The Occupational Safety and Health Act clearly places the burden of Minneapolis, Minnesota, and the submission of reasonable proof to seller prior to
compliance on the user of the equipment and the act is generalized to the extent that return of the equipment that the defect is due to a matter embraced within seller's
determination of compliance is a judgment decision on the part of the local warranty hereunder. Any such defect in material and workmanship shall be presented
inspection. Hence, Thermo Fisher Scientific will not be responsible for meeting the to seller as soon as such alleged errors or defects are discovered by purchaser and
full requirements of OSHA in respect to the equipment supplied or for any penalty seller is given opportunity to investigate and correct alleged errors or defects and in
assessed for failure to meet the requirements, in respect to the equipment supplied, as all cases, buyer must have notified seller thereof within one (1) year after delivery, or
interpreted by an authorized inspector. Thermo Fisher Scientific will use their best one (1) year after installation if the installation was accomplished by the seller.
efforts to remedy such violation at a reasonable cost to the buyer. Said warranty shall not apply if the equipment shall not have been operated
Safety in Transportation and Handling and maintained in accordance with seller's written instructions applicable to such
The Micro-Tech is an integral part of your plant and when transporting, equipment, or if such equipment shall have been repaired or altered or modified
handling, and installing the unit, your own plant safety instructions must be applied. without seller's approval; provided, however, that the foregoing limitation of warranty
Because your Micro-Tech and associated systems are tailored to application insofar as it relates to repairs, alterations, or modifications, shall not be applicable to
requirements, it is impossible to be precise about product mass/weight. If precise routine preventive and corrective maintenance which normally occur in the operation
values are required, the shipping crate will be marked with the overall shipping mass of the equipment.
of the product and this may be used as a reasonable guideline. “EXCEPT FOR THOSE WARRANTIES SPECIFICALLY CONTAINED
Safe Practices During Use, Maintenance, and Repair HEREIN, SELLER DISCLAIMS ANY AND ALL WARRANTIES WITH
This manual contains details, as appropriate, including the appropriate tools. RESPECT TO THE EQUIPMENT DELIVERED HEREUNDER, INCLUDING
However, because of its importance, the warning contained in the installation section THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR
is repeated here. USE. THE SOLE LIABILITY OF SELLER ARISING OUT OF THE WARRANTY
TO GUARANTEE PERSONAL SAFETY, CARE MUST BE TAKEN WHEN CONTAINED HEREIN SHALL BE EXCLUSIVELY LIMITED TO BREACH OF
WORKING ON OR AROUND THE MICRO-TECH. AS WITH ALL SUCH THOSE WARRANTIES. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH
DEVICES THE MAIN SUPPLIES (ELECTRICAL AND OTHER) TO THE OF THE WARRANTIES SET OUT ABOVE SHALL BE LIMITED TO THE
SYSTEM MUST BE LOCKED OFF WHEN PERFORMING REPAIR OR REPAIR OR REPLACEMENT OF ANY DEFECTIVE ACCESSORY, PART OR
MAINTENANCE WORK. MATERIAL WITH A SIMILAR ITEM FREE FROM DEFECT, AND THE
All of the recommendations for LVD apply to the prevention of electrical SELLER BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL
shock. If access to the electronics enclosure is required, the incoming AC power DAMAGES.”
supply should be isolated remotely and locked-off. Access to the electronics Purchaser agrees to underwrite the cost of any labor required for replacement;
enclosure by untrained personnel is not recommended. including time, travel, and living expenses of a Thermo Fisher Scientific Field
ensure that when installing the Micro-Tech, a switch or circuit breaker is used and is Thermo Fisher Scientific Bulk Weighing and Monitoring 501 90th Avenue NW
positioned close to the Micro-Tech in easy reach of the operator. The switch or circuit Minneapolis, MN 55433 Phone: (800) 445-3503 Fax: (763) 783-2525
isolator. Though the information provided herein is believed to be accurate, be advised that the
Thermo Fisher Scientific Warranty information contained herein is not a guarantee for satisfactory results. Specifically,
The seller agrees, represents, and warrants that the equipment delivered this information is neither a warranty nor guarantee, expressed or implied, regarding
hereunder shall be free from defects in material and workmanship. Such warranty performance, merchantability, fitness, or any other matter with respect to the
shall not apply to accessories, parts, or material purchased by the seller unless they products, and recommendation for use of the product/process information in conflict
are manufactured pursuant to seller's design, but shall apply to the workmanship with any patent. Please note that Thermo Fisher Scientific reserves the right to change
incorporated in the installation of such items in the complete equipment. To the and/or improve the product design and specifications without notice.
Conventions
The following conventions are used in this manual.
Safety Precautions
Listed below are the safety messages for your Micro-Tech and its
associated scale system. Please read all safety messages very carefully,
because this information is important—for your own personal safety and
the safety of others.
Table of Contents
Glossary ...................................................................................... 1
List of Figures
Figure 1–1. Field-Mounted Version of the Micro-Tech .............. 1-2
Figure 1–2. Panel-Mounted Version of the Micro-Tech ............ 1-2
Figure 1–3. Connectors on Underside of Enclosure ................. 1-6
Figure 2–1. Main Features of the Micro-Tech Console ............. 2-1
Figure 2–2. Location of Load-Cell Data .................................... 2-4
Figure 2–3. Run Screen .......................................................... 2-21
List of Tables
Table 1–1. Symbol Identification ............................................. 1-10
Table 1–2. Micro-Tech Technical Specifications ..................... 1-11
Table 4–1. Micro-Tech Parts List .............................................. 4-3
Table A–1. Load-Cell Technical Specifications .......................A-10
Chapter 1
Introduction
This manual provides the information you need to install, operate, and
troubleshoot the Micro-Tech. Please read the entire manual before
installing your Micro-Tech. For personal and system safety, and for the
best product performance, make sure you thoroughly understand the
manual before installing or using your Micro-Tech. If you have a
question not covered in this manual please refer to the Reference
manual.
The Micro-Tech has been properly packaged for shipment at the factory.
