Anda di halaman 1dari 5

ARTICLE IN PRESS

Atmospheric Environment 38 (2004) 6425–6429


www.elsevier.com/locate/atmosenv

A field evaluation of a SO2 passive sampler in tropical


industrial and urban air
Lı́cia P.S. Cruz, Vânia P. Campos, Adriana M.C. Silva, Tania M. Tavares
Analytical Chemistry Department, Chemistry Institute, Federal University of Bahia, Salvador, Bahia 40210-340, Brazil

Received 28 November 2003; received in revised form 22 June 2004; accepted 12 July 2004

Abstract

Passive samplers have been widely used for over 30 years in the measurement of personal exposure to vapours and
gases in the workplace. These samplers have just recently been applied in the monitoring of ambient air, which presents
concentrations that are normally much smaller than those found in occupational environments. The locally constructed
passive sampler was based on gas molecular diffusion through static air layer. The design used minimizes particle
interference and turbulent diffusion. After exposure, the SO2 trapped in impregnated filters with Na2CO3 was extracted
by means of an ultrasonic bath, for 15 min, using 1.0  102 mol L1 H2O2. It was determined as SO2 4 by ion
chromatography. The performance of the passive sampler was evaluated at different exposure periods, being applied in
industrial and urban areas. Method precision as relative standard deviation for three simultaneously applied passive
samplers was within 10%. Passive sampling, when compared to active monitoring methods under real conditions, used
in urban and industrial areas, showed an overall accuracy of 15%. A statistical comparison with an active method was
performed to demonstrate the validity of the passive method. Sampler capacity varied between 98 and 421 mg SO2 m3
for exposure periods of one month and one week, respectively, which allows its use in highly polluted areas.
r 2004 Elsevier Ltd. All rights reserved.

Keywords: Passive sampling; Sulphur dioxide; Diffusive sampler; Environment monitoring; Brazil

1. Introduction Adverse health effects, associated with SO2 concentra-


tions in ambient air, include respiratory and cardiovas-
Sulphur dioxide is emitted into the atmosphere by cular diseases. The maximum permissible limits for SO2
natural sources (volcanoes, sulphur gases oxidation in Brazil is 365 mg m3 as a 24 h average and 80 mg m3
produced by plant decomposition) and also by anthro- as mean annual (CONAMA, 1990), although
pogenic sources, such as the combustion of sulphur- the guideline from World Health Organization
containing fossil fuels (coal, oil and natural gas), the (WHO, 2000) recommends daily levels not exceeding
smelting of ores (mostly copper, lead and zinc), and 125 mg m3 and mean annual levels below 50 mg m3.
other industrial processes (Cullis and Hirschler, 1980). SO2 is oxidized in the atmosphere to sulphuric acid,
by reactions occurring in the gaseous phase, in the
Corresponding author. Tel./fax: +55-71-237-4024. liquid phase, or on the surfaces of solids. It contributes
E-mail addresses: lila@ufba.br (L.P.S. Cruz), to acidic deposition in the form of dry-deposited
vaniaroc@ufba.br (V.P. Campos), gas or wet deposition form, when dissolved in clouds,
adriana@laquam.qui.ufba.br (A.M.C. Silva), fog, rain, or snow. All of these deposition forms
ttavares@ufba.br (T.M. Tavares). impact humans, aquatic ecosystems, forests and

1352-2310/$ - see front matter r 2004 Elsevier Ltd. All rights reserved.
doi:10.1016/j.atmosenv.2004.07.022
ARTICLE IN PRESS
6426 L.P.S. Cruz et al. / Atmospheric Environment 38 (2004) 6425–6429

