Anda di halaman 1dari 43

1 Anatomy and physiology

of the heart

Description and location of the heart Function of the heart


The heart is a hollow, four-chambered muscular organ that The heart is the hardest working organ in the body. The heart
lies in the middle of the thoracic cavity between the lungs, functions primarily as a pump to circulate blood and supply
behind the sternum, in front of the spinal column, and just the body with oxygen and nutrients. Each day the average
above the diaphragm (Figure 1-1). The top of the heart (the heart beats over 100,000 times. During an average lifetime,
base) is at approximately the level of the second intercostal the human heart will beat more than 3 billion times.
space. The bottom of the heart (the apex) is formed by the The heart is capable of adjusting its pump performance
tip of the left ventricle and is positioned just above the dia- to meet the needs of the body. As needs increase, as with
phragm to the left of the sternum at the fifth intercostal exercise, the heart responds by accelerating the heart rate
space, midclavicular line. There, the apex can be palpated to propel more blood to the body. As needs decrease, as
during ventricular contraction. This physical examination with sleep, the heart responds by decreasing the heart rate,
landmark is referred to as the point of maximal impulse resulting in less blood flow to the body.
(PMI) and is an indicator of the heart’s position within the The heart consists of:
thorax.  four chambers
The heart is tilted forward and to the left so that the — two atria that receive incoming blood
right side of the heart lies toward the front. About two- — two ventricles that pump blood out of the heart
thirds of the heart lies to the left of the body’s midline and  four valves that control the flow of blood through the heart
one-third extends to the right. The average adult heart is  an electrical conduction system that conducts electrical
approximately 5″ (12 cm) long, 3½″ (8 to 9 cm) wide, and impulses to the heart, resulting in muscle contraction.
2½″ (6 cm thick) — a little larger than a normal-sized fist.
The heart weighs between 7 and 15 oz (200 and 425 grams). Heart surfaces
Heart size and weight are influenced by age, weight, body
build, frequency of exercise, and heart disease. There are four main heart surfaces to consider when dis-
cussing the heart: anterior, posterior, inferior, and lateral
(Figure 1-2). The heart surfaces are explained below:
 anterior — the front
Clavicle  posterior — the back
 inferior — the bottom
1
 lateral — the side.
Rib 2

Heart
3 Structure of the heart wall
4
Sternum
The heart wall is arranged in three layers (Figure 1-3):
/
5  the pericardium — the outermost layer
6
 the myocardium — the middle muscular layer
Diaphragm
 the endocardium — the inner layer.
12
7
Enclosing and protecting the heart is the pericardium,
8
which consists of an outer fibrous sac (the fibrous pericar-
9
10 dium) and an inner two-layered, fluid-secreting membrane
Xiphoid process 12th thoracic
11
(the serous pericardium). The outer fibrous pericardium
of sternum vertebra comes in direct contact with the covering of the lung (the
pleura) and is attached to the center of the diaphragm infe-
Figure 1-1. Location of the heart in the thorax. riorly, to the sternum anteriorly, and to the esophagus,
trachea, and main bronchi posteriorly. This position

ECG workout_Chap01.indd 1 4/28/2011 2:04:23 AM


2 Anatomy and physiology of the heart

friction as the heart beats. In certain conditions, large


accumulations of fluid, blood, or exudates can enter the
pericardial space and may interfere with ventricular filling
and the heart’s ability to contract.
The myocardium is the thick, middle, muscular layer
0 ), that makes up the bulk of the heart wall. This layer is com-
J posed primarily of cardiac muscle cells and is responsible
I Lateral for the heart’s ability to contract. The thickness of the
/ myocardium varies from one heart chamber to another.
Chamber thickness is related to the amount of resist-
ance the muscle must overcome to pump blood out of the
chamber.
/
The endocardium is a thin layer of tissue that lines the
Anterior
-
s
/ ;
Posterior inner surface of the heart muscle and the heart chambers.
\ /
Extensions and folds of this tissue form the valves of the
vz: heart.

Inferior Circulatory system


Figure 1-2. Heart surfaces. The circulatory system is required to provide a continuous
flow of blood to the body. The circulatory system is a closed
system consisting of heart chambers and blood vessels.
anchors the heart to the chest and prevents it from shift- The circulatory system consists of two separate circuits,
ing about in the thorax. The serous pericardium is a con- the systemic circuit and the pulmonary circuit. The sys-
tinuous membrane that forms two layers: the parietal layer temic circuit is a large circuit and includes the left side of
lines the inner surface of the fibrous sac and the visceral the heart and blood vessels, which carry oxygenated blood
layer (also called epicardium) lines the outer surface of the to the body and deoxygenated blood back to the right heart.
heart muscle. Between the two layers of the serous peri- The pulmonary circuit is a small circuit and includes the
cardium is the pericardial space, or cavity, which is usually right side of the heart and blood vessels, which carry deox-
filled with 10 to 30 mL of thin, clear fluid (the pericardial ygenated blood to the lungs and oxygenated blood back to
fluid) secreted by the serous layers. The primary function the left heart. The two circuits are designed so that blood
of the pericardial fluid is to provide lubrication, preventing flow is pumped from one circuit to the other.

Epicardium (visceral layer


of serous pericardium)
Myocardium
r
-

'
Endocardium
i
i
(4 v

i
1 1 Pericardial cavity

fir
i \i
Parietal layer
of serous pericardium

. Fibrous pericardium

Figure 1-3. Heart wall.

ECG workout_Chap01.indd 2 4/28/2011 2:04:23 AM


Heart valves 3

Heart chambers much greater resistance to flow (the arterial pressure in


the systemic circulation).
The interior of the heart consists of four hollow chambers
(Figure 1-4). The two upper chambers, the right atrium Heart valves
and the left atrium, are divided by a wall called the inter-
atrial septum. The two lower chambers, the right ven- There are four valves in the heart: the tricuspid valve,
tricle and the left ventricle, are divided by a thicker wall separating the right atrium from the right ventricle; the
called the interventricular septum. The two septa divide pulmonic valve, separating the right ventricle from the
the heart into two pumping systems — a right heart and pulmonary arteries; the mitral valve, separating the left
a left heart. atrium from the left ventricle; and the aortic valve, sepa-
The right heart pumps venous (deoxygenated) rating the left ventricle from the aorta (Figure 1-5). The
blood through the pulmonary arteries to the lungs primary function of the valves is to allow blood flow in
(Figure 1-5). Oxygen and carbon dioxide exchange takes one direction through the heart’s chambers and prevent
place in the alveoli and arterial (oxygenated) blood a backflow of blood (regurgitation). Changes in cham-
returns via the pulmonary veins to the left heart. The ber pressure govern the opening and closing of the heart
left heart then pumps arterial blood to the systemic valves.
circulation, where oxygen and carbon dioxide exchange The tricuspid and mitral valves separate the atria from
takes place in the organs, tissues, and cells; then venous the ventricles and are referred to as the atrioventricular
blood returns to the right heart. Blood flow within the (AV) valves. These valves serve as in-flow valves for the ven-
body is designed so that arteries carry oxygen-rich blood tricles. The tricuspid valve consists of three separate cusps
away from the heart and veins carry oxygen-poor blood or leaflets and is larger in diameter and thinner than the
back to the heart. This role is reversed in pulmonary mitral valve. The tricuspid valve directs blood flow from
circulation: pulmonary arteries carry oxygen-poor blood the right atrium to the right ventricle. The mitral valve (or
into the lungs, and pulmonary veins bring oxygen-rich bicuspid valve) has only two cusps. The mitral valve directs
blood back to the left heart. blood flow from the left atrium to the left ventricle. Both
The thickness of the walls in each chamber is related valves are encircled by tough, fibrous rings (valve rings).
to the workload performed by that chamber. Both atria The leaflets of the AV valves are attached to thin strands
are low-pressure chambers serving as blood-collecting of fibrous cords called chordae tendineae (heart strings)
reservoirs for the ventricles. They add a small amount of (Figure 1-6). The chordae tendineae are then attached to
force to the moving blood. Therefore, their walls are rela- papillary muscles, which arise from the walls and floor of
tively thin. The right ventricular wall is thicker than the the ventricles. During ventricular filling (diastole) when
walls of the atria, but much thinner than that of the left the AV valves are open, the valve leaflets, the chordae
ventricle. The right ventricular chamber pumps blood a tendineae, and the papillary muscles form a funnel, pro-
fairly short distance to the lungs against a relatively low moting blood flow into the ventricles. As pressure increases
resistance to flow. The left ventricle has the thickest wall, during ventricular contraction (systole), the valve cusps
because it must eject blood through the aorta against a close. Backflow of blood into the atria is prevented by con-
traction of the papillary muscles and the tension in the
chordae tendineae. Dysfunction of the chordae tendineae
Interatrial septum or a papillary muscle can cause incomplete closure of an AV
valve. This may result in a regurgitation of blood from the
ventricle into the atrium, leading to cardiac compromise.
Left atrium The first heart sound (S1) is the product of tricuspid and
mitral valve closure. S1 is best heard at the apex of the heart
Right atrium
located on the left side of the chest, fifth intercostal space,
Left
heart midclavicular line.
Right The aortic and pulmonic valves have three cuplike cusps
heart
Left
shaped like a half-moon and are referred to as the semi-
Right ventricle lunar (SL) valves. These valves serve as out-flow valves
ventricle
for the ventricles. The cusps of the SL valves are smaller
and thicker than the AV valves and do not have the sup-
port of the chordae tendineae or papillary muscles. Like
the AV valves, the rims of the semilunar valves are sup-
ported by valve rings. The pulmonary valve directs blood
Interventricular septum
flow from the right ventricle to the pulmonary artery.
Figure 1-4. Chambers of the heart. The aortic valve directs blood flow from the left ventri-
cle to the aorta. As pressure decreases during ventricular

ECG workout_Chap01.indd 3 4/28/2011 2:04:25 AM


4 Anatomy and physiology of the heart

Alveolus of lung

O3
S
Pulmonary arteries
(to lungs)
4
Aorta (to body)

Superior vena cava * Pulmonary veins


(from upper body) (from lungs)
>
LA
Aortic valve

Pulmonic valve \

RA i
Mitral valve
Tricuspid valve v

LV
V
' Septum
Inferior vena cava k
RV
(from lower body)

Figure 1-5. Chambers, valves, blood flow.


RA, right atrium; RV, right ventricle;
LA, left atrium; LV, left ventricle.

relaxation (diastole), the valve cusps close. Backflow of Blood flow through the
blood into the ventricles is prevented because of the cusps’
fibrous strength, their close approximation, and their
heart and lungs
shape. The second heart sound (S2) is produced by closure Blood flow through the heart and lungs is traditionally
of the aortic and pulmonic SL valves. It is best heard over described by tracing the flow as blood returns from the sys-
the second intercostal space on the left or right side of the temic veins to the right side of the heart, to the lungs, back
sternum. to the left side of the heart, and out to the arterial vessels

Superior vena cava

Branches of right pulmonary artery


u Aortic arch
Pulmonic valve

Branches of left pulmonary artery


Right atrium
Left atrium
Right pulmonary veins c 1
Left pulmonary veins
I
Mitral valve
Tricuspid valve
Chordae tendineae 1
A\l
S3
v Myocardium

Interventricular septum

,
I Aortic valve

Right ventricle
Papillary muscle Left ventricle
/

Inferior vena cava

Descending aorta

Figure 1-6. Papillary muscles and chordae tendineae.

ECG workout_Chap01.indd 4 4/28/2011 2:04:25 AM


Coronary circulation 5

of the systemic circuit (Figure 1-5). The right atrium right coronary artery supplies the right side of the heart and
receives venous blood from the body via two of the body’s the left coronary artery supplies the left side of the heart.
largest veins (the superior vena cava and the inferior vena The right coronary artery arises from the right side
cava) and from the coronary sinus. The superior vena cava of the aorta and consists of one long artery that travels
returns venous blood from the upper body. The inferior downward and then posteriorly. The major branches of the
vena cava returns venous blood from the lower body. The right coronary artery are:
coronary sinus returns venous blood from the heart itself.  conus artery
As the right atrium fills with blood, the pressure in the  sinoatrial (SA) node artery (in 55% of population)
chamber increases. When pressure in the right atrium  anterior right ventricular arteries
exceeds that of the right ventricle, the tricuspid valve  acute marginal artery
opens, allowing blood to flow into the right ventricle. As  AV node artery (in 90% of population)
the right ventricle fills with blood, the pressure in that  posterior descending artery with septal branches
chamber increases, forcing the tricuspid valve shut and the (in 90% of population)
pulmonic valve open, ejecting blood into the pulmonary  posterior left ventricular arteries (in 90% of population).
arteries and on to the lungs. In the lungs, the blood picks Dominance is a term commonly used to describe coro-
up oxygen and excretes carbon dioxide. nary vasculature and refers to the distribution of the terminal
The left atrium receives arterial blood from the pulmo- portion of the arteries. The artery that gives rise to both the
nary circulation via the pulmonary veins. As the left atrium posterior descending artery with its septal branches and the
fills with blood, the pressure in the chamber increases. posterior left ventricular arteries is considered to be a “domi-
When pressure in the left atrium exceeds that of the left nant” system. In approximately 90% of the population, the
ventricle, the mitral valve opens, allowing blood to flow into right coronary artery (RCA) is dominant. The term can be
the left ventricle. As the left ventricle fills with blood, the confusing because in most people the left coronary artery is of
pressure in that chamber increases, forcing the mitral valve wider caliber and perfuses the largest percentage of the myo-
shut and the aortic valve open, ejecting blood into the aorta cardium. Thus, the dominant artery usually does not perfuse
and systemic circuit, where the blood releases oxygen to the the largest proportion of the myocardium. The left coronary
organs, tissues, and cells and picks up carbon dioxide. artery arises from the left side of the aorta and consists of the
Although blood flow can be traced from the right side of left main coronary artery, a short stem, which divides into
the heart to the left side of the heart, it is important to realize the left anterior descending artery and the circumflex artery.
that the heart works as two pumps (the right heart and the left The left anterior descending (LAD) travels downward over
heart) working simultaneously. As the right atrium receives the anterior surface of the left ventricle, circles the apex, and
venous blood from the systemic circulation, the left atrium ends behind it. The major branches of the LAD are:
receives arterial blood from the pulmonary circulation. As  diagonal arteries
the atria fill with blood, pressure in the atria exceeds that of  right ventricular arteries
the ventricles, forcing the AV valves open and allowing blood  septal perforator arteries.
to flow into the ventricles. Toward the end of ventricular fill- The circumflex artery travels along the lateral aspect of
ing, the two atria contract, pumping the remaining blood the left ventricle and ends posteriorly. The major branches
into the ventricles. Contraction of the atria during the final of the circumflex are:
phase of diastole to complete ventricular filling is called the  SA node artery (in 45% of population)
atrial kick. The ventricles are 70% filled before the atria con-  anterolateral marginal artery
tract. The atrial kick adds another 30% to ventricular capac-  posterolateral marginal artery
ity. In normal heart rhythms, the atria contract before the  distal left circumflex artery.
ventricles. In abnormal heart rhythms, the loss of the atrial In 10% of the population, the circumflex artery gives
kick results in incomplete filling of the ventricles, causing a rise to the posterior descending artery with its septal
reduction in cardiac output (the amount of blood pumped branches, terminating as the posterior left ventricular
out of the heart). Once the ventricles are filled with blood, arteries. A left coronary artery with a circumflex that gives
pressure in the ventricles increases, forcing the AV valves rise to both the posterior descending artery and the pos-
shut and the SL valves open. The ventricles contract simul- terior left ventricular arteries is considered a “dominant”
taneously, ejecting blood through the pulmonary artery into left system. When the left coronary artery is dominant, the
the lungs and through the aortic valve into the aorta. entire interventricular septum is supplied by this artery.
Table 1-1 summarizes the coronary artery distribution to
the myocardium and the conduction system.
Coronary circulation The right and left coronary artery branches are intercon-
The blood supply to the heart is supplied by the right cor- nected by an extensive network of small arteries that provide
onary artery, the left coronary artery, and their branches the potential for cross flow from one artery to the other.
(Figure 1-7). There is some individual variation in the These small arteries are commonly called collateral vessels
pattern of coronary artery branching, but in general, the or collateral circulation. Collateral circulation exists at birth

ECG workout_Chap01.indd 5 4/28/2011 2:04:26 AM


6 Anatomy and physiology of the heart

Right coronary artery Left coronary artery

Left main coronary artery


Sinoatrial node artery
Conus artery Left circumflex coronary artery
Left anterior descending artery
Acute marginal artery Anterolateral marginal artery
Posterolateral marginal artery
Anterior right ventricular
Distal left circumflex artery
artery
Diagonal arteries
AV node artery

Posterior left Right ventricular artery


ventricular arteries Septal perforator artery

Posterior descending
coronary artery

Septal branch

Figure 1-7. Coronary circulation.

