DYNATRIAX
Cyclic Triaxial Software
MANUALE DI ISTRUZIONI
INSTRUCTION MANUAL
Manuale di Istruzioni DYNATRIAX Software
DYNATRIAX Software Instruction Manual
CONTENTS
This instruction manual is an integral part of the machine and should be read before using the machine
and be safely kept for future reference.
CONTROLS reserves all rights of this manual; no part or whole can be copied without the written
permission of CONTROLS.
The proper use of this machine must be strictly adhered to, any other use must be considered as
incorrect.
The manufacturer cannot be held responsible for damage caused by incorrect use of the machine.
The machine must not be tampered with for any reason. In case of tampering, the manufacturer
declines any responsibility of functioning and safety of the machine.
CONTROLS reserve the right to update its manuals without notification in order to correct possible
typing errors, mistakes, updating of information and/or updating of programs and/or accessories.
Such changes will be inserted in the latest edition of the current manual.
1.0 INTRODUCTION
From the start to the end of a test, the software has total control of the
compression frame, the pressure systems, the actuator, and the volume change
device, allowing a test to be completed with minimal intervention from the
operator.
The software displays live transducer readings, graphical test data and
calculated specimen dimensions throughout the test.
Current or historic test data can also be easily reviewed and imported into a
spreadsheet program (e.g. Microsoft Excel) for processing.
The software also contains functions for managing the DYNATRIAX system,
allowing the user to calibrate transducers, set system limits and tune the control
channels.
This icon indicates a NOTE; please read the items marked by this icon
carefully.
This icon indicates a WARNING message; the items marked by this icon
refer to the safety aspects of the operator and / or the service engineer.
2.0 DESCRIPTION
2.1 MAIN FEATURES
The following stages of a cyclic triaxial test can be performed::
Saturation by increments of cell pressure with B value checks
Saturation by back pressure increments with volume change measurement
Saturation by a drained ramp to a pre-defined cell pressure and back
pressure, with volume change measurement
Isotropic consolidation with automatic management of the volume change
device to ensure continuous measurement of volume change
Undrained or drained load controlled ramp to pre-defined values of stress
path parameters p and q or s and t
Undrained or drained load controlled ramp to pre-defined values of total
vertical and horizontal stresses
Determination of K0 under drained or undrained conditions by vertically
ramping at a pre-defined rate of strain, with specimen diameter control
Load controlled cyclic shear to ASTM D 5311 standard
Load or displacement controlled modulus and damping to ASTM D 3999
standard
User-defined load/stress or displacement/strain controlled cyclic shear
Undrained or drained monotonic shear in compression or extension
With the exception of cyclic stages, the entire testing sequence can be
performed automatically with no need for user intervention
Communication and control is over a high speed Ethernet link
A calibration facility is included for performing digital calibrations and
verification, and producing calibration certificates for the system’s transducers
Transducer limits can be set to prevent damage to the system by overloading
Historical data can be easily loaded in and reviewed
Measured and calculated test data for the cyclic stages is displayed, including
permanent and recoverable deformation and strain, and resilient modulus
The DYNATRIAX software is pre-installed on the system’s PC. If the software needs
to be re-installed for any reason, it should be done by following Steps 1 to 2 below:
Fig 3-1
2. Open Windows Explorer and open the folder on the CD called ‘TriaxSetup’.
Double-click the application called ‘setup.exe’ (see Fig 3-1).
3. Click the ‘Next >>’ button on the next two windows that are shown, wait for the
installer to complete and then click the ‘Finish’ button on the final screen.
Fig 3-2
Fig 3-3
NOTE:
Clicking the Unload test button before a test has finished will cause the
test to be aborted. You will be unable to restart or recover the test after the
unload command has been confirmed.
Test data up to the unload point will be saved in the location specified in the
main panel.
4.2.2 4.2.1
Fig 4-2
Displays a brief description of the conditions under which the stage is being
carried out.
4.4.20 Logo
Clicking on the logo opens an ‘About’ screen containing details of the
software and firmware version numbers. Click on the ‘About’ screen to close
it.
Fig 4-5
4.5.3.1 Saturation method control – for selecting the method that will be used for
the saturation.
Click on the arrow to the right of the cell to select from:
‘Cell pressure increment’ to saturate the specimen by increments of cell
pressure only;
‘Back pressure increment’ to saturate the specimen by increments of cell
pressure and back pressure;
4.5.3.2 Auto stage check box – for specifying whether or not the stage will be
performed automatically (i.e. without any user input).
Tick the box to let the software perform the stage automatically.
Leave the box un-ticked to perform the stage manually.
4.5.3.3 Cell pressure increment cell – for entering the amount, in kPa, that the
cell pressure will be increased by during a cell pressure increment.
This parameter is only needed for performing an automatic stage and so will
only be available if the Auto stage check box (Item 4.5.3.2) is ticked.
4.5.3.4 Differential pressure cell – for entering the difference, in kPa, that you
want between the cell pressure and the back pressure during a back
pressure increment.
This parameter is only needed for performing an automatic stage and so will
only be available if the Auto stage check box (Item 4.5.3.2) is ticked.
4.5.3.5 Minimum back pressure cell – for entering the minimum back pressure
that must be reached by the end of the saturation stage.
This parameter is only needed for performing an automatic stage and so will
only be available if the Auto stage check box (Item 4.5.3.2) is ticked.
4.5.3.6 Ramp time cell – for entering the time, in seconds, that you want the
system to take to ramp to the next pressure increment.
This parameter is only needed for performing an automatic stage and so will
only be available if the Auto stage check box (Item 4.5.3.2) is ticked.
4.5.3.7 Data logging method control – for selecting the logging method for the
saturation stage.
Click on the arrow to the right of the cell to select from ‘Linear (time)’ or
‘Logarithmic (time)’.
If ‘Linear (time)’ is selected then the saturation data will be logged at the
rate entered in the Linear time interval cell (Item 4.5.3.8).
If ‘Logarithmic (time)’ is selected then the saturation data will be logged at
the rate given by the formula (A x reading number B), where A and B are the
values entered in the Logarithmic factor cells (Item 4.5.3.9).
4.5.3.8 Linear time interval cell – for entering the time interval, in minutes, that
you want between each logged reading.
This parameter is only available when you have selected ‘Linear (time)’ for
the Data logging method control (Item 4.5.3.7).
4.5.3.9 Logarithmic factor cells – for entering the factors A and B that give your
required logging rate when used in the formula (A x reading number B).
These parameters are only available when you have selected ‘Logarithmic
(time)’ for the Data logging method control (Item 4.5.3.7).
4.5.4.1 Effective consolidation pressure cell – for entering the effective pressure
that the specimen will be consolidated to (i.e. the difference between the
cell pressure and the back pressure), in kPa.
In order to achieve the required effective consolidation pressure, an
increment of cell pressure will be added on to the back pressure.
This parameter is only needed for performing an automatic stage and so will
only be available if the Auto stage check box (Item 4.5.4.2) is ticked.
4.5.4.2 Auto stage check box – for specifying whether or not the stage will be
performed automatically (i.e. without any user input).
Tick the box to let the software perform the stage automatically.
Leave the box un-ticked to perform the stage manually.
4.5.4.3 Back pressure check box – for specifying whether or not the system will
use the back pressure from the previous stage for the consolidation.
Tick the box to use the value from the previous stage.
Leave the box un-ticked to allow the system to choose the back pressure –
the system will round the current pore pressure to the nearest 10kPa and
use this value for the back pressure.
This parameter is only needed for performing an automatic stage and so will
only be available if the Auto stage check box (Item 4.5.4.2) is ticked.
4.5.4.4 Target ramp time cell – for entering the time, in seconds, that you wants
the system to take to ramp to the consolidation pressures.
This parameter is only needed for performing an automatic stage and so will
only be available if the Auto stage check box (Item 4.5.4.2) is ticked.
4.5.4.5 Initial stage time interval cell – for entering the time interval, in seconds,
that you want between each logged reading of the initialisation stage.
The initialisation stage is the period before the start of consolidation when
the system is ramping to the target consolidation pressures.
4.5.4.6 Logarithmic factor cells - for entering the factors A and B that give your
required logging rate when used in the formula (A x reading number B).
4.5.4.5 4.5.4.6
Fig 4-7
4.5.5.1 K0 consolidation method control – indicates that the stage will be vertical
strain controlled.
4.5.5.2 Auto stage check box - for specifying whether or not the stage will be
performed automatically (i.e. without any user input).
Tick the box to let the software perform the stage automatically.
Leave the box un-ticked to perform the stage manually.
4.5.5.3 Rate of strain cell – for entering the rate of strain that the K0 will run at, in
percent per hour.
This parameter is only needed for performing an automatic stage and so will
only be available if the Auto stage check box (Item 4.5.5.2) is ticked.
4.5.5.4 Loading method control – for selecting whether the K0 will run in
compression or extension.
Click on the arrow to the right of the cell to select from ‘Compression’ or
‘Extension’.
This parameter is only needed for performing an automatic stage and so will
only be available if the Auto stage check box (Item 4.5.5.2) is ticked.
4.5.5.5 Drainage method control – for selecting whether the K0 stage will be
drained or undrained.
Click on the arrow to the right of the cell to select from ‘Drained’ or
‘Undrained’.
The K0 stage can only be undrained if a radial on-specimen transducer is
being used to measure specimen diameter.
This parameter is only needed for performing an automatic stage and so will
only be available if the Auto stage check box (Item 4.5.5.2) is ticked.
4.5.5.6 Diameter control method – for specifying the method that will be used to
control the specimen diameter during the K0 stage.
Click on the arrow to the right of the cell to select from:
‘Disp / Vol’ to monitor and control the diameter as calculated from the
change in volume and height;
‘Radial belt’ to monitor and control the diameter as measured by the radial
on-specimen transducer. (This option is only available if you have a radial
on-specimen transducer enabled in the System configuration panel, Item
4.16)
This parameter is only needed for performing an automatic stage and so will
only be available if the Auto stage check box (Item 4.5.5.2) is ticked.
