Anda di halaman 1dari 7

Effects of Prolonged Reading on Dry Eye

Sezen Karakus, MD,1 Devika Agrawal, BS,1 Holly B. Hindman, MD, MPH,2 Claudia Henrich, MD,1,3
Pradeep Y. Ramulu, MD, PhD,1 Esen K. Akpek, MD1

Purpose: To demonstrate the effects of prolonged silent reading on tear film and ocular surface parameters.
Design: Prospective, observational clinical study.
Participants: A total of 177 patients with dry eye and 34 normal controls aged 50 years and older.
Methods: After evaluating symptoms using the Ocular Surface Disease Index (OSDI) questionnaire, the
following tests were performed in consecutive order: automated noninvasive tear break-up time (TBUT), surface
asymmetry and regularity indices, Schirmer’s testing without anesthesia, corneal staining using fluorescein, and
conjunctival staining using lissamine green. The participants were then asked to read a 30-minute validated
passage silently. The tests were repeated after the reading task.
Main Outcome Measures: Changes in tear film and ocular surface parameters after reading.
Results: All parameters, with the exception of surface asymmetry index, worsened after the reading task in
patients with dry eye and in controls. The worsening reached a statistical significance for corneal and conjunctival
staining in the dry eye group (P < 0.001) and for corneal staining in the control group (P < 0.01). At baseline, OSDI
scores correlated only with corneal and conjunctival staining scores (r ¼ 0.19, P ¼ 0.006 and r ¼ 0.27, P < 0.001).
Among postreading measurements, baseline OSDI scores correlated with TBUT (r ¼ 0.15, P ¼ 0.03) in addition
to corneal and conjunctival staining (r ¼ 0.25, P < 0.001 and r ¼ 0.22, P ¼ 0.001). Changes in TBUT and
Schirmer’s test correlated significantly with their respective baseline values (r ¼ 0.61, P < 0.001 and r ¼ 0.44,
P < 0.001), indicating that the more unstable the tear film and the lower the aqueous tear secretion, the worse
they became after the prolonged reading task. Worsening in corneal staining directly correlated with the baseline
conjunctival staining (r ¼ 0.17, P ¼ 0.02) and surface regularity index (r ¼ 0.21, P ¼ 0.01).
Conclusions: Evaluating tear film and ocular surface parameters at rest may miss clinical findings brought
about by common everyday tasks such as reading, leading to discordance between patient-reported symptoms
and clinician-observed signs. Quantifying dry eye after visually straining activities such as prolonged silent
reading may help better understand patient symptomatology. Ophthalmology 2018;-:1e7 ª 2018 by the
American Academy of Ophthalmology

Modern lifestyle requires increasingly demanding visual with dry eye with perfect visual acuity and mild corneal
tasks such as book reading for leisure or use of electronic staining report that they are unable to read comfortably
devices for work-related productivity.1 Previous because of their symptoms. We hypothesized that quanti-
questionnaire-based studies indicated self-reported diffi- fying dry eye at rest may miss clinical findings brought about
culty with reading as one of the most common symptoms of by common daily tasks such as reading. Therefore, in this
dry eye, suggesting that reading difficulty is an important study, we aimed to quantify the objective dry eye pa-
component of how dry eye affects the quality of life of rameters measured at baseline and after sustained reading,
affected patients.2-6 In fact, symptoms after sustained and correlate them with baseline patient symptoms.
reading are not unique to patients with dry eye, and even
healthy individuals with no prior dry eye diagnosis experi-
ence visual symptoms after prolonged reading.7,8 Computer Methods
work is particularly problematic, with prolonged use of
computer displays causing significant worsening in dry eye The study protocol was approved by the Johns Hopkins University
symptoms.6,8-10 and the Rochester University Institutional Review Boards in
One shortcoming with regard to the diagnosis and accordance with the Declaration of Helsinki, and the study proced-
treatment of dry eye has been the absence of objective ures were performed according to Health Insurance Portability and
Accountability Act. New and return patients with dry eye, older than
measurable parameters that can help quantify the disease 50 years of age and with a binocular visual acuity of 20/25 or better,
state and evaluate the efficacy of treatments. As clinicians who presented to the Ocular Surface Diseases and Dry Eye Clinic at
who care for patients with dry eye, we are aware of the lack the Wilmer Eye Institute, the Johns Hopkins University (Baltimore,
of correlation between patient-reported symptoms experi- MD), and the Dry Eye Clinic at the Flaum Eye Institute, Rochester
enced on an everyday basis and objective findings as University (Rochester, NY) were identified by a study coordinator.
currently measured in the clinic. For example, many patients Patients with self-reported illiteracy or language problems, contact