Table 1-1.
Coronary arteries
Coronary artery and its branches Portion of myocardium supplied Portion of conduction system supplied
Right coronary artery
Right atrium Sinotrial (SA) node (55%)*
Right ventricle Atrioventricular (AV) node and bundle of His (90%)*
Inferior wall of left ventricle (90%)*
Posterior one-third of interventricular septum (90%)*

Left coronary artery


Left anterior descending (LAD) Anterior wall of left ventricle Right and left bundle branches
Anterolateral wall of left ventricle
Anterior two-thirds of interventricular septum

Circumflex Left atrium SA node (45%)*


Anterolateral wall of left ventricle AV node and bundle of His (10%)*
Posterolateral wall of left ventricle
Posterior wall of left ventricle
Inferior wall of left ventricle (10%)*
Posterior one-third of interventricular septum (10%)*

* = of population

ECG workout_Chap01.indd 6 4/28/2011 2:04:26 AM


Cardiac innervation 7

but the vessels do not become functionally significant until


the myocardium experiences an ischemic insult. If a block-
age occurs in a major coronary artery, the collateral vessels
enlarge and provide additional blood flow to those areas of
reduced blood supply. However, blood flow through the col-
lateral vessels isn’t sufficient to meet the total needs of the
myocardium in most cases. In other vascular beds of the body,
arterial blood flow reaches a peak during ventricular contrac-
tion (systole). However, myocardial blood flow is greatest dur-
ing ventricular diastole (when the ventricular muscle mass
is relaxed) than it is during systole (when the heart’s blood
vessels are compressed). The blood that has passed through
the capillaries of the myocardium is drained by branches of
the cardiac veins whose path runs parallel to those of the
coronary arteries. Some of these veins empty directly into the
right atrium and right ventricle, but the majority feed into
the coronary sinus, which empties into the right atrium.

Cardiac innervation
The heart is under the control of the autonomic nerv-
ous system located in the medulla oblongata, a part of
the brain stem. The autonomic nervous system regu-
lates functions of the body that are involuntary, or not
under conscious control, such as blood pressure and
heart rate. It includes the sympathetic nervous system
and the parasympathetic nervous system, each produc-
ing opposite effects when stimulated. Stimulation of
the sympathetic nervous system results in the release
of norepinephrine, a neurotransmitter, which acceler-
ates the heart rate, speeds conduction through the AV
node, and increases the force of ventricular contrac-
tion. This system prepares the body to function under
stress (“fight-or-flight” response). Stimulation of the
parasympathetic nervous system results in the release
of acetylcholine, a neurotransmitter, which slows the
heart rate, decreases conduction through the AV node,
and causes a small decrease in the force of ventricular
contraction. This system regulates the calmer functions
of the body (“rest-and-digest” response). Normally a bal-
ance is maintained between the accelerator effects of
the sympathetic system and the inhibitory effects of the
parasympathetic system.

ECG workout_Chap01.indd 7 4/28/2011 2:04:26 AM


2
Cardiac cells
Electrophysiology

water, producing positively and negatively charged ions.


The heart is composed of thousands of cardiac cells. The An ion with a positive charge is called a cation. An ion with
cardiac cells are long and narrow, and divide at their ends a negative charge is called an anion. Potassium (K+) is the
into branches. These branches connect with branches of primary ion inside the cell and sodium (Na+) is the primary
adjacent cells, forming a branching and anastomosing ion outside the cell.
network of cells. At the junctions where the branches join A membrane separates the inside of the cardiac cell
together is a specialized cellular membrane of low electri- (intracellular) from the outside (extracellular). There is a
cal resistance, which permits rapid conduction of electrical constant movement of ions across the cardiac cell mem-
impulses from one cell to another throughout the cell net- brane. Differences in concentrations of these ions deter-
work. Stimulation of one cardiac cell initiates stimulation mine the cell’s electric charge. The distribution of ions
of adjacent cells and ultimately leads to cardiac muscle on either side of the membrane is determined by several
contraction. factors:
There are two basic kinds of cardiac cells in the heart:  Membrane channels (pores) — The cell membrane has
the myocardial cells (or “working” cells) and the pace- openings through which ions pass back and forth between
maker cells. The myocardial cells are contained in the the extracellular and intracellular spaces. Some channels
muscular layer of the walls of the atria and ventricles. The are always open; others can be opened or closed; still others
myocardial “working” cells are permeated by contractile can be selective, allowing one kind of ion to pass through
filaments which, when electrically stimulated, produce and excluding all others. Membrane channels open and
myocardial muscle contraction. The primary function of close in response to a stimulus.
the myocardial cells is cardiac muscle contraction, fol-  Concentration gradient — Particles in solution move,
lowed by relaxation. The pacemaker cells are found in the or diffuse, from areas of higher concentration to areas of
electrical conduction system of the heart and are primar- lower concentration. In the case of uncharged particles,
ily responsible for the spontaneous generation of electrical movement proceeds until the particles are uniformly dis-
impulses. tributed within the solution.
Cardiac cells have four primary cell characteristics:  Electrical gradient — Charged particles also diffuse, but
 automaticity — the ability of the pacemaker cells to the diffusion of charged particles is influenced not only by
generate their own electrical impulses spontaneously; this the concentration gradient, but also by an electrical gradi-
characteristic is specific to the pacemaker cells. ent. Like charges repel; opposite charges attract. Therefore,
 excitability — the ability of the cardiac cells to respond positively charged particles tend to flow toward negatively
to an electrical impulse; this characteristic is shared by all charged particles and negatively charged particles toward
cardiac cells. positively charged particles.
 conductivity — the ability of cardiac cells to conduct  Sodium-potassium pump — The sodium-potassium
an electrical impulse; this characteristic is shared by all pump is a mechanism that actively transports ions across
cardiac cells. the cell membrane against its electrochemical gradient.
 contractility — the ability of cardiac cells to cause car- This pump helps to reestablish the resting concentrations
diac muscle contraction; this characteristic is specific to of sodium and potassium after cardiac depolarization.
myocardial cells. Electrical impulses are the result of the flow of ions (pri-
marily sodium and potassium) back and forth across the
cardiac cell membrane (Figure 2-1). Normally there is an
Depolarization and repolarization ionic difference between the two sides. In the resting car-
Cardiac cells are surrounded and filled with an electrolyte diac cell, there are more negative ions inside the cell than
solution. An electrolyte is a substance whose molecules outside the cell. When the ions are so aligned, the rest-
dissociate into charged particles (ions) when placed in ing cell is called polarized. During this time, no electrical

ECG workout_Chap02.indd 8 4/28/2011 1:09:30 AM


Electrical conduction system of the heart 9

+ + + + + + + + + + +
Electrical conduction system
Na* of the heart
Resting cell
( polarized state ) The heart is supplied with an electrical conduction system
that generates and conducts electrical impulses along
+ + + + + + + + + + specialized pathways to the atria and ventricles, causing
+ + + + + + + + them to contract (Figure 2-2). The system consists of the
+ + sinoatrial node (SA node), the interatrial tract (Bach-
Depolarization
beginning
! K*
Na" mann’s bundle), the internodal tracts, the atrioventricular
node (AV node), the bundle of His, the right bundle branch,
( stimulated state ) j the left bundle branch, and the Purkinje fibers.
\ + + +
+
•• •••••••••••• ••••••••••••••••••••••••••••••••a
+ + + + + + + The SA node is located in the wall of the upper right
atrium near the inlet of the superior vena cava. Special-
'

I + + + + + + + + + + + -
% ized electrical cells, called pacemaker cells, in the SA node
discharge impulses at a rate of 60 to 100 times per minute.
Depolarization Na* I Pacemaker cells are located at other sites along the con-
complete | K+
! duction system, but the SA node is normally in control and
is called the pacemaker of the heart because it possesses
the highest level of automaticity (its inherent firing rate
is greater than that of the other pacemaker sites). If the

Repolarization
beginning
s a a « a•••••••
+ + + + + +
••••••
+ + r
+•••••+
+•••••••
- '

Na - n +
SA node fails to generate electrical impulses at its normal
rate or stops functioning entirely, or if the conduction
1-K of these impulses is blocked, pacemaker cells in second-
( recovery state )
+
M » » aaMM
+ + + + + + + v+ “
» aa « aaaaMaatMaMaMa > MMM » waa » >aa > » iaaM>WW » mwnnaWF
* ) ary pacemaker sites can assume control as pacemaker of
+ + the heart, but at a much slower rate. Such a pacemaker is
+ + + + + + + + + + + called an escape pacemaker because it usually only appears
Na* (“escapes”) when the faster firing pacemaker (usually the
Repolarization SA node) fails to function. Pacemaker cells in the AV junc-
K+
complete tion generate electrical impulses at 40 to 60 times per
minute. Pacemaker cells in the ventricles generate elec-
+ + + + +
•••••••••••••!••••••••
+ + + + + + trical impulses at a much slower rate (30 to 40 times per
minute or less). In general, the farther away the impulse
Figure 2-1. Depolarization and repolarization of a cardiac cell.
originates from the SA node, the slower the rate. A beat or
series of beats arising from an escape pacemaker is called
activity is occurring and a straight line (isoelectric line) is an escape beat or escape rhythm and is identified according
recorded on the ECG (Figure 2-5). to its site of origin (for example, junctional, ventricular).
Once a cell is stimulated, the membrane permeability As the electrical impulse leaves the SA node, it is con-
changes. Potassium begins to leave the cell, increasing ducted through the left atria by way of Bachmann’s bundle
cell permeability to sodium. Sodium rushes into the cell, and through the right atria via the internodal tracts, caus-
causing the inside of the cell to become more positive ing electrical stimulation (depolarization) and contraction
than negative (cell is depolarized). Muscle contraction of the atria. The impulse is then conducted to the AV node
follows depolarization. Depolarization and muscle con- located in the lower right atrium near the interatrial sep-
traction are not the same. Depolarization is an electrical tum. The AV node relays the electrical impulses from the
event that results in muscle contraction, a mechanical atria to the ventricles. It provides the only normal conduc-
event. tion pathway between the atria and the ventricles. The AV
After depolarization, the cardiac cell begins to recover. node has three main functions:
The sodium-potassium pump is activated to actively trans-  To slow conduction of the electrical impulse through the
port sodium out of the cell and move potassium back into AV node to allow time for the atria to contract and empty
the cell. The inside of the cell becomes more negative than its contents into the ventricles (atrial kick) before the ven-
positive (cell is repolarized) and returns to its resting state. tricles contract. This delay in the AV node is represented on
Depolarization of one cardiac cell acts as a stimulus on the ECG tracing as the flat line of the PR interval.
adjacent cells and causes them to depolarize. Propagation  To serve as a backup pacemaker, if the SA node fails, at a
of the electrical impulses from cell to cell produces an rate of 40 to 60 beats per minute
electric current that can be detected by skin electrodes and  To block some of the impulses from being conducted to
recorded as waves or deflections onto graph paper, called the ventricles when the atrial rate is rapid, thus protecting
the ECG. the ventricles from dangerously fast rates.

ECG workout_Chap02.indd 9 4/28/2011 1:09:30 AM


10 Electrophysiology

Interatrial tract (Bachmann's bundle)

f/ I >
N S
I \ s
S
% N
/ \
/ / \
/ t
SA node /
/
/
/ \ \ f\
\
\ v7
\
/ / \\ '/ Interatrial septum
I I \\
I
I l I
\\
• I
I
t
I I ' II
I I
I r/ Anterior fascicle of left bundle branch
Internodal I l
tracts I \
l/ \ \ \
N \ \ \
'
C ss \
V
\

s
\ Posterior fascicle of left bundle branch
\ s \
I
\ \ \
\
\
\ \ \
AV node s
\ V Interventricular septum
s
s
s
\ N
\ V
s
/

s \
s
Bundle of His ss s I /
\
/
sV
N .V /
/

* Purkinje fibers
Right bundle branch

Figure 2-2. Electrical conduction system of the heart.

