Anda di halaman 1dari 6

CLINICAL CHEMISTRY

Original Article

Accuracy of Portable Blood


Glucose Monitoring
Effect of Glucose Level and Prandial State
DINO A. VALLERA, M.D., MICHAEL G. BISSELL, M.D., PH.D., M.P.H.,
AND WILLIAM BARRON, M.D.

Glucose was determined on capillary and venous blood samples the measurement was performed within 4 hours of eating,
from 274 adult diabetics by three different methodologies: the whereas no such effect was seen with Accu-Check error. Multiple
Glucoscan 2000® and Accu-Check II® portable glucose meters regression analysis revealed that the Glucoscan measurement
(capillary) and the Kodak Ektachem 700® analyzer (venous). was independently influenced by both venous plasma glucose
Both glucose meters correlated significantly with the Ektachem and prandial state, whereas the Accu-Check II measurement
results. A significant positive bias was found for the Glucoscan was not dependent on either variable. The within-run precision
compared with Ektachem, not found with the Accu-Check II. for both glucose meters were comparable. (Key Words: Portable
The Accu-Check performed better than the Glucoscan at venous glucose monitoring; Accuracy.) Am J Clin Pathol 1991;95:247-
plasma glucose levels < 1 g/L. The mean error of Glucoscan 252
determinations was significantly greater and biased positive when

Portable capillary blood glucose meters have been used termined by a laboratory reference methodology.12 Con-
extensively in both inpatient and outpatient settings for sideration of this factor is important because there may
many reasons, including patient and physician conve- be significant peripheral tissue glucose extraction in the
nience and cost.1,2 Although the accuracy and precision postprandial state,13 resulting in venous plasma levels that
of this technology have been addressed by a number of are considerably lower than capillary values.
previous investigators,3,4 many of these studies have had The present investigation was designed to address a
considerable limitations. For example, sample sizes are number of these issues by performing a large number of
often small,5"7 meter accuracy frequently is addressed only capillary blood glucose determinations on two currently
through determination of the correlation coefficient be- popular instruments, analyzing differences between por-
tween meter and reference laboratory methods,5,8"12 and table meter and standard laboratory measurements as a
the impact of ambient plasma glucose levels on accuracy function of both ambient plasma glucose level and the
often is neglected. Furthermore, we are aware of only one time since the patient's last meal.
study that has addressed the impact of prior feeding on
the concordance between capillary blood glucose mea- METHODS
sured by a portable meter and venous plasma glucose de-
Study subjects included 274 adult diabetics on whom
a blood glucose level had been requested by their primary
From the Departments of Pathology, Medicine, and Obstetrics and
care physician in the University of Chicago Health Service,
Gynecology, University of Chicago Hospitals, Chicago, Illinois. a large internal medicine clinic with more than 50,000
Received September 18, 1989; accepted for publication July 24, 1990. outpatient visits per year. Participants included patients
Dr. Vallera is currently with the Department of Pathology, Hinsdale on dietary therapy only, on oral hypoglycemic agents and
Hospital, Hinsdale, Illinois; Dr. Bissell is currently with the Department on insulin. Each subject was asked when he had last eaten,
of Clinical Pathology, City of Hope National Medical Center, Duarte,
California. and the time was recorded to the nearest 30 minutes.
Supported by Boehringer Mannheim Diagnostics, Indianapolis, In- Blood was obtained from an antecubital vein and placed
diana. immediately into a tube containing potassium oxalate
Address reprint requests to Dr. Bissell: Department of Clinical Pa-
thology, City of Hope National Medical Center, 1500 East Duarte Road, (14.0 mg) and sodium fluoride (17.5 mg) (Becton Dick-
Duarte, California 91010-0269. inson Vacutainer Systems, Rutherford, NJ). Plasma ob-

