7 A2 15 16 /IORQ 19 3 19 3
9 ZC/TO 0 D7 D7 A7 D7 A7
+5V A1 17 18 D7 25 25
10 +5V A8 +5V A8
33 6 A0 A0 19 20 D6 24 24
Circuitdiagram 1/3

RXC A A0 A9 A9
34 5 A1 D0 21 22 D5 J1 21 J2 21
1 TXC A A1 A10 A10
30 4 A2 D1 23 24 D4 23 23
2 W/RDY A A2 A11 A11
31 3 A3 D2 25 26 D3 32K 1 2 8K 26 2
3 SYNC A A3 Vpp/A15 A12 CE2/A13 A12
32 2 A4 /NMI 27 28 /INT 64K 26 32K 1
4 RXD A A4 A13 A14
35 1 A5 /RES 29 30 IEOUT 27
5 TXD A A5 A14 22 /RD
L2 36 100 A6 31 32 IC4 IC5 OE
6 DTR A A6 22 /RD 27 /WR
37 70 A7 A15 EPROM OE A13 RAM WE
7 RTS A A7 RF L5 20 20
38 98 A8 CE CE1

serielle Schnittstelle
serial interface
8 CTS A A8 EXPANSION 28 14 28 14
39 97 A9 +5V +5V Vcc GND +UB Vcc GND
9 DCD A A9
+5V 96 A10
10 A10
49 95 A11 +5V +UB +5V
RXC B A11
48 94 A12
1 TXC B A12 +5V
52 93 A13
2 W/RDY B A13 20 R3
51 92 A14 IC3
3 SYNC B A14 A4 1 Vcc 12 /C0 100K /RES 12

100nF
100nF
100nF
100nF
10uF Ta
100nF
100nF
100nF
10uF Ta
100nF
100nF
50 91 A15 I0 F0 11
4 RXD B A15 A5 2 13 /C1 13 D
CPU Board CL Z80 ADC12

47 +5V + + I1 F1
5 TXD B A6 3 14
L3 46 99 R5 I2 F2 74HC00
6 DTR B A7 A7 4 15 /M1
45 100K I3 F3
7 RTS B A12 5 16 /C90

C3
C4
C5
C6
C7
C8
C9
C11
44 72 IEI I4 F4 14

C10
C13
C14
8 CTS B IEI A13 6 17 +UB 1

Drucker
printer
43 71 IEO DGnd I5 F5 3
9 DCD B IC1 IEO A14 7 18 2 A
RN1 5*100K I6 IC6 F6 IC3
10 Z84C15 A15 8 19
19 /INT 2 I7 GAL F7 7 74HC00
INT /MREQ 9
63 3 /NMI I8 16V8
NMI /IORQ 11
11 /WAIT 4 I9 ESCHWEILER CombiLine
WAIT GND
DA0 29 10 5 CL Z80ADC Rev. 1.2
PA 0 BUSREQ 10 CL Z80ADC 12a.tcd / 01.02.2006
DA1 28 9 6 /RES
PA 1 RESET Ident-No. 2061
DA2 27
PA 2 +5V
DA3 26 8 /M1 +5V +UB
PA 3 M1

Service- Manual - Version 4.01


DA4 25 13 /WR
PA 4 WR R4
DA5 24 14 /RD
PA 5 RD 20 18 100K
DA6 23 15 /IORQ
PA 6 IORQ A0 1 Vcc 12 SEL0 D0 14 Vdd 10 /WR
DA7 22 17 /MREQ I0 F0 D0 WR
PA 7 MREQ A1 2 13 SEL1 D1 13 8 /RD
21 7 +5V I1 F1 D1 RD
+5V ASTRB RFSH A2 3 14 SEL2 D2 12 15 /RES
R7 100K 20 12 R6 I2 F2 D2 CS1
ARDY BUSACK J4 A3 4 15 /RD D3 11
DB0 53 18 100K I3 F3 D3
PB 0 HALT IC3 IC3 A4 5 16 SEL3 A0 4 1
DB1 54 67 4 9 I4 F4 A0 STD.P
PB 1 EV 6 8 A5 6 17 SEL4 A1 5 16
DB2 55 40 5 B 10 C I5 F5 A1
PB 2 ICT1 A6 7 18 SEL5 A2 6 17
DB3 56 42 J3 I6 IC7 F6 A2 IC8
PB 3 ICT2 74HC00 74HC00 A7 8 19 SEL6 A3 7
DB4 57 73 I7 GAL F7 A3 RTC
PB 4 WDTOUT /IORQ 9 3
DB5 58 68 CLK +5V I8 16V8 +5V ALE 72421
PB 5 CLKOUT /WR 11 2
DB6 59 69 B1 I9 CS0
PB 6 CLKIN 2 7 GND GND
DB7 60 65 Vcc RES +5V
PB 7 XTAL1 8 5 10 9
61 Q1 Vbat PFO 1
+5V BSTRB 3 6
R8 100K 62 66 GND WDI
BRDY XTAL2 R1 4 1 +UB
C2 C1 PFI Vout
9.8304
33pF MHz 33pF
+5V R2 IC2 MAX690 = LEGENDE

CPU-PLATINE / CPU BOARD CL Z80ADC REV. 1.2 ESCHWEILER GmbH & CO. KG KIEL GERMANY IDENT Nr.: 2061

SCHALTPLAN / CIRCUIT DIAGRAMS DATUM / DATE : 01.02.2006 NAME : CL Z80ADC 12a.TCD BLATT / PAGE : 1/4

page 37
+5V
+5V
+5V

page 38
22 53 61 66 78 20 10 L11 L2
1 Vcc D0 3 Vcc GND 2 D0
72 62 D0 9 11
2 CLK/TRG 0 D0 D1 4 5
73 63 D1 10 12
3 CLK/TRG 1 D1 D2 7 IC10 6
74 64 D2 7 13
4 CLK/TRG 2 D2 D3 8 74HC273 9
75 65 D3 8 14
5 CLK/TRG 3 D3 D4 13 12
L6 47 68 D4 D Q 5 15
6 ZC/TO 3 D4 D5 14 15
48 69 D5 6 16
7 ZC/TO 2 D5 D6 17 16
49 70 D6 3 17
8 ZC/TO 1 D6 D7 18 19 D7 GRAPHIC-DISPLAY
50 71 D7 4 18
9 ZC/TO 0 D7 CL
15 2
10 SEL0 11 1
1 A0 11 19
Circuitdiagram 2/3

A0 /RES /RES
84 A1 10
A1
+5V 83 A2 +5V +5V 16 3
A2
17 82 A3
RXC A A3
16 16 8
1 TXC A
32 80 /M1 3 Vcc GND 2 /WR
2 W/RDY A M1 12 5
31 79 /RD 4 5 /RS
3 SYNC A RD 14 6
18 77 /IORQ 6 IC11 7 C/D
4 RXD A IORQ 13 8
15 76 /RES 11 74HC174 10 /CE
5 TXD A RESET 1 7
L7 30 57 CLK 13 12
6 DTR A CLOCK D Q
29 54 /INT 14 15
7 RTS A INT DISPLAY ADAPTER DISPINT
9 52 IEOUT
8 CTS A IEO
8 51 IEO
9 DCD A IEI
81 /C90 CL
10 CS
SEL1 9 1
60
XTALI
CPU Board CL Z80 ADC12

59
XTALO +5V
56 +5V
CLKOUT +5V
55
10 OSC 20 10
PC7 1
9 26 18 Vcc GND 2 2
PC6 2
8 IC9 25 5 15 5
PC5 19 3
7 Z84C90 PA0 17 14 IC12 6 6 ESCHWEILER CombiLine
PC4 20 4
6 KIO PA1 15 9 74HC244 11 8 CL Z80ADC Rev. 1.2
L8 PC3 21 5 L13 FREI / FREE CL Z80ADC 12b.tcd / 01.02.2006
5 PA2 13 12 8 9
PC2 23 6 Ident-No. 2061
4 PA3 11 7 13 7
+5V PC1 25 7
3 PA4 9 16 4 4
PC0 26 8
2 PA5 7 3 17 3