After the delay in the AV node, the impulse moves because atrial repolarization occurs during ventricular
through the bundle of His. The bundle of His divides into depolarization and is hidden in the QRS complex. The PR
two important conducting pathways called the right bundle interval represents the time from the onset of atrial depo-
branch and the left bundle branch. The right bundle branch larization to the onset of ventricular depolarization. The PR
conducts the electrical impulse to the right ventricle. The segment, a part of the PR interval, is the short isoelectric
left bundle branch divides into two divisions: the anterior line between the end of the P wave to the beginning of the
fascicle, which carries the electrical impulse to the anterior QRS complex. It is used as a baseline to evaluate elevation
wall of the left ventricle, and the posterior fascicle, which or depression of the ST segment. The QRS complex depicts
carries the electrical impulse to the posterior wall of the ventricular depolarization, or the spread of the impulse
left ventricle. Both bundle branches terminate in a network throughout the ventricles. The ST segment represents
of conduction fibers called Purkinje fibers. These fibers early ventricular repolarization. The T wave represents
make up an elaborate web that carry the electrical impulses
directly to the ventricular muscle cells. The ventricles are
capable of serving as a backup pacemaker at a rate of 30 to R
40 beats per minute (sometimes less). Transmission of the
electrical impulses through the conduction system is slow-
est in the AV node and fastest in the His-Purkinje system
(bundle of His, bundle branches, and Purkinje fibers).
The heart’s electrical activity is represented on the PR interval
monitor or ECG tracing by three basic waveforms: the < > ST segment
P wave, the QRS complex, and the T wave (Figure 2-3). T
k
A U wave is sometimes present. Between the waveforms P U
are the following segments and intervals: the PR interval,
the PR segment, the ST segment, and the QT interval.
Although the letters themselves have no special signifi- Q
S
cance, each component represents a particular event in the r° r
*
QT interval
>
depolarization–repolarization cycle. The P wave depicts PR segment
atrial depolarization, or the spread of the impulse from Figure 2-3. Relationship of the electrical conduction system to
the SA node throughout the atria. A waveform represent- the ECG.
ing atrial repolarization is usually not seen on the ECG

ECG workout_Chap02.indd 10 4/28/2011 1:09:31 AM


Refractory and supernormal periods of the cardiac cycle 11

^A
M

QS
T
1r \
Biphasic
deflections

f-

Figure 2-4. The cardiac cycle.

Negative
Electric
current
-A Positive
ventricular repolarization. The U wave, which isn’t always deflection deflection
present, represents late ventricular repolarization. The QT – +
interval represents total ventricular activity (the time from
Figure 2-6. Relationship between current flow and waveform
the onset of ventricular depolarization to the end of ven-
deflections.
tricular repolarization).

The cardiac cycle can be applied to the P wave, the QRS complex, and the
A cardiac cycle consists of one heartbeat or one PQRST T wave deflections.
sequence. It represents a sequence of atrial contraction
and relaxation followed by ventricular contraction and
relaxation. The basic cycle repeats itself again and again
Waveforms and current flow
(Figure 2-4). Regularity of the cardiac rhythm can be A monitor lead, or ECG lead, provides a view of the heart’s
assessed by measuring from one heartbeat to the next electrical activity between two points or poles (a positive
(from one R wave to the next R wave, also called the R-R pole and a negative pole). The direction in which the elec-
interval). Between cardiac cycles, the monitor or ECG tric current flows determines how the waveforms appear
recorder returns to the isoelectric line (baseline), the flat on the ECG tracing (Figure 2-6). An electric current flow-
line in the ECG during which electrical activity is absent ing toward the positive pole will produce a positive deflec-
(Figure 2-5). Any waveform above the isoelectric line is tion; an electric current traveling toward the negative pole
considered a positive (upright) deflection and any wave- produces a negative deflection. Current flowing away from
form below this line a negative (downward) deflection. the poles will produce a biphasic deflection (both positive
A deflection having both a positive and negative compo- and negative). Biphasic deflections may be equally positive
nent is called a biphasic deflection. This basic concept and negative, more negative than positive, or more positive
than negative (depending on the angle of current flow to
the positive or negative pole).
R R The size of the wave deflection depends on the magni-
tude of the electrical current flowing toward the individual
pole. The magnitude of the electrical current is determined
P T P T by how much voltage is generated by depolarization of a par-
ticular portion of the heart. The QRS complex is normally
Q S Q larger than the P wave because depolarization of the larger
S
Isoelectric line muscle mass of the ventricles generates more voltage than
does depolarization of the smaller muscle mass of the atria.

A Refractory and supernormal


periods of the cardiac cycle
Positive deflection Negative deflection Biphasic deflection There is a period of time in the cardiac cycle during which
Figure 2-5. Relationship between waveforms and the isoelectric the cardiac cells may be refractory, or unable to respond,
line. to a stimulus. Refractoriness is divided into three phases
(Figure 2-7):

ECG workout_Chap02.indd 11 4/28/2011 1:09:31 AM


12 Electrophysiology

QRS complex

P wave T wave supernormal


period

absolute relative
refractory refractory
period period
Figure 2-7. Refractory and supernormal periods.

 Absolute refractory period — During this period the  Supernormal period — During this period the cardiac
cells absolutely cannot respond to a stimulus. This period cells will respond to a weaker than normal stimulus. This
extends from the onset of the QRS complex to the peak of period occurs during a short portion near the end of the
the T wave. During this time the cardiac cells have depolar- T wave, just before the cells have completely repolarized.
ized and are in the process of repolarizing. Because the car-
diac cells have not repolarized to their threshold potential
(the level at which a cell must be repolarized before it can
ECG graph paper
be depolarized again) they cannot be stimulated to depolar- The PQRST sequence is recorded on special graph paper
ize. In other words, the myocardial cells cannot contract, made up of horizontal and vertical lines (Figure 2-8). The
and the cells of the electrical conduction system cannot horizontal lines measure the duration of the waveforms in
conduct an electrical impulse during the absolute refrac- seconds of time. Each small square measured horizontally
tory period. represents 0.04 second in time. The width of the QRS com-
 Relative refractory period — During this period the plex in Figure 2-9 extends across for 2 small squares and
cardiac cells have repolarized sufficiently to respond to represents 0.08 second (0.04 second × 2 squares). The ver-
a strong stimulus. This period begins at the peak of the tical lines measure the voltage or amplitude of the wave-
T wave and ends with the end of the T wave. The relative form in millimeters (mm). Each small square measured
refractory period is also called the vulnerable period of vertically represents 1 mm in height. The height of the
repolarization. A strong stimulus occurring during the QRS complex in Figure 2-9 extends upward from baseline
vulnerable period may usurp the primary pacemaker of 16 small squares and represents 16 mm voltage (1 mm ×
the heart (usually the SA node) and take over pacemaker 16 squares).
control. An example might be a premature ventricular con-
traction (PVC) that falls during the vulnerable period and
takes over control of the heart in the form of ventricular 1

tachycardia.
I
ir- is T ~
:
3
-
. . 1
4
s
\ e -1 5
tffl
Voltage e
\ \ E I : r:
\ £ \ -
V

\
Lf> H H- 3 f
Ml
S I
I 0.04 ; .I t -
- IS
I
I
second
1
n

mam
I i
I
I
0.20 second i
L

Time t

Figure 2-8. Electrocardiographic paper. Figure 2-9. QRS width: 0.08 second; QRS height: 16 mm.
3 Waveforms, intervals,
segments, and
complexes
Much of the information that the ECG tracing provides is and peaked. The abnormal P wave in right atrial enlarge-
obtained from the examination of the three principal wave- ment is sometimes referred to as p pulmonale because the
forms (the P wave, the QRS complex, and the T wave) and atrial enlargement that it signifies is common with severe
their associated segments and intervals. Assessment of this pulmonary disease (for example, pulmonary stenosis and
data provides the facts necessary for an accurate cardiac insufficiency, chronic obstructive pulmonary disease,
rhythm interpretation. acute pulmonary embolism, and pulmonary edema).
Impulses traveling through an enlarged left atrium (left
atrial hypertrophy) result in P waves that are wide and
P wave notched. The term p mitrale is used to describe the abnormal
The first deflection of the cardiac cycle, the P wave, P waves seen in left atrial enlargement because they were first
is caused by depolarization of the right and left atria seen in patients with mitral valve stenosis and insufficiency.
(Figure 3-1). The first part of the P wave represents depo- Left atrial enlargement can also be seen in left heart failure.
larization of the right atrium; the second part represents  Ectopic P wave — The term ectopic means away from its
depolarization of the left atrium. The waveform begins as normal location. Therefore, an ectopic P wave arises from a
the deflection leaves baseline and ends when the deflection site other than the SA node. Abnormal sites include the atria
returns to baseline. A normal sinus P wave originates in and the AV junction. P waves from the atria may be positive
the sinus node and travels through normal atria, resulting or negative; some are small, pointed, flat, wavy, or sawtooth
in normal depolarization. Normal P waves are smooth and in appearance. P waves from the AV junction are always neg-
round, positive in lead II (a positive lead), 0.5 to 2.5 mm ative (inverted) and may precede or follow the QRS complex
in height, 0.10 second or less in width, with one P wave or be hidden within the QRS complex and not visible.
to each QRS complex. More than one P wave before a Examples of P waves are shown in Figure 3-2.
QRS complex indicates a conduction disturbance, such as
that which occurs in second and third-degree heart block PR interval
(discussed in Chapter 8).
There are two types of abnormal P waves: The PR interval (sometimes abbreviated PRI) represents
 Abnormal sinus P wave — An abnormal sinus P wave the time from the onset of atrial depolarization to the onset
originates in the sinus node and travels through enlarged of ventricular depolarization. The PR interval (Figure 3-3)
atria, resulting in abnormal depolarization of the atria. includes a P wave and the short isoelectric line (PR seg-
Abnormal atria depolarization results in abnormal-looking ment) that follows it. The PR interval is measured from the
P waves. beginning of the P wave as it leaves baseline to the begin-
Impulses traveling through an enlarged right atrium ning of the QRS complex. The duration of the normal PR
(right atrial hypertrophy) result in P waves that are tall interval is 0.12 to 0.20 seconds.
Abnormal PR intervals may be short or prolonged:
 Short PR interval — A short PR interval is less than
0.12 seconds and may be seen if the electrical impulse

I
originates in an ectopic site in the AV junction. A short-
ened PR interval may also occur if the electrical impulse
progresses from the atria to the ventricles through one
of several abnormal conduction pathways (called acces-
sory pathways) that bypass a part or all of the AV node.
Wolff-Parkinson-White syndrome (WPW) is an example of
such an accessory pathway.
 Prolonged PR interval — A prolonged PR interval is
Figure 3-1. The P wave. greater than 0.20 seconds and indicates that the impulse

13

ECG workout_Chap03.indd 13 4/28/2011 5:23:15 AM


14 Waveforms, intervals, segments, and complexes

p
;3 P
't

r
Eit : ; '
rrr
r ai - 1?
7 » IH
y J - i^xt* S
IP
= ;: j
r

.
W

I R.

4'
+
,i l
r.
s: i 1_. f
* - .i t± ± nh r Tt :
T? r
r r•>
- - i; a x
H I? I
:: t T
S55E
T

t ::: ?
:: • X-
tii
*

-»- - —
u £
r*
SHSHE h -
; :
a - t- P
1

i I X ::- T r ::; S3: *


-ii i t::
1
-Jfl- ^«
r -: - 4 :
: IX - ; i =t FT TJ
i
1 E 4 **

A Normal P wave B Inverted P wave


£ T F +P M

If rr I
ill!
a M
KJ
+
«
£
c ::; : E £ -
r 4- 4 EH? U t. : - n si;
C No visible P waves D Two P waves to each QRS

iiii 111
E Wide, notched P wave F Tall, peaked P wave

Ii;
ill
m r -
Small, pointed
G Flat P wave H P wave I Sawtooth P waves J Wavy P waves

Figure 3-2. P wave examples.

ECG workout_Chap03.indd 14 4/28/2011 5:23:15 AM


QRS complex 15

J point

Figure 3-3. The PR interval.


Figure 3-5. The QRS complex.

was delayed longer than normal in the AV node. Prolonged


PR intervals are seen in first-degree AV block. Finding the beginning of the QRS complex usually isn’t
Examples of PR intervals are shown in Figure 3-4. difficult. Finding the end of the QRS complex, however,
is at times a challenge because of elevation or depres-
sion of the ST segment. Remember, the QRS complex
QRS complex ends as soon as the straight line of the ST segment
The QRS complex (Figure 3-5) represents depolarization begins, even though the straight line may be above or
of the right and left ventricles. The QRS complex is larger below baseline.
than the P wave because depolarization of the ventricles Although the term QRS complex is used, not every QRS
involves a larger muscle mass than depolarization of the complex contains a Q wave, R wave, and S wave. Many
atria. variations exist in the configuration of the QRS complex
The QRS complex is composed of three wave deflec- (Figure 3-6). Whatever the variation, the complex is still
tions: the Q wave, the R wave, and the S wave. The R wave called the QRS complex. For example, you might see
is a positive waveform; the Q wave is a negative waveform a QRS complex with a Q and an R wave, but no S wave
that precedes the R wave; the S wave is a negative wave- (Figure 3-6, example B), an R and S wave without a Q wave
form that follows the R wave. The normal QRS complex (Figure 3-6, example C), or an R wave without a Q or an
is predominantly positive in lead II (a positive lead) with a S wave (Figure 3-6, example D). If the entire complex is
duration of 0.10 second or less. negative (Figure 3-6, example F), it is termed a QS com-
The QRS complex is measured from the beginning plex (not a negative R wave because R waves are always
of the QRS complex (as the first wave of the complex positive). It’s also possible to have more than one R
leaves baseline) to the end of the QRS complex (when wave (Figure 3-6, example I) and more than one S wave;
the last wave of the complex begins to level out into the (Figure 3-6, example J). The second R wave is called R prime
ST segment). The point where the QRS complex meets and is written R'. The second S wave is called S prime and
the ST segment is called the J point (junction point). is written S'. To be labeled separately, a wave must cross

U-)
T - +
. .. .
j
i 4- V • T

12= ^ -

t -*
• •
: IL :: r : r •:
1
c
I L: 1? • r
.
s
11
: .
i: U hi

ft r = mE _= ' :- f c;

— r- - ::
C
A B
Normal PR interval of 0.20 Short PR interval Long PR interval of 0.38
second (0.04 second ´ 5 of 0.08 second second (0.04 second ×
squares). (0.04 second ´ 9½ squares)
2 squares) Figure 3-4. PR interval examples.

ECG workout_Chap03.indd 15 4/28/2011 5:23:17 AM


16 Waveforms, intervals, segments, and complexes

A+\
R R R the baseline. A wave that changes direction but doesn’t
cross the baseline is called a notch. (Figure 3-6, example E,
shows a notched R and Figure 3-6, example K, shows a
A B C notched S.)
q q
S S Capital letters are used to designate waves of large
amplitude (5 mm or more) and lowercase letters are used
R Notched R to designate waves of small amplitude (less than 5 mm).
This allows you to visualize a complex mentioned in a
textbook when illustrations aren’t available. For example,
D E F
if a complex is described in a text as having an rS wave-
form, the reader can easily picture a complex with a small
QS r wave and a big S wave.
R′ An abnormal QRS complex is wide with a duration of

tf
r r r 0.12 second or more. An abnormally wide QRS complex
G
i Q
~ H I

S
may result from:
 a block in the conduction of impulses through the right
or left bundle branch (bundle-branch block)
S  an electrical impulse that has arrived early (as with pre-
R′
mature beats) at the bundle branches before repolariza-
tion is complete, allowing the electrical impulse to initiate
r r depolarization of the ventricles earlier than usual, result-
J K ing in abnormal (aberrant) ventricular conduction and
causing a wide QRS complex
S S′  an electrical impulse that has been conducted from
Notched S the atria to the ventricles through an abnormal accessory
Figure 3-6. QRS variations. conduction pathway that bypasses the AV node, allow-
ing the electrical impulse to initiate depolarization of

0.12 second 0.08 second 0.04 second


(3 squares x 0.04 second) (2 squares x 0.04 second) (1 square x 0.04 second)

0.10 second 0.06 second 0.08 second


(21/2 squares x 0.04 second) (11/2 squares x 0.04 second) (2 squares x 0.04 second)

Figure 3-7. QRS examples.