agriculture, monuments and buildings (Finlayson-Pitts


and Pitts Jr., 1986).
Passive samplers are devices capable of taking samples
of gas or vapour pollutants from the atmosphere,
without involving air active movement through them.
The fixation rate is controlled by a physical process,
which can be diffusion through a static air layer or
permeation through a membrane (Brown et al., 1984).
The average atmospheric concentration is calculated by
Fick’s first law, using the exposure period (Palmes and
Lindenboom, 1979). Several passive samplers have been
developed since Palmes and Gunnison (1973) published
the principles of passive sampling. The use of these
samplers for gas or vapour monitoring has become an
attractive alternative, due to the advantages that they
present when compared to conventional active techni-
ques. In addition to being simple and cost-effective, this
device presents the following benefits: the possibility of
simultaneous sampling of numerous interesting areas;
easy transportation to difficult locations, such as forests
and mountains, due to its small dimensions and light
weight; silent operation; reusability; a 100% time Fig. 1. Passive sampler diagram.
coverage is obtainable; and does not need field calibra-
tion, sampled air volume measurements, electricity nor
technical personnel at the sampling site. These char-
acteristics indicate that passive samplers are suitable for sampler model has been used in cold climates, with
determining the spatial distribution of gases over large temperatures ranging between 15 and +161C, and in
areas, checking atmospheric transport and deposit dry climates, with very low relative humidity, to
models, screening studies, establishing atmospheric determine SO2. The sorbent used was an NaOH 1%
monitoring networks, personal monitoring, mapping solution in methanol (Ferm and Svanberg, 1998; Ayers
concentrations in cities, and measuring concentrations et al., 1998; Makkonen and Juntto, 1997). The sampler
in remote sites (Carmichael et al., 2003; Cox, 2003; Cruz consists of a cylindrical polyethylene tube (internal
and Campos, 2002). diameter 21 mm, length 12 mm) sealed at the bottom to
Several types of SO2 passive samplers have been avoid convective transport, which produces positive
described with variations regarding dimensions, diffu- errors in determined pollutant concentration. To mini-
sion barriers and pollutant trapping reagents. Ferm and mize internal turbulent diffusion and particle interfer-
Svanberg (1998) carried out comparisons between SO2 ence, a Teflon membrane (pore 0.5 mm, diameter 25 mm)
passive samplers and active techniques at both urban was used in the air inlet. The membrane was protected
and remote sites in Sweden. The results demonstrated from mechanical damage by a stainless steel screen
close agreement between the two methods (within (thread diameter 0.08 mm and mesh size 0.125 mm).
715%). Nevertheless, it lacks validation for tropical After migration through the diffusion layer, SO2 is
environments, where high temperatures and relative trapped on a cellulose filter impregnated with a Na2CO3
humidity are present at all times. solution. Several tests were carried out to compare
This paper evaluated passive sampler performance in absorption efficiency and select the most adequate
the determination of SO2 concentrations in a tropical solution to impregnate the filters, used in the passive
environment. Passive sampler precision was determined sampler that would be exposed under conditions of
and its accuracy was estimated by comparison with relative temperature and humidity present in typically
active methods. tropical air. These tests consisted in bubbling a standard
SO2 atmosphere produced by a commercial permeation
tube (Dynacal), gravimetrically calibrated (permeation
2. Materials and methods rate: 141713 ng min1 SO2) and kept in a thermostatic
bath (25.070.1 1C). Three serial bubblers containing
2.1. Description and preparation of the passive sampler 10 mL of test solution each were used. As carrier gas,
commercial nitrogen was used at a rate of 40 L h1 for
In this experiment, a wide and shorter passive sampler periods of 1 h. The tested solutions and their respective
first designed by Ferm (1991) was tested (Fig. 1). This SO2 absorption were: 1.0  102 mol L1 potassium
ARTICLE IN PRESS
L.P.S. Cruz et al. / Atmospheric Environment 38 (2004) 6425–6429 6427