Unpacking the Please inspect all packages for damage before opening the shipping
Micro-Tech package, because the carrier is likely responsible for any damage. Once
removed from the package, the Micro-Tech can be safely stored with its
cover and latches secured and with the hole plugs installed. During
storage, do not expose the Micro-Tech to moisture or to temperatures
outside the range of –22 to +158°F (–30° to +70°C).
The Micro-Tech has provisions for four outputs on the digital output
board, plus one DC output from the mother board—making a total of
five, one of which can be defined as a Fault output. In addition, many
automatic and check functions are available to monitor its calibration
functions and maintenance schedule.
There are two models of Micro-Tech: the field-mounted version
(figure 1–1) and the panel-mounted version (figure 1–2). For the panel-
mounted version, provide a cut-out (see figure 1–2 for dimensions) in
the panel and, after removing the holding brackets and installing the
gasket, insert the Micro-Tech.
Please read the following warnings and cautions before installing, operating, or maintaining the Micro-Tech.
General Safety Precautions Incoming Power Safety
CAUTION. Do not install, operate, or perform any maintenance Please read the following warnings and cautions, when working
procedures until you have read all the safety precautions listed below. with incoming power to the Micro-Tech or its associated systems.
CAUTION. Do not connect power to the electronics or turn on the unit
CAUTION. Do not connect power until you have read and
until you have read and understood this entire manual. The precautions
understood this entire section. Improper connection may result in
and procedures presented in this manual must be followed carefully in
damage to your Micro-Tech.
order to prevent equipment damage and protect the operator from
possible injury. WARNING. All wiring must be in accordance with standards
CAUTION. For North America locations a certified Nema 4/4X (IEC, EN) national and local codes (NEC, VDE, and so forth)
bushing must be used for openings. For other locations see your local outline provisions, for safely installing electrical equipment.
Electrical Authorities. Installation must comply with specifications regarding wire types,
WARNING. Covers over the electronics should always remain in place conductor sizes, branch circuit protection, and disconnect devices.
during operation. They should be removed only for maintenance Failure to do so may result in personal injury and/or equipment
procedures with the machine’s power OFF. Be sure to replace all covers damage.
before resuming operation.
WARNING. Ground impedance must conform to the requirements
WARNING. All switches (such as control or power) must be OFF when
of national and local industrial safety regulations and/or electrical
checking input AC electrical connections, removing or inserting printed
codes. The integrity of all ground connections should be
circuit boards, or attaching voltmeters to the system.
periodically checked. For installations within a cabinet, a single
WARNING. Incoming voltages must be checked with a voltmeter
safety ground-point or ground bus-bar connected directly to
before being connected to the electronics.
building steel should be used. All circuits including the AC input
WARNING. Extreme caution must be used in testing in, on, or around
ground conductor should be grounded independently and directly to
the electronics, PC boards, or modules. There are voltages in excess of
this point/bar. Grounding all enclosures and conduits is strongly
115V or 230V in these areas. Avoid high voltage and static electricity
recommended.
around the printed circuit boards.
WARNING. Maintenance procedures should be performed only by CAUTION. Verify that the input voltage is correct with an AC
qualified service personnel and in accordance with voltmeter before you connect it to the Micro-Tech.
procedures/instructions given in this manual.
CAUTION. Earth ground must be provided to the Micro-Tech. Do
WARNING. During maintenance, a safety tag (not supplied by Thermo not use conduit to provide this ground.
Fisher Scientific) should be displayed in the ON/OFF switch areas as a
CAUTION. A readily accessible disconnect device (maximum 20
precaution instructing others not to operate the unit.
amp) must be incorporated in the field wiring. This disconnect
WARNING. Only qualified service technicians should be allowed to device should be within easy reach of the operator and must be
open and work in the electronics, power supply, control, or switch marked as the disconnecting device for the equipment.
boxes. EMC Instructions
The Micro-Tech may cause radio interference if used in a
WARNING. This equipment should not be operated or utilized in
residential or domestic environment. The installer is required to take
applications other than those stated in the original order.
measures to prevent interference, in addition to the essential
WARNING. All panels covering the electronics must be in place and requirements for CE compliance provided in this manual, if
tight before wash down procedures. Damage to the electronics could necessary.
result from water, moisture, or contamination in the electronics
Conformity of the Micro-Tech with CE/EMC requirements does not
housing.
guarantee an entire machine or installation complies with CE/EMC
requirements.
Hardware This section tells you how to complete the hardware installation for your
Micro-Tech. Please go to the appropriate section, depending on which
Installation model of Micro-Tech you purchased (field-mounted or panel-mounted).
Important Wiring Before installing the Micro-Tech, please read the following important
and Safety safety information about wiring up the Micro-Tech.
Information
l Ensure power is OFF at the main disconnect.
l Do not route load-cell and signal cables in the same conduit with
power cables or any large source of electrical noise.
l Earth ground all instrument chassis’ and conduits. A ground
connection between all conduits is required.
l Connect the shields only where shown.
l Check that all wires are tight in their connections.
l Never use a “megger” to check the wiring.
l A readily accessible disconnect device must be incorporated in the
field wiring. This disconnect should be within easy reach of the
operator and must be marked as the disconnecting device for the
Micro-Tech and associated equipment.
l All conduits should enter the bottom of the enclosure. Do not run
conduit through the top or sides of the enclosure.
Installing the The integrator should not be exposed to excessive vibration, heat, direct
sunlight, or moisture. The ideal mounting location would be on a
Field Model separate wall or beam in view of the device being monitored. Refer to
system wiring diagram for the maximum allowed distance from the
monitored device to the Micro-Tech.
Mounting Mount the Micro-Tech to a rigid, flat, vertical surface using four
mounting holes provided on the back of the enclosure. Care should be
taken to ensure the mounting surface is flat, so as not to twist or warp
the fiberglass enclosure when tightening the mounting bolts.
Connecting the To connect the incoming power, use the following procedure. Please
Incoming Power note that all units shipped from the factory are configured for 100 to 240
Supply VAC.