ECG workout_Chap03.indd 16 4/28/2011 5:23:18 AM


ST segment 17
WM
ill
0.10 second 0.08 second 0.18 second
(21/2 squares x 0.04 second) (2 squares x 0.04 second) (41/2 squares x 0.04 second)
1
::i n
t :i
I
:: l
tSE!S!!!Stj3SS!!S ?
'
-*- l - - :
TT :J i :: :.r ,
T r 7
11 :
i* ^F

r
rf

111:13
uLiiutEHEEEu:
. :

-
it
1 iliLJ

itii
IlIKHIHUEMUli: atiliLiiiii JlliiiiE
!!!»!!!»!!!£!!! .i
(!
a
. r
Jstf lit: t
\\ rtt
0.12 second 0.08 second 0.16 second
(3 squares x 0.04 second) (2 squares x 0.04 second) (4 squares x 0.04 second)
Figure 3-7. (continued)
the ventricles earlier than usual, resulting in abnormal ment may be displaced above baseline (elevated ST seg-
(aberrant) ventricular conduction and causing a wide QRS ment) or below baseline (depressed ST segment). The PR
complex segment is normally used as a baseline reference to evalu-
 an electrical impulse that has originated in an ectopic ate the degree of displacement of the ST segment from
site in the ventricles. the isoelectric line. An ST segment is abnormal when it is
Examples of QRS complexes are shown in Figure 3-7. elevated or depressed 1 mm or more, measured at a point
0.04 second past the J point (the point where the QRS com-
plex and the ST segment meet).
ST segment Elevated ST segments may be horizontal (straight
The ST segment represents early ventricular repolariza- across), convex (rounded upward), or concave (rounded
tion. The ST segment is the flat line between the QRS com- inward). Common causes include ST elevation myocardial
plex and the T wave (Figure 3-8). Normally the ST segment infarction (STEMI), coronary artery spasm (Prinzmetal’s
is positioned at baseline (the isoelectric line). The ST seg- angina), acute pericarditis, ventricular aneurysm, early
repolarization pattern (a form of myocardial repolariza-
tion seen in normal healthy individuals that produces
ST-segment elevation closely mimicking that of acute
myocardial infarction [MI] or pericarditis), hyperkalemia,
and hypothermia.
J point Depressed ST segments may be horizontal, downslop-
ing, upsloping, or sagging. Common causes include
myocardial ischemia, non-ST elevation MI (non-
STEMI), reciprocal ECG changes associated with STEMI,
hypokalemia, and digitalis effect. Digitalis causes a sagging
ST-segment depression, with a characteristic “scooped-
Figure 3-8. The ST segment. out” appearance. Examples of ST segments are shown in
Figure 3-9.
18 Waveforms, intervals, segments, and complexes

t
£ --
It : !: :
rrri n
rrmr £irrui
: ;; :
•i-i-*-
— -—ri
i:|
t v

- --n
- T fi
k

: z 11
T
L L i i-4 4 a
LI
- t
i
i - t
n

ri

rrr .. ±5 :1 ; :
- ^
. .- r

r -r: +
i :
1 •
ft
iitiilvivSSSlMHMlfP
4 BfjB BB 3 B B B <1 B H H 44 : 4. JL
'

6
1

c
A a
r T- itrr r ir it r :
. i . i . ..
_
u iriii i t T. r :
P- 7
4 44 LLLi M

- u-
A Normal ST segment B
d a rr: G
Normal ST segment
h4
4
44 4
t
r!
— t
4 rr r
. i
4 4

I- II
4 4*
iL
r — 4 rt

is 3
4 4 .
1:
1
lllli: :&
i r f -V 1 -
r:

t:
r ;:n
i f

m ..
M - it TI ! 4 LJ

[
- - r-
-.
r
1
4. :
rr
i
f
*

ti
I-:3
±
:S £
1:
l 1 --
• r
• -
. ri
4 I .1? r —- i
r: i i I-
,

ir llMi) : -
=T 4 in dri-CT a tl
•r
n I 4-
44
I ff

iJ
T
r::
-
Ft !
orr
tH

44-
j
-
-- 4
lr~ r4- t -
r
4
4

t-

i

+
J
I
a-
'

t
L

- :;z
I
-

r
4

E 4
J - 4
r
tl
4
-- -
-TL.
N
4
:
IT if . 14
h
1!
.

4T - N.
4! t
: r.xit t
[ IT U!
-

T. 1-
TXT
-:tr
!

=
r
"
4n

4!
tf
,
= rz :: z

::
-
+- I

:
r

St :
4

4.1:
MI
i-
-mTfi : t m
H

fi
-*-1-1
!

.:.ri. :
m EEffll
rr
444 fill
ttt
ui : 4
*
rirr '
;* 1
44:
t

C Convex elevation
t - I
- *
i
J;
:: t
4

m D Concave elevation
tr
1
T
- rt: TtT-u

ulR
: .L :: t
i: : M : :i
=Htz= ri Mr:
J L ••
-
i ii lit
4ii4- 4i4
1 k
1: i rrrt rrr: rnr
- -- .. rrr !
rr :r ;
-
» J
44

1
3 rtn
44
n si- - r *
r. f -
-
: :

; a rn:-
ri:
::
TL
rm
4
4

3
1: 4;
rrr;
I
rm •
k
j
.
. 1: ::- rrr St
1 II 4
4 HT; 3! 4

HBI
rrn rrr:

IllilMII
4: . MT
rrtr
n . rr
«6 mffl Si
t
T r: 41 Li.44: ;
1

4 - n 4: r1
MM
44
fe W
rr:
-
'4
Hr • : 1 :i :: 1 ::
r
1!
H
r;
r
rrr
-: i-:::P
:r
4
a
111
L4TT1
- rr
1
r

Pr
i in4
. 1
H
•T
. ; -ill Bimm t
"
: r~
if r
I
RBrr; H -- I 1
+
r
ri 1
L:

rr 7 1 it ; - -
11
rM
E Horizontal elevation Horizontal depression

7.11 4i± :: :ss


2
r T . T -.
mwru
TTTT '
£
tl
4
ism
imi
;s;
vmii

111 ittrtlit :::: mil ri


::: ::::::::::: nui


K UB

-
M

mm
L:
mmm
«

t
r
M
ET ===::
S
iiBi

RS

SI 3
MB 5*1
s: :i ...J
Mr
'IMNI

1: i
; -
i TTTJ
1: 1::i rr E::U inn
rL 4r r r -rrt Sii
mu
H- as ±
MB
- M
nmmmm

I
MB

5! - T I 4
1 saaa
Mil
II

m
1- :l
TZ
r
IBM!
MB
BBBII
BMB £
:as
11-L 4
^ nnrtTiTTnrTTnm ::J: . * MMI
BIBIB

14 L H - -S
a
IT. i r j
rrrrr rtt
.i . i: 4 HUE lit .Tilt
.
1
-
Hi:
i 4

iltttlh: :
tz Li
ft 1
H f
G Downsloping depression H Sagging depression

Figure 3-9. ST segment samples.

ECG workout_Chap03.indd 18 4/28/2011 5:23:19 AM


T wave 19

T wave
The T wave represents ventricular repolarization. The nor-
mal T wave begins as the deflection gradually slopes upward
from the ST segment, and ends when the waveform returns
to baseline (Figure 3-10). Normal T waves are rounded and
slightly asymmetrical (with the first part of the T wave grad-
ually sloping to the peak and returning more abruptly to
baseline), positive in lead II (a positive lead), with an ampli-
Figure 3-10. The T wave. tude less than 5 mm. The T wave always follows the QRS
complex (repolarization always follows depolarization).

rrn tr:: 5!
m n: - 4«J .
: :: .»
-
m *
: ^ ' i
ft *- : itrfil
rrrr 4 44
rrn
* 1
r.
&
- iiif n
\y :+ 44 44 i
rrrr rT T
~ — pr
*i tt
E
tt
n
9
: ]?•
r rr fT

-- a
ill;
t I
P H 4
s
Ifili
7* T

53 rrn u trir 1

SI
r;:i . n;- j n ; ;
_
r H
:.; ; rrrt
Jrrrr :t :TS
rr: P •
u fH P4
: I M
H
Ar *

iw., I
Sr .: il i i • : r: F
rrn rrr:
£1 - Hi:: ip ! L:
::: - •i iT ::
* . : n:
•t :: P P f ••

tr ±r+r tn -
JriJT
: ; i: rm S3 rrr:
id lip : :: :
r: rr i a ::
r:: : :
: in.: -4 , 4 4 r 4
-> w P
rr: rH . . : •• >
h 44
: i : n : :: l:
*4
r.:.L .4
: .
4n -U . J

. -. m
1' T .
^m« rtT tUi
rr. : !ni I : rrr :
it 4
T: : r ; ::
: .
n -4 i
H
ti: .
. .J M p:
l •
4i 4.:::

r* i t ' •
••
t
: ::: 5
"
r•

A Normal T wave B Biphasic T wave

i h r mmmmM
+
:vz ~
MM
3
ir
rt
T MBS
MM
lassi
- tr
f
£ n
Pi
L
- iEn
sma
i «::
|
:
j :J.: JLL: IUI: Mill

rr 1 r i : M l 4
7: [Hi : : BBI
; jijri '|
jiagujoc mpj. m|j{uiji| j j][ijji|jjjriiDujE
i|| smu
; i: T
^-
4 r
HUl
nm
a ii rrn
iiii 7 •
i me
44 u. H -
3dttH
..
ii I inn i
4
n
t
»11 i:; : it •-

M4
r
r.
:n: :r m rr. - it n : ::
. r .
rrt
!
=
t
r+rl 7 F bsi rrr
iii |3|j{Ejmjjj|jiiBSgjjE|ngj{jDijio)isgi EH nns r• •7 P ? i i - H
r

C Tall, peaked T wave D Inverted T wave

t
T r
n n

J r: rzsrxassx
: t
- j- .
!
t:

r r 51
m
i i
I
lilliygR
E Flat T wave

Figure 3-11. T wave examples.

ECG workout_Chap03.indd 19 4/28/2011 5:23:22 AM


20 Waveforms, intervals, segments, and complexes
Abnormal T waves may be abnormally tall or low, flattened,
biphasic, or inverted. Common causes include myocardial
ischemia, acute MI, pericarditis, hyperkalemia, ventricular
enlargement, bundle-branch block, and subarachnoid hem-
orrhage. Significant cerebral disease, such as subarachnoid
hemorrhage, may be associated with deeply inverted T waves

I
(called cerebral T waves).

"
Examples of T waves are shown in Figure 3-11.
QT interval Figure 3-12. QT Interval.
The QT interval represents the time between the onset of
ventricular depolarization and the end of ventricular repo- Duration of the QT interval can be determined by multiply-
larization. The QT interval is measured from the beginning ing the number of small squares in the QT interval by 0.04
of the QRS complex to the end of the T wave (Figure 3-12). second (Figure 3-13). The length of the QT interval normally
l : i • I - 1
t Hi 5
5 t
1
J: ..: n
:rrn
t-
fie i :
:
.
1 4 -
H
* :::t
t
i

s t ; r: irr : ;
.1 M
V •• •
i •

m m& m i l
i u
n fi
t:
i: rr T:
:
+
r. - n rr rr
<\ I
iii : H
i


:
: :: : . J i I Zl I rf
--
A 1. Number of small squares between R waves = 31. Half of B 1. Number of small squares between R waves = 38. Half of
31 = 15. 38 = 19.
2. Number of small squares in QT interval = 11 2. Number of small squares in QT interval = 13
3. Compare the difference: QT interval is less than half the 3. Compare the difference: QT interval is less than half the
R-R interval (11 small squares are less than 15 small R-R interval (13 small squares are less than 19 small
squares); QT interval is normal for this heart rate. squares); QT interval is normal for this heart rate.
(Duration of QT interval: 11 squares ´ 0.04 second = 0.44 (Duration of QT interval: 13 small squares ´ 0.04
second.) second = 0.52 second.)
i :
'
1 :: i l l! ! !
.
! ! ::: U -
i ' I * i I !
1 :
M i! i ; . :
- I !* ! : '
' I!i '
i . i ! •*
>
:: ':
i 1
' L .: JJ : L i .1
• 1 i
- i :; . .
£
A

-
L
r
i.-:
.1 i M S!• .!
1
:: . :: .• ;.
i i!

I 1 i
N -
. ; ! , :i ,!
i
itJ
r ?n tirprawM m n ?rrn
C 1. Number of small squares between R waves = 18. Half of
18 = 9.
2. Number of small squares in QT interval = 13.
3. Compare the difference: QT interval is more than half the
R-R interval (13 small squares are more than 9 small
squares); QT interval is prolonged for this heart rate.
(Duration of QT interval: 13 squares ´ 0.04 second =
0.52 second.)
Figure 3-13. QT interval examples.
U wave 21

J[
varies according to age, sex, and particularly heart rate. The
QT interval is more prolonged with slow heart rates.
Generally speaking, the normal QT interval should be
less than half the R-R interval (the distance between two
consecutive R waves) when the rhythm is regular. The
determination of the QT interval should be made in a lead
where the T wave is most prominent and shouldn’t include
the U wave. Accurate measurement of the QT interval can
be done only when the rhythm is regular for at least two
cardiac cycles before the measurement.
~
^
Figure 3-14. The U wave.
To determine if the QT interval is normal or prolonged:
 Count the number of small boxes in the R-R interval
and divide by two. ventricular tachycardia (discussed in Chapter 9). Com-
 Count the number of small boxes in the QT interval. mon causes include electrolyte imbalances (hypokalemia,
 Compare the difference. If the QT interval measures less hypomagnesemia, hypocalcemia), hypothermia, brady-
than half the R-R interval, it’s probably normal. If the QT arrhythmias, liquid protein diets, myocardial ischemia,
interval measures the same as half the R-R interval, it’s antiarrhythmics, psychotropic agents (phenothiazines,
considered borderline. If the QT interval measures longer tricyclic antidepressants), and hereditary long-QT syn-
than half the R-R interval, it’s prolonged. drome. It can also occur without a known cause (idiopathic).
A prolonged QT interval indicates a delay in ventricular Examples of QT intervals are shown in Figure 3-13.
repolarization. The prolongation of the QT interval length-
ens the relative refractory period (the vulnerable period
of repolarization), allowing more time for an ectopic
U wave
focus to take control and putting the ventricles at risk for The U wave is a small deflection sometimes seen follow-
life-threatening arrhythmias such as torsades de pointes ing the T wave (Figure 3-14). Neither its presence nor its

Ijjjjj!:::::!::::!

