Systems Data
No. MSDXR0106EA MGU-1000A
APPLICATION
This unit is the MGU-1000A, an inverter-controlled X-ray
mammography unit with a cathode-grounded X-ray tube.
FEATURES
• The source-to-image-receptor distance (SID) is 70 cm.
– The 70-cm SID creates a large effective space, facilitat-
ing the compression operation.
• An isocentric C-arm is used.
– The C-arm can be rotated around the breast (about the
rotation center) regardless of the C-arm position (CC,
MLO, ML, ISO, FB, LMO, LM, or SIO), facilitating posi-
tioning.
• C-arm operating switches with distinctive shapes are
used.
– The distinctive shapes of the C-arm operating switches
allow them to be identified easily, permitting rapid
switch operation.
• C-arm up/down movement can be performed using the • Additional molybdenum (30 µmMo) and rhodium (25
footswitch. µmRh) filters are provided to lower unnecessary soft X-
– When positioning of breast is performed using both ray exposure.
hands, the C-arm can be moved up/down using the – The filter is switched automatically.
footswitch. – The rhodium filter is useful for increasing image con-
• The X-ray beam limiting device is motor-driven. trast in large breasts and in breasts with large amounts
– The exposure field is set automatically according to the of glandular tissue.
selected radiography technique. • An inverter type X-ray high-voltage generator and an X-
• A hand rest is provided for the examinee. ray tube with high-speed anode rotation (9700 rpm) are
– Examinees can undergo breast compression in a com- used.
fortable position. – X-ray output suitable for mammography can be stably
• The automatic exposure control (AEC) function is provid- obtained.
ed. – Magnification radiography (1.5 times) can be per-
– The AEC detector position can be selected from formed with a small focus (0.1 mm) (option).
among 10 options, permitting optimum position selec-
tion according to the projection direction.
– The AEC detection field size can be switched between
large and small, permitting optimum size selection
according to breast size (stable AEC is possible even
for mammography of small breasts).
• Either electrical or manual operation can be selected for
breast compression.
– The maximum compression force for electrical opera-
tion can be set.
– In order to minimize examinee discomfort from com-
pression, compression can be released automatically
when X-ray generation is completed.
– The thickness of the compressed breast and compres-
sion force are displayed digitally.
MGU-1000A
COMPOSITION – mAs range that can
be controlled: 2 mAs to 600 mAs (large focus)
Standard configuration 2 mAs to 200 mAs (small focus)
• C-arm stand........................................................................1 – Film density setting: 15 steps
• X-ray tube assembly (DRX-B3856HED-Mo) ....................1 (degree of change: 6%)
• X-ray high-voltage generator ............................................1 • X-ray exposure control
• Accessories – Automatic kV/mAs control (Full-auto)
– Compression plate – kV pre-setting/automatic mAs control (Semi-auto)
(for 18-cm × 24-cm contact radiography) ....................1 – kV/mAs pre-setting (Manual)
– 18-cm × 24-cm Bucky device ........................................1 • Self-diagnostic function
– 18-cm × 24-cm film marker ............................................1 Unsuitable operating operations are indicated by the
– Footswitches for C-arm operation error messages and codes.
(for breast compression and up/down operations) ......2
– Face guard ......................................................................1 C-arm stand
– Compression plates for replacement • Movement of the C-arm stand
(for 18-cm × 24-cm contact radiography) ....................1 – Up/down movement
stroke: Approx. 69 cm (by motor-driven
Optional items movement)
• Magnification radiography unit (MGMK-1000A) – Rotation: Approx. 180° to -150° (by motor-
• X-ray protective baffle (XAPS-1000A) driven movement)
• 24-cm x 30-cm radiography unit (MGAK-1500A) – Locking method: Self-locking type
• Display panel on the base of the stand (MGSP-1000A) – Rotation angle display:180° to -150° (in steps of 1°)
• Biopsy compression plate (MGCD-1000B) – Rotation angle setting
• Handswitch for X-ray exposure operation (MGHS-1000A) · Manual setting:
Set the C-arm rotation angle using the C-arm operat-
PERFORMANCE AND SPECIFICATIONS ing switches at either side of the C-arm.
Operation ratings · Automatic setting:
• Small focus: 2 mAs to 200 mAs, The CC position is set automatically. For the left/right
22 kV to 35 kV MLO positions, after an MLO position is set and radi-
• Large focus: 2 mAs to 600 mAs, ography is performed at the position, the other (oppo-
22 kV to 39 kV site) MLO position is set automatically. For the ML
Note: The kV setting range when the Mo filter is selected is from positions, after radiography at the MLO position is
22 kV to 33 kV for both the large and small focuses. performed and the C-arm operation switch is pressed
in the same direction, the positions are set.
Additional filters • SID: Approx. 70 cm
• X-ray detection field for AEC
Half value layer – Detection field
Tube voltage (kV) Additional filter
(minimum) movement range: 10 locations (by motor-driven
0.03 mmMo 0.35 mmAl movement)
33 kV
0.025 mmRh 0.41 mmAl
39 kV 0.025 mmRh 0.44 mmAl
Note: The half value layer shown above was measured using an
aluminum plate with a purity of at least 99.9%.
Standard configuration
X-ray high-voltage generator
• Output rating: 2.8 kW (30 kV, 92 mA, 0.1 s)
• X-ray exposure condition setting
– kV setting range: 22 kV to 39 kV (in steps of 1 kV)
– mAs setting range: 2 mAs to 600 mAs
(51-step setting)
• Automatic exposure control (AEC) function
This function is used in contact radiography and magnifi-
cation radiography.