4. Locate the wiring panel (see figure 1-3 below), which lies on the
underside of the electronics enclosure. The wire-safety ground-
terminal is located on the enclosure back panel.
Mounting Provide a cut-out (see figure 1-2 for dimensions) in the panel and, after
removing the holding brackets, and installing the gasket, insert the
Micro-Tech. From the back, insert the holding brackets on both sides of
Connecting the To connect the incoming power to the Micro-Tech, use the following
Incoming Power procedure. Please note that all units are 24VDC only.
Supply
1. For input power, use 16 AWG / 1.5 mmsq standard wires.
2. Wire the safety ground to the terminal labeled “E” on the Power
Input Terminal.
3. Wire the +24VDC to the terminal labeled “+” on the Power Input
Terminal.
Configuring In most instances, your Micro-Tech is shipped to you from the factory
with all the needed jumpers installed and the switches set in the correct
Jumpers and positions for your particular installation and application. As a result, you
Switches should not need to connect any jumpers or set any switches but, if you
do, all the appropriate settings are shown in the specific model reference
manual.
Standard The Micro-Tech has many hardware and software features. The standard
features of the Micro-Tech are listed below.
Features
l Menu-driven scroll entries on a four line display.
l Four LED status indicators.
l Audit trail.
l Automatic zero and span calibration.
l Auto zero tracking (where applicable).
Micro-Tech Each Micro-Tech has been designed for a specific application and is
capable of performing all of the necessary measuring functions. All of
Menus and the required functions are resident in the software of the microprocessor.
Functions Optional functions are automatically turned on when the relevant
hardware is installed, or after the operator has selected them through the
Monitoring The Integrator includes internal diagnostics that generate alarms in case
Functions of hardware failures or programming errors.
Alarms are visible on the display and can be acknowledged and reset
through the keypad, digital input, or serial line. Alarms can be delayed
to avoid intervention in case of short time peaks. Each individual alarm
can be programmed to operate as alarm, shut down, or none. Two LED’s
indicate the cumulative status of alarms and shut down. Digital outputs
are also provided for the following:
• Hardware failure
• Alarm cumulative
l Serial Communications
The communication protocol allows a remote intelligent device to
read the contents of the registers and write to some registers. During
the communication activity, the Micro-Tech always acts as a Slave,
responding to a request from a Master device on the line, but never
attempting to send messages out. One electrical interface may be
selected and accessed through one communication port.
l Ethernet Port
The Micro-Tech has a built-in Ethernet port. Communications
protocols Ethernet/IP and Modbus/TCP can be used. The Micro-
Tech is a slave device only, and cannot initiate messages.
Symbol Here are the details of the symbols used on the Micro-Tech.
Identification
Table 1–1. Symbol Identification
Symbol Description
Alternating current
PROTECTIVE
CONDUCTOR TERMINAL
Caution (refer to
accompanying documents)
Conformity with the Low Voltage (LVD) Directive and Electromagnetic Compatibility
Standards (EMC) Directive has been demonstrated using harmonized European Norm (EN)
Applied standards published in the Official Journal of the European Communities, and
International (IEC) applicable standard used in North America.
The Micro-Tech™ 9000 series comply with the EN and IEC standards listed below,
when properly installed in accordance with this and other relevant manuals.
l CAN/CSA-C22.2 No.61010.1-04
Safety Requirements for Electrical Equipment for Measurement, Control, and
Laboratory Use.
Part 1: General Requirements.
l UL 6101-1(2nd Edition)
Safety Requirements for Electrical Equipment for Measurement, Control, and
Laboratory Use.
Part 1: General Requirements.
l UL 60950-1
Information Technology Equipment—Safety
Part 1: General Requirements.
l IEC/EN 61010-1:2001
Safety requirements for electrical equipment for Measurement, Control, and
laboratory use.
Part 1: General requirements.
The Micro-Tech™ 9000 series has been tested with the EN and IEC standards listed
below.
l IEC/EN 61326-1
Electrical equipment for measurement, control and laboratory use—EMC
requirements.
Part 1: General requirements
l EN 55011
Limits and methods of measurement of radio disturbance characteristics of
industrial, scientific and medical (ISM) radio-frequency equipment.
l EN 55022
Information technology equipment. Radio disturbance characteristics. Limits and
methods of measurement.
l 2004/108/CE—EMC Directive.
Specifications
Table 1–2. Micro-Tech Technical Specifications
Description Specification
Field Mount Enclosure NEMA 4X, IP66, dust and watertight, 17.01 [432] x 14.18 [360] x
6.59 [167] inches. Fiberglass reinforced polyester.
Steel chassis providing EMI/RFI shielding.
Panel Mount Enclosure Size: 12.11 [308] x 4 [102] x 7.95 [202] inches.
Material: Zinc-plated mild steel.
Environmental Conditions Mounting Mount as close to the measuring device as possible without
exposing to excessive heat or moisture.
Field Mount suitable for outdoor mounting.
Temperature (Ambient) Storage: -22° to +158° F (-30° to +70° C).
Operating: –4° to +140° F (-20° to +60° C).
Relative Humidity Maximum relative humidity 80% for temperatures up to 31°C
decreasing linearly to 50% humidity at 40°C.
Pollution Degree Level 2 per IEC 61010-1
Altitude Up to 6,561 ft (2000m)
Installation Category 2
Shock 15G peak for 11ms duration (±1.0 ms)
Vibration 0.006 in./0.152 mm displacement, 1G peak
Emission Limitation According to IEC/EN 61326-1, Class A
Noise Immunity According to IEC/EN 61326-1, Industrial Environmental
Nominal Voltage Field Mount: 100-240 VAC.
Panel Mount: 24VDC +10%,-15% (user supplied).
Nominal Frequency Field Mount: 50-60 Hz.
Panel Mount: DC only.
Fusing 250VAC, 2A fast acting, on motherboard
Power Consumption 50 VA max.
Maximum Non-Destructive Input Voltage Field Mount: 265 VAC.
Panel Mount: 28VDC.
