Anda di halaman 1dari 12

Anestesi, 2011, 66, halaman 40-51 doi: 10,1111 / j.1365-2044.2010.06564.

x
.....................................................................................................................................................................................................................

MENGULAS ARTIKEL

Panas dan kelembaban exchanger dan lters sistem pernapasan fi: penggunaannya dalam
anestesi dan perawatan intensif.
Bagian 2 - penggunaan praktis, termasuk masalah, dan menggunakan mereka dengan
pasien anak-anak

AR Wilkes

Senior Research Fellow, Anestesi, Perawatan Intensif dan Pain Medicine, Universitas Cardiff, Cardiff, Wales, UK

Ringkasan
lters panas dan penukar kelembaban dan sistem pernapasan fi dimaksudkan untuk menggantikan normal pemanasan,
pelembab dan penyaringan fungsi saluran napas atas. Yang pertama bagian dari ulasan ini dianggap sejarah,
prinsip-prinsip operasi dan efisiensi dari perangkat ini. Tujuan dari ini bagian dari tinjauan ini adalah untuk meringkas
pedoman baru pada penggunaan perangkat ini dan menguraikan masalah yang dapat terjadi. Secara khusus, efek dari
perangkat ini pada analisis gas, ruang mati, ketahanan terhadap gas aliran dan penyumbatan sistem pernapasan
dianggap. Pada anak-anak, penting untuk mempertimbangkan penambahan ruang mati dan ketahanan terhadap fl ow
gas. Sebuah berat badan 2,5 kg mungkin batas berat yang lebih rendah untuk digunakan dengan panas dan kelembaban
penukar, dan 3 kg untuk lters fi.

.......................................................................................................

Korespondensi: Dr Antony R. Wilkes Email:


wilkes@cf.ac.uk yang diterima: 19 Oktober 2010

Seperti yang dibahas sudah dalam bagian pertama dari ulasan ini [1], panas mungkin juga terjadi. Pedoman telah dirumuskan untuk penggunaan
dan kelembaban penukar (HMEs) dimaksudkan untuk melestarikan sebagian perangkat ini untuk pasien dengan spesifik penyakit fi c, atau mencegah
dari panas dan kelembaban dan kondisi terinspirasi gas dihembuskan pasien pasien dari mengalah pada penyakit tertentu. Beberapa pedoman ini
dengan pemanasan dan pelembab itu [2]; bernapas lters sistem fi membedakan antara dua jenis utama dari sistem pernapasan fi lters: 'lipit'
dimaksudkan untuk mengurangi transmisi mikroba dan partikel lainnya dalam dan 'elektrostatik'. Untuk penjelasan dari istilah-istilah ini, lihat bagian
sistem pernapasan [3] ketika saluran udara bagian atas pasien telah dilewati pertama dari ulasan ini [1].
selama anestesi dan perawatan intensif. Panas- dan kelembaban-bertukar
lters fi (HMEFs) menyediakan kedua fungsi. Dalam hal ini bagian dari tinjauan, masalah yang dapat terjadi
dengan penggunaan HMEs dan lters fi dijelaskan, panduan tentang
penggunaan perangkat ini ditinjau dan rekomendasi lebih lanjut
Untuk melembabkan gas dan melindungi sistem pernapasan dari tetesan dikembangkan.
infektif kedaluwarsa, perangkat ditempatkan antara pasien dan sistem
pernapasan [4]. Dalam posisi ini, hal itu menambah ruang mati, sehingga
Komplikasi yang terkait dengan penggunaan HMEs dan lters fi
meningkatkan rebreathing (memerlukan peningkatan ventilasi atau
memungkinkan hiperkapnia diperbolehkan), dan juga menambah ketahanan
terhadap gas mengalir, sehingga meningkatkan kerja pernapasan. Efek pada ketahanan terhadap gas fl ow dan ventilasi
Penyumbatan juga dapat terjadi jika cairan memasuki perangkat. Jadi, Ketika HMEs atau lters fi ditambahkan ke sistem pernapasan, mereka
meskipun pasien mungkin manfaat dari penambahan perangkat ke sistem meningkatkan ketahanan terhadap gas mengalir dan karenanya kerja pernapasan.
pernapasan, beberapa bahaya Dalam sebuah penelitian membandingkan HMEF elektrostatik besar dengan
dipanaskan humidi fi er pada pasien

2010 Penulis The


40 Anestesi 2010 Asosiasi Dokter-dokter anestesi dari Britania Raya dan Irlandia
Anestesi, 2011, 66, halaman 40-51 AR Wilkes Æ HMEs di anestesi dan perawatan intensif - penggunaan praktis
.....................................................................................................................................................................................................................

menerima ventilasi non-invasif yang memiliki kegagalan pernafasan ventilasi mekanik di ICU [12], ventilasi meningkat dengan penggunaan
akut hiperkapnia, itu menunjukkan bahwa menit ventilasi secara HME dan fi lter, dan peningkatan ventilasi berkorelasi dengan ukuran
signifikan lebih tinggi ketika fi lter digunakan dan bahwa ini juga terkait ruang mati. perangkat yang lebih kecil dapat memiliki output kelembaban
dengan peningkatan yang ditandai dalam karya bernapas [5]. rendah dari perangkat yang lebih besar. Dalam penelitian ini, dapat
disimpulkan bahwa, ketika memilih fi humidi pasif er, orang yang dipilih
Iotti et al. dibandingkan sebuah HMEF besar-lipit dengan perangkat harus memiliki ruang mati terkecil tetapi memenuhi persyaratan keluaran
HME-satunya dan dipanaskan humidi fi er pada pasien yang menerima tekanan air yang diinginkan.
dukungan ventilasi untuk kegagalan pernafasan akut [6] dan menunjukkan
bahwa ventilasi meningkat secara signifikan dengan HMEF, dan bahwa Untuk mengurangi risiko trauma paru-paru atau cedera selama ventilasi
resistensi saluran napas meningkat secara signifikan dengan kedua HME dan mekanik yang berkepanjangan di ICU, target volume tidal harus dikurangi. Ini
yang HMEF. Iotti et al. menyimpulkan bahwa 'arti Pejabat hidung' tetap akan memiliki efek pada ukuran sistem pernapasan fi lter yang dapat
merupakan alternatif yang menarik untuk dipanaskan ers fi humidi, tapi digunakan (lihat di bawah).
preferensi yang harus diberikan kepada penggunaan lowvolume dan resistansi
rendah perangkat. Penyumbatan dengan cairan

Cair dapat mengalir ke HMEs dan lters fi baik dari pasien (sputum atau
Resistensi terhadap gas fl ow dapat meningkat tajam selama edema paru), atau dari sistem pernapasan (jika kondensasi hadir). Hal ini
ventilasi mekanik paru-paru di unit perawatan intensif (ICU), terutama dapat menyebabkan peningkatan resistensi terhadap gas fl ow dan, dalam
jika koleksu cair dalam fi lter [7], meskipun, dalam penelitian khusus ini, beberapa kasus, oklusi lengkap, mencegah ventilasi yang memadai dari
tidak ada gejala sisa penting yang timbul dari ini. paru-paru [13, 14]. The resistensi yang tinggi terhadap fl ow gas dari HME
atau fi lter dapat mencukupi untuk mencegah aktivasi alarm ventilator
Namun, dalam sebuah studi yang menyelidiki efek dari tiga HMEs putuskan [15]. Identifikasi penyebab peningkatan yang substansial dalam
dan HMEFs pada tingkat tekanan auto positif akhir ekspirasi (PEEP) tekanan udara puncak karena dari fi lter diblokir tidak selalu segera jelas
dan hyperin dinamis inflasi paru-paru, dinilai dengan mengukur sisa sebagai fi lter mungkin tersembunyi di bawah tirai [16]. Bahkan jika fi lter
kapasitas fungsional (FRC) pada pasien ICU dengan penyakit terlihat, mungkin sulit untuk mengidentifikasi penumpukan cairan di
paru-paru obstruktif kronis menerima ventilasi mekanik, perangkat tidak dalamnya [17, 18]. Jika fi lter ditempatkan di bawah tingkat paru-paru
memiliki efek yang signifikan di kedua auto PEEP atau FRC setelah 12 pasien, cairan dapat mengalir dengan mudah ke dalam fi lter dan
jam penggunaan [8]. Selain itu, tidak ada gejala sisa lainnya. menyebabkan obstruksi. Suf Volume fi sien cairan dapat mencegah
kedaluwarsa pasien, tetapi memungkinkan ventilator untuk memberikan
napas berikutnya, sehingga fi lter bertindak sebagai katup satu arah [19].
Untuk mengurangi risiko ini, fi lter harus ditempatkan pada tingkat lebih
Ruang mati tinggi dari paru-paru pasien, dengan lapisan fi lter dalam vertikal, bukan
Panas dan kelembaban exchanger dan lters fi menambah ruang mati dari horizontal, orientasi [19].
sistem pernapasan ketika mereka terhubung antara pasien dan sistem
pernapasan, sehingga proporsi yang lebih besar dari karbon dioksida yang
dihembuskan dikembalikan dalam napas berikutnya. Menambahkan fi lter
dengan volume internal yang besar, daripada fi er humidi, dapat Filter pada anggota tubuh ekspirasi dapat menyumbat jika pasien
menyebabkan peningkatan yang signifikan dalam tingkat pernapasan expectorates sekresi berdarah [20, 21] dan jika kondensasi
spontan dan arteri tekanan parsial karbon dioksida [9]. Selama percobaan mengumpulkan dalam fi lter [22]. Hal ini terutama masalah jika fi lter
penyapihan, ruang mati tambahan dari fi lter dibandingkan dengan digunakan bersama dengan dipanaskan humidi fi er [23]. Penyumbatan
dipanaskan humidi fi er perlu diperhitungkan [10]. Ketika membandingkan lters fi juga dapat terjadi jika fi lter digunakan bersama dengan obat
perangkat, beberapa perangkat yang lebih kecil cenderung memiliki nebulised [24].
ketahanan yang lebih besar untuk fl ow gas daripada yang lebih besar.
Dengan demikian, meskipun perangkat yang lebih besar dapat Efek pada kapnografi
menyebabkan peningkatan ventilasi, perangkat yang lebih kecil Panas dan kelembaban exchanger dan lters fi dapat mempengaruhi
meningkatkan drive ventilasi dan kerja pernapasan [11]. Namun, dalam gelombang kapnografi. Hal ini sangat mungkin di mana volume internal
studi tertentu, 2 O kompensasi untuk peningkatan kerja pernapasan. dari fi lter adalah proporsi yang signifikan dari volume tidal: dalam
kasus ekstrim kapnografi jejak dapat hilang sama sekali [25]. Sampel
gas untuk analisis dapat diambil baik dari pasien-atau machineside dari
fi lter. Bisa ada perbedaan dalam tingkat ditampilkan karbon dioksida
Dalam sebuah penelitian yang membandingkan efek dari panas humidi fi tergantung pada sisi mana fi lter sampel diambil dari [26]; efek ini
er, sebuah HME (dengan ruang mati dari 28 ml) dan fi lter (dengan ruang tampaknya lebih besar untuk lters fi yang lebih besar atau ketika
mati dari 90 ml) pada variabel pernafasan pada pasien spontan bernapas volume tidal adalah
dan lumpuh menerima