ª 2018 by the American Academy of Ophthalmology https://doi.org/10.1016/j.ophtha.2018.03.039 1


Published by Elsevier Inc. ISSN 0161-6420/18
Ophthalmology Volume -, Number -, Month 2018
lens wearers, and patients who had any ocular surgery within the within the study groups. Pearson and Spearman’s rank correlation
preceding 3 months were not considered. Patients who did not take coefficients were used to analyze the associations between variables.
any prescription eye drops (including topical cyclosporine, steroids, Values of P < 0.05 were considered statistically significant. All
and antiglaucoma eye drops) within 30 days of the study visit were statistical analyses were performed using IBM SPSS Statistics
consented by a study investigator. Individuals who were using over- version 23 (IBM Corp., Armonk, NY).
the-counter drops (including artificial tears) were asked to refrain for
24 hours before the study visit. Similarly aged accompanying friends
or family members, as well as healthy volunteers, with no history of Results
dry eye or current dry eye symptoms, or other ocular diseases
requiring topical treatment were recruited as controls at both sites. A total of 224 participants completed the study-related procedures.
First, the Ocular Surface Disease Index (OSDI) questionnaire was Thirteen participants who finished the reading task in less than 20
administered to evaluate dry eyeerelated symptoms. A total OSDI minutes (8 patients with dry eye and 5 controls) were excluded.
score and 3 subscoresd(1) ocular symptoms subscore (questions 1 to The remaining 211 patients (177 patients with dry eye and 34
5), (2) vision-related function subscore (questions 6 to 9), and (3) controls) were included in the analyses.
environmental triggers subscore (questions 10 to 12)deach ranging Table 1 displays the baseline clinical characteristics in each
from 0 to 100 were calculated as previously described.11 group. There were no significant differences between patients
The following tests were then administered to both eyes in the with dry eye and controls with regard to age or sex. Mean
order listed with at least 5-minute breaks in between. First, the Tear duration of reading was 28.3 minutes (20.6e31.6 minutes; standard
Stability Analysis System incorporated into the Tomey Top-Ref- deviation, 2.2) in the dry eye group and 27.7 minutes (20.2e30.5
Keratometer RT-7000 (Tomey, Phoenix, AZ) corneal topography minutes; standard deviation, 2.8) in the control group (P ¼ 0.15).
machine was used to measure noninvasive tear break-up time At baseline, patients with dry eye had a higher mean symptom
(TBUT). Values greater than 3.0 seconds were considered normal as score and objective clinical parameters compared with controls
previously recommended.12 The surface irregularity indices (surface except for the surface regularity indices (Table 1).
asymmetry index and surface regularity index) were measured by the All measured parameters worsened in the dry eye group and in the
Tomey TMS-4 videokeratoscopy (Tomey, Waltham, MA). Greater controls (with the exception of surface asymmetry index); however,
values represent greater irregularity of the ocular surface.13 Finally, not all of these changes were statistically significant (Table 2). The
Schirmer’s testing was performed without anesthesia using sterile worsening reached a statistical significance for corneal and
strips (Tear Flo; Sigma Pharmaceuticals, Monticello, IA), and the conjunctival staining in patients with dry eye (P < 0.001 and P ¼
length of paper wetting was measured after 5 minutes. 0.001, respectively) and for corneal staining in the control group
Ocular surface staining was graded according to the Sjögren’s (P ¼ 0.003).
International Collaborative Clinical Alliance grading system.14 Table 3 displays correlations between OSDI scores and tear film
Conjunctival staining was evaluated first, using a neutral density and ocular surface parameters at baseline, postreading, and the
filter over the light source immediately after instillation of change from baseline. The total OSDI score and all 3 OSDI
lissamine green dye (Green-Glo; HUB Pharmaceuticals, LLC., subscores (ocular symptom, vision-related function, and environ-
Rancho Cucamonga, CA). The nasal and temporal conjunctiva, mental triggers) correlated significantly with baseline corneal and
within the interpalpebral fissure, were graded separately with a conjunctival staining. The strongest correlation was observed between
maximum score of 3 for each area. The area where the Schirmer’s the environmental triggers subscore and baseline temporal conjunctival
test strips were placed was not taken into consideration when staining (r ¼ 0.32, P < 0.001). Although there was no significant
grading the temporal conjunctiva. Corneal staining was evaluated correlation at baseline, the TBUT measured after the prolonged reading
using the cobalt blue filter at least 2 minutes after instillation of test correlated with the baseline total OSDI score as well as ocular
fluorescein dye (Ful-Glo; Akorn, Inc., Lake Forest, IL). The symptom and environmental triggers subscores. In addition, correla-
maximum possible fluorescein score (the punctate epithelial tions between all OSDI scores and the corneal staining became more
erosions grade plus any extra points for modifiers [central corneal robust after the reading task. When correlations between baseline OSDI
staining, confluent staining, and filaments]) was 6 for each cornea. scores and worsening of tear film or ocular surface parameters after
The total possible maximum ocular staining score, derived by reading were analyzed, the increase in corneal staining correlated well
summing the corneal and conjunctival scores, was 12 for each eye.14 with the total OSDI score and environmental triggers subscore. Among
At this point, participants were given a standardized, validated the 3 modifiers of corneal staining (presence of central staining,
reading task consisting of a 7200-word story to be read silently for 30 confluent staining, and presence of filaments), confluent staining at
minutes or until finished.15,16 Participants were allowed to wear their baseline significantly correlated with total OSDI score and all 3 sub-
habitual correction for near vision. Reading was performed in a 12 scores. Central staining was found to correlate with all OSDI scores not
feet wide by 17 feet long room with standardized lighting between at baseline but postreading. Moreover, correlations between confluent
400 and 600 lux15 and environmental conditions (relative humidity staining and OSDI scores became stronger after reading.
level between 30% and 50% and temperature level between 73 F Table 4 displays correlations between baseline tear film
and 75 F). parameters and changes from baseline after the sustained reading
After completion of the reading task, the following tests were task. Changes in TBUT and the Schirmer’s test correlated
repeated as described previously in this order: noninvasive TBUT, significantly with their respective baseline values, indicating that
the irregularity indices (surface asymmetry index and surface the more unstable tear film and the lower aqueous tear secretion,
regularity index), Schirmer’s test, conjunctival lissamine green the worse they became after the prolonged reading task. Of note,
staining, and corneal fluorescein staining. worsening in corneal staining directly correlated with the
baseline conjunctival staining and surface regularity index.
Statistical Methods
Measurements only from the right eye were used for all analyses. A t Discussion
test was used to compare the continuous variables, and a chi-square
test was used to compare categoric variables between groups. A
This study demonstrates that prolonged reading causes
paired t test was used to compare before and after observations
significant alterations on the ocular surface not only in