Service- Manual - Version 4.01


27 TASTATUR / KEYBOARD 9
1 PA6 5 G1 G2
28 RN2 10
PA7 3 1 19 8*100K
46 SEL2
ASTB 19
+5V 45
ARDY 23 L10
3 42 +5V
RXC B PB0 18 FÜLLSTAND / LEVEL CONTROL
4 41 +5V
1 TXC B PB1 16 L12
11 40 20 10
2 W/RDY B PB2 14 1
12 39 18 Vcc GND 2 2
3 SYNC B PB3 12 2 SCHIEBER / PUSH BOTTOM BGA 1 / 3
2 38 5 15 6
4 RXD B PB4 10 3 ABFALLFLASCHE / WASTE
5 37 14 IC13 6 5
5 TXD B PB5 8 4 SCHIEBER / PUSH BOTTOM BGA 2 / 4
L9 13 36 9 74HC244 11 9
6 DTR B PB6 6 5 SPÜLLÖSUNG / WASH
14 35 12 8 7
7 RTS B PB7 4 6 FREE
6 44 7 13 8
8 CTS B BSTB 21 7 FREE
7 43 16 4 3
9 DCD B BRDY 24 8 FREE
3 17 4
10 GND 2 9 FREE
G1 G2
10 24 33 34 58 67 1 RN3 10
1 19 8*100K
SEL3
pull up widerstände auf
ECO-TAST (4 x 100K)

pull up resistors on
ECO-TAST (4 x 100K)

CPU-PLATINE / CPU BOARD CL Z80ADC REV. 1.2 ESCHWEILER GmbH & CO. KG KIEL GERMANY IDENT Nr.: 2061
SCHALTPLAN / CIRCUIT DIAGRAM DATUM / DATE : 01.02.2006 NAME : CL Z80ADC12b.TCD BLATT / PAGE : 2/4

combiline
+5V
RN7 8*30K
ANALOG-INPUT L4 ANALOG-GND 10 20 RN4 1 8*330 1
D0 3 GND Vcc 9 8 9 RN10 4*4.7K M1

1
3
5
7
9
2
4
6
8
D

11
13
15
17
19
21
23
25
10
12
14
16
18
20
22
24
26
D1 4 2 7 10 2 8 11
1 ISE 1 / 3
D2 7 5 6 11 5 7 12
IC19 2 ISE 2 / 4

combiline
ESCHWEILER CombiLine D3 8 8 5 12 C 13
D 3 ( HB-LAMP )
CL Z80ADC Rev. 1.2 D4 13 7 4 13 14
CL Z80ADC 12c.tcd / 10.01.2006 ILQ1 6 C 4 LAMP - THERMOSTATE
D5 14 IC16 6 3 14 6 15
Ident-No. XXXX 5 PIEP - SAMPLE SOUND

19
20
21
22
23
24
25
26
11
10
9
8
7
6
5
4
D6 17 9 2 15 9 5 16
6 MZ - SAMPLE COUNTER
D7 18 74HC273 6 1 16 D IC25 17
7 VE2
RN11 18

IN1
IN2
IN3
IN4
IN5
IN6
IN7
IN8
IN9
ULN 8 AP - SUCTION PUMP

IN10
IN11
IN12
IN13
IN14
IN15
IN16
Q
5 1 16 RN10 2804
DB0 17 12 SEL4 11 B
A0 IC15 GND 12 2 15 3 4
DB1 16 10
A1 15 3 14 4 3 V+ 9
DB2 15 HI506 1 IC20
Circuitdiagram 3/3

A2 V++ +15V 4 4 13 A
DB3 14 /RES 1
A3 CL 3 5 12
27 ILQ1 B
V-- -15V 16 6 11 7 2
9
EN ANA IN 19 7 10 8 1 GND 10
2 8 9 A
18 28
RN11 4*4.7K

+15V +5V
+5V RN8 8*30K
+15V 10 20 RN5 1 8*330 1
2 VR1 1 -5V + C15 3 GND Vcc 9 8 9 RN12 4*4.7K M2
-15V D
LM79L05 10µF Ta 4 2 7 10 2 8 11
1 EV - INPUT VALVE
+ C12 7 5 6 11 5 7 12
2 ABS - SUCTION VALVE
3 10µF Ta + C16 8 8 5 IC21 12 13
D C 3 SP - WASH
10µF Ta 13 7 4 13 14
4 PV - SAMPLE
+5V -15V 14 6 3 ILQ1 14 6 C 6 15
IC17 5 BGA 1 / 3
-15V 17 9 2 15 9 5 16
6 BGA 2 / 4
18 6 1 16 17
CPU Board CL Z80 ADC12

74HC273 D IC26 7 LV -
RN13 18
8 VE1 -
1 40 28 Q ULN
5 1 16 RN12 2804
GND V+ V- SEL5 11 B
0.15µF CI1 12 2 15 3 4
DB4 18 32 10
LBEN INT 15 3 14 4 3 V+ 9
DB5 19 IC22
HBEN 0.33µF CAZ1 4 4 13 A
DB6 2 31 1
STATUS AZ CL 3 5 12
DB7 26 ILQ1 B
RUN/HOLD 200K RI1 16 6 11 7 2
30 9
BUF 19 7 10 8 1 GND 10
DA0 16
B1 2 8 9 A
DA1 15 38
B2 REFC- RN13 4*4.7K
DA2 14
B3 1uF CR1
DA3 13 37 +5V
B4 REFC+ RN9 8*30K
DA4 12 10 20 RN6 1 8*330 1

Service- Manual - Version 4.01


B5
DA5 11 IC14 35 3 GND Vcc 9 8 9 RN14 4*4.7K M3
B6 INHI D
DA6 10 4 2 7 10 2 8 11
B/ ICL7109 1 ROLLER PUMP RIGHT
DA7 9 34 7 5 6 11 5 7 12
B8 INLO GND IC23 2 ROLLER PUMP LEFT
33 8 8 5 12 C 13
COMMON D 3 ROLLER PUMP SLOW
8 39 13 7 4 13 14
B9 -REFIN ILQ1 6 C 4 ROLLER PUMP FAST
7 14 IC18 6 3 14 6 15
B10 TR1 5 FREE
6 17 9 2 15 9 5 16
B11 36 6 FREE
5 +REFIN 20K 3 18 74HC273 6 1 16 D IC27 17
B12 7 FREE
4 RN15 18
OR UREF 1 +5V ULN 8 FREE
3 2.048V 2 UR1 Q
POL 5 1 16 RN14 2804
17 24 AD580 SEL6 11 B
TEST OCSEL 12 2 15 3 4
20 21 10
CE MODE 15 3 14 4 3 V+ 9
4 4 IC24 13
OSCIN OSCOUT The reference voltage can be adjusted with TR1 to that 1 A
value which is indicated as EREF in the testprogram. CL 3 5 12
22 23 ILQ1 B
Testpoint UREF; GND 16 6 11 7 2
9
19 7 10 8 1 GND 10
Q2
2 8 9 A
4.0 MHz
RN15 4*4.7K

CPU-PLATINE / CPU BOARD CL Z80ADC REV. 1.2 ESCHWEILER GmbH & CO. KG KIEL GERMANY IDENT Nr.: xxxx

SCHALTPLAN / CIRCUIT DIAGRAM DATUM / DATE : 10.01.2006 NAME : CL Z80ADC 12c.TCD BLATT / PAGE : 3/4

page 39
CPU Board CL Z80 ADC12

Specification

AD-conversion: (2 x 8 Bit parallel) 12 Bit AD-converter, ICL7109 int., range -4.096V to +4.096V, resolution
1mV. 16 channel multiplexer, HI506