tetrachloromercurate (TCM: K2HgCl4), 100%; 7 mmol L1 sampling period. To determine if contamination oc-
buffered formaldehyde, 99%; 1.0  102 mol L1 curred during sampler preparation, transport or analy-
Na2CO3, 97%, and 1.0  101 mol L1 triethanolamine sis, the passive samplers’ field blanks followed the same
(TEA: 2,20 ,200 – nitrilotriethanol), 88%. The Na2CO3 trajectory as exposed samples, that is, they were taken to
solution was chosen as the trapping reagent due to other the field and returned to the laboratory at the same
advantages it offers, when compared to other reagents periods. Comparisons between passive sampling and
tested, aside from presenting high SO2 absorption: it is a active techniques were performed in both urban and
stable, non-toxic reagent and the analysis method industrial areas.
utilized, i.e. ion chromatography, is simple and fast.
Samplers were prepared in the following manner: 2.3. Analysis
cellulose filters with 25 mm in diameter were cut to fit
into the sampler bottom. Then, they were submitted to After exposure of the passive samplers, the filters were
three washings with deionized water in an ultrasonic transferred to micro tubes, and the trapped SO2 was
bath, and one with ethanol, for 15 min each. After extracted in an ultrasonic bath, using a 1.5 mL of
drying in a stove at 50 1C, the filters were impregnated 1.0  102 mol L1 H2O2 solution, for 15 min, to ensure
with 1.0  102 mol L1 Na2CO3 solution, for 2 h, and complete oxidation to sulphate. The extract was
dried again before being put into the samplers. All other centrifuged for 5 min, and analysed by suppressed ion
components of the passive sampler, as well as poly- chromatography. The analysis conditions were as
ethylene vials used for storing and transport, underwent follows: 120  6 mm BT I AN 5311111 anion analytical
a similar cleaning treatment. Rubber gloves and tongs column, 1.8 mL min1 flow rate, eluent 2 mmol L1
were used to handle passive samplers and avoid Na2CO3/0.75 mmol L1 NaHCO3 and ion suppresser
contamination. Dionex regenerated with 40 mmol L1 solution of
Passive sampler performance in the laboratory was sulphuric acid. To determine the precision of the
checked by exposing sets of six passive samplers in a analysis (defined as the relative standard deviation) each
Teflon chamber (Cruz et al., 2004). Several experiments sulphate standard solution of the analytical calibration
were carried out with the purpose of studying face curve was analysed four times. An average value of 2%
velocity, exposure time, pollutant concentration, relative was obtained for the precision of the analysis. Its
humidity, exposure to zero concentration and inter- accuracy was estimated at 4% through two artificial rain
ference effects. Storage effect of exposed samplers was samples containing 2.5 and 8.5 mg SO2 1
4 mL . The
also studied. All the tests were based on the protocol detection limit of the method was estimated at the
recommended by the Comité Européen de Normalisation threefold standard deviation of the average field blanks
(CEN, 1998) for passive sampler validation. The for each exposure time. For a 4-week period, the
sampler presented good performance, except for air in detection limit was 0.10 mg SO2 m3.
total stillness condition (0 m s1), for very short ex-
posure periods (0.5 h), for atmosphere with 2.4. Sampler capacity
O3X150 mg m3 and high concentrations of reduced
sulphur compounds (H2SX7 mg m3 and others VSCs Based on the quantity of CO23 used to impregnate the
20–30 mg m3). SO2 sorbent in the passive sampler, and considering the
stequiometry of the reaction, it is noted that the device is
2.2. Sampling capable of absorbing up to 95 mg SO2. Taking different
exposure periods into account, it is estimated that the
Two different tropical sites with varied sources and sampler can be exposed in an atmosphere containing up
levels of SO2 were selected for passive sampler exposure: to 98 mg SO2 m3 and 421 mg SO2 m3 during periods
Vitória, a wind-tunnel type street, in the city of that last 1 month and 1 week, respectively.
Salvador, Bahia, located on the eastern coast of Brazil;
and Lamarão do Passé, a district 60 km NW from the
city of Salvador, downwind of a copper smelter and of 3. Results and discussion
the largest petrochemical complex (Camac- ari, Bahia) in
the Southern Hemisphere. Sets of 3 samplers were 3.1. Field validation
established under a PVC shelter, with concave borders
for protection against rain and sun, at a minimum height When SO2 levels were high enough, stable electric
of 2 m. After an exposure period of 7, 14, 21 or 28 days, power was available and equipment security could be
the samplers were closed, placed into polyethylene vials, assured, continuous monitoring was applied. These
taken to the laboratory, and stored in a refrigerator at conditions were only achieved at Lamarão do Passé,
4 1C until analysis. The SO2 diffusion coefficient was where a SO2 continuous analyser, belonging to the air
corrected to average ambient temperature for each monitoring network of the petrochemical complex, was
ARTICLE IN PRESS
6428 L.P.S. Cruz et al. / Atmospheric Environment 38 (2004) 6425–6429