2
Karakus et al Prolonged Reading and Dry Eye

Table 1. Comparison of Baseline Characteristics According to Dry Eye Status

Variables Dry Eye (n[177) Controls (n[34) P Value


Demographics
Age, yrs 63.2 (8.2) 60.4 (7.9) 0.07
Female, n (%) 131 (74) 23 (67) 0.44
Duration of reading, min 28.3 (2.2) 27.7 (2.8) 0.15
OSDI total score, 0e100 32.0 (22.0) 4.8 (5.6) <0.001
Discomfort-related subscore, 0e100 33.8 (22.8) 6.6 (7.9) <0.001
Vision-related subscore, 0e100 26.1 (24.7) 2.3 (5.6) <0.001
Environmental triggers subscore, 0e100 37.5 (31.1) 5.5 (8.7) <0.001
TBUT, sec 3.4 (3.5) 4.8 (3.7) 0.04
Surface irregularity indices
Surface asymmetry index 0.49 (0.29) 0.49 (0.45) 0.94
Surface regularity index 0.48 (0.30) 0.43 (0.33) 0.49
Schirmer’s test, mm 12.9 (9.0) 17.5 (9.9) 0.009
Total corneal staining, 0e6 1.8 (1.7) 0.4 (0.7) <0.001
Corneal staining score, 0e3 1.4 (1.2) 0.4 (0.6) <0.001
Central corneal staining present, n (%) 33 (19) 0 0.006
Confluent corneal staining present, n (%) 42 (24) 1 (3) 0.006
Filaments present, n (%) 4 (2) 0 >0.99
Total conjunctival staining, 0e6 1.9 (2.2) 0.5 (1.1) <0.001
Nasal conjunctival staining score, 0e3 1.1 (1.2) 0.3 (0.9) <0.001
Temporal conjunctival staining score, 0e3 0.9 (1.1) 0.1 (0.4) <0.001
Total OSS, 0e12 3.8 (3.3) 0.9 (1.5) <0.001

