Mechanical
Laser localization 2 mm 1.5 mm 1 mm
Distance indicator @ Iso 2 mm
Collimator size indicator 2 mm 1 mm
Safety
Door interlock (beam off)
Door closing safety Functional
Audio-visual monitor
Stereotactic interlocks (lockout) NA Functional
Radiation area monitor (if used)
Functional
Beam on indicator
Mechanical
Light/ radiation field coincidence 2 mm or 1% on a side
Light/ radiation field coincidence (asymmetric) 1 mm or 1% on a side
Distance check device for lasers compared with front pointer
1 mm
Gantry/ collimator angle indicators (@cardinal angles) (digital
1°
only)
Accessory trays
2 mm
Jaw position indicators (Symmetric)
Jaw position indicators (Symmetric)
1 mm
Cross-hair centering (walkout)
Treatment couch position indicators 2 mm/ 1° 1 mm/ 0.5°
Wedge placement accuracy 2 mm
Compensator placement accuracy 1 mm
Latching of wedges, blocking tray Functional
Localizing lasers ±2 mm ± 1 mm ≤ ± 1mm
Safety
Laser guard-interlock test Functional
Respiratory Gating
Beam output constancy 2%
Phase, amplitude beam control
In-room respiratory monitoring system Functional
Gating interlock
Mechanical
Collimator rotation isocenter
Gantry rotation isocenter ±1 mm from baseline
Couch rotation isocenter
Electron applicator interlock Functional
Coincidence of radiation and mechanical isocenter ±2 mm from baseline ±1 mm from baseline
Table top sag 2 mm from sagline
Table angle 1º
Table travel maximum range movement in all direction ±2 mm
Stereotactic accessories, lockouts, etc NA Functional
Safety
Follow manufacturer’s test procedures Functional
Respiratory Gating
Beam energy constancy 2%
Temporary accuracy of phase/ amplitude gate on
100 ms of expected
Calibration of surrogate for respiratory phase/ amplitude
Interlock testing Functional
Table 4. Imaging1
Procedure Non-SRS/SBRT SRS/SBRT
Daily
Planar kV and MV (EPID) imaging
Collision interlocks Functional
Positioning/ repositioning ≤2 mm ≤1 mm
Imaging and treatment coordinate coincidence (single gantry ≤2 mm ≤1 mm
angle)
Monthly
Planar MV imaging (EPID)
Imaging and treatment coordinate coincidence (four cardinal ≤2 mm ≤1 mm
angles)
Scaling ≤2 mm
Spatial resolution
Contrast Baseline
Uniformity and noise
Planar kV imaging
Imaging and treatment coordinate coincidence (four cardinal ≤2 mm ≤1 mm
angles)
Scaling ≤2 mm ≤1 mm
Spatial resolution
Contrast Baseline
Uniformity and noise
Annual
Planar MV imaging (EPID)
Full range of travel SSD ±5 mm
Imaging dose Baseline
Planar kV imaging
Beam quality/ energy
Baseline
Imaging dose
Cone-beam CT (kV and MV)
Imaging dose Baseline
Monthly
Orientation of gantry lasers with respect to the imaging plane ±2 mm over the length of laser projection
Annual
Table indexing and position ±1 mm over the scan range
Gantry tilt accuracy ±1 mm over the gantry tilt range
Gantry tilt position accuracy ±1° or ±1 mm from nominal position
Scan localization ±1 mm over the scan range
Monthly
CT number accuracy (4 to 5 different materials) Baseline
In plane spatial integrity (x and y direction) ±1 mm
Field uniformity (most commonly used kVp) Within ±5 HU
Annual
CT number accuracy (electron density phantom) Baseline
Field uniformity (other used kVp settings) Within ±5 HU
Electron density to CT number conversion Consistent with commissioning results and test phantom
manufacturer specifications
Spatial resolution
Manufacturer specification
Contrast resolution
Monthly
Geometric calibration maps OR Replace/ refresh
kV/ MV/ laser alignment
±1 mm
Couch shift: accuracy of motions
Scale, distance, and orientation accuracy
Baseline
Uniformity, noise
High contrast spatial resolution ≤2 mm (or ≤5 lp/cm)
Low contrast detectability Baseline
Annual
CT number accuracy and stability
Imaging dose
Baseline
X-ray generator performance (kV system only): tube
potential, mA, ms accuracy, and linearity
Anteroposterior, mediolateral, and craniocaudal orientations
Accurate
are maintained (upon upgrade from CT to IGRT system)
Long and short term planning of resources (disk spaces, Support clinical use and current imaging policies and procedures
manpower, etc)
Address Areas of Uncertainty Innacuracies in measured data (beam and patient specific) Ongoing
Innacuracies of data entry: digitizer accuracy Weekly ± 2 mm
Innacuracies of data output Weekly
Test of reproducibility
Algorithm Inaccuracy Annually, post-updates ± 2% or ± 2mm in regionsof high dose gradients
Test evaluating calculation accuracy
Function Checks Image registration, autocontouring, deformable registration, etc. Initially, ongoing
Monthly
Setting vs radiation field for two patterns (non-IMRT)
2 mm
Backup diaphragm setting (Elekta only)
Travel speed (IMRT) Loss of leaf speed > 0.5 cm/s
Leaf position accuracy (IMRT) 1 mm for leaf positions of an IMRT field for four cardinal gantry
angles. (Picket fence test may be used, test depends on clinical
planning-segment size)
Annually
MLC transmission (average of leaf and interleaf ±0.5% from baseline
transmission), all energies
Leaf position repeatability ±1.0 mm
MLC spoke shot ≤1.0 mm radius
Coincidence of light field and x-ray field (all energies) ±2.0 mm
Segmental IMRT (step and shoot) test
<0.35 cm max. error RMS, 95% of error counts <0.35 cm
Moving window IMRT (four cardinal gantry angles)
Table 9. Brachytherapy pretreatment QA checks. For routine applications, these tests can be performed by the radiation therapists or source curator.3
System tested Test end point Test methodology
Reusable applicators, e.g., intracavitary applicators All applicator components, caps, cylinder segments, clamps,
tools, and other accessories available. Plastic components not
Completeness, operable condition, and structural integrity
cracked or abraded and fit tightly over metal tubes. Applicator
appropriately sterilized.
Single-use interstitial and transluminal applicators Identity Correct diameter and length
Buttons, connectors, etc, fit on applicators. Applicators not
Correct function
kinked. If indicated, perform simulated treatment on machine
Completeness All accessories present
If active source positioning requires fixed length, sound each
Positional accuracy
applicator
Templates and other custom devices Correct function Correct hole diameter
Identity Review spacing, dimensions, etc., with physician or physicist
Waiting room readiness All required equipment and medical resources available Arrangement in advance. Verify on day of procedure
Remote afterloader Correct function of device and safety accessories Execute daily QA protocol the morning of the procedure
1. Khan F, Gibbons J. Quality Assurance In: Khan F, Gibbons, eds. Khan’s The Physics of Radiation Therapy. 5th ed. Philadelphia, PA: Lippincott Williams and Wilkins; 2014: 371-403.
2. Thomason C, Vann AM, Lenards N. QA TPS, IMRT and SRS. [SoftChalk]. La Crosse, WI
3. Kubo HD, Glasgow GP, Pethel TD, et al. High-dose rate brachytherapy treatment delivery: Report of the AAPM radiation therapy committee task group no. 59. Med Phys. 1998;(25):375-403.