Installation/Service

&
Operators Manual

Model 13-FBT-4WFT-E2

4 Way Float Top X-Ray Table

Tingle X-Ray, LLC


5481 Skyland Boulevard East
Cottondale, Alabama 35453

Telephone (205) 556-3803

DO NOT INSTALL OR OPERATE THIS FLOAT TOP TABLE WITHOUT


FIRST READING AND UNDERSTANDING THIS MANUAL.

1|Page
TXR
Title: User/Assembler Manual Technical Publication
Model 13-FBT-4WFT-E2 780.205.00

Revision Date Implemented Reason for Approved By


Change
0 6-7-2007 New Edition Jerry Tingle
1 8-10-2009 Table Redesign Jerry Tingle
2 10-12-2009 Various Changes RD/BB/DH
3 6-10-2011 Table Top Redesign RD/BB/DH
4 1-20-2012 Revised Content B Barnhill
5 3-16-2013 Review & Dana Emakoji
Specification Revision Larry West

2|Page
Table of Contents
DISCLAIMER.............................................................................................................5
SECTION 1 PRE-INSTALLATION...........................................................................5
SECTION 2 INTRODUCTION..................................................................................5
SECTION 3 SAFETY................................................................................................6
SECTION 4 INSTALLATION....................................................................................7
SECTION 5 OPERATING PROCEDURES............................................................15
SECTION 6 SPECIFICATIONS..............................................................................16
SECTION 7 ROUTINE MAINTENANCE................................................................17
SECTION 8 SERVICE........................17
Section 9 Schematics.........................................................................................17

3|Page
IMPORTANT! … X-Ray Protection

X-RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATOR UNLESS


MEASURES OF PROTECTION ARE STRICTLY OBSERVED

X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein should
be thoroughly read and understood before attempting to place this equipment in operation. We will
be glad to assist and cooperate in placing this equipment in use.

Although this apparatus is built to the highest safety standards and incorporates a high degree of
protection against X-radiation other than the useful beam, no practical design of equipment can
provide complete protection. No can any practical design compel the operator to take adequate
precautions to prevent the possibility of any persons carelessly, unwisely, or unknowingly exposing
themselves of others to X-radiation.

It is important that everyone working with X-radiation be properly trained and take adequate steps to
insure protection against injury.

The manufacturer assumes that all operator and service personnel authorized to use, install,
calibrate and maintain this equipment is cognizant of the danger of excessive exposure to X-
radiation, is sufficiently trained and has the required knowledge for it. The equipment herin
described is sold with the understanding that the manufacturer, its agents, and representatives are
not liable for injury or damage which may result from exposure to X-radiation.Various protective
material and devices are available. It is recommended that such materials and devices be used.
------------------------------------------------------------------------------------------------------------------------
!IMPORTANTE! …Proteccion ante los rayos-X
LOS EQUIPOS DE RAYOS-X SON PELIGROSOS PARA EL PACIENTE Y EL OPERADOR A MENOS QUE LAS
MEDIDAS DE PROTECCION SEAN ESTRICTAMENTE OBSERVADAS

Si el equipo de rayos-X no se usa adecuadamente, puede causar lesions. Por este motive, las instrucciones
aqui incluidas se deben leer y comprender en su totalidad antes de intentar poner el equipo en
funcionamiento. Estaremos gustosos de asistir y cooperar en poner el equipo en marcha.

Aunque el equipo esta construido segun las normas de seguridad mas estrictas y presenta un alto grado de
proteccion contra las radiaciones-X, ningun diseno practico puede ofecer una proteccion complete. Tampoco
ningun diseno practico puede obligor al operador a tomar las precauciones adecuadas para prevenir la
posibilidad de que caulquier persona de manera descuidada, poco sensate o ignorante, se exponga a
radiaciones directas o indirectas. Es importante que caulquier persona relacionada con radiaciones-X este
debidamente entrenada y torne las medidas adecuadas para asegurar la proteccion contra posibles lesions.

El fabricante assume que todo operador y personal de servicio authorizado para manejar, instalar, calibrar o
mantener este equipo, es consciente del peligro que conlieva la exposicion excesiva a las radiaciones-X, esta
suficientemente entrenado y posee los conocimiento necesarios para ello. Por lo tanto, el equipo aqui
descrito se vende entendiendo que el fabricante, sus agents y representantes no tienen ninguna
responsabilidad en case de lesions o danos que puedan resultar de las exposicion a dichas radiaciones.

Existen diversos materials y dispositivos protectors, cuyo uso es recomendavie.

4|Page
Section 1
Pre-Installation
Prior to installation, perform any necessary floor preparations to guarantee the floor
is structurally sound and level. Review the table specifications from page 15 of this
manual and map out the position the table will be placed within the room. This
method will ensure you are able to take full advantage of the full range of table top
motion allowed.

Section 2
Introduction
The 13-FBT-4WFT-E2 Float Top Table with Aluminum Extruded Rails has been
designed for use in medical x-ray environments. This table provides cross and
lateral movement of the table-top and lateral movement of a Grid Cabinet or Bucky.
Magnetic locks are provided for cross and lateral table-top movement along with
Bucky/Grid Cabinet movement.

