Anda di halaman 1dari 42

27

CHAPTER 7
RESULT AND DISCUSSION

7.1. The study of the change of cyclic voltammogram and the relationship between scan
rate and peak current.
When the ferricyanide solution is used as the electrolyte, the anodic and cathodic currents are
given at specific potential value because the redox couples are defined as ferricyanide and
ferrocyanide. The changes in the cyclic voltammograms would reflect the diffusion coefficient
between the solution and the surface of the electrode by using the cyclic voltammetric
measurement. For this, the cyclic voltammetric analysis was performed with the various types
of electrodes for the changes in cyclic voltammograms determination including the anodic peak
current, cathodic peak current and the diffusion coefficient in each type of electrodes.
7.1.1. Carbon fiber electrode

A) B)

C) D)
28

E)
Figure19: The changes in the cyclic voltammograms of carbon fiber as a working electrode
for (A) 5, (B) 10, (C) 50, (D)75, and (E) 100 mV/s with the potential between -1 to 1 V vs.
the Ag/AgCl reference electrode, pH = 7

A) B)
29

C) D)

E) F)
30

G)

Figure20: The anodic and cathodic peak currents vs. square root of the scan rate that were
caused by the bacteria adhesion on the carbon fiber electrode for (A) 1 hr.,(B) 5 hrs., (C) 9 hrs.,
(D) 13 hrs., (E) 17 hrs., (F) 21 hrs., and (G) 25 hrs. with the potential between -1 to 1 V vs. the
Ag/AgCl reference electrode, pH = 7

Peak slope (A)


Time
Anodic Cathodic
1 0.0162 0.0103
5 0.0156 0.0098
9 0.0157 0.0106
13 0.0195 0.0107
17 0.0191 0.0097
21 0.0183 0.0095
25 0.0179 0.0091
Table6: Peak slope of carbon fiber without modified condition

Fig19 Indicates that the current increase in the range of -1 to 1 V with time, this change of
current indicates the ability of the bacteria to digest the organic matters in the electrolyte and
the electrons participated in the redox reaction. (S. Kurissery, N. Kanavillil, K. Leung, A. Chen,
L. Davey, H. Schraft, Biofouling 26 (2010) 799.)One interpretation for the current increase
shown in fig.19 is that an electrochemical interaction like the electron transfer between the
bacterial that has its ability to digest the organic matters in the synthesis wastewater and gives
out the electron to the surface of electrodes, which considered as oxidized layer, enhanced the
electrons to the surface and generated the desorption current as known as anodic current The
glucose molecule acts as the electron donor, and appears as oxidation reaction while the
ferricyanide as the electron acceptor received the electrons from the surface electrode that acts
as reduced layer which leads to the adsorption current, as known as cathodic current, the
ferricyanide as the electron acceptor so Fe3+ accept electron and undergoes to Fe2+, which leads
to the solution that turns blue (Prussian blue) in the presence of Fe2+ ions of ferrocyanide ions.
31

A) B)
Figure21: A) Initial solution for carbon fiber without modification and B) Final solution for
carbon fiber without modification

Figure 19 (a) to (e) presents the changes in the cyclic voltammograms by the bacteria that digest
organic matters (glucose) in synthesis wastewater, which acts as electron donor in the
ferricyanide solution as an electron acceptor together with the anodic and cathodic peak
currents with respect to time clearly demonstrates that the longer time, the ability of the bacteria
to digest the glucose molecule was increased, which leads to the absolute values of anodic and
cathodic currents increase as shown in Fig.20 (a) to (g).

Figure22: The anodic and cathodic peak currents vs. Time at the various scan rates
5,10,50,75,100 mV/s for carbon fiber electrodes as the working electrode( with the potential
between -1 to 1 V vs. the Ag/AgCl reference electrode, pH = 7).
Fig.22 shows that the current is proportional to the scan rate, according to the Randles-Sevcik
equation (eq.1), the current was increased as well as the scan rate that was increased. Moreover
the current was increased until the specific time that it was stable and eventually decreased.

The current was decreased since the changes of cyclic voltammograms were induced not only
by the enhanced electron transfers but also the decrease of the surface area. The previous
studies (R.A. Illsley, S.G. Roscoe, E.D. Jackson, T.J. Hughes, Biofouling 11 (1997) 191.), S.
Kurissery, N. Kanavillil, K. Leung, A. Chen, L. Davey, H. Schraft, Biofouling 26(2010) 799.
Which reported that microbial adsorption on the electrode contributed to the current decrease
in cyclic voltammogram as a result of the diminishing of the surface coverage by bacterial
adhesion.
32

Figure 23: The difference between the


anodic and cathodic peak potential (Epa-
Epc)( with the potential between -1 to 1 V
vs. the Ag/AgCl reference electrode, pH =
7)

Additionally, the difference between the anodic and cathodic peak potential, which indicated
the reversibility of the electrochemical reaction, increased with time as shown in Fig.23

For the peak potential, which is anodic peak potential (Epa) indicates the adsorption of the
electron for the carbon fiber working electrode that undergoes the oxidation reaction, which
has glucose in the solution as the electron donor. Also, the cathodic peak potential (Epc)
indicates the desorption of electron for the carbon fiber working electrode, which undergoes
the reduction reaction, which has ferricyanide ions in the solution as the electron acceptor.
Desorption current is relevant to the reductive process of oxidized layer, whereas the adsorption
current is relevant to the oxidative process of the reduced layer. These currents are familiar
patterns in typical cyclic voltammograms of working electrodes in aqueous electrolytes.(A.J.
Bard, L.R. Faulkner, Electrochemical Methods, Wiley, 2001.)

The relationship between the current and potential for the changes in cyclic voltammograms
for 1, 5, 9, 13, 17, 21, and 25 hrs. With scan rate 5,10,50,75,100 mV/s as shown in the table
below, corresponded to the adsorption (Eq.1) and desorption (Eq.2) of the electron for the
carbon fiber working electrode that leads to the adsorption and desorption currents (ipa, ipc) and
potentials (Epa, Epc).

C6H12O6 + 6H2O  6CO2 + 24H+ + 24e- (1)


24(Fe(CN)6)3- + 24e- 24(Fe(CN)6)4- (2)

Scan rate Time Current (A) Potential (V) Epa-


ipa/ipc
(mV/s) (hours) ipa ipc epa epc Epc
1 0.0309 -0.0420 0.5289 0.0300 -0.73571 0.4989
5 0.0366 -0.0385 0.5692 -0.0974 -0.95065 0.6666
9 0.0397 -0.0391 0.5875 -0.1063 -1.01535 0.6938
5 13 0.0468 -0.0434 0.6408 -0.1424 -1.07834 0.7832
17 0.0479 -0.0439 0.6591 -0.1456 -1.09112 0.8047
21 0.0476 -0.0437 0.6721 -0.1601 -1.08924 0.8322
25 0.0458 -0.0426 0.6857 -0.1641 -1.07512 0.8498
33

1 0.0460 -0.0560 0.6490 -0.2670 -0.82143 0.9160


5 0.0510 -0.0520 0.6850 -0.2090 -0.98077 0.8940
9 0.0630 -0.0580 0.7750 0.2630 -1.08621 0.5120
10 13 0.0650 -0.0590 0.7930 -0.2720 -1.10169 1.0650
17 0.0650 -0.0590 0.8110 -0.2710 -1.10169 1.0820
21 0.0640 -0.0580 0.8250 -0.2890 -1.10345 1.1140
25 0.0620 -0.0570 0.8290 -0.3030 -1.08772 1.1320
1 -0.1030 -0.6434 0.6434
5 -0.0950 -0.5629 0.5629
9 -0.1040 -0.5990 0.5990
50 13 -0.1070 -0.6030 0.6030
17 -0.1020 -0.6070 0.6070
21 -0.1000 -0.6030 0.6030
25 -0.0970 -0.6000 0.6000
1 -0.1137 -0.1463 0.1463
5 -0.1050 -0.6200 0.6200
9 -0.1150 -0.6520 0.6520
75 13 -0.1170 -0.6710 0.6710
17 -0.1125 -0.6520 0.6520
21 -0.1090 -0.6700 0.6700
25 -0.1060 -0.6740 0.6740
1 -0.1212 -0.6687 0.6687
5 -0.1130 -0.6790 0.6790
9 -0.1187 -0.6795 0.6795
100 13 -0.1250 -0.6800 0.6800
17 -0.1176 -0.6972 0.6972
21 -0.1160 -0.7010 0.7010
25 -0.1120 -0.6830 0.6830
Table7: Peak current and potential results for carbon fiber without modification
34

7.1.2. Carbon fiber with nitric treatment

A) B)

C) D)

E) F)

Figure24: the changes in the cyclic voltammograms of carbon fiber modified with Nitric acid
as a working electrode for A) 1, B) 3, C) 6, D)9, E) 12, and F) 24 hrs. with the potential between
-1 to 1 V vs. the Ag/AgCl reference electrode, pH = 7
35