DC Power Supply Required for Panel Mount Output voltage: 24 VDC
Isolation: No.
Output current: 2A minimum, short circuit protected.
Description Specification
Processor Coldfire MCF5234 32-bit microprocessor
2 MB Flash memory
128K NVRam
2 Integrated UARTs and Ethernet communication peripherals.
Removable Storage USB flash driver port
RAM Battery Life expectancy of the RAM support battery is a minimum of 10
years, if power is not applied. Under normal operation where
power is on continuously, life expectancy is much longer.
Inputs #1, #2 Optically isolated. Powered by + 24VDC supply.
Built-in current source for dry contact use.(Gold plated contacts
recommended)
Frequency range Voltage/current type sensor: 0.25 to 2.0 kHz.
Contact closure type sensor: 0.25 to 30 Hz.
Low threshold: +1.3 VDC min.
High threshold: +2.2 VDC max.
Low or High Pulse Duration Voltage/current type sensor: 200 us min.
Contact closure type sensor: 15 ms min.
Hysteresis 0.8 VDC minimum.
Input impedance 10 k-ohm typical, 500 ohm minimum.
Input source current -2 mA nom. at 0 VDC.
Max. non-destructive input voltage ±28 peak, continuous.
Digital Output (Output #5) Able to drive TTL, CMOS, or relay solenoids.
Current sinking driver.
+24 VDC internal supply, 100mA DC maximum.
Standard Communication Serial Interface UART 0 RS-232C provides support for modem.
RS-485; 2 and 4 wire multi-drop.
Data rate: 110 to 19200 bits/second, operator selectable from
the keypad. Data format: Asynchronous, bit-serial, selectable
parity, data length, and stop bits.
Optical isolation: 250 VRMS max. Input Voltage: ±30 Vdc max.
(RS-232C) ±15/-10 Vdc max. (RS-485).
Cable length: RS-232C, 50ft [15m] max; RS-485, 4000 ft
[1219m] max.
Standard Communication Serial Interface UART 2 RS-485; 2 and 4 wire multi-drop in RS- 485.
(For use with Thermo Fisher Scientific equipment Data rate: 110 to 19200 bits/second, operator selectable from
only.) the keypad. Data Format: Asynchronous, bit-serial, selectable
parity, data length, and stop bits.
Isolation: Non-Isolated.
Cable Length: 4000ft [1219m] max.
Description Specification
Physical: 100baseT, RJ45 Ethernet port
Embedded Web server
Ethernet Communication Supported Protocols: Modbus TCP, Ethernet IP.
l Keypad
l Soft keys
l Status LEDs
Display Screen This displays the built-in Micro-Tech menus as well as any entries you
make using the keypad. The display also shows the current functions
(such as Edit, Enter, and Clear) that are assigned to the four Micro-Tech
soft keys situated below the display.
Keypad The keypad allows you to scroll through the Micro-Tech menus, enter
numbers and letters into the Micro-Tech’s menus, and control the
operation of the Micro-Tech using the Run button. As you will already
have noticed—similar to the keys on a cell phone—the Micro-Tech’s
number keys have multiple uses. All are context sensitive, meaning, for
example, that when the Micro-Tech is displaying a menu, the number
“8” key operates as a down-arrow key, but when the Micro-Tech is
expecting you to enter a number, it operates as an “8” key. Similarly, in
the print menu, when you are naming your output, repeatedly pressing
the “8” key brings up, in succession, the letters V and W.
l Arrow Keys
The up-arrow and down-arrow keys allow you to scroll through the
Micro-Tech menu screens—up and down as well as left and right in
some menus.
l Control Keys
The Micro-Tech has two control keys—the Menu button and the
Run button. Once the Micro-Tech is up and running, pressing the
Menu button brings up the menu screens. Pressing the Run button
returns the Micro-Tech to its normal operating mode.
The four blue keys below the display screen are “soft keys,” that is, they
Soft Key Buttons have different functions depending on which menu you are using. The
soft keys are assigned to various menu-selection and data-entry
functions—such as Edit, Clear, Reset, Totals, and so forth.
Status LEDs The four indicators show the status of the Loss-In-Weight Feeder
Controller.
• REMOTE
• AUTO
• ALARM
• READY
The unit can perform automatic zero and span calibrations. Analog
(current) output signals or communications can be used to transmit rate,
net and gross weight to other control devices. Displayed variables and
analog outputs can be smoothed via damping filters, individually
programmable.
The system fully handles the refill of the bin, by acting in volumetric
mode during the refilling time and adjusting the volumetric constants
according to the current net weight. One single instrument can control
one or two completely independent loss in weight feeders.
Scale Capacity Determine the maximum scale capacity in one of the following weight
units (Pounds, Tons, LTons, kg, t) and record the capacity and weight
units below. (Example: 400.0 pounds)
Scale #1
Scale #2
Load Cell Capacity, All Thermo Fisher load cells have the capacity, sensitivity, and
Sensitivity, and resistance marked (as shown below) on the end of the cable. In case the
Resistance label is not present please refer to the data sheet supplied with the load
cell. Enter the capacity, sensitivity, and resistance below.
Load Cell Determine the load cell size in weight units (Pounds, Tons, LTons, kg,
t). Record the weight and units below.(Example: 250.0 Pounds)
Capacity
(Load Cell Capacity) Scale #1
(Load Cell Capacity) Scale #2
Load Cell From the load cell nameplate, determine the load cell sensitivity in
mV/V. Record the sensitivity below. (Example 3.000 mV/V)
Sensitivity
mV/V (Load Cell Sensitivity) Scale #1
mV/V (Load Cell Sensitivity) Scale #2
Load Cell Determine the load cell resistance from the label or data sheet or
measure the signal (output) resistance with a digital VOM. You must
Resistance
determine the resistance separately for each individual load cell. Record
the individual resistances below. (Example: 350.000)
(Ohms) Scale #1
(Ohms) Scale #2
Using the Please see page 2-1 of the manual to familiarize yourself with the Micro-
Tech’s console and how to use it.