247
248 CLINICAL CHEMISTRY
Original Article

tained from the specimen was analyzed for glucose in the indicated that results for groups 1 and 2 were similar, as
General Clinical Chemistry Laboratory of the University were those for groups 3 and 4; therefore, data were com-
of Chicago Hospitals on the Ektachem 700® analyzer bined into two groups, 0-4.0 hours and >4.0 hours since
(Eastman Kodak Co., Rochester, NY), which uses a glu- the last meal, and two-tailed, paired and two-sample t-
cose oxidase-based reflection densitometric method. tests were performed to assess group differences. Finally,
Within 4 minutes after venipuncture, the patient's finger a multiple regression analysis was performed to assess the
was punctured with a spring-loaded lancet to obtain a independent contributions of (1) the Ektachem value, and
drop of capillary whole blood. This was analyzed for glu- (2) the time since the last meal, on meter error. Data are
cose with either the Glucoscan 2000® with Glucoscan test reported as mean ± 1 SD. A P value of <0.05 was con-
strips (Lifescan, Inc., Mountain View, CA) or the Accu- sidered significant, except where multiple testing was per-
Check II® with Chemstrip bG test strips (Boehringer formed, in which case the critical P value was adjusted
Mannheim Diagnostics, Indianapolis, IN). All determi- by the Bonferroni method (both uncorrected and Bon-
nations were performed by a single nurse who was trained ferroni P values are presented). All calculations were per-
in the operation and quality control of both instruments formed on a microcomputer using SYSTAT software.
as directed by the manufacturers. One instrument of each Precision evaluation experiments were performed on
type was used for the entire study. both the Accu-Check II and the Ektachem according to
The precision of the Ektachem 700 was assessed by the National Committee for Clinical Laboratory Stan-
performing 20 repeated glucose determinations on a single dards (NCCLS) Guidelines EP5T. Briefly, manufacturers'
sample over 20 consecutive minutes and by analyzing the supplied glucose control material was analyzed in dupli-
same sample once daily on 20 consecutive days. Precision cate, in two runs per day, for 20 days. Two analyte con-
of each portable meter was evaluated by daily determi- centrations were tested for each method. This data was
nations of the glucose concentration of a standard test analyzed to provide an estimate of within-run and total
solution (provided by the manufacturer) for 20 days. In imprecision. For the Accu-Check II, the determinations
addition, the precision of the entire portable glucose meter were performed by the same trained nurse who performed
methodology was assessed for each instrument by per- the patient determinations. The Ektachem study was per-
forming 20 repeated capillary blood glucose determina- formed by trained medical technologists in the General
tions on a single subject over a 30-60 minute period. Clinical Chemistry Laboratory.
The statistical approach used to assess the agreement The Glucoscan 2000 could not be included in this phase
between each meter and the laboratory reference method of the study because of the nature of its control materials.
is a modification of that suggested by Bland and Altman.14 These were packaged in small plastic vials and, according
First, least squares linear regression analysis was used to to the manufacturer's instructions, are to be used within
assess correlations between meter and Ektachem deter- 24 hours of opening. This would necessitate the use of 20
minations. Next, two-tailed paired /-tests were performed different vials of control material for each analyte con-
to evaluate the difference between meter and Ektachem centration to complete the study and would introduce
determinations (referred to here as "meter bias" or "er- another source of variability into the data.
ror") and to compare the magnitude of these two differ-
ences for the two meters. In addition, 95% confidence RESULTS
intervals (CI) were calculated for the mean bias for each
meter. A total of 277 determinations were made with the Glu-
coscan 2000, and 232 were performed on the Accu-Check
To determine whether meter bias was dependent on
II. There was a highly significant correlation between cap-
plasma glucose level, we performed a linear regression of
illary (meter) and venous (Ektachem) glucose values for
the difference between meter and Ektachem measure-
both instruments:
ments on the Ektachem value. To further clarify this re-
lationship, data were stratified in 1 g/L increments ac- Glucoscan: meter value = 0.84 X (Ektachem value)
cording to the Ektachem value, and an analysis of variance + 37.1, r = 0.94
was performed, followed by two-tailed two sample Mests Accu-Check: meter value = 0.92 X (Ektachem value)
to assess differences between the meters at various levels + 14.9, r = 0.94
of venous plasma glucose.
To analyze the impact of prior meal ingestion on mon- Despite these results, further evaluation disclosed sig-
itor accuracy, meter and Ektachem values were grouped nificant differences in performance of the two meters.
by the length of time that had passed since the last meal. Capillary glucose values measured with the Glucoscan
Preliminary analysis separating data into four groups (0- averaged 1.91 ± 0.84 g/L (range = 0.29-4.50 g/L; n
2.0 hours, 2.5-4.0 hours, 4.5-6.0 hours, and >6.0 hours) = 277), whereas venous plasma glucose values on the same