4.5.5.7 Termination / hold conditions – for specifying the conditions that you
want the K0 to stop at.
For each termination condition that you want to be taken into consideration
during the K0 stage, the relevant box must be ticked and a value entered. If
more than one condition is specified, the one that is reached first will
determine when the stage stops.
The system will continue ramping until one of the termination conditions is
met. In a manual stage, when the first termination condition is reached the
system will stop ramping and hold all the stresses until you end the stage.
In an automatic stage, the system will end the stage and proceed to the
next stage when the first termination condition is met.
4.5.5.8 Diameter control setup cells – for specifying ranges of tolerances for
diameter change and the cell pressure responses to them.
The tolerances must be entered, in millimetres, in the ‘Range’ cells. Your
required cell pressure responses must be entered, in kPa, in the ‘Change’
cells.
The diameter control works in the following way during the K0 stage:
If the specimen diameter deviates by less than the value entered in ‘Range
4’ there will be no change in the cell pressure;
If the specimen diameter deviates by more than the value entered in ‘Range
4’ but less than the value entered in ‘Range 3’, the cell pressure will be
changed by the value in ‘Change 4’;
If it deviates by more than the value entered in ‘Range 3’ but less than the
value entered in ‘Range 2’, the cell pressure will be changed by the value in
‘Change 3’;
If it deviates by more than the value entered in ‘Range 2’ but less than the
value entered in ‘Range 1’, the cell pressure will be changed by the value in
‘Change 2’;
If it deviates by more than the value entered in ‘Range 1’, the cell pressure
will be changed by the value in ‘Change 1’.
The values must be entered in descending order of size, with the largest
values in ‘Range 1’ and ‘Change 1’ cells. The values are absolute - a
negative change in diameter will result in a positive change in cell pressure
and a positive change in diameter will result in a negative change in cell
pressure.
4.5.5.9 Data logging method control – for selecting the logging method for the K0
stage.
Click on the arrow to the right of the cell to select from ‘Linear (time)’ or
‘Linear (strain)’.
If ‘Linear (time)’ is selected then the K0 data will be logged at the rate
entered in the Linear time interval cell.
If ‘Linear (strain)’ is selected then the K0 data will be logged at the rate
entered in the Linear strain interval cell (Item 4.5.5.11).
4.5.5.10 Linear time interval cell – for entering the time interval, in minutes, that
you want between each logged reading.
This parameter is only available when you have selected ‘Linear (time)’ for
the Data logging method (Item 4.5.5.9).
4.5.5.11 Linear strain interval cell – for entering the strain interval, in percent, that
you want between each logged reading.
This parameter is only available when you have selected ‘Linear (strain)’ for
the Data logging method (Item 4.5.5.9).
4.5.6.1 Stress path method control – for selecting which stresses will be used as
the targets for the stress path and whether the stage will be stress or strain
controlled.
Click on the arrow to the right of the cell to select from:
‘p & q – stress control’ to ramp to target values of p (1/3 ( v+2 h)) and q
( v- h);
‘s & t – stress control’ to ramp to target values of s (½ ( v+ h)) and t (½ ( v-
h));
‘Total horiz. & vert. stress – stress control’ to ramp to target values of total
horizontal and vertical stress;
‘Total vert. stress – strain control’ to ramp the total vertical stress at a
specified rate of strain.
4.5.6.2 Auto stage check box - for specifying whether or not the stage will be
performed automatically (i.e. without any user input).
Tick the box to let the software perform the stage automatically.
Leave the box un-ticked to perform the stage manually.
4.5.6.3 Target abscissa cell – for entering the abscissa (x-axis) stress path point
that you want the system to reach by the end of the ramp.
The abscissa will be either s, p or h, depending on which stress path
method you have selected.
This parameter is only needed for performing an automatic stress controlled
stress path and so will only be available if the Auto stage check box (Item
4.5.6.2) is ticked and you have selected a stress control option for the
Stress path method control (Item 4.5.6.1).
4.5.6.4 Target ordinate cell – for entering the ordinate (y-axis) stress path point
that you want the system to reach by the end of the ramp.
The ordinate will be either t, q or v, depending on which stress path
method you have selected.
This parameter is only needed for performing an automatic stress controlled
stress path and so will only be available if the Auto stage check box (Item
4.5.6.2) is ticked and you have selected a stress control option for the
Stress path method control (Item 4.5.6.1).
4.5.6.5 Ramp time cell – for entering the time, in minutes, that you want the
system to take to ramp to the stress entered in the target ordinate cell and
the target abscissa cell.
This parameter is only needed for performing an automatic stress controlled
stress path and so will only be available if the Auto stage check box (Item
4.5.6.2) is ticked and you have selected a stress control option for the
Stress path method control (Item 4.5.6.1).
4.5.6.6 Rate of strain cell – for entering the rate of strain, in percent per hour, that
you want the stress path to run at.
Only a positive number may be entered – the strain direction is determined
by the loading method.
This parameter is only needed for performing an automatic strain controlled
stress path and so will only be available if the Auto stage check box (Item
4.5.6.2) is ticked and you have selected the strain control option for the
Stress path method control (Item 4.5.6.1).
4.5.6.7 Loading method control – for selecting whether the stress path will run in
compression or extension.
Click the arrow to the right of the cell to select from ‘Compression’ or
‘Extension’.
This parameter is only needed for performing an automatic strain controlled
stress path and so will only be available if the Auto stage check box (Item
4.5.6.2) is ticked and you have selected the strain control option for the
Stress path method control (Item 4.5.6.1).
4.5.6.8 Drainage method control – for selecting whether the stress path will be
drained or undrained.
Click the arrow to the right of the cell to select from ‘Drained’ or ‘Undrained’.
This parameter is only needed for performing an automatic stress path and
so will only be available if the Auto stage check box (Item 4.5.6.2) is
ticked.
4.5.6.9 Hold time cell – for entering the time, in minutes, that you want the system
to hold the target stress for, once they have been reached at the end of the
stage.
This parameter is only needed for performing an automatic stress path and
so will only be available if the Auto stage check box (Item 4.5.6.2) is
ticked.
4.5.6.10 Termination / hold conditions – for specifying the conditions that you
want the stress path stage to stop at.
For each termination condition that you want to be taken into consideration
during the stress path stage, the relevant box must be ticked and a value
entered. If more than one condition is specified, the one that is reached first
will determine when the stage stops.
If you have selected a stress control method option, the system will
continue ramping until either the targets entered in the Target abscissa
cell and the Target ordinate cell (Items 4.5.6.3 & 4.5.6.4) are reached or
until one of the termination conditions is met, whichever is reached first. If
you have selected the strain control method, the system will continue
ramping until one of the termination conditions is met.
When the first termination condition is reached the system will stop ramping
and hold all the stresses for the hold time specified (in an automatic stage)
or until you end the stage (in a manual stage).
4.5.6.11 Data logging method control – for selecting the logging method for the
stress path stage.
Click on the arrow to the right of the cell to select from ‘Linear (time)’ or
‘Linear (strain)’.
If ‘Linear (time)’ is selected then the stress path data will be logged at the
rate entered in the Linear time interval cell (Item 4.5.6.12).
If ‘Linear (strain)’ is selected then the stress path data will be logged at the
rate entered in the Linear strain interval cell (Item 4.5.6.13).
4.5.6.12 Linear time interval cell – for entering the time interval, in minutes, that
you want between each logged reading.
This parameter is only available when you have selected ‘Linear (time)’ for
the Data logging method control (Item 4.5.6.11).
4.5.6.13 Linear strain interval cell – for entering the strain interval, in percent, that
you want between each logged reading.
This parameter is only available when you have selected ‘Linear (strain)’ for
the Data logging method control (Item 4.5.6.11).
4.5.7.1 Drainage method control – for selecting whether the shear stage will be
drained or undrained.
Click the arrow to the right of the cell to select from ‘Drained’ or ‘Undrained’.
This parameter is only needed for performing an automatic stage and so will
only be available if the Auto stage check box (Item 4.5.7.2) is ticked.
4.5.7.2 Auto stage check box - for specifying whether or not the stage will be
performed automatically (i.e. without any user input).
Tick the box to let the software perform the stage automatically.
Leave the box un-ticked to perform the stage manually.
4.5.7.3 Loading method control – for selecting whether the shear stage will run in
compression or extension.
Click the arrow to the right of the cell to select from ‘Compression’ or
‘Extension’.
The stage can only be run in extension if ‘Actuator’ is selected for the
Displacement control (Item 4.5.7.7).
This parameter is only needed for performing an automatic stage and so will
only be available if the Auto stage check box (Item 4.5.7.2) is ticked.
4.5.7.4 Rate of strain cell – for entering the rate of strain, in percent per hour, that
you want the shear stage to run at.
Only a positive number may be entered – the strain direction is determined
by the loading method.
This parameter is only needed for performing an automatic stage and so will
only be available if the Auto stage check box (Item 4.5.7.2) is ticked.
4.5.7.5 Failure criteria control – for selecting the criterion at which the shear
stage will be considered to be finished.
Click on the arrow to the right of the cell to select from:
‘Maximum Effective Principle Stress Ratio’ to stop the shear stage when the
maximum effective principle stress ratio is reached (only applicable for an
undrained stage);
‘Constant Shear Stress / Pore Pressure’ to stop the shear stage when there
is no further significant change in both shear stress and pore pressure (only
applicable for an undrained stage);
‘Constant Volume / Shear Stress’ to stop the shear stage when there is no
further significant change in both shear stress and volume (only applicable if
for a drained stage);
‘Use Termination Strain only’ to stop the shear stage when the axial strain
reaches the value entered in the Termination strain cell (Item 4.5.7.6).
If the selected failure criterion is not reached during the shear stage, the
stage will be stopped when the termination strain is reached.