Console
This section gives you step-by-step instructions to guide you through the
Initializing the software-initialization process.
Software
NOTE. You must complete the entire software initialization and scale-
calibration procedure before putting the Micro-Tech into operation.
There are no shortcuts!
Overview There are four steps in the software initialization process, as follows.
Cold-Starting the The first time you power up the Micro-Tech, you are doing what is
known as a “cold-start.” Once the Micro-Tech is up and running, you
Micro-Tech can use the cold-start procedure (described below) to—in computer
terms—“reboot” the Micro-Tech. In other words, when you do a cold
start, the Micro-Tech’s RAM memory is erased and everything is
returned to its initial start-up state. As a result, cold-starts are used, for
example, to restore all the Micro-Tech settings from a previously made
back-up flash drive.
To cold-start the Micro-Tech, do the following.
2. The Micro-Tech starts up, and the Alarm LED will light to indicate
that the Micro-Tech has not yet been initialized or calibrated. After
a brief delay the Default screen appears, as shown in the section
below.
Setting the Date You are now ready to set the current date and time. (In the following
example we are going to set the date to May 21, 2013.)
Install Factory
Defaults?
NO YES
Exact date?
- Date 01 – 01 – 2012
DAY 1
EDIT
3. Press the Edit button. (The Micro-Tech clears the Day entry field
leaving just the underline.)
4. Use the keypad to enter the correct day. Remember to enter two
numbers for the day. If you make a mistake, press the Clear button.
(We entered 21 for day, as shown below.)
Exact date?
- Date 01 – 01 – 2012
DAY 21
EDIT
5. Press the Enter button. Follow steps 3 and 4 above to enter the
correct month and year.
6. Press the Enter button. The display should now look something
like this. (You may have to repeatedly press the Edit and Enter
buttons on start up, scrolling through the fields again, to get to this
screen.) Either way, make sure this screen is displayed before
proceeding.
Exact date?
- Date 05 – 21 – 2013
YEAR 2013
EDIT
7. You are now ready to enter the correct time, as described below.
Setting the Time In the following example we are going to set the time to 2:09 p.m. To set
the correct time, do the following.
1. Press the down-arrow button (see figure 2–1). The display should
currently look like this.
Exact time?
- Time 12:00 am
EDIT AM/PM
2. Press the Edit button. (The Micro-Tech clears the hour entry field
leaving just the underline.)
3. Use the keypad to enter the correct hour. Remember to enter two
numbers for the hour.
Exact time?
- Time 2:00 am
EDIT AM/PM
6. Press the Edit button. (The Micro-Tech clears the minute entry
field leaving just the underline.)
10. The time is now set. Press the down-arrow button to bring up the
USB screen. The Micro-Tech pauses for about 10 seconds, while it
checks for the presence of a flash drive in the USB port. (If you
were rebooting the Micro-Tech to restore your previously saved
settings, this is where you would insert the back-up flash drive into
the USB port.)
Choosing a The default language shown in the Micro-Tech display is English. You
can, however, choose other languages.
Language
1. The Micro-Tech display should currently look like this.
- MEMORY ERASED -
key to continue to
NOTE. Ignore the “Memory Erased” message. The date and time you
already entered have been retained.
2. To select the current language, press the Enter button and the scale
set-up screen appears.
and calibration
6. Press the down-arrow button (or Scroll button) and the “Display
Scroll 1” screen appears.
Selecting This menu allows you to choose what units of measurement the Micro-
Tech uses when displaying its results. The Micro-Tech can display
English/Metric information using the following units of measurement.
Units
l Standard English units—such as pounds, standard tons, and long
tons.
l Metric units—such as kilograms and tonnes.
l Both English and metric units. (The “Mixed” option.)
- DISPLAY SCROLL 1 -
Measure Units
CHOICE
Setting the Rate This menu allows you to choose the rate units used by the Micro-Tech.
Clearly, which rate units are available in this menu depends on the
Units choices you made in previous menus (English, Metric, Mixed).
English Rate Units The Micro-Tech display should currently look like this, if you chose
English units.
- DISPLAY SCROLL 2 -
Rate Units
CHOICE
Metric Rate Units The Micro-Tech display should currently look like this, if you chose
metric units.
- DISPLAY SCROLL 2 -
Rate Units
CHOICE
Mixed Rate Units The Micro-Tech display should currently look like this, if you chose
mixed units.
- DISPLAY SCROLL 4 -
Rate Units
CHOICE
Setting the This menu allows you to choose the weight units used by the Micro-
Tech. Clearly, which weight units are available in this menu depends on
Weight Units the choices you made in previous menus (English, Metric, Mixed).
English Weight Units The Micro-Tech display should currently look like this, if you chose
English units.
- DISPLAY SCROLL 3 -
Weight Units
CHOICE
Metric Weight Units The Micro-Tech display should currently look like this, if you chose
metric units.
- DISPLAY SCROLL 3 -
Weight Units
> kg <
CHOICE
c. “percent %”
Mixed Weight Units The Micro-Tech display should currently look like this, if you chose
mixed units.
- DISPLAY SCROLL 3 -
Weight Units
> kg <
CHOICE
Setting the Total This menu allows you to select the specific units of measure that are
displayed by the Micro-Tech when reporting its results (known as
Units
“Totalization”). Clearly, which units of measure are available in this
menu depends on the choice you made in previous topic (“Selecting
English/Metric Units).
English Total Units The Micro-Tech display should currently look like this, if you chose
English units.
- DISPLAY SCROLL 4 -
Total Units
CHOICE
Tons (the standard U.S. ton, 2,000 lbs—also known as the British “short
ton”) is the default value.
Metric Total Units The Micro-Tech display should currently look like this, if you chose
metric units.
- DISPLAY SCROLL 4 -
Total Units
>t<
CHOICE
Press the down-arrow button (or Scroll button) and the “Number of
Entering Scale scales” screen appears. On this scroll you can choose between 1 to 2
Data scales.