OSDI ¼ Ocular Surface Disease Index; OSS ¼ Ocular Staining Score; TBUT ¼ tear break-up time.
Results are represented as mean (standard deviation) for continuous variables and number (percentage) for a binary variable. The t test was used for
comparison of continuous variables between groups and the chi-square testing for categoric variables. Bolded values represent P < 0.05.

patients with dry eye but also in older individuals with a Environmental factors such as temperature, humidity
healthy ocular surface. The baseline patient symptoms changes, and airflow/wind are well known to exacerbate
measured using the OSDI correlated well with the tear film clinical signs and symptoms in patients with dry eye.17,18 To
parameters and ocular surface staining measured after the quantify the effects of adverse environmental conditions on the
sustained silent reading task, including some of the dry eye ocular surface, the controlled-environment chamber has been
signs that did not seem to correlate at rest (e.g., TBUT). developed, and various versions have been used in clinical
These findings may explain the puzzling discrepancy trials since then.19-22 However, an adverse environmental
between physician-measured dry eye and patient-reported chamber is expensive to build and provides exaggerated
symptoms often observed in clinical practice. environmental conditions rather than mimicking a real-life

Table 2. Changes in Ocular Surface and Tear Film Parameters after Sustained Reading Task

Dry Eye (n[177) Control (n[34)


Variables Baseline After Reading P Value Baseline After Reading P Value
TBUT, sec 3.4 (3.6) 3.1 (3.3) 0.23 4.8 (3.7) 4.3 (3.5) 0.52
Surface irregularity indices
Surface asymmetry index 0.49 (0.29) 0.54 (0.40) 0.21 0.50 (0.45) 0.45 (0.18) 0.62
Surface regularity index 0.48 (0.31) 0.51 (0.32) 0.12 0.43 (0.34) 0.44 (0.26) 0.83
Schirmer’s test, mm 12.6 (8.9) 11.8 (8.1) 0.05 17.5 (9.9) 16.2 (9.7) 0.31
Total corneal staining, 0e6 1.8 (1.7) 2.6 (1.9) <0.001 0.4 (0.7) 1.0 (1.4) 0.002
Corneal staining score, 0e3 1.4 (1.2) 1.8 (1.2) <0.001 0.4 (0.7) 0.7 (1.0) 0.003
Central corneal staining present, n (%) 34 (19) 64 (36) <0.001 0 5 (15)
Confluent corneal staining present, n (%) 42 (24) 75 (43) <0.001 1 (3) 4 (12) 0.12
Filaments present, n (%) 4 (2) 4 (2) 0 0
Total conjunctival staining, 0e6 1.9 (2.2) 2.1 (2.2) 0.001 0.5 (1.1) 0.6 (1.2) 0.13
Nasal conjunctival staining score, 0e3 1.06 (1.2) 1.14 (1.2) 0.03 0.3 (0.9) 0.4 (0.9) 0.18
Temporal conjunctival staining score, 0e3 0.86 (1.1) 0.98 (1.2) 0.001 0.1 (0.4) 0.2 (0.5) 0.16
Total OSS, 0e12 3.8 (3.3) 4.7 (3.5) <0.001 0.9 (1.5) 1.6 (2.0) 0.001

OSS ¼ Ocular Staining Score; TBUT ¼ tear break-up time.


Results are represented as mean (standard deviation) for continuous variables and number (percentage) for a binary variable. The t test was used for
comparison of continuous variables between groups and the chi-square testing for categoric variables. Bolded values represent P < 0.05.