Driver Output (24 pieces) 3 x 8 Bit outputs, with D-Latch (74HC273), Optocoupler (galvanic insulation)
and driver circuits for loads up to 1A (ULN2803)

Display: Graphic-display NLC240x128BTGLE at L11 via ECODISP

Level control: (1 x 8 Bit) 8 Input-Ports for checking the filling levels of reagents via L12 CL FÜLLSTAND

Serial interface: 1. External interface RS-232 SD INT at L2

2. Thermo printer at L3

EPROM: 64KB Program (27C512)

RAM: 8KB (8K x 8)

Keyboard: Two parallel Ports at L10

Reset-component: battery operation, automatic reset generation, switch over to battery operation

4 Timer and counter channels: Baudrate generation (2 channels) Timer (multitasking software modul)

Remark:

An automatic reset is generated if the operation voltage of the CPU board becomes lower than 4.75V or
higher than 5.25V.

page 40 Service- Manual - Version 4.01 combiline


Serial Interface RS-232 SD INT 1.4

Serial interface with SD-Card Reader RS-232 SD INT 1.4

Technical Data:

Baudrate: 9600 bit/sec

Databit: 8

Parity: no

Startbit: 1 (STX = 02HEX))

Stopbit: 1 (ETX = 03HEX)

no handshake (unidirectional)

connector: terminal RS232 at the rear panel of the instrument

Description:

All data which is sent to the printer after completition of a sample measurement, will also be transmitted to
the RS232 interface at the same time together with CR (carriage return) and LF (line feed) at the end of the
line. The transmission is unidirectional, there is no dialog between COMBILINE and PC. The data has to be
received in a polling mode after starting a measurement. The data is given out in standard ASCII code.

Data recalled from the memory (SD-Card) will not be sent via RS232 !

The data transmission is unidirectional only.

Remark:

The data given by the interface can be detected by any terminal program for example TERMINAL.

combiline Service- Manual - Version 4.01 page 41


Serial Interface RS-232 SD INT 1.4

Printout:

Example of print out of measurering results with corresponding data output

Data output Meaning

(position no.,Value) (Parameter unit)


STX Start of Text (02HEX)
80,CL3000 Type of analyser
81,0945250808 Sample number
82,09:45,25.08.08 Time and Date
0,171.1 pO2 mmHg or kPa
1,1.8 pCO2 mmHg or kPa
2,7.373 pH
3,5.9 K+ mmol/l
4,104 Na+ mmol/l
5,1.38 Ca++ mmol/l
6,141 Cl- (QC) mmol/l
7,0.92 Li+ mmol/l
8,0.0 GLU mmol/l
9,0.0 LAC mmol/l
10,15.0 Hb g/dl
11,45.0 HCT %
12,42.4 H+ nmol/l
13,1.36 Ca++7.4 mmol/l
14,126 Cl- (Blood / Serum) mmol/l
15,37.0 Temperature °C
16,20.9 FIO2 %
17,0.85 RQ
18,1.0 HCO3A mmol/l
19,13.3 HCO3S mmol/l
20,-19.0 BE mmol/l
21,-21.5 SBE mmol/l
22,1.1 TCO2 mmol/l
23,29.0 BB mmol/l
24,99.4 O2SAT %
25,20.5 O2-CT %
26,27.67 P50 mmHg
27,-24.20 AADO2 mmHg
28,-17.1 A-GAP mmol/l
29,4.4 SHUNT %
30,171.1 pO2, temp. correct. mmHg
31,22.81 pO2, temp. correct. kPa
32,1.8 pCO2, temp. correct. mmHg
33,0.25 pCO2, temp. correct. kPa
34,7.373 pH, temp corrected
35,0.00 GLU mg/dl
36,0.00 LAC mg/dl
37,760 Barometric Pressure mmHg
38,101.3 Barometric Pressure kPa
65 Sample type (MESART)*
0 Timeouts (ADCMSK)*
255 Active Sensors (MESMSK)*
* see next page for description ETX End of Text (03HEX)

page 42 Service- Manual - Version 4.01 combiline


Serial Interface RS-232 SD INT 1.4

Meaning of the last three Values of a data output:

1. Value - MESART= kind of measurement

2. Value - ADCMSK = sensors which exceeded the permissible time for measurement (measure-value with ?)

3. Value - MESMSK = sensors which were ACTIVE during measurement

All values are available in ASCII-Format as hexadecimal value. The value range is from 0 to 32768.

Explanation:

Bit Valence MESART ADCMSK, MESMSK

00 1 Syringe PO2

01 2 Capillary PCO2

02 4 BGA only PH

03 8 ISE only K

04 16 QC NA

05 32 Respiratory Gas CA

06 64 Serum, Plasma CL

07 128 Blood LI

08 256 - HB

09 512 - -

10 1024 - -

11 2048 - -

12 4096 - -

13 8192 - -

14 16384 - -

15 32768 - -

combiline Service- Manual - Version 4.01 page 43


Serial Interface RS-232 SD INT 1.4

Example for MESART:

65 64 + 1 Serum, Plasma ; Syringe

82 64 + 16 + 2 Serum, Plasma ; QC ; Capillary

133 128 + 4 + 1 Blood ; BGA only ; Syringe

Example for ADCMSK:

0 0 All Sensors comply with permissible measure times

3 2+1 pO2 + pCO2 did not comply with permissible measure times

Examples for MESMSK:

255 128+64+32+16+8+4+2+1 pO2, pCO2, pH, K+, Na+, Ca++, Cl-, Li+ were ACTIVE during measurement

7 4+2+1 pO2 + pCO2 + PH were ACTIVE during measurement

page 44 Service- Manual - Version 4.01 combiline


Serial Interface RS-232 SD INT 1.4

Layout

RS 232 Interface with SD Card-Reader

combiline Service- Manual - Version 4.01 page 45


Serial Interface RS-232 SD INT 1.4

Circuit Diagram

page 46 Service- Manual - Version 4.01 combiline


DSUB Buchse, 9 polig DSUB Buchse, 9 polig
DSUB plug female, 9 pin DSUB plug female, 9 pin

combiline
1 1
6 RX RX 6
2 2
RX = Receive Data / Daten empfangen 7 7
TX = Transmit Data / Daten senden 3 3
Interface Cable

8 TX TX 8
4 4
9 9
5 5
GND
Computer Analysator / Analyser

ACHTUNG: Brücken nur auf der Computerseite !


ATTENTION: Bridge connections on computer side only !

DSUB Buchse, 25 polig DSUB Buchse, 9 polig


DSUB plug female, 25 pin DSUB plug female, 9 pin

1
Serial Interface RS-232 SD INT 1.4

1
14 RX RX 6
2 2

Service- Manual - Version 4.01


15
3
7
16 3
4
TX
17
TX 8
5 4
18
6
9
19 5
7
20
8
GND
21 Analysator / Analyser
9
22
10
23
11
24
12
25
13

Computer

RS232 INTERFACEKABEL / RS232 INTERFACE-CABLE ESCHWEILER GmbH & CO. KG KIEL GERMANY
ECOSYS II, ECOLYTE, COMBISYS II, COMBILINE DATUM: 01.02.2006 BEZ.: RS232 Kabel.TCD BLATT: 1/1

page 47
Fill Level Control CL FÜLLSTAND

Description:
Via L2 the waste bottle sensor is connected. The control of the Calibrants is made by the Sample Light
Barrier in the Sensorholder (Thermostat).