4.5.7.6 Termination strain cell – for entering the axial strain, in percent, at which
the shear stage should be stopped.
This value will only be used to stop the shearing if the selected failure
criterion has not been reached, or if ‘Use termination strain only’ has been
selected as the failure criterion.
4.5.7.7 Displacement control – for selecting whether the system uses the actuator
or the Tritech compression frame to shear the specimen.
Click the arrow to the right of the cell to select from choice of ‘Actuator’ or
‘Tritech’.
It is not possible to use Tritech control if you have chosen to do an
automatic stage – in an automatic shear stage, the displacement control is
fixed to ‘Actuator’.
It is only possible to shear the specimen in extension using the actuator.
The actuator has a maximum loading capacity of 5kN.
4.5.7.8 Data logging method control – for selecting the logging method for the
shear stage.
Click on the arrow to the right of the cell to select from ‘Linear (time)’,
‘Linear (strain)’ or ‘Logarithmic (time)’.
If ‘Linear (time)’ is selected then the shear data will be logged at the rate
entered in the Linear time interval cell (Item 4.5.7.9).
If ‘Linear (strain)’ is selected then the shear data will be logged at the rate
entered in the Linear strain interval cell (Item 4.5.7.11).
If ‘Logarithmic (time)’ is selected then the saturation data will be logged at
the rate given by the formula (A x reading number B), where A and B are the
values entered in the Logarithmic factor cells (Item 4.5.7.10).
4.5.7.9 Linear time interval cell – for entering the time interval, in minutes, that
you want between each logged reading.
This parameter is only available when you have selected ‘Linear (time)’ for
the Data logging method control (Item 4.5.7.8).
4.5.7.10 Logarithmic factor cells – for entering the factors A and B that give your
required logging rate when used in the formula (A x reading number B).
These parameters are only available when you have selected ‘Logarithmic
(time)’ for the Data logging method control (Item 4.5.7.8).
4.5.7.11 Linear strain interval cell – for entering the strain interval, in percent, that
you want between each logged reading.
This parameter is only available when you have selected ‘Linear (strain)’ for
the Data logging method control (Item 4.5.7.8).
4.5.8.1 Cyclic shear method control – for selecting the method for the cyclic
stage.
Click on the arrow to the right of the cell to select from:
‘ASTM D 5311 (Load Controlled Cyclic Strength Test)’ to cycle the
specimen in accordance with the ASTM standard D 5311, which specifies a
load controlled sinusoidal wave form;
‘ASTM D 3999 (Load Controlled Modulus and Damping Test)’ to cycle the
specimen in load control, in accordance with the ASTM standard D 3999,
which specifies a sinusoidal wave form with a maximum of 40 cycles. With
this method selected, the damping ratio, Young’s modulus and shear
modulus will be calculated and shown on the Cyclic Shear panel (see Item
4.12.10);
‘ASTM D 3999 (Displacement Controlled Modulus and Damping Test)’ to
cycle the specimen in displacement control, in accordance with the ASTM
standard D 3999, which specifies a sinusoidal wave form with a maximum
of 40 cycles. With this method selected, the damping ratio, Young’s
modulus and shear modulus will be calculated and shown on the Cyclic
Shear panel (see Item 4.12.10);
‘Non standard (multi cycle test)’ to cycle the specimen with user-defined
control modes, wave shapes and number of cycles.
‘Non standard (single cycle test)’ to cycle the specimen with user-defined
control modes and wave shapes, with the number of test cycles fixed to
one.
4.5.8.2 Control mode – for selecting whether the axial cycling will be under load
control or displacement control.
Click the button to the right of the cell to select from:
‘Stress’ to specify the extension and compression values for the cyclic
loading in deviator stress (kPa);
‘Strain’ to specify the extension and compression values for the cyclic
loading in axial strain (%);
‘Load’ to specify extension and compression values for the cyclic loading in
load (N);
‘Displacement’ to specify extension and compression values for the cyclic
loading in deformation (mm).
If you have selected ‘ASTM D 5311 (Load Controlled Cyclic Strength Test)’
or ‘ASTM D 3999 (Load Controlled Modulus and Damping Test)’ for the
Cyclic shear method control (Item 4.5.8.1), this parameter will be fixed as
‘Load’. If you have selected ‘ASTM D 3999 (Displacement Controlled
Modulus and Damping Test)’, it will be fixed as ‘Displacement’.
4.5.8.3 Drainage method control – for selecting whether the cyclic stage will be
drained or undrained.
Click the arrow to the right of the cell to select from ‘Drained’ or ‘Undrained’.
4.5.8.4 Reference stress cell – for entering the reference shear stress (if the
specimen is clay) or the effective consolidation stress (if the specimen is
sand).
This value is used to calculate the average and cyclic shear stress ratios on
the Cyclic Shear panel (Item 4.12.10).
The label for this cell depends on the soil type that you have chosen on the
Specimen Setup panel (Item 4.6.2).
4.5.8.5 Waveshape control – for selecting a wave shape for the cyclic loading.
Click the button to the right of the cell to select from ‘Sinusoidal’, ‘Triangle’,
‘Square’ or ‘User defined’.
(A user defined wave shape is selected using the User waveform cell
(Item 4.5.8.6.))
This function is only available if you have selected a ‘Non-standard’ option
for the Cyclic shear method control (Item 4.5.8.1).
4.5.8.6 User waveform cell – for loading in a pre-defined wave shape file.
A wave shape file is a .udw file, which is created using the program ‘UDW
Generator’ that accompanies this software.
Click the open folder icon to the left of the cell to locate your required .udw
file.
This function is only available if you have selected ‘User defined’ for the
Waveshape control (Item 4.5.8.5).
4.5.8.7 Test frequency cell - for entering the frequency (or number of complete
cycles per second) for the cyclic wave form, in hertz.
Frequency (Hz) = number of cycles / time (seconds).
Any value between 0.008Hz and 10Hz may be entered.
The time period for one cycle at the frequency that has been entered is
shown on the x-axis of the Cycle chart (Item 4.5.8.12).
4.5.8.8 Sample rate cell – for entering the number of readings that you want to be
logged for each cycle, unless you have selected a user-defined wave-
shape, in which case it is for entering the number of readings per second.
The maximum number that may be entered depends on the test frequency
and is indicated in the Maximum sample rate indicator (Item 4.5.8.9). If a
user-defined wave shape has been loaded, the maximum value is 2000
readings per second.
Every channel will be logged at the rate that you enter.
4.5.8.9 Maximum sample rate indicator – indicates the maximum value that can
be entered in the Sample rate cell (Item 4.5.8.8), the value is the maximum
number of readings that can be logged per cycle.
This parameter is not available if you have selected ‘User defined’ for the
Waveshape control (Item 4.5.8.5).
4.5.8.10 Axial amplitude cell – for entering the required amplitude of the axial
cycles – see Fig 4-12.
The units for this parameter depend on which Control mode (Item 4.5.8.2)
has been selected.
Fig 4-12
Here are two examples:
Example 1 - The previous stage was an isotropic consolidation, the load at
the start of the cyclic stage is 0N, you want to cycle +/- 100N. You should
enter 100 in the Amplitude cell. The peak to peak amplitude will be 200N
and the cycle will peak at -100N in extension to +100N in compression.
Example 2 - The previous stage was a stress path, the load at the start of
the cyclic stage is 100N, you want to cycle +/- 100N. You should enter 100
in the Amplitude cell. The peak to peak amplitude will be 200N and the
cycle will peak at 0N in extension to 200N in compression.
4.5.8.11 Axial peak to peak amplitude - indicates the peak to peak amplitude of
the axial cycle, i.e. the difference between the peak compression value and
the peak extension value.
4.5.8.12 Cycle chart – displays the appearance of a single axial cycle using the
parameters that you have entered. The x-axis displays the duration, in
seconds, of one cycle.
4.5.8.13 Radial amplitude cell – for entering the required amplitude of the cell
pressure cycles – see Fig 4-12, in kPa.
If you don’t want any radial cycling, enter a zero in this cell and the Radial
extension cell (Item 4.5.8.14).
This parameter is only available if you have selected ‘Non standard (multi
cycle test)’ or ‘Non standard (single cycle test)’ for the Cyclic shear
method control (Item 4.5.8.1).
4.5.8.14 Radial peak to peak amplitude - indicates the peak to peak amplitude of
the radial cycle, in kPa, i.e. the difference between the peak compression
value and the peak extension value.
This value will be applied to the specimen in addition to any cell pressure
that has already been applied to the specimen before the start of the stage.
If you don’t want any radial cycling, enter a zero in this cell and the Radial
compression cell (Item 4.5.8.13).
This parameter is only available if you have selected ‘Non standard (multi
cycle test)’ or ‘Non standard (single cycle test)’ for the Cyclic shear
method control (Item 4.5.8.1).
NOTE:
Radial cycling should only be used together with axial cycling and should be
carried out only at frequencies of less than 1 Hz.
4.5.8.15 Invert radial waveshape check box – for specifying whether or not the
radial cycle is in phase with the axial cycle (see Fig 4-13).
Tick the box to invert the wave shape, so that the radial cycle is 180º out of
phase with the axial cycling.
Leave the box un-ticked for the cycle to be in phase with the axial cycling.
This parameter is only available if you have selected ‘Non standard (multi
cycle test)’ or ‘Non standard (single cycle test)’ for the Cyclic shear
method control (Item 4.5.8.1).
Here is an example:
The cell pressure at the start of the cyclic stage is 400kPa, you want to
cycle the cell pressure by +/- 50kPa. You should enter 50 in the radial
compression cell and 50 in the radial extension cell. The peak to peak
amplitude will be 100kPa and the cycle will peak at 350kPa in extension to
450kPa in compression.
4.5.8.16 Poisson’s ratio cell – for entering the Poisson’s ratio which will be used to
calculate shear strain and shear modulus during the cyclic stage.
Poisson’s ratio = radial strain/axial strain. The default value is 0.5.
