Anda di halaman 1dari 30

HEALTHCARE Maintenance

Global Services Organisation

Order-No.: DD+DIS211.05E

*1YZSZR1*
1 Piece YZSZR MA1
DRYSTAR 5500
Type 5364/100

Edition 1, Revision 8

This documentation is separately available. Order No: DD+DIS211.05E

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.

printed in Germany 2005-09-12

Agfa Company Confidential


Maintenance DD+DIS211.05E

Caution:

This system uses high voltage. Please consider the respective safety regulations.

These instructions describe adjustments and routines which must only be performed by qualified
technical personnel.

Note:

Electrical repairs and connections must only be performed by a qualified electrician.


Mechanical repairs and connections must only be performed by a qualified technician.

CE Declaration:

The CE Declaration (CE Conformity) becomes invalid if the product is changed without explicit
consent of the manufacturer! This applies to all parts, not only to safety elements.

We reserve the right to change data and characteristics in the light of technical
progress.

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 DRYSTAR 5500 Edition 1, Revision 8
Agfa Company Confidential (Type 5364/100) 2005-09-12
DD+DIS211.05E Maintenance

Chapter 9:

List of contents

1 Safety ........................................................................................ 1
2 Introduction.............................................................................. 3
2.1 Maintenance frequency............................................................................ 3
2.2 Required time ........................................................................................... 3
2.3 Required Tools ......................................................................................... 3
2.4 Required cleaning material...................................................................... 4
2.5 Required Spare Parts ............................................................................... 4
3 Diagnostics .............................................................................. 5
3.1 Questioning of the customer................................................................... 5
3.2 Infocounter analysis................................................................................. 5
3.2.1 Infocounter Analysis via floppy and PC ................................................................. 5
3.2.2 Infocounter analysis via browser ........................................................................... 6
3.2.3 How to evaluate the Infocounter............................................................................ 6
3.3 Image quality check ................................................................................. 8
3.4 Extended System check .......................................................................... 9
4 Maintenance points - Power off.............................................. 9
4.1 Check of general condition ................................................................... 10
4.2 Check of condition of belts and gears.................................................. 10
4.3 Check of Drum motor and replacement if necessary.......................... 10
4.4 Visual check of the cables..................................................................... 10
4.5 Visual check of discharge brushes ...................................................... 11
4.6 Replacement of vacuum filters ............................................................. 11
4.7 Cleaning of the suction cups ................................................................ 11
4.8 Cleaning of Air Circulation Paths and Ventilation Openings ............. 12
4.9 Cleaning of Electronics Compartment ................................................. 12
4.10 Cleaning of Power supply...................................................................... 13
4.11 Check and Cleaning of Drum ................................................................ 13
4.12 Cleaning of the cleaning rollers ............................................................ 14
4.13 Cleaning of Thermal print head............................................................. 14
4.14 Electrical check ...................................................................................... 15
4.14.1 Basic Protective Earth Test ................................................................................. 15
4.14.2 Electrical Check according to National Regulations. ........................................... 16

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 8 DRYSTAR 5500 Chapter 9 / I
2005-09-12 (Type 5364/100) Agfa Company Confidential
Maintenance DD+DIS211.05E

5 Maintenance points - Power on ............................................ 16


5.1 Check of safety switches....................................................................... 16
5.2 SW patch installation ............................................................................. 17
5.3 Film calibration....................................................................................... 17
5.4 Density Reproduction Check ................................................................ 17
5.5 Test image evaluation............................................................................ 19
5.6 Host system image quality check ......................................................... 20
6 Completion of the maintenance ........................................... 20
6.1 Backup Data on Floppy ......................................................................... 20
6.2 Reset of the relative counters in the Infocounter ................................ 20
6.3 Customer Conversation......................................................................... 20

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / II DRYSTAR 5500 Edition 1, Revision 8
Agfa Company Confidential (Type 5364/100) 2005-09-12
DD+DIS211.05E Maintenance

1 Safety
⇒ The DRYSTAR 5500 is suitable for continuous operation.
⇒ The DRYSTAR 5500 enables or helps in the diagnosis of a patient.

When operating or maintaining the DRYSTAR 5500 always observe the


following safety guidelines:
• Have electrical or mechanical defects repaired by skilled personnel
only!
Do not override or disconnect the integrated safety features.
• Ventilation openings may not be covered.
• Always switch off the DRYSTAR 5500 and disconnect the power cord
from the outlet before carrying out any maintenance work.
Always take into account the markings provided on the inside and outside of
the printer. A brief overview of these markings and their meaning is given
below:

Safety warning, indicating that the DRYSTAR 5500 User and


Reference manuals should be consulted before making any
connections to other equipment. The use of accessory equipment
not complying with the equivalent safety requirements of this
printer may lead to a reduced level of safety of the resulting
system. Consideration relating to the choice of accessory
equipment shall include:
- Use of the accessory equipment in the patient vicinity.
- Evidence that the safety certification of the accessory
equipment has been performed in accordance with the
appropriate IEC 601-1 and IEC 601-1-2 harmonized national
standard.
In addition, all configurations must comply with the medical
electrical systems standard IEC 601-1-2. The party that makes
the connections acts as system configurator and is responsible
for complying with the systems' standard. If required contact your
Agfa Service Organization.
In order to reduce the risk of electric shock, do not remove any
covers.