A) B)

C) D)

E) F)

G)

H)
36

I)
Figure25: The anodic and cathodic peak
currents vs. square root of the scan rate that was caused by the bacteria adhesion on the carbon
fiber modified with Nitric acid electrode for (a) 1 hr. (b) 3 hrs. (c) 6 hrs. (d) 9 hrs. (e) 12 hrs.
(f) 15 hrs. (g) 18 hrs.(h) 21 hrs. (i) 24 hrs. with the potential between -1 to 1 V vs. the Ag/AgCl
reference electrode, pH = 7

Peak slope (A)


Time
Anodic Cathodic
1 0.0111 -0.0095
3 0.0104 -0.0091
6 0.0089 -0.0039
9 0.0065 -0.0042
12 -0.008
15 -0.0078
18 -0.0081
21 -0.0086
24 -0.0085
Table8: Peak slope of carbon fiber with nitric acid treatment condition

Figure 24(a) to (f) presents the changes in the cyclic voltammograms by the bacteria that digest
organic matters (glucose) in synthesis wastewater, which acts as electron donor in the
ferricyanide solution as an electron acceptor together with the anodic and cathodic peak
currents with respect to time clearly demonstrates that the longer time, the ability of the bacteria
to digest the glucose molecule was increased, which leads to the absolute values of anodic and
cathodic currents increase. For the anodic peak current, it was observed to disappear after 12
hrs. operation whereas the cathodic peak current still increased concerning time as shown in
Fig.25 (a) to (i). Moreover, Fig.20 shows that from 6 hrs. , the higher scan rate cannot detect
the anodic peak current, while at low scan rate still able to detect. The assumption is that at low
scan rate the current response to the electrode is almost at the steady-state condition since the
electrons have sufficient time to penetrate the pores of the material, whereas, for higher scan
rates, only the electrons accumulate on the outer surface. Therefore, it is considered as transient
steady state condition, which leads to the low detection. Also, it shows that the
electrochemical reaction has a low reaction rate. Then, when applying low scan rates the
reaction has a higher time to occur and, consequently, the peak associated with it grows.
37

One assumption for the disappearing of the anodic current is the glucose in the solution was
used up. Therefore the electrons were no more enhanced to the surface of carbon fiber modified
with the nitric acid electrode. Another assumption is the treatment for nitric acid to the carbon
fiber electrode was practically incorrect. Since the nitric acid causes the formation of oxygen
surface complex at the edge sites of carbon structures, which also can hinder the electron
enhancement during oxidation process in case of the wrong treatment was applied.

Figure26: The anodic and cathodic peak currents vs. Time at the various scan rates 3,5,7,10
mV/s for carbon fiber modified with Nitric acid electrodes as the working electrode( with the
potential between -1 to 1 V vs. the Ag/AgCl reference electrode, pH = 7)

Fig.26 shows that the current is proportional to the scan rate, according to the Randles-Sevcik
equation (eq.1). The current was increased as well as the scan rate that was increased. For the
anodic peak current, the current was increased until 10 hrs. Then it was decreased, whereas the
cathodic peak current, was observed that the current was decreased until 10 hrs. That it was
started to increase according to the redox couple reaction.

Figure27: The difference between the


anodic and cathodic peak potential (Epa-
Epc)( with the potential between -1 to 1 V vs.
the Ag/AgCl reference electrode, pH = 7)

Additionally, the difference between the anodic and cathodic peak potential, which indicated
the reversibility of the electrochemical reaction, increased with time as shown in Fig.27
The relationship between the current and potential for the changes in cyclic voltammograms
for 1, 3, 6, 9, 12, 15, 18, 21, and 24 hrs. With scan rate 3,5,7,10 mV/s as shown in the table
below, corresponded to the adsorption (Eq.1) and desorption (Eq.2) of the electron for the
carbon fiber working electrode that leads to the adsorption and desorption currents (ipa, ipc) and
potentials (Epa, Epc).
38

Scan rate Time Current (A) Potential (V) Epa-


ipa/ipc
(mV/s) (hours) ipa ipc epa epc Epc
1 0.0134 -0.0276 0.6000 -0.4109 -0.48551 1.0109
3 0.0164 -0.0324 0.6904 -0.4600 -0.50617 1.1504
6 0.0167 -0.0240 0.6053 -0.3571 -0.69583 0.9624
9 0.0179 -0.0189 0.7975 -0.3837 -0.94709 1.1812
3 12 -0.0172 -0.4736 0.4736
15 -0.0188 -0.4198 0.4198
18 0.0215 -0.0190 0.9052 -0.3973 -1.13158 1.3025
21 0.0221 -0.0194 0.8780 -0.3795 -1.13918 1.2575
24 0.0221 -0.0195 0.8467 -0.3659 -1.13333 1.2126
1 0.0195 -0.0324 0.6904 -0.4600 -0.60185 1.1504
3 0.0214 -0.0312 0.6768 -0.4239 -0.68590 1.1007
6 0.0214 -0.0288 0.7217 -0.4286 -0.74306 1.1503
9 0.0206 -0.0235 0.8200 -0.5049 -0.87660 1.3249
5 12 -0.0224 -0.5943 0.5943
15 -0.0235 -0.5494 0.5494
18 -0.0239 -0.5316 0.5316
21 -0.0244 -0.4961 0.4961
24 -0.0247 -0.4961 0.4961
1 0.0235 -0.0369 0.7620 -0.5540 -0.63686 1.3160
3 0.0258 -0.0348 0.7535 -0.4914 -0.74138 1.2449
6 0.0248 -0.0267 0.7975 -0.5404 -0.92884 1.3379
9 0.0239 -0.0262 0.8915 -0.5765 -0.91221 1.4680
7 12 -0.0256 -0.6345 0.6345
15 -0.0267 -0.6120 0.6120
18 -0.0275 -0.6032 0.6032
21 -0.0274 -0.5896 0.5896
24 -0.0283 -0.5676 0.5676
1 0.0295 -0.0410 0.8531 -0.5943 -0.71951 1.4474
3 0.0312 -0.0395 0.8513 -0.5540 -0.78987 1.4053
6 -0.0306 -0.5896 0.5896
9 -0.0246 -0.6970 0.6970
10 12 -0.0286 -0.7019 0.7019
15 -0.0299 -0.6836 0.6836
18 -0.0305 -0.6925 0.6925
21 -0.0312 -0.6709 0.6709
24 -0.0317 -0.6617 0.6617
Table9: Peak current and potential results for carbon fiber with nitric acid treatment
39

7.1.3. Carbon fiber with Teflon treatment

A) B)

C) D)

E)
F)

Figure28: The changes in the cyclic voltammograms of carbon fiber as a working electrode for
A) 3, B) 5, C) 7, D)10, E) 30 and F) 50 mV/s with the potential between -1 to 1 V vs. the
Ag/AgCl reference electrode, pH = 7
40

A) B)

C)
Figure29: The anodic and cathodic peak currents vs. square root of the scan rate that was caused
by the bacteria adhesion on the carbon fiber electrode for A) 1 hr. B) 12hrs. moreover, C) 24
hrs. with the potential between -1 to 1 V vs. the Ag/AgCl reference electrode, pH = 7

Peak slope (A)


Time
Anodic Cathodic
1 0.00848 -0.0072
12 0.00868 -0.0072
24 0.00824 -0.0078
Table10: Peak slope of carbon fiber with Teflon treatment condition

Figure30: Carbon fiber with Teflon treatment solution for an initial and final period
41

Figure 28(a) to (f) presents the changes in the cyclic voltammograms by the bacteria that digest
organic matters (glucose) in synthesis wastewater, which acts as electron donor in the
ferricyanide solution as an electron acceptor together with the anodic and cathodic peak
currents with respect to time clearly demonstrates that the longer time, the ability of the bacteria
to digest the glucose molecule was increased, which leads to the absolute values of anodic and
cathodic currents increase as shown in Fig.29(a) to (c).

Figure31: The anodic and cathodic peak currents vs. Time at the various scan rates 3, 5, 7, 10,
30, and 50 mV/s for carbon fiber electrodes as the working electrode( with the potential
between -1 to 1 V vs. the Ag/AgCl reference electrode, pH = 7)

Fig.31 shows that the current is proportional to the scan rate, according to the Randles-Sevcik
equation (eq.1), the current was increased as well as the scan rate that was increased. Moreover
the current was increased until the specific time that it was stable and eventually decreased.

Figure32: The difference between


the anodic and cathodic peak
potential (Epa-Epc) ( with the
potential between -1 to 1 V vs. the
Ag/AgCl reference electrode, pH =
7).

Additionally, the difference between the anodic and cathodic peak potential, which indicated
the reversibility of the electrochemical reaction, increased with time as shown in Fig.32.
42

The relationship between the current and potential for the changes in cyclic voltammograms
for 1,12,24 hrs. With scan rate 3, 5, 7, 10, 30, and 50 mV/s as shown in the table below,
corresponded to the adsorption (Eq.1) and desorption (Eq.2) of electron for the carbon fiber
working electrode that leads to the adsorption and desorption currents (ipa, ipc) and potentials
(Epa, Epc).

