aerosols of 2 agonists
Otto G Raabe, PhD; Timothy M Wong, BS; Garrett B Wong; James W Roxburgh, MBA, RRT;
S David Piper; and James I C Lee, BS
Background: Various studies have demonstrated the benefits of continuous nebu- diovascular risks such as arrhythmias
lization therapy for delivering aerosols of the 2 agonists such as terbutaline sulfate or and myocardial ischemia.2– 4 Develop-
albuterol sulfate to patients with severe asthma and/or impending respiratory failure. ment of more selective bronchodilators
Objective: The purpose of this investigation was to explicate the operational has decreased potential side effects be-
factors associated with the use of nebulizers for extended aerosol respiratory therapy cause of their selective receptor activ-
including those factors that affect the prescribed aerosol dosages and the relation- ity; repeated treatment has been shown
ship to actual delivery of prescribed drugs to the respiratory airways of the lungs of to be effective.5–13 Various studies
a patient under treatment conditions. have demonstrated the efficacy and
Methods: Operational characteristics and methods have been investigated for use safety of continuous nebulization ther-
of long-running nebulizers for continuous nebulization therapy. Factors considered apy for delivering aerosols of the 2
were particle size distribution, setup conditions, aerosolization concentrations and agonists terbutaline sulfate or albuterol
rates, delivery fraction of aerosol reaching patient, and changes in medication sulfate to patients with severe asthma
concentration during extended operation. With a large volume nebulizer, aerosols and/or impending respiratory fail-
can be delivered to the patient without dilution via a standard open mask for up to ure.14 –27 For example, physicians at the
eight hours without refill. The pneumatic HEART nebulizer with 240 mL reservoir University of Michigan Medical Cen-
was evaluated. ter and at the University of Missouri-
Results: The nebulizer was operated from a single compressed air or oxygen Kansas City School of Medicine found
source and found to provide from 10 to 15 L/min of aerosol with 38 to 50 L of continuous nebulization therapy with
aerosolized medicine per liter of air (or oxygen) and utilize from 30 to 56 mL/hour terbutaline sulfate to be effective for
of medicinal liquid. The mass median aerodynamic diameter of the aerosol droplets the treatment of status asthmaticus in
was found to be about 2.0 m (g ⫽ 2.7). Delivery efficiency to the patient mask pediatric patients who did not improve
was about 90%. The aerosolized medicine delivered to the patient can be increased with the standard therapy of amino-
by adjusting the flow rate of the gas source or changing the solution concentration phylline, methylprednisolone, and in-
of medicine. Typically, several milligrams of drug can be delivered to the patient as termittent nebulized terbutaline sul-
inhaled aerosol per hour of treatment of which about one-quarter can be expected to fate.15,16
be deposited in the lungs. During eight hours of operation the concentration of
Terbutaline sulfate, a 2 agonist, can
medicinal solution increased by about a factor of two because of water evaporation.
be administered by oral, inhalation,
Conclusions: Continuous nebulization therapy is an important means of treating
subcutaneous, or intravenous routes.
patients with severe asthma. Dosage criteria can be established based on the
Undesirable side effects may occur de-
operating characteristics of the nebulizer system, drug solution concentration, and
pending upon the serum level. When
patient respiration.
Ann Allergy Asthma Immunol 1998;80:499–508.
delivered to the patient via aerosol in-
halation, there are fewer side effects
compared with oral or parenteral ad-
INTRODUCTION impending respiratory failure associ- ministration since there is less drug
Continuous intravenous isoproterenol ated with severe asthma.1 This drug absorption into the systemic circula-
has been administered previously with and route of administration, however, tion.17,18 Also, the delivery of aerosol-
has been associated with increased car- ized medicines such as terbutaline sul-
fate results in greater therapeutic
From the Vortran Medical Technology, Inc, efficacy at lower dosage levels because
Sacramento, California. Mercy General Hospital. James I C Lee is an
The work described in this manuscript was employee of Vortran Medical Technology. S
of its direct effect on 2 receptors in
performed at Mercy General Hospital, 3957 J David Piper is a private consultant. Otto G the lung. The biologic half-life of ter-
Street, Sacramento, California and in the labo- Raabe, who supervised this study, is a professor butaline sulfate or other 2 agonists,
ratories of Vortran Medical Technology, Inc, at the University of California, Davis. Vortran however, is relatively short. Peak bron-