2010 Penulis Anestesi 2010 Asosiasi Dokter-dokter anestesi dari Britania Raya dan Irlandia
41
AR Wilkes Æ HMEs di anestesi dan perawatan intensif - penggunaan praktis Anestesi, 2011, 66, halaman 40-51
.....................................................................................................................................................................................................................

kecil [27, 28]. Namun, sampel gas harus dihapus dari mesin-sisi lters fi baik terjajah atau terinfeksi Pseudomonas aeruginosa.
di mana dimungkinkan karena sampel ini akan disaring, mengurangi Pada periode Januari-Juni tahun 1985, 384 pasien dirawat; 75
resiko kontaminasi, dan umumnya akan lebih kering. Efek yang sama diperlukan ventilasi dan menerima humidi fi kasi dari fi lter saja, di
dapat terjadi dengan pengukuran volume tidal mana penggunaan HME antaranya hanya 16 yang baik terjajah atau terinfeksi Pseudomonas
dapat menyebabkan meremehkan volume tidal dibandingkan dengan aeruginosa.
dipanaskan humidi fi er [29]. Sejumlah calon percobaan terkontrol acak kemudian telah dilakukan
membandingkan efek dari perangkat fi kasi humidi (HME, HMEF atau
dipanaskan humidi fi er) terhadap kejadian ventilator-associated
efek lainnya pneumonia (Tabel 1). Satu-satunya studi menunjukkan keuntungan yang
Ada bukti yang menunjukkan bahwa kontak yang terlalu lama (> 4 jam) signifikan dari HMEFs adalah bahwa dari Kirton et al. [49], dan
untuk tingkat tinggi des fl URANE (2 MAC) mengurangi kinerja filtrasi dari satu-satunya studi menunjukkan keuntungan yang signifikan dari
beberapa lters fi elektrostatik dimaksudkan untuk digunakan dengan orang dipanaskan ers fi humidi adalah Lorente et al. [53]. Namun, dalam studi
dewasa [30] dan anak-anak [31]. Lawes dijelaskan efek lain bahwa khusus ini, perangkat pasif adalah produk HME-satunya dan bukan fi lter.
penggunaan HMEs dan lters fi dapat memiliki pada pasien pemantauan
selama anestesi [32]. Namun, anggapan bahwa penggunaan HMEs dan
lters fi di anestesi dapat menyebabkan pengeringan penyerap karbon Empat meta-analisis dan dua ulasan bukti telah dilakukan
dioksida tidak mungkin, sebagai reaksi dari karbon dioksida yang membandingkan kejadian VAP ketika baik HMEs atau ers fi humidi
dihembuskan dengan penyerap yang menghasilkan air, sehingga penyerap digunakan (Tabel 1). Kesimpulan yang dicapai tergantung pada
yang sangat tidak mungkin mengering ketika digunakan dengan pasien. penelitian yang termasuk dalam analisis. Namun, terlepas dari fakta
Oleh karena itu masalah lebih lanjut terkait dengan pengeringan penyerap bahwa beberapa kesimpulan yang mendukung penggunaan perangkat
(ketidaktelitian pulse oximetry dan produksi senyawa A dari agen anestesi fi kasi humidi pasif, bukti tidak memiliki kekuatan.
volatil) juga sangat mungkin terjadi.
Kola et al. melaporkan bahwa penurunan kejadian VAP ketika
menggunakan HMEF lebih ditandai selama ventilasi berkepanjangan
[39]. Hal ini juga ditemukan dalam penelitian oleh Kranabetter et al., Di
mana perbedaan antara fi lter dan kelompok er humidi fi hanya signifikan
pedoman
untuk pasien yang memerlukan ventilasi selama lebih dari 2 hari [55].
ventilator-associated pneumonia
Antara 8% dan 28% dari pasien yang membutuhkan ventilasi mekanis Pedoman telah diterbitkan pada metode untuk mencegah VAP. pedoman
mengembangkan ventilator-associated pneumonia (VAP) [33]. Tingkat baru-baru ini yang saling bertentangan dalam beberapa saran mereka. Sehubungan
kematian untuk VAP relatif tinggi, mulai from24% sampai 50%, sedangkan dengan panas dan kelembaban penukar dan lters sistem pernapasan fi:
tingkat themortality dapat setinggi 76% dalam beberapa situasi c fi spesifik atau
bila infeksi paru-paru disebabkan oleh sangat patogen berisiko tinggi. • Pusat Pengendalian dan Pencegahan Penyakit [56] membuat
beberapa rekomendasi:
Organisme dominan yang bertanggung jawab adalah Staphylococcus aureus, Pseudomonas
aeruginosa dan Enterobacteriaceae. - Menggunakan defisiensi lters bakteri fi tinggi-ef di berbagai posisi
di sirkuit pernafasan (misalnya di Y-piece atau pada inspirasi dan
ventilator-associated pneumonia dikaitkan dengan peningkatan ekspirasi sisi sirkuit) telah menganjurkan dan terbukti
durasi ventilasi mekanis. Dalam studi kasus kontrol besar, pasien menurunkan kontaminasi sirkuit. Namun, efficacy dari lters fi
dengan VAP memiliki mean (SD) durasi secara signifikan lebih lama bakteri dalam mencegah kesehatan pneumonia terkait perawatan
ventilasi (14,3 (15,5) hari dibandingkan dengan 4,7 (7,0) hari) [34]. belum terbukti.
Safdar dkk. Diperkirakan VAP menyebabkan peningkatan durasi tinggal
di ICU 6,1 hari [35]. - Tidak ada rekomendasi dapat dibuat untuk penggunaan preferensial baik HMEs

atau dipanaskan ers fi humidi untuk mencegah pneumonia pada pasien yang

Gallagher et al. merekomendasikan penggunaan sistem pernapasan membutuhkan ventilasi mekanis dibantu, dan ini masih merupakan masalah yang

fi lter, ditempatkan di antara pasien dan Y-piece dari sistem belum terselesaikan.

pernapasan, sebagai satu-satunya sarana humidi fi kasi pada pasien - Sebuah HME yang di gunakan pada pasien harus diubah jika
yang menerima ventilasi mekanik di ICU berdasarkan pengurangan malfungsi mekanis atau menjadi terlihat kotor, tetapi HMEs tidak
infeksi dan kolonisasi pasien dengan Pseudomonas aeruginosa dalam harus rutin diganti lebih sering daripada setiap 48 jam.
sebuah studi kasus-kontrol yang tak tertandingi [36]. Pada periode
Januari-Juni 1984, 380 pasien dirawat ICU; 66 diperlukan dikontrol - Dengan tidak adanya kontaminasi kotor atau kerusakan, sirkuit
ventilasi dan menerima humidi fi kasi dari mandi air panas humidi fi er, pernafasan melekat pada HME tidak harus diubah secara rutin
di antaranya 35 adalah ketika sedang digunakan pada pasien.

2010 Penulis The


42 Anestesi 2010 Asosiasi Dokter-dokter anestesi dari Britania Raya dan Irlandia
Anestesi, 2011, 66, halaman 40-51 AR Wilkes Æ
HMEs di anestesi dan perawatan intensif - penggunaan praktis
.....................................................................................................................................................................................................................