A.J.C.P. • February 1991


VALLERA, BISSELL, AND BARRON 249
Accuracy of Portable Blood Glucose Monitoring

subjects measured with the Ektachem averaged 1.85 Ektachem = -0.11 X (Ektachem) + 28.6, r = 0.477, P
± 0.96 g/L (range = 0.24-5.16 g/L), a highly significant < 0.001]. This effect was much less evident for the Accu-
difference (P = 0.001). The average absolute difference Check II, and was significant only for the error measured
between the Glucoscan and Ektachem was 0.065 ± 0.326 in absolute terms [difference (g/L) = -0.08 X (Ektachem)
g/L (95% confidence interval = 0.026-0.104), and the + 15.6, r = 0.229, P< 0.001; % difference = -0.02 X (Ek-
average percentage difference was 8.8 ±21.6% (95% con- tachem) + 4.8, r = 0.09, P = 0.18].
fidence interval = 6.2-11.3). The observation that neither This effect is better appreciated when the data are
of these 95% confidence intervals includes 0 supports the grouped in 1-g/L intervals of the Ektachem value (Table
given result (i.e., the mean Glucoscan 2000 and Ektachem 1). The average bias of the Glucoscan is more than 20%
values are significantly different from one another). (95% confidence interval = 19-28) when the Ektachem
Mean capillary glucose values determined with the value is less than 1 g/L, becomes minimal in the range
Accu-Check II were 1.95 ± 0.99 g/L (range = 0.22-5.00 2-3 g/L, and is more than - 1 0 % when the venous plasma
g/L; n = 232), and paired venous samples measured on level is >4 g/L. In contrast, the error of the Accu-Check
the Ektachem averaged 1.94 ± 1.01 g/L (range = 0.24- at the lowest glucose values averaged - 6 ± 15% (95%
6.05 g/L), a difference that was not significant (P = 0.93). confidence interval = - 2 — 11; P < 0.005 vs. Glucoscan
The mean difference between the Accu-Check and Ek- error, Bonferroni critical P = 0.01) and varied from —6.3%
tachem was 0.002 ± 0.35 g/L (95% confidence interval to +7.8% in other ranges of venous glucose (= not sig-
= -0.017-+0.39), data which confirm the lack of differ- nificant vs. Glucoscan for each of these ranges). Thus, the
ence between Accu-Check and Ektachem determinations. Accu-Check performed better (i.e., had less bias) than the
For the Glucoscan, 79.8% of determinations were Glucoscan primarily when the venous plasma glucose was
within 20% of the correct value, whereas 87.9% of the <lg/L.
Accu-Check measurements were in this range. Fourteen It also is of interest to note that all meter determinations
Glucoscan (5.0%) and five Accu-Check (2.2%) values de- with a difference from the Ektachem value of greater than
viated more than 50% from that determined on the Ek- 90% occurred in the most commonly encountered venous
tachem. Direct comparison of the average bias of the two plasma glucose range (1-2 g/L) for both the Glucoscan
meters indicated a highly significant difference when (n = 4) and Accu-Check (n = 3). These seven determi-
measured as a percentage of the Ektachem value. (Glu- nations were made on four different subjects; on two of
coscan: 8.8 ± 21.6%; Accu-Check: 1.1 ± 21.8%, P these, other meter measurements were within 20% of Ek-
< 0.001), and a trend in the same direction when the bias tachem values. When these seven outliers were excluded,
was expressed in g/L (Glucoscan: 0.065 ± 0.326 g/L; Accu- the mean measurement error in the 1-2 g/L range was 8
Check: 0.002 ± 0.351 g/L, P = 0.038, Bonferroni critical ± 13% (95% confidence interval = 6-11) for the Glucoscan
value: P = 0.025). and 3 ± 16% (95% confidence interval = -0.3-+7.0) for
Next, we analyzed the effect of the venous plasma glu- the Accu-Check (P = 0.023, Bonferroni critical P = 0.01).
cose level (Ektachem measurement) on the accuracy of Assessment of the effect of prior meal ingestion on
each meter. For the Glucoscan there was a significant lin- measurement bias indicated further differences in meter
ear correlation between the meter error expressed in either performance (Table 2). The mean error of Glucoscan de-
absolute [(Glucoscan value) - (Ektachem value) = -0.17 terminations was significantly greater and biased in a pos-
X (Ektachem value) + 37.1, r = 0.486, P < 0.001] or itive direction when the measurement was performed
percentage terms [% difference between Glucoscan and within 4 hours of eating, whereas recent food intake did