3941 J Street, #354, Sacramento, California. Medical Technology supported this research.
Timothy M Wong and Garrett B Wong were Received for publication February 4, 1997.
chodilation from the action of drugs
interns at Vortran Medical Technology. James Accepted for publication in revised form such as terbutaline sulfate on the 2-
W Roxburgh and Brian King are employees of January 29, 1998. receptor sites, occurs for up to one and
THEORY
Nebulizer Operation
The operational characteristics of med-
ical nebulizers are described by the
aerosol droplet size distribution using
aerosol particle mass median aerody-
namic diameter [MMAD, m] and
geometric standard deviation [g],
aerosol droplet output concentration
[a, L, or mL of aerosolized liquid per
L of air or oxygen mixture], evapora-
tion loss [w, L, or mL of evaporated
water per L of air or oxygen mixture],
volumetric flow rate of aerosol [Q,
L/min], nebulized liquid consumption
[A, mL/h] and operating gage pressure
of compressed air or oxygen mixture
[p, psig].31,32 The volumetric flow rate Figure 1. Photograph of the HEART nebulizer (left) and the Mini HEART nebulizer (right) used in
[Q] was measured using rotameters this study.
METHODS
Nebulizer
The nebulizer setup is shown in Figure
3. This arrangement for continuous
nebulization therapy utilizes a standard
open mask and a single-patient, dis-
Figure 2. Overview of the deposition in the respiratory tract of aerosol particles inhaled by mouth as posable nebulizer having a high-output
a function of particle geometric diameter for spherical particles of density 1 g/cm3 equal to the pneumatic nebulization nozzle oper-
aerodynamic equivalent diameter for particles larger than 0.5 m and the diffusive diameter for particles ated in a large volume reservoir (240
smaller than 0.5 m (Adapted and reproduced with permission from Bronchial Asthma, Principles of mL) to produce 10 to 15 L/min of
Diagnosis and Treatment, 2nd ed, Grune & Stratton, 1986, from Raabe et al).35
relatively high concentrations of ther-
apeutic aerosol in the respirable drop-
let diameter range.35,36 The arrange-
that prescribed treatment is applicable where Vt (mL), is the total volume of ment requires only a single source of
even though the actual delivery of drug solution to be nebulized over the time compressed air or oxygen mixture. No
to the lungs may be far less than the duration of treatment, T (h). The vol- humidifier is used because high-output
amount nebulized. ume of physiologic saline required to nebulizers produce an output flow of
Preparation of medicinal solution is dilute the drug stock solution is given liquid aerosol that is large enough to
based on nebulization conditions. For by: maintain humidity near saturation as a
an aerosolization dosage of D (mg/h), result of the excess water present and
the concentration of the solution, Cn Vs ⫽ 共A ⫻ T兲 ⫺ Vm 共mL兲 (6) the naturally elevated vapor pressure
(mg/mL), is given by: of the small droplets. At room temper-
where Vs (mL) is the volume of saline ature, 100% relative humidity is
D and Vm (mL) is the volume of medic- reached when the absolute humidity
Cn ⫽ 共mg/mL兲 (3)
A inal drug stock solution combined to exceeds about 20 L of liquid water
prepare the total volume of solution. evaporated in each liter of air.31 More
for an average liquid output of A
(mL/h) from the nebulizer. In this
working calculation the evaporation of
liquid is tacitly but intentionally ig-
nored. To prepare the solution at this
concentration, the stock solution of the
medicinal drug at a concentration of
Cm (mg/mL) must be diluted with sa-
line.