Tabel 1 Perbandingan panas dan kelembaban penukar (HMEs) dan dipanaskan ers fi humidi (HH) terhadap kejadian VAP (ventilator-associated pneumonia). Nilai
adalah jumlah (proporsi) dan risiko relatif (95% CI).

review
Perbandingan Hasil Termasuk dalam meta-analisis sistematis

Masak et Hess et Kola et al. ‡ Siempos et Collard et Dodek et al.


al. * [37] al. † al § [40] al. - ** [42]
Belajar HME (F) vs HH Kejadian VAP [38] [39] [41]

3 3 3 3 3 3
Martin et al. †† [ 43] PUBCF vs dipanaskan humidi fi er HME 2 / 31 (6%) vs HH 8 / 42 (19%),
(unspeci fi ed) 0,34 (0,08-1,49)
3 3 3 3 3 3
Roustan et al. BB22-15 vs Aquaport HME 5 / 55 (9%) vs HH 9 / 61 (15%),
[44] 0,62 (0,22-1,73)
3 3 3 3 3 3
Dreyfuss et al. Hygrobac vs berbagai KK HME 6 / 61 (10%) vs HH 8 / 70 (11%),
[45] 0.86 (0,32-2,34)
3 3 3
Branson et al. Baxter HME ‡‡ vs MR730 HME 3 / 49 (6%) vs HH 3 / 54 (6%),
[46] 1.10 (0,23-5,21)
3 3 3
Boots et al. [47] Lembab-Vent Filter Cahaya vs HME 14 / 75 (19%) vs HH 7 / 41 (17%),
MR730 1,09 (0,48-2,49)
3 3 3
Hurni et al. [48] Hygroster vs berbagai HH HME 5 / 59 (8%) vs HH 7 / 56 (13%),
0,68 (0,23-2,01)
3 3 3 3 3 3
Kirton et al. [49] BB100 vs Marquest HME 9 / 140 (6%) vs 22 / 140 (16%),
0,41 (0,20-0,86)
3 3 3 3
Kollef et al. [50] Durasi vs MR730 HME 15 / 163 (9%) vs HH 15 / 147 (10%),
0,90 (0,46-1,78)
3 3 3
Memish et al. Hudson HME Filter vs HH HME 14 / 123 (11%) vs HH 19 / 120 (16%),
[51] 0,72 (0,38-1,37)
3
Lacherade et al. Hygrobac vs MR730 HME 47 / 185 (25%) vs HH 53 / 184 (29%),
[52] 0.88 (0,63-1,23)
3
Lorente et al. Edith Flex vs MR850 & HME 21 / 53 (40) vs HH 8 / 51 (16),
[53] Aerodyne 2000 2,53 (1,23-5,18)

3
Boots et al. [54] Lembab-Vent Filter Compact HME 24 / 190 (13%) vs HH 23 / 191 (12%),
vs MR730 1,05 (0,61-1,79)

VAP, ventilator-associated pneumonia; PUBF, Pall Ulpitor bernapas sirkuit fi lter.


* tarif VAP lebih rendah mungkin terkait dengan penghindaran ers fi humidi dan penggunaan HMEs.
† Meskipun bukti menunjukkan tingkat VAP lebih rendah dengan pasif dari humidi aktif fi kasi masalah lain menghalangi rekomendasi umum untuk penggunaan umum ers fi humidi pasif.

‡ Pengurangan risiko VAP pada kelompok HME khususnya di ventilasi mekanis untuk setidaknya 7 hari.
§The bukti yang tersedia tidak mendukung kinerja preferensial baik ers humidi fi pasif atau aktif dalam ventilasi mekanik dalam hal VAP kejadian, kematian atau morbiditas.

- Tidak terdaftar di antara praktik pencegahan dengan bukti pendukung terkuat. Meta-analisis akan membantu untuk menentukan apakah HMEs terkait dengan tingkat yang lebih rendah dari VAP.

* * Direkomendasikan: HMEs dengan tidak adanya kontra-indikasi.


†† Studi dihentikan lebih awal karena kematian seorang pasien dalam kelompok HME dari tabung trakea diblokir.
‡‡ HME-satunya perangkat.

- Tidak ada rekomendasi dapat dibuat untuk menempatkan fi lter bakteri ing kelompok [59], tercatat (dengan mengacu Masak et al. [37]) bahwa pasien yang

dalam pernapasan systemor sirkuit pasien peralatan anestesi; ini adalah menggunakan HMEs memiliki insiden yang secara signifikan lebih rendah dari VAP bila

masalah yang belum terselesaikan. dibandingkan dengan mereka yang menggunakan dipanaskan ers fi humidi.

• The American Thoracic Society menyatakan bahwa penukar


heatmoisture menurunkan ventilator sirkuit kolonisasi, tapi tidak secara • Partai bekerja pada pneumonia didapat di rumah sakit dari British
signifikan mengurangi insiden VAP. Oleh karena itu mereka tidak dapat Society for Antimicrobial Chemotherapy merekomendasikan
dianggap sebagai alat untuk mencegah VAP [57]. penggunaan HMEs daripada dipanaskan ers fi humidi 'sebagai HMEs
lebih efektif dalam mengurangi kejadian VAP' [60]. Ketika memilih
• Eropa Perawatan Bundle untuk pencegahan VAP peringkat 'HMEs sebuah HME, pihak bekerja direkomendasikan bahwa output
lebih disukai' (lebih dipanaskan ers fi humidi) hanya kesebelas dalam kelembaban yang memadai untuk meminimalkan risiko obstruksi jalan
urutan kepentingan dari 16 langkah yang disarankan [58]. napas diperlukan. Selanjutnya, HMEs tidak perlu diubah secara rutin.
Partai kerja merekomendasikan bahwa lters fi yang tepat digunakan
• Meskipun tidak salah satu dari 20 poin konsensual dicapai oleh Eropa untuk melindungi mekanik
didapat di rumah sakit pneumonia dengan pekerjaan

2010 Penulis Anestesi 2010 Asosiasi Dokter-dokter anestesi dari Britania Raya dan Irlandia
43
AR Wilkes Æ HMEs di anestesi dan perawatan intensif - penggunaan praktis Anestesi, 2011, 66, halaman 40-51
.....................................................................................................................................................................................................................