Via L1 the voltage supply (+5V / GND digital) and the outputs of the comparators are connected. The switch
thresholds of the light barrier LS1 for WASH is adjusted with potentiometer TR1.

The light barriers must fulfil the following conditions:

1. Rinsing solution/calibration solution in tube Umax = 0.300 V

2. Air in tube Umin = 1.000 V

The outputs of the comparators supply the following signals:

empty full

rinsing solution T8 0V +5 V

Push bar BGA 1/3 T4 0V +5 V (released / pressed)

Push bar BGA 2/4 T3 0V +5 V (released / pressed)

waste bottle T5 +5 0V

Check of the waste bottle control

1. Reference voltage at T6 = 2.500 V

2. Sensors of waste bottle: in air, T7 approx. 3.500 V

in tap water, T7 approx. 2.000 V

in distilled water T7 approx. 2.400 V

For testpoints (T...) see circuit diagram CL FÜLLSTAND

Transport of calibration solutions CAL3, CAL4, BGA1, BGA2, BGA3 and BGA4 is controled by the Sample
Light Barrier during calibration procedure!

page 48 Service- Manual - Version 4.01 combiline


Fill Level Control CL FÜLLSTAND

Layout

Filling-Level Control Board CL-FÜLLSTAND

combiline Service- Manual - Version 4.01 page 49


T1
L1.1 +5V

page 50
R16 R17
200 4,7K
R1 R10 TS1 R8
S1.4 S1.1
100K 100K 9.1K
L1 VERBUNDEN MIT Z80 ADC L12
OPB 875 T51 L1 CONNECTED TO Z80 ADC L12
E S
R3 LS2
10 3
- T4
S1.3 S1.2 13
30K TLC272CD IC2_D L1.2 BIT 0
11
Circuit Diagram

+ SCHIEBER FÜR KALIBRIERLÖSUNG BGA 1 / 3


6 BIT1
T6 - T5 L1.3 PUSH BUTTON FOR SOLUTION BGA 1 / 3
2 8 D1 7
- R6 T7 IC1_B
1 5
IC1_A + T9 0.5V
3 R14 R15
+ 4 100K 200 4,7K
1N4148 AUSGANG - ABFALLFLASCHE LM 3302
OUTPUT - WASTE BOTTLE TS2 8
C2 S2.4 S2.1 -
14 T3
R4 0.33µF IC2_C L1.4 BIT 2
R11 9
E S OPB 875 T51 + SCHIEBER FÜR KALIBRIERLÖSUNG BGA 2 / 4
L2.1 100K LS3
10K PUSH BUTTON FOR SOLUTION BGA 2 / 4
L2.3 S2.3 S2.2
C1 R2 R7 C3

100nF 100K 1M 100nF


SENSOR TR1
5K

ABFALLFLASCHE 4
WASTE BOTTLE R12 R13 - T8
1,1K 3K 2
IC2_B L1.5 BIT 3
5
TLS + 12 AUSGANG SPÜLLÖSUNG
Testpunkte / testpoints : OUTPUT WASH SOLUTION

T1 - +5V E S OPB 350


LS1 R9
T2 - Masse / ground 1K
T3 / T4 - +5V = BGA3 / 4 - Schieber gedrückt / push bar pressed
0V = BGA 3 / 4 - Schieber nicht gedrückt / push bar released T2
L1.10 GND - DIGITAL
T5 - +5V = Abfallflasche leer / waste bottle empty
Fill Level Control CL FÜLLSTAND

0V = Abfallflasche voll / waste bottle full

Service- Manual - Version 4.01


T6 - 2.5V = Referenzspannung für Abfallflaschensensor / reference voltage for waste sensor Die Lichtschranken LS 2 und LS 3 befinden sich im Beuteladapter. Sie erzeugen ein Signal bei
Betätigung der Schieber
T7 - Sensorsignal Abfallflasche / sensor signal waste-bottle The light barriers LS 2 and LS 3 are located inside the bag adapter. They produce a signal when
3.500V = Luft oder offen / air or open pressing the push buttons.

2.400V = destilliertes Wasser / distilled water


2.000V = Leitungswasser / tap water Bei Verwendung der Platine im Gerätetyp COMBILINE ISE, werden
an den Steckern S1 und S2 jeweils die Pins 1 und 2 mit einer Steckbrücke überbrückt!
T8 - +5V = Spüllösung vorhanden / WASH solution available
For use in Type COMBILINE ISE, the pins 1 and 2 of the plugs S1 and S2
0V = Spüllösung nicht vorhanden / WASH solution not available are bridged with a jumper-plug!

T9 - ca. 0,5V Referenzspannung für Lichtschranken / reference voltage for light barriers

ESCHWEILER CombiLine2
CL FÜLLSTAND REV 2.0
CL FÜLLSTAND 20.tcd / 01.02.2010
Ident-No. 3011
Blatt 1 / 1

FÜLLSTANDSKONTROLLE / LEVEL CONTROL CL FÜLLSTAND REV 2.0 ESCHWEILER GmbH & CO. KG KIEL GERMANY IDENT Nr. : 3011

FÜR ABFALLFLASCHE UND REAGENZIEN / FOR WASTE BOTTLE AND REAGENTS DATUM / DATE : 01.02.2010 NAME : CL FÜLLSTAND 20.TCD BLATT / PAGE : 1/1

combiline
Keyboard

Specification
The keyboard is operated directly at a parallel 8-bit port. Debouncing of the contacts takes place by by a
progranmable Controller (PIC)

The PIC scans the keyboard-matrix permanently. If a key is pressed, the corresponding value is sent to the
port. By an interrupt the information is interpreted by the analyser software. The corresponding action will be
performed then.

Cleaning

Use a suitable cleaning solution as described in chapter !2 Maintenance and Hygiene and pay attention to
the notes.

Keyboard Adaptor CL-TASTATUR

+5V

AR1 C1
TASTATUR 4*10K 100nF
1 2 28
TOCKI +U MCLR
F1 F2 F4 F3
6
9 RA0 IC1
1 4 C 7 PIC 16C57C
7 25
8 RA1 26
ENTER 2 5 0 8
8 18
7 RA2 RC0 17
3 6 . 9
9 19
6 RA3 RC1 15
20
10 RC2 13
5 RB0
21 L10
11 RC3 11 Z80ADC
4 RB1
22
12 RC4 9
3 RB2
23
13 RC5 7
2 RB3
24
14 RC6 5
1 RB4
25
RC7 3
L2 27
OSC1 1,2

26 L1
OSC2
Q1
4.0 MHz GND
4

ESCHWEILER CombiLine
Tastaturadapter CL TASTATUR Rev. 1.2
CL TASTATUR 12.tcd / 01.02.2010
Ident-No. 3031

TASTATURADAPTER CL TASTATUR Rev. 1.2 ESCHWEILER GmbH & CO. KG KIEL GERMANY IDENT Nr.: 3031

SCHALTBILD UND BESTÜCKUNGSPLAN / CIRCUIT DIAGRAM AND LAYOUT DATUM / DATE : 01.02.2010 BEZ.: CL TASTATUR 12.TCD BLATT / PAGE : 1 / 1

Keyboard Adaptor, Circuit Diagram and Layout


combiline Service- Manual - Version 4.01 page 51
LCD-Display

Specification
Type: NLC240x128BTGLE

15 lines LCD-Display 30 characters / line

Contrast Adjustment

The trim potentiometer for contrast adjustment is located above the roller pump, as shown in figure 42a.

• Turn the potentiometer knob until the contrast of the display is pleasant for the operator.

