PORTABLE DETECTOR BUMP TEST & CALIBRATION

REPORT
THE OWNERS MANUAL FOR THE DETECTOR PROVIDES ACCURATE INSTRUCTION AS TO HOW THE UNITS IS CALIBRATED. THE
MANUAL INSTRUCTIONS SHOULD BE PRESENT AND FOLLOWED. BE SURE TO REMOVE THE FLOW VALVE AFTER USE, AND CHECK
THAT THE CYLINDER IS NOT LEAKING.

CALIBRATION & BUMP TEST


Date: 16th July-2018 Location: RSC-Islamabad Client: WDI - 812

Cal Gas Expiry Date: DEC 2019 Cylinder Pressure on Completion: 800 psi

Calibration Gas Information Used:

Cylinder Number Production Date Expiry Date


Cylinder Lot No.: 641144 Dec, 2017 Dec , 2019
Cylinder No.:A0446904

Calibration Gas Components Component Mixture


Hydrogen Sulfide 25 PPM
Carbon Monoxide 100 PPM
Methane 50% LEL (2.5% Methane)
Oxygen 18.0%
Any other (if more than 4 gas detectors) NA
Balance Nitrogen
Gas Type and
Response Time (0 Stabilize Time
No. MODEL Serial No. ppm./% for Remarks
to 1ppm) (Return to zero)
LEL
ALARMS
10114837
MSA ALTAIR QUAD AND
01 00051212 4sec 6
5X GAS READINGS
ARE OK

Comment:
MSA Altair 5x has been calibrated according to the specification. Humidifier tube was used during test
for accuracy. All sensor elements are working properly

Calibrated By: Tabish Khattak (Operations Coordinator)

Signed:

*Note: If a sensor does not give an initial response to gas within 10 seconds, the sensor fails. Recalibrate and
Retest. If the sensor still fails, complete an AFN and take out of service.
Use a separate form if 2 different gas bottles are used (ie. Quad Gas/Single gas)

FR – 08 – 4.2.1.3f - 01