Calibrate, pg. 3
Find Center of Symmetry, pg. 9
Definition of Structure and Indexing, pg. 9
Measure, pg. 14
Snap Function, pg. 15
This is how you show the Diffraction button bar should it not be visible:
1) Open the Special menu.
2) Select the Edit Button Bars... command.
3) Select the check box in front of the Diffraction entry located in the But-
ton bars list.
4) Close the dialog box by clicking OK.
Installing add-in
2
Should your current software configuration not offer the Diffraction button
bar, you must follow these steps in order to activate the add-in:
1) Open the Special menu.
2) Select the Add-In Manager... command.
3) Select the check box in front of the Electron Diffraction Analysis entry
located in the Available add-ins list.
4) Click the Close button.
Diffraction image calibration
1) Load the diffraction pattern you would like to calibrate into the active im-
age buffer or acquire a diffraction image.
This diffraction pattern of
monocrystalline aluminum
has not been calibrated. The
scale bar depicted therefore
shows the unit "Pixel"
(image points).
2) Click the Snap Function button located in the Diffraction button bar
when wanting to apply this function (see pg. 15). The button appears
clicked if the snap function has been activated.
3) Click the Calibrate button located in the Diffraction button bar.
" The Calibration of the diffraction pattern dialog box is opened.
3
Electron Diffraction Analysis • Diffraction image calibration
Enter lattice spacing 4) In order to be able to calibrate a diffraction pattern, you must know at
least one pair of reflexes in the diffraction pattern, respectively the Mill-
er indices of at least one diffraction ring, as well as the corresponding
lattice spacing.
Enter the lattice spacing for the reflex and its counterpart into the
d-spacing field.
Example calculation If you want to calibrate a diffraction pattern of mono-crystalline aluminum,
you can use the (200) lattice planes for the calibration. You will need the fol-
lowing information to be properly able to specify the lattice spacing:
• Miller Indices
For aluminum, you can use h=2, k=0 and l=0.
• G-value(s)
Aluminum only has one g-value a=0,405 nm.
• Formula used to determine the lattice spaces dhkl
The formula for aluminum is:
a
d hkl = ---------------------------------
2 2 2
h +k +l
The specified values for aluminum result in a lattice spacing of 0,205 nm.
0, 405nm 0, 405nm
d 200 = ----------------------------------- = ------------------------- = 0,2025 nm
2 2 2 2
2 +0 +2
Note Note that you must enter the distance in "real space" into the d-spacing field.
Enter the value 0,202 (nm) for aluminum if you would like to apply the reflex
pair [002] and [002] for the calibration.
Set the unit 5) [nm] is offered as a unit by default. You can change this unit however
you like. To do so, click the Set Unit... button.
In the Set Unit dialog box, select the Å (Angström) entry from the Basic
unit list, if the lattice spacing is to be specified in Angström.
Find Center of Symmetry 6) Enter the camera length used to acquire the current diffraction pattern
into the Current field located in the Camera Length Calibration group.
The camera length is always in [mm].
The camera length is not necessary for the calibration of a single dif-
fraction pattern. It is, however, saved as additional information together
4
with the diffraction pattern. Therefore, the diffraction pattern will still be
properly calibrated even if you do not know the camera length.
Enter the camera length 7) Select the Apply to camera length check box to add the current camera
length to the Available list. The camera length is added to the list as
soon as the calibration has been successfully completed by clicking the
OK button.
Calibrate 8) Click the Calibrate button.
" The mouse pointer jumps into the active image.
9) Position the pointer onto the center of the first reflex, e.g., the [002] re-
flex and click the mouse using the left mouse button.
Position the pointer onto the center of the opposite reflex, in this case
[002], and confirm with the left mouse button.
Diffraction image calibration
10) Click OK to close the dialog box and to conclude the calibration.
11) Use the [Shift + F4] shortcut to show the scale bar in the image. If the
diffraction pattern has been correctly calibrated, the scale bar's unit is
a reciprocal length unit, i.e., 1/nm.
This diffraction pattern of
monocrystalline aluminum
has been calibrated. There-
fore, the unit of the scale bar
has a reciprocal length, e.g.,
"1/nm".