4.5.8.17 Axial strain reference control – for selecting whether axial strain during
the cyclic stage is calculated using data measured by the external
displacement transducer or the actuator displacement transducer.
Click the arrow to the right of the cell to select from ‘External’ or ‘Act Pos’ for
the actuator displacement.
4.5.8.18 Maximum test cycles cell – for entering the maximum number of cycles
that you want to be applied during the cyclic stage.
The maximum number that may be entered is 9,999,999.
If you have selected ‘Non standard (single cycle)’ for the Cyclic shear
method control (Item 4.5.8.1), this parameter will be fixed at 1.
If you have selected ‘ASTM D 3999 (Load Controlled Modulus and Damping
Test)’ or ASTM D 3999 (Displacement Controlled Modulus and Damping
Test)’ for the Cyclic shear method control, this parameter will be fixed at
40.
This value will only be used to stop the cycling if none of the selected
termination conditions have been reached, or if you have not selected any
other Termination conditions (Item 4.5.8.19).
Fig 4-13
4.5.8.19 Termination conditions - for specifying the conditions that you want the
cycling to stop at.
For each termination condition that you want to be taken into consideration
during the cyclic loading, the relevant box must be ticked and a value
entered. If more than one condition is specified, the one that is reached first
will determine when the cycling is stopped.
If none of the selected termination conditions are reached during the cyclic
loading, the cycling will stop when the number of cycles reaches the
number entered in the maximum test cycles cell.
The system must calculate the peak and valley data after each complete
cycle before it can determine whether termination conditions have been
met. Therefore, when the system stops cycling, it will stop after the final
cycle is complete.
4.5.8.20 Graph data memory management – for entering how frequently you want
to save the test data for loading back into the review panel after the test is
finished. The first value, N1, is fixed as ‘1’ so for cycle numbers 1 to 100,
data for every cycle will be stored. The default values for N2, N3 and N4
are 10, 100 and 1000 respectively.
This function is intended to save computer memory in cyclic stages with
very large numbers of cycles where it is not necessary to recall data for
every cycle.
NOTE:
If you enter 1 for all the values when performing a cyclic stage with very
large cycle counts (e.g. 1,000,000) large amounts of the system memory
may be consumed by the software and could cause the Windows operating
system to become unstable.
The following functions are only available during an automatic stage (see Fig 4-16):
4.7.24
Fig 4-16
4.9.9 K0 summary
Displays the following live calculated values for the stage:
Elapsed time – the amount of time that the stage has been running for, in
hours, minutes and seconds
Change in length – the change in specimen height, in millimetres, as
measured by the displacement transducer
A positive number indicates a decrease in length; a negative number
indicates an increase.
Axial strain – the change in specimen height as a percentage of the
specimen height at the start of the stage
Change in pore pressure – the amount of excess pore pressure that has
been generated since the start of the stage, in kPa
Change in volume – the amount of water that has moved into or out of the
specimen, as measured by the volume change device, in cubic centimetres
A negative number indicates that water has moved into the specimen; a
positive number indicates that water has moved out.
Volumetric strain – the amount of water that has moved into or out of the
specimen, as a percentage of the specimen volume at the start of the stage
Required volume – a live estimate of the total amount of water that must
move into or out of the specimen in order to keep the specimen diameter
constant, in cubic centimetres
A negative number indicates that water must move into the specimen; a
positive number indicates that water must move out.
This value is only during a drained stage.
Change in diameter – the change in specimen diameter, in millimetres,
either as measured by the radial on-specimen transducer or as calculated
from the change in specimen volume and height, depending on which option
is selected in the Setup K0 consolidation panel (Item 4.5.5.6)
CP target – the current target for the cell pressure, in kPa.
The target is calculated using the current cell pressure +/- whatever is
needed to correct for any change in the specimen diameter, using the
parameters entered in the Diameter control setup cells (Item 4.5.5.8) of the
Setup K0 consolidation panel.
Rate of displacement – the rate of displacement, in millimetres per minute,
that the actuator is set to move at.
4.9.15 K0 graph
Displays graphical K0 consolidation data.
Select which data series are displayed on the graph by using the graph
selection control and the graph legend.
Allows you to view specific areas of the graph in more detail by zooming in
and out and moving the region of the graph that is displayed.
The function of each item is as follows:
Cursor movement tool – not used in this software
Zoom tool – click it to enable zooming in and out of the graph
Panning tool – click it to enable panning the display (dragging the
contents of the graph to view different areas)
4.12.9.1 4.12.9.4
4.12.10.1 4.12.10.3
4.12.10.6 Closure error – the change in deformation between two successive cycle
peaks
This value is only relevant for tests being carried out to ASTM 3999
standard.
Displays final conditions for the specimen at the end of the cyclic shear
stage.
Click the tab to bring it to the front.
The tab will only be available after the stage has ended.
4.13.1
4.13.2
4.13.3
4.13.4
4.13.5
4.13.6
4.13.12
4.13.7
4.13.8
4.13.9
4.13.10
4.13.11
4.13.13
Fig 4-24
4.14.1 4.14.3 4.14.8 4.14.4 4.14.2 4.14.9 4.14.6 4.14.5 4.14.10 4.14.7
Fig 4-25
Click the arrow to the right of the cell to select from ‘Linear’, ‘Polynomial’ or
‘Linearization’.
When the fit method is changed, the data in the fitted data column of the
Calibration table is updated.
4.14.23.2 Verification results table – displays the following data calculated from the
transducer readings in the Verification data table (Item 4.14.23.1):
Mean – the arithmetic mean of the measured transducer readings at
each verification point, in engineering units
RE – the error between the value in the Mean column and the reference
reading (in the Ref. column of the Calibration table) at each verification
point, as a percentage of the reference reading
The calculation used for RE is: RE = (M – R)/R%
Where: M = Mean transducer reading at the verification point, in
engineering units;
R = Reference reading at the verification point, in engineering
units
RRE – the relative repeatability of the measured transducer readings at
each verification point, as a percentage of the reference reading
The calculation used for RRE is: RRE = (Mmax – Mmin)/R %
Where: Mmax = highest measured transducer reading at the
verification point, in engineering units
4.14.23.5
4.14.23.6
4.14.23.10
4.14.23.7
4.14.23.8
4.14.23.9
4.14.23.11 4.14.23.12
Fig 4-27
4.14.23.5 Reference output cell – displays the value that the reference equipment
should be set to for the next verification point.
4.14.23.6 Apply button – captures the live transducer reading and adds the value to
the Verification data table (Item 4.14.23.1).
4.14.23.8 Raw transducer reading indicator – displays the live transducer reading,
in volts.
4.14.23.9 Series indicator – displays the verification series number that you are
currently performing.
The maximum number of series that may be performed is three.
4.14.23.11 Exit (no save) button – closes the panel without saving the verification
data.
4.14.23.12 Save and exit button – closes the panel and saves the verification data,
entering it into the Verification data table on the Verification panel (Item
4.14.23).
NOTE:
Any movement of the actuator or increase / decrease in pressure will be
immediately auctioned by the system after the buttons of Items Error!
Reference source not found. to Error! Reference source not found. are
clicked. Care must be taken to ensure that the specimen is not damaged
by the action.
Fig 4-30
NOTE:
Entering inappropriate values in the transducer limit cells or selecting ‘Do
nothing’ for the transducer limit controls may result in damage to your
triaxial system and may compromise the accuracy of your test data.
Fig 4-32
4.17.13.1 4.17.13.5
4.17.13.2
4.17.13.6
Fig 4-33
NOTE:
Entering inaccurate text may cause misunderstandings when using the
software. If you change any of the existing languages or enter your own
language, it is your responsibilty to enter the correct wording.
4.19.2
4.19.1 4.19.3 4.19.4 4.19.5
4.19.8 4.19.7
4.19.6
Fig 4-35
The other items displayed in the Active panel area are the same as can be
found on the Test setup panel (Item 4.5), the Specimen setup panel (Item
4.6) and the test stage panels (Items 4.7 to 4.12).
The following Sections (5.0 to 5.17) are intended as a step-by-step user guide to help
you perform a cyclic triaxial test with the DYNATRIAX system. It is recommended that
before you start running real tests, you run a few dummy tests while following these
instructions in order to familiarise yourself with the system.
The software is very flexible and allows you to perform a test in many different ways –
the instructions given for each test stage may not cover all the possible options but
they should help you initially to understand how to use the software.
To perform a test, first follow Sections 5.0 and 5.2 which describe how to run the
software and set up a test. Then follow the instructions in the sections that describe
each of the stages that you have chosen for your test. For example, if you wanted to
do a test starting with an automatic saturation stage, then an automatic consolidation
stage, then a cyclic shear stage, followed by a manual monotonic shear stage, you
would follow Section 5.2 first, then 5.8, 5.9, 5.13, then 5.12, and finally 5.14 which
describes how to end a test.
To find out more information about any component of the software, refer to the
reference sections 4.1 to 4.12 which contain descriptions of all the screens and
functions within the software.
1. Run the software by double clicking the Triax_main icon (Fig 5-1) on the desktop
of the DYNATRIAX computer.
Fig 5-1
The software will open showing the Main panel (Fig 5-2) and all the hardware
components will initialise. The system will take control, holding the current
actuator position and holding the cell and back pressures at 0kPa.
Fig 5-2
1. Enter a name for the test data file in the Test name cell on the Main panel and
click the New test button (see Fig 5-3).
The location that the test file will be saved in is shown in the Log directory
indicator. If you want to change the location, enter a new filepath in the Log file
path cell in the System configuration panel (see Item 4.16).
2. Add the stages that you want to perform during the test to the Stage list by
selecting a stage in the Stage type cell and then clicking the Add stage button.
The selected stage will be added to the bottom of the Stage list. If you make a
mistake, click the Remove stage button to delete the last stage from the list.
If you want to load a set of stages in from a test that saved previously (using the
Save template button), click the Load template button.