Scan rate Time Current (A) Potential (V) Epa-


ipa/ipc
(mV/s) (hours) ipa ipc epa epc Epc
1 0.0046 -0.0105 0.3723 -0.0306 -0.4381 0.4029
3 12 0.0129 -0.0124 0.4514 -0.0123 -1.0403 0.4637
24 0.0143 -0.0133 0.4574 0.0055 -1.0752 0.4519
1 0.0073 -0.0133 0.4261 -0.0306 -0.5489 0.4567
5 12 0.0164 -0.0159 0.5065 -0.0394 -1.0314 0.5459
24 0.0185 -0.0173 0.5201 -0.0525 -1.0694 0.5726
1 0.0207 -0.0191 0.5603 -0.0619 -1.0838 0.6222
7 12 0.0194 -0.0186 0.5468 -0.0708 -1.0430 0.6176
24 0.0219 -0.0206 0.5739 -0.0974 -1.0631 0.6713
1 0.0241 -0.0228 0.5875 -0.1199 -1.0570 0.7074
10 12 0.0258 -0.0238 0.6000 -0.1199 -1.0840 0.7199
24 0.0261 -0.0244 0.6319 -0.1377 -1.0697 0.7696
1 0.0414 -0.0384 0.7307 -0.3258 -1.0781 1.0565
30 12 0.0447 -0.0405 0.7798 -0.3435 -1.1037 1.1233
24 0.0000
1 0.0507 -0.0491 0.7750 -0.3973 -1.0326 1.1723
50 12 -0.0504 -0.4600 0.4600
24 0.0000
Table11: Peak current and potential results for carbon fiber with Teflon treatment

7.1.4. Carbon fiber with nitric acid and Teflon treatment

A)
B)
43

C) D)

E) F)

G) H)
Figure33: the changes in the cyclic voltammograms of carbon fiber modified with PTFE as a
working electrode for A) 1, B) 3, C) 6, D)9, E) 12, F) 18, G)21, and H)24 hrs. with the potential
between -1 to 1 V vs. the Ag/AgCl reference electrode, pH = 7.

A) B)
44

C) D)

E) F)

H)
I)
Figure34: The anodic and cathodic peak currents vs. square root of the scan rate that were
caused by the bacteria adhesion on the carbon fiber modified with Nitric acid and PTFE
electrode for (a) 1 hr. (b) 3 hrs. (c) 6 hrs. (d) 9 hrs. (e) 12 hrs. (f) 15 hrs. (g) 18 hrs. (h) 21 hrs.
(i) 24 hrs. with the potential between -1 to 1 V vs. the Ag/AgCl reference electrode, pH = 7.

Peak slope (A)


Time
Anodic Cathodic
1 0.01114 -0.0095
3 0.01038 -0.0091
6 0.00888 -0.0039
9 0.00652 -0.0042
12 -0.008
45

15 -0.0078
18 -0.0081
21 -0.0086
24 -0.0085
Table12: Peak slope of carbon fiber with nitric and Teflon treatment condition

A) B)
Figure35: A) Final solution for carbon fiber with nitric and Teflon treatment, B) Electrode
after the experiment for carbon fiber with nitric and Teflon treatment

Figure 33(a) to (h) presents the changes in the cyclic voltammograms by the bacteria that digest
organic matters (glucose) in synthesis wastewater, which acts as electron donor in the
ferricyanide solution as an electron acceptor together with the anodic and cathodic peak
currents with respect to time clearly demonstrates that the longer time, the ability of the bacteria
to digest the glucose molecule was increased, which leads to the absolute values of anodic and
cathodic currents increase as shown in Fig.34 (a) to (i).

Figure36: The anodic and cathodic peak currents vs. Time at the various scan rates 3,5,7,10
mV/s for carbon fiber modified with Nitric acid and PTFE electrodes as the working electrode(
with the potential between -1 to 1 V vs. the Ag/AgCl reference electrode, pH = 7).

Fig.36 shows that the current is proportional to the scan rate, according to the Randles-Sevcik
equation (eq.1), the current was increased as well as the scan rate that was increased. For the
anodic peak current the current was increased. Eventually it was decreased, whereas the
46

cathodic peak current, was observed that the current was decreased, eventually it was started
to stable according to the redox couple reactions.

Figure37: The difference between the anodic and


cathodic peak potential(Epa-Epc)( with the potential
between -1 to 1 V vs. the Ag/AgCl reference
electrode, pH = 7).

Additionally, the difference between the anodic and cathodic peak potential, which indicated
the reversibility of the electrochemical reaction, increased with time as shown in Fig.37
The relationship between the current and potential for the changes in cyclic voltammograms
for 1, 3, 6, 9, 12, 15, 18, 21, and 24 hrs. With scan rate 3,5,7,10 mV/s as shown in the table
below, corresponded to the adsorption (Eq.1) and desorption (Eq.2) of the electron for the
carbon fiber working electrode that leads to the adsorption and desorption currents (ipa, ipc) and
potentials (Epa, Epc).

Scan rate Time Current (A) Potential (V) Epa-


ipa/ipc Eo
(mV/s) (hours) ipa ipc Epa epc Epc
1 0.0139 -0.0259 0.4083 -0.0394 -0.53668 0.4477 0.1845
3 0.0153 -0.0244 0.4125 -0.0572 -0.62705 0.4697 0.1777
6 0.0165 -0.0232 0.4308 -0.0394 -0.71121 0.4702 0.1957
9 0.0235 -0.0200 0.5337 -0.0394 -1.17500 0.5731 0.2472
3 12 0.0228 -0.0187 0.5337 -0.0436 -1.21925 0.5773 0.2451
15 0.0221 -0.0185 0.5337 -0.0483 -1.19459 0.5820 0.2427
18 0.0218 -0.0185 0.5201 -0.0436 -1.17838 0.5637 0.2383
21 0.0214 -0.0184 0.5201 -0.0347 -1.16304 0.5548 0.2427
24 0.0210 -0.0184 0.5024 -0.0436 -1.14130 0.5460 0.2294
1 0.0195 -0.0309 0.4438 -0.1063 -0.63107 0.5501 0.1688
3 0.0209 -0.0291 0.4621 -0.1021 -0.71821 0.5642 0.1800
6 0.0313 -0.0243 0.6053 -0.0974 -1.28807 0.7027 0.2540
9 0.0288 -0.0244 0.5875 -0.1110 -1.18033 0.6985 0.2383
5 12 0.0279 -0.0238 0.5971 -0.0933 -1.17227 0.6904 0.2519
15 0.0274 -0.0233 0.5917 -0.1021 -1.17597 0.6938 0.2448
18 0.0271 -0.0235 0.5781 -0.1110 -1.15319 0.6891 0.2336
21 0.0266 -0.0232 0.0578 -0.0974 -1.14655 0.1552 -0.0198
24 0.0264 -0.0231 0.5781 -0.1063 -1.14286 0.6844 0.2359
1 0.0241 -0.0345 0.4976 -0.1335 -0.69855 0.6311 0.1821
7
3 0.0255 -0.0326 0.4935 -0.1246 -0.78221 0.6181 0.1845
47

6 0.0346 -0.0287 0.6544 -0.1424 -1.20557 0.7968 0.2560


9 0.0335 -0.0287 0.6544 -0.1554 -1.16725 0.8098 0.2495
12 0.0323 -0.0279 0.6408 -0.1513 -1.15771 0.7921 0.2448
15 0.0317 -0.0276 0.6455 -0.1513 -1.14855 0.7968 0.2471
18 0.0310 -0.0276 0.6366 -0.1424 -1.12319 0.7790 0.2471
21 0.0307 -0.0272 0.6230 -0.1424 -1.12868 0.7654 0.2403
24 0.0305 -0.0271 0.6277 -0.1466 -1.12546 0.7743 0.2406
1 0.0297 -0.0390 0.5337 -0.1915 -0.76154 0.7252 0.1711
3 0.0308 -0.0373 0.5565 -0.1690 -0.82574 0.7255 0.1938
6 0.0403 -0.0340 0.7307 -0.2051 -1.18529 0.9358 0.2628
9 0.0393 -0.0333 0.7218 -0.2092 -1.18018 0.9310 0.2563
10 12 0.0379 -0.0329 0.7082 -0.2092 -1.15198 0.9174 0.2495
15 0.0369 -0.0325 0.7035 -0.1915 -1.13538 0.8950 0.2560
18 0.0365 -0.0323 0.6996 -0.2139 -1.13003 0.9135 0.2429
21 0.0360 -0.0319 0.6810 -0.1962 -1.12853 0.8772 0.2424
24 0.0356 -0.0316 0.6846 -0.2000 -1.12658 0.8846 0.2423
Table13: Peak current and potential results for carbon fiber with nitric acid and Teflon
treatment

7.1.5. Carbon fiber electrode with Grinded sludge

A) B)

C) D)
48

E) F)

G) H)
Figure38: the changes in the cyclic voltammograms of carbon fiber with grinded sludge as a
working electrode for (a) 1 (b) 3 (c) 6 (d)9 (e) 12 (f) 18 (g)21 (h)24 hrs. with the potential
between -1 to 1 V vs. the Ag/AgCl reference electrode, pH = 7.