ECG without U wave

w . ..
.i . H
1 r?
: '

I; til E
: : :;:

S FM
mmm
llfl

raa
I T
MS
III
MB P r i- 4
3J

itr t: -
*•
MB
H. L r+ t
M
rt : - ::1
r ti t 4

Siif f ::
1 * ! » - am ? 1 ^i T

ECG with U wave


Figure 3-15. U wave examples.

ECG workout_Chap03.indd 21 4/28/2011 5:23:26 AM


22 Waveforms, intervals, segments, and complexes

absence is considered abnormal. The U wave represents late


repolarization of the ventricles, probably a small segment
of the ventricles.
The waveform begins as the deflection leaves baseline
and ends when the deflection returns to baseline. Normal U
waves are small, rounded, and symmetrical, positive in lead
II (a positive lead), and 2 mm or less in amplitude (always
smaller than the preceding T wave). The U wave can best be
seen when the heart rate is slow.
Abnormal U waves are tall (greater than 2 mm in
height). Common causes include hypokalemia, cardio-
myopathy, and left ventricular enlargement, among other
causes. A large U wave may occasionally be mistaken for
a P wave, but usually a comparison of the morphology of
both waveforms will help differentiate the U wave from the
P wave.
Examples of U waves are shown in Figure 3-15.

ECG workout_Chap03.indd 22 4/28/2011 5:23:27 AM


Waveform practice: Labeling waves 23

Waveform practice: Labeling waves


For each of the following rhythm strips (strips 3-1 through 3-14), label the P, Q, R, S, T, and U waves. Some of the strips
may not have all of these waveforms. Check your answers with the answer key in the back of the book.

a H

4 -:. -J -

K:
RE
i

r:

Strip 3-1. Strip 3-2.

, . 4. itrtiti : t+ r i- *
- 4 .
H
fc: d —
8 [
i

a t aa::: :::ri:i-:-r - === Wtl I

. .
T - ~

an : ±m
*
- - V
<
M
a:
1« ::
i
S* - E3
i: M P 4-
H
:: -T h -
r -?
=
: n : r :::::::

i
: : : :: z ± ± i

r- :
>
t r -

H- — Hr — t t - -4- H
"

Strip 3-3. Strip 3-4.

ESSE
BMBB
c
! t ••

im:
flHI
H t- -
i :s:sn: -
1:'

=* 4^
1

E I!

—-
BBBB
H •
:: c u T 1J

P
12»
;
n 3 r
h u 3
. - T
r *-
BMl

;;su
» a H-

M
tt
-f 4:

:s K
4* ft -4
i
• V
1

BHKB 12
aseasnli :: § i
i E aaaaa -, am *

-
i ii i!
l » i B P B i a a l B B H B B i i a a a a i a i B B B M B M|i d

::Ru::;i EEES:ERS:S::;R:K]EEBIK::!EI5:::::::!:!
I
i .:. . r TT.
3+
r I:
:i
I :4 ‘

L!
f t :x

:rs:i I-
i 3rt t "
tl
T h' 7
L

R
4
imi

.:
BBBB
BMBB r

r
t: >
-rtr TH : ; 4P ! - 4' 443
Strip 3-5. Strip 3-6.

ECG workout_Chap03.indd 23 4/28/2011 5:23:27 AM


24 Waveforms, intervals, segments, and complexes

Strip 3-7. Strip 3-8.

Strip 3-9. Strip 3-10.

f• - H +
4 ::
:i

! r :: r r :L . .
-•
4
I - -- -
i -i
F: 4
n
il 4

m 44-
4
& r
t —
mm
"Riivaii!

t T -»
4; 1 r
4
4
irri
H -
t
i
r +
I :
i . *-
.. ‘U

^ i-
r Li
*
i

: =x
Strip 3-11. Strip 3-12.

m- m m::: m J
::
a !
-
I Ml : -U J

14
3
E
r
h
r
M
zrq

i=
H- ± + p: m\
a
Bo
M 1W
+ IF
OB
—tttrrrrrn Li : ::
Ei
: * t

ft
-- 1 *

: 40: tr rrtr:
TT - -
: .... if n ••
M
L
FHiS 4
ZW.z ' i
: siM h ir -
r
; sSIVr.:
*

;; iri 4 : 44 d3p z O\
TO: :: r
F F “
O TO
t -t • 11 i
:: 4 LHT M
:i: i +1 t

2.
on
: ;•

till
n ::
- •

t:a .n :
• rr T - r •”
± ±L :112

—— -
Ti
is
U
-IF 4 iri 44p OlSon ? 4
H
!: On
rt

r- i
nmLilli

M
r - !!
H -H
if E 3Ei
rr jit:
:::
: :: : :
rt
MI; -
••
-
1 ::M:
::: t
:
£
::
r - r .444

-r +
Ip
FO :
4
E
— TT
- - —a m OF

Strip 3-13. Strip 3-14.

ECG workout_Chap03.indd 24 4/28/2011 5:23:29 AM


4
Purpose of ECG monitoring
Cardiac monitors

The electrocardiogram (ECG) is a recording of the electrical


activity of the heart. The ECG records two basic electrical /? %
processes: RA LA
 Depolarization — the spread of the electrical stimulus
through the heart muscle, producing the P wave from the
atria and the QRS complex from the ventricles. C
 Repolarization — the recovery of the stimulated mus-
cle to the resting state, producing the ST segment, the T A \

wave, and the U wave. / a \


The depolarization-repolarization process produces RL LL
electrical currents that are transmitted to the surface of
the body. This electrical activity is detected by electrodes
attached to the skin. After the electric current is detected, Figure 4-1. Hardwire monitoring — Five leadwire system.
it’s amplified, displayed on a monitor screen (oscilloscope), This illustration shows you where to place the electrodes and
and recorded on ECG graph paper as waves and complexes. attach leadwires using a five-leadwire system. The leadwires are
The waveforms can then be analyzed in a systematic man- color-coded as follows:
ner and the “cardiac rhythm” identified.  white — right arm (RA)
Bedside monitoring allows continuous observation  black — left arm (LA)
of the heart’s electrical activity and is used to identify  green — right leg (RL)
arrhythmias (disturbances in rate, rhythm, or conduction),  red — left leg (LL)
evaluate pacemaker function, and evaluate the response  brown — chest (C).
to medications (for example, antiarrhythmics). Continu- Leads placed in the arm and leg positions as shown allow you
ous cardiac monitoring is useful in monitoring patients in to view leads I, II, III, aVR, aVL, and aVF. To view chest leads V1–V6,
critical care units, cardiac stepdown units, surgery suites, the chest lead must be placed in the specific chest lead position
outpatient surgery departments, emergency departments, desired. In this example, the brown chest lead is in V1 position.
and postanesthesia recovery units.

midclavicular line), one below the left clavicle (2nd inter-


Types of ECG monitoring space, left midclavicular line), one on the right lower rib
There are two types of ECG monitoring: hardwire and cage (8th interspace, right midclavicular line), one on the
telemetry. With hardwire monitoring (bedside monitor- left lower rib cage (8th interspace, left midclavicular line),
ing), electrode pads (conductive gel discs) are placed and one in a chest lead position (V1 to V6). The six chest lead
on the patient’s chest and attached to a lead-cable sys- positions (Figure 4-2) include:
tem and then connected to a monitor at the bedside.  V1 – 4th intercostal space, right sternal border
With telemetry monitoring (portable monitoring), elec-  V2 – 4th intercostal space, left sternal border
trode pads are attached to the patient’s chest and con-  V3 – midway between V2 and V4
nected to leads that are attached to a portable monitor  V4 – 5th intercostal space, left midclavicular line
transmitter.  V5 – 5th intercostal space, left anterior axillary line
 Hardwire monitoring — Hardwire monitoring uses  V6 – 5th intercostal space, left midaxillary line
either a five-leadwire system or a three-leadwire system. The right arm (RA) lead is attached to the electrode pad
With the five-leadwire system (Figure 4-1), five elec- below the right clavicle; the left arm (LA) lead to the elec-
trode pads and five leadwires are used. One electrode trode pad below the left clavicle; the right leg (RL) lead
is placed below the right clavicle (2nd interspace, right to the electrode pad on the right lower rib cage; the left

25

ECG workout_Chap04.indd 25 4/29/2011 4:31:36 PM


26 Cardiac monitors

I y
/ ft % V2
RA LA

V1 t \ V3

\ a. V6
\

/ a V4 / a \
LL
V5

Figure 4-2. Chest lead positions. Figure 4-3. Hardwire monitoring — Three-leadwire system.
This illustration shows you where to place the electrodes and attach
leadwires using a three-leadwire system. The lead wires are color-
coded as follows:
leg (LL) lead to the electrode pad on the left lower rib cage;
 white — right arm (RA)
and the chest lead to the electrode pad of the specific chest
 black — left arm (LA)
position desired (V1 through V6).
 red — left leg (LL).
With the five-leadwire system for hardwire monitor-
ing, you can continuously monitor two leads using a Leads placed in this position will allow you to monitor leads I,
lead selector on the monitor. Leads placed in the arm II, or III using the lead selector on the monitor.
and leg positions allow you to view leads I, II, III, AVR,
AVL, and AVF (Figure 4-1). To view chest lead V1 to V6, the LA lead is attached to the electrode pad below the left
the chest lead must be placed in the specific chest lead clavicle, and the LL lead is attached to the electrode pad on
position desired. Generally, a limb lead (usually I, II, or the left lower rib cage. You can monitor either limb leads
III) and a chest lead (usually V1 or V6) are chosen to be I, II, or III by turning the lead selector on the monitor.
monitored. Although you can’t monitor chest leads (V1 to V6) with a
With the three-leadwire system (Figure 4-3), three elec- three-leadwire system, you can monitor modified chest
trode pads and three leadwires are used. One electrode pad leads that provide similar information. To monitor any of
is placed below the right clavicle (2nd interspace, right these leads, reposition the LL lead to the appropriate posi-
midclavicular line), one below the left clavicle (2nd inter- tion for the chest lead you want to monitor, and turn the
space, left midclavicular line), and one on the left lower rib lead selector on the monitor to lead III. Examples of modi-
cage (8th interspace, left midclavicular line). The RA lead fied chest lead V1 (MCL1) and modified chest lead V6 (MCL6)
is attached to the electrode pad below the right clavicle, are shown in Figure 4-4.

y
/?-
RA LA RA
^ LA

LL \
ib A <S
V

LL
\ & \

/ a \ / ft \

Modified Chest Lead V1 (MCL1) Modified Chest Lead V6 (MCL6)

Figure 4-4. Hardwire monitoring — Three-leadwire system: Leads MCL1 and MCL6. Modified chest leads can be monitored with the three-
leadwire system by repositioning the left leg (LL) lead to the chest position desired and turning the lead selector on the monitor to lead III.

ECG workout_Chap04.indd 26 4/29/2011 4:31:36 PM


Troubleshooting monitor problems 27

) monitored at a time, and a lead selector on the monitor


/cL % isn’t available.

RA LA Applying electrode pads


Proper attachment of the electrode pads to the skin is the
C most important step in obtaining a good quality ECG trac-
ing. Unless there is good contact between the skin and the
\
\ electrode pad, distortions of the ECG tracing (artifacts)
/ 0 N may appear. An artifact is any abnormal wave, spike, or
RL LL movement on the ECG tracing that isn’t generated by the
electrical activity of the heart. The procedure for attaching
the electrodes is as follows:
Figure 4-5. Telemetry monitoring — Five-leadwire system.  Choose monitor lead position. It’s helpful to assess the
This illustration shows you where to place the electrodes and 12-lead ECG to ascertain which lead provides the best QRS
attach leadwires using a five-leadwire system. The leadwires are complex voltage and P wave identification.
color-coded as follows:
 Prepare the skin. Clip the hair from the skin using a
 white — right arm (RA)
clipper; hair interferes with good contact between the
 black — left arm (LA)
electrode pad and the skin. Using a dry washcloth, wipe
 green — right leg (RL)
site free of loose hair. If the patient is perspiring and the
 red — left leg (LL)
electrodes won’t stay adhered to skin, apply a thin coat of
 brown — chest (C).
tincture of benzoin and allow to dry.
With the five-leadwire system for telemetry monitoring you  Attach the electrode pads. Remove pads from packag-
can monitor any one of the 12 leads using a lead selector on the ing and check them for moist conductive gel; dried gel
monitor. Leads placed in the conventional limb positions allow can cause loss of the ECG signal. Place an electrode pad
you to view leads I, II, III, aVR, aVL, and aVF. To view chest leads on each prepared site, pressing firmly around periphery of
V1–V6, the chest lead must be placed in the specific chest lead the pad and avoiding bony areas, such as the clavicles or
desired. prominent rib markings.
 Connect the leadwires. Attach appropriate leadwires to
 Telemetry monitoring — Wireless monitoring, or the electrode pads according to established electrode-lead
telemetry, gives your patient more freedom than hardwire positions.
monitoring. Instead of being connected to a bedside
monitor, the patient is connected to a portable monitor
transmitter, which can be placed in a pajama pocket or
Troubleshooting monitor
in a telemetry pouch. Telemetry monitoring systems are problems
available in a five-leadwire system and a three-leadwire Many problems may be encountered during cardiac
system. monitoring. The most common problems are related to
The five-leadwire system for telemetry (Figure 4-5) is patient movement, interference from equipment in or
connected in the same manner as the five-leadwire sys- near the patient’s room, weak ECG signals, poor choice
tem for hardwire monitoring with the four limb positions of monitor lead or electrode placement, and poor contact
(RA, LA, RL, and LL) in the conventional locations and the between the skin and electrode-lead attachments. Moni-
chest leads placed in the desired V1 to V6 location. With tor problems can cause artifacts on the ECG tracing,
this system you can monitor any one of the 12 leads using making identification of the cardiac rhythm difficult or
a lead selector on the monitor. Leads placed in the limb triggering false monitor alarms (false high-rate alarms
positions as shown in Figure 4-5 allow you to view leads and false low-rate alarms). Some problems are poten-
I, II, III, AVR, AVL, or AVF. To view chest leads V1 through tially serious and require intervention, whereas others
V6, the chest lead must be placed in the specific chest lead are temporary, non-life-threatening occurrences that will
position desired. correct themselves. The nurse and monitor technician
The three-leadwire system for telemetry (Figure 4-6) need to be proficient in recognizing monitoring prob-
uses three electrodes and three leadwires. The leadwires lems, identifying probable causes, and seeking solutions
are connected to positive, negative, and ground connec- to correct the problem. The most common monitoring
tions on the telemetry transmitter and attached to elec- problems are:
trode pads placed in specific chest lead positions (leads  False high-rate alarms — High-voltage artifact potentials
I, II, III, MCL1, and MCL6). Only one lead position can be are commonly interpreted by the monitor as QRS complexes

ECG workout_Chap04.indd 27 4/29/2011 4:31:37 PM


28 Cardiac monitors

/ / /
yp %
– + – –
\ t i

a
/ a / a / a \
G G + G +

Lead I Lead II Lead III

Negative lead – 2nd interspace Negative lead – 2nd interspace Negative lead – 2nd interspace
right midclavicular line right midclavicular line left midclavicular line

Positive lead – 2nd interspace Positive lead – 8th interspace Positive lead – 8th interspace
left midclavicular line left midclavicular line left midclavicular line

Ground lead – 8th interspace Ground lead – 8th interspace Ground lead – 8th interspace
right midclavicular line right midclavicular line right midclavicular line

/
% /p
^ – –

+ i f /
+
<0
/ s / a \

G G

Modified Chest Lead V1 (MCL1) Modified Chest Lead V6 (MCL6)

Negative lead – 2nd interspace Negative lead – 2nd interspace


left midclavicular line left midclavicular line

Positive lead – 4th interspace Positive lead – 5th interspace


right sternal border left midaxillary line

Ground lead – 8th interspace Ground lead – 8th interspace


right midclavicular line right midclavicular line

Figure 4-6. Telemetry monitoring: Three-leadwire system.