Each table has a name plate (Certification Label) which provides the serial number,
date of manufacture and model identification number. This label is located on the
back right-hand side of the table if you are facing the table.

It is imperative you locate the certified label and have all the information available
from that label when you contact your original installer for service or parts.

Notice: If unique situations or applications arise that are not


covered in this manual, please call for assistance.

5|Page
Section 3
Safety
The Model 13-FBT-4WFT-E2 has a Table-Top capacity of 500 pounds. Great care
must be exercised during the installation, operation and maintenance of this
device.

The Model 13-FBT has a table-top capacity of 300 pounds. Great care must be exercised during the
installation, operation and maintenance of this device.

DO NOT INSTALL OR OPERATE THIS TABLE WITHOUT FIRST


READING AND UNDERSTANDING THIS MANUAL.

In order to provide for the safe operation of this device, Tingle X-Ray, LLC., has identified certain
hazards which may occur during the installation, maintenance and use of this table. The following
procedures have been written to implement safety while using this device. Please read each
procedure below and adhere to these safety guidelines.

 This medical table MUST BE BOLTED to the floor before using.


 If the locks are not working properly, do not place a patient on the table.
 Do not place hands, fingers or hair over the side or underneath the table top.
 Do not put hands or feet under the bottom of the table base.
 Do no perform any repair work while a patient is on the table.
 Do not work under a table, or if the tabletop is in any extended position, without first
centering and securing all movements.
 All personnel and patients must stand clear of the table when the table-top is in any
extended position.
 Do not stand, sit or climb onto the table. PATIENTS MUST BE HELPED AND NOT LEFT
UNATTENDED AT ANY TIME.
 Do not exceed the 500 pound load capacity.
 Do not load a patient onto the table from the ends.
 Do not allow a patient to sit on the ends of the table
 Center the table top, lock it into place, then, assist patient onto the table
from the front and center, and secure them properly before leaving them
for x-ray.

6|Page
Section 4
Installation

A. Uncrating And Inspection

This table assembly has been specially packaged and shipped in a prefabricated
Nefab crate to prevent damage in shipment. Upon receipt of the table, follow all
instructions written on the shipping crate’s panels and on the contents.

TO PREVENT TABLETOP CROSS (TRANSVERSE), AND GRIDCABINET


MOVEMENT DURING SHIPPING: CROSS TRAVEL STRUCTURE AND GRID
CABINET SLIDES ARE PINNED AND STRAPPED. DO NOT REMOVE PIN OR
STRAP UNTIL THE TABLE HAS BEEN MOVED INTO ITS ASSIGNED
FLOOR SPACE, CONNECTED TO 24VAC POWER AND IS CONNECTED TO
EARTH GROUND.

1. Check for signs of damage to the table-top. (All assemblies are factory
inspected then bubble wrap is applied to prevent damage to the top during
shipping or while uncrating.)
2. Check for loose screws, bolts or nuts.
3. Inspect for bent or damaged metal parts.

THE TABLE BASE MUST BE BOLTED TO THE FLOOR WITH SIX


FASTENERS WHICH HAVE A MINIMUM 1275 POUNDS PULLOUT
STRENGTH FOR EACH BOLT. REFERENCE: ASTM STANDARD E 488.

A. The floor in the installation area should be flat, stable and free from surface
defects.
B. Place the table base in the exact operating position. Make sure that all
levelers are in contact with the floor.
C. Wiring circuits must conform to local codes and be installed by qualified
personnel.
D. Tables are pre-wired at the factory for magnetic lock controls, safety and
performance standards.

B. Table Top Preparation

7|Page
The Table Top, when shipped, is locked down by two shipping brackets. These
brackets must be removed prior to operation. The table top panel is attached to the
Table Frame by Velcro. Carefully separate the hooks and loops of the Velcro down
the long side of one side of the table, then, slowly lift the panel top up and away
from the Table Frame.

Next, remove and discard the two shipping brackets shown here in the photo.

Shipping Bracket

Shipping Bracket

8|Page
Remove foam blocks to free the bucky/cabinet movement as shown.

Foam Block

Foam Block

After the brackets and foam blocks have been removed, check for full motion of
table frame and receptor cabinet. If both are moving smoothly with no binding,
replace the table top.

C. Leveling

Depending on the floor conditions, it may be necessary to level the table to the
floor. The table is provided with four leveling feet for this purpose. Using a wrench,
the levelers may be adjusted accordingly. (See Photo). Levelers are shipped in the
fully retracted position.

9|Page
The base must be anchored to the floor. Minimum requirements (4) fasteners with
a 1275 pound pullout strength for each bolt. (4) Anchoring Plates have been
provided for this purpose. See Photo. (Reference: ASTM Standard E 488.)

D. Power Supply

10 | P a g e
The Power Supply Board is located inside the table base under a cover.(See
Photos).

The Power Supply Board supplies all three lock assemblies and requires 24 VAC
Input. The circuit is protected by a 3 amp fuse. If replacement is necessary, replace
with only an equivalent fuse.