Stage list Stage type cell Add stage button
Any of the test stages can either be performed manually, where you have full
control of the test, or automatically, where the system runs the stage without any
user intervention (except cyclic shear which is only manual). The test
parameters for each different stage must be entered before the previous stage
ends (for the first stage, they must be entered before you start the test).
The parameters that you set up for each stage (whether automatic or manual)
can be changed at any time during the test before the stage has started. Once
the stage is live, the parameters will be locked.
Steps 3 to 14 describe which test parameters to enter for each different stage
type.
3. Open the Test Setup panel by clicking on the Test setup tab (see Fig 5-4) and
enter the stage parameters as follows:
o To set up a manual saturation stage, follow Step 4
o To set up a manual isotropic consolidation stage, follow Step 5
o To set up a manual stress path stage, follow Step 6
o To set up a manual K0 consolidation stage, follow Step 7
o To set up a manual monotonic shear stage, follow Step 8
o To set up a automatic saturation stage, follow Step 0
o To set up a automatic isotropic consolidation stage, follow Step 10
o To set up a automatic stress path stage, follow Step 11
o To set up a automatic K0 consolidation stage, follow Step 12
o To set up a automatic monotonic shear stage, follow Step 13
o To set up a cyclic shear stage, follow Step 14
The parameters will be saved automatically when you exit the individual panels.
Test Setup tab Auto stage check box Stage list
Data logging method control Logarithmic factor cells Linear time interval cell
Fig 5-4
4. Enter parameters for the manual saturation stage(s) (if applicable) - click the
saturation stage that you want to edit in the Stage list to view the Setup
Saturation panel (Fig 5-4).
Enter only the following parameters on the panel:
o Auto stage check box – leave this box un-ticked
o Data logging method control – select a method from the options for logging
the saturation data
o Linear time interval cell – if you selected ‘Linear (time)’, enter the time
interval you want between logged readings, in seconds
o Logarithmic factor cells – if you selected ‘Logarithmic (time)’, enter
appropriate factors A and B to obtain your required logging rate
The other (unavailable) parameters are for an automatic stage and do not need
to be entered.
5. Enter parameters for the manual isotropic consolidation stage(s) (if applicable) -
click the isotropic consolidation stage that you want to edit in the Stage list to
view the Setup Consolidation Isotropic panel (Fig 5-5).
Auto stage check box Stage list
The other (unavailable) parameters are for an automatic stage and do not need
to be entered.
6. Enter parameters for the manual stress path stage(s) (if applicable) - click the
stress path stage that you want to edit in the Stage list to view the Setup Stress
Path panel (Fig 5-6).
Stress path method control Auto stage check box Stage list
Data logging method control Termination / hold conditions Linear time interval cell
Linear strain interval cell
Fig 5-6
o Linear time interval cell – if you selected ‘Linear (time)’, enter the time
interval you want between logged readings, in seconds
o Linear strain interval cell – if you selected ‘Linear (strain)’, enter the strain
interval you want between logged readings, in percent
The other (unavailable) parameters are for an automatic stage and do not need
to be entered.
7. Enter parameters for the manual K0 consolidation stage(s) (if applicable) - click
the K0 consolidation stage that you want to edit in the Stage list to view the
Setup K0 Consolidation panel (Fig 5-7).
Auto stage check box Termination / hold conditions Stage list
Data logging method control Diameter control setup cells Linear strain interval cell Linear time interval cell
Fig 5-7
8. Enter parameters for the manual monotonic shear stage(s) (if applicable) - click
the monotonic shear stage that you want to edit in the Stage list to view the
Setup Monotonic Shear panel (Error! Reference source not found.).
Failure criteria control Auto stage check box Termination strain cell Stage list
Data logging method control Displacement control Linear strain interval cell Linear time interval cell
Logarithmic factor cells
Fig 5-8
If the failure criterion that you selected is not reached during the stage, or
you selected ‘Use Termination Strain only’, the system will stop the shear
stage when the strain value entered here is reached.
o Displacement control – select whether you want the specimen to be sheared
using the actuator or the Tritech compression frame.
If you want to shear the specimen in extension, you must select ‘Actuator’.
o Data logging method control – select a method from the options for logging
the shear data.
o Linear time interval cell – if you selected ‘Linear (time)’, enter the time
interval you want between logged readings, in seconds;
o Logarithmic factor cells – enter appropriate factors A and B to obtain your
required logging rate for the shear stage.
o Linear strain interval cell – if you selected ‘Linear (strain)’, enter the strain
interval you want between logged readings, in percent.
The other (unavailable) parameters are for an automatic stage and do not need
to be entered.
Enter parameters for the automatic saturation stage(s) (if applicable) - click the saturation
stage that you want to edit in the Stage list to view the Setup Saturation panel (
9. Fig 5-9).
Saturation method control Differential pressure cell Minimum back pressure cell
Cell pressure increment cell Auto stage check box Ramp time cell
Data logging method control Logarithmic factor cells Linear time interval cell Stage list
Fig 5-9
Back pressure check box Initial stage time interval cell Logarithmic factor cells
Fig 5-10
11. Enter parameters for the automatic stress path stage(s) (if applicable) - click the
stress path stage that you want to edit in the Stage list to view the Setup Stress
Path panel (Fig 5-11).
Stress path method control Auto stage check box Termination / hold conditions Stage list
Data logging method control Hold time cell Linear time interval cell
Drainage method control Linear strain interval cell
Fig 5-11
o Stress path method control – select from the options whether you want the
stress path to be stress or strain controlled and, if stress controlled, which type
of stresses you want to be used as the targets.
o Drainage method control – select whether you want the stress path to be
drained or undrained.
o Hold time cell – if you want the system to hold the stresses once they are
reached at the end of the stress path ramp, enter the number of minutes that
you want the hold to last for before the system moves on to the next stage.
o Termination / hold conditions – if required, tick the box next to the condition
that you want the stress path to stop at and enter a value.
If you selected ‘Total vert. stress – strain control’ for the Stress path method
control, you must select a termination condition from the list or the system
will continue ramping until a system limit is reached.
If you selected one of the stress control options for the Stress path method,
the system will continue ramping until either the target stresses are met or
until one of the ticked termination conditions is met, whichever is reached
first.
If more than one condition is specified, the one that is reached first will
determine when the stage stops.
o Data logging method control – select a method from the options for logging
the stress path data.
o Linear time interval cell – if you selected ‘Linear (time)’, enter the time
interval you want between logged readings, in seconds;
o Linear strain interval cell – if you selected ‘Linear (strain)’, enter the strain
interval you want between logged readings, in percent.
o Target abscissa cell – if you selected a stress control method, enter the
abscissa (x-axis) stress path point that you want to reach by the end of the
stress path ramp (see Fig 5-12).
o Target ordinate cell – if you selected a stress control method, enter the
ordinate (y-axis) stress path point that you want to reach by the end of the
stress path ramp.
o Ramp time cell – if you selected a stress control method , enter the time
period that you want the system to take to ramp to the target stresses.
Target abscissa cell Target ordinate cell
Fig 5-12
o Rate of strain cell – if you selected the strain control method, enter the rate of
strain that you want the stress path to run at (see Fig 5-13).
o Loading method control – if you selected the strain control method, select
whether you want the stress path to be in compression or extension.
Rate of strain cell
12. Enter parameters for the automatic K0 consolidation stage(s) (if applicable) - click
the K0 consolidation stage that you want to edit in the Stage list to view the
Setup K0 Consolidation panel (Fig 5-14).
Rate of strain cell Auto stage check box Stage list
Loading method control Termination / hold conditions
Data logging method control Linear time interval cell Diameter control setup cells
Diameter control method Drainage method control Linear strain interval cell
Fig 5-14
13. Enter parameters for the automatic monotonic shear stage(s) (if applicable) -
click the monotonic shear stage that you want to edit in the Stage list to view the
Setup Monotonic Shear panel (Fig 5-15).
Enter only the following parameters on the panel:
o Auto stage check box –tick this box
o Drainage method control – select whether you want the stage to be drained
or undrained.
o Loading method control – select whether you want the specimen to be
sheared in compression or extension.
o Rate of strain cell – enter the rate of strain that you want the specimen to be
sheared at.
o Failure criteria control – select the criterion at which you want the shear
stage to be stopped.
If you select ‘Use Termination Strain only’, the shear stage will be stopped
when the value in the Termination strain cell is reached.
o Termination strain cell – enter the axial strain at which you want the shear
stage to be stopped;
If the failure criterion that you selected is not reached during the stage, or
you selected ‘Use Termination Strain only’, the system will stop the shear
stage when the strain value entered here is reached.
o Data logging method control – select a method from the options for logging
the shear data.
o Linear time interval cell – if you selected ‘Linear (time)’, enter the time
interval you want between logged readings, in seconds;
Loading method control Rate of strain cell Stage list
Drainage method control Auto stage check box Termination strain cell
14. Enter parameters for the cyclic shear stage(s) (if applicable) - click the cyclic
stage that you want to edit in the Stage list to view the Setup Cyclic Shear
panel (Fig 5-16).
Stage list
Fig 5-16
o Sample rate cell – if you selected ‘User defined’ for the Waveshape control,
enter the number of readings that you want to be logged per second. For all
other waveshapes, enter the number of readings that you want to be logged
per cycle.
o Axial amplitude cell – enter the amplitude for the axial cyclic loading.
o Radial amplitude cell – if you selected a ‘Non standard’ control mode, you
can enter an amplitude for a radial cyclic loading.
o Invert radial waveshape check box - if you selected a ‘Non standard’ control
mode, you can tick this box to apply the radial cycle 180º out of phase with the
axial cycle.