The three-leadwire system uses three electrode pads and three leadwires. The leadwires are connected to positive, negative, or ground
connections on the telemetry transmitter and attached to specific lead positions (lead I, lead II, lead III, lead MCL1, or lead MCL6). Only one
lead position can be monitored at a time. A lead selector isn’t available.

and activate the high-rate alarm. Most high-voltage arti- gel, a loose electrode, or a disconnected leadwire. Low-
facts are related to muscle movements from the patient voltage QRS complexes can also activate the low-rate
turning in bed or moving the extremities (Figure 4-7). alarm; if the ventricular waveforms aren’t tall enough,
Seizure activity can also produce high-voltage artifact the monitor detects no electrical activity and will sound
potentials (Figure 4-8). the low-rate alarm.
 False low-rate alarms — Any disturbance in the trans-  Muscle tremors — Muscle tremors (Figures 4-13 and
mission of the electrical signal from the skin electrode to 4-14) can occur in tense, nervous patients or those shiver-
the monitoring system can activate a false low-rate alarm ing from cold or having a chill. The ECG baseline has an
(Figures 4-9, 4-10, 4-11, and 4-12). This problem is usu- uneven, coarsely jagged appearance, obscuring the wave-
ally caused by ineffective contact between the skin and the forms on the ECG tracing. The problem may be continuous
electrode-leadwire system, resulting from dried conductive or intermittent.

ECG workout_Chap04.indd 28 4/29/2011 4:31:37 PM


Troubleshooting monitor problems 29

1
t

t
P
-

i
r
ri

--
PP
rrr .

V
i TT

- -
rr : :

r-

r :H

M
.

rr
TT:

j
rrr irr

'


-

;r i
Ui

r $3 - 12
. .
;

MW
.

m
n
rrr

31
m

5
=
-i
.
i i
,
rr

t
i
-
r*T
: ::: ::

a
p:

. 11 J

f ::
t
rp: mi

-
.

**
i
P

LP
P
*

4
r rrr
2:

-

+

T’
P

Pi;
f ::
r rr

H
L
+

i
-f

U
rrr
: rr
£E

t-
r;•
th r: s :_

:. .
_

UP

: hH rrr m

PT
Ml

:n
. :: r :

: :

t
P
rr
IKS

isi 3

!i
r:

- T
l

h

i
_
rg-
a
.
>
i
iam*

t
IT

:fc :
If

&
P

± -p

Figure 4-7. Patient movement. Cause: Strips above show patient turning in bed or extremity movement. Solution: Problem is usually
intermittent and no correction is necessary. Movement artifact can be reduced by avoiding placement of electrode pads in areas where
extremity movement is greatest (bony areas such as the clavicles).

Figure 4-8. Seizure activity can activate the high-rate alarm on the monitor.
30 Cardiac monitors

J F ±i
i 4 .
- nXt : -- :
:
tri t -- - -- rr T f -
4

J ±
t
H n t i - •j
- -rf
• 1
ll
:
tv
! 4-

S!KM! .
4
Is**
t .
i
7

i •. 1
r 4 -+
J

-
< it
l - t
"
p::: -
l
j ‘ i
i t
11

i

1 iii
•t r
r
iniiiiiiiiiiiiiilili t -
II t

Figure 4-9. Continuous straight line. Cause: Dried conductive gel, disconnected lead wire, or disconnected electrode pad. Solution: Check
electrode-lead system; re-prep and re-attach electrodes and leads as necessary. Note: A straight line may also indicate the absence of
electrical activity in the heart; the patient must be evaluated immediately for the presence of a pulse.

3
il ! liii!

Figure 4-10. Intermittent straight line. Cause: Ineffective contact between skin and electrode pad. Solution: Make sure hair is clipped
and electrode pad is placed on clean, dry skin; if diaphoresis is a problem, prep skin surface with tincture of benzoin solution.

-
.
mr
i
r
:
.
- 71 t* T r
L
i
t •<
.
t -
-f -* 1- -r-rr 4
r U-r -
• *
a
kJ
C ltd

-t i
a r
.-
i ;:::: 4- -
iff r -
r
5
rr
H

HTI
i -
i iT
2
n li
n
mrrrr -
4 4 4 1 >1 1

t
rrrrTTTTrr
•l l

n
rrr T r r 4
H
!
r
rrV
1* rr• *
" .
rT * T

1
HE - ; rm 1

n

nr
: fj

rr m
tn nj rH r? H
L fc
trj
3
*

i t:
J
i f S ’
: r
i
j* ”
tin ; i ;i E; F i- [V
BS
4 Hi 4
+r inr rrrr rr-H
ti fr v? 4-
H
4-
T
rH hi
r<
t -+-r 4

I n
. j i
in
-
i

1=-34 iil
t
ilr Jl'hitiJi:
tf
n
r
UBilsi
x ti
i +
h - 1

1&
.
H- 4 - t i- drr -1- m

Vk flli
irj
- ' I ni t. nr t f n: -r -
1 I 4T ' nr nrc: i 11 h
nc in rI * “
tn
n 4 t : HI I: t
1 3 il
!
iSli
B
r J!
-4
sn r
-1
i
rr-rr
-
v
t:
- El
3 A
tr
4 n 3 =n p H
l rH
3
Tt 4 IA S3 hrH rn t mi :r r
Figure 4-11. Continuous low waveform voltage. Cause: Low-voltage QRS complexes. Solution: Turn up amplitude (gain) knob on monitor
or change lead positions.

ECG workout_Chap04.indd 30 4/29/2011 4:31:39 PM


Troubleshooting monitor problems 31
:::::::::::::::

11
I
£
111
Lai
M

•aaa

- Hi
ill :!

trr
§

T
r

i- -i
I

is
A
*
- -UU-J

u
u

ZZ11
4

4
3
c

:
» :::8
KSB] B I
i

^
!

U
h
y

j
i
»

:
X

K
n


- *.

:=
tt

7
Figure 4-12. Intermittent low waveform voltage. Cause: Intermittent low-voltage QRS complexes are seen in both strips above.
Solution: If the problem is frequent and activates the low-rate alarm, change lead positions. r
tt

-T z :
t

saxaKT‘ iiiuusuniis
T.-.i
ir

1
-H
L
ims

a ^
aiincu

r:

*
H

t: i
i
- ti

t
4-

: r :: xi
H

Li

a
r. t
r
Figure 4-13. Continuous muscle tremor. Cause: Muscle tremors are usually related to tense or nervous patients or those shivering from
cold or a chill. Solution: Treat cause.
32 Cardiac monitors

«
t
w
M
m
m*

1
r

:i

i
i

i
;

-

Figure 4-14. Intermittent muscle tremor. Cause: Muscle tremors that occur intermittently. Solution: Correction is usually unnecessary.
Note: In this strip, the patient has two P waves preceding each QRS complex (second-degree atrioventricular block, Mobitz II). If the muscle
tremors were continuous (as in Figure 4-13), you would be unable to identify this serious arrhythmia.

-
:

44

4
T
+r

-f

f
t
m
.11 ]
'11

: ::

inz
dzv
r
-i

> I

::

L

Loss of Loss of reception Loss of


:

reception reception
-

X
r

.. J

i
Trr t: ::

:i

f M
.

::

-
-

- rrTi:- T
-

4-
i

r
rr -rti

+
JT +1 :

:
=4
T
1.

i
I

- »

1
4

:::: 4
-

T-
t

-e
r

-
-

£
=

t *
4

4
-
ii
11
t

Spike Spike
+-

r
4

-r
rrt
r+
..

h;
-

rrtii
-

-f - 44
4

i
r

P
4
T
..

—•
- 4-
-t - -

•f

4
if
t
L-
;u
x.t:

4
Loss of
i ::
- -

L -:

- -:

reception

H-

id
4

4
:t::

A- . ii

X
r
L
^
mm

cscsaiie iu

.isiii
»

4:

-
1-

**

r.
h

r
r

li !

L
VT

4 Muscle movement Spikes


t

Spikes
’H
H4

7
:x
T:

-
44
H
-I
U
:i ±

-
3

.
"

r
4

j
-

ITT
:r
::tr TIT
"itrx i

iiiiiin iiiiri
iflBBiBrsssEBES(B;s::r B«isss::s:»KKi::»

Figure 4-15. Telemetry-related interference. Cause: ECG signals are poorly received over the telemetry system causing sharp spikes
and sometimes loss of signal reception. This problem is usually related to weak batteries or the transmitter being used in the outer fringes of
the reception area for the base station receiver. Solution: Change batteries; keep patient in reception area of base station receivers.

 Telemetry-related interference — Telemetry-related of the base station receiver, resulting in sharp spikes or
artifacts occur when the ECG signals are poorly received straight lines on the ECG tracing.
over a telemetry monitoring system (Figure 4-15). Weak  Electrical interference (AC interference) — Electrical
ECG signals are caused by weak batteries or by the trans- interference (Figure 4-16) can occur when multiple pieces
mitter being used in the outer fringes of the reception area of electrical equipment are in use in the patient’s room;
Troubleshooting monitor problems 33
Figure 4-16. Electrical
interference (AC interference).
Cause: Patient using electrical
equipment (electric razor, hair
ii
IIUIH

dryer); multiple electrical equip-


ML ± :

ment in use in room; improperly


grounded equipment; loose
electrical connections or ex-
brain

posed wiring. Solution: If patient


is using electrical equipment,
problem is transient and will
correct itself. If patient is not us-
ing electrical equipment, unplug
all equipment not in continu-
ous use, remove from service
and report any equipment with
t:

breaks or wires showing, and


ask the electrical engineer to
check the wiring.
Kiiiiiiiiiiiiliiilll

X ::t
tPH
:r.
!.
n

• [
H
i-

r
l
.

!ii!iiK5»* imiisn
hiiimii
faiiwiiinMigiM

1 '|
t

-
I

"'
i!ii; 3;;;;!;ii;;nij ;i5iiMii

I
aaBMii
lJ L

1!

n
1:1
II
mi
ii

ii
-. .

x

t
i J

m
-

: I^t
*-

iW
.

••
1.

wifi iiaBiiiNi - k
k
!.

r
'
r
14
lli


±
-T-

!
-

i
1
y

imiffliiii
.

r
::

,, i

r
^
..
. -i -
l
i

..
-

(
.1
: X

Figure 4-17. Wandering baseline. Cause: Exaggerated respiratory movements usually seen in patients in respiratory distress (patients
with chronic obstructive pulmonary disease). Solution: Avoid placing electrode pads in areas where movements of the accessory muscles are
most exaggerated (which can be anywhere on the anterior chest wall). Place the pads on the upper back or top of the shoulders if necessary.
when the patient is using an electrical appliance (such as
an electric razor or hair dryer); when improperly grounded
equipment is in use; or when loose or exposed wiring is
present. This type of interference results in an artifact with
a wide baseline consisting of a continuous series of fine,
even, rapid spikes, which can obscure the waveforms on
the ECG tracing.
 Wandering baseline — A wandering baseline (Figure 4-17)
is a monitor pattern that wanders up and down on the mon-
itor screen or ECG tracing and is caused by exaggerative
respiratory movements commonly seen in patients with
severe pulmonary disease (for example, chronic obstructive
pulmonary disease). This type of artifact makes it difficult
to identify the cardiac rhythm as well as changes in the ST
segment and T wave.
5 Analyzing a rhythm
strip

There are five basic steps to be followed in analyzing a calipers, a variation in the R-wave regularity may be noted,
rhythm strip. Each step should be followed in sequence. but without marking and measuring between the short-
Eventually this will become a habit and will enable you to est and longest R-wave variation, there is no way to deter-
identify a strip quickly and accurately. mine how irregular the rhythm is. Examples of rhythm
measurement are shown in Figures 5-2, 5-3, and 5-4.
Step 1: Determine the regularity
(rhythm) of the R waves Step 2: Calculate the heart rate
Starting at the left side of the rhythm strip, place an index This measurement will always refer to the ventricular rate
card above the first two R waves (Figure 5-1). Using a sharp unless the atrial and ventricular rates differ, in which case
pencil, mark on the index card above the two R waves. both will be given. The ventricular rate is usually deter-
Measure from R wave to R wave across the rhythm strip, mined by looking at a 6-second rhythm strip. The top of the
marking on the index card any variation in R wave regular- electrocardiogram paper is marked at 3-second intervals;
ity. If the rhythm varies by 0.12 second (3 small squares) two intervals equal 6 seconds (Figure 5-5). Several methods
or more between the shortest and longest R wave variation can be used to calculate heart rate. These methods differ
marked on the index card, the rhythm is irregular. If the according to the regularity or irregularity of the rhythm.
rhythm doesn’t vary or varies by less than 0.12 second, the
rhythm is considered regular. Regular rhythms
Calipers may also be used, instead of an index card, to Two methods can be used to calculate heart rate in regular
determine regularity of the rhythm strip. R wave regularity rhythms:
is assessed in the same manner as with the index card, by  Rapid rate calculation — Count the number of R waves
placing the two caliper points on top of two consecutive R in a 6-second strip and multiply by 10 (6 seconds × 10 = 60
waves and proceeding left to right across the rhythm strip, seconds, or the heart rate per minute). This method pro-
noting any variation in the R-R regularity vides an approximate heart rate in beats per minute, is
The author prefers the index card method, because each fast and simple, and can be used with both regular and
R wave variation (however slight) can be marked and meas- irregular rhythms.
ured to determine if a 0.12-second or greater variance exists  Precise rate calculation — Count the number of small
between the shorter and longer R-wave variations. With squares between two consecutive R waves (Figure 5-6) and
refer to the conversion table printed on the inside back
cover of the book. A removable conversion table is also pro-
vided. Although this method is accurate, it can be used only
for regular rhythms. If a conversion table isn’t available,
divide the number of small squares between the two con-
Shortest R wave Longest R wave secutive R waves into 1500 (the number of small squares
variation variation in a 1-minute rhythm strip). The heart rates for regular
Measure rhythms in the answer keys were determined by the precise
between rate calculation method.
here
Irregular rhythms
Only rapid rate calculation is used to calculate heart rate
I R wave variations in irregular rhythms. Count the number of R waves in a
1st two R waves 6-second strip and multiple by 10 (Figure 5-7), or count
the number of R waves in a 3-second strip and multiply
Figure 5-1. Index card. by 20 (3 seconds × 20 = 60 seconds, or the heart rate per
minute).
34

ECG workout_Chap05.indd 34 4/28/2011 1:18:21 AM


Step 2: Calculate the heart rate 35

.
• t

1.
t :T
J li

1
j
m u t
4
SI
nr 17
f
i t U -L
r: :
f 44

nr!