Save calibrated 12) Use the File > Save As... command to save the diffraction image as a
diffraction patterns file.
A calibrated diffraction pattern should be saved in the TIF file format.
Only then is the calibration data preserved.
Electron Diffraction Analysis • Diffraction image calibration
Testing remote control In order to test the remote control, activate the Image > Configure Input >
Magnification tab. Here, click the Read button. The current microscope mag-
nification will now be read out and displayed in the Magnification field.
To test whether or not your system can read out camera lengths, switch your
microscope to diffraction mode and acquire a diffraction pattern. Doubleclick
the image buffer which contains the diffraction pattern in order to open the
Image Information dialog box. Check to see if the camera length is listed in
the transferred microscope information located in the Microscope tab.
7
Electron Diffraction Analysis • Diffraction image calibration
2) Enter the lattice spacing for the reflex and its counterpart into the
d-spacing field.
3) Enter the current camera length into the Current field located in the
Camera Length Calibration group.
4) Select the Apply to camera length check box.
This step is absolutely necessary to calibrate the input channel
and not only the current diffraction pattern.
5) Click the Calibrate button.
6) First click the left mouse button on the first reflex and then on the sec-
ond reflex you want to use for the calibration.
7) Close the dialog box by clicking OK.
8) Carry out the calibration on all the camera lengths.
9) Now, acquire a diffraction pattern with any camera length.
" The acquired diffraction pattern is automatically calibrated correctly.
10) If you want to index or measure the diffraction pattern, click subse-
quently the Find Center of Symmetry button.
Find Center of Symmetry
Note Please note that the calibration data for the camera lengths are saved to-
gether with the input channel. If you are using another input channel for the
acquisition of the diffraction pattern, the already calibrated camera lengths
are no longer available.
Note If you use the Calibrate command to calibrate a single diffraction pattern, the
center of symmetry is usually determined during the calibration. In this case,
you do not need the Find Center of Symmetry command.
Showing the center of The center will be marked by a little red circle. The circle is
symmetry located in the diffraction pattern's overlay, so that you can fade
it in or out as much as you like.
5) Select the Single crystal option if the current diffraction pattern shows
individual diffraction reflexes.
Select the Polycrystalline material if the current diffraction image con-
tains rings.
6) For aluminium, select for example the entries Cubic and Face centered
(F) from the Crystal type group.
" The lattice parameters are automatically adapted to the selected
lattice type. Thus, one does not require three different g-values for
all of the lattice types. In the dialog box, only the fields that must be
filled out for the selected lattice type can be edited.
For example, there is only a single g-value (a) for the "face-cen-
tered cubic" lattice type. Angles α, β and γ are always 90°.
7) Enter the g-values. Aluminum has only one g-value (a = 0,405 nm).
Index reflexes 8) Activate the Indexing diffraction pattern tab.
Diffraction image indexing
11
11) Load the image to be able to view the crystal parameters again. Click
the Definition of Structure and Indexing button again. The Define
crystal tab now shows the structure parameters of the current diffrac-
tion pattern.
6) Position the mouse pointer onto a reflex which you are sure to be able
to identify and click the left mouse button.
" The Select reflexes dialog box will be opened. All of the reflexes are
listed in the dialog box which can be theoretically located near the
clicked position in the diffraction pattern. If you can not find the
reflex which you want to use for calibration, you should recheck the
entries for the crystal structure in the Define crystal tab or re-cali-
brate the diffraction pattern.
7) Enter the Miller indices for the first basis vector or select the appropri-
ate Miller indices from the dialog box's list. Confirm the first basis vector
by clicking OK.
8) Define the rest of the basis vectors.
" All of the reflexes of the current diffraction pattern will be reindexed
with the defined basis vectors.
Two basis vectors are suffi-
cient enough to clearly de-
fine the lattice for a face-
centered cubic lattice.
You define the basis vectors
by clicking two known
reflexes. If the diffraction
pattern has been correctly
calibrated, the clicked reflex,
in this case [111], will be of-
fered in the Select reflexes
dialog box.
imagine the overlay as being a transparent foil which is placed over the
image. Because of this, the image is not altered by the indexing, but rather
only supplemented in the image overlay. In addition, you can always hide or
delete the image overlay. To do so, use the functions of the Overlay button
bar.
In the Standard button bar, click the Overlay Button Bar button to show the
Overlay button bar. To display or hide the overlay of an image, use the Layer
button, for example.
The image overlay only then remains separate from the image, if you have
saved the image in TIF format. All other image formats do not support such
an image overlay. For this reason, the image overlay is automatically burned
into the image when saving the image in another image file format.
Note You should not permanently burn the overlay into the image. The overlay will
then become a permanent part of the image. The image information which
was located below the drawing object, are lost when burning.
Diffraction patterns often have a bit depth of 16 bits and cannot be burned
Diffraction image indexing
Color Select one of 16 possible colors from the Display color palette. This color
then solely effects the color of the labeling when indexing a diffraction
image.
Labeling size You can alter the font with which the reflexes are to be indexed. Open the
Special > Preferences > View tab. Enter the desired font into the Overlay >
Pen size field.
Export to image Select the Show indices in Overlay check box to show all of the Miller indices
overlay in the overlay of the diffraction pattern.
Clear the Show indices in Overlay check box, so that only the position of the
located diffraction reflexes, respectively rings are shown in the diffraction
pattern's overlay instead of the Miller indices.
Select the Show only basis vectors check box to show only Miller indices
which belong to the basis vectors. This button is not available when working
with polycrystalline material.
Export to sheet Select the Results to sheet check box to create a sheet in addition to the in-
dexing. The sheet is created as soon as you click on one of the following but-
tons: Auto-indexing, Define g-vectors, or Refresh.
The status of the Lattice spacing [nm] and Reciprocal lattice spacing [1/nm]
determines which values are to be transferred to the output sheet.
13
Note In doing so, an already existing sheet will be overwritten. If you would like to
maintain the sheet, save it as a file using the File > Save As... command be-
fore creating the next sheet.
The sheet contains all of the
located Miller indices and
the lattice spacing, respec-
tively the reciprocal lattice
spacing pertaining to them.
Electron Diffraction Analysis • Measuring diffraction images
You can only then measure a diffraction pattern, if the following require-
ments have been fulfilled:
• The diffraction pattern has been correctly calibrated.
• The center of the diffraction pattern has been properly defined.
This is how you measure the radius of a diffraction ring or the
distance between the primary beam and a reflex
1) Click the little arrow located next to the Measure button and select the
desired measurement functions. In this case, select the Distances
entry.
" The mouse pointer moves to the image window.
2) Position the pointer onto the diffraction reflex or onto any point on the
diffraction ring and click the left mouse button. The primary beam is
automatically the start point of the measurement.
" The first distance to be measured is shown in the image overlay.
Select the distances you
want to measure directly on
the image. The origin for
each of the measurements
is automatically the primary
beam, respectively the
center of symmetry.
3) Repeat the last steps for each measurement you want to execute.
4) Click the middle mouse button or the [Esc] key to end the measure-
ment.
5) Use the File > Save As... command to save the image together with the
measurement as one file. The image should be saved in the TIF file for-
mat.
Snap Function
Snap Function
Click this button in the Diffraction button bar to activate or deactivate the snap function. The but-
ton appears clicked if the snap function has been activated.
Some of the functions used for the analysis of diffraction patterns require
you to define a reflex on the diffraction pattern via mouse click. The accuracy
of all the results depends on how well the reflex's center has been hit. The
snap function assists you.
If the snap function is active, you simply have to click in the vicinity of a dif-
fraction reflex or ring. Your image analysis program then automatically
searches for a diffraction reflex or ring located in the vicinity of your mouse
15
click.
Index
A
Activate remote control 6
Automatic calibration 8
C
Calibrate 3
D
Definie crystal structure 10
Diffraction button bar 2
Display and Export Options, Indexing 12
Display and Export Options, Measuring 15
F
Find center of symmetry 9
I
Image overlays 13
Indexing 9
Indexing via basis vectors 11
Install 2
M
16
Measure 14
S
Saving diffraction pattern 11
Snap Function 15
T
Test remote control 7