Waveshape control User waveform cell Sample rate cell
Axial amplitude cell Fig 5-18 Invert radial waveshape Radial amplitude
check box cell
o Poisson’s ratio cell – enter the Poisson’s ratio value you want the software to
use to calculate shear strain and shear modulus during the stage (see
o Fig 5-19).
o Axial strain reference control – select whether you want the software to use
external displacement data (‘External’) or actuator position data (‘Act Pos’) to
calculate axial strain during the stage.
o Maximum test cycles cell – enter the maximum number of cycles that you
want to be applied during the stage.
o Termination conditions – if required, tick the box next to the condition that
you want the cycling to stop at and enter a value.
If none of the conditions are reached, the cycling will stop when the number
of cycles reached the value entered in the Maximum test cycles cell.
Maximum test cycles cell Axial strain reference control
Fig 5-19
Fig 5-20
17. Open the Manual control panel (Fig 5-23) by clicking the Manual control
button on the Main panel (see Fig 5-21).
Fig 5-21
18. Ensure that the volume change device is set well within the minimum and
maximum limits that you specified in the Transducer limit cells in the System
configuration panel (see Item 4.16).
Actuator controls
20. With the cell body placed onto the cell base, hold the load cell up so it is not in
contact with the specimen and lock it in position with the piston locking bracket.
21. If you need to zero the load cell or the pressure transducers (or any on-specimen
transducers if you are using them), do it now by clicking the Real-time display
button (see Fig 5-21) and using the Zero buttons (see Fig 5-23). Close the
panel by clicking the Close button.
Actuator assembly
Vacuum valve
Cell body
Fig 5-24
22. Unlock the piston locking bracket and gently lower the load cell until the two
sections of the vacuum top cap are in contact with each other.
23. Apply a vacuum to the top cap through the vacuum valve in the top of the cell
and then close the valve.
24. Carefully lower the actuator assembly into contact with the load cell assembly
using the Actuator controls in the Manual control panel (see Fig 5-22). When
the ball joint is seated in the socket, secure the connection by screwing the ball
joint collar onto the load cell assembly (see Fig 5-24).
Ensure that load is not applied to the specimen during this operation by
monitoring the load cell reading in the Manual control panel.
26. Click the Start new test button on the Main panel (see Fig 5-25).
The software will show the panel for the first stage of the test.
Fig 5-25
If Section 5.2 and the instructions for the previous stage (if applicable) of your
test have been followed and the current stage of the test is a saturation stage,
the Saturation panel should now be displayed.
Fig 5-26
2. Apply a cell pressure increment:
o In the Cell pressure increment cell (see Fig 5-27), enter your required
increment (i.e. the amount that you want the cell pressure to increase by for
this pressure step).
o Click the Cell pressure increment button to apply the pressure.
Cell pressure increment cell
Cell pressure increment button Fig 5-27 Cell pressure target indicator
The cell pressure that the system must reach after applying the increment will be
displayed in the Cell pressure target indicator.
The cell pressure will be increased by the required amount and when the
pressure is reached, logging will start.
Test data for the increment will be logged at the rate that you specified in the
Setup saturation panel and plotted on the Saturation graph (see Fig 5-29).
The data series can be changed using the Graph selection control.
o When you want to end the increment (i.e. when the pore pressure has reached
equilibrium), proceed to Step 3.
3. Apply the next increment – to apply a cell pressure increment, repeat Step 2. To
apply a back pressure increment, go to Step 4. To end the saturation stage, go
to Step 5.
Back pressure increment button Fig 5-28 Back pressure target indicator
The back pressure that the system must reach after applying the increment will
be displayed in the Back pressure target indicator.
The back pressure will be increased by the required amount and when the
pressure is reached, logging will start.
Test data for the increment will be logged at the rate that you specified in the
Setup saturation panel and plotted on the Saturation graph (see Fig 5-29).
The data series can be changed using the Graph selection control.
o When you want to end the increment (i.e. when the pore water pressure has
become equal to the back pressure and volume change is negligible), return to
Step 3.
6. Click the Exit saturation button to proceed to the next stage in the Stage list.
Exit saturation button End saturation button End of stage values tab
The next step only is only applicable if you want to apply a cell and back pressure
ramp:
o In the Ramp time cell, enter the time, in seconds, that you want the ramp to
last for (i.e. how long you want the system to take to reach the target
pressures entered).
o Click the Cell and back pressure increment button to start the ramp.
The cell pressure will increase at a constant rate over the time period you
specified until the value in the Cell pressure target indicator is reached. The
back pressure will increase at the same rate to maintain the differential entered in
the Back pressure differential cell.
Test data for the ramp will be logged at the rate that you specified in the Setup
saturation panel and plotted on the Saturation graph (see Fig 5-29). The data
series can be changed using the Graph selection control.
Cell and back pressure increment button Cell pressure target indicator
o If you want to do another pressure ramp, repeat Step 7. If you want to end the
saturation stage, go to Step 5.
If Section 5.2 and the instructions for the previous stage (if applicable) of your
test have been followed and the current stage of the test is an isotropic
consolidation stage, the Isotropic Consolidation panel should now be
displayed.
1. Enter the cell and back pressures that you want to use for the consolidation in
the Target cell pressure cell and the Target back pressure cell (see Fig 5-31).
2. In the Target ramp time cell, enter the number of seconds that you want the
system to reach the target pressures.
Target ramp time cell Target cell pressure cell Target back pressure cell
With the back line valve closed, the cell and back pressures will increase or
decrease to meet the targets (if necessary) over the time period that you entered
in the Target ramp time cell. When the target pressures are reached, the Start
consolidation button will be enabled.
The initialize stage will be logged at the rate that you specified in the Setup
consolidation isotropic panel (Section 5.2, Step 5) and the data will be
displayed on the Consolidation graph (see Fig 5-32). The data series can be
changed using the Graph selection control.
4. When the specimen is ready to consolidate (i.e. when the pore pressure has
reached equilibrium), click the Start consolidation button.
The back line valve will open and the pressure will be applied to the specimen.
Test data for the stage will be logged at the rate you specified in the Setup
consolidation isotropic panel (Section 5.2, Step 5) and plotted on the
Consolidation graph (see Fig 5-32).
5. When you are ready to end the consolidation stage, click the End consolidation
button.
The back line valve will close.
Values for the specimen at the end of the stage will be calculated and displayed
on the End of stage values tab (see Fig 5-32).
6. Click the Exit consolidation button to proceed to the next stage in the Stage
list.
If Section 5.2 and the instructions for the previous stage (if applicable) of your
test have been followed and the current stage of the test is a stress path stage,
the Stress path panel should now be displayed.
1. Enter the stresses that you want to be applied to the specimen in the Target
abscissa cell and the Target ordinate cell. Enter the number of minutes that
you want the system to take to reach those stresses in the Ramp time cell. If
you selected ‘Total vert. stress – strain control’ for the Stress path control
method, these values are not required and so will not be visible – instead, you
need to enter the rate of strain that you want the system to ramp at, in percent
per hour (see Fig 5-33).
2. Select whether you want the stress path to be drained or undrained from the
Drainage method control.
Target abscissa cell Start stress path button Rate of strain cell
3. If you selected ‘Total vert. stress – strain control’ for the Stress path control
method, select whether you want the stress path to be in compression or
extension from the Loading method control (see Fig 5-34).
4. Click the Start stress path button.
The system will take a few seconds to initialise the panel.
If you selected ‘Drained’ for the Drainage method control, the back line valve
will open.
If you selected a stress controlled method, the system will ramp to the targets
over the time period that you specified by controlling the movement of the
actuator and by increasing or decreasing the cell pressure, as necessary. The
cell pressure, back pressure and load required to achieve the target stress are
displayed in the Stress path summary (see Fig 5-35).
If the stress path is strain controlled, the system will ramp at the rate of strain that
you specified in Section 5.2, Step 6.
Test data for the stage will be logged at the rate you specified in the Setup
stress path panel (Section 5.2, Step 6) and plotted on the Stress path graph
(see Fig 5-35). The data series can be changed using the Graph selection
control.
The stress path will stop ramping automatically when the target stresses are
reached or when one of the termination conditions that you set in the Setup
Stress Path panel (Section 5.2, Step 6) is met, whichever is reached sooner.
The system will then hold the stresses until you end the stage.
5. To end the stress path stage, click the End stress path button.
Values for the specimen at the end of the stage will be calculated and displayed
on the End of stage values tab (see Fig 5-35).
6. Click the Exit stress path button to proceed to the next stage in the Stage list.
If Section 5.2 and the instructions for the previous stage (if applicable) of your
test have been followed and the current stage of the test is a K 0 consolidation
stage, the K0 consolidation panel should now be displayed.
1. Enter your required rate of strain in the Rate of strain cell, in percent per hour
(see Fig 5-36).
Loading method control Rate of strain cell
2. Select whether you want the K0 stage to be in compression or extension from the
Loading method control.
3. Select whether you want the K0 stage to be drained or undrained from the
Drainage method control.
The K0 stage can only be undrained if a radial on-specimen transducer is being
used to measure specimen diameter.
4. Click the Start K0 consolidation button.
The system will take a few seconds to initialise the panel.
If you selected ‘Drained’ for the Drainage method control, the back line valve
will open.
The system will ramp at the rate of strain that you specified in the Setup K0
consolidation panel (Section 5.2, Step 7). The cell pressure will be increased
or decreased as necessary, in order to keep the specimen diameter constant.
The target cell pressure is displayed in the K0 summary. (see Fig 5-37).
Test data for the stage will be logged at the rate you specified in the Setup K0
Consolidation panel (Section 5.2, Step 7) and plotted on the K0 graph (see Fig
5-37). The data series can be changed using the Graph selection control.
The K0 stage will stop automatically when one of the termination conditions that
you set in the Setup K0 consolidation panel (Section 5.2, Step 7) is met,
whichever one is reached first. The system will then hold the stresses until you
end the stage.
6. Click the Exit K0 consolidation button to proceed to the next stage in the Stage
list.
If Section 5.2 and the instructions for the previous stage (if applicable) of your
test have been followed and the current stage of the test is a monotonic shear
stage, the Monotonic shear panel should now be displayed.