'
: : rr::
nr
- i
: i3 r •

. t . i: H •
: irrtr:
WE K
j .. ; l ut r t r u
il- .
r
,
i * i
L
:i
- - * •n
I
»
M
l i :
n n
tl n

*
. rr

r h - > . • • • \ .2 , . • i .

'

T
Nr * *
i&Z :: ..
r • s •!

r. E r .; ::
-t -
i f
rrt
r.
r-
-.
j n:
:t r..1 ‘
nn : i f il
n t LG
nf
1 4.12 A In
I - nr:
- 1
t:
-
t - H V-
• r n r‘ ‘ •v •• 1
11

Figure 5-2. Regular rhythm; R-R intervals do not vary.

n. nr Ut rr: T.
4
It i
k *- 4- .1.. 1 ,
. . : tl : n!t -tit
a-
Si HE£L Lt LG • rr L •- rrn •

s

;: :U J rr Tl 1 1 :::

m:
- nm

nt u n; i - LXJ
n rr
IT ILL: 3 I1

.. tt l HHIT
- -t t* -
T t
G
r :i r
:u nt

:
li t
I
nm
fffi - :.—t: r ::
J J M •I
4 T

r. tit: I.

4 4
-- —- *
- ill w : rut t :: rrJ
I 1


HU 3
t
i
it

TT
i
3: tnrt .::::r
*U r
r

u
rt rr
•11

J . H hu - 1 •

in;
? it -
» *
i
nn r : T: t r r :: . n: n: :r. r : . Jr: r


LG . T
• .
G I
- r
LT
:
-h I --
, . . .. n
. r T i : rr n 1
HU
.
in : I: J . .
: • '•
t .. i

: : -r-
.. +
nr L n I A A i
J :: rr: :r J: +

LI
i

-t
.» ,
rrrfr rn
i

r it: :.r: r
. L:: n:
i rr- 1
r
rU lr:r
>

*
in

hH-

Figure 5-3. Irregular rhythm; R-R intervals vary by 0.32 second.

r ±j 3 t ? t ::
: m
h -G- 4 - t 4 .1 >•

n m-

--
i"
1 :r
n
T
± I;L: i :
rr:
t:
4 l

l
4-

rt
rt
4
*
" '

— t
x
LL
:4

n
rr r
3
h
4

a +
4

U: -X :: L
:r is !' "
tn r-f -r - G r
r
t- nr
PW
L: L ii i-
rl r
r L 3 :T T
t.
t
t - HG
rr
-= IH -* •
r.

r
n - r-
LG
t nn IHTrr im IIP
Hr :
r
4l r
: ::
; I!:
. :: t:
a •i
4
::
::: :1j
n r
i r
.1 .t rt t! •
- 4

mi r H
rr :nz
5?
t
Lt: UL4
n i- i t
1
: r
Mil
BIAS - rm - •-
:
a:: -I
trr - r fr . ’ 1 ff ? ’ :: t ' 1

Figure 5-4. Regular rhythm; R-R intervals vary by 0.04 second.

Other hints rhythm. In the example, the first rhythm is irregular


When rhythm strips have a premature beat (Figure 5-8), and the heart rate is 140 beats per minute (7 R waves in
the premature beat isn’t included in the calculation of the 3 seconds × 20 = 140). The second rhythm is regular and
rate. In this example the first rhythm is regular and the the heart rate is 250 beats per minute (6 small squares
heart rate is 68 beats per minute (22 small squares between between R waves = 250).
R waves = 68). When a rhythm covers less than 3 seconds on a rhythm
When rhythm strips have more than one rhythm on a strip (Figure 5-10), rate calculation is difficult, but not
6-second strip (Figure 5-9), rates must be calculated for impossible. In the example, the first rhythm takes up
each rhythm. This will aid in the identification of each half of a 3-second interval. There are only two R waves.

ECG workout_Chap05.indd 35 4/28/2011 1:18:21 AM


underlying rhythm first, then add additional information,
rately. When interpreting a rhythm strip, describe the basic
answers. Each rhythm on the strip must be analyzed sepa-
have one answer or several answers. Figures 5-8, 5-9, and
5-10 have two different rhythms and thus two different
or several rhythms. Therefore, each rhythm strip may
As you have seen, rhythm strips may have one rhythm
p
H - TTTI „ t: :: :
Hi - —
r 111

= 7733 —
:
1 H a .-p i :

—- PL:
+ I

n Hd: 17 ” T’
3
t - - 4
r: i : L: ! r
:r-: tri
!ti j :: I
‘‘ .- f
* •
“ n • •ii I :
rrr. i .
i
=* = - 3 l
:
71
P
1
:
:7 7i:; :;z +H 777

.1! • Tt r — t “
:: 1777
J
-- !tr;
[7
r- - :zn ±rh
l.
77
- m= P- — rr
3 seconds

i i" llll

± in:i .. a 77
I : :•:. J :
: L — Tprr

I
I ::
i I u
i - r : i:
33 737
: iiPt
- 71:
l


..i
t
7TTZ
ti :: Ti
474

-L
_ 1777 3
7 -tp;

77
i ITT . J ..
**

tr PP. : ::. " V ffl- p .a ; :: IT:


• 1

:: i
-
T - rrrJ
37 M

n±rtn - 7:1 J : IP ::: : n r:

Figure 5-6. Regular rhythm; 25 small squares between R waves = 60 heart rate.
H' ••
.i
i : : rr
- Ur:
rrarn :. i tPtPtr: :Jp
*• L7 f : . i :tz
7 rr

bid f 4 jtrfp- Hii :r::


:+ n
7
7z ::: r
P
.7:;.3 3TL =rr

:,
4 MU : 77 r.
:: •

rate of 167 beats per minute (9 small squares between R


per minute. The second rhythm is regular, with a heart
40 (1½ second × 40 = 60 seconds, or the heart rate per
minute) to obtain an approximate heart rate of 80 beats
irregular. In this situation, multiply the two R waves by
Therefore, you can’t determine if the rhythm is regular or
Figure 5-7. Irregular rhythm; 11 R waves × 10 = 110 heart rate.
zzi:
iPVP ii
i

J
*
..
V* ::: rr:
- U: MrrthpP:: TP
U «-. ..
- - UH
H 7337
3
u J
7 IP J . j j ..

11
•*
"» r ::
:: i :3 : j
— 77 i :
-
*::
: r
— -
i :: r *
LLL .17::77 37
7
-
1
f777 71 rr 7:: : *:::
L f
;h n “ rrrr.
J * ii : :: 7 7 S rm
-
< t

id 77 - 1J _. - iSlIErX! - r l T
If : :
.1
fHiPTT
- r a £ iiu !i 1 If :r ::.i: ife iti i!:r iirT
i
L 77 : ff r: ;7 fp-
Ti :i :
-
i
13 u : «
T: : - H t: IP
..
-.
Tf drr
Analyzing a rhythm strip

; 1

.
r:
::T
U ii: . i MP
- -
i
4Tt 73 : ;
t zb:
4r:P
: J

; I
r*i
;?:
:::
il .. .
P
H -+ -
' *t :itr.
Figure 5-5. ECG graph paper.

:: n r4 r : :

’ ”
r' * 1* 37 J :
ii: 3 rt 773 •- 1
Uiii i: : i: 4~ -4 U 13i7 ::rr

i
i i :: :
• • i

3
3 a i . , rU ?
i
b

it
IP
tn m
L. .
M: T <*
ti t

waves = 167).
i
n
i
.1
[ tt
It H
TV.
73
ZZZX \
4
LI

a -»
J.
U
h*
J .
7 ml 4 mm m rt
r
j in
t i
fell
36
Figure 5-10. Calculating rate when a rhythm covers less than 3 seconds.

Figure 5-9. Rhythm strip with two different rhythms.

Figure 5-8. Rhythm with premature beat.


I

- H

Sc
2 r-1

=rH
J

ag a-
iJ

± n.:
•T' 9 r .!rt• t

.
MM
MM m
.
L
ss

First rhythm
lii i
d :s

First rhythm
1.5 seconds
First rhythm
^4

itr 111
I M
..
rn
t

*- t L
.. .i
" r
-t
— rr:
4
Fi >I

i --
rri'd Y ::: .: r
'

11
r rr:
1 r UKBESibC

*
Si
n
t tj

I -
i:lid
M .
H4.
••r
: ::t

iSmJi M
>k
Second rhythm

j .. 3 s: 1 14
u: TtrttH
r.sn
L

(Second rhythm)
Premature beat
xir: ui
3 seconds
i -
1: 99
h T\

L: rt i ^

5 f
rt

Second rhythm
±:: 1

Step 2: Calculate the heart rate


>r 4
4 n T rq:
4 ... . : J- r
;
did Y * » • [
. - - .1.LJ - 1
1
*

i
a
..
. : :.1i:.:
:
^ -4433 1
1
rrn
-
"
ddt: .i
.

V 4 ft r •
4 4 .
4
-- J

rr mi
i •t

37
38 Analyzing a rhythm strip

3r
I

r
-
4

M
S
IS
•*•

i
i
:
1

::i»3::u
i

'
7 7

[
i -
Figure 5-11. Normal P waves.

sn
M
H
3
Ml

••III
III!
P
•»
•r

Figure 5-12. Abnormal P waves.

-
r
! 1
-H-f-

44 =
t
= ft
fenJii
±
r i: : ::rrl
kUji
E 2
- .
i11' »4 I i h
•«
•*

- .
IBM
::•»«
III

BBIB
a
!us

Ball

AMI
1M
B

LC i 114
HI;

III .44
••
1

::
.
St
A

-m at- irrr
H+ - H t-
fnpE

Figure 5-13. PR interval 0.16 second. Figure 5-14. QRS complex 0.12 second.

Box 5-1.
such as normal sinus rhythm with one premature ventric-
Rhythm strip analysis ular contraction (PVC) (Figure 5-8).
1. Determine regularity (rhythm).
2. Calculate rate. Step 3: Identify and examine P waves
3. Examine P waves.
4. Measure PR interval. Analyze the P waves; one P wave should precede each
5. Measure QRS complex. QRS complex. All P waves should be identical (or near
identical ) in size, shape, and position. In Figure 5-11
Step 5: Measure the QRS complex 39

3i : =
i

Figure 5-15. QRS complex 0.10 second.

there is one P wave to each QRS complex, and all


P waves are the same in size, shape, and position. In
Figure 5-12 there is one P wave to each QRS complex,
but the P waves vary in size, shape, and position across
the rhythm strip.

Step 4: Measure the PR interval


Measure from the beginning of the P wave as it leaves
baseline to the beginning of the QRS complex. Count the
number of small squares contained in this interval and
multiply by 0.04 second. In Figure 5-13 the PR interval is
0.16 second (4 small squares × 0.04 second = 0.16 second).

Step 5: Measure the QRS complex


Measure from the beginning of the QRS complex as it leaves
baseline until the end of the QRS complex when the ST
segment begins. Count the number of small squares in this
measurement and multiply by 0.04 second. In Figure 5-14
the QRS complex takes up 3 small squares and represents
0.12 second ( 3 small squares × 0.04 second = 0.12 second).
In Figure 5-15 the QRS complex takes up 2½ small
squares and represents 0.10 second (2½ small squares x
0.04 second = 0.10 second).
If rhythm strips are analyzed using a systematic step-
by-step approach (Box 5-1), accurate interpretation will be
achieved most of the time.

ECG workout_Chap05.indd 39 4/28/2011 1:18:27 AM


40 Analyzing a rhythm strip

Rhythm strip practice: Analyzing rhythm strips


Analyze the following rhythm strips using the five-step process discussed in this chapter. Check your answers with the
answer key in the appendix.

-
ii f n = r

§
*
*: HE
t .. : 1
1 -rr- E - - frr
hH
\
I
f
TI; rxx iiri
c H
Ksipssipnjn H
TOPTiTi £ """
- * !!
MI
" " "i
P
mmmmmmM mmmmm m
rj

- TI i. XT I .. r
r* s tr pit
1

ii l
r
E £
PP E = i
rf s it L

H £
- PI tx 11 j«iiii»
- h:::
:
.
ri P
I |iiiii««i|i|8m«iit »» i
rta
T:
^ :t. :
ta
rr; X u: xrx T a
*
£ £ - •? - • -
r
Hf" t T I
: g
i
32
! I::: g
a
4
x it :: ::: :
g Ft
mm m
Strip 5-1. Rhythm: _____________________________ Rate: ________________________ P wave: __________________
PR interval: ___________________________ QRS complex: __________________

c :: ±i± ti lie ft n XXtt ti


H T
:: rr P i t:
r
1
H-f - 1
P =
g ~ t rXffnXwXpwS
X
- * 4.
t t .
a: 1
r i

* XX - - r ::::
Ei
fm: ± ttxuixixxp nTnxxxT-ixj- u^u

rxi
t -
•*
Cl
n
-t « ri * •+ Pxxxi
K:
trxr : ± :: x sag
i r e:
F:
rfr
H r- j
U Pt = £ XIX :: .
F- Xil T 1 H P fx;
Strip 5-2. Rhythm: _____________________________ Rate: ________________________ P wave: __________________
PR interval: ___________________________ QRS complex: __________________

*. ..
1. ! a : 1 i rf H 4 tr ]
- a
XI
a
Uit ±±H L:
11 -
11
rt
t :r
l -i u=
i:
xil
i j

4 ,4
*

m
14 <*


XJ i
1 4 4 i
It t+ .. rt 4- -
4*
::i: J4 : 4 4
Ti ' n 1 r
mmm
"* t

m -xj:
i
i x cr - BH
j
I u -vtftp
: - 4 v * J t *
:r
• ]
Hri
- Pi i
Ti
'

i : xx 4

alii mn
* 4
XXIX
rrrx
. ••* r Pt
: ::: t -r * -H r •-
:* . xrttxxnr 4

£ * a r L
IX
r•
4
4
'
XXX
4
Xi rr. : .rr 1
m 1 - cxx : X +
t txx P :- '
X

rr
3 : UP WT; ;
TTT XXXI
ir: rt f •
t
I
xn
r 4 »•
L
r
- tr
r
-
4 4

I i-t
1-
rx LX IT
rn t tn rxx
r x
TT
4 i
r "

fit -r
T
4 4
: ::
tm +T4
uti
..
. i .. : •
ft T
X t :: T* r t * *

xx tx4;i
-
|r
r:r: -r: :44: • I •4 * - t 4
xv
. rt ;; :w
+
4X. :x:x
a rf
4 4 4 t
xj

Pi ir -m
' * »
Xr V1 1 T r *
t : X. r
BxH; :nr. nLi:i. :rti n : f xxiU XT •

e’ u
t 4 '

.
4i
mi t r xxxx 131 i rttri :I n PT 44 r
l
:: •r i 12
. !Hfn
t
: ; P JT:
xxxr
xx: ::.r . : rx :
Ei ua - V

-
4
4 H
En BT fiH t pp
t- rxnn rt :x I Hr
-
4

P n P U
t-rr4 + T i x x i a Blx ’
ITTX :
t-
4-

Strip 5-3. Rhythm: _____________________________ Rate: ________________________ P wave: __________________


PR interval: ___________________________ QRS complex: __________________

ECG workout_Chap05.indd 40 4/28/2011 1:18:28 AM


Rhythm strip practice: Analyzing rhythm strips 41

••r J Vi
1 H
tr - -
i . *
+
m I
-
it f
'
Tit
it it
U
-U
± T T,
m
E -- P

k i t-
T
3k -
J
r
tl*- - :
t
4
-
ii
t
iv
-- t

i
i
: r -.. :
Tt t4
i- - .