available. Measurements were made during the rainy results obtained showed accuracy for the SO2 passive
season, with temperature varying between 21 and 27 1C samplers ranging between 10%–32% when applied to
and relative humidity between 85% and 100%. Passive heavily industrial areas and compared to readings
sampler precision, defined as the relative standard produced by the continuous analyser. In the urban
deviation for three simultaneously exposed samplers, area, the accuracy was between 8% and 14%, when
varied between 2.4% and 10%, with an average of 5.1% parallel measurements were made using active sampling
for air concentration levels ranging from 6 to 13 mg with a 1.0  101 mol L1 potassium tetracloromercurate
SO2 m3. In the urban area of Salvador (Vitória street) (TCM) coated diffusion tube (length 35 cm, diameter
the measurements were taken during a period of intense 0.6 cm), air flow of 40 L h1 for 24 h (Table 1). This
rain, with temperatures ranging between 22 and 31 1C, method was chosen because the available SO2 continuous
and relative humidity of 70% and 100%. Passive analyser has a detection limit of 2.6 mg SO2 m3, and
sampler precision was within 4–10%, with an average lower values were expected. Overall, the passive technique
value of 6.2% for SO2 levels ranging between 1.9 and agreed with active techniques within ca. 715%, which is
3.0 mg SO2 m3. considered a reasonable percentile for routine passive
The accuracy of the passive sampler measurements monitoring (Ferm and Svanberg, 1998).
was determined, in the industrial area, by comparison to Based on data regarding passive sampler accuracy
readings performed by a continuous SO2 analyser (Table 1), it is evident that sampler performance was
(Environnement, Mod. AF 21M), and in the urban best in the urban area, as opposed to the industrial one,
area, by comparison to a standard denuder technique in which SO2 concentrations were, on average, 4 times
for SO2. The continuous analyser was calibrated with a higher. A least-square regression analysis of the results
dynamic gas calibrator (Environment, Mod. VE–3M) for the passive sampler versus concentrations measured
and provided readings every 15 min. These were used to by the SO2 continuous analyser showed a good, positive,
calculate the averages corresponding to the passive and linear correlation between the two methods
samplings periods that were performed in parallel. The (R=0.9784, po0.0001).

Table 1
Accuracy and precision of field measurements with SO2 diffusive passive sampler

Duration of Exposure period Site type/location Average concentration (mg SO2 m3) Accuracy Precision
exposure of (%) (%)
passive sampler
Passive sampler Active method

1 week 07/12–19/1999 Industrial 6.070.6 7.8 23 10


07/19–26/1999 1370.5 18 28 3.9
07/26–08/02/1999 7.070.3 7.8 10 4.3
08/02–09/1999 8.270.6 12 32 7.3
2 weeks 07/12–26/1999 Lamarão do 9.070.4 11 18 4.4
Passé—BA
07/26–08/09/1999 T: 21–27 1C 8.270.4 9.1 9.9 4.9
RH: 85–100%
3 weeks 07/12–08/02/1999 8.870.3 10 15 3.4
4 weeks 07/12–08/09/1999 8.470.2 10 20 2.4
03/06–04/03/2003 1070.5 12 17 5.0
04/03–30/2003 5.570.2 6.7 18 3.6
04/30–05/29/2003 7.370.5 8.5 14 6.9
1 week 07/30–08/06/1999 Urban 1.970.2 2.2 14 10
08/06–13/1999 2.170.1 1.9 10 4.8
08/20–27/1999 Vitória St— 3.070.3 2.8 7.1 10
Salvador—BA
T: 22–31 1C
RH: 80–100%
2 weeks 07/30–08/13/1999 2.570.1 2.2 14 4.0
3 weeks 07/30–08/20/1999 2.470.1 2.5 4.0 4.2
4 weeks 07/30–08/27/1999 2.370.1 2.5 8.0 4.4
Average accuracy in comparison to the continuous analyser (industrial area) 1976.9 —
Average accuracy in comparison to the TCM-denuder (urban area) 9.574.0 —
Average global accuracy 1577.4 —
ARTICLE IN PRESS
L.P.S. Cruz et al. / Atmospheric Environment 38 (2004) 6425–6429 6429