sirkuit ventilator dari kontaminasi bakteri dan, dalam anestesi, lters fi salah satu yang lters fi partikel 0,3 l m dalam ukuran dengan e fi siensi
harus diubah antara pasien. lebih dari 95% di kedua negara dibongkar dan dimuat di maksimum aliran
tingkat ventilator. ow fl ini akan lebih besar dari mengalir umumnya
• Masyarakat untuk Kesehatan Epidemiologi Amerika / digunakan untuk menentukan kinerja filtrasi (15 dan 30 l.min) 1 untuk lters fi
Penyakit Infeksi Society of America tidak secara khusus menyebutkan ditujukan untuk digunakan dengan pasien anak dan dewasa,
penggunaan HMEs atau lters fi [61]. masing-masing [3]); meningkatkan aliran mengurangi infiltrasi efisiensi
• VAP Pedoman Komite dan Kanada Critical Care Trials Grup tidak dari sistem pernapasan fi lters [71].
membuat rekomendasi untuk jenis humidi fi kasi, tapi tidak
merekomendasikan penggunaan lters fi [62]. Jika HMEs digunakan,
dianjurkan untuk mengubah mereka untuk setiap pasien, setiap 5-7 Parah sindrom pernapasan akut
hari dan sebagai indikasi klinis. Namun, pedoman sebelumnya dari Berat akut sindrom pernafasan (SARS) adalah pneumonia atipikal yang
kelompok yang sama [42] telah merekomendasikan penggunaan menyebar dengan cepat dari Asia selatan-timur ke Kanada dan banyak
HMEs dengan tidak adanya kontraindikasi dengan perubahan negara lain di Amerika, Eropa dan Australasia pada akhir tahun 2002 dan
mingguan, meskipun kelompok itu menyimpulkan bahwa hanya ada awal 2003. Dengan 7 Agustus
insiden sedikit menurun dari VAP dibandingkan dengan dipanaskan 2003, sudah ada 8422 kasus yang dilaporkan fi nite de pasien dengan
ers fi humidi. pertimbangan biaya disukai penggunaan HMEs. SARS, dari yang 916 telah meninggal [72]. SARS tidak biasa bahwa
sebagian besar orang-orang yang terinfeksi adalah pekerja kesehatan:
dalam satu kohort dari 144 pasien dengan SARS di Greater Toronto
Area, 73 (51%) adalah pekerja kesehatan [73]. Penyebaran virus itu
tuberkulosis didominasi oleh tetesan menular dan kontak. Di Toronto, enam pekerja
Tuberkulosis (TB) bertanggung jawab untuk diperkirakan 1,3 juta kesehatan dikontrak SARS setelah berpartisipasi dalam sulit dan
kematian di seluruh dunia pada tahun 2008 [63]. Diperkirakan bahwa berkepanjangan intubasi trakea. Beberapa pekerja kesehatan terinfeksi
satu orang yang baru terinfeksi setiap detik dan bahwa sepertiga dari meskipun memakai alat pelindung diri penuh, termasuk N95 respirator
populasi dunia saat ini terinfeksi dengan basil TB. Infeksi silang terjadi lters fi.
oleh penularan melalui udara dari basil ketika pasien terinfeksi batuk,
bersin dan berbicara. Setiap orang dengan penyakit TB aktif akan
menginfeksi rata-rata antara 10 dan 15 orang setiap tahun jika mereka Darurat Perawatan Research Institute merekomendasikan bahwa lters sistem
tidak diobati [63]. Dosis menular diperkirakan kurang dari 10 basil [64]. pernapasan fi dimasukkan dalam anggota tubuh ekspirasi dari ventilator setiap
Perawatan dan pengobatan pasien dengan TB mungkin melibatkan digunakan pada pasien dengan SARS [74]. Mereka menyatakan bahwa bernapas
penggunaan ventilasi mekanik. sirkuit lters fi memiliki filtrasi ketidakefisienan ef fi bakteri dan virus dari 99,97%
atau lebih besar akan menawarkan perlindungan yang sama atau lebih baik dari
defisiensi partikulat lters udara fi high-ef. Selain itu, mereka menyarankan
Infeksi silang dari TB di fasilitas pelayanan kesehatan telah mempertimbangkan penggunaan HMEs (atau sirkuit kawat dipanaskan) untuk
dilaporkan. Jereb et al. menggambarkan bagaimana 10 pasien di unit meminimalkan kelembaban dalam sistem pernapasan dan dengan demikian beban
transplantasi ginjal terinfeksi TB, lima di antaranya meninggal [65]. kelembaban pada fi lter. Hal ini karena kelembaban terakumulasi di fi lter dapat
Masa inkubasi rata-rata untuk TB dalam kelompok menyebabkan peningkatan resistensi ekspirasi atau menyebabkan obstruksi.
immunocompromised pasien ini adalah 7,5 minggu. Penularan juga
terjadi antara pasien dan kesehatan pekerja [66]. Basil TB dapat
melewati sirkuit pernapasan anestesi, termasuk tabung soda kapur, Thiessen et al. merekomendasikan penggunaan lters fi dengan
dan tetap layak [67]. Basil tetap di udara kecuali aliran gas melalui kinerja filtrasi setara dengan respirator facemask N100, yaitu 99,97%
sistem pernapasan berhenti selama lebih dari 1 jam. efisiensi [75]. Seperti fi lter ditempatkan di antara pasien dan Y-piece
dari sistem pernapasan harus memiliki output kelembaban minimal 32
gm) 3. Atau, sebuah HME dapat digunakan di Y-piece dan N100-setara
Lipit lters fi hidrofobik mencegah penularan Mycobacterium bovis lters fi ditempatkan pada inspirasi dan ekspirasi anggota badan dari
(pengganti tes untuk Mycobacterium tuberculosis) [ 68] dan Mycobacterium sistem pernapasan pada ventilator.
chelonae [ 69]. Dua jenis elektrostatik fi lter memiliki filtrasi ef defisiensi
dari> 99,999% ketika ditantang dengan Mycobacterium chelonae [69].
Dalam sebuah editorial, Kamming et al. menekankan pentingnya
perencanaan ke depan ketika intubasi trakea pasien dengan SARS
Pusat Pengendalian dan Pencegahan Penyakit (CDC) diperlukan [76], termasuk penggunaan tinggi-efisiensi hidrofobik fi lter
merekomendasikan menempatkan fi lter bakteri baik pada tabung trakea antara sungkup muka dan pernapasan sistem atau antara sungkup
atau ekstremitas ekspirasi dari sistem pernapasan [70]. Ketika memilih fi muka dan self-dalam tas Ating fl. Di teater operasi, Kamming et al.
lter, CDC merekomendasikan

2010 Penulis The


44 Anestesi 2010 Asosiasi Dokter-dokter anestesi dari Britania Raya dan Irlandia
Anestesi, 2011, 66, halaman 40-51 AR Wilkes Æ HMEs di anestesi dan perawatan intensif - penggunaan praktis
.....................................................................................................................................................................................................................

direkomendasikan menempatkan defisiensi bakteri tinggi-ef / virus 'Lters fi lipit dari benar spesifik kasi' pada kateter gunung dan port
hidrofobik fi lter di sirkuit ekspirasi dari ventilator. pernafasan.
Ramsey et al. direkomendasikan bahwa manajemen ventilasi
Peng et al. melaporkan praktek mereka untuk mencegah penyebaran kegagalan pernafasan hypoxaemic disebabkan infeksi virus H1N1
SARS ketika berhadapan dengan pasien yang terinfeksi dalam anestesi harus didasarkan pada sindrom gangguan pernapasan akut (ARDS)
[73]. Tinggi efisiensi, volume rendah, lters fi hidrofobik yang digunakan protokol jaringan, dengan volume tidal target 6 ml.kg) 1 [ 81]. Ini karena
pada kedua inspirasi dan anggota badan ekspirasi dari mesin anestesi. itu memiliki efek pada HMEs dan lters fi yang dapat digunakan dengan
lters fi elektrostatik tidak dianggap sebagai, menurut Peng et al., mereka pasien-pasien ini, dalam perangkat yang dengan ruang mati besar akan
memungkinkan lewatnya virus saat basah. tidak cocok.

Dalam influenza fl (H1N1 dan H5N1) cedera akut paru-paru dan ARDS, dan protokol jaringan ARDS
Avian fl u (H5N1) dan babi fl u (H1N1) adalah kedua jenis influenza A
virus yang telah menyebar dengan cepat di seluruh dunia dalam Pasien yang menerima ventilasi mekanis berkepanjangan rentan terhadap
beberapa tahun terakhir. Beberapa pasien tertular fl burung atau babi u cedera paru akut (ALI). Ini dapat berkembang lebih lanjut ke ARDS.
telah memerlukan perawatan di ICU. Modus penularan virus ini Penggunaan volume tidal besar selama ventilasi mekanik secara signifikan
umumnya dianggap oleh droplet pernapasan dari orang ke orang. meningkatkan risiko mengembangkan ALI [82] selain meningkatkan angka
kematian pada mereka yang sudah menderita ALI atau ARDS [83, 84].
Penggunaan volume tidal yang lebih kecil memiliki implikasi untuk
petugas kesehatan dan pasien lainnya dalam satu meter dari pasien penggunaan HMEs dan lters fi, yang, ketika terhubung antara pasien dan
yang terinfeksi berada pada risiko kontaminasi karena pelepasan tetesan sistem pernapasan, menambah ruang mati dari sistem pernapasan. Jika
yang terinfeksi [77]. Penggunaan osilasi ventilasi frekuensi tinggi (HFOV) volume tidal berkurang, sebagian kecil dari ruang mati disebabkan HME
dapat memperoleh keuntungan beberapa pasien, tapi sirkuit HFOV atau fi lter meningkat dan dapat menjadi tidak dapat diterima.
standar tanpa fi lter virus dapat menjadi bahaya infeksi [77]. Dengan
demikian, sirkuit HFOV dengan fi lter virus lebih disukai.
The ARDSNet ventilasi mekanik Ringkasan protokol
Dalam laporan khusus tentang epidemiologi, pencegahan dan implikasi (http://www.ardsnet.org) merekomendasikan pengaturan ventilator
untuk mencapai volume tidal awal 8 ml.kg) 1
untuk anestesi dari fl burung u, Edler et al. melaporkan bahwa, ketika
anestesi umum diperlukan untuk pasien dengan fl burung u, lters fi diprediksi berat badan dan kemudian mengurangi volume tidal oleh 1
hidrofobik ditempatkan pada kedua inspirasi dan anggota badan ekspirasi ml.kg) 1 pada interval kurang dari 2 jam sampai volume tidal adalah 6
dari sistem pernapasan untuk mengurangi penyebaran tetesan [78]. ml.kg) 1 diprediksi berat badan. Tekanan puncak dataran tinggi harus
nomore dari 30 cmH 2 O. Jika lebih besar dari ini, maka volume tidal
Daugherty et al., Dalam sebuah artikel di persiapan untuk menanggapi harus dikurangi, untuk minimal 4 ml.kg) 1. Selama penyapihan dari
pasien dengan virus H1N1, menyarankan bahwa bukti lemah pada ventilasi mekanik, tujuan untuk volume tidal spontan setidaknya 4
penggunaan lters fi di ventilator, tetapi mengakui bahwa lters fi harus ml.kg) 1.
digunakan pada port inlet ventilator yang menarik di kamar udara, dan
bahwa panas dan kelembaban penukar dengan lters fi dapat mengurangi Sebagai contoh, 'standar' pasien laki-laki dengan berat badan prediksi 70
pencemaran lingkungan [79]. Namun, ini adalah pada peningkatan risiko kg akan memiliki volume tidal tidak lebih dari 420 ml selama ventilasi
dari trakea tabung oklusi dan meningkatkan ketahanan terhadap gas fl ow. mekanik. Jika tekanan puncak dataran tinggi atau selama penyapihan dari
ventilasi mekanik, volume tidal bisa serendah 280 ml. Beberapa HMEs dan
lters fi memiliki volume internal 100 ml atau lebih [85]. Ini lebih dari
Rekomendasi telah diterbitkan baru-baru ini oleh European Society sepertiga dari volume tidal pasien dalam kasus ini. Penggunaan jenis
of Task Force Intensif CareMedicine untuk ICU triase untuk digunakan perangkat akan meningkatkan rebreathing, sehingga meningkatkan P aCO2 [ 86,
selama epidemi influenza fl atau bencana massal [80]. Rekomendasi 87]. Jika HMEs atau HMEFs digunakan, volume internal mereka harus
termasuk penggunaan 'bakteri berkualitas tinggi / virus fi lter 'pada port sekecil praktis, meskipun output kelembaban tinggi masih diperlukan. Salah
ekspirasi dari ventilator, sebuah' berkualitas tinggi bakteri / panas virus satu metode untuk mengekspresikan ini adalah untuk membagi output
dan penukar kelembaban dan fi lter' pada trakea atau trakeostomi kelembaban oleh volume internal, untuk memperoleh output kelembaban
tabung, dan bakteri yang / virus fi lter melekat ke port ekspirasi per ml volume internal yang [88, 89]. Panas dan kelembaban penukar
perangkat ventilasi kantong-mask dengan yang lain fi lter antara masker cenderung memiliki output kelembaban yang lebih tinggi per ml volume
dan katup. Penggunaan dipanaskan ers fi humidi pada ventilator harus internal yang dari lters fi [88, 89].
dihindari. Selanjutnya, rekomendasi termasuk penggunaan