A) B)

C) D)
49

E)
F)

G) H)

Figure39: The anodic and cathodic peak currents vs. square root of the scan rate that was
caused by the bacteria adhesion on the carbon fiber with grinded sludge electrode for (a) 1 hr.
(b) 3 hrs. (c) 6 hrs. (d) 9 hrs. (e) 12 hrs. (f) 18 hrs. (g) 21 hrs.(h) 24 hrs. with the potential
between -1 to 1 V vs. the Ag/AgCl reference electrode, pH = 7

Peak slope (A)


Time
Anodic Cathodic
1 0.01114 -0.0095
3 0.01038 -0.0091
6 0.00888 -0.0039
9 0.00652 -0.0042
12 -0.008
50

15 -0.0078
18 -0.0081
21 -0.0086
24 -0.0085
Table14: Peak slope of carbon fiber with nitric and Teflon treatment condition

A) B)
Figure40: A) Final solution of carbon fiber electrode with grinded sludge, B) Electrode of
carbon fiber with grinded sludge experiment

Figure 38(a) to (h) presents the changes in the cyclic voltammograms by the bacteria that digest
organic matters (glucose) in synthesis wastewater, which acts as electron donor in the
ferricyanide solution as an electron acceptor together with the anodic and cathodic peak
currents with respect to time clearly demonstrates that the longer time, the ability of the bacteria
to digest the glucose molecule was increased, which leads to the absolute values of anodic and
cathodic currents increase. For the anodic peak current, it was observed to disappears after 18
hrs. Operation whereas the cathodic peak current still increased concerning time as shown in
Fig.38 (a) to (h).Fig.39shows that from 18 hrs. , the higher scan rate cannot detect the anodic
peak current, while at low scan rate still able to detect, so it shows that the electrochemical
reaction has a low reaction rate. Then, when applying low scan rates the reaction has a higher
time to occur and, consequently, the peak associated with it grows.

Figure41: The anodic and cathodic peak currents vs. Time at the various scan rates 3,5,7
mV/s for carbon fiber with grinded sludge electrodes as the working electrode( with the
potential between -1 to 1 V vs. the Ag/AgCl reference electrode, pH = 7).
51

Fig.41 shows that the current is proportional to


the scan rate, according to the Randles-Sevcik
equation (eq.1), the current was increased as well
as the scan rate that was increased. For the anodic
peak current, the current was increased until 10
hrs. That it was decreased, whereas the cathodic
peak current, was observed that the current was
decreased until 10 hrs that it was started to
increase according to the redox couple reactions.

Figure42: The difference between the anodic and cathodic peak potential (Epa-Epc) ( with the
potential between -1 to 1 V vs. the Ag/AgCl reference electrode, pH = 7).

The relationship between the current and potential for the changes in cyclic voltammograms
for 1, 3, 6, 9, 12, 15, 18, 21, and 24 hrs. with scan rate 3, 5, 7 mV/s as shown in the table below,
corresponded to the adsorption (Eq.1) and desorption (Eq.2) of the electron for the carbon fiber
working electrode that leads to the adsorption and desorption currents (ipa, ipc) and potentials
(Epa, Epc).

Scan Current (A) Potential (V)


Time Epa-
rate ipa/ipc Eo
(hours) ipa ipc Epa epc Epc
(mV/s)
1 0.0158 -0.0275 0.6319 -0.4914 -0.57455 1.1233 0.0703
3 0.0174 -0.0246 0.6904 -0.4376 -0.70760 1.1280 0.1264
6 0.0184 -0.0237 0.6993 -0.3796 -0.77637 1.0789 0.1599
9 0.0190 -0.0220 0.7082 -0.3393 -0.86364 1.0475 0.1845
3 12 0.0186 -0.0215 0.7124 -0.3211 -0.86512 1.0335 0.1957
15 0.0189 -0.0211 0.7307 -0.3211 -0.89573 1.0518 0.2048
18 -0.0149 -0.5583
21 -0.0144 -0.5990
24 0.0210 -0.0184 0.5024 -0.0436 -1.14130 0.5460 0.2294
1 0.0211 -0.0316 0.7751 -0.6079 -0.66772 1.3830 0.0836
3 0.0222 -0.0293 0.8111 -0.5629 -0.75768 1.3740 0.1241
6 0.0235 -0.0281 0.7975 -0.5000 -0.83630 1.2975 0.1488
9 0.0237 -0.0268 0.8153 -0.4511 -0.88433 1.2664 0.1821
5 12 0.0235 -0.0261 0.8247 -0.4464 -0.90038 1.2711 0.1892
15 0.0237 -0.0254 0.8247 -0.4334 -0.93307 1.2581 0.1957
18 -0.0176 -0.6836 0.00000 0.6836 -0.3418
21 -0.0171 -0.0171 0.00000 0.0171 -0.0086
24 -0.0169 -0.7422 0.00000 0.7422 -0.3711
7 1 0.0247 -0.0350 0.8561 -0.6925 -0.70571 1.5486 0.0818
52

3 0.0264 -0.0328 0.8963 -0.6392 -0.80488 1.5355 0.1286


6 0.0275 -0.0315 0.8916 -0.5541 -0.87302 1.4457 0.1688
9 0.0276 -0.0303 0.9000 -0.5227 -0.91089 1.4227 0.1887
12 0.0273 -0.0295 0.9000 -0.5180 -0.92542 1.4180 0.1910
15 0.0273 -0.0289 0.9052 -0.5000 -0.94464 1.4052 0.2026
18 -0.0196 -0.7866 0.00000 0.7866 -0.3933
21 -0.0191 -0.8137 0.00000 0.8137 -0.4069
24 -0.0186 -0.8226 0.00000 0.8226 -0.4113
Table15: Peak current and potential results for carbon fiber with grinded sludge

7.2. Calculation of diffusion coefficient using cyclic voltammetry by Randles-Sevcik


equation

The diffusion coefficient can be calculated from the Randles-Sevcik equation. The Randle-
Sevcik equation expresses that the current of the cyclic voltammograms is proportional to the
square root of the scan rate.

ip = (2.69x105)n3/2AactD1/2C0v1/2 (1)

Where, ip : peak current (A) ; n:number of electron (l) ; Aact: active surface area (cm2); D:
diffusion coefficient (cm2/s) ; C0: bulk concentration (mol/cm3) ; v: scan rate (mV/s)

From eq. (1) the diffusion coefficient can be calculated using the slope of curves representing
the relationship between the peak current (ip) and the square root of various scan rate (v1/2)
To obtain the diffusion coefficient, the active surface area was fixed at 6 cm2. Since the
diffusion coefficient is mainly dependent on temperature which operates at room’s temperature
at 25 C. Thus, the influence of Aact affecting on diffusion coefficient is thought to be minimal
at the condition of fixed temperature. Therefore, the diffusion coefficient of the electrode was
calculated.

7.2.1 Carbon fiber electrode

The diffusion coefficient of this experiment as shown in the table below;

Peak
Bulk Diffusion
Time current Number of Active surface
concentration coefficient
(hours) slope, ip electrons, n area (cm2)
(mol/cm3) (cm2/s)
(A)
1 0.0162 2.732E-16
5 0.0156 2.534E-16
9 0.0157 2.566E-16
13 0.0195 24 6 0.005164 3.959E-16
17 0.0191 3.798E-16
21 0.0183 3.487E-16
25 0.0179 3.336E-16
53

Table16: Diffusion coefficient of oxidation reaction for carbon fiber working electrode

Peak
Bulk Diffusion
Time current Number of Active surface
concentration coefficient
(hours) slope, ip electrons, n area (cm2)
(mol/cm3) (cm2/s)
(A)
1 0.0103 1.105E-16
5 0.0098 9.999E-17
9 0.0106 1.170E-16
13 0.0107 24 6 0.005164 1.192E-16
17 0.0097 9.796E-17
21 0.0095 9.396E-17
25 0.0091 8.622E-17
Table17: Diffusion coefficient of reduction reaction for carbon fiber working electrode

A) B)
Figure43: Shows the change in the diffusion coefficient calculated from the results of cyclic
voltammograms in A) cathodic and B) anodic diffusion coefficient that indicates how fast the
electrons transfer to the surface of the carbon fiber electrodes.