– kV setting range: 22 kV to 39 kV Chest wall side
LARGE 25 mm
· Magnification spot
radiography: Approx. 13 cm × 13 cm
SMALL
12 mm
(1.5 times) (when optional magnification
spot radiography has been
37 mm selected)
3
MGU-1000A
X-ray protection baffle MASS
XAPS-1000A
• C-arm stand: Approx. 350 kg
• Size of the main unit: 800 mm (width) × 1,680 mm
• X-ray high-voltage
(height) × 12 mm (depth)
generator (control unit): Approx. 190 kg
• Mass: Approx. 58 kg
• Maximum usable tube
voltage: Less than 150 kV
INSTALLATION CONDITIONS
• Lead equivalent of Power conditions
protective plate: 0.5 mmPb (when tube voltage is • Type: Single-phase AC
49 kV) • Nominal line voltage: 200, 220, 230, 240 V
• Line frequency: 50 Hz/60 Hz
24 cm × 30 cm radiography unit • Allowable voltage
MGAK-1500A fluctuation range: Within ±10% of the nominal line
• Configuration voltage indicated above
– 24 cm × 30 cm Bucky device • Allowable power
– Compression plate impedance: 0.33 Ω or less for 200 V
– Film marker 0.36 Ω or less for 220 V
• 24 cm × 30 cm Bucky device 0.38 Ω or less for 230 V
– Grid 0.40 Ω or less for 240 V
· Grid density: 41 lines/cm • Recommended capacity
· Grid ratio: 3.5 : 1 of the power transformer: 4.2 kVA or more
· Focusing distance: 70 cm
· Interspace material: Fiber Grounding
– Movement method: Reciprocating method Grounding must be provided in accordance with all applic-
– Tabletop: CFRP able legal requirements for medically used electrical
– Cassette size: 24 cm × 30 cm equipment.
(applicable to Kodak Min-R2 or
equivalent) Minimum ceiling height
• Compression plate 2,300 mm
– Material: Polycarbonate
• Film marker Ambient conditions
– Applicable unit: 24 cm × 30 cm Bucky device • Operating conditions
– Ambient temperature: 10°C to 35°C
Display panel on the base of the stand – Relative humidity: 30% to 80% (no condensation)
MGSP-1000A – Atmospheric pressure: 860 hPa to 1,060 hPa
• Applicable unit: C-arm stand of the MGU-1000A – Atmosphere: Do not use the unit in an atmos-
phere containing explosive or
Biopsy compression plate corrosive gases.
MGCD-1000B • Storage conditions
• Material: Polycarbonate – Ambient temperature: -10°C to 50°C
– Relative humidity: 30% to 80% (no condensation)
Handswitch for X-ray exposure operation – Atmospheric pressure: 860 hPa to 1,060 hPa
MGHS-1000A
• Applicable unit: MGU-1000A Leakage current value
Grounding Enclosure Patient leakage
Power supply
leakage current leakage current current
AC: 0.1 mA
Normal state 5 mA or less 0.1 mA or less or less
DC: 0.01 mA
or less
AC: 0.5 mA
Single-fault or less
10 mA or less 0.5 mA or less
state DC: 0.05 mA
or less
4
MSDXR0106EA
COMPLIANCE
• IEC60601-1 (1988)/A2 (1995)
• IEC60601-1-2 (2001)/A1 (2004)
• IEC60601-1-3 (1994)
• IEC60601-1-4 (1996)/A1 (1999)
• IEC60601-1-6 (2004)
• IEC60601-1-8 (2003)
• IEC60601-2-28 (1993)
• IEC60601-2-32 (1994)
• IEC60601-2-45 Ed.2 (2001)
5
MGU-1000A
OUTLINE DRAWINGS
18
15
0°
(40.3)
(27.6)
1,024
700
1,520 to 2,210
(59.8 to 87.0)
(27.2 to 54.3)
690 to 1,380
Film
(Bucky device)
692 735
(27.2) (28.9)
C-arm stand
Unit: mm (in)
6
MSDXR0106EA
OUTLINE DRAWINGS
(54.4)
1,381
(39.8)
1,012
810
(31.9)
930 408
(36.6) (16.1)
Unit: mm (in)
7
MGU-1000A
OUTLINE DRAWINGS
(10.0)
(7.0)
253
178
(3.9)
99
(1.6)
40
298
(11.7)
Compression plate
For 18-cm × 24-cm contact radiography
(11.1)
282
340
(13.4)
(2.0)
(1.8)
50
45
8
MSDXR0106EA
OUTLINE DRAWINGS
3,000 ±100
(118.1 ±3.9)
475
(18.7)
(7.1)
181
2.5 34
(0.1) (1.3)
234 110
(9.2) (4.3)
Face guard
Unit: mm (in)
9
MGU-1000A
OUTLINE DRAWINGS
(11.1)
282
(10.8)
274
346
(13.6)
(73.2)
1860
342 44
(13.5) (1.7)
808 215
(31.8) (8.5)
(2.0)
52
Unit: mm (in)
10
MSDXR0106EA
OUTLINE DRAWINGS
(2.8)
72
198
(7.8)
233
(9.2)
(1.3)
33.1
Display panel on the base of the stand
MGSP-1000A
φ39
(10)
254
(7.0)
178
(φ1.5)
(5.3)
134
298
(11.7)
(3.9)
99
(1.6)
40
178
(7.0)
Unit: mm (in)
11
MGU-1000A MSDXR0106EA
2008-6 TME/KI Made for Life is a trademark of Toshiba Medical Systems Corporation.
Printed in Japan