2010 Penulis Anestesi 2010 Asosiasi Dokter-dokter anestesi dari Britania Raya dan Irlandia
45
A. R. Wilkes Æ HMEs in anaesthesia and intensive care - practical use Anaesthesia, 2011, 66, pages 40–51
.....................................................................................................................................................................................................................

hepatitis C The AAGBI reiterated its advice on the use of a breathing system
In 1994, Chant et al. described an incident of suspected cross-infection filter for each patient in guidelines published in 2002 [94] and updated
of hepatitis C virus (HCV) during a minor surgical list in New South in 2008 [95]. In 2002, it noted that ‘although it appears that pleated
Wales, Australia [90]. Two patients, who had been operated on during hydrophobic filters have a better filtration performance than most
the list, were subsequently diagnosed with hepatitis. Neither patient electrostatic filters, the clinical relevance of this has yet to be
was at a high risk of developing the disease. During further established’.
investigation, it was found that three other patients out of a total of 13
patients on the same list had also developed hepatitis C, with all five In its publication, ‘Checking anaesthetic equipment’, in 2003 the
patients having the same genotype (1a) of hepatitis C. This cluster of AAGBI reiterated its recommendation that ‘A new, single-use bacterial ⁄ viral
patients was extremely unlikely to have occurred by chance. All five had filter and angle piece ⁄
shared the same breathing system, as it was common practice to only catheter mount must be used for each patient’, although the type or
change the breathing system after each list. Breathing system filters performance was not specified [96].
had not been used. It was hypothesised that the first patient had More recently, the German Society of Hospital Hygiene and the
coughed infected secretions into the breathing system which then acted German Society for Anaesthesiology and Intensive Care have issued a
as a reservoir of infection for the remaining four patients. The two joint recommendation on the use of filters to prevent infection during
patients presented with acute hepatitis C five and seven weeks ventilation in anaesthesia [97]. If a filter is used, the breathing system
following surgery. can be used for up to 7 days provided that it is checked and is
functioning as intended and that the manufacturer has declared this in
the instructions for use. If a filter is used, its filtration performance for
airborne particles, measured using particles of themost penetrating
A second incident of cross-infection of hepatitis C was reported in size, should be better than 99%, and it should be able to withstand an
2000 [91]. In this case, a patient presented with clinical symptoms of applied pressure of 60 hectopascals ( 60 cmH 2 O) without allowing liquid
hepatitis seven weeks after surgery. A previous patient on the same list to pass through, or 20 hectopascals above the set pressure limit in the
was positive for HCV antibody and positive for HCV polymerase chain breathing system.
reaction. Again, no filters were used for these patients.

The case described by Chant et al. highlights the importance of


protecting the patient from a contaminated breathing system, either by
Use of HMEs and filters with paediatric patients
using a new breathing system for each patient, or by placing an
appropriate filter between the patient and the breathing system. Provision of humidification and filtration of gases delivered to paediatric
patients using HMEFs is controversial. This patient group is particularly
In addition, these cases highlight the risk of crossinfection from the susceptible to increases in dead space and resistance to gas flow.
transmission of liquid-borne microbes present in sputum particularly, Furthermore, children are more likely to be susceptible to infection and
blood-stained. Hence, breathing system filters should also prevent the less able to cope with reduced levels of humidity. In addition, the ability
transmission of liquid from the patient to the breathing system and vice of filters intended for paediatric use to prevent the transmission of
versa. microbes effectively has been questioned. However, there is evidence
that HMEs and HMEFs can be used successfully with this patient
Following the suspected cross-infection of hepatitis C described by group.
Chant et al., the Blood Borne Viruses Advisory Panel of the Association
of Anaesthetists of Great Britain and Ireland (AAGBI) published The incidence of nosocomial pneumonia in children is about 22% in
guidelines in 1996 [92], recommending that ‘either an appropriate filter paediatric ICUs [98]. In intensive care, as with adult patients, children
should be placed between the patient and the breathing system, a new who succumb to an incident VAP have a greater need for mechanical
filter being used for each patient, or that a new breathing system be ventilation (12 vs 22 ventilator-free days), longer ICU stay (6 vs 13
used for each patient.’ median ICU-free days), higher total median hospital costs ($308 534 vs
$252 652) and increased absolute hospital mortality (10.5% vs 2.4%)
It was also noted that ‘of the filters currently available, only those than those without VAP [99]. However, the effect of the use of filters on
which use a pleated, hydrophobic membrane reliably prevent the incidence of VAP in this group of patients has not been studied.
contamination of the breathing system’.
Pleated hydrophobic filters prevent the transmission of HCV in vitro
[93]. Electrostatic filters do not prevent transmission. This is caused by Use of ineffectively reprocessed breathing systems can lead to an
the ability of liquid to pass through electrostatic filters when pressure is increase in morbidity and mortality. In one reported this may outbreak of
applied, which would take viruses (and bacteria) through as well. Bacillus cereus respiratory tract infections associated with the
breakdown of a low

2010 The Author


46 Anaesthesia 2010 The Association of Anaesthetists of Great Britain and Ireland
Anaesthesia, 2011, 66, pages 40–51 A. R. Wilkes Æ HMEs in anaesthesia and intensive care - practical use
.....................................................................................................................................................................................................................

temperature steam disinfector used to reprocess breathing systems, six with paediatrics. Six filters, all electrostatic with an internal volume of 10
of thirteen neonates became infected, three of whom died [100]. to 16 ml, were suitable for use with small infants. The penetration of
particles through adult and paediatric filters was measured using a flow
Breathing system filters are available for use with children weighing of air of 30 and 15 l.min) 1, respectively. However, the Association of
at least 3 kg to reduce the transmission of microbes [101]. Luchetti et Paediatric Anaesthetists considered that a flow of 15 l.min) 1 was too
al. demonstrated that two small electrostatic HMEFs, the Hygrobaby high for filters intended for use with small infants [117]. Further work
and Hygroboy, provided humidified gases to children weighing up to 10 has now demonstrated that penetration of particles through filters
kg and more than 10 kg, respectively, with moisture contents of more intended for use with small infants can be substantially lower (and
than 20 g.m) 3 throughout the operation, increasing up to 26 and 27 g.m) 3, hence the filtration efficiency is higher) when the flow is reduced [118,
119].

respectively, by the end of the study period [102]. The moisture content
of gases delivered to patients using a pleated filter-only product in a
different study with older and larger children (mean (SD) 48 (20)
Conclusions
months; 16 (3.5 kg) was about 22 g.m) 3 [ 103]. Heat and moisture
exchanges also provide adequate humidification to children during Patients can acquire infections fromequipment used during anaesthesia
anaesthesia [104, 105]. and intensive care. Breathing systems can be a source of contamination.
The use of filters is recommended to reduce the risk of cross-infection
With very small children, the practice has been to use uncuffed during anaesthesia if the breathing system is used for more than one
tracheal tubes. This means that some of the moisture in the expired gas patient.
does not reach the HME because of the leak around the tracheal tube, During anaesthesia with circle breathing systems, where
and is therefore not available to humidify the gas delivered to the condensation can occur, the use of electrostatic filters might increase
patient. This causes a reduction in the performance of HMEs [106] the risk of blockage because of ingress and absorption of water.
although this may still be adequate [107]. However, in a large However, this risk can be reduced by placing the filter above the level of
randomised controlled trial comparing cuffed and uncuffed tracheal the patient and the breathing system so that any liquid flows away from
tubes, the use of cuffed tracheal tubes did not cause an increase in the filter. Condensation is easier to detect visually in pleated filters and
stridor and hence, by implication, did not cause an increase in trauma these filters also prevent the transmission of water under normal
[108]. The routine use of cuffed tracheal tubes would improve the ventilatory pressures. However, viscous sputum and nebulised drugs
moisture-conserving performance of HMEs. can block these filters. It is important that users are aware of this
scenario and that the high pressures found in such situations are not
misdiagnosed.
The performance of HMEs has been compared with heated
humidifiers in intensive care. Fassassi et al. demonstrated that HMEs
can provide moisture outputs of at least 28 g.m) 3 [ 109]; Schiffmann et al. The use of HMEs or filters, rather than heated humidifiers, does not
measured moisture outputs of 34 g.m) 3 [ 110]. reduce the incidence of ventilatorassociated pneumonia. However, the
use of filters with patients known to have particularly infectious diseases
It is also important to consider the addition of dead space [111] and is recommended to prevent cross-infection, although there is no
resistance to gas flow when using HMEs or filters in children. The lower evidence that the incidence of cross-infection is reduced by using filters.
weight limit of patients suitable for use with HMEs is probably 2.5 kg, As an adequate moisture output is essential for patients receiving
and 3 kg for filters [112]. The work of breathing increases when HMEFs prolonged mechanical ventilation, a useful performance indicator for
are used [113]. The resistance to gas flow of an HMEF added to a filters and HMEs is the moisture output per ml of internal volume.
Mapleson F breathing system can cause a delay in the induction of
anaesthesia [114]. The addition of an HME can cause hypercapnia if
the internal volume of the HME is too large [115]. Loading an HME with
water can cause an excessive increase in resistance to gas flow [116].
Competing interests

This review was based on a project carried out for Covidien plc,
funded though a contract between
The Medicines and Healthcare products Regulatory Agency Covidien and Cardiff University. AW has received conference expenses
published a report in 2004 on the filtration performance of 104 different from Pall Medical, and Cardiff University has received payment from
breathing system filters available on the UK market [85]. Thirty-four of Pall Medical for AW’s speaking at a meeting. AW did not gain financially
the filters were intended for, or were an appropriate size for, use from either arrangement.