During the experiment, Fig.43 (b) shows the diffusion coefficient rate for reduction process,
which has ferricyanide as the electron acceptor, increased until 9 hrs. At the rate 1.1958x10-
16
cm2/s, which indicates the maximum value that was occurred, eventually, the rate was
decreased. For the diffusion coefficient rate for oxidation process, which has the glucose as the
electron donor, Fig.43(a) shows the rate was increased until 13 hrs, which is 3.98x10 - 16cm2/s,
which indicates the maximum value that was occurred, eventually the rate was also decreased
similar to the reduction reaction.

7.2.2. Carbon fiber electrode modified with Nitric acid


The diffusion coefficient of this experiment as shown in the table below;
54

Active Bulk Diffusion


Time Peak current Number of
surfa9ece area concentration coefficient
(hours) slope, ip (A) electrons, n
(cm2) (mol/cm3) (cm2/s)

1 0.0111 1.29E-16
3 0.0104 1.12E-16
6 0.0089 8.21E-17
9 0.0065 4.43E-17
12 24 6 0.005164
15
18
21
24
Table18: Diffusion coefficient of oxidation reaction for carbon fiber with the nitric acid
treatment working electrode

Bulk Diffusion
Time Peak current Number of Active surface
concentration coefficient
(hours) slope, ip (A) electrons, n area (cm2)
(mol/cm3) (cm2/s)

1 -0.00949 9.38E-17
3 -0.00913 8.68E-17
6 -0.00387 1.56E-17
9 -0.00417 1.81E-17
12 -0.00795 24 6 0.005164 6.58E-17
15 -0.00776 6.27E-17
18 -0.00809 6.81E-17
21 -0.00863 7.75E-17
24 -0.00854 7.59E-17
Table19: Diffusion coefficient of reduction reaction for carbon fiber with the nitric acid
treatment working electrode

Figure43 shows the change in the diffusion coefficient calculated from the results of cyclic
voltammograms in (a) cathodic and (b) anodic diffusion coefficient, that indicates how fast the
electrons transfer to the surface of the carbon fiber modified with Nitric acid and Teflon
electrodes

During the experiment, Fig.31vshows the diffusion coefficient rate for reduction process,
which has ferricyanide as the electron acceptor, increased until 3 hrs. At the rate 8.68x10-17
cm2/s, which indicates the maximum value that was occurred, eventually, the rate was
decreased. For the diffusion coefficient rate for oxidation process, which has the glucose as the
electron donor, Fig.31 shows the rate was increased until 1 hrs, which is 1.29x10-16cm2/s, which
55

indicates the maximum value that was occurred. Eventually the rate was also decreased
similarly to the reduction reaction.

7.2.3. Carbon fiber electrode modified with Teflon


The diffusion coefficient of this experiment as shown in the table below;

Scan Bulk Diffusion


Time Peak current Number of Active surface
rate concentration coefficient
(hours) slope, ip (A) electrons, n area (cm2)
(mV/s) (mol/cm3) (cm2/s)

1 0.00848 7.487E-17
3 12 0.00868 24 6 0.005164 7.844E-17
24 0.00824 7.069E-17
Table20: Diffusion coefficient of reduction-oxidation for carbon fiber with Teflon treatment
working electrode

Scan Number of Bulk Diffusion


Time Peak current Active surface
rate electrons, concentration coefficient
(hours) slope, ip (A) area (cm2)
(mV/s) n (mol/cm3) (cm2/s)

1 -0.00721 5.412E-17
3 12 -0.00722 24 6 0.005164 5.427E-17
24 -0.00775 6.253E-17
Table21: Diffusion coefficient of reduction for carbon fiber with Teflon treatment working
electrode

Figure44: shows the change in the diffusion coefficient calculated from the results of cyclic
voltammograms in (a) cathodic and (b) anodic diffusion coefficient that indicates how fast
the electrons transfer to the surface of the carbon fiber modified with Teflon electrodes.

During the experiment, Fig.15shows the diffusion coefficient rate for reduction process, which
has ferricyanide as the electron acceptor, increased until 12 hrs. at the rate 5.427x10-17 cm2/s,
which indicates the maximum value that was occurred. Eventually the rate was decreased. For
the diffusion coefficient rate for oxidation process, which has the glucose as the electron donor,
56

Fig.15 shows the rate was increased until 12 hrs, which is 7.844x10 -17cm2/s, which indicates
the maximum value that was occurred, eventually the rate was also decreased similar to the
reduction reaction.

7.2.4. Carbon fiber electrode modified with nitric acid and Teflon
The diffusion coefficient of this experiment as shown in the table below;

Bulk Diffusion
Time Peak current Number of Active surface
concentration coefficient
(hours) slope, ip (A) electrons, n area (cm2)
(mol/cm3) (cm2/s)

1 0.01114 1.29E-16
3 0.01038 1.12E-16
6 0.00888 8.21E-17
9 0.00652 4.43E-17
12 24 6 0.005164
15
18
21
24
Table22: Diffusion coefficient of reduction-oxidation for carbon fiber with Teflon treatment
working electrode

Active Bulk Diffusion


Time Peak current Number of
surface area concentration coefficient
(hours) slope, ip (A) electrons, n
(cm2) (mol/cm3) (cm2/s)

1 -0.00949 9.38E-17
3 -0.00913 8.68E-17
6 -0.00387 1.56E-17
9 -0.00417 1.81E-17
12 -0.00795 24 6 0.005164 6.58E-17
15 -0.00776 6.27E-17
18 -0.00809 6.81E-17
21 -0.00863 7.75E-17
24 -0.00845 7.43E-17
Table23: Diffusion coefficient of reduction for carbon fiber with Teflon treatment working
electrode

\
57

Figure45: shows the change in the diffusion


coefficient calculated from the results of cyclic
voltammograms in (a) cathodic and (b) anodic
diffusion coefficient that indicates how fast the
electrons transfer to the surface of the carbon fiber
modified with Nitric acid and Teflon electrodes.

During the experiment, Fig.30vshows the diffusion coefficient rate for reduction process,
which has ferricyanide as the electron acceptor, increased until 18 hrs. At the rate 9.47x10-17
cm2/s, which indicates the maximum value that was occurred, eventually, the rate was
decreased. For the diffusion coefficient rate for oxidation process, which has the glucose as the
electron donor, Fig.30 shows the rate was increased until 6 hrs, which is 2.65x10-16cm2/s, which
indicates the maximum value that was occurred, eventually the rate was also decreased similar
to the reduction reaction.

7. 2.5. Carbon fiber electrode with grinded sludge


The diffusion coefficient of this experiment as shown in the table below;

Bulk Diffusion
Time Peak current Number of Active surface
concentration coefficient
(hours) slope, ip (A) electrons, n area (cm2)
(mol/cm3) (cm2/s)

1 0.01114 1.29E-16
3 0.01038 1.12E-16
6 0.00888 8.21E-17
9 0.00652 4.43E-17
12 24 6 0.005164
15
18
21
24
Table24: Diffusion coefficient of reduction-oxidation for carbon fiber with the grinded sludge
working electrode

Peak Bulk Diffusion


Time Number of Active surface
current concentration coefficient
(hours) electrons, n area (cm2)
slope, ip (A) (mol/cm3) (cm2/s)

1 -0.00949 9.38E-17
3 -0.00913 8.68E-17
24 6 0.005164
6 -0.00387 1.56E-17
9 -0.00417 1.81E-17
58

12 -0.00795 6.58E-17
15 -0.00776 6.27E-17
18 -0.00809 6.81E-17
21 -0.00863 7.75E-17
24 -0.00845 7.43E-17
Table25: Diffusion coefficient of reduction for carbon fiber with the grinded sludge working
electrode

Figure46: shows the change in the


diffusion coefficient calculated from the
results of cyclic voltammograms in (a)
cathodic and (b) anodic diffusion
coefficient that indicates how fast the
electrons transfer to the surface of the
carbon fiber modified with Nitric acid
and Teflon electrodes.

During the experiment, Fig.32 shows the diffusion coefficient rate for reduction process, which
has ferricyanide as the electron acceptor, increased until 3 hrs. At the rate 8.68x10-17 cm2/s,
which indicates the maximum value that was occurred, eventually, the rate was decreased. For
the diffusion coefficient rate for oxidation process, which has the glucose as the electron donor,
Fig.32 shows the rate was increased until 1 hrs, which is 1.29x10-16cm2/s, which indicates the
maximum value that was occurred. Eventually, the rate was also decreased similarly to the
reduction reaction.

7.3. The study of the early and later stage of the bacteria and its roles in current
generation by the change in diffusion coefficient and current.