Contrast Potentiometer for LCD-Display

Display contrast adjustment

page 52 Service- Manual - Version 4.01 combiline


LCD-Converter

Circuit-Diagram and Layout

combiline Service- Manual - Version 4.01 page 53


Thermoprinter GPT-4352-60

Pin configuration for Printer GPT-4352-60

Figure 44 printer board


Voltage supply plug: Interface (TTL/ 4,5V Pegel ):

Pin 1: GND Pin 1:GND

Pin 2:GND Pin 2:TXD (input)

Pin 3:GND Pin 3:RXD (output)


Pin 4:+24V Pin 4:RTS (input)
Pin 5:CTS (output) Pin 5:CTS (output)

Paper specifications:
Roll diameter: up to 60mm
Roll width: 57mm
Paper thickness: 0.04mm

Key-Functions
FEED / ENTER key (1)
Printer self test:
Press key 1 then switch analyser ON.
Keep key 1 pressed for at least 3 seconds.
The printer software version and the character
set will be printed

Paper feed will be performed during normal


operation

1 LED

page 54 Service- Manual - Version 4.01 combiline


Liquid-Valve SMC
Description
Liquids will be sucked in by the control of Liquid Valves. 5 of them are sam-
pled together to an array back beside of the Sample Port inside the housing.
The Liquid Valves are controlled by the NT-Move board.

Following liquids are controlled by the Liquid Valves (depending


onconfiguration)
BGA 1 for Calibration Solution BGA 1
BGA 2 for Calibration Solution BGA 2
BGA 3 for Calibration Solution BGA 3
BGA 4 for Calibration Solution BGA 4
CAL 3 for Calibration solution CAL 3
CAL 4 for Calibration solution CAL 4 (+M)
LV for Air
Sedimentation
After a longer period of use, or the COMBILINELayout has been parked for a
longer time, it can be possible that a sedimentation on the seal can cause a
leakage.
Error evidence:
Valve does not respond of commands: Check magnetic coil with a digital
multimeter. In case of trouble, replace valve.
In case of sedimentations inside of the valve, the liquid cannot be trans-
ported properly. The valve must be cleaned inside.

Replacement of a Liquid Valve


Order no.:
40 8 20 10 Liquid-valve block SMC complete, for COMBILINE BGA+E and
meta versions
40 8 20 15 Liquid-valve block SMC complete, for COMBILINE BGA
40 8 20 20 Liquid-valve block SMC complete, for COMBILINE ISE
Order no. of a single valve:
50 8 30 20 Liquid-valve SMC
• Proceed as follows:
• Power OFF COMBILINE and disconnect it from Mains.
• Unscrew the three slotted-head retaining screws at the right side of the
reagent wall.
• Remove all tubes from the valves.
Each of the valves is assembled together with two screws.
• Unscrew the two two for dismantling.
• Replace the valve to be replaced.
• Plug in the cables at the new valve (don't misplace cable!)
• Assemble the valve group in reverse order.
continued next page
• Power ON COMBILINE and open REAGENT TRANSPORT in SERVICE
menu
• Ventilate the replaced Valve by pressing the corresponding function-key

combiline Service- Manual - Version 4.01 page 55


Liquid-Valve SMC

Cleaning of a Liquid Valve


Proceed as described as follows:
• Dismantle the Valve Group.
• Dismantle the Valve as described above
• Clean the internal parts with suitable cleaning solution.
• For assembling of the valve proceed in reverse order.
• For reassembling of the Valve Group proceed as described b

Liquid Valve - SMC Type

air

CAL3

CAL4

BGA3/1

BGA4/2

outlet

Liquid Valve SMC Type - Fluidics

page 56 Service- Manual - Version 4.01 combiline


Ventilation-Valve VE1

Delay-circuit, Layout and Circuit-Diagram


The delay-circuit CL ASV is mounted on the ventilation valve VE2

combiline Service- Manual - Version 4.01 page 57


Suction-Pump

Replacement of the Suction Pump

The Suction Pump is mounted with an aluminium angle beside the trans-
former and is controlled by the Main Board M1.

For a replacement proceed as follows:


• Switch OFF COMBILINE and disconnect it from Mains.
• Unscrew the two retaining screws on the aluminium angle.
• Remove angle from pump.
• Exchange the two cables to the new motor (don't misplace cables!).
• Exchange the two tubes (don't misplace in/out).
• Assemble the new pump with aluminium angle to its place.
• Switch ON COMBILINE.
• Open the dialogue REAGENTTRANSPORT in SERVICE-TEST menu
• Press 1-key for SUCTION to check its functioning.

Cleaning of Pump's Head

The internal parts of the pump could properly be contaminated


with biological material. There is a risk of infection!

• Wear gloves!
• Uninstall the Suction Pump as described in the chapter above.
• Remove lid (B) from pump head, see figure 49 on next page.Liquid-Valve
• Remove the two screws (A) and the front panel (don't loose centre ring for
motor axis).

• Loosen the allen screw (D) at motor axis.


• Remove both retaining screws for motor and pull pump head from motor
axis.

• Clean the corresponding internal parts with a suitable solution, see chapter
!2 Hygiene.

Attention: There is no spare part available!

• Proceed in reverse order to install the pump head.


• Check the functioning as described in the above.

page 58 Service- Manual - Version 4.01 combiline


Suction-Pump

G
D

E F

AS
F

A B

Figure 49 Suction Pump explored

combiline Service- Manual - Version 4.01 page 59


Sensor Spare Parts
ESCHWEILER combi line ESCHWEILER
automatic analysing systems
Sensor Spare Parts and average lifetime information
Ersatzteil-Liste für BGA-Sensoren
spare part list for BGA sensors

Ref. Haltbarkeit
(Monate)
pO2 Haltbarkeit
(Monate)
pCO2 Haltbarkeit
(Monate)
pH Haltbarkeit
(Monate)
durability durability durability durability
1 (month) (month) (month) (month)

12 - 36
1 1

2 1
2 12 - 48 2 12 - 24

3 3 3
12 - 36
2 12 - 24

4 3-6 4
4 3
3-6

5 12 - 48 5 12 - 48 5 12 - 48 3 12 - 24

6 6 6 4

Referenzsensor kompl. pO2 Sensor kompl. pCO2 Sensor kompl. pH Sensor kompl.
reference sensor compl. pO2 sensor compl. pCO2 sensor compl. pH sensor compl.
50 5 10 00 50 1 10 00 50 1 20 00 50 1 30 00

1 Ref. Ableitsystem 1 Schraubkopf 1 Schraubkopf 1 Schraubkopf


ref. conduct. system screw cap screw cap screw cap
50 5 10 10 50 1 10 21 50 1 10 21 50 1 10 21

2 Schraubkopf 2 Sensoreinheit 2 Sensoreinheit 2 Sensoreinheit


screw cap sensor unit sensor unit sensor unit
50 1 10 21 50 1 10 10 50 1 20 10 50 1 30 10

3 Ref.-Aufsatz 3 O-Ring 8.0 x 1.0 3 O-Ring 8.0 x 1.0 3 Sensorgehäuse


ref.- extension o-ring 8.0 x 1.0 o-ring 8.0 x 1.0 sensor casing
50 5 10 50 50 1 10 50 50 1 10 15 50 1 30 20

4 Hülse mit Membran 4 Hülse mit Membran 4 Hülse mit Membran 4 O-Ring 1.8 x 1.0
membrane shell membrane shell membrane shell o-ring 1.8 x 1.0
50 5 10 30 50 1 10 30 50 1 20 30 50 1 10 25

5 Sensorgehäuse 5 Sensorgehäuse 5 Sensorgehäuse


sensor casing sensor casing sensor casing
50 5 10 20 50 1 10 20 50 1 20 20

6 O-Ring 1.8 x 1.0 6 O-Ring 1.8 x 1.0 6 O-Ring 1.8 x 1.0


o-ring 1.8 x 1.0 o-ring 1.8 x 1.0 o-ring 1.8 x 1.0
50 1 10 25 50 1 10 25 50 1 10 25