TABLE 1. METER BIAS IN ABSOLUTE (mg/dl) AND PERCENTAGE TERMS AS A FUNCTION


OF VENOUS PLASMA GLUCOSE LEVEL

Venous
Plasma Glucoscan 2000 Bias Accu-Check II Bias
Glucose
(q/L) 3'L % H 9IL % n

<1.0 0.18 ±0.13 23 ± 17 55 -0.05 ±0.12 - 6 ± 15 44


1.0-1.9 0.16 + 0.34* 12 ± 2 5 * 125 0.11 ±3.9| 8±30f 90
2.0-2.9 -0.05 ± 0.28 -2+11 63 0.01 ± 0.26 1 ± 11 61
3.0-3.9 -0.11 ±0.25 -3 ±8 23 -0.23 ± 0.38 6+10 28
^4.0 -0.52 ± 0.29 -11 ± 6 11 -0.12 ±0.67 1 ± 13 9

• When four outliers {% difference from Ektachem >90%) were excluded, differences were 0.11 t When three outliers were excluded, differences were 0.05 ± 0.25 g/L and 3 ± 16%.
± 0.20 g/L and 8 ± 13%. Results are expressed as mean ± SD; n = number of paired determinations.

Vol. 95 • No. 2
250 CLINICAL CHEMISTRY
Original Article
TABLE 2. EFFECT OF PRIOR MEAL INGESTION GLUCOSCAN 2000
ON METER BIAS

Time Since Last Meal

Instrument Units 0-4 hours n >4 hours n

Glucoscan g/L 0.102 ± 0.34 171 0.005 ± 0.28* 106


% 11.4 ±23.6 171 4.5 ± 16.9f 106
Accu-Check g/L
%
0.016
2.1
± 0.38* 149 -0.024 ±
±23.4§ 149 -0.8 ±
0.28
18.6
83
83 I
* P < 0.02 for the comparison with the 0-4 hour time period.
t P < 0.001 for the comparison with the 0-4 hour time period.
X P = 0.038 for the comparison with Glucoscan.
§ P < 0.001 for the comparison with Glucoscan.
Bonferroni critical P = 0.0125.
Results are expressed as mean ± SD; n = number of paired determinations. -60 -50 -40 -30 -20 -10 10 20 30 40

DIFFERENCE (X)

not appear to have a significant effect on the Accu-Check ACCUCHECK II


meter error. Furthermore, the Accu-Check performed
significantly better (i.e., less bias) compared with the Glu-
coscan only for measurements made less than 4 hours
after the last meal.
Finally, multiple regression analysis was performed to
determine whether the level of venous plasma glucose
and the time since the last meal had independent effects
on meter bias. Regression coefficients (Table 3) demon-
strate that the bias of the Glucoscan, in both g/L and
percent, is independently influenced by both plasma glu-
cose and prandial state. In contrast, the bias of the Accu-
Check, as a percentage of the Ektachem value, was de-
pendent upon neither of these variables, although the bias -60 -50 -40 -30 -20 -10 10 20 30 40 50 60

in g/L was significantly influenced, albeit to a smaller de- B DIFFERENCE (%)