3
Ophthalmology Volume -, Number -, Month 2018

Table 3. Correlations between Ocular Surface Disease Index Scores and Baseline Tear Film Parameters in All Participants
Ocular Symptom Vision-Related Function Environmental Triggers OSDI
r P Value r P Value r P Value r P Value
TBUT

Baseline e0.05 0.50 e0.09 0.20 e0.09 0.28 e0.08 0.27


After reading e0.14 0.04 e0.09 0.18 e0.15 0.04 e0.15 0.03
Change from baseline e0.09 0.19 e0.01 0.87 e0.03 0.63 e0.07 0.35
SAI
Baseline e0.03 0.69 e0.003 0.97 0.04 0.65 e0.001 0.99
After reading e0.007 0.93 e0.01 0.86 0.007 0.93 e0.01 0.90
Change from baseline 0.03 0.72 0.002 0.98 e0.02 0.85 0.003 0.97
SRI
Baseline e0.12 0.15 e0.09 0.29 e0.03 0.72 e0.10 0.25
After reading e0.07 0.39 e0.03 0.69 e0.03 0.73 e0.05 0.54
Change from baseline 0.03 0.71 0.05 0.58 e0.002 0.98 0.03 0.72
Schirmer’s test
Baseline e0.003 0.96 e0.04 0.60 0.009 0.89 e0.009 0.90
After reading e0.05 0.48 e0.12 0.01 e0.07 0.32 e0.09 0.21
Change from baseline e0.03 0.68 e0.05 0.45 e0.08 0.28 e0.06 0.39
Corneal staining
Baseline 0.19 0.006 0.15 0.03 0.17 0.01 0.19 0.006
After reading 0.23 0.001 0.19 0.006 0.25 <0.001 0.25 <0.001
Change from baseline 0.13 0.05 0.12 0.09 0.19 0.006 0.16 0.02
Central staining
Baseline 0.12 0.08 0.10 0.13 0.09 0.17 0.12 0.09
After reading 0.18 0.007 0.17 0.01 0.17 0.02 0.20 0.004
Change from baseline 0.11 0.10 0.11 0.11 0.12 0.10 0.13 0.05
Confluent staining
Baseline 0.21 0.002 0.14 0.04 0.16 0.02 0.19 0.006
After reading 0.24 0.001 0.16 0.02 0.25 <0.001 0.23 0.001
Change from baseline 0.08 0.26 0.05 0.46 0.16 0.02 0.10 0.16
Conjunctival staining
Baseline 0.25 <0.001 0.21 0.002 0.29 <0.001 0.27 <0.001
After reading 0.19 0.006 0.18 0.01 0.25 <0.001 0.22 0.001
Change from baseline e0.17 0.02 e0.09 0.17 e0.11 0.11 e0.14 0.04

OSDI ¼ Ocular Surface Disease Index; SAI ¼ surface asymmetry index; SRI ¼ surface regularity index; TBUT ¼ tear break-up time.
Pearson correlation coefficient (r) was used to analyze the associations between variables. Bolded values represent P < 0.05.

situation. Therefore, the utility of end points measured using an believe that conjunctival staining needs particular attention
environmental chamber in dry eye clinical trials is question- when evaluating patients for dry eye.
able. Our findings indicate that prolonged reading represents Several noninvasive tests have been proposed to evaluate
an inexpensive tool that simulates a real-life desiccating stress the tear film stability, including automated TBUT and
with quantifiable changes on the ocular surface correlating well corneal surface irregularity indices using computerized
with subjective patient symptoms measured at baseline. videokeratoscopy.12,13,27 Noninvasive TBUT was proposed
Indeed, a previous study hypothesized that a reading challenge to be a useful tool to evaluate tear film in dry eye with the
can function similarly to an environmental challenge and in- disadvantage of its variability, especially in patients with
crease the magnitude of dry eye parameters to observable severe dry eye.12 Although in our study the baseline TBUT
clinical findings.23 did not differ between the patients with dry eye versus
Of all the available dry eye tests, corneal fluorescein controls at rest, significant correlations were found
staining is reportedly the most commonly performed, and between baseline OSDI scores and postreading TBUT
the conjunctival lissamine is the least commonly used values, indicating the variability/instability of the tear film,
test.24,25 This could be due to ease/difficulty of access to as previously suggested to be the main mechanism
these dyes or perhaps lack of knowledge or awareness underlying dry eye.28
regarding the significance of each.26 Our study showed that Surface regularity index, but not surface asymmetry
the degree of baseline conjunctival staining was a significant index, was previously reported to be significantly higher in
predictor of the worsening in corneal staining after sustained subjects older than 50 years of age with no significant dif-
reading. In addition, subjective symptoms, particularly ference between subjects with dry eye versus controls.13 In
environmental trigger-related symptoms, showed the stron- our study, although all participants were 50 years of age or
gest correlation with baseline conjunctival staining of all the older, the levels of both indices measured in either group at
dry eye parameters. On the basis of these findings, we any time point were much lower than the levels previously