One interpretation for the diffusion coefficient and the current that was decreased after a while
of the experiment that obtained from the result is related to the diminishment of the active
surface area of the electrodes as more bacterial cells cover wider surface, which interferes with
the electrochemical reaction of the electrode. Since the limited amount of the organic matters
(COD ca. 2000 mgO2/L) was added to this experiment, the ability of bacteria to digest the
organic matters was decreased as time passed by. Although, the surface area decreases with
some bacteria growth on the electrode.
Then, there may be two putative actions in opposite directions that change the current in the
cyclic voltammogram with the bacteria roles and attachment; the enhanced electrochemical
reaction like electron transfer between bacteria and electrode surface due to its ability to digest
the organic matters (glucose) in the synthesized wastewater and gives out the electron then it
was enhanced to the electrode surface that increases the current.
Another action is the surface covering action to diminish the area available for the
electrochemical interaction that lowers the current. Moreover, the organic matters (glucose)
that present in the synthesis wastewater also decreased from being digested by the bacteria and
the electrons were less being generated. Therefore, the enhancement of electron to the electrode
surface was decreased, which also leads lowers the current.
59

A)

B)

Figure47: The anodic and cathodic peak currents vs. Time for the carbon fiber-based
electrode with various treatment includes nitric acid, PTFE, nitric acid and PTFE, grided
sludge, as the working electrode( with the potential between -1 to 1 V vs. the Ag/AgCl
reference electrode, scan rate 5 mV/s and pH = 7).

From Fig.47(a), For the anodic peak current it shows that the current was increased concerning
time, eventually stable and decreased as the pattern of an electrochemical redox reaction as
mentioned in eq.(1,2) which has glucose as electron donor and ferricyanide as an electron
acceptor. For carbon fiber electrode, which represents as the material that gives the highest
anodic peak current (0.0478A, 17 hrs) > carbon fiber with Nitric acid and PTFE electrode
(0.0314A, 6 hrs.) > carbon fiber electrode with grinded sludge (0.0237, 9 hrs.) > carbon fiber
with Nitric electrode (0.0215A, 3 hrs.)> carbon fiber with PTFE electrode (0.0185 A,24 hrs.)
as shown in the table below.

From Fig.47(b), For the cathodic peak current, it shows that the current was decreased in the
beginning concerning time, eventually stable as the pattern of an electrochemical redox
reaction as mentioned in eq.(1,2) which has glucose as electron donor and ferricyanide as an
electron acceptor. For carbon fiber electrode, which represents as the material that gives the
maximum cathodic peak current (0.043A, 13 hrs) > carbon fiber with Nitric acid electrode
(0.0325A, 1 hrs.) > carbon fiber electrode with grinded sludge (0.0316, 1 hrs.) > carbon fiber
with Nitric and PTFE electrode (0.0309A, 1 hrs.)> carbon fiber with PTFE electrode (0.0174
A,1 hrs.) as shown in the table below.
60

Anodic Cathodic
Material Peak Peak
Time Time
current current
(hour) (hour)
(A) (A)

Carbon fiber 17 0.0478 13 0.0430

Nitric acid and


6 0.0314 1 0.0309
PTFE
Carbon fiber with
9 0.0233 1 0.0316
grinded sludge

Nitric acid 3 0.0215 1 0.0325

PTFE 24 0.0185 24 0.0174

Table26: Cathodic and anodic peak current

From Figs.47(a) and (b), it shows that the carbon fiber electrode gives the highest for both
anodic and cathodic peak current compares to other types of electrode. In the other hands, the
carbon-based electrode with PTFE appears to give the lowest for both anodic and cathodic peak
current compares to other types of electrode. For the carbon-based electrode with Nitric acid,
grinded sludge, nitric acid PTFE do not give a vastly difference from each other.

A) B)
Figure48: The anodic and cathodic diffusion coefficient vs. Time for the carbon fiber-based
electrode with various treatment includes nitric acid, PTFE, nitric acid and PTFE, grided
sludge, as the working electrode( with the potential between -1 to 1 V vs. the Ag/AgCl
reference electrode, scan rate 5 mV/s and pH = 7).

From Fig.48(a), For the anodic diffusion coefficient, it shows that the diffusion coefficient was
increased concerning time, eventually stable and decreased as the pattern of an electrochemical
redox reaction as mentioned in eq.(1,2) which has glucose as electron donor and ferricyanide
as an electron acceptor. For carbon fiber electrode, which represents as the material that gives
the highest anodic diffusion coefficient (6.868x10-16 cm2/s, 9 hrs) > carbon fiber with Nitric
61

acid and PTFE electrode (2.65x10-16 cm2/s, 6 hrs.) > carbon fiber electrode with nitric acid
(1.325x10-16 cm2/s, 1 hrs.) > carbon fiber electrode with grinded sludge(1.056x10-16 cm2/s,
6hrs.)> Carbon fiber with PTFE electrode (7.48x10-17 cm2/s,24 hrs.) as shown in the table
below.

From Fig.48(b), For the cathodic diffusion coefficient, it shows that the diffusion coefficient
was decreased concerning time, eventually stable as the pattern of an electrochemical redox
reaction as mentioned in eq.(1,2) which has glucose as electron donor and ferricyanide as an
electron acceptor. For carbon fiber electrode, this represents as the material that gives the
highest anodic diffusion coefficient (1.2x10-16 cm2/s, 9 hrs) > carbon fiber with Nitric acid
and PTFE electrode (1.028x10-16 cm2/s, 12 hrs.) > carbon fiber electrode with nitric acid
(9.35x10-17 cm2/s, 1 hrs.) > carbon fiber electrode with grinded sludge(8.62x10-17 cm2/s,
12hrs.)> Carbon fiber with PTFE electrode (6.22x10-17 cm2/s,24 hrs.) as shown in the table
below.

Anodic Cathodic
Material
Time Diffusion Time Diffusion
(hour) coefficient (hour) coefficient

Carbon fiber 9 6.69E-16 9 1.20E-16

Nitric acid and


6 2.65E-16 12 1.03E-16
PTFE
Carbon fiber
with grinded 6 1.06E-16 9 8.62E-17
sludge

Nitric acid 1 1.33E-16 1 9.35E-17

PTFE 12 7.48E-17 24 6.22E-17

Table27: Diffusion coefficient

From Figs.48(a) and (b), it shows that the carbon fiber electrode gives the highest for both
anodic and cathodic diffusion coefficient compares to other types of electrode. In the other
hands, the carbon-based electrode with PTFE appears to give the lowest for both anodic and
cathodic peak current compares to other types of electrode. For the carbon-based electrode with
Nitric acid, grinded sludge, nitric acid PTFE do not give a vast difference from each other.

The assumption for this phenomena, that the carbon fiber gives the highest for both current
generation and diffusion coefficient in anodic (oxidation reaction) and cathodic (reduction
reaction) is the impractical treatment. Since the treatment for carbon-based with nitric acid
0.3M then dried in an oven at 50 C for 2 hrs. Whereas, the previous studies (G. M. JENKINS
and K. KAWAMURA, “Polymeric Carbon-Carbon Fibre, Glass and Char” (Cambridge
University Press, London, 1976).) used carbon fiber electrode treated with 3M, 2x10-4 m3 Nitric
acid then dried in oven for a few hours and it shows that the ability for adsorption and electrical
properties increase more than 40%, As a result of treating with nitric acid, can lower the surface
resistance for ion migration. For carbon-based with PTFE electrode, during the experiment 40
62

%wt. PTFE was used with carbon fiber, then dried with 100 C in the oven to get rid of the
moisture. Whereas, the previous studies, suggested the optimum amount of PTFE was 20 %wt.
For activated carbon material. Also, adding more than 30 %wt of binder resulted in pore
blockage and reduced the available site on the activated carbon electrode. For carbon-based
with Nitric acid + PTFE electrode, during the experiment 0.3M nitric acid and 30 %wt. PTFE
was used with carbon fiber, then dried with 100 C in the oven to get rid of the moisture. For
carbon-based electrode with grinded sludge, since all the previous batches were varied the
types of electrode and the granular sludge was used, but this batch was focused on the bacteria
which considered as the digestion matters that digested the glucose in the solution, were
grinded. The assumption for this batch is the grinded bacteria has more active surface area
compared to the granular sludge, this means that the effect by the diminished of electrode
surface area is predominant due to the grinded sludge was attached on the surface of the
electrode, which leads to the current and diffusion coefficient were also decreased.

In conclusion, in early stage of the experiment, the effect by the enhanced electron to the
electrode surface interaction is superior to the counter effect by the diminished surface area of
electrode due to the presence of the organic matters (glucose) in the synthesis wastewater, but
in the later stage, the effect by the diminished surface area is predominant due to the decreased
amount of the organic matters (glucose) in the synthesis wastewater.

This means that the change of cyclic voltammogram is dependent on the amount of the organic
matters that present in the synthesis wastewater, the degree of biofilm development, and the
type of electrodes due to its ability of an enhanced electron to the surface of electrodes.