gree than for the Glucoscan, by the venous plasma glucose. FIG. 1. A (upper). Percent differences between EKTACHEM and
For the Accu-Check II meter, the overall mean of all GLUCOSCAN 2000 values grouped according to the level of venous
determinations using the high glucose control material glucose determined by the EKTACHEM. Note that the ambient level
of venous glucose has a significant effect on meter bias and that when
was 3.093 g/L. The within-run precision was 0.177 g/L, venous glucose is less than 1 g/L, bias is approximately +20%. B (lower).
whereas the between-run precision was 0.268 g/L (coef- Percent differences between EKTACHEM and ACCUCHECK II values
ficient of variation = 8.7%). With the low concentration grouped according to the level of venous glucose determined by the Ek-
tachem. There is no significant relationship between meter bias and ve-
glucose control, the overall mean was 0.409 g/L, with the nous plasma glucose levels.
within-run precision being 0.0185 g/L and the between-
run precision being 0.0236 g/L (coefficient of variation
= 6.4%). With the Ektachem, using the high concentration glu-
cose control, the overall mean was 4.28 g/L, with the
within-run precision being 0.0172 g/L, between-run pre-
TABLE 3. COEFFICIENT OF MULTIPLE REGRESSION cision being 4.08 g/L (coefficient of variation = 0.95%).
ANALYSIS EVALUATING THE DEPENDENCE OF
METER BIAS ON PLASMA VENOUS GLUCOSE For the low glucose control, the overall mean was 0.827
AND TIME SINCE LAST MEAL g/L, with within-run precision being 0.0052 g/L and be-
tween-run precision being 0.0121 g/L (coefficient of vari-
Glucose
Instrument Units Level Time Interval
ation = 5.5%).
In the whole method precision study, both the Accu-
Glucoscan g/L -0.17 ±0.02 <0.001 -8.94 ±3.50 0.011 Check II and Glucoscan 2000 were tested. Using the Glu-
% -0.11 ±0.01 <0.001 -6.47 ±2.32 0.006
Accu-Check g/L -0.08 ± 0.02 <0.001 -4.87 ± 4.69 0.30 coscan 2000, the first volunteer's mean blood glucose
% -0.02 + 0.01 0.16 -3.12 ±2.98 0.296 concentration was 0.94 g/L, with SD of 0.0615 g/L (coef-
ficient of variation = 6.5%). The second volunteer was
Results are expressed as mean ± SE. "Time period since last meal" is categorized into "0-4
hours" or ">4 hours." tested with the Accu-Check II, yielding a mean blood glu-

