Please cite this article in press Devikasubramaniyan et al., Method Development and Validation for Estimation of
Dosulepin in Pure and Dosage Form by Using HPLC, Indo Am. J. P. Sci, 2018; 05(04).
INTRODUCTION:
Instrumentation of HPLC are usually constructed of glass or stainless steel
The basic liquid chromatograph consists of six basic materials which are chemically resistant to mobile
units. The mobile phase supply system, the pump and phase.
programmer, the sample valve, the column, the
detector and finally a means of presenting and Mobile phase
processing the results [1-6]. Mobile phases in HPLC are usually mixtures of two
or more individual solvents. The usual approach is to
Mobile phase (solvent) reservoirs and solvent choose what appears to be the most appropriate
degassing column, and then to design a mobile phase that will
The mobile phase supply system consists of number optimize the retention and selectivity of the system.
of reservoirs (200 mL to 1,000 mL in capacity). They The two most critical parameters for nonionic mobile
phases are strength and selectivity [7-9].
detection, Limit of Quantitation, Robustness, System The standard deviation SD is given by,
suitability and Stability criteria.
N
1
Accuracy √ ∑(xi − x)2
The accuracy of measurement is defined as the N−1
i=1
closeness of the measured value to the true value. In a
method with high accuracy, a sample (whose “true Intermediate precision:
value” is known) is analyzed and the measured value It is the agreement of complete measurements
should ideally be identical to the true value.Typically, (including standards) when the same method is
accuracy is represented and determined by recovery applied many times within the same laboratory.
studies but there are three ways to determine Intermediate precision expresses within-laboratories
accuracy: variations: different days, different analysts, different
1. Comparison to a reference standard equipment etc.
2. Recovery of the analyte spiked into blank Reproducibility:
matrix or Reproducibility expresses the precision between
3. Standard addition of the analyte. laboratories and is often determined in collaborative
studies or method transfer experiments.
The ICH documents recommended that accuracy
should be assessed using a minimum of nine The precision of an analytical procedure is usually
determinations over a minimum of three expressed as the variance, standard deviation or
concentrations levels the specified range (i.e., three coefficient of variation of a series of measurements. 22
concentrations and three replicates of each Acceptance criteria: The % RSD should be less than
concentration). Accuracy was tested (%Recovery and 2.
%RSD of individual measurements) by analyzing Linearity:
samples at least in triplicate, at each level (80%, The linearity of an analytical procedure is its ability
100% and 120% of label claim) is recommended. For (within a given range) to obtain test results, which are
each determination fresh samples were prepared and directly proportional to the concentration (amount) of
assay value is calculated [4-8]. analyte in the sample. It is a measure of how well a
Acceptance criteria: The accuracy should be within calibration plot of response vs. concentration
98-102%. approximates a straight line. Linearity can be
Precision assessed by performing single measurements at
Precision can be defined as “the degree of agreement several analyte concentrations. The data are then
among individual test results when the procedure is processed using a linear least squares regression. The
applied repeatedly to multiple samplings of a Resulting plot slope, intercept and correlation
homogeneous sample”. A more comprehensive coefficient provide the desired information on
definition proposed by the International Conference linearity [9-15].
on Harmonization (ICH) divides precision into three Acceptance criteria: Correlation coefficient
types: (R2)>0.998.
1. Repeatability Specificity
2. Intermediate precision Specificity is the ability to assess unequivocally the
3. Reproducibility analyte in the presence of components which may be
expected to be present; these include impurities,
Repeatability:
Repeatability expresses the precision under the same degradants and matrix etc. Lack of specificity of an
operating conditions over a short interval of time. individual analytical procedure may be compensated
Repeatability is also termed intra-assay precision and by other supporting analytical procedures. This
involves multiple measurements of the same sample definition has the following implications:
(different preparations) by the same analyst under the Identification: To ensure the identity of an analyte.
same conditions.At least Purity Tests: To ensure that all the analytical
1. 5 or 6 determinations procedures performed allow an accurate statement of
2. At two or three different concentrations the content of impurities of an analyte, i.e. related
Should be done and the relative standard substances test, heavy metals, residual solvents
content etc.
deviations were calculated. The % RSD can be
calculated by, Assay (content or potency): To provide an exact
𝐒𝐃 × 𝟏𝟎𝟎 % result which allows an accurate statement on the
% 𝐑𝐒𝐃 = content or potency of the analyte in a sample [16-20].
𝐌𝐄𝐀𝐍 Acceptance criteria: No interference should be
Where, RSD = relative standard deviation
SD = standard deviation present.
VALIDATION
PREPARATION OF MOBILE PHASE: Diluent Preparation:
Preparation of mobile phase: The Mobile phase was used as the diluent.
Accurately measured 1000ml of HPLC Water Linearity
(100%) were mixed and degassed in a digital The linearity of was obtained in the concentration
ultrasonicater for 10 minutes and then filtered ranges from 5-25 g/ml
through 0.45 µ filter under vacuum filtration.
Table 1: Linearity data of Dosulepin
Concentration
Concentration Level (%)
g/ml
60 5
80 10
100 15
120 20
140 25
700000
600000
Dosulepin
500000
Area(AU)
400000
300000 y = 24149x - 216.43
200000 R² = 0.9982
100000
0
0 5 10 15 20 25 30
Concentartion(ppm)
Accuracy
Dosulepin tablets content were taken at various concentrations ranging from 50 % to 150 % (50 %, 75 %, 100 %,
125 %, and 150 %) to accurately quantify and to validate the accuracy. The assay was performed in triplicate. The
results were shown in Table-3
Table 3.The accuracy results for Dosulepin
15. Draft ICH Guidelines on Validation of 20. Skoog D A, West D M, Holler FJ: Introduction of
Analytical Procedures Definitions and analytical chemistry. Sounder college of
terminology. Federal Register, vol 60. IFPMA, publishing, Harcourt Brace college publishers.
Switzerland, (1995), PP 1126. (1994), PP 1-5.
16. Code Q2B, Validation of Analytical Procedures; 21. Sharma B K, Instrumental method of chemical
Methodology. ICH Harmonized Tripartite analysis Meerut. (1999), PP 175-203.
Guidelines, Geneva, Switzerland, (1996), PP 1- 22. Breaux J and Jones K: Understanding and
8. implementing efficient analytical method
17. Introduction to analytical method validation development and validation. Journal of
(online), available from: Pharmaceutical Technology (2003), 5, PP 110-
URL: http://www.standardbase.hu/tech/HPLC% 114.
20validation%20PE.pdf. 23. Willard, H. y. Merritt L.L, Dean J.A and Settle
18. Data elements required for assay validation, F.A “Instrumental methods of analysis” 7th
(online) available from: edition CBS publisher and distributors, New
URL: http://www.labcompliance.com/tutorial/m Delhi, (1991), PP 436-439.
ethods/default.aspx. 24. ICH Q2A, “validation of analytical methods,
19. Snyder LR practical HPLC method definitions and terminology”, ICH Harmonized
development, 2nd edition. John Wiley and sons, tripartite guideline, (1999).
New York, (1997), PP 180-182.