Anda di halaman 1dari 1

On-chip spectroscopy for the

18
quality control testing of F radiotracers
Mark D. Tarn, Stephen J. Archibald and Nicole Pamme
Department of Chemistry and Positron Emission Tomography Research Centre,
University of Hull, Cottingham Road, Hull, HU6 7RX, UNITED KINGDOM

We present the use of miniaturised absorbance and Raman spectroscopic techniques for performing several on-chip quality control (QC)
tests on the 2-[18F]fluoro-2-deoxy-D-glucose ([18F]FDG) radiotracer for positron emission tomography (PET) medical imaging. Absorbance
spectroscopy was employed for the determination of appearance/optical clarity and colourimetric pH determination, while Raman
spectroscopy is being developed for the analysis of residual solvents and for the identification of unpurified [18F]FDG. The platforms reduce
analysis times, radioactive exposure to personnel, and require minimal sample volumes, making them suitable for dose-on-demand
radiotracer synthesis.

PET imaging Quality control of [18F]FDG Microfluidic chip setup


 [18F]FDG is a glucose analogue used to image  three layer glass
glucose metabolism (suitable for cancer imaging) chips
 microfluidics enables “dose-on-demand” synthesis;  top layer:
requires miniaturisation of other production steps mixing of reagents
 [18F]FDG must follow stringent QC testing  middle layer:
3.1 mm pathlength
 bottom layer: outlet
 injection of positron  chip clamped into
emitting radiotracer chip holder
(e.g. [18F]FDG)
 optical fibres/Raman
 positron annihilation probe aligned with
yields emission of  here, we employ on-chip spectroscopy for the pathlength channel
gamma radiation testing of three important QC parameters:
 miniaturised Ocean
 detection of γ rays allows construction of image
 appearance: clear, colourless, free of particles Optics spectrometer
based on intensity and location of tracer
 pH: must be in the range of pH 4.5 – 8.5
 different tracers allow analysis for oncology,  simple setup for absorbance spectroscopy and
neurology, cardiology and inflammation  residual solvents: must be below threshold levels Raman spectroscopy in a miniaturised format

Absorbance spectroscopy pH determination: V-shaped plots and star plots


 Universal  star plots of
indicator and signal intensity
sample/standards generated for
pumped into chip each pH standard
and mixed from the spectra
at wavelengths
 spectra of
from 520–640 nm
coloured product
recorded  more accurate pH
determination
 absorbance values than “yes/no”
of pH standards V-plot method
plotted at 551 nm
 gives V-shaped plot
 absorbance detection used to determine  a line is drawn  shape
pass/fail based on appearance/optical clarity across pH 4.5 and recognition
 doses of [18F]FDG pumped directly into chip pH 8.5 values used to judge QC
pass/fail of
 identified failed dose ([18F]FDG sample #5387);  [18F]FDG samples below the line will pass the [18F]FDG
exhibited yellow-green colour to the eye QC test, those above the line will fail. samples

Raman spectroscopy Conclusions


 fingerprinting of  Absorbance-based detection employed for on-
a range of chip QC testing for appearance/optical clarity.
solvents that
can be used for  Colourimetric assay with absorbance detection
[18F]FDG used for pH determination by V-plots & star plots.
synthesis and  Raman spectroscopy being investigated for
for the cleaning residual solvent analysis; suitable for ethanol but
of synthesis further optimisation required for acetonitrile.
apparatus
 Ethanol  Acetonitrile  Raman also used to identify unpurified samples.
LOD = 2410 ppm LOD = 870 ppm
 Tests take only a few min and <2 µL of sample.
 solvents used in [18F]FDG production must be
below threshold levels in the final injectable dose  Eliminates subjectivity of some QC tests and
 Raman also used would reduce radiation exposure to technician.
 most common solvents and allowed concentrations
to identify purified
 Developed alongisde other tests towards an
ethanol: used for washing; max. 5000 ppm vs. unpurified
integrated microfluidic QC platform.
[18F]FDG samples
acetonitrile: used for synthesis; max. 410 ppm
 unpurified sample Acknowledgements
 calibration curves prepared for ethanol and exhibited increase The authors thank the Daisy Appeal (UK) (grant: DAhul0211) and the
acetonitrile using on-chip Raman spectroscopy University of Hull for financial support, and Dr Assem Allam for his generous
in background support of the project.

Anda mungkin juga menyukai