A.J.C.P. • February 1991


VALLERA, BISSELL, AND BARRON 251
Accuracy of Portable Bh <d Glucose Monitoring

cose concentration of 0.735 g/L, with SD of 0.0405 g/L ent study allow one to be 95% certain that a given glu-
and coefficient of variation of 5.5%. Repetitive sampling coscan measurement was between - 3 4 % and +52% of
(n = 20) of the Ektachem high glucose control resulted the laboratory reference value and that an Accu-Check
in a mean concentration of 4.398 g/L, with SD of 0.0405 determination was between —43% and +45% of this value.
g/L and coefficient of variation of 0.53%. For the low When viewed in this light, the two meters do not appear,
control, the mean concentration was 0.817 g/L, with SD on average, to differ as much as suggested by comparison
of 0.0047 g/L and coefficient of variation of 0.58%. of the mean bias. Furthermore, the range of these limits
make it clear that portable blood glucose meter results
DISCUSSION must be interpreted with caution, particularly in those
ranges where a change in therapy may have substantial
This study critically evaluated the performance of two adverse effects.21
popular blood glucose meters in a busy clinical setting Overall estimates of glucose meter bias as described
and attempted to identify variables that might account may be misleading if this bias is significantly dependent
for any observed bias. As reported by others, 815 we found on the level of venous plasma glucose (e.g., the bias may
that both meters' readings were highly correlated (r = 0.94) be large at one glucose level and minimal at another).
with the laboratory reference method; however, additional Indeed, we found a significant linear relationship between
analysis disclosed that, on average, the Accu-Check II Glucoscan 2000 meter error and venous plasma glucose,
provided more accurate determinations, with a mean bias with highest errors at the lowest levels of plasma glucose,
of only 1. i % compared with 8.8% for the Glucoscan 2000. a phenomenon not observed for the Accu-Check II. Two
Yet more detailed evaluation disclosed that the greater previous studies of earlier models of the Accu-Check meter
bias of the latter instrument occurred primarily at lower provided similar results.6,20 When we stratified venous
levels of plasma glucose and when the patient had eaten glucose levels (Table 1), it was discovered that the major
within several hours of capillary glucose measurement. difference between the two meters occurred at concentra-
There have been numerous published reports that pur- tions of <1 g/L, at which level the mean bias of the Glu-
portedly have investigated the accuracy of portable blood coscan was +23% (95% limits of agreement: +6-+40%),
glucose monitors.5"8,10,11,15"20 However, as discussed in whereas the average error for the Accu-Check II was —6%
detail by Bland and Altman,14 many investigators inap- (95% limits of agreement: -21-+9%).
propriately report only the correlation coefficient between These observations may have important clinical im-
measurements made by meter and some reference meth- plications. For example, when rigid metabolic control is
odology.5"12 Such a statistic fails to provide clinically rel- being attempted, such as is standard during diabetic
evant information {i.e., how much does the meter deter- pregnancy10,22 (for which desired fasting and postprandial
mination deviate from the true [laboratory reference] glucose levels are <1 and <1.2 g/L, respectively), Glu-
value, and does this deviation change at various levels of coscan 2000 results may lead to excessive insulin admin-
venous plasma glucose). istration. On the other hand, use of Accu-Check II would
Godine and colleagues15 evaluated the accuracy of the be somewhat more likely to lead to unacceptably high
Accu-Check meter when used by staff nurses and found blood glucose levels. Such conclusions highlight the need
a mean error of 8 ± 7%, whereas Landon and colleagues,20 to understand glucose meter performance at levels of
studying the same meter, observed a mean deviation of blood glucose that are likely to lead to changes in therapy.
approximately 15%. Brooks and co-workers'8 found a Another factor that might affect meter accuracy is the
mean bias of—0.06 ± 0.30 g/L for the Accu-Check II and proximity of glucose measurement to the patient's last
0.08 ± 0.27 g/L for the Glucoscan 2000 (percentage data meal because significant forearm glucose uptake occurs
not available). Results of average bias for the two meters in the postprandial state.13 This could lead to capillary
in the present study were in the ranges cited, with a mean (meter) glucose levels that are significantly higher than
error for the Accu-Check II of 1 ± 22% (0 ± 0.35 g/L) venous (Ektachem) if food had been consumed shortly
and 9 ± 22% (0.06 ± 0.33 g/L) for the Glucoscan 2000. before measurement. 23 We are aware of only one study
Although the mean measurement error (as a % of the that addressed the impact of this phenomenon on the
venous glucose level) of both glucose meters was relatively apparent bias of portable meter capillary glucose deter-
small (<10%; 95% confidence interval for mean bias: minations. 12 In the fasting state, capillary determinations
Glucoscan: 6.2-11.3%; Accu-Check II: -1.7-4-3.9%), this exceeded venous glucose levels by 14%; however, 1 and
does not provide an estimate of a given capillary glucose 2 hours after a 100-g glucose load, capillary measurements
determination, a parameter of considerable clinical im- were, respectively, 20% and 19% higher than venous de-
portance. The latter is better estimated by the 95% limits terminations. Surprisingly, we found a similar effect only
of agreement (i.e., mean error ± 2 SD), which in the pres- with the Glucoscan 2000 and not with the Accu-Check