- SC DATA SCROLL 1 -
Number of scales
EDIT
Entering the type When the Number of scales is set to 1, the following “Type of scale”
screen is shown. When using only 1 scale you have the capability to
of Scale utilize multiple loadcell A/D channels (2,3,4) to match the number of
load cells on your scale.
> One A/D Channel < > Two A/D Channel <
CHOICE CHOICE
In you have an optional Dual A/D expansion board installed you also
have the following choices:
> Three A/D Channel < > Four A/D Channel <
CHOICE CHOICE
The default value is “One” A/D channel. However, if your scale has 2,
3, or 4 load cells, choose the “Two”, “Three”, or “Four” A/D channels
option. Review the system-specific wiring diagram to determine the
number of load cells on your scale.
Entering the This menu allows you to enter the maximum scale capacity of the
particular scale you are using in your facility. Please note that the
Maximum Scale maximum scale capacity is expressed as a weight, Pounds (lbs), Tons,
Capacity (T), and so on.
The Micro-Tech display should look something like this, depending on
the choices you made in the menus above.
- SC DATA SCROLL 2 -
100.00 lbs
EDIT
- SC DATA SCROLL 2 -
1750 lbs
ENTER • CLEAR
Entering the This menu allows you to tell the Micro-Tech how to report the quantity
of material that crosses the scale in one hour. For example, if 1,750 tons
Scale Divisions cross the scale in an hour and you want the results reported to one
decimal place (that is, to the nearest 200 lbs.), you would choose a scale
division of 0.1. As a result, hourly rates would be reported as—for
example—1742.8 Tph (tons per hour).
Please note that the choice of division has no bearing on the accuracy of
the underlying numbers, and that if your control system contains a PLC
- SC DATA SCROLL 3 -
Scale divisions
CHOICE
Entering Number Enter the number of load cells on each scale, then press the down-arrow
button to move on.
of Load Cells
READY AUTO ALARM REMOTE
- SC DATA SCROLL 4 -
# of load cells
EDIT
Load Cell Capacity, Enter the load cell Capacity, Sensitivity and Resistance for each scale.
Sensitivity, and Note: You must enter the resistance separately for each load cell
Resistance
- SC DATA SCROLL 8 -
Lever Ratio
EDIT
Calibration The following screen appears. The Micro-Tech is now setting the
appropriate span number for the scale.
CALIBRATION
IN
PROGRESS
1. When the span number has been set, the following screen appears briefly
in the display. Notice that the red “Alarm” LED in the console goes off
and the green “Ready” LED comes on.
S1 calibrated
If you get an “S1 not calibrated” message, check all the numbers you
entered in the previous steps. Then go back to the “Entering the Scale
Data” section above (see page 2-14) and carefully re-enter all the data
into the Micro-Tech. If the calibration fails again, check the load cell (or
cells) are working and sending signals to the Micro-Tech.
3.Press the Run button and the Micro-Tech Run screen appears, which
looks like this.
READY AUTO ALARM REMOTE
PV 0.0 lb/h
TOTALS
2. Make sure the Run screen (see below) is currently being displayed.
Run Screen The screen below is known as the Run screen. This screen shows the
process variable (rate).
PV 0.0 lb/h
TOTALS
Line 1 of the display always displays the Process Variable. A “R” will
appear when the system is in refill.
Lines 2 and 3 of the display are by default blank, but can be selected to
show:
Weight, Reset tot, Master tot, Date/Time, Bargraph
R PV 50.0 Lb/h
Gross 146.0 lbs
R. Tot 142.0 lbs
TOTALS
2. Press the Menu button and the “Main Menu 1” screen appears.
MAIN MENU 1
ZERO SPAN
CAL CAL
3. Press the Zero Calibration button and the following screen appears.
ZERO CAL
START MANUAL
4. Make sure your scale is completely empty, then press the Start
button. The count-down screen appears. (The data in your screen
will, of course, be different.)
AUTO ZEROING
Time remaining 59
Gross 0.00 Tons
END ABORT
AUTOZERO COMPLETE
Error 0.01%
Change zero?
YES NO
In our example, the display shows that the newly established zero is just
0.01% different from the previous zero, meaning that both zeros are
essentially the same and the scale is performing consistently. However,
as there has been a small amount of drift, we decide to reset the zero to
the newly established zero point.
6. Press the Yes button to accept the new zero, and the zero-changed
screen appears.
ZERO # CHANGED
RUN MENU
7. Press the Run soft key in the display to return the Micro-Tech to
the Run mode.
Doing an R-Cal To perform an R-Cal span calibration for your scale, do the following.
Span Calibration
1. Make sure the Run screen is currently being displayed.
2. Press the Menu button and the “Main Menu 1” screen appears.
MAIN MENU 1
ZERO SPAN
CAL CAL
press START
START MANUAL
4. Make sure the scale is empty, then press the Start button. The
count-down screen appears. (The data in your screen will, of
course, be different.)
AUTO SPANNING
Time remaining 59
0.000 Tons
END ABORT
5. The calibration time (60 seconds) will start counting down. The
End key can be used to conclude the function in less time. When
AUTOSPAN COMPLETE
Error 0.01%
Change span?
YES NO
7. Press the Run button (in the display not the keypad) to return the
Micro-Tech to the Run mode. Congratulations! You are now ready
to put your Micro-Tech into operation.
In the example screen above, the display shows that the established span
is just 0.01% different from the previous span, meaning that both spans
are essentially the same and the scale is performing consistently. This
error is below the critical threshold error of 0.5%. As a result, the span
should not be changed. Record the span results for future reference.
6. This is important! Press the No button (that is, do not change the
span) and the following screen appears. (Your numbers will, of
course, be different.)
SPAN # UNCHANGED
7. Press the Run soft key in the display to return the Micro-Tech to
the Run mode.
Chapter 3
Maintenance and
Troubleshooting
The maintenance information in this manual should meet your service
needs. If problems occur requiring technical assistance, please call 1-
800-445-3503 or the local Thermo contact listed in Chapter 4. Thermo
Scientific has a repair center located at our plant in Minneapolis,
Minnesota. Contact one of our technical representatives at 1-800-445-
3503, or the local Thermo contact listed in Chapter 4 for assistance. To
expedite your service request, please have your Micro-Tech model,
serial number, and scale data available.