7.4. The %COD removal and coulombic efficiency (CE)

The coulombic efficiency describes the efficiency with which electrons are transferred in a
system facilitating a reaction. It is a fraction of electrons, after the organic matters (glucose)
that present in the synthesis wastewater and were digested by the bacteria then gives out the
electrons, eventually converted to electricity. The coulombic efficiency was calculated by using

CS − CO
CE = × 100
CS
Where,
CE = Coulombic Efficiency
CS = Initial COD
CO = Final COD

The percentage of COD removal and coulombic efficiency (CE) parameters are not available.
Lately, there is the chemical which added as a mediator to the system to increase the electrons
transferring performance, potassium ferricyanide. Besides oxygen which is an electron
acceptor due to its high oxidation potential, potassium ferricyanide is another electron donor
used in MFC studies since its concentration is no limited to solubility like in the case of oxygen
and the result is not only a faster reaction but also get a high power output. From the previous
reasons, potassium ferricyanide was added to the system and later was found that the previous
process of COD measurement may not suit for the solution contains potassium ferricyanide.
Another discussion is glucose left in the solution might too less to use the previous process of
measurement.
63

The table below shows the %COD removal for 24 hours without adding potassium
ferricyanide;

7.5. pH of the synthesis wastewater at the early and later stage of the experiment

pH
Condition
Before After
Carbon fiber without
7 7
modification
Carbon fiber with
7 7
nitric treatment
Carbon fiber with
7 7
Teflon treatment
Carbon fiber with
nitric and Teflon 7 7
treatment
Carbon fiber with
7 7
grinded sludge
Table28: Initial and final pH for all conditions

Table28 shows that pH of the solution in the system did not change along the operating time
which is possible because in the solution there is a buffer which is added in order to stabilize
the pH of the solution in order to avoid the acidic or basic change which can affect the
performance and reaction of chemicals in the system and too basic or acidic condition might
affect or cause toxic to the bacteria also.
64

CHAPTER 8
CONCLUSION
The objectives of this study are to compare the performance in current generation and the
diffusion coefficient, which are a critical physical parameters of the species involved in an
electrochemical reaction that describes diffusion transport, which indicates the performance of
electrons migration between the solution and the electrode surface, for various types of carbon
fiber based electrode and its modification. The result shows that the carbon fiber electrode
gives the highest anodic current is 0.0478A at time 17 hours for operation and also the highest
cathodic peak current is 0.043A at 13 hours for operation. For the effects of the electrode
materials on diffusion coefficient, the carbon fiber electrode gives the highest anodic diffusion
coefficient which is 6.868x10-16 cm2/s with 9 hours of operation and also the highest cathodic
diffusion coefficient is 1.2x10-16 cm2/s with 9 hours operation.

In conclusion, in the early stage of the experiment, the effect by the enhanced electron to the
electrode surface interaction is superior to the counter effect by the diminished surface area of
electrode due to the presence of the organic matters (glucose) in the synthesis wastewater, but
in the later stage, the effect by the diminished surface area is predominant due to the decreased
amount of the organic matters (glucose) in the synthesis wastewater. This means that the
change of cyclic voltammogram is dependent on the amount of the organic matters that present
in the synthesis wastewater, the degree of biofilm development, the type of electrodes due to
its ability of an enhanced electron to the surface of electrodes, and the optimum electrode
treatment condition.
65

REFERENCES
[1] Zhuang L, Feng CH, Zhou SG, Li YT, Wang YQ. Comparison of membrane- and cloth-
cathode assembly for scalable microbial fuel cells: construction, performance and cost. Process
Biochem. 2010;45:929–934.
[2] Yu T, Bishop PL. Stratification of microbial metabolic processes and redox potential
change in an aerobic biofilm studied using microelectrodes. Water SciTechnol. 1998;37:195–
198.
[3] Yi HN, Nevin KP, Kim BC, Franks AE, Klimes A, Tender LM, Lovley DR. Selection
of a variant of Geobactersulfurreducens with enhanced capacity for current production in
microbial fuel cells. BiosensBioelectron. 2009;24:3498–3503.
[4] Wang X, Feng YJ, Ren NQ, Wang HM, Lee H, Li N, Zhao QL. Accelerated start-up of
two-chambered microbial fuel cells: effect of anodic positive poised
potential. ElectrochimActa. 2009;54:1109–1114.
[5] Varela H, Malta M, Torresi RM. Low cost in situ techniques in electrochemistry: the
quartz crystal microbalance. Quim Nova. 2000;23:664–679.
[6] Strycharz-Glaven SM, Snider RM, Guiseppi-Elie A, Tender LM. On the electrical
conductivity of microbial nanowires and biofilms. Energ Environ Sci. 2011;4:4366–4379.
[7] Coursolle D, Baron DB, Bond DR, Gralnick JA. The Mtr respiratory pathway is
essential for reducing flavins and electrodes in Shewanellaoneidensis. J
Bacteriol. 2010;192:467–474.
[8] Manohar AK, Bretschger O, Nealson KH, Mansfeld F. The use of electrochemical
impedance spectroscopy (EIS) in the evaluation of the electrochemical properties of a
microbial fuel cell. Bioelectrochemistry. 2008;72:149–154.
[9] Reguera G, McCarthy KD, Mehta T, Nicoll JS, Tuominen MT, Lovley DR.
Extracellular electron transfer via microbial nanowires. Nature. 2005;435:1098–1101.
[10] Coursolle D, Baron DB, Bond DR, Gralnick JA. The Mtr respiratory pathway is
essential for reducing flavins and electrodes in Shewanellaoneidensis. J
Bacteriol. 2010;192:467–474.
[11] Li FX, Sharma Y, Lei Y, Li BK, Zhou QX. Microbial fuel cells: the effects of
configurations, electrolyte solutions, and electrode materials on power
generation. ApplBiochemBiotechnol. 2010;160:168–181.
[12] Marsili E, Rollefson JB, Baron DB, Hozalski RM, Bond DR. Microbial biofilm
voltammetry: direct electrochemical characterization of catalytic electrode-attached
biofilms. Appl Environ Microbiol. 2008b;74:7329–7337.
[13] Rabaey K, Boon N, Siciliano SD, Verhaege M, Verstraete W. Biofuel cells select for
microbial consortia that self-mediate electron transfer. Appl Environ
Microbiol. 2004;70:5373–5382.
[14] Wei JC, Liang P, Huang X. Recent progress in electrodes for microbial fuel
cells. Bioresource Technol. 2011;102:9335–9344.
[15] Call DF, Logan BE. A method for high throughput bioelectrochemical research based
on small scale microbial electrolysis cells. BiosensBioelectron. 2011;26:4526–4531.
[16] Kissinger PT, Heineman WR, editors. Laboratory techniques in electroanalytical
chemistry. Marcel Dekker, Inc; New York (NY): 1996. p. 1008.
[17] Logan BE, Hamelers B, Rozendal RA, Schrorder U, Keller J, Freguia S, Aelterman P,
Verstraete W, Rabaey K. Microbial fuel cells: methodology and technology. Environ Sci
Tecnol. 2006;40:5181–5192.
66

[18] Clauwaert P, Aelterman P, Pham TH, De Schamphelaire L, Carballa M, Rabaey K,


Verstraete W. Minimizing losses in bio-electrochemical systems: the road to
applications. ApplMicrobiolBiotechnol. 2008;79:901–913.
[19] Fricke K, Harnisch F, Schroder U. On the use of cyclic voltammetry for the study of
anodic electron transfer in microbial fuel cells. Energy Environ Sci. 2008;1:144–147.
[20] Logan BE, Call D, Cheng S, Hamelers HVM, Sleutels THJA, Jeremiasse AW,
Rozendal RA. Microbial electrolysis cells for high yield hydrogen gas production from organic
matter. Environ Sci Technol. 2008;42:8630–8640.
[21] Wang HY, Bernarda A, Huang CY, Lee DJ, Chang JS. Micro-sized microbial fuel cell:
a mini-review. Bioresource Technol. 2011;102:235–243.
[22] Aelterman P, Freguia S, Keller J, Verstraete W, Rabaey K. The anode potential
regulates bacterial activity in microbial fuel cells. ApplMicrobiolBiotechnol. 2008;78:409–
418.
[23] Lee HS, Torres CI, Rittmann BE. Effects of substrate diffusion and anode potential on
kinetic parameters for anode-respiring bacteria. Environ Sci Technol. 2009b;43:7571–7577.
[24] Kim JR, Min B, Logan BE. Evaluation of procedures to acclimate a microbial fuel cell
for electricity production. ApplMicrobiolBiotechnol. 2005;68:23–30.
[25] Larrosa-Guerrero A, Scott K, Katuri KP, Godinez C, Head IM, Curtis T. Open circuit
versus closed circuit enrichment of anodic biofilms in MFC: effect on performance and anodic
communities. ApplMicrobiolBiotechnol. 2010;87:1699–1713.
[26] Banin E, Vasil ML, Greenberg EP 2005. Iron and Pseudomonas aeruginosa biofilm
formation. ProcNatlAcadSci 102: 11076–11081
[27] Oldham KB, Myland JC. Fundamentals of electrochemical science. Academic Press,
Inc.; San Diego (CA): 1994. p. 496.
[28] Clauwaert P, Aelterman P, Pham TH, De Schamphelaire L, Carballa M, Rabaey K,
Verstraete W. Minimizing losses in bio-electrochemical systems: the road to
applications. ApplMicrobiolBiotechnol. 2008;79:901–913.
[29] Mehanna M, Kiely PD, Call DF, Logan BE. Microbial electrodialysis cell for
simultaneous water desalination and hydrogen gas production. Environ Sci
Technol. 2010;44:9578–9583.
[30] Marsili E, Baron DB, Shikhare ID, Coursolle D, Gralnick JA, Bond
DR. Shewanella secretes flavins that mediate extracellular electron transfer. P NatlAcadSci
USA. 2008a;105:3968–3973.
[31] Richter H, McCarthy K, Nevin KP, Johnson JP, Rotello VM, Lovley DR. Electricity
generation by Geobactersulfurreducens attached to gold electrodes. Langmuir. 2008;24:4376–
4379.
[32] Yang SN, Lewandowski Z. Measurement of local mass-transfer coefficient in
biofilms. BiotechnolBioeng. 1995;48:737–744.
[33] Xia FH, Beyenal H, Lewandowski Z. An electrochemical technique to measure local
flow velocity in biofilms. Water Res. 1998;32:3631–3636.
[34] Strycharz SM, Malanoski AP, Snider RM, Yi H, Lovley DR, Tender LM. Application
of cyclic voltammetry to investigate enhanced catalytic current generation by biofilm-modified
anodes of Geobactersulfurreducens strain DL1 vs. variant strain KN400. Energ Environ
Sci. 2011;4:896–913.
[35] Nguyen HD, Renslow R, Babauta J, Ahmed B, Beyenal H. A voltammetricflavin
microelectrode for use in biofilms. Sensor Actuat B-Chem. 2012b;161:929–937.
[36] Bard AJ, Faulkner LR, editors. Electrochemical methods: fundamentals and
applications. John Wiley & Sons, Inc.; New York (NY): 2001. p. 856.
[37] Bard AJ, Faulkner LR, editors. Electrochemical methods: fundamentals and
applications. John Wiley & Sons, Inc.; New York (NY): 2001. p. 856.
67