2010 The Author Anaesthesia 2010 The Association of Anaesthetists of Great Britain and Ireland
47
A. R. Wilkes Æ HMEs in anaesthesia and intensive care - practical use Anaesthesia, 2011, 66, pages 40–51
.....................................................................................................................................................................................................................

References 14 Williams DJ, Stacey MRW. Rapid and complete occlusion


of a heat and moisture exchange filter by pulmonary edema (clinical report). Canadian
1 Wilkes AR. Heat and moisture exchangers and breathing
Journal of Anesthesia 2002; 49:
system filters: their use in anaesthesia and intensive care. Part 1 – History,
126–31.
principles and efficiency. Anaesthesia 2011;
15 Milligan KA. Disablement of a ventilator disconnect alarm
66: 31–39.
by a heat and moisture exchanger. Anaesthesia 1992; 47:
2 British Standards Institution. Anaesthetic and Respiratory
279.
Equipment – Heat and Moisture Exchangers (HMEs) for Humidifying
16 Garnerin P, Schiffer E, Van Gessel E, Clergue F.
Respired Gases in Humans - Part 1: HMEs for Use with Minimum Tidal
Root-cause analysis of an airway filter occlusion: a way to improve the
Volumes of 250 ml (BS EN ISO 93601:2009). London: British Standards
reliability of the respiratory circuit. British Journal of Anaesthesia 2002; 89: 633–5.
Institution, 2009. 3 British Standards Institution. Breathing System Filters for
17 Morgan-Hughes NJ, Mills GH, Northwood D. Air

Anaesthetic and Respiratory Use – Part 1: Salt Test Method to Assess


flow resistance of three heat and moisture exchanging filter designs
Filtration Performance (BS EN ISO 23328-1:2008).
under wet conditions: implications for patient safety. British Journal of
London: British Standards Institution, 2008. 4 Wilkes AR. Breathing
Anaesthesia 2001; 87:
filters, humidifiers and nebulizers.
289–91.
In: Davey AJ, Diba A eds. Ward’s Anaesthetic Equipment
18 Turnbull D, Fisher PC, Mills GH, Morgan-Hughes NJ.
5th edn. London: WB Saunders 2005: 291–303. 5 Lellouche F,
Performance of breathing filters under wet conditions: a laboratory
Maggiore SM, Deye N, et al. Effect of the
evaluation. British Journal of Anaesthesia 2005; 94:
humidification device on the work of breathing during noninvasive
675–82.
ventilation. Intensive Care Medicine 2002; 28:
19 Wilkes AR. The ability of breathing system filters to
1582–9.
prevent liquid contamination of breathing systems: a laboratory study. Anaesthesia
6 Iotti GA, Olivei MC, Palo A, et al. Unfavorable
2002; 57: 33–9. 20 Kopman AF, Glaser L. Obstruction of bacterial filters by
mechanical effects of heat and moisture exchangers in ventilated
patients. Intensive Care Medicine 1997; 23:
edema fluid. Anesthesiology 1976; 44: 169–70. 21 Mason J,
399–405.
Tackley R. An acute rise in expiratory
7 Chiaranda M, Verona L, Pinamonti O, Dominioni L,
resistance due to a blocked ventilator filter. Anaesthesia
Minoja G, Conti G. Use of heat and moisture exchanging (HME) filters in
1981; 36: 335.
mechanically ventilated ICU patients: influence on airway flow-resistance. Intensive
22 Buckley PM. Increase in resistance of in-line breathing
Care Medicine
filters in humidified air. British Journal of Anaesthesia 1984;
1993; 19: 462–6.
56: 637–43.
8 Conti G, De Blasi RA, Rocco M, et al. Effects of
23 McEwan AI, Dowell L, Karis JH. Bilateral tension
heat-moisture exchangers on dynamic hyperinflation of mechanically
pneumothorax caused by a blocked bacterial filter in an anesthesia
ventilated COPD patients. Intensive Care Medicine 1990; 16: 441–3.
breathing circuit. Anesthesia and Analgesia 1993;
76: 440–2.
9 Jaber S, Chanques G, Matecki S, et al. Comparison of the
24 Stacey MRW, Asai T, Wilkes A, Hodzovic I. Obstruction
effects of heat and moisture exchangers and heated humidifiers on
of a breathing system filter. Canadian Journal of Anaesthesia
ventilation and gas exchange during non-invasive ventilation. Intensive
1996; 43: 1276.
Care Medicine 2002; 28:
25 Costigan SN, Snowdon SL. Breathing system filters can
1590–4.
affect the performance of anaesthetic monitors. Anesthesia
10 Le Bourdelle`s G, Mier L, Fiquet B, et al. Comparison of the
1993; 48: 1015–6.
effects of heat and moisture exchangers and heated humidifiers on
26 Puolakka JJ, Jousela IT. The effect of a heat and moisture
ventilation and gas exchange during weaning trials from mechanical
exchange filter on sidestream spirometry in critically ill patients. International
ventilation. Chest 1996;
Journal of Clinical Monitoring and Computing 1994; 11: 217–22. 27
110: 1294–8.
Hardman JG, Curran J, Mahajan RP. End-tidal carbon
11 Pelosi P, Soica M, Ravagnan I, Tubiolo d, Ferrsio L,
Gattinnoni L. Effects of heat and moisture exchangers on minute
dioxide measurement and breathing system filters.
ventilation, ventilatory drive, and work of breathing during pressure-support
Anaesthesia 1997; 52: 646–8. 28 Hardman JG, Mahajan RP, Curran J.
ventilation in acute respiratory failure. Critical Care Medicine 1996; 24: 1184–8.
The influence of
12 Campbell RS, Davis K Jr, Johannigman JA, Branson RD.
breathing system filters on paediatric capnography. Pediatric Anesthesia 1999; 9:
35–8.
The effects of passive humidifier dead space on respiratory variables in
29 Fujita Y, Imanaka H, Fujino Y, et al. Effect of humidifying
paralyzed and spontaneously breathing patients.
devices on the measurement of tidal volume by mechanical ventilators. Journal
Respiratory Care 2000; 45: 306–12.
of Anesthesia 2006; 20: 166–72. 30 Poonawala Y, Wilkes AR, Cann C, Hall
13 Schummer W, Schummer C, Fuchs J, Voigt R. Sudden
JE. The effect of
upper airway occlusion due to invisible rain-out in the heat and moisture
desflurane on filtration performance of breathing system filters. Anaesthesia
exchanger. British Journal of Anaesthesia 2002;
2010; 65: 831–5.
89: 335–6.

2010 The Author


48 Anaesthesia 2010 The Association of Anaesthetists of Great Britain and Ireland
Anaesthesia, 2011, 66, pages 40–51 A. R. Wilkes Æ HMEs in anaesthesia and intensive care - practical use
.....................................................................................................................................................................................................................