1. In the Rate of strain cell, enter the rate of strain that you want the specimen to
be sheared at, in percent per hour (see Fig 5-38).
Start shear button Rate of strain cell
Fig 5-39
5. If you selected ‘Tritech’ for the Displacement control, you will see the message
shown in Fig 5-39 telling you to lock the piston.
Lock the piston by screwing the piston locking collar up until it is in contact with
the top beam of the Tritech frame (see Fig 5-40). When you have done this, click
the OK button on the message.
Fig 5-40
The Tritech compression frame or the actuator (depending on what you selected
for the Displacement control in Section 5.2, Step 8) will start moving at the rate
entered in the Rate of strain cell, in the direction selected from the Loading
method options. If you selected ‘Drained’ from the Drainage options, the back
line valve will open.
Exit shear button End shear button End of stage values tab
Test data for the stage will be logged at the rate you specified in the Setup
Monotonic shear panel (Section 5.2, Step 8) and plotted on the Shear graph
(see Fig 5-41). The data series can be changed using the Graph selection
control.
The shear stage will stop automatically when one of the termination conditions
that you set in the Setup monotonic shear panel (Section 5.2, Step 8) is met,
whichever one is reached first.
6. If you want to end the stage before any of your selected failure criteria are
reached, click the End shear button.
Values for the specimen at the end of the stage will be calculated and displayed
on the End of stage values tab (see Fig 5-41).
Fig 5-42
7. If the specimen was sheared using the Tritech, you will see the message shown
in Fig 5-42 telling you to unlock the piston.
Run the Tritech down a little (using the ‘fast down’ button on the front panel) to
release the load until it is possible to unscrew the piston locking collar.
Unlock the piston by screwing the piston locking collar down until it is down as far
away from the top beam of the Tritech as it will go (see Fig 5-43). When you
have done this, click the OK button on the message.
Fig 5-43
8. Click the Exit shear button to proceed to the next stage in the Stage list.
If Section 5.2 and the instructions for the previous stage (if applicable) of your
test have been followed and the current stage of the test is a saturation stage,
the Saturation panel should now be displayed (Fig 5-44) and the saturation
stage should have started.
If you selected ‘Cell pressure increment’ for the Saturation method control, the
software will apply increments (in accordance with the parameters that you
entered in the Setup Saturation panel) in the following way:
When the system is ready to apply a cell pressure increment, the value in the
Cell pressure target indicator will be updated to show the cell pressure that the
system is aiming to reach. The cell pressure will then increase steadily over the
time period that you entered in the Ramp time cell until the target pressure is
reached and then logging will start.
When the pore pressure has levelled off or starts to decrease, the system will
apply the next increment.
The system will continue applying increments until either the B value reaches
0.95 or a transducer limit is reached.
If you selected ‘Back pressure increment’ for the Saturation method control,
the software will apply increments, in accordance with the parameters that you
entered in the Setup Saturation panel, in the following way:
The system will start the saturation with a cell pressure increment and then apply
alternate increments of cell pressure and back pressure.
When the system is ready to apply a cell pressure increment, the value in the
Cell pressure target indicator will be updated to show the cell pressure that the
system is aiming to reach. The cell pressure will then increase steadily over the
time period that you entered in the Ramp time cell until the target pressure is
reached, and then logging will start.
When the pore pressure has levelled off or starts to decrease, the system will
apply a back pressure increment:
The values in the Cell pressure target indicator and the Back pressure target
indicator will be updated to show the cell pressure that the system is aiming to
reach and the back pressure that is needed to give the differential that you
entered in the Differential pressure cell. The cell and back pressures will then
increase steadily over the time period that you entered in the Ramp time cell
until the target pressures are reached, then the back line valve will open and
logging will start.
When the change in volume is constant and the pore pressure have levelled off,
the system will apply a cell pressure increment.
The system will continue applying alternate increments until either the B value
reaches 0.95 or a transducer limit is reached.
If you selected ‘Cell + back pressure ramp’ for the Saturation method control,
the software will ramp the cell and back pressures simultaneously and
continuously over the time period that you entered in the Ramp time cell until
the target pressures are reached.
When the change in volume is constant and the pore pressure have levelled off,
the system will end the saturation stage.
If, at the end of the saturation, the pore pressure has not reached the value that
you entered in the Minimum back pressure cell, the system will continue to
apply increments until the pore pressure equals or exceeds that value.
Test data for the stage will be logged at the rate you specified in the Setup
Saturation panel (Section 5.2, Step 0) and plotted on the Saturation graph
(see Fig 5-44). The data series can be changed using the Graph selection
control.
Next increment button End saturation button Cell pressure target indicator Back pressure target indicator
1. If at any time during an increment you want to over-ride the system and end the
increment manually, click the Next Increment button and the system will then
apply the next increment.
2. If you want to over-ride the system and end the saturation stage manually, click
the End saturation button.
When the saturation stage has ended, the system will proceed automatically to
the next stage of your test.
If Section 5.2 and the instructions for the previous stage (if applicable) of your
test have been followed and the current stage of the test is an isotropic
consolidation stage, the Isotropic consolidation panel should now be displayed
(Fig 5-45) and the system should be initializing the pressures for the
consolidation.
The system will initialise the cell and back pressures by increasing or decreasing
them (if necessary) from the values at the end of the previous stage to the values
that are needed to achieve the consolidation pressure that you entered for the
Effective consolidation pressure. If you ticked the Back pressure check box,
the system will use the back pressure from the previous stage and set the cell
pressure to equal the back pressure plus the effective consolidation pressure. If
you did not tick the Back pressure check box, the system will round the current
pore pressure up to the nearest 10kPa and use this value for the back pressure,
and set the cell pressure to equal the back pressure plus the effective
consolidation pressure.
Target back pressure cell
Target cell pressure cell Target ramp time cell
The target pressures are indicated in the Target cell pressure cell and the
Target back pressure cell and the amount of time that the system will take to
reach them is shown in the Target ramp time cell.
The initialise stage will be logged at the rate you entered in the Setup isotropic
consolidation panel (Section 5.2, Step 10) and the data will be plotted on the
Consolidation graph. The data series can be changed using the Graph
selection control.
When the specimen is ready to consolidate (i.e. when the pore pressure has
reached equilibrium), the back line valve will open to apply the back pressure to
the specimen and logging will start.
Test data for the stage will be logged at the rate you specified in the Setup
Isotropic consolidation panel (Section 5.2, Step 10) and plotted on the
Consolidation graph (see Fig 5-45).
The system will end the consolidation stage when there is no further significant
volume change and the pore pressure has levelled off.
If you want to over-ride the system and end the consolidation manually, click the
End consolidation button.
When the consolidation stage has ended, the system will proceed automatically
to the next stage of your test.
If Section 5.2 and the instructions for the previous stage (if applicable) of your
test have been followed and the current stage of the test is a stress path stage,
the Stress path panel should now be displayed (Fig 5-46) and the system
should have started to apply the stress path ramp.
The system will ramp to the values displayed in the Target abscissa cell and
Target ordinate cell over the time period in the Ramp time cell. The ramp will
be acheived by controlling the direction and speed of the actuator movement and
by increasing or decreasing the cell pressure, as necessary. The cell pressure,
back pressure and load required to achieve the target stresses are dispayed in
the Stress path summary. If you selected ‘Total vert. Stress – strain control’ for
the Stress path method control, the system will ramp using the actuator at the
rate you entered in the Rate of strain cell.
If you selected ‘Drained’ for the Drainage method control, the back line valve
will open at the start of the stage.
Ramp time cell Graph selection control Fig 5-46 Stress path graph
Test data for the stage will be logged at the rate you specified in the Setup
stress path panel (Section 5.2, Step 11) and plotted on the Stress path graph
(see Fig 5-46). The data series can be changed using the Graph selection
control.
The system will end the stress path stage when the target stresses are reached
or when one of the termination conditions that you set in the Setup stress path
panel is met, whichever is reached sooner. The system will hold the stresses
that it has reached for the time period entered in the Hold time cell, then it will
end the stage.
If you want to over-ride the system and end the stage manually, before the
targets or the termination conditions are reached, or before the hold time period
is finished, click the End stress path button.
When the stress path stage has ended, the system will proceed automatically to
the next stage of your test.
If Section 5.2 and the instructions for the previous stage (if applicable) of your
test have been followed and the current stage of the test is a K 0 consolidation
stage, the K0 consolidation panel should now be displayed (Fig 5-47) and the
system should have started to apply the K0 ramp
The system will ramp at the rate that you entered in the Rate of strain cell. The
cell pressure will be increased or decreased as necessary in order to keep the
specimen diameter constant. The target cell pressure and volume change are
displayed in the K0 summary.
If you selected ‘Drained’ for the Drainage method control, the back line valve
will open at the start of the stage.
Test data for the stage will be logged at the rate you specified in the Setup K0
consolidation panel (Section 5.2, Step 12) and plotted on the K0 graph (see Fig
5-47). The data series can be changed using the Graph selection control.
The system will end the K0 stage when one of the termination conditions that you
set in the Setup K0 consolidation panel is met, whichever one is reached first.
If you want to over-ride the system and end the stage manually, before a
termination condition is reached, click the End K0 consolidation button.
When the K0 stage has ended, the system will proceed automatically to the next
stage of your test.
If Section 5.2 and the instructions for the previous stage (if applicable) of your
test have been followed and the current stage of the test is a monotonic shear
stage, the Monotonic shear panel should now be displayed and the system
should have started shearing the specimen.
End shear button Rate of strain cell
NOTE:
The specimen can only be sheared using the actuator in an automatic
monotonic shear stage. (This is because using the Tritech to shear the
sample requires user intervention at the start of the stage and so it cannot
be performed automatically.) The actuator has a maximum capacity of 5kN.
The actuator will move at the rate displayed in the Rate of strain cell, in the
direction appropriate for the selection made for the Loading method control. If
you selected ‘Drained’ for the Drainage method control, the back line valve will
open at the start of the stage.
