Anda di halaman 1dari 3

Halaman : 1 dari 1

Nomor : FRM/QCD/E.045/00
DAFTAR CONSUMABLE PART
Tanggal : 30/06/2018

No Instrumen Spare Part Tipe Spesifikasi Ket

1. Printer Port - Paralel port


2. Lamps - Deuterium Lamp & Tungsten Halogen Lamp
1 Spektrofotometer 3. Detector - Silicon Photodiode
4. Glass
- Glass
Cuvettes
Rechargeable betteries = 4 x 1.2 V NiMH
1. Batteries AA rechargeable batteries, type AA (no charging
2 Conductivity meter function)
2. Usb USB
Cable length = Max 3m
interface 1.1
CRYS
OLYT
3 pH meter Crison 1. Electrode Sleeved diaphgragm
Electrl
yte
2. Magnetic Crison
-
Stirrer Stirrer
1. Online
4 TOC Analyzer - Flow rate =m0.5 mL/min
detector
2. Oxidation
- UV lamp 185 nm & 254 nm
reactor
3. Sampling
- -
tube
4. Water
- -
sample bottle
HPLC (High 1. High 1.Infusion system = dual piston, floating
NP-
5 Performance Liquid Pressure piston, single stroke liquid discharge volume
7000C
Chromatography) Infusion Pump
of 30ul.
2. Flow Range = 0,001 – 9,999 ml/min,
increment 0,001 ml/min.
3. Flow Accuracy = ± 0,5% or 10 µL/min.
4.Flow Repeatability = RSD ≤ 0,1%
5. Pressure Fluctuation = ≤ 0,5%
6. Pressure Range = ≤ 42 Mpa (316L pump
head), ≤ 30 Mpa (PEEK pump head)
7.Pipe Link = 1/16” Standard pipe connection
8.Display Parameter = 320x240 graphics
display
9.Power Supply = 85 – 264 VAC, power
150W
2. UV Visible NU- 1. Detection System = 190 – 700 nm
Light Detector 3000C
wavelength,optional single or double
wavelength mode
Catatan : Diperiksa oleh :

Nama :
Tgl :
Halaman : 1 dari 1
Nomor : FRM/QCD/E.045/00
DAFTAR CONSUMABLE PART
Tanggal : 30/06/2018

2. Wavelength range = 190 – 700 nm,


incremental 1 nm, deuterium lamp tungsten
lamps optional
3. Wavelength accuracy = 1 nm
4. Wavelength precision = ± 0,1nm
5. Noise = 1 x 10-5 AU
6. Drift = 2 x 10-4 AU
7. Flow cell volume = depend on model
3. Injection VI-11
-
Valve Japan
Heder
a
4. ODS-
Chromatograp 2 10 cm x 4,6 mm
hy Column H1810
0405.
25
1. Ca
wavel
AAS (Atomic
ength
6 Absorption 1. Light Source
=
Spectrophotometer)
422,7
nm
1. Modulation frequency = 100 Hz (Self –
2. Na
wavel Reversal background correction mode,
ength
400Hz (D2 lamp background correction
=
589n mode)
m
2. Ampere = 1mA – 12mA
3. K
wavel 3. Photometer type = single beam
ength
=
766,5
nm

Catatan : Diperiksa oleh :

Nama :
Tgl :
Halaman : 1 dari 1
Nomor : FRM/QCD/E.045/00
DAFTAR CONSUMABLE PART
Tanggal : 30/06/2018

Czern 1. Chromatic/optical dispersing component =


y-
planar diffraction grating
2. Optics Turner
Mono 2. Consist = entrance slit , collimating mirror ,
chrom
diffraction grating and exit slit
ator
3. Grating = 1800 grooves/mm
4. Blazed wavelength = 250nm
5. Focus = 300mm
6. Bandwidth = 0.1 0.2 0.3 0.4 1.0 2.0 nm
7. Scan mode = automatic
3. Burner 1. Three kinds of titanium metal burner head,
Assembly rot-proof spray chamber, high efficiency
glass nebulizer, burner height auto-
adjustable.
2. Has 100mm,50mm wide slot and 3 slot in
the top for the flame.
3. Photometer type = single beam
4. Premix chamber from an inert material
(PTFE)
5. Temperature flame = 3000°C
Photo 1. Consist of photocathode and a series of
multipl
4. Detector ier dynodes
Tube 2. Voltage = 300 VDC – 1000 VDC
(PMT)
3. Wavelength range = 190nm – 900nm

Catatan : Diperiksa oleh :

Nama :
Tgl :