Vol. 95 • No. 2
252 CLINICAL CHEMISTRY
Original Article

II (Table 2). Furthermore, multiple regression analysis for self-monitoring of serum glucose levels. Can Med Assoc J
indicated that this was not simply due to the positive bias 1984;130:38-43.
9. Drucker RF, Williams DRR, Price CP. Quality assessment of blood
of the Glucoscan 2000 at certain levels of venous glucose. glucose monitors in use outside the hospital laboratory. J Clin
Thus, we have no explanation for this observation, but Pathol 1983;36:948-953.
the data indicate that a greater allowance for error must 10. Marquette GP, Dillard T, Bietla S, Niebyl JR. The accuracy of visual
and meter determinations of blood glucose with the use of Chem-
be taken into account if the Glucoscan 2000 is used to strip bG. Am J Obstet Gynecol 1985;153:883-884.
measure capillary glucose in the postprandial state. 11. Shapiro B, Savage PJ, Lomatch D, et al. A comparison of accuracy
and estimated cost of methods for home blood glucose monitoring.
In considering all of these statistics, one must balance
Diabetes Care 1981;4:396-403.
factors such as availability, turnaround time, convenience, 12. Weiner CP, Faustich M, Burns J, Fraser L, Whitaker L, KJugman
and portability against the difference in accuracy and pre- M. The relationship between capillary and venous glucose con-
centration during pregnancy. Am J Obstet Gynecol 1986; 155:
cision in various testing systems. These factors may be- 61-64.
come more or less important in different clinical settings. 13. Jackson RA, Peters N, Advani U, et al. Forearm glucose uptake
For example, accuracy and precision may be more im- during the oral glucose tolerance test in normal subjects. Diabetes
portant in measurements used to adjust insulin doses in 1973;22:442-458.
14. Bland JM, Altman DG. Statistical methods for assessing agreement
a "rigidly" controlled diabetic, whereas ease of operation between two methods of clinical measurement. Lancet 1986; 1:
and availability may have increased importance in less 307-310.
tightly controlled patients. 15. Godine JE, Hurxthal K, Nathan DM. Bedside capillary glucose
measurement by staff nurses in a general hospital. Am J Med
1986;80:803-806.
16. Aziz S, Hsiang YH. Comparative study of home blood glucose mon-
itoring devices: Visidex, Chemstrip bG, Glucometer, and Accu-
REFERENCES Check bG. Diabetes Care 1983;6:529-532.
17. Barr B, Leichter SB, Taylor L. Bedside capillary glucose monitoring
1. American Diabetes Association. Bedside blood glucose monitoring in the general hospital. Diabetes Care 1984;7:261-264.
in hospitals. Diabetes Care 1986;9:89. 18. Brooks KE, Rawal N, Henderson AR. Laboratory assessment of
2. Belsey R, Morrison JI, Whitlow KJ, Baer DM, Nelson S, Hardwick three new monitors of blood glucose: Accu-Check II, Glucometer
DF. Managing bedside glucose testing in the hospital. JAMA II, and Glucoscan 2000. Clin Chem 1986;32:2195-2200.
1987;258:1634-1637. 19. Clements RS, Keane NA, Kirk KA, Boshell BR. Comparison of
3. Rasaiah B. Self-monitoring of the blood glucose level: potential various methods for rapid glucose estimation. Diabetes Care
sources of inaccuracy. Can Med Assoc J 1985;132:1357-1361. 1981;4:392-395.
4. Villeneuve ME, Murphy J, Mazze RS. Evaluating blood glucose 20. Landon MB, Cembrowski GS, Gabbe SG. Capillary blood glucose
monitors. Am J Nurs 1985;11:1258-1259. screening for gestational diabetes: a preliminary investigation. Am
5. Gifford-Jorgensen RA, Borchert J, Hassanein L, Tilzer GA, Eaks J Obstet Gynecol 1986;155:717-721.
GA, Moore WV. Comparison offiveglucose meters for self-mon- 21. Cohen FE, Sater B, Feingold KR. Potential danger of extending
itoring of blood glucose by diabetic patients. Diabetes Care 1986;9: SMBG techniques to hospital wards. Diabetes Care 1986;9:320-
70-76. 322:
6. James K, Latham D, Marrero J, Barrett DA. Capillary blood glucose 22. Langer O, Mazze RS. Diabetes in pregnancy: evaluating self-mon-
measured by the laboratory phlebotomy team. Laboratory Med- itoring performance and glycemic control with memory-based
icine 1985;16:417-421. reflectance meters. Am J Obstet Gynecol 1986;155:635-637.
7. Webb DJ, Lovesay JM, Ellis A, Knight AH. Blood glucose monitors: 23. Eriksson KF, Fex G, Trell E. Capillary-venous differences in blood
a laboratory and patient assessment. Br Med J 1980;1:362-364. glucose values during oral glucose tolerance test. Clin Chem
8. Chiasson JL, Morrisset R, Hamet P. Precision and costs of techniques 1983;29:993.

A.J.C.P. • February 1991

Anda mungkin juga menyukai