31 Froom SR, Cann C, Wilkes AR, Gildersleve C, Hall JE. 46 Branson RD, Davis K Jr, Brown R, Rashkin M.
The effect of volatile anaesthetic agents on the filtration performance of Comparison of three humidification techniques during mechanical
paediatric breathing system filters. ventilation: patient selection, cost, and infection consideration. Respiratory
Anaesthesia 2008; 63: 77–81. Care 1996; 41: 809–16. 47 Boots RJ, Howe S, George N, Harris FM,
32 Lawes EG. Hidden hazards and dangers associated with the Faoagali J.
use of HME ⁄ filters in breathing circuits. Their effect on toxic metabolite Clinical utility of hygroscopic heat and moisture exchangers in intensive care
production, pulse oximetry and airway resistance. British Journal of patients. Critical Care Medicine 1997; 25:
Anaesthesia 2003; 91: 249–64. 33 Chastre J, Fagon J-Y. 1707–13.
Ventilator-associated pneumonia. 48 Hurni JM, Feihl F, Lazor R, Leuenberger P, Perret C.
American Journal of Respiratory and Critical Care Medicine Safety of combined heat and moisture exchanger filters in long-term
2002; 165: 867–903. mechanical ventilation. Chest 1997; 111:
34 Rello J, Ollendorf DA, Oster G, et al. Epidemiology and 686–91.
outcomes of ventilator-associated pneumonia in a large US database. Chest 2002; 49 Kirton OC, DeHaven B, Morgan J, Morejon O, Civetta J.
122: 2115–21. A prospective, randomized comparison of an in-line heat moisture
35 Safdar N, Dezfulian C, Collard HR, Saint S. Clinical and exchange filter and heated wire humidifiers: rates of ventilator-associated
economic consequences of ventilator-associated pneumonia: a systematic early-onset (communityacquired) or late-onset (hospital-acquired)
review. Critical Care Medicine 2005; 33: pneumonia and incidence of endotracheal tube occlusion. Chest 1997; 112:
2184–93.
36 Gallagher I, Strangeways JEM, Allt-Graham J. Contami- 1055–9.
nation control in long-term ventilation. Anaesthesia 1987; 50 Kollef MH, Shapiro SD, Boyd V, et al. A randomized
42: 476–81. clinical trial comparing an extended-use hygroscopic condenser
37 Cook D, De Jonghe B, Brochard L, Brun-Buisson C. humidifier with heated-water humidification in mechanically ventilated
Influence of airway management on ventilator-associated pneumonia. patients. Chest 1998; 113:
Evidence from randomized trials. Journal of the American Medical 759–67.
Association 1998; 279: 781–7. 38 Hess DR, Kallstrom TJ, Mottram CD, Myers 51 Memish ZA, Oni GA, Djazmati W, Cunningham G, Mah
TR, MW. A randomized clinical trial to compare the effects of a heat and
Sorensen HM, Vines DL. Care of the ventilator circuit and its relation to moisture exchanger with a heated humidifying system on the occurrence
ventilator-associated pneumonia. Respiratory Care 2003; 48: 869–79. rate of ventilator-associated pneumonia. American Journal of Infection
Control 2001; 29:
39 Kola A, Eckmanns T, Gastmeier P. Efficacy of heat and 301–5.
moisture exchangers in preventing ventilator-associated pneumonia: 52 Lacherade JC, Auburtin M, Cerf C, et al. Impact of
meta-analysis of randomized controlled trials. humidification systems on ventilator-associated pneumonia: a randomized
Intensive Care Medicine 2005; 31: 5–11. 40 Siempos II, Vardakas KZ, multicenter trial. American Journal of Respiratory and Critical Care Medicine 2005;
Kopterides P, Falagas ME. 172: 1276–82. 53 Lorente L, Lecuona M, Jime´nez A, Mora ML, Sierra A.
Impact of passive humidification on clinical outcomes of mechanically
ventilated patients: a meta-analysis of randomized controlled trials. Critical Ventilator-associated pneumonia using a heated humidifier or a heat and
Care Medicine 2007; 35: moisture exchanger: a randomized controlled trial. Critical Care 2006; 10: R116.
2843–51. 54 Boots RJ, George N, Faoagali JL, Druery J, Dean K, Heller
41 Collard HR, Saint S, Matthay MA. Prevention of ventila-
tor-associated pneumonia: an evidence-based systematic review. Annals of RF. Double-heater-wire circuits and heat-and-moisture exchangers and the
Internal Medicine 2003; 138: 494–501. 42 Dodek P, Keenan S, Cook D, et al. risk of ventilator-associated pneumonia.
Evidence-based clinical Critical Care Medicine 2006; 34: 687–93. 55 Kranabetter R, Leier M,
practice guideline for the prevention of ventilatorassociated pneumonia. Annals Kammermeier D, Just HM,
of Internal Medicine 2004; 141: Heuser D. The effects of active and passive humidification on
305–13. ventilation-associated nosocomial pneumonia.
43 Martin C, Perrin G, Gevaudan M-J, Saux P, Gouin F. Heat Anaesthesist 2004; 53: 29–35.
and moisture exchangers and vaporizing humidifiers in the intensive care 56 Tablan OC, Anderson LJ, Besser R, Bridges C, Hajjeh R.
unit. Chest 1990; 97: 144–9. 44 Roustan JP, Kienlen J, Aubus P, Aubus S, du Guidelines for Preventing Health-care-associated Pneumonia,
Cailar J. 2003. Recommendations of CDC and the Healthcare Infection Control
Comparison of hydrophobic heat and moisture exchangers with heated Practices Advisory Committee. Morbidity and Mortality Weekly Report
humidifier during prolonged mechanical ventilation. Intensive Care Medicine 1992; 2004; http:// www.cdc.gov/hicpac/pdf/guidelines/HApneu2003guide
18: 97–100. 45 Dreyfuss D, Djedaı lines.pdf, accessed 14 ⁄ 04 ⁄ 2010.
¨ni K, Gros I, et al. Mechanical ventila-
tion with heated humidifiers or heat and moisture exchangers: effects on 57 American Thoracic Society. Guidelines for the manage-
patient colonization and incidence of nosocomial pneumonia. American ment of adults with hospital-acquired, ventilator-associated, and
Journal of Respiratory Care Medicine 1995; 151: 986–92. healthcare-associated pneumonia. American Journal of Respiratory and
Critical Care Medicine 2005; 171: 388–416.

2010 The Author Anaesthesia 2010 The Association of Anaesthetists of Great Britain and Ireland
49
A. R. Wilkes Æ HMEs in anaesthesia and intensive care - practical use Anaesthesia, 2011, 66, pages 40–51
.....................................................................................................................................................................................................................

58 Rello J, Lode H, Cornaglia G, Masterton R, The VAP 73 Peng PW, Wong DT, Bevan D, Gardam M. Infection
Care Bundle Contributors. A European care bundle for prevention of control and anesthesia: lessons learned from the Toronto SARS outbreak. Canadian
ventilator-associated pneumonia. Intensive Care Medicine 2010; 36: 773–80. Journal of Anesthesia 2003; 50:
989–97.
59 Torres A, Ewig S, Lode H, Carlet J. Defining, treating and 74 ECRI. Mechanical ventilation of SARS patients. Safety
preventing hospital acquired pneumonia: European perspective. Intensive issues involving breathing-circuit filters. Health Devices
Care Medicine 2009; 35: 9–29. 60 Masterton RG, Galloway A, French G, et al. 2003; 32: 220–2.
Guidelines for 75 Thiessen RJ. The impact of severe acute respiratory
the management of hospital-acquired pneumonia in the UK: Report of the syndrome on the use of and requirements for filters inCanada.
Working Party on Hospital-Acquired Pneumonia of the British Society for Respiratory Care Clinics of North America 2006; 12: 287–306. 76 Kamming D,
Antimicrobial Chemotherapy. Journal of Antimicrobial Chemotherapy 2008; Gardam M, Chung F. Anaesthesia and SARS.
British Journal of Anaesthesia 2003; 90: 715–8. 77 Patel M, Dennis A,
62: 5–34. Flutter C, Khan Z. Pandemic (H1N1)
61 Coffin SE, Klompas M, Classen D, et al. Strategies to 2009 influenza. British Journal of Anaesthesia 2010; 104:
prevent ventilator-associated pneumonia in acute care hospitals. Infection 128–42.
Control and Hospital Epidemiology 2008; 78 Edler AA. Avian flu (H5N1): its epidemiology, prevention,
1: S31–40. and implications for anesthesiology. Journal of Clinical Anesthesia 2006;
62 Muscedere J, Dodek P, Keenan S, et al. Comprehensive 18: 1–4.
evidence-based clinical practice guidelines for ventilatorassociated 79 Daugherty EL, Branson RD, Deveraux A, Rubinson L.
pneumonia: prevention. Journal of Critical Care Infection control in mass respiratory failure: preparing to respond to H1N1. Critical
2008; 23: 126–37. Care Medicine 2010; 38: e103–9. 80 Sprung CL, Zimmerman JL, Christian
63 World Health Organization (WHO). Tuberculosis Fact Sheet MD, et al.
No 104. Geneva: WHO, March 2010. 64 Ryan KJ. Sherris Medical Recommendations for intensive care unit and hospital preparations for an
Microbiology: An Introduction to influenza epidemic or mass disaster: summary report of the European
Infectious Diseases, 3rd edn. Norwalk, CT: Appleton and Lange, 1994. Society of Intensive Care Medicine’s Task Force for intensive care unit
triage during an influenza epidemic or mass disaster. Intensive Care
65 Jereb JA, Burwen DR, Dooley SW, et al. Nosocomial Medicine 2010; 36: 428–43.
outbreak of tuberculosis in a renal transplant unit: application of a new
technique for restriction fragment length polymorphism analysis of Mycobaterium 81 Ramsey CD, Funk D, Miller RR III, Kumar A. Ventilator
tuberculosis isolates. Journal of Infectious Diseases 1993; 168: management for hypoxemic respiratory failure attributable to H1N1 novel
swine origin influenza virus. Critical Care Medicine 2010; 38: e58–65. 82
1219–24. Gajic O, Frutos-Vivar F, Esteban A, Hubmayr RD,
66 Teo DTW, Lim TW. Transmission of tuberculosis from
patient to healthcare workers in the anaesthesia context. Anzueto A. Ventilator settings as a risk factor for acute respiratory
Annals of the Academy of Medicine Singapore 2004; 33: 95–9. 67 Langevin distress syndrome in mechanically ventilated patients. Intensive Care
PB, Kenneth HR, Layon AJ. The potential for Medicine 2005; 31: 922–6. 83 Amato MB, Barbas CS, Medeiros DM, et al.
dissemination of Mycobacterium tuberculosis through the anesthesia Effect of a
breathing circuit. Chest 1999; 115: 1107–14. 68 Aranha-Creado H, Prince protective-ventilation strategy on mortality in the acute respiratory
D, Greene K, Brandwein H. distress syndrome. New England Journal of Medicine 1998; 338: 347–54.
Removal of Mycobacterium species by breathing circuit filters. Infection
Control and Hospital Epidemiology 1997; 18: 84 Brower RG, Matthay MA, Morris A, et al. Ventilation with
252–4. lower tidal volumes as compared with traditional tidal volumes for acute lung
69 Vezina DP, Tre´panier CA, Lessard MR, Gourdeau M, injury and the acute respiratory distress syndrome. New England Journal of
Tremblay C, Guidoin R. An in vivo evaluation of the mycobacterial Medicine 2000; 342:
filtration efficacy of three breathing filters used in anesthesia. Anesthesiology 1301–8.
2004; 101: 104–9. 70 Centers for Disease Control and Prevention (CDC). 85 Medicines and Healthcare products Regulatory Agency.
Breathing system filters. An assessment of 104 different breathing
Guidelines for preventing the transmission of Mycobacterium tuberculosis system filters. Evaluation 2004; 04005: 1–52. London: Medicines and
in health-care settings, 2005. Morbidity and Mortality Weekly Report 2005; 54( Healthcare products Regulatory Agency.
RR-17): 1–141. 71 Wilkes AR. The effect of gas flow on the filtration
86 Prin S, Chergui K, Augarde R, Page B, Jardin F,
performance of breathing system filters. Anaesthesia 2004; Vieillard-Baron A. Ability and safety of a heated humidifier to control
59: 278–82. hypercapnic acidosis in severe ARDS. Intensive Care Medicine 2002; 28: 1756–60.
72 World Health Organization. Summary Table of SARS Cases 87 Branson RD. Humidification of respired gases during
by Country, 1 November 2002–7 August 2003. Available at:
http://www.who.int/csr/sars/country/country2003_08_ mechanical ventilation: mechanical considerations. Respiratory Care Clinics
15.pdf, accessed 22 ⁄ 04 ⁄ 2010. of North America 2006; 12: 253–61.