Test data for the stage will be logged at the rate you specified in the Setup
monotonic shear panel (Section 5.2, Step 13) and plotted on the Shear graph
(see Fig 5-48). The data series can be changed using the Graph selection
control.
The system will end the shear stage when the failure conditions that you
specified in the Setup monotonic shear panel are met.
If you want to over-ride the system and end the stage manually, before the failure
conditions are reached, click the End shear button.
When the shear stage has ended, the system will proceed automatically to the
next stage of your test.
If Section 5.2 and the instructions for the previous stage (if applicable) of your
test have been followed and the current stage of the test is a cyclic shear stage,
the Cyclic shear panel should now be displayed.
1. In the Cycles to view cell, enter the number of cycles that you want to see on
the Cyclic shear graph.
2. Click the Start cyclic shear button.
The system will take a few seconds to initialise the panel.
Start cyclic shear button
The system will start cycling the specimen according to the parameters that you
entered in Section 5.2, Step 14.
If you selected ‘Drained’ for the Drainage method control, the back line valve
will open at the start of the stage.
Test data for the stage will be logged at the rate you specified in the Setup
cyclic shear panel (Section 5.2, Step 14) and plotted on the Cyclic shear
graph (see Fig 5-50). The type of data for the graph (actual values, peak and
valley data or a hysteresis loop of the last cycle) can be selected using the
Graph data type control and the data series can be changed using the Graph
selection control.
The cycling will stop automatically when one of the termination conditions that
you set in the Setup cyclic shear panel (Section 5.2, Step 14) is met, or the
number of cycles reaches the value entered in the Maximum test cycles cell,
whichever one occurs first.
(The system must calculate the peak and valley data after each complete cycle
before it can determine whether termination conditions have been met.
Therefore, when the system stops cycling, it will be after the final cycle is
complete.)
3. If you want to end the cycling before any of the termination conditions are
reached, click the End cyclic shear button.
NOTE:
If you click the End cyclic shear button mid-cycle, the system will
complete the cycle before it stops.
Values for the specimen at the end of the stage will be calculated and displayed
on the End of stage values tab (see Fig 5-50).
Exit cyclic shear button End cyclic shear button End of stage values tab
4. Click the Exit cyclic shear button to proceed to the next stage in the Stage list.
When the last stage in the Stage list has finished, the system will wait for you to
either add another stage or to end the test. To end the test when it has finished,
proceed as follows:
1. Check that the locking collar has been released and is not in contact with the top
beam of the Tritech compression frame (see Fig 5-51).
Fig 5-51
Fig 5-52
The pressures will be reduced to zero and the actuator will return to the load zero
position. The Main panel will be displayed and control of the system will be
returned to you.
o Open the bleed valve on the top of the cell and let the water drain out of the
cell.
o When the cell is empty, release the push fitting to remove the waste line from
the cell tap and re-connect the cell pressure line to the cell.
If you need to end a test for any reason before it has finished, follow Steps 1 to 3
above but be aware of the following note.
NOTE:
Clicking the Unload test button before a test has finished will cause the
test to be aborted. You will be unable to restart or recover the test after the
unload command has been confirmed.
Test data up to the unload point will be saved in the location specified in the
Main panel.
Table 5-1
1. On the Main panel, click the Manual control button (see Fig 5-53) to open the
Manual control panel.
Fig 5-53
2. Make a note of the Actuator reading and enter 1.00 in the Movement cell (see
Fig 5-54).
3. Click the Actuator down button. If the actuator moves down in one smooth
action, with no ringing (vibrating/juddering) and the Actuator reading has
Fig 5-55
7. Open the Manual control panel and make a note of the Actuator reading.
Click the Actuator up button and when the movement has finished, check the
change in the Actuator reading. If the target movement was not achieved open
the Tuning panel and increase the Proportional Gain by 50. Repeat this step
until the target movement is achieved, in which case the tuning is complete and
you can close the panel, or until there is slight ringing, in which case continue to
the next step.
8. Open the Tuning panel and increase the Derivative Gain by 50. Close the
panel and then check the response using the Manual control panel. If the
actuator moves smoothly with no ringing, go to the next step. If there is still
ringing, repeat this step until it is eliminated.
9. Make a note of the Actuator reading in the Manual control panel. Click the
Actuator up button and when the movement has finished, check the change in
the Actuator reading. Close the panel. If the target 1.00mm movement was not
achieved, go to the next step. If it was achieved then the tuning is completed.
10. Open the Tuning panel and increase the Integral Gain by 50. Close the panel
and repeat Step 9. Repeat until the movement is 1.00mm +/-0.02mm. The
tuning is then complete.
1. On the Main panel, click the Tuning button (see Fig 5-56) to open the Tuning
panel.
Fig 5-56
2. Change the values in the Actuator PID cells to P = 500, I = 0 and D = 0 (see Fig
5-55) and observe the behaviour of the control on the graph and in the Reading
indicators. If the cycle shape is poor and the Target amplitude is not being
achieved, increase the Proportional Gain in steps of 50 until the Reading
amplitude matches the Target amplitude, in which case the procedure is
finished and you can close the panel, or until there is slight ringing and/or
overshoot (see Fig 5-57), in which case go to step 3.
Fig 5-57
3. Now increase the Derivative Gain in steps of 50 until the ringing has been
eliminated and the cycle is a good shape.
4. Check the target and reading amplitude values. If there is an unacceptable error,
increase the Integral Gain in steps of 50 until the Amplitude error is reduced to
an acceptable value.
5. The tuning is now complete. Click the Close button to close the Tuning panel.
Fig 5-58
Fig 5-59
1. On the Main panel, click the Tuning button (see Fig 5-56) to open the Tuning
panel. The percentage error between the Target Amplitude and Reading
Amplitude is shown in the Amplitude error cell (see Fig 5-61).
Fig 5-60
2. To reduce the error to an acceptable level using manual controls, see step 3. To
use the automatic control, see step 4.
3. Click the Manual mode button to turn on manual control (see Fig 5-61). The
Mean adjustment control and Amplitude adjustment control will display the
relevant values from the cyclic stage that is currently running. Use the arrows to
the left of the controls to adjust the values up or down as necessary to correct the
error in amplitude and/or mean. When you have finished, click the Close button
to close the panel.
Mean
adjustment
control
4. Enter the maximum error between the target amplitude and the reading
amplitude that is acceptable for your requirements in the Allowable error cell.
Select how frequently you want the error to be automatically checked and
corrected in the Cycle check interval cell. Click the Amplitude control button
to turn the function on and the software will then adjust the amplitude and mean
to keep the error below the allowable level. When you have finished, click the
Close button to close the panel.
1. Click the Main tab to view the Main panel (Fig 5-62), and then click on the
Calibration button to open the Calibration panel (Fig 5-63).
Main tab
Fig 5-62
2. Click the Edit calibrations button and enter the password to unlock the panel.
The password is set in the System configuration panel.
Edit calibrations button
Calibration table Transducer selection control Transducer details
3. Select the type of transducer that you want to calibrate from the Transducer
selection control.
4. If necessary, change the values for the full scale and the number of decimal
places, and update the information in the Transducer details.
5. Click the New calibration button to clear the current data (see Fig 5-64).
6. Put the transducer into position for the first calibration point (usually zero), using
an appropriate piece of reference equipment.
New calibration button Delete calibration point button
7. Click the Add calibration point button, and in the window that appears (Fig
5-65), enter the reading on the reference equipment (in engineering units). Click
the Apply button to accept the value and close the window.
Fig 5-65
8. The next window that appears shows the transducer reading in volts (Fig 5-66).
When the reading is steady, click the Read button to accept the value and close
the window.
Fig 5-66
The reference value will be entered in the ‘Ref.’ column of the Calibration table
and the transducer reading will be entered in the ‘Measured’ column.
If you make a mistake, the point can be deleted by clicking the Delete
calibration point button (see Fig 5-64) and then repeated.
9. Move the transducer into position for the next calibration point.
10. Repeat Steps 7 to 9 for calibration points at suitable intervals across the whole
range of the transducer.
The minimum number of calibration points that you can do is two; the maximum
is ten, including zero.
11. Select a suitable fit method for the data using the Fit method control (see Fig
5-64).
The fitted data is displayed in the ‘Fitted data’ column of the Calibration table.
The following steps are optional, and describe how to carry out a verification. If you
don’t wish to verify the calibration, go straight to Step 18 to end the calibration.
12. Click the Verification button (see Fig 5-62) to open the Verification panel (Fig
5-67).
Verification data table Verification results table
13. Click the Run verification button to open a new panel (see Fig 5-68).
14. Put the transducer into position using the reference equipment, according to the
value indicated in the Reference output cell. When the transducer reading is
steady (it is displayed in the Transducer reading indicator), click the Apply
button.
15. Repeat Step 14 for all the subsequent points of the calibration.
16. When Series 1 is completed, click the Save and exit button to save the data and
close the panel, or continue with another verification series.
A maximum of three verification series may be carried out.
The verification data will be entered into the Verification data table. The mean,
the error (RE) and the repeatability (RRE) are calculated for each point and
displayed in the Verification results table.
Fig 5-68
If you need to calibrate another transducer on the same channel, repeat the
procedures from Step 4 for the new transducer.
19. When you have finished calibrating all the transducers, if you want to save the
calibration data in a certificate format, click the Export calibration to Excel
button (see Fig 5-63). The data for all the currently loaded transducers will be
saved in an Excel file with one certificate worksheet for each transducer.
If you have calibrated more than one transducer on the same channel, make
sure that the calibration for the transducer that is currently connected to the
system is loaded in before the next step. Use the Import Transducer cal button
to load in data that you saved in Step 18.
If a test is running, the software cannot be closed. Wait until the test has finished
or end the test following the instructions in Section 5.14 before closing the
software.
1. Click the Main tab to open the Main panel and then click the Quit button (see Fig
5-69).
Main tab
Fig 5-69
Notes:
Notes: