Anda di halaman 1dari 3

NATIONAL METROLOGY INSTITUTE OF ETHIOPIA (NMIE)

INDUSTRIAL METROLOGY DIRETCORATE

This calibration certificate documents the


CERTIFICATE traceability to national standards, which
realize the units of measurement according
OF CALIBRATION to the international system of units
PO BOX 5722
Addis Ababa, Ethiopia
Tel: 251-11- 6517985
Fax: 251-11-6460534
e-mail: info @nmie.net
website : http://www.nmie.net

Date of Issue: 2018-12-01 Certificate number:PTH-1044 Page 1 of 3

Object Moisture Tester

Manufacturer DRAMINSKI

Serial Number 18228

Customer Mamo Kacha Plc

Registration No.: 11-0619

Number of pages of the certificate 3

Date of Calibration: 2018-11-27

This calibration certificate may not be reproduced other than in full except with the permission of
the issuing authority. Calibration certificates without signature and seal are not valid.

Seal Date Approved by Calibrated by

2018-12-01 Henok Kebede Ephrem w/gebreal


_____________________________________________________________________________________________
NATIONAL METROLOGY INSTITUTE OF ETHIOPIA (NMIE)
INDUSTRIAL METROLOGY DIRETCORATE

This calibration certificate documents the


CERTIFICATE traceability to national standards, which
realize the units of measurement according
OF CALIBRATION to the international system of units
PO BOX 5722
Addis Ababa, Ethiopia
Tel: 251-11- 6517985
Fax: 251-11-6460534
e-mail: info @nmie.net
website : http://www.nmie.net

Date of Issue: 2018-12-01 Certificate number:PTH-1044 Page 2 of 3

Calibrated Manufacturer: DRAMINSKI


Object
Type: -

Serial No.: 18228

Identification No.: --
Resolution: 0.1%

Standards  Type: Oven, Digital Balance and Standard method (ISO 6675).

Calibration procedure  The instrument is calibrated using oven resolution of 0.1oc and
balance resolution 0.1mg. one measurements were taken out for
each sample and average results were reported using Standard
method according to (ES 25:2001).
Environmental conditions Temperature: 22.0 oC
Humidity: ---
Validity of calibration The values in this certificate are correct at the time of calibration.
Subsequently the accuracy will depend on such factors as the care
exercised in handling and use of the instrument and the frequency of
use. Recalibration should be performed after a period, which has been
chosen to ensure that the instrument’s accuracy remains within the
desired limits.
NATIONAL METROLOGY INSTITUTE OF ETHIOPIA (NMIE)
INDUSTRIAL METROLOGY DIRETCORATE

This calibration certificate documents the


CERTIFICATE traceability to national standards, which
realize the units of measurement according
OF CALIBRATION to the international system of units
PO BOX 5722
Addis Ababa, Ethiopia
Tel: 251-11- 6517985
Fax: 251-11-6460534
e-mail: info @nmie.net
website : http://www.nmie.net

Date of Issue: 2018-12-01 Certificate number:PTH-1044 Page 3 of 3

Measuring Results

Estimated
Standard reading Instrument
Correction% uncertainty
sample designation % according to , reading%
of
ISO 6675
measurement
Coffee 9.9 8.3 1.6 0.58

Note: - Actual Value = Instrument Reading + correction

The uncertainties are based on root sum square of the contributions with a confidence interval
Not less than 95% and coverage factor k=2.

_____________________________End of Certificate_________________________________