BGA_combiline/15012009HEU

page 60 Service- Manual - Version 4.01 combiline


Sensor Spare Parts
Sensor Spare Parts and average lifetime information

Haltbarkeit Haltbarkeit
Na (Monate) ISE (Monate)
durability durability
(month) (month)

12 - 36

K+ 3 - 6
1 12 - 36 2 Ca++ 3 - 6
2 Li+ 3 - 6
Cl- 3 - 6

Na+12-24
3 3 12 - 24

4 4

Na+ Sensor kompl. ISE Sensoren:


Na+ sensor compl. ISE sensors:
50 3 20 00 K+ / Ca++ / Cl- / Li+

1 Ableitsystem K+ Sensor kompl.


conducting system K+ sensor compl.
50 3 20 10 50 3 10 00

2 O-Ring 3.2 x 1.8 Ca++ Sensor kompl.


o-ring 3.2 x 1.8 Ca++ sensor compl.
50 3 20 15 50 3 30 00

3 Na+ Sensorgehäuse Cl- Sensor kompl.


Na+ sensor casing Cl- sensor compl.
50 3 20 20 50 3 40 00

4 O-Ring 1.8 x 1.0 Li+ Sensor kompl.


o-ring 1.8 x 1.0 Li+ sensor compl.
50 1 10 25 50 3 50 00

Ersatzteile ISE Sensoren / spare parts ISE sensors K+ / Ca++ / Cl- / Li+
1 Ableitsystem 3 Sensorgehäuse
conducting system sensor casing
K+ 50 3 10 10 K+ 50 3 10 20
Ca++ 50 3 30 10 Ca++ 50 3 30 20
Cl- 50 3 40 10 Cl- 50 3 40 20
Li+ 50 3 50 10 Li+ 50 3 50 20

2 Hülse mit Membran 4 O-Ring 1.8 x 1.0


membrane shell o-ring 1.8 x 1.0
K+ 50 3 10 30 50 1 10 25
Ca++ 50 3 30 30
Cl- 50 3 40 30
Li+ 50 3 50 30

combiline Service- Manual - Version 4.01 page 61


Lifetime of Sensors

Lebensdauer von Sensoren / Lifetime of sensors


Ableitsystem Membran
conducting system membrane

12 - 36 Monate 6 - 9 Monate
Ref. Referenzsensor
reference sensor 12 - 36 months 6 - 9 months
50 5 10 00

Sensor-Einheit Membran
sensor unit membrane

pO2 pO2 Sensor 12 - 48 Monate 3 - 6 Monate


pO2 sensor 12 - 48 months 3 - 6 months
50 1 10 00
12 - 24 Monate 3 - 6 Monate
pCO2 pCO2 Sensor
pCO2 sensor 12 - 24 months 3 - 6 months
50 1 20 00

12 - 24 Monate
pH pH Sensor
pH sensor 12 - 24 months
50 1 30 00

Ableitsystem 1 Membran
conducting system 1 membrane

+ K+ Sensor 12 - 36 Monate 3 - 6 Monate


K K+ sensor 12 - 36 months 3 - 6 months
50 3 10 00
+ Na+ Sensor 12 - 36 Monate 12 - 24 Monate
Na Na+ sensor 12 - 36 months 12 - 24 months
50 3 20 00
++ Ca++ Sensor 12 - 36 Monate 3 - 6 Monate
Ca Ca++ sensor 12 - 36 months 3 - 6 months
50 3 30 00
- Cl- Sensor 12 - 36 Monate 3 - 6 Monate
Cl Cl- sensor 12 - 36 months 3 - 6 months
50 3 40 00
+ Li+ Sensor 12 - 36 Monate 3 - 6 Monate
Li Li+ sensor 12 - 36 months 3 - 6 months
50 3 50 00

GLU Sensor 30 Tage oder ca. 1000 Proben


GLU GLU sensor
30 days or approx. 1000 samples
50 4 20 00
LAC Sensor 15 Tage oder ca. 400 Proben
LAC LAC sensor
15 days or approx. 400 samples
50 4 10 00
1
Das Ableitsystem der ISE-Sensoren kann bei Beschädigung der Chloridschicht werksseitig regeneriert werden.
In case of damaged chlorid layer the conducting system of ISE sensors can be regenerated by manufacturer.

The lifetimes given in the above table are average lifetimes

page 62 Service- Manual - Version 4.01 combiline


List of Consumeables

Article no.: description packing/unit


Calibration- and Rinsing solutions
for combiline
line BGA
3-4.100 WASH 1 (Rinsing solution for BGA) 6x330 ml
40 6 10 20 BGA 1 (Calibration solution for BGA / low) 12x130 ml
40 6 10 30 BGA 2 (Calibration solution for BGA / high) 12x130 ml

for combiline
line BGA+E
40 6 10 10 WASH 2 (Rinsing solution for BGA plus E) 6x250 ml
40 6 10 40 BGA 3 (Calibration solution BGA plus E/low) 12x130 ml
40 6 10 50 BGA 4 (Calibration solution BGA plus E/high) 12x130 ml
40 6 10 45 CAL 3 (Calibration solution for BGA plus E/low) 12x150 ml
40 6 10 55 CAL 4 (Calibration solution for BGA plus E/high) 12x150 ml

for COMBILINE meta


40 6 10 11 WASH 2+M Rinsing solution 8x250 ml
40 6 10 40 BGA 3 Calibration solution BGA plus E/low 12x130 ml
40 6 10 50 BGA 4 Calibration solution BGA plus E/high 12x130 ml
40 6 10 45 CAL 3 Calibration solution for BGA plus E/low 12x150 ml
40 6 10 56 CAL 4+M Calibration solution for BGA plus E/high 12x150 ml

for combiline
line ISE
3-4.110 WASH 2 (Rinsing solution for ISE) 6x330 ml
3-4.400 CAL 3 (Calibration solution for ISE/low) 6x330 ml
3-4.500 CAL 4 (Calibration solution for ISE/high) 6x330 ml

Special Solutions
50 6 10 80 Protein remover 1 x 100 ml
50 6 10 84 Electrode cleaner 50 ml
50 6 10 88 Activating solution for pH-sensor 20 ml

Quality Controls
Control Solutions for BGA
50 6 20 10 Level I BGA (acidosis) 30 ampoules
50 6 20 20 Level II BGA (normal) 30 ampoules
50 6 20 30 Level III BGA (alkalosis) 30 ampoules

Control Solutions for BGA plus E


50 6 20 40 Level I BGA+E (acidosis) 30 ampoules
50 6 20 50 Level II BGA+E (normal) 30 ampoules
50 6 20 60 Level III BGA+E (alkalosis) 30 ampoules
50 6 20 70 Level IV (high pO2) 30 ampoules

Control Solutions for BGA plus E+Li+


50 6 20 45 Level I EGL (acidosis) 30 ampoules
50 6 20 55 Level II EGL (normal) 30 ampoules
50 6 20 65 Level III EGL (alkalosis) 30 ampoules

Sampling Systems
Capillaries
50 6 30 08 hep. capillaries 8 cm/lg 250 pcs.
50 6 30 10 hep. capillaries 10 cm/lg 250 pcs.
50 6 30 12 hep. capillaries 12 cm/lg 250 pcs.
50 6 30 14 hep. capillaries 14 cm/lg for instruments with Hb sensor 250 pcs.
50 6 30 16 hep. capillaries 16 cm/lg for instr. with meta and Hb sensor 250 pcs.
50 6 30 30 metal stirrers for capillaries 250 pcs.
50 6 30 35 magnet for stirrer
50 6 30 40 end-caps for capillaries 500 pcs.

combiline Service- Manual - Version 4.01 page 63


List of Consumeables

Article no.: description packing/unit


Aspiration Pipes
3-7.200 pipe for control solution, activating solution etc. 10 pcs.
50 6 30 55 ampoule holder for aspiration pipe 1 pc.