24 VDC to
Locks

Foot Switch
Cable

11 | P a g e
24 VAC Input

E. Lock Location & Adjustment

NOTE: You can easily break the banks of the longitudinal locks if you remove or
replace the tabletop improperly. Check the guide supplied on your table top during
shipment and follow proper procedure if adjustments are needed.

After power has been applied to the table, verify lock operation. Locks are adjusted
at the factory for optimum performance, however, if one or more of the locks
require adjustment, complete the following steps:

Longitudinal Locks: The longitudinal locks are located in the table support frame
at the rear of the table. To gain access, remove Table Top Panel (see photos) and
move the table top toward the rear of the table. The set screws which adjust the
table are located underneath each lock assembly left and right. An Allen head
wrench will be needed to adjust both locks. When adjustments are complete,
replace the tabletop carefully.

DO NOT ATTEMPT TO MODIFY OR CHANGE THE ELECTRICAL


CIRCUIT. ANY ALTERATIONS WILL CAUSE DAMAGE TO THE POWER
SUPPLY AND POSSIBLE PHYSICAL INJURY. ONLY AUTHORIZED
QUALIFIED PERSONNEL SHOULD SERVICE THE ELECTRICAL
CIRCUITS.

12 | P a g e
Transverse Lock: The transverse lock is located on the left side of the table base,
below the table top.
1) Remove the cover.

All locks on the table adjust in a similar fashion. Each magnetic lock can be raised
or lowered by loosening the hex nut then turning the adjusting screw in or out for
the appropriate height. The distance between the magnetic lock and the strip it is
adhering to should only be a hairline distance. When power is placed on the lock,
the flex plate on the magnetic lock will close the gap and stop movement of the
area in which the lock is placed.

After all adjustments have been made retighten each hex nut to secure the lock in
the proper position. (See Photo).

13 | P a g e
Locknut

Adjusting
Nut

Replace the covers.

Bucky or Grid Cabinet Lock: The bucky or grid cabinet lock is located underneath
the table top and mounted directly to the middle left-hand side of the cabinet. The
lock can be adjusted to the lock plate using the screws located on the top of the
lock box, as shown.

Receptor Lock

Rear Base Cover

14 | P a g e
Section 5
Operating Procedures
In order to operate the float top table, follow these operating procedures:

1. Read and understand all of the instructions before operating. If the table has
been modified or has accessories, read and understand their instructions.
2. Load the patient, from the front and center of the table, onto the table-top.
3. To position the table top, depress the treadle and move to the desired
position. Releasing the treadle activates all table motion locks.
4. The Receptor Cabinet has its own lock release button located on the side of
the cabinet.
5. Before unloading the patient be sure all locks are engaged. Carefully slide
feet and legs toward the front and center of the table and lift patient up to a
sitting position.

15 | P a g e
Treadle

A. DO NOT LOAD OR UNLOAD A PATIENT ONTO OR OFF OF A MOVING


TOP.
B. IF TABLE LOCKS DO NOT WORK PROPERLY DO NOT USE THE TABLE.
C. DO NOT EXCEED THE RATED LOAD(500lbs).
D. MAINTAIN PATIENT LOAD, CENTER OF GRAVITY, WITHIN 8 INCHES OF
THE LONG AND CROSS CENTER OF THE TABLETOP.
E. DURING EXAMINATIONS: DO NOT ALLOW PATIENTS TO EXTEND
HANDS, FINGERS, HAIR OR BODY THE SIDES OF THE TABLE-TOP.

Section 6
Specifications

Model No. 13-FBT-4WFT-E2


Measurements Inches Measurements
mm
Tabletop Length 85” 2159 mm
Tabletop Width 29 ¼” 743 mm
Height 30” Without Levelers 762 mm
Longitudinal 38” (+/- 19”) 965 +/-
Tabletop Travel 483mm
Transverse 10” (+/- 5”) 254 +/-
Tabletop Travel 127mm
Bucky Travel 11” Total Travel 279 mm

16 | P a g e
Tabletop/Cassette 3.35 85 mm
(International Standard)
Power Required 24VAC 75 VA
Patient Weight 500 pounds 227 kgs
maximum
Table Weight 252 pounds 114.5 kgs
Weight in Crate 550 pounds 250 kgs
Crate Dimensions 37x40x90 940x1016x2286
inches (minimum) mm

Specifications and Features are subject to change without notice

Certification:
This product meets 21CFR Subchapter J

Section 7
Routine Maintenance
All routine maintenance should be performed on a yearly basis. Before
performing any maintenance, shut the power off. Center the table-top over the
base and secure. Perform the routine listed checks.
1. Inspect all bearings for signs of wear. Replace if damaged.
2. Inspect wiring for damage or wear. Replace damaged wire immediately.
2. Check all table base fasteners.
3. Inspect table top for any physical damage.

Section 8
Service

17 | P a g e
There are no user serviceable components/parts on this table. If service is
required, it should only be performed by qualified service personnel.

Section 9
Schematic

18 | P a g e