Frequently
Here is a list of frequently-asked questions (FAQs) to help you resolve
Asked common problems and concerns about operating, calibrating, and
Questions maintaining your scale.
Question Answer
What is the best way to calibrate my The best way to calibrate your scale is to use the “Test Weight Calibration”
scale? method.If, for whatever reason, you cannot use the Weights method
(capacity of the system too big) another suggested method is to load the
scale with a known weight of material. As a last possibility you can run a
“simulated-calibration” test. For example, all Micro-Techs can be calibrated
using the “R-Cal” procedure.
How often should the zero and span As a general rule, if you make or receive payments based upon the weight
be calibrated? readings from your scale, the scale should be zeroed daily and the span
checked weekly.
• Your scale is only as good as the repeatability of your error on
repeated zero calibrations.
• The span should never change drastically, if the zero is properly
maintained.
How often should I check the The scale should only need weekly preventive-maintenance checks for
mechanical installation of the scale? material build up in critical areas, such as where the load cells are located.
This is schedule is dependent on the type of operations that are performed
over the scale (e.g. manual loading of product, high possibility of material
escape, and so forth.)
Why do I need to see repeatability The repeatability of a scale is important for the accuracy of the scale. If the
during calibrations? scale cannot repeat tests within the scale’s percentage of accuracy, then
you should investigate why the scale is not repeating within the appropriate
tolerances.
Every time I complete a zero or span The only time you should change zero or span is on the initial, start-up test.
test and get a percentage of error, do I Press the “Yes” button to change the span. In any subsequent test, even
say “Yes” each time to change the when there is an error, press the “No” button—as any additional tests are
zero or span? for repeatability, which is a maintenance feature of calibrations.
Question Answer
What kind of information is available The programmable outputs are as follows.
from the digital output?
Alarm cumulative Totalization
Shutdown cumulative Custom
Ready Fault
High Weight Load Weights
Low Weight Deviation alarms
High Rate
Low Rate
Chapter 4
Service, Repair, and
Replacement Parts
Listed below is information about how to get help servicing, repairing,
and obtaining replacement parts for your Micro-Tech. In addition,
Thermo Fisher Scientific provides experienced, on-site service
technicians who can assist you with installing, setting up, calibrating,
maintaining, and repairing your Micro-Tech. They can also help you
train your operators and solve virtually any Micro-Tech-related problem.
For a detailed list of the spare parts available for your Micro-Tech,
Overview please see the Parts List in table 4–1 on page 4-3.
Getting Ready The quickest way to get the parts you need for your Micro-Tech is to do
the following.
to Order
1. Identify the broken or faulty parts.
2. See whether the part is shown in table 4–1, which lists parts that
may need to be replaced. Note the part number from the table. If
the part is not listed in the table, contact Thermo Fisher Scientific
directly (as described in the following pages).
Please verify and write down your Micro-Tech model number and serial
Contacting number before contacting us. Things will go a lot more quickly and
Thermo Fisher efficiently once we know this information.
Scientific
North America Brazil
1-800-445-3503 +55-11-2367-2192
1-763-783-2525 +55-11-2367-2192 fax
Service.bulk.us@thermofisher.com
parts.bulk.us@thermofisher.com
Germany Chile
+49 (0) 208-824930 +56 2 2378 5080
+49 (0) 208-852310 fax +56 2 2370 1082 fax
service.oberhausen@thermofisher.com
Italy China
+39 02-959514-1 +86 10-8419-3588
+39 02-953200-15 fax +86 10-8419-3580 fax
service.bulk.emea@thermofisher.com
Spain India
+34 91-484-5965 +91-22-4157-8800
+34 91-484-3597 fax +91-22-4157-8801 fax
- MAIN MENU 2 -
Press MENU for more
SCALE CALIB
DISPLAY DATA DATA
2. Press the Calibration Data button and the following screen appears.
- START OF SCROLL -
Use SCROLL keys to view
selections.
3. Press the down-arrow button repeatedly until the “Cal Data Scroll
5” screen appears.
EDIT
5. Use the keypad to enter a zero value (0). Then press the Enter
button. Your screen will now look like this.
EDIT
- MAIN MENU 1 -
Press MENU for more
ZERO SPAN
CAL CAL
AUTOSPAN R Cal
Empty scale, then
press START
START MANUAL
3. Make sure the scale is empty, then press the Start button. The
following Auto Spanning screen appears.
AUTO SPANNING
Time remaining 59
44.12 lbs
END ABORT
AUTOSPAN COMPLETE
Error -0.07 %
Change Span?
Yes NO FACTOR
SPAN UNCHANGED
Old span # 1002010
New span # 1002010
MENU REPEAT
Set R-Cal Factor Now to set the new R-Cal factor, do the following.
- MAIN MENU 2 -
Press MENU for more
SCALE CALIB
DISPLAY DATA DATA
EDIT
3. Press the Edit button and use the keypad to enter the R-Cal error.
Then press the Enter button. The error you entered is displayed in
the screen below.
READY AUTO ALARM REMOTE
4. Factoring R-Cal is now complete. Press the Run button and the
Micro-Tech Run screen reappears, as shown below.
PV xxxxx.x lb/h
TOTALS
COMM A, (UART 0)
A/D Jumpers— Load-cell sense is controlled by selectable jumpers (J14 and J15 for
Load-Cell Sense channel 1, and J19 and J20 for channel 2) located on the motherboard.
The jumpers should be in position “1-2” local sense, if the distance is
less than 200 feet between the load cell and the Micro-Tech. For
distances greater than 200 feet and less than 3,000 feet, the jumper
should be in position “2-3” and a special 6-wire cable is required. Refer
to the field wiring diagram that is appended to this manual, for jumper
requirement in the scale junction box.
Load-Cell
Table A–1. Load-Cell Technical Specifications
Specifications
Load Cell Excitation Power Supply 5 VDC ±10%, 90 mA, minimum load impedance (58 ohms).