4
Karakus et al Prolonged Reading and Dry Eye

Table 4. Correlations between Baseline and Change from Baseline Tear Film Parameters in All Participants
DTBUT DSAI DSRI DSchirmer’s DCorneal Staining DConjunctival Staining DOSS
TBUT

r e0.61 e0.01 0.15 e0.07 e0.07 e0.08 e0.11


P value <0.001 0.91 0.08 0.32 0.29 0.23 0.12
SAI
r 0.16 e0.56 e0.44 0.01 0.09 0.05 0.11
P value 0.06 <0.001 <0.001 0.89 0.27 0.51 0.17
SRI
r 0.24 e0.17 e0.50 0.06 0.21 0.11 0.25
P value 0.005 0.04 <0.001 0.49 0.01 0.17 0.002
Schirmer’s test
r 0.00 e0.11 e0.07 e0.44 e0.01 e0.13 e0.09
P value 0.10 0.20 0.40 <0.001 0.89 0.06 0.21
Corneal staining
r 0.03 0.00 0.03 0.07 e0.02 0.04 0.01
P value 0.68 0.10 0.70 0.29 0.77 0.58 0.92
Conjunctival staining
r 0.09 e0.01 e0.01 0.07 0.17 e0.10 0.07
P value 0.19 0.92 0.90 0.31 0.02 0.13 0.31
Total OSS
r 0.08 e0.006 0.01 0.08 0.10 e0.05 0.05
P value 0.28 0.95 0.91 0.23 0.15 0.48 0.44

OSS ¼ Ocular Staining Score; SAI ¼ surface asymmetry index; SRI ¼ surface regularity index; TBUT ¼ tear break-up time.
Pearson correlation coefficient (r) was used to analyze the associations between variables. Bolded values represent P < 0.05.

reported in the literature.13,29 In addition, these indices did ocular surface as controls. Last, the tests were performed
not differ between patients with dry eye and controls at in a nonmasked fashion with the exception of corneal and
baseline or after prolonged reading, and there was no conjunctival staining, which could have affected the results.
correlation with subjective symptoms. However, baseline In conclusion, this study demonstrates that prolonged
surface regularity index measurement correlated signifi- silent reading for as little as 30 minutes causes measurable
alterations on the tear film and ocular surface in individuals
cantly with the worsening in TBUT and corneal staining,
aged 50 years and older. This effect is more pronounced in
indicating that a higher surface regularity index may predict patients with dry eye. The magnitude of these changes
alteration of the ocular surface after visually straining correlates well with the baseline OSDI scores, arguing
activities such as reading. Further studies are needed to against the common notion that dry eye signs and symptoms
clarify these findings. do not correlate. Because assessing dry eye at rest may miss
One of the limitations of this study is that we included clinical findings brought about by common everyday tasks
only older subjects (aged 50 years). The effects of such as reading, we propose this previously validated test as
prolonged reading remain unknown in younger individuals, a tool to quantify dry eye, particularly in a clinical trial
such as students or individuals who work at an office setting setting to assess effects of various treatment modalities.
and perform reading-related functions daily. Second, we do
not know exactly how prolonged reading causes alterations
in the tear film and ocular surface because we did not record References
the blink rate or pattern or measure the tear film secretion
rate or evaporation. Abnormal blink patterns, such as an
incomplete blink or a reduced blink rate during near tasks, 1. Uchino M, Uchino Y, Dogru M, et al. Dry eye disease and
have been studied8,30,31 and may have an effect on the work productivity loss in visual display users: the Osaka study.
Am J Ophthalmol. 2014;157:294e300.
magnitude of presumed desiccation effect during reading. 2. Miljanovic B, Dana R, Sullivan DA, Schaumberg DA. Impact
We do not know whether the effects of reading on the ocular of dry eye syndrome on vision-related quality of life. Am J
surface are due to changes occurring in the composition of Ophthalmol. 2007;143:409e415.
the tear film, such as osmolarity or inflammatory markers, 3. Tong L, Waduthantri S, Wong TY, et al. Impact of symp-
because those were not measured.32 In addition, 13 patients tomatic dry eye on vision-related daily activities: the
Singapore Malay Eye Study. Eye (Lond). 2010;24:
who read the passage in less than 20 minutes were excluded 1486e1491.
(8 patients with dry eye and 5 controls). Although it is 4. Pouyeh B, Viteri E, Feuer W, et al. Impact of ocular surface
entirely possible that those excluded participants read symptoms on quality of life in a United States veterans affairs
more quickly because they had better ocular surface, this population. Am J Ophthalmol. 2012;153:1061e1066.
is not relevant to our research question because our aim 5. van Landingham SW, West SK, Akpek EK, et al. Impact of
was to use prolonged reading as a stressor on the ocular dry eye on reading in a population-based sample of the elderly:
surface and we included normal individuals with healthy

