Anda di halaman 1dari 3

The Gradient Approach to Thermocouple

Circuitry
Part II-Illustrations of the Method

ROBERT J. MOFFAT, Mechanical Engineering Department. Stanford University

The first part of this paper described the basic premise markedly.* The thermocouple is assumed to be connected
of the 'gradient approach,' (I e , that all of the emf can be to a readout instrument with cold-junction compensation
accounted for by considering only the portions of wire so that if there is no external signal, the instrument will
which are in regions of temperature gradient) In this part, read ambient temperature, 75°F in this case as shown
let us examine some rather complex circuits to illustrate in Fig l(a)
the ease with which their apparently anomalous behavior With the tip of the probe in the furnace, as in Fig l(b),
can be explained using this approach the instrument will read 1000°F If all of the chromel-
These circuits represent a simple demonstration alumel is inside the furnace, as in Fig l(c), then the
apparatus which was used at the Research Laboratories instrument will again read 75"F, since all of the chromel
to demonstrate the validity and usefulness of this approach and alurnel is at a uniform temperature and hence does
not generate any emf This can be deduced from eq (1)'
Illustrations since dl/dx = 0 for all the chromel and alumel The only
material in a gradient is copper integration from 75" to
Consider a thermocouple circuit composed of chromel 1000" to 75°F in the same material yields no emf
and alumel wire, but interrupted midway by an equal If the thermocouple is pushed through the furnace so
length of copper wire in each leg When this thermocouple that half of the chromel-alumel tip protrudes as in Fig
is inserted into a muffle furnace to various depths, as l(d), then the instrument will read -850"F, although the
shown in Fig 1, the indicated temperature will change apparent junction is now at 7 5 ° F There I S a temperature
gradient in a pair of copper wires and in a chromel-
alumel pair The copper wires yield no net emf, while the
chromel-alumel is in a negative temperature gradient

This article was originally published by Reinhold Publishing ___


*Everything inside the furnace is assumed to be at 1000°F
Company (New York) in 'Temperature-Its Measurement and
Control in Science and Industry, 3 (2-Applied Methods and everything outside is assumed to be at 7 5 ° F
instruments), (1962) ' See Part 1. March EIT, p 25

cu

cu

cu

w
+- -+ I

cu I
0 2M) 600 IWO 1400 1800 2200 2600

Fig. 1-The indjcated temperalure of a compound


thermocouple with different locations of the Fig. 2- Thermai-EMF relarimships for thermocouple
gradient materials vs plaiinum

Experimental Techniques 23
0 200 400 600 Mx) lo00
TEW. - * F

$0 --

EMF
I..)

32 400 800 I200 1600 IBM)


TEW.-
.F ICI

Fig. 3-Graphical solution for the net EMF of a Fig. 4-Graphical analysis of calibration drift due
compound circuit to service-induced distributed inhomogeneity

(temperature decreasing with distance from the instrument). (1000°F). This is represented by a line parallel to the
Thus we have essentially a chromel-alumel thermocouple chromel vs. platinum emf curve between 500°F and
connected to the instrument backward. The chromel- 1000°F. The line is extended from point 3 to point 4
alumel couple is generating an emf proportional to 925" F, parallel to the alumel vs. platinum curve over the tempera-
the temperature difference across the chromel-alumel ture span from 1000°F to 500°F. The circuit is completed
wires, which is now subtracted from the 75°F reference by the last segment parallel to constantan between 500°F
rather than being added to it. and 32OF.
Each of these examples could be handled junction by The magnitude and direction of the net emf of this
junction if tabular data were available for chromel-copper circuit can be determined from the ordinate of point 5 on
and alumelcopper and if the proper sign conventions were the completed analysis. Magnitude is straightforward. The
used. These illustrations were intended to show, however, sign convention is not. The electrical polarity of point 5 is
the relative speed and ease with which these circuits can the opposite of the sign of its ordinate on this chart. The
be handled by considering only the materials in a tem- ordinate, here, is positive, meaning that point 5 is elec-
perature gradient. trically negative with respect to point 1. This sign reversal
results from the original construction of Fig. 2 and the
Graphical Analysis of Compound Circuits definitions of the words 'positive' and 'negative' as applied
to thermocouples. When chromel is referred to as 'positive'
The gradient analysis can be made even more useful with respect to alumel, this means that the cold end of the
through its application to the graphical analysis of circuits. chromel is positive with respect to the cold end of the
Resolving a complex circuit into a graphical solution alumel. Figure 2 is drawn to yield the intuitive polarity
provides a certainty and an intuitive understanding which understanding which agrees with tabular data. When
is hard to equal by any other means. As an example, a Fig. 2 is used in a graphical analysis, a sign reversal results.
leadwire-error problem is solved graphically in Fig. 3 based
on the data presented in Fig. 2. The 'integration' is accom-
plished by moving clockwise around the circuit. Point 1 Graphical Analysis of the Calibration
represents one terminal of the measuring instrument, and of Used Thermocouples
represents 0 emf. A copper wire extends from point 1
(32°F) to point 2 (500°F). This is represented graphically The gradient approach also makes crystal clear the
by a line drawn parallel to the copper vs. platinum emf reason why used thermocouples should not be recalibrated
curve between 32°F and 500°F. starting at 0 emf at In a furnace Consider a thermocouple removed from
32OF. A chromel wire connects point 2 (500°F) to point 3 service because solrie evidence suggests !hat its output