Printer Paper
50 6 50 00 printer paper pack.=2 rolls

Cleaning Paste
50 1 10 41 cleaning paste (silicon carbide)

Accessories for Maintenance


40 6 40 00 service-set complete for COMBILINE without Metabolites
consisting of:
40 6 40 10 set tubings for roller-pump (reagents) 2 pcs.
40 6 40 15 tubing for magnetic valve VE1 1 pc.
40 6 40 20 tubing for magnetic valve VE2 1 pc.
40 6 40 25 tubing for magnetic valve ABS 1 pc.

40 6 40 01 service-set complete for COMBILINE with Metabolites


consisting of:
40 6 40 11 set tubings for roller-pump (reagents) 2 pcs.
40 6 40 15 tubing for magnetic valve VE1 1 pc.
40 6 40 20 tubing for magnetic valve VE2 1 pc.
40 6 40 25 tubing for magnetic valve ABS 1 pc.

1-8.106 sample port insert (red) 1 pc.

Sensors
pO2-Sensor
50 1 10 00 pO2-Sensor complete
50 1 10 10 pO2-Sensor-unit with membrane shell
50 1 10 15 gasket for Sensor-unit (8.0 x 1.0)
50 1 10 20 pO2-Sensor-casing with screw-cap
50 1 10 21 screw cap for Sensor-casing
50 1 10 25 gasket for Sensor-casing (1.8 x 1.0)
50 1 10 30 pO2-membrane shells pack.= 5 pcs.
50 1 10 35 pO2-fill solution bot.=20 ml
50 1 10 50 polishing kit for pO2 sensors

pCO2-Sensor
50 1 20 00 pCO2-Sensor complete
50 1 20 10 pCO2-Sensor-unit with membrane shell
50 1 10 15 gasket for Sensor-unit (8.0 x 1.0)
50 1 20 20 pCO2-Sensor-casing
50 1 10 21 screw cap for Sensor-casing
50 1 10 25 gasket for Sensor-casing (1.8 x 1.0)
50 1 20 30 pCO2-membrane shells pack.=5 pcs.
50 1 20 35 pCO2-fillsolution bot.=20 ml

pH-Sensor
50 1 30 00 pH-Sensor complete
50 1 30 10 pH-Sensor-unit
50 1 30 20 pH-Sensor-casing
50 1 10 21 screw cap for Sensor-casing
50 1 10 25 gasket for Sensor-casing (1.8 x 1.0)

page 64 Service- Manual - Version 4.01 combiline


List of Consumeables

Article no.: description packing/unit


Hb-Sensor
50 2 10 01 Hb-Sensor Type II complete

K+-Sensor
50 3 10 00 K+-Sensor complete
50 3 10 10 K+-conducting system
50 3 10 20 K+-Sensor casing
50 1 10 25 gasket for Sensor-casing 1.8 x 1.0)
50 3 10 30 K+-membrane shell with membrane
50 3 10 35 K+-fillsolution (organic tenside, buffer substances) bot.=20 ml

Na+-Sensor
50 3 20 00 Na+-Sensor complete btl.= bottle
50 3 20 10 Na+-conducting system
50 3 20 15 gasket for conducting system (3.2 x 1.8)
50 3 20 20 Na+-Sensor-casing
50 1 10 25 gasket for Sensor-casing 1.8 x 1.0)
50 3 20 35 Na+-fillsolution bot.= 20 ml

Ca++-Sensor
50 3 30 00 Ca++-Sensor complete
50 3 30 10 Ca++-conducting system
50 3 30 20 Ca++-Sensor casing
50 1 10 25 gasket for Sensor-casing 1.8 x 1.0)
50 3 30 30 Ca++-membrane shell with membrane
50 3 30 35 Ca++- fillsolution bot.=20 ml

Cl--Sensor
50 3 40 00 Cl--Sensor complete
50 3 40 10 Cl-- conducting system
50 3 40 20 Cl--Sensor casing
50 1 10 25 gasket for Sensor-casing 1.8 x 1.0)
50 3 40 30 Cl--membrane shell with membrane
50 3 40 35 Cl--fillsolution bot.=20 ml

Li+-Sensor
50 3 50 00 Li+-Sensor complete
50 3 50 10 Li+-conducting system
50 3 50 20 Li+-Sensor casing
50 1 10 25 gasket for Sensor-casing 1.8 x 1.0)
50 3 50 30 Li+-membrane shell with membrane
50 3 50 35 Li+-fillsolution bot.=20 ml

Reference-Sensor
40 5 10 00 reference-sensor complete
50 5 10 10 reference-conducting system
50 3 20 15 gasket for conducting system (3.2 x 1.8)
50 5 10 20 reference-Sensor casing
50 1 10 25 gasket for Sensor-casing 1.8 x 1.0)
40 5 10 50 reservoir for reference-Sensor casing
50 5 10 30 ref.-membrane shell with membrane
50 5 10 35 reference-fillsolution bot.=50 ml

combiline Service- Manual - Version 4.01 page 65


List of Spare Parts

Article no.: description


System casing
60 7 10 00 system casing, complete
60 7 10 10 front cover complete, painted, with print

Sensor fixing device / Thermostate


50 7 20 00 sensor fixing device, complete
40 7 10 30 device for specimen input
50 7 20 10 closing mechanism
40 7 10 20 sample Light Barrier COMBILINE
40 7 10 25 capillary for sample light barrier
50 7 20 20 board for thermostate illumination, (LED)
50 7 20 21 panel for thermostate illumination (LED), complete
50 7 20 05 temperature safety switch
50 7 20 90 temperature regulation sensor NTC (for Thermostat or Meta block)

Sensor fixing device for Metabolites


50 7 20 60 thermostate block, for Metabolite sensors, complete
50 7 20 70 fixing screw, white, for GLU-Sensor
50 7 20 75 fixing screw, black, for LAC-Sensor
50 7 20 80 holding device for GLU-Sensor
50 7 20 85 holding device for LAC-Sensor
50 7 20 95 peltier-element for Meta-block

Calibration-bag adapter
40 7 10 35 bag-adapter COMBILINE, complete
consisting of:
G 40.1 metal-casing
40 7 10 40 slide-bar holder COMBILINE
G 41.3 slide-bar COMBI 1 set
G 40.4 spring for slide-bar 1 set
G 40.5 cannula 1set

40 7 10 36 Suspension for Calibration-Bags

Peristaltic pump
50 7 40 00 peristaltic pump, complete
50 7 40 10 micromotor with gear 97.3:1
50 7 40 20 bush for bearing
50 7 40 30 bearing
50 7 40 40 roller device for peristaltic pump
50 7 40 50 clutch

Reagent-wall
50 7 40 60 tube connector, black
50 7 40 61 tube connector, white, reagent-wall
50 7 40 62 tube connector, white, bottle compartment
50 7 40 70 holder for roller pump tubing, reagents (low)
40 7 40 70 holder for roller pump tubing WASH (high)

40 7 10 50 intermediate vessel
40 7 10 55 holder for intermediate vessel

page 66 Service- Manual - Version 4.01 combiline


List of Spare Parts

Article no.: description


Bottles for COMBILINE BGA+E
Rm 100.24 waste-bottle 250ml, empty
Rm 100.30 suction tube for reagent bottle WASH (250ml)
Rm 100.31 suction tube for reagent bottle CAL3/4(+M) (150ml)
R 100.21 screw cap for reagent- WASH- or waste-bottle (150 / 250ml)
60 8 20 80 waste bottle device, complete