2010 The Author


50 Anaesthesia 2010 The Association of Anaesthetists of Great Britain and Ireland
Anaesthesia, 2011, 66, pages 40–51 A. R. Wilkes Æ HMEs in anaesthesia and intensive care - practical use
.....................................................................................................................................................................................................................

88 Lellouche F, Taille´ S, Lefranc¸ois F, et al., Groupe de travail 103 Monrigal JP, Granry JC. The benefit of using a heat and
sur les Respirateurs de l’AP-HP Humidification performance of 48 passive moisture exchanger during short operations in young children. Pediatric
airway humidifiers: comparison with manufacturer data. Chest 2009; 135: 276–86. Anesthesia 1997; 7: 295–300. 104 Bissonnette B, Sessler DI, LaFlamme P.
89 Branson RD, Davis K Jr. Evaluation of 21 passive humid- Passive and active
inspired gas humidification in infants and children.
ifiers according to the ISO 9360 standard: moisture output, dead space and Anesthesiology 1989; 71: 350–4. 105 Bissonnette B, Sessler DI. Passive
flow resistance. Respiratory Care 1996; 41: or active inspired gas
736–43. humidification increases thermal steady-state temperatures in anesthetized
90 Chant K, Kociuba K, Munro R, et al. Investigation of infants. Anesthesia andAnalgesia 1989; 69: 783–7. 106 Tilling SE, Hayes B. Heat
possible patient-to-patient transmission of Hepatitis C in a hospital. New and moisture exchangers in
South Wales Public Health Bulletin 1994; 5: artificial ventilation. An experimental study of the effect of gas leakage. British
47–51. Journal of Anaesthesia 1987; 59: 1181–8. 107 Gedeon A, Mebius C, Palmer K.
91 Heinsen A, Bendtsen F, Fomsgaard A. A phylogenetic Neonatal hygroscopic
analysis elucidating a case of patient-to-patient transmission of hepatitis C condenser humidifier. Critical Care Medicine 1987; 15: 51–4. 108 Weiss M,
virus during surgery. Journal of Hospital Infection 2000; 46: 309–13. 92 Blood Dullenkopf A, Fischer JE, Keller C, Gerber AC,
Borne Viruses Advisory Panel. A Report Received by European Paediatric Endotracheal Intubation Study Group. Prospective
randomized controlled multi-centre trial of cuffed or uncuffed endotracheal
Council of the Association of Anaesthetists on Blood Borne Viruses and tubes in small children.
Anaesthesia: An Update (January 1996). London: The Association of British Journal of Anaesthesia 2009; 103: 867–73. 109 Fassassi M, Michel
Anaesthetists of Great Britain and Ireland, F, Thomachot L, et al. Airway humid-
1996. ification with a heat and moisture exchanger in mechanically ventilated
93 Lloyd G, Howells J, Liddle C, Klineberg PL. Barriers to neonates: a preliminary evaluation. Intensive Care Medicine 2007; 33: 336–43.
hepatitis C transmission within breathing systems: efficacy of a pleated 110 Schiffmann H, Rathgeber J, Singer D, Harms K, Bolli A,
hydrophobic filter. Anaesthesia and Intensive Care
1997; 25: 235–8. Zu¨chner K. Airway humidification in mechanically ventilated neonates and
94 Association of Anaesthetists of Great Britain and Ireland. infants: a comparative study of a heat and moisture exchanger vs. a heated
Infection Control in Anaesthesia. London: The Association of Anaesthetists humidifier using a new fast-response capacitive humidity sensor. Critical
of Great Britain and Ireland, 2002. 95 Association of Anaesthetists of Great Care Medicine 1997; 25: 1755–60.
Britain and Ireland.
Infection control in anaesthesia. Anaesthesia 2008; 63: 1027– 111 Chau A, Kobe J, Kalyanaraman R, Reichert C, Ansermino
36. M. Beware the airway filter: deadspace effect in children under 2 years. Pediatric
96 Association of Anaesthetists of Great Britain and Ireland. Anesthesia 2006; 16: 932–8. 112 Carter JA. Filters and small infants. Anaesthesia
Checking Anaesthetic Equipment, 3rd edn. London: The Association of 2004; 59:
Anaesthetists of Great Britain and Ireland, 719–20.
2003. 113 Bell GT, Martin KM, Beaton S. Work of breathing in
97 Kramer A, Kranabetter R, Rathgeber J, et al. Infection anesthetized infants increases when a breathing system filter is used. Pediatric
prevention during anaesthesia ventilation by the use of breathing system Anesthesia 2006; 16: 939–43. 114 Da Fonseca JMG, Wheeler DW, Pook JAR.
filters (BSF): joint recommendation by German Society of Hospital Hygiene The effect of a
(DGKH) and German Society for Anaesthesiology and Intensive Care heat and moisture exchanger on gas flow in a Mapleson F breathing
(DGAI). system during inhalational induction. Anaesthesia
GMS Krankenhaushygiene Interdisziplina¨r 2010; 5: 1–19. 98 Richards MJ, 2000; 55: 571–3.
Edwards JR, Culver DH, Gaynes RP. Nos- 115 Briassoulis G, Paraschou D, Hatzis T. Hypercapnia due to a
ocomial infections in pediatric intensive care units in the United States. heat and moisture exchanger. Intensive Care Medicine 2000;
National Nosocomial Infections Surveillance System. Pediatrics 1999; 103: e39. 26: 147.
116 Barnes SD, Normoyle DA. Failure of ventilation in an
99 Srinivasan R, Asselin J, Gildengorin G, Wiener-Kronish J, infant due to increased resistance of a disposable heat and moisture
Flori HR. A prospective study of ventilator-associated pneumonia in exchanger. Anesthesia and Analgesia 1996; 83: 193.
children. Pediatrics 2009; 123: 1108–15. 100 Gray J, George RH, Durbin GM, 117 Bennett N, Bingham R. Association of Paediatric
Ewe AK, Hocking MD, Anaesthetists of Great Britain and Ireland. Statement on paediatric filters
Morgan MEI. An outbreak of Bacillus cereus respiratory tract infections on a and breathing systems. Anaesthesia News
neonatal unit due to contaminated ventilator circuits. Journal of Hospital 2005; 213: 24–5.
118 Malan CA, Wilkes AR, Hall JE, Gildersleve C. An evalu-
Infection 1999; 41: 19–22. 101 Moores A, Bell GT. Filters and small infants. Anaesthesia
ation of the filtration performance of paediatric breathing system filters at
2004; 59: 719. low flows. Anaesthesia 2007; 62: 504–8. 119 Wilkes AR, Malan CA, Hall JE.
102 Luchetti M, Pigna A, Gentili A, Marraro G. Evaluation of The effect of flow on the
the efficiency of heat and moisture exchangers during paediatric filtration performance of paediatric breathing system filters.
anaesthesia. Pediatric Anesthesia 1999; 9: 39–45. Anaesthesia 2008; 63: 71–6.

2010 The Author Anaesthesia 2010 The Association of Anaesthetists of Great Britain and Ireland
51

Anda mungkin juga menyukai