[38] Snider RM, Guissepi-Elie A, Strycharz-Glaven S, Tender L. Abstracts of Papers of the


American Chemical Society. American Chemical Society; Washington (DC): 2011. 241. On
the conductive nature of biofilms of Geobactersulfurreducens. Abstract No. 371-BIOT.
[39] Strycharz-Glaven SM, Snider RM, Guiseppi-Elie A, Tender LM. On the electrical
conductivity of microbial nanowires and biofilms. Energ Environ Sci. 2011;4:4366–4379.
[40] Beyenal H, Lewandowski Z. Mass-transport dynamics, activity, and structure of
sulfate-reducing biofilms. AIChE Journal. 2001;47:1689–1697.
[41] McLean JS, Wanger G, Gorby YA, Wainstein M, McQuaid J, Ishii SI, Bretschger O,
Beyenal H, Nealson KH. Quantification of electron transfer rates to a solid phase electron
acceptor through the stages of biofilm formation from single cells to multicellular
communities. Environ Sci Technol. 2010;44:2721–2727.
[42] Nevin KP, Kim BC, Glaven RH, Johnson JP, Woodard TL, Methe BA, DiDonato RJ,
Covalla SF, Franks AE, Liu A, et al. Anode biofilm transcriptomics reveals outer surface
components essential for high density current production in Geobactersulfurreducens fuel
cells. Plos One. 2009;4(5):e5628.
[43] Majors PD, McLean JS, Scholten JCM. NMR bioreactor development for live in-situ
microbial functional analysis. J MagnReson. 2008;192:159–166.
[44] McLean JS, Majors PD, Reardon CL, Bilskis CL, Reed SB, Romine MF, Fredrickson
JK. Investigations of structure and metabolism within Shewanellaoneidensis MR-1 biofilms. J
Microbiol Meth. 2008;74:47–56.
[45] Cao B, Shi LA, Brown RN, Xiong YJ, Fredrickson JK, Romine MF, Marshall MJ,
Lipton MS, Beyenal H. Extracellular polymeric substances from Shewanella sp. HRCR-1
biofilms: characterization by infrared spectroscopy and proteomics. Environ
Microbiol. 2011;13:1018–1031.
[46] Sand W, Gehrke T. Extracellular polymeric substances mediate
bioleaching/biocorrosion via interfacial processes involving iron(III) ions and acidophilic
bacteria. Res Microbiol. 2006;157:49–56.
[47] A. Bressel, J. Schultze, W. Khan, G. Wolfaardt, H.P. Rohns, R. Irmscher, M. Schoning,
Electrochim. Acta 48 (2003) 3363.
[48] L. Tamachkiarow, H. Flemming, Water. Sci. Technol.: J. Int. Assoc. Water Pollut. Res.
47 (2003) 19.
[49] Wrana N, Sparling R, Cicek N, Levin DB. Hydrogen gas production in a microbial
electrolysis cell by electrohydrogenesis. J Clean Prod. 2010;18:S105–S111.
[50] C. Punckt, M.A. Pope and I.A. Aksay 2013 The Journal of Physical Chemistry C 117
16076- 16086
[51] S. Cheng and J. Wu 2013 Bioelectrochemistry 92 22-26
[52] H. Dong, H.B. Yu, X. Wang, Q.X. Zhou and J.L. Feng 2012 Water Research 46 5777-
5787
[53] H. Dong, H.B. Yu and X. Wang 2012 Environmental Science and Technology 46
13009-13015
[54] X.Y. Zhang, X. Xia, I. Ivanov, X. Huang and B.E. Logan 2014 Environmental Science
and Technology 48 2075-2081
[55] A. Gambou-Bosca and D. Belanger 2015 Journal of The Electrochemical Society 162
A5115-5123
[56] Z.T. Zhu, S.H. Tang, J.W. Yuan, X.L. Qin, Y.X. Deng, R.J. Qu and G.M. Haarberg
2016 International Journal of Electrochemical Science 11 8270-8279
[57] T. Shimizu, Y. Higashihata and T. Nakamura (1998). United States Patent No.
5707763.
[58] I.I. Misnon, N.K.M. Zain, R.A. Aziz, B. Vidyadharan and R. Jose 2015
ElectrochimicaActa 17478-86
68

[59] Q. Abbas, D. Pajak, E. Frąckowiak and F. Béguin 2014 ElectrochimicaActa 140 1328
[60] Hawaiah Imam Maarof et al 2017 IOP Conf. Ser.: Mater. Sci. Eng. 210 012011
[61] B.R. Corry, Carbon black, in: H.S. Katz, J.V. Milewski (Eds.), Handbook of Fillers For
Plastics. Section VI. Carbon Black and Organic Fillers, Van Nostrand Reinhold, New York,
1987 (Chapter 19).
[62] M. Uchida, Y. Aoyama, M. Tanabe, N. Yanagihara, N. Eda,A. Ohta, Influences of both
carbon supports and heat-treatment of supported catalyst on electrochemical oxidation of
methanol, J. Electrochem. Soc. 142 (1995) 2572–2576.
[63] S. Mrozowski, Studies of carbon powders under compression,in: Proceedings of the
Third Biennial Conference on Carbon, University of Buffalo, Buffalo, New York, June 17–27,
1957, 495 pp.
[64] D. Pantea et al., Applied Surface Science 217 (2003) 181–193
[65] H. Darmstadt, C. Roy, Comparative investigation of defects on carbon black surfaces
by nitrogen adsorption and SIMS, Carbon 39 (2001) 841–848.)
[66] F. Duan, Y. Li, H. Cao, Y. Wang, J.C. Crittenden and Y. Zhang 2015 Chemosphere
125 205-21
[67] T. Kim, J. Kang, J. Lee, J. Yoon, Water Res. 45 (2011) 4615.
[68] S. Kurissery, N. Kanavillil, K. Leung, A. Chen, L. Davey, H. Schraft, Biofouling 26
(2010) 799.
[69] M. Tian, N. Kanavillil, L. Davey, K. Leung, H. Schraft, A. Chen, Electroanal. Chem.
J. 611 (2007) 133.
[70] A.J. Bard, L.R. Faulkner, Electrochemical Methods, Wiley, 2001.) +(J. Wang, J.
Farrell, Environ. Sci. Technol. 37 (2003) 3891.) +( T. Kim, J. Kang, J. Lee, J. Yoon, Water
Res. 45 (2011) 4615
[71] R.M. Landry, D. An, J.T. Hupp, P.K. Singh, M.R. Parsek, Mol. Microbiol. 59 (2006)
142
[72] Dreiner, S.; Schürmann, M.; Westphal, C.; Zacharias, H.; Phys. Rev. Lett. 2001, 86,
4068
[73] Dreiner, S.; Schürmann, M.; Westphal, C.; Zacharias, H.; Phys. Rev. Lett. 2001, 86,
4068.
[74] T. Ito, D. M. N. T. Perera and S. Nagasaka, Journal of Chemical Education 2008, 85,
1112-1115.

Anda mungkin juga menyukai