Output short circuit, 0.5 A maximum.
Load Cell Number: Up to six (6) 350-ohm load cells in parallel. Cable
distance: 200ft [61m] or less without sense, or 3000ft [914m]
with sense.
Load cell input circuits
Sensitivity: 0.5mV/V to 3.5 mV/V (keypad selectable).
(2 each)
Input Impedance: 1M-ohm minimum.
Maximum Usable Signal: 114% of 3mV/V.
Internal A/D counts: (3mV/V): 6,440,000.
Isolation: Non-isolated.
Max non-destructive input voltage: ± 6 V relative to ground.
Load Cell Cable Shield: Connected to earth ground.
Load Cell 4 wire system: cable distance not exceed 200ft [61m].
6 wire system: cable distance not to exceed 3000ft [914m].
Excitation-Sense Circuitry Nominal input voltage: 5 VDC.
(2 each) Input impedance: 100 k-ohm minimum.
Jumper selectable: Local or remote sense.
l Digital Inputs
l Two (2) (DC) inputs on the motherboard. (See the Specifications
in chapter 1.)
l Three (3) programmable dry-contact inputs on the optional DC
Input Board, or three (3) programmable opto-22 inputs modules
on the optional Opto22 Input Board.
l Eight (8) programmable inputs on the optional Digital I/O
8in/8out Board. Two of these boards may be installed for a total
of sixteen (16) inputs.
l Digital Outputs
l One (1) Digital Pulse Output on the motherboard. (See the
Specifications in chapter 1.)
l Four (4) programmable relay outputs on the optional Relay
Output Board, or four (4) programmable opto-22 output modules
on the optional Opto22 Output Board.
l Eight (8) programmable outputs on the optional Digital I/O
8in/8out Board. Two of these boards may be installed for a total
of sixteen (16) outputs.
DC Input Board This is an optional board with three inputs (inputs #3–5).
Type: Current sourcing to common ground. Designed for dry-contact
input. Rated: 24VDC, 5mA typical. Input function is assigned by user.
Opto22 Input This is an optional board with three inputs (inputs #3–5). The module
Board options are as follows.
Opto22 Output This is an optional board. There are four (4) outputs (outputs #1–4).
Modules are installed in slots U2–U5 on the output board. One spare
Board fuse is located on the output board. The module options are as follows.
This is an optional board with eight (8) inputs (inputs #6–13) and eight
DIO 8in/8out (8) outputs (outputs #6–13). Install in one of the motherboard expansion
Board slots J10–J13. Up to two boards may be installed for a total of 16
inputs/16 outputs.
The DIO board provides isolated contact closure inputs and 24-volt
current sinking or current sourcing isolated outputs. Output current must
be limited to 80 mA maximum, continuous.
Board Diagram
Analog I/O The analog I/O board is available in two configurations described below.
Type A has one current output only, whereas, Type B has two voltage
Boards inputs and two current outputs. The Micro-Tech can support up to four
analog inputs and four analog outputs.
Type A: 4–20mA This is an optional board. Install in one of the motherboard expansion
slots J10–J13. Single channel high-level current output.
Output Board
l Outputs
l Rate
l Load
l Optically isolated
l Isolated power source
l Voltage output by adding an internal dropping resistor
Board Diagram
Type B: Analog This is an optional board. Install in one of the motherboard expansion
slots J10–J13.
I/O Board
l Inputs
l Incline compensation
l Moisture compensation
l Outputs
l Rate
l Load
Board Diagram
“Channel 1,” top connector has jumpers J14 and J15 that allow selection
of either external excitation sense (6-wire LC hook-up) or internal
excitation sense (4-wire LC hook-up).
“Channel 2,” bottom connector has jumpers J19 and J20 that allow
selection of either external excitation sense (6-wire LC hook-up) or
internal excitation sense (4-wire LC hook-up).
One of four different interrupt channels can be selected from J1 jumper.
J1 is board interrupt 1-2 this is a factory installed jumper, do not move.
Jumper Locations
Board Diagram
1. Select the jumper positions on the COMM board for the desired
communication standard. Below is a table which summarizes the
jumper positions for selection of the electrical interface. The
jumper locations are shown below.
Jumpers
Mode OP1 OP2 OP3 OP4 OP5 OP6
RS-232 “A” “A” “A” “A” “A” “B”
RS-485* “B” “A” “B” “B” “MDP” “TRM”
20 mA “B” “B” “A” “A” “A” “C”
* Default
“MDP” “TRM”
For RS-485 only For RS-485 only
OP5 OP6
“A” Normal “A” Terminated
“B” Multi-drop “B” Not terminated
2. Open the Micro-Tech wall mount enclosure and turn power off at
the mains, or remove panel mount enclosure from the panel and
remove top cover allowing access to the motherboard.
3. Remove the field mating connector. Wire the connector per the
supplied field-wiring diagram at the end of the manual.
4. Remove the hex head mounting screw from the connector end of
the COMM board.
Board Diagram
LTons The “long ton,” equivalent to 2,240 Tph Tons per hour.
lbs. Totalizer The Totalizer shows the total tons
LTph Long tons per hour. accumulated by the Micro-Tech.
Mixed units A menu choice that allows the
Micro-Tech to display a mixture of
English and metric units.
mV/V Millivolts per volt. A measure of the
sensitivity of a load cell.
pcba Printed-circuit board assembly.
PEIC Periodic-error-integrating control.
Description Document
Field-Wiring Diagrams
Micro-Tech 9104 D07392B-E030
Analog I/O Board B07392B-E003
8-In/8-Out Digital Board B07392B-E005
Serial Communication C07392B-E008
Siemens Profibus Board C07392B-E011
Communication Board C07392B-E017
Notes—Micro-Tech 9000 C07392B-E018
Digital Output Boards C07392B-E021
Anybus Comm for Device Net B07392B-E022
Digital Input Boards B07392B-E025
4–20mA Out Board B07392B-E026
Dual Plant LC A/D Board B07392B-E027