Bottles for COMBILINE BGA and COMBILINE ISE


Rm 100.23 waste-bottle 330ml (empty)
Rm 100.21 suction tube for reagent bottle WASH / CAL (330ml)
R 100.20 screw cap for CAL- WASH or waste-bottle (330ml)
60 8 20 70 waste bottle device, complete for 330ml bottle

Sample port
Rm 20.80 sample port device, complete with insert
consisting of:
R 20.81 (1-8.105) sample port insert with integr. Steel canula (white)
R 20.83 (1-8.106) sample port insert (red)

G 20.51 sample port holder


G 20.52 sample port cover
G 20.53 sample port support

Wiring
SEm 100.21 sensor-cable pO2
SEm 200.21 sensor-cable pCO2
SEm 300.21 sensor-cable pH
SEm 400.21 sensor-cable Na+
SEm 500.21 sensor-cable K+
SEm 600.21 sensor-cable Ca++
SEm 700.21 sensor-cable Li+
SEm 800.21 sensor-cable Cl-
SEm 900.22 sensor-cable reference

Valves
F 200.25 tube valve (normally open)
F 200.28 sample port valve PV (normally open)

50 8 30 20 liquid Valve SMC


40 8 20 11 liquid-valve block SMC complete for COMBILINE BGA+E
40 8 20 16 liquid-valve block SMC complete for COMBILINE BGA
40 8 20 21 liquid-valve block SMC complete for COMBILINE ISE
40 8 20 25 multilayer valve carrier SMC

combiline Service- Manual - Version 4.01 page 67


List of Spare Parts

Article no.: description


Electronic boards
(ESW- ANALOG)
60 8 10 10 analog-board for pO2 -pCO2-pH-baro.pressure, temperature
60 8 10 11 analog-board for pO2 -pCO2-pH-baro-pressure-temperature-Hb
60 8 10 15 analog-board for BGA, BP, K-temperature
60 8 10 20 analog-board for BGA, BP, K-Na-temperature
60 8 10 25 analog-board for BGA, BP, K-Na-Ca-temperature
60 8 10 26 analog-board for BGA, BP, K-Na-Cl-temperature
60 8 10 30 analog-board for BGA, BP, K-Na-Ca-Cl-temperature
60 8 10 35 analog-board for BGA, BP, K-Na-Ca-Cl-Li-temperature

60 8 11 10 analog-board for pO2 -pCO2-pH-Meta-baro.pressure, temperature


60 8 11 11 analog-board for pO2 -pCO2-pH-Meta-baro-pressure-temperature-Hb
60 8 11 15 analog-board for BGA, Meta, BP, K-temperature
60 8 11 20 analog-board for BGA, Meta, BP, K-Na-temperature
60 8 11 25 analog-board for BGA, Meta, BP, K-Na-Ca-temperature
60 8 11 26 analog-board for BGA, Meta, BP, K-Na-Cl-temperature
60 8 11 30 analog-board for BGA, Meta, BP, K-Na-Ca-Cl-temperature
60 8 11 35 analog-board for BGA, Meta, BP, K-Na-Ca-Cl-Li-temperature

60 8 10 40 analog-board for K-Na-temperature


60 8 10 45 analog-board for K-Na-Ca-temperature
60 8 10 50 analog-board for K-Na-Ca-Cl-temperature
60 8 10 55 analog-board for K-Na-Ca-Cl-Li-temperature

40 8 10 60 air pressure sensor 1240-015A-3N

60 8 10 70 CPU board (Z80 ADC)


60 8 10 80 power supply board for valves, motors and
temperature regulation (CL TEMP NETZ)
60 8 10 90 switching power supply PM110 45-2A for +5V, -12V, +12V, +24V compl.
40 8 20 00 board for filling level control (CL FÜLLSTAND)
E 210.5 displayadapter (ECODISP)
60 8 20 40 interface RS232 for data output, incl. SD-Card Reader

Various electronical and electrical devices


50 8 20 65 temperature measuring sensor
50 8 20 70 keyboard, complete
50 8 20 71 keyboard adapter CL TASTATUR
F 100.4.2 LCD Display 240x128 dpi
60 8 20 50 LCD Converter
60 8 20 90 thermal printer GBT-4352
F 10.5.1 heating resistor for thermostate heating device
60 8 20 60 fan for power supply 24V, 1W
F 10.19 suction pump

page 68 Service- Manual - Version 4.01 combiline


combiline Service- Manual - Version 4.01 page 69
ESCHWEILER
„ automatic analysing systems
Eschweiler GmbH & Co. KG z Holzkoppelweg 35 z D - 24118 Kiel phone: ..49-(0)431-546580 e-mail: info@eschweiler-kiel.de
fax: ..49-(0)431-549423 homepage: www.eschweiler-kiel.de

Declaration of EC-Conformity
as defined by EC In-Vitro Diagnostica (IVD) directive 98/79/EG

Eschweiler GmbH & Co. KG


Holzkoppelweg 35 / D-24118 Kiel / Germany

declares that the product

Description: blood gas-/ electrolyte- and metabolite - analysers

Model: ESCHWEILER combi line


to which this declaration relates is in accordance with the provisions of the following directives

2004/108/EG Electromagnetic Compatibility


2006/95/EG Low Voltage Directive

Applied harmonised standards:

EN 61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use:
General requirements
EN 61010-2-08/A1 Safety requirements for electrical equipment for measurement, control, and laboratory use:
Particular requirements for laboratory equipment for analysis
EN 61010-2-101 Safety requirements for electrical equipment for measurement, control, and laboratory use:
Particular requirements for in vitro diagnostic (IVD) medical equipment
Emission:
EN 55011 A2 Industrial, scientific and medical radio-frequency equipment – Emission: Class B
EN 61000-3-2 Harmonic current emissions
EN 61000-3-3 Voltage fluctuation and flicker

Immunity:
EN 61326-1 Electrical equipment for measurement, control and laboratory use – EMC requirements:
Industrial requirement
EN 61000-4-2 Electrostatic discharge immunity test
EN 61000-4-3 +A1 Radiated, radio-frequency electromagnetic field – immunity test
EN 61000-4-4 Electrical fast transient/burst immunity test
EN 61000-4-5 Surge immunity tests
EN 61000-4-6 Immunity to conducted disturbances, induced by radio frequency fields
EN 61000-4-11 Voltage dips, short interruptions and voltage variations immunity test

ESCHWEILER GmbH & Co. KG

Kiel, 01.04.2010

.....................................
i.A Dr. H.-J. Heupke
(QCM)
EC declaration combilineV2 IVD engl/01012010

ESCHWEILER Telefon: 0431 / 546580 Deutsche Bank AG, Kiel Commerzbank, Kiel Postbank
GmbH & Co. KG Telefax: 0431 / 549423 Kto.-Nr. 0324 426 Kto.-Nr. 7224488 Hamburg 24676 - 204
Holzkoppelweg 35 info@eschweiler-kiel.de (BLZ 210 700 20) (BLZ 210 400 10) (BLZ 200 100 20)
D-24118 Kiel www.eschweiler-kiel.de SWIFT: DEUTDEHH 210 SWIFT: COBADEFFXXX SWIFT: PBNKDEFF
IBAN: DE46 2107 0020 0032 4426 00 IBAN: DE66 2104 0010 0722 4488 00 IBAN: DE56 2001 0020 0024 6762 04

Gerichtsstand / Place of Jurisdiction: Kiel USt.-ID Nr.: DE 188048227


Handelsregister Kiel HRA 3736 KI Kompl.: Eschweiler Verwaltungs GmbH Kiel HRB 4610 KI Geschäftsführer: Horst H. Kock

page 70 Service- Manual - Version 4.01 combiline