•• f t r H
T rt
r
.

: :: : i rt
a i -
rr
1
- PW- -
rm
-t -i
_
r. i
r m
- fl m t

-.
hr: t - k n
4 1
if J:
r - •
r t -
rt
::::
"
rr r
Tt
-

i t
1 it
u
11

Tl
t
Z

t
t
7: T > •r
tl: f
T

M!i :: t t
t

iff •• rr atT
k - 4 U

m -1 i ••* *
"
n4
TT :
- -
•*

' :r .iii
i

If? . .i •r - - i-
r 4
n

* H- *
K
3P
iitiiiifiiiti r f •j - rT rr
P - 4- - -
-t- i

Strip 5-4. Rhythm: _____________________________ Rate: ________________________ P wave: __________________


PR interval: ___________________________ QRS complex: __________________

:a
» «
M
Blfl
IM u
p
t
f m t:
I
a lk
r.r
ia : t
ill

ns
HI j]
d
1=
111
J
n B a n nr: t+4 -
5 1 l
r *f

Tt - 4 ff ZJ
fi ! H

s1
r 1
7 i.
H £
4.- P
: U.
i-k I
+T7 r : i: K 4.

: 4
1
r - 4
J 5! T trr
- S
i 35
::
5T 1 la +

::: 4 roili n: mi
1 E i » D i d d d » < 3i > i l S > » M i « « «!i a 3 M i I > N i *i a|EuiEl ( !
Strip 5-5. Rhythm: _____________________________ Rate: ________________________ P wave: __________________
PR interval: ___________________________ QRS complex: __________________

* •
- :1. - ro
a wmm I *
-M t i
i ,
t
14 t Tt-
l XI .

t
t
.. 5

a
4

a 4
p:-: k V
II
•J

rrrr -rr:
r:;t
:: :::: i ::t J :: :
•• ::
a --
11
S g T:

+4 4

H 4
-

5
m c
H
n
rr:
XL
1 ?
3
r - T-
»
H n: t ;n
J i •

v:
L:; r- ‘
ii 4k
rl
J: ; : t:
4
"
i
ff i
-- 5 1- i ifrH Tl 4 P;T
. i

5 7:: i;
I
:T M ;
n
1
.. rt : WTH : :: 1 r o t t i
-
5
- j

rr. r
1r 1i s
I ::r ; h i ::i .
1

s •v- . ill

1
‘f 1
3 l
rrt :i .
-T‘
IX ... XI
:: tr- t
: 5k
3
I
li H
; 5:
k. t
rtr :::5 rrt

::
t -
- 33

::
.;.5
-P rm Iii ’1 1«
: — Irr Eli; m4
t
[-3-1 11
l i l t'

:4

ut
i
r

rr- -
i - 4:
\
1

I:
3333
r-t -u :
i i

-1 « 4
rr (•
:: •
::4 : rrr : : r :TT
r. -
4 i
LM 4* -
r
rt
ga HT: :ait rr.T : 1A ::x -t r :; r il
35
4 it alga " XXI t . -. zzi . i -:
;; 3 ;r r
a
U pi . -
rt trt- 1
::r Llk . -U .
: 45
i rk 781
t:+
tr : .
*4

474 E if .
n : H : :• 5 111 M
i:
I rt iiiiii
Strip 5-6. Rhythm: _____________________________ Rate: ________________________ P wave: __________________
PR interval: ___________________________ QRS complex: __________________
Strip 5-7. Rhythm: _____________________________ Rate: ________________________ P wave: __________________

Strip 5-8. Rhythm: _____________________________ Rate: ________________________ P wave: __________________

Strip 5-9. Rhythm: _____________________________ Rate: ________________________ P wave: __________________


TZ . + xx

ilili
a
•Pi::'! i

4! ill
tin £111:4
'
ats
f
45
5
-H
a - i555
0
+:
i - -4
K >
t- t
1
5
?
:

u
XHHSHS
a 15 n Hat : ftft
i SE
i
c 15 : 1ixn
ixi ; t
* : m
ini 44 i
±: t :
rt
U r:
H “ :

iTiif 1 xrrt xtxr


-
LX:. t tH
!
xrn
- :
rir;
U Tpipif - T -•-•! —- * ' •

£
5
Ft
r-
irr
;:
< • +
i r 1
H
TTT1 4
55551-—
r*

m—
t 4 J

si
. t i*
H -+X4 +4 I aatii
- 44 HH
+• •*! • ft
•» "
4 +
t n -
-r Itii: i f
it
r rrr
n
J

-r
M

PR interval: ___________________________ QRS complex: __________________

PR interval: ___________________________ QRS complex: __________________

PR interval: ___________________________ QRS complex: __________________


n ^
rxxt
BJ
i
n Ml:: '
i
txtr
t T P nrt rrz
:
111! nU H
tltli i

S5 r l TJ
p ?T - >

lit
an tit 4 - .t

i ttt ::::KH
a V
rir : 4H
H i :: TH 4
4
ft

4? r
- m xt:U ::; t: .1.1 Ml ! iliiaHiigiiiicS
i

=
'
4 h
rrr Tn AAltS-t : : :i
-4 .

t±d: Ir
p a +
it
+
tt x xx ij
it -i ia H* t t E mmmmm
HIIV
M
U
HI HK
win

E
in
it = rt ll\\\
an :
3
lit t r -
4 1 4

v
ti i
wmmmm
nm
aai:
H
mm

V
=4 F:
i niiililillllliilinlilil|;ijii:!iiiiiiij
31
5 r ta
I: ¥ xx
m f H;
i
KHS
3»! iifiMiiilflflaiAi
la! « *«*•
5
tt 4 - •- i TTT| 3si:
2 iSn 23»:K
sa !
^^
k.

4 + t t# <
>••
naa
3HH
±- i+ -I X + * l
J 4 Sill
5
u= rr : t tnr
At a -» -» 4
Uli
Nil

44
.i
a 1 '

xx;

+
t naa
«uaa
aaaa
M 5S:
1 an 4
3 A E: Ft 6 xxt
ra 55
!
!
MUI •
— B—
aaaaa
aaaaa
yi !t -- -, i : - 4 4 aaaa
Analyzing a rhythm strip

nil it; 4
5
hH
It •riaai
a:::
PII
ITFA
TT
itxBTr
?
aiTuti t - IF ? tt
<
xftftTftftftfti
Til - ttf 1. ; - r- -t i r •• r# r t
"

i E *
Nflfll
aaaaa

it a
r?
tu i iilliliiillllliigil 'inmnoniiloi Hi:
•aaaaa

:a
r i f : aaaa
; E
•aaa
aaaa
pxx l i t Tt TP i •
aaa
aHia
a aaaa
Jf :± 4 m *NI «
aaaa

Si TIT:
4 i E E:
m
r •

44 I H 4
?rx::: r ! L ITTitr
4 rm 1
4
4 4
;
X
x
:r *
Uf .

S£ u
Ft ?
T

full i
ti m
m
- t

4 ±E
u ii
: Xt ix; m THE
nix : X
IT
:ss Xttt
iEffi -i-
j
:t r
45
*
I
Tt
1
-

WH 1155535 if !
R i ±i :: ft X
n
42
Rhythm strip practice: Analyzing rhythm strips 43

I II* B f l B H t#H V a B ««« a l l l l i a B a « V «« B V B B B B «« a H B B V r t B B f c B « M A i t K I U4l i i A Bi B lAiiiiifl 1 1 lilAfeiVB A l C B B B B B B B B B B B BI B B B B I B B I B P B B f l B l I I I B B BlflfetBBBB BAIIfB I B B B B B B B B B»B B A B B B B A


*
•§ M I M M I I I £ ••
»» M f gHBB V B I«4 VBBBBaaillBIBBJaBaBBBBBJaBBBBBBJBflBBBBBJaftaflBBBgJBaaBBBBBBBjaftaBBBBBaBBBftBBBaBBBBflftBBBBB gAJ
fllJJBBBBPtllMVifVtta !
^
!

jliii!j!i[iij!!ii!![ljiiii!ii:!!!!!iH!!!ii!jjjii!!!
i i!ili!il!!!li:iii:!!!!ji!lij:!i![!!!!jl!j!!:!iliiii!! i!fi!iii!p!!!jiii3iiii!liji!jiiiiii]i[j!iiiiiiijjji
i iil|!
IscsfKSBassssssisiasss^isiiaeiHssiiHiiaassessiaaiaJisssiEuSiiataiihhssss a sxisisiisasisszhzssEiizissshiihitEhEsiiiaesssiiisshsssiizIssSiEsi::::!:!
^^
'lagi •• •»
(
•• •
'’ i ia laaai
| a*v ,,

|
•ifaaaaaa aagappi < jaaaaaapaaaaaBaaaaaaaaaaaa »•
ifiaaiaaiapig > pm iaaaaaai
Hll *,9
' II« «9I
*aM(i*•
« • galii
«iiiaM|
•• • • •• '
jtta
iaagn ia • •tmaaiaaiamtii
ia •
« •Maaf
•• •«•
•ai •
! «•>•!
(it

-•• • « »* < •ig


•• ••iiaii p <
••ai
i
ni
aaap
(i>aiilitBBtitiai< iiiiaiakiii «|
aaiBnit
•• |
MB
iaaaaiiajiKai
*llHflllMM
! 9ii
a «*
a'
iaa <**i*iai! *if*i >aiaaaaai
| fi M Biiiaiii 1 aiiaaaagaaaaaa
in

••
B9HiCiiihaiaaa
•*taaiaaaaB
9f4
• •
a a•apapa •|
aaaaaaaaa
| aaaaiaiaaaaaaaaaiaaiaaiaaaaaaaaaiiaaa| | |
aaaBaii
| aaaaBB
| flaalf
iifiaaBBaaaiaaaBBBaaaaaaaaBBaaaaaaaaBfBBaBBaaaBaBggaBaaaBaggBaBaaagBBiaaaaaBBiBaaaaaBigaaaaaBBigaBa
« « i • iiiluiiiiittiif
» BaaiBBBBBiUBBPBg > BBB « BBnggnianniB| Baa
»• •ia «
! aaa a | AGAAAAAAAAAAAAAABBBBBA
aBaaaaaaaaa a a a AAAAAAAPAAAAAABBBBAAABAAAABI
| aa «aaa alaafPWapaaaiai aaagaagi
aftaaaBiaiaiaaiaaaflaaiiBiaaaaiaaafliaaiaiiaaaaiaaaBgaaaliaiiiiaaaaaaaflaiiaaiaaiflailaaiiaiaaaliBa9iiilafliSfiA*3* BBaaaaa aaaB

•BaaaaaaaaBBBBBBaaaaaaBBBaaBaa
• H « |
|l HMM
|
•aaaaaiiaaaaiaagaaaaaaaaaaa
••aaifaaagaagaaaaaBaaiaiaaaaaaaaaaaaaaaaaaaBaaaaaaaa
•gaaaaaaaaaiaaaaBBBagaaaaaaiaaaiaaaaaaa •aMIB> MI<•• » | aaaaaaaaaaaaaaaaaaaaaaaiaai > aaaaaaaaa « iJiaaa
<« BII I IM «IMIIIM « B B*BIBBM I>BIBI IMIBB «IMIMBBB*B
H p
| IMl ((
|
iaiaaaaaaaaaaaiaaaariaaaaaaaaaaAaaaaaaaaaaaaaaaaaaaaaagiaaaalllaaa
]44 ia
1 ( 4 ( (
P BPPlPPflllPP 1 PBBBPBBP

iiiittia • SifliSS 222 a 2 * 22222!ZIA 22 » S S!S 3SSiSSSS aaBBi 2 ai 2 a A *9222222325 i « 22 Z 2 I 3 *2*2222323!22222222332222222212222222 222
IBVfViBIVI BIBIBtillIIIHIBBBBIIllBlIlflflBfllVBI BillIBIII BilliflflBB liflflllAlIBBai
• BlfllAIBIB1 BflflfltBBB 11 BIBBIII1 B
BIIIBIPBBBBBB !§
!
22922222222222222222 ililllilliiiiliiiilii
szas3iii»is»»2iiii

Strip 5-10. Rhythm: ____________________________ Rate: ________________________ P wave: __________________


PR interval: __________________________ QRS complex: __________________

ni t I
Hi H A
i .in .T : r r
-n i-
YU e H

I
H i

.... i
—- ft f- r

\
-
11
P

9
F4
n n
U 3 P
aiaaa
ifiaa .X
1 rr P
:
P
n - B! n
if
;i. H
u
:

"ri
dr . U
rr f
- HlliPP
lillllBMl

. - - -;
r4> , M ,
- rr t
: •4 •••a r
-
a a a a
i
a
?
r - I p 1
4-;4^:4l 4 ^nr:t7: r p.v. ::
V* ff ’ TT
•r
4 ::: •
•. UiiJitt --
I’
-
t 7 t ±rtz Si T KU r

Strip 5-11. Rhythm: ____________________________ Rate: ________________________ P wave: __________________


PR interval: __________________________ QRS complex: __________________

ECG workout_Chap05.indd 43 4/28/2011 1:18:32 AM

Anda mungkin juga menyukai