5
Ophthalmology Volume -, Number -, Month 2018
the Salisbury Eye Evaluation. Br J Ophthalmol. 2014;98: 19. Barabino S, Shen L, Chen L, et al. The controlled-environment
639e644. chamber: a new mouse model of dry eye. Invest Ophthalmol
6. van Tilborg M, Kort H, Murphy P, Evans K. The influence of Vis Sci. 2005;46:2766e2771.
dry eye and office environment on visual functioning. Stud 20. Alex A, Edwards A, Hays JD, et al. Factors predicting the
Health Technol Inform. 2015;217:427e431. ocular surface response to desiccating environmental stress.
7. Blehm C, Vishnu S, Khattak A, et al. Computer vision syn- Invest Ophthalmol Vis Sci. 2013;54:3325e3332.
drome: a review. Surv Ophthalmol. 2005;50:253e262. 21. Tesón M, González-García MJ, López-Miguel A, et al. Influ-
8. Chu CA, Rosenfield M, Portello JK. Blink patterns: reading ence of a controlled environment simulating an in-flight
from a computer screen versus hard copy. Optom Vis Sci. airplane cabin on dry eye disease. Invest Ophthalmol Vis Sci.
2014;91:297e302. 2013;54:2093e2099.
9. Yazici A, Sari ES, Sahin G, et al. Change in tear film char- 22. Ousler 3rd GW, Rimmer D, Smith LM, Abelson MB. Use of
acteristics in visual display terminal users. Eur J Ophthalmol. the Controlled Adverse Environment (CAE) in clinical
2015;25:85e89. research: a review. Ophthalmol Ther. 2017;6:263e276.
10. van Tilborg MM, Murphy PJ, Evans KS. Impact of dry eye 23. Ousler 3rd GW, Rodriguez JD, Smith LM, et al. Optimizing
symptoms and daily activities in a modern office. Optom Vis reading tests for dry eye disease. Cornea. 2015;34:917e921.
Sci. 2017;94:688e693. 24. Behrens A, Doyle JJ, Stern L, et al. Dysfunctional tear syndrome
11. Schiffman RM, Christianson MD, Jacobsen G, et al. Reli- study group. Dysfunctional tear syndrome: a Delphi approach to
ability and validity of the Ocular Surface Disease Index. Arch treatment recommendations. Cornea. 2006;25:900e907.
Ophthalmol. 2000;118:615e621. 25. Bunya VY, Bhosai SJ, Heidenreich AM, et al, Sjögren’s In-
12. Gumus K, Crockett CH, Rao K, et al. Noninvasive assessment ternational Collaborative Clinical Alliance (SICCA) Study
of tear stability with the tear stability analysis system in tear Group. Association of dry eye tests with extraocular signs
dysfunction patients. Invest Ophthalmol Vis Sci. 2011;52: among 3514 participants in the Sjögren’s Syndrome Interna-
456e461. tional Registry. Am J Ophthalmol. 2016;172:87e93.
13. de Paiva CS, Lindsey JL, Pflugfelder SC. Assessing the 26. Wolffsohn JS, Arita R, Chalmers R, et al. TFOS DEWS II
severity of keratitis sicca with videokeratoscopic indices. Diagnostic Methodology report. Ocul Surf. 2017;15:539e574.
Ophthalmology. 2003;110:1102e1109. 27. Liu Z, Pflugfelder SC. Corneal surface regularity and the effect
14. Whitcher JP, Shiboski CH, Shiboski SC, et al. A simplified of artificial tears in aqueous tear deficiency. Ophthalmology.
quantitative method for assessing keratoconjunctivitis sicca 1999;106:939e943.
from the Sjögren’s Syndrome International Registry. Sjögren’s 28. Craig JP, Nichols KK, Akpek EK, et al. TFOS DEWS II
International Collaborative Clinical Alliance Research Groups. Definition and Classification Report. Ocul Surf. 2017;15:
Am J Ophthalmol. 2010;149:405e415. 276e283.
15. Ramulu PY, Swenor BK, Jefferys JL, Rubin GS. Description 29. Goto T, Klyce SD, Zheng X, et al. Gender- and age-related
and validation of a test to evaluate sustained silent reading. differences in corneal topography. Cornea. 2001;20:270e276.
Invest Ophthalmol Vis Sci. 2013;54:673e680. 30. Portello JK, Rosenfield M, Chu CA. Blink rate, incomplete
16. Mathews PM, Ramulu PY, Swenor BS, et al. Functional blinks and computer vision syndrome. Optom Vis Sci.
impairment of reading in patients with dry eye. Br J Oph- 2013;90:482e487.
thalmol. 2017;101:481e486. 31. Argilés M, Cardona G, Pérez-Cabré E, Rodríguez M. Blink
17. Paschides CA, Stefaniotou M, Papageorgiou J, et al. Ocular rate and incomplete blinks in six different controlled hard-copy
surface and environmental changes. Acta Ophthalmol Scand. and electronic reading conditions. Invest Ophthalmol Vis Sci.
1998;76:74e77. 2015;56:6679e6685.
18. Doughty MJ. Consideration of three types of spontaneous 32. Messmer EM, von Lindenfels V, Garbe A, Kampik A. Matrix
eyeblink activity in normal humans: during reading and video metalloproteinase 9 testing in dry eye disease using a
display terminal use, in primary gaze, and while in conversa- commercially available point-of-care immunoassay. Ophthal-
tion. Optom Vis Sci. 2001;78:712e725. mology. 2016;123:2300e2308.