24 April 1984
The circuit, and its operating temperature profile are
shown in Fig. 4(b). The effect of this temperature distri-
bution is to change the thermoelectric power of the
material 8 so that the wire in each temperature interval
from 1400°F to 1700°F has a different calibration. This
will be represented, for simplicity, by three uniformly
changed segments, based on the mean temperature in
each segment.
The emf of this thermocouple under operating con-
ditions is shown in Fig. 4(c). When first installed, no shift
in calibration has occurred. After aging, the different
calibrations shown in Fig. 4(a) begin to appear. Beginning
at point 1, a single line represents the unchanged A
material between 32'F and 1700°F. From point 2 to
point.3, a line is constructed parallel to the B, line, for
material 8 ; from 3 to 4 the line is parallel to the €I2 line;
from 4 to 5 it is parallel to the B , line; from 5 to 6 it re-
presents B o , the unchanged material. The dashed line
represents the behavior on first immersion, before material
B had been changed in calibration. The difference in emf is
the 'drift' related to a distributed change in calibration
along the wire when the thermocouple is still in the tem-
Fig. 5-Graphical analysis of furnace calibration perature gradient which caused the calibration change.
01 athermocouple with distributed inhomogeneity Now let us attempt a furnace calibration of this thermo-
Thermocouple o f Fig 4 couple as in Fig. 5(a). Figure 5(b) shows the temperature
distribution in the thermocouple due to its insertion into a
muffle furnace with a long region of uniform temperature.
All of the gradient occurs in material which had never
changed its calibration. The emf is derived in Fig. 5(c).
has changed. Somewhere in the thermocouple the emf where the integration through the uniform temperature
characteristics of the wire have changed, perhaps due to region is shown by the single point representing all the
a change in composition by selective oxidation or by material which is at 1700°F. Note that there is no emf
alloying with environmental contaminants. The changed contributed by the changed material since it is not in a
material must have been in a region of temperature temperature gradient. To the casual investigator this probe
gradient when the thermcouple was in service in order to would appear to be perfectly good and would go back into
cause the change in output. From the gradient approach service. Once back in its original environment, with a
it is obvious that this damaged material must be placed temperature gradient in the changed region, it would
under exactly the same temperature gradient condition if continue to give faulty information.
the thermocouple recalibration is to have any significance This example is intended only to provide an under-
In terms of service operation. If the damaged portion of the standing of the effects of a distributed inhomogeneity on
wire is put into a uniform temperature region (such as a service and calibration results. There is little or no informa-
muffle furnace) for calibration, no effect will be seen. tion available from which to construct actual 'service
Furnace calibrations are, in reality, calibrations only of the calibration' solutions such as shown in Fig. 4(c).
length of material at the entrance to the furnace; that is.
the material in the temperature gradient. If the wire is not
homogeneous, a furnace calibration is useless. References
This is illustrated in Fig. 4 using a pair of wires A and B
I . Peltier. M., Ann. chrm. etphys.. 56, 371 (7834).
which have the characteristic that B loses its thermo- 2. Dike, P. H., Thermoelectric Thermometry, Leeds and Northrup
electric power after exposure to a certain set of environ- Co., Philadefph,a (1954).
mental conditions. In Fig. 4(a) are shown the emf vs. 3. Somerfefd, A,, Thermodynamics and Statistical Mechanics,
temperature curves for these hypothetical materials. The Academic Press, fnc , Publishers, New Yark. 156-163 and 342
( 1 956).
positive element is unaffected by its environment. The 4 . Seitz, f . , The Modern Theory of Solids, McGraw-Hill Book Co.,
negative material could show any one of four calibration lnc., New York and London, 178-187 (1940).
lines depending on the temperature to which it had been 5. Roeser, W.F. and Dahl, A.I., "Standard Tables for Chromel-
exposed. The B , curve represents the calibration after Afumel Thermocouples," J. Res. NES, RP 767, 74, 239(1935).
6. Roeser, W.F. and Dahl, A.I., "Reference Tables for lron-
exposure to 1450' F, 8, after exposure to 1550' F and & Constantan and Copper-Constantan Thermocouples, ' ' J Res
after 1650" F service. These values are hypothetical for NBS, RP 1080, 20, 337(1938).
illustration of the technique and do not reflect measured 7. Shenker, H., Reference Tables for Thermocouples, NES
Performance for any material. Circular 561 (1955)

Experimental Techniques 25

Anda mungkin juga menyukai