4. Conclusions and South America using passive samplers. Atmospheric


Environment 37, 1293–1308.
The SO2 diffusive passive sampler using Na2CO3 filter CEN/TC264/WG11, 1998. Ambient air quality. Diffusive
impregnation was tested under ambient conditions in samplers for the determination of gases or vapour—
Brazil, during periods of exposure ranging from 1–4 requirements and test methods. Comité Européen de
Normalisation, Brussels.
weeks. Its precision varied between 2.4% and 10% for a
CONAMA—Conselho Nacional do Meio Ambiente, 1990.
SO2 concentration range of 1.9–13 mgSO2 m3, when
Resoluc- ão No. 003, Diário Oficial da República Federativa
applied to two different types of tropical environments. do Brasil, Brası́lia, Brazil. Available: http://www.mma.
The field measurement results showed good agreement gov.br.
between passive and active methods, during the same Cox, R.M., 2003. The use of passive sampling to monitor forest
exposure period. The overall accuracy of the measure- exposure to O3, NO2 and SO2: a review and some case
ments was 15%. Considering the growing demands for studies. Environmental Pollution 126, 301–311.
environmental monitoring, passive samplers represent a Cruz, L.P.S., Campos, V.P., 2002. Amostragem passiva de
cost-effective tool for SO2 monitoring. poluentes atmosféricos. Aplicac- ão ao SO2. Quı́mica Nova
25, 406–411.
Cruz, L.P.S., Campos, V.P., Novaes, J.A.P., Tavares, T.M.,
Acknowledgements 2004. Laboratory validation of a passive sampler for SO2
atmospheric monitoring. Journal of the Brazilian Chemical
Society, in press.
Authors acknowledge CAPES for fellowship, FINEP Cullis, C.F., Hirschler, M.M., 1980. Atmospheric sulphur:
for grants and CETREL for supplying the SO2 natural and man-made sources. Atmospheric Environment
continuous analyser’s data from the station located at 14, 1263–1278.
Lamarão do Passé. Ferm, M., 1991. A sensitive diffusional sampler. IVL Report
L91-172. Swedish Environmental Research Institute,
Göteborg, Sweden.
References Ferm, M., Svanberg, P.A., 1998. Cost-efficient techniques for
urban and background measurements of SO2 and NO2.
Ayers, G.P., Keywood, M.D., Gillett, R., Manins, P.C., Atmospheric Environment 32, 1377–1381.
Malfroy, H., Bardsley, T., 1998. Validation of passive Finlayson-Pitts, B.J.F., Pitts Jr, J.N., 1986. Atmospheric
diffusion samplers for SO2 and NO2. Atmospheric Envir- Chemistry—Fundaments and Experimental Techniques.
onment 32, 3587–3592. Wiley, New York.
Brown, R.H., Harvey, R.P., Purnell, C.J., Saunders, K.J., 1984. Makkonen, U., Juntto, S., 1997. Field comparison of measure-
A diffusive sampler evaluation protocol. American Indus- ment methods for sulphur dioxide and aerosol sulphate.
trial Hygiene Association Journal 45, 67–75. Atmospheric Environment 31, 983–990.
Carmichael, G.R., Ferm, M., Thongboonchoo, N., Woo, J.-H., Palmes, E.D., Gunnison, A.F., 1973. Personal monitoring
Chan, L.Y., Murano, K., Viet, P.H., Mossberg, C., Bala, device for gaseous contaminants. American Industrial
R., Boonjawat, J., Upatum, P., Mohan, M., Adhikary, S.P., Hygiene Association Journal 34, 78–81.
Shrestha, A.B., Pienaar, J.J., Brunke, E.B., Chen, T., Jie, T., Palmes, E.D., Lindenboom, R.H., 1979. Ohm’s law, Fick’s law,
Guoan, D., Peng, L.C., Dhiharto, S., Harjanto, H., Jose, and diffusion samplers for gases. Analytical Chemistry 51,
A.M., Kimani, W., Kirouane, A., Lacaux, J.-P., Richard, 2400–2401.
S., Barturen, O., Cerda, J.C., Athayde, A., Tavares, T., WHO—World Health Organization, 2000. Air Quality Guide-
Cotrina, J.S., Bilici, E., 2003. Measurements of sulfur lines for Europe, Second ed. WHO Regional Publications,
dioxide, ozone and ammonia concentrations in Asia, Africa European Series, No. 91, Copenhagen.

Anda mungkin juga menyukai