The volume of stock solution of me-
dicinal drug, Vm (mL), that is required
for T (h) of treatment is given by:
Vm ⫽ 共D ⫻ T兲/Cm 共mL兲 (4)
A detailed summary of a selected dosage calculations instead of the aero- via tubing and mask showed that only
example of the operation of the high sol concentration, a. 10.3% ⫾ 1.0% SD was lost in carrying
output nebulizer at 10 L/min for eight The results of the experimental mea- the aerosols from the nebulizer to the
hours beginning with a full reservoir of surement of evaporation effects with mask. The fluorometric measurement
240 mL is given in Table 3. The ex- three nebulizer units equipped with indicated that only 10.1% ⫾ 1.9% SD
pected increased concentration with tubing and masks is summarized in of the aerosol was lost. This showed
time of operation due to water evapo- Figure 4 showing the measured solu- that about 90% of the aerosol gener-
ration will result in a higher delivery tion concentrations (and standard devi- ated by the nebulizer is actually deliv-
rate of medication during the latter part ation, SD, among the three units) as a ered to the patient. For the measured
of the treatment than during the begin- function of time. The calculated and particle size distribution, the expected
ning. The change of the solution con- measured values were found to be in regional deposition of inhaled aerosol
centration is calculated based on an good agreement. The observed maxi- in the human airways for transoral in-
evaporation loss of 12.1 L/L (Table mum increase in medication concen- halation would be about 15% in the
1). The average liquid consumption tration over eight hours of operation of tracheobronchial conductive airways
rate [A] for this nebulizer is 30 mL/h at the nebulizer was about a factor of and about 50% in the alveolar gas-
a flow rate of 10 L/min which will two. exchange region (Fig 2).36 This equates
remain constant over the treatment du- Gravimetric measurements of the to about one-quarter of the aerosoliza-
ration. Table 3 assumes a start with delivery of aerosol from the nebulizers tion dosage being deposited in the
240 mL with drug solution concentra-
tion of 1 mg/mL. After one hour of Table 3. Calculated Aerosolization Characteristics of a Typical HEART Nebulizer*
nebulization, this nebulizer delivered
25 mg of medication, which leaves 215 Start Beginning Liquid Medication Aerosolization Medication
mg of medication in 210 mL of solu- Time Volume Consumed Concentration Dosage Remaining
(hour) (mL) (mL/h) (mg/mL) (mg/h) (mg)
tion. The concentration therefore, in-
creased to 1.03 mg/mL after one hour 0 240 30 1.00 22.8 240
of nebulization. The medication aero- 1 210 30 1.03 23.5 216
solization rate will continue to increase 2 180 30 1.07 24.4 193
gradually as calculated in Table 3. 3 150 30 1.12 25.5 168
4 120 30 1.18 26.9 142
During the full eight hours of opera-
5 90 30 1.27 28.9 114
tion, essentially all of the medication is 6 60 30 1.40 31.9 84
aerosolized so that the average aerosol- 7 30 30 1.65 37.7 50
ization dosage rate [D] of medication 8 ⬇0 [⬇2.00] 0
can be assumed to be equal to the total
liquid consumption rate [A] of 30 Mean 30 1.32 30
mL/h times the initial solution concen- * Conditions: 10 L/min over eight hours beginning with a normalized medication concentration
tration [Cn] of 1 mg/mL ⫽ 30 mg/h. of 1 mg/mL using Eq. 2. Lower concentrations for aerosolization dosages smaller than 30 mg/h
For practical purposes the liquid con- can be derived from this table by multiplying the medications columns (three right columns) by
sumption rate, A, as given in Tables 1 the actual drug concentration of the nebulized solution in mg/mL. Experimentally measured
and 2 can be used for aerosolization values of concentration are compared to the calculated values in Figure 4.