• LED Intensity - This routine uses the Yellow tile on the WR Roll to record the scanner
response to the lamp at FWA installation or Lamp replacement.
• Gain/Offset - This routine calibrates scanner response to Lamp intensity.
• Shading Profile - This routine sets correction values in the Scanner to compensate for
variations in the Scanner CCD light response.
• Parameters - This routine calibrates the read position, focal point and depth of field for the
Scanner.
• Check Paper Measurement GP 7 - This procedure establishes accurate paper measure-
ment points for the FWA in order to enable accurate performance by FWA applications.
Several other separate routines are embedded in dC 970 and they are performed for the fol-
lowing purpose:
• Calibrate Color Sensor - This routine uses the Yellow tile on the WR Roll to calibrate color
response of the FWA.
• Adjust Replaced Lamp - Performed when the Lamp Assembly is replaced.
• Adjust Replaced Color Calibration Plate - Performed when the WR Roll assembly is
replaced.
Prerequisites
Before performing this routine at installation, ensure that the WR Roll tiles (PL20.1) and both
sides of the Lamp Window (GP9) are clean. Figure 1 WR Bar Code Values
Ensure that there are no IQ defects with the system and that the xerographic subsystem is 3. Check that the values on the Lamp Assembly match the values in NVM. In the example in
operating properly. Figure 2, the value in [988-137] should be 0004 (a zero is added to make four digits).
Procedure
Perform each of the following in order.
3. Press Next and then press Start to adjust LED 2 (Figure 5).
4. Press Next and then press Start to adjust LED 1/2 (Figure 6).
5. If there are no NGs, press Save to save the adjustment values. If NGs appear, values are
displayed with a red bar. Refer to Table 1 to troubleshoot the error code stored in [988-
006].
Table 1
Error Code
Error and Stored in [988-
description Component Action 006]
LED intensity too LED 1 Increase Gain 51
low NVM [988-143]
and readjust Figure 7 dC 970 Main Screen
LED intensity too LED 2 Replace the Illumi- 54
low nation Module 2. Press Start to perform the Adjustment with Gain selected in the pulldown menu (Figure
LED intensity too LED 1&2 Increase Gain 57 8).
low NVM [988-143]
and readjust
LED intensity too LED 1 Decrease Gain 52
high NVM [988-143]
and readjust
LED intensity too LED 2 Replace the Illumi- 55
high nation Module
LED intensity too LED 1&2 Decrease Gain 58
high NVM [988-143]
and readjust
RGB unbalanced LED 1 Replace the Illumi- 50
LED 1/2 nation Module
3. Select Gain from the pulldown menu and ensue that Current Values fall between the
Minimum and Maximum values (Figure 9).
Figure 10 Offset Values After Adjustment
NOTE: NGs will be displayed in Red.
6. If the values are OK, press Save to save the values and proceed to Update Shading
Profile. If the values are NG, continue with this procedure.
7. Remove the Lamp Window and clean both sides of the glass with Lens and Mirror
Cleaner and a lint free cloth GP9.
8. Access the dC330 and ensure that the Lamps light [045-002]. Troubleshoot electrical con-
nections or replace the Assembly (BSD 6.11).
9. Repeat the Gain / Offset adjustment.
Update Shading Profile
1. Enter dC 970 and press the Update Shading Profile button (Figure 11).
4. If the values are OK, proceed to step 5. If the values are NG, continue with step 7.
5. If there are values greater than 1, perform GP15 (WR Roll Evaluation) and GP6 Window
Cleaning and repeat the adjustment. If the adjustment fails again, replace the WR Roll
Assembly (REP 6.5).
Adjust Parameters
1. Enter dC 970 and select Adjust Parameters (Figure 13)
4. If there are no NGs, press Save to save the adjustment values and proceed to Adjust
Parameters.
4. Ensure that there are no NGs displayed in the Illumination Depth Detection Surface val- 6. Ensure that there are no NGs displayed in the Focus Detection Surface values are (Fig-
ues (Figure 15). If NGs are displayed, go to steps 7and 8 to resolve. ure 17). If NGs are displayed, go to steps 7and 8 to resolve.
Figure 15 Illumination Depth Detection Surface Values Figure 17 Focus Detection Surface Values
5. Ensure that there are no NGs displayed in the Scan Position Detection Surface values 7. If the any of the adjustments fail after the first attempt perform the following:
(Figure 16). If NGs are displayed, go to steps 7and 8 to resolve. • Clean FWA window per GP9
• Clean the WR Roll