Anda di halaman 1dari 3

VENTILATOR

Brand Maquet
Type Servo-i Infant
Origin Sweden
Technical Data of Servo- i Infant
1 Applicable to : Infant (Can be upgraded to universal level for all patient categories)
Patient weight : 0.5 - 30 Kg

2 General
Modular and upgradeable system : Interchangeable module with other system
Dimension : User interface : 355 x 53 x 295 mm
: Patient Unit : 300 x 205 x 415 mm
Weight (without stand) : ± 20 kg (without trolley)
Method of triggering : Flow and Pressure (can be upgraded to neural trigger)
Max. Operating Pressure : 115 cmH2O
Bias Flow : Infant : 0.5 l/min
Inlet Gas Pressure : 200-650 kPa/ 2.0 - 6.5 bar/ 29-94 PSI
Unavailable Gas/loss of Gas pressure : The flow from an unavailable gas (air/O2) is automatically compensated for so
that patient gets the preset volume
Power Supply : 100-120 VAC 10 % 50-60Hz or 220 - 240 10% 50-60 Hz
External : 12 - 15 VDC, 10 A
Battery backup : 2 Battery module, up to 6 battery can be included
Battery Capacity : Rechargeable, 12 V, 3.5 Ah each
Battery Backup Time : at least 3 H (when Using 6 Batteries)
Max Power Consumption : 140 W
Operating temperature : +10 to + 40 ºC
Relative humidity : 15 -95 % non-condensing
Atmospheric pressure : 660 to 1060 hPa
Oxygen sensor lifetime : > 5 years
3 General Features - The Suction Support function pauses the ventilator from cycling during a
tracheal suction procedure including configurable pre/post oxygen
concentration.
- User interface can be operated via touch screen or by means of the Main
Rotary Dial. Direct access knobs for vital settings such as PEEP, O2
concentration.
- Left & Right Orientation
- Possibility to save event log, trends, data of screen pictures (bitmap)
- Will automatically prompt a Pre-use check where all important parameters are
calibrated or checked and documented in the memory. Patient circuit can be
tested separately from the Pre-use check.
- In supported modes the termination criteria for the breath is set by the user
(Inspiratory Cycle off), in order to overcome causes of patient – ventilatory
asynchrony

- Cleaning procedure involves only one single interchangeable component.


- The flow sensor is compatible for all categories patient (do not need to be
replaced when changing patient category)
- Ventilator will return automatically to the supported mode, after an apnea
period that has caused the ventilator to start Backup ventilation
- Including Nebulizer
4 User Interface - Can be rotated and tilted for maximum flexibility. Can be removed and
clamped for flexibility in placement e.g. bed, wall mount, IV pole, ceiling supply
unit and a 2.9 m cable including cable management.
- Display 12,1 inch, Color, Touch Screen, high visibility from any position
- Automatic brightness adjusted
5 Ventilation Modes
Pressure Controlled Ventilation : PC
Volume Controlled Ventilation : VC
Pressure Supported Ventilation/ Continous Positive Airway : PS/CPAP
Pressure Ventilation
Synchronized Intermittent Mandatory Ventilation Based on : SIMV (PC) + PS
Pressure Controlled Ventilation with Pressure Support

Synchronized Intermittent Mandatory Ventilation Based on : SIMV (VC) + PS


Volume Controlled Ventilation with Pressure Support

Pressure Regulated Volume Controlled Ventilation : PRVC


Synchronized Intermittent Mandatory Ventilation Based on : SIMV (PRVC) + PS
Pressure Regulated Volume Controlled Ventilation with
Pressure Support
Non- Invasive pressure controlled ventilation : NIV Pressure Control
Non- Invasive pressure Support ventilation : NIV Pressure Support

6 Advanced Ventilation Modes


: Nasal CPAP
: Open Lung Tool Software
: CO2 Analyzer

7 Ventilation Setting
Tidal volume : 2 to 350 mL
Inspiratory Minute Volume : 0.3 - 20 l/min
Apnea, time to alarm : 5-45 s
PC/PS above PEEP (cmH2O) : 0 - (80- PEEP)
PC/PS above PEEP in NIV (cmH2O) : 0 - (32-PEEP)
PEEP : 0 - 50 cmH2O
CMV Frequency : 4 - 150 bpm
SIMV Frequency : 1 - 60 bpm
Breath cycle time : 0.5 - 15 s
P high : (PEEP+1) - 50
T high : 0.2 - 10 s
T PEEP : 0.2 - 10 s
PS above Phigh : 0 - (80 - Phigh)
O2 Concentration : 21 -100 %
I : E Ratio : 1:10 - 4:1
T Insp. : 0.1 - 5 s
T pause : 0 - 30 % of breath cycle time
T pause : 0 - 1.5 s
Flow trigger sensitivity level : 0 -100%
Pressure trigger sensitivity : -20 - 0 cmH2O
Insp. Rise time : 0 - 0.2 s
Insp. Cycle off : 1 - 70 % of peak flow
Nebulizer time : 5 -30 min
Oxygen Breath : 100 % for 1 minute
Start Breath : initiation of 1 breath
Alarm silence / reset : 2 minutes silence and reset of latched alarms
Compliance Compensation : On/Off
Nebulizer : On/Off

8 Alarms
: Airway pressure max
: Exp. Minute volume max/min
: Apnea
: Respiratory Frequency
: Exp. Pressure max/min
: High Continous Pressure
: O2 concentration
: Gas supply max/min
: Battery
: Leakage
: No patient effort in NIV
: Technical Alarm

9 Waveform and loop presentation


Real time waveform : Pressure Waveform
: Flow Waveform
: Volume Waveform
: Realtime combination : PV Loop and flow, Pressure, Volume waveform
Loops : Volume/Pressure
: Flow/Volume

10 Log function
Event Log : Alarm
: Ventilator Setting
: Apnea Periods
: Immediate function
Service Log : Technical Alarm
: Test Result
: Preventive Maintenance
: Service History
: Configuration log.

11 Communication/Interface
: Serial Port

12 Accessories :
Test lung 1 pcs
Test Tube 1 pcs
Expiratory cassete 2 pcs
AGA O2 Hose, 3m 1 pc
AGA Air Hose, 3m 1 pc
Mobile cart 1 set
Support Arm 1 set
Nebulizer 1 set
Bacterial filter 50 Pcs 1 box
Humidifier Holder 1 pc
F & P Humidifier MR 850, English 1 set
Temperature & Flow Probe 1 pc
Reuseable kit MR 850 1 set
Humidifier Chamber Infant 1 pc
NIV Reusable Mask 1 Pc
Breath syst 10mm (Aktive Humid)Sili 1 set
Angled adapter 22-15mm/15mm 1 pack
CO2 Analyzer Module 1 Pc
CO2 Analyzer Software 1 Pc
etCO2 CAPNOSTAT III sensor 1 Pc
Adap etCO2 Airway Neo 1 Pc
Miniflow Adapter 1 Pc
Bonnet with Measuring Tape 1 set
Nasal Prong or Nasal Mask 1 Pc

Anda mungkin juga menyukai