Footnotes and Financial Disclosures


Originally received: December 9, 2017. Supported in part by an investigator-initiated research grant from Allergan,
Final revision: March 13, 2018. Inc., and a grant provided by Jerome L. Greene Sjögren’s Center, Johns
Accepted: March 20, 2018. Hopkins University (Baltimore, MD). The TMS-4 videokeratoscopy was
Available online: ---. Manuscript no. 2017-2809. provided by Tomey (Waltham, MA) for the duration of the study. The
1
The Wilmer Eye Institute, Johns Hopkins University, Baltimore, funding organizations had no role in the design or conduct of this research.
Maryland. HUMAN SUBJECTS: Human subjects, human-derived materials, or hu-
2 man medical records were part of this study protocol. No animal subjects
Flaum Eye Institute, University of Rochester, Rochester, New York. were used in this study. Study protocol was approved by the Johns Hopkins
3
Department of Ophthalmology, University of Ulm, Ulm, Germany. University and the Rochester University Institutional Review Boards in
accordance with the tenets of the Declaration of Helsinki. The study pro-
Presented at: the American Society of Cataract and Refractive Surgery
cedures were performed according to Health Insurance Portability and
Annual Meeting, April 13e17, 2018, Washington DC.
Accountability Act
Financial Disclosure(s):
Author Contributions:
The author(s) have made the following disclosure(s): P.Y.R.: Grants
National Eye Institute, outside the submitted work. Conception and design: Karakus, Henrich, Ramulu, Akpek
Data collection: Karakus, Agrawal, Hindman, Henrich, Akpek

6
Karakus et al Prolonged Reading and Dry Eye
Analysis and interpretation: Karakus, Ramulu, Akpek Abbreviations and Acronyms:
Obtained funding: Not applicable OSDI ¼ Ocular Surface Disease Index; TBUT ¼ tear break-up time.
Overall responsibility: Karakus, Agrawal, Hindman, Henrich, Ramulu, Correspondence:
Akpek Esen K. Akpek, MD, Ocular Surface Diseases and Dry Eye Clinic, The
Wilmer Eye Institute, 600 North Wolfe Street, Woods 372, Baltimore, MD
21287-9238. E-mail: esakpek@jhmi.edu.

Anda mungkin juga menyukai