Caution hot:
Keep hands clear from the Thermal Print Head.

Type B equipment:
Indicates that the DRYSTAR 5500 complies within the limits for
type B equipment.

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 8 DRYSTAR 5500 Chapter 9 / 1
2005-09-12 (Type 5364/100) Agfa Company Confidential
Maintenance DD+DIS211.05E

Supplementary protective earth connector:


Provides a connection between the DRYSTAR 5500 and the
potential equalization busbar of the electrical system as found in
medical environments. This plug should never be unplugged
before the power is turned off and the power plug has been
removed.
Intergrounding connector:
i Provides a connection between the printer and other equipment
that might exhibit minor ground potential differences. These
differences may degrade the quality of communication between
different equipment. Never remove connections to this terminal.
Protective earth (ground)
Provides a connection between the printer and the protective
earth of the mains. Do not remove this connection, because this
will have a negative influence on the leakage current.
Power/Reset Button
Note that the power cord has to be disconnected from the wall
outlet in order to disconnect the unit entirely from the mains.
Precautions for use in USA only:
Make sure that the circuit is single-phase center-tapped when the
printer is connected to a 240 V/60 Hz source instead of a
120 V/60 Hz source.
Transport after installation
The user has to be very cautious concerning stability, when moving the
DRYSTAR 5500. When doing this, he has to take into account the condition
and the structure of the subsoil, obstructions and slopes. The DRYSTAR
5500 can only be transported with all covers closed. The brakes have to be
loose. The DRYSTAR 5500 may not be transported continuously from one
location to the other.

Waste disposal and environmental regulations


In most countries Drystar film is considered industrial waste and is not
allowed to be disposed of as household waste. Please consult your local
waste disposal regulations. Agfa recommends having Drystar film waste
hauled away by a licensed company.
After its life span, do not dispose of the DRYSTAR 5500 without
consideration of local waste disposal regulations. Please consult your local
service organization.

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 2 DRYSTAR 5500 Edition 1, Revision 8
Agfa Company Confidential (Type 5364/100) 2005-09-12
DD+DIS211.05E Maintenance

2 Introduction
The maintenance steps described in this document are based on the
technically required measures to guarantee the proper operation of the
printer.
The maintenance points have been arranged in a chronologically suitable
order to make the work routines as efficient as possible. The sequence of the
maintenance points in the checklist (see appendix) is identical with the
maintenance instructions.
Additional maintenance steps can be required according to national
regulations.

2.1 Maintenance frequency


The maintenance has to be carried out:
• every 25.000 printed films, or
• every 12 months, whichever comes first.

2.2 Required time


• Time required for the maintenance: ≤ 2 hours

2.3 Required Tools


Description Order Number / Comment
Calibrated densitometer recommended types are
• MacBeth TR-924
• X-Rite 301
• X-Rite 310
• X-Rite 331
• X-Rite 341
• Gretag D200-2
Flash light commercially available
Loopback connectors (1 piece for CM+339977.0
network; 1 piece for RS232 interface)
Magnifying glass 8x CM+9.9579.9904.0
Service PC with Drystar 5500 Density For Service PC specifications refer to
Reproduction Check files chapter 3.2, Tools.
Density reproduction check files are
available via MEDNET, GSO Library
Waterproof pen commercially available
Wrist strap CM+9.9999.0830.0

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 8 DRYSTAR 5500 Chapter 9 / 3
2005-09-12 (Type 5364/100) Agfa Company Confidential
Maintenance DD+DIS211.05E

2.4 Required cleaning material


Cleaning material Order Number
Agfa Cleaning Tissue (as alternative to CM+309931.0
Isopropyl alcohol or Ethanol and Lint
free cloth). Box of 50 pieces.
Ink eraser 'Pelikan' 10+9.9999.1176.0
Grease: Kluber 46MR401 EB+53610820
Isopropyl alcohol or Ethanol Commercially available
Lint free cloth Commercially available
Vacuum cleaner
220/240V, 50Hz CM+9.9999.0895.0
110/115V, 60Hz CM+9.9999.0896.0

2.5 Required Spare Parts


The following spare parts are part of the Maintenance-assortment
CM+0 5364 0999 733

Description Order Number Case of Exchange/ Usage

Discharge Brush CM+309798.X In case of wear

Air Filter CM+339257.X Exchange each maintenance

The following spare parts only have to replaced in case certain signs of wear
are visible:
Description Order Number Case of Exchange/ Usage
Drum Motor CM+326021.X In case 3.46 mm banding
visible in STI2 and medical
image

Always have the latest Drystar 5500 software patches available on the
Service PC.
Depending of the urgency and classification of a software patch it might be
necessary to update the printer software.

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 4 DRYSTAR 5500 Edition 1, Revision 8
Agfa Company Confidential (Type 5364/100) 2005-09-12
DD+DIS211.05E Maintenance

3 Diagnostics
The diagnostics before the maintenance comprise following tasks:
Task Details see..
Questioning of the customer 3.1
Infocounter analysis 3.2
Image quality check 3.3
Extended System check 3.4

3.1 Questioning of the customer


• Ask the customer for any problem that appeared since the last
maintenance.

3.2 Infocounter analysis


The Infocounter can be obtained via floppy drive or browser.
Type of Infocounter analysis: Refer to ...
Infocounter analysis via floppy and PC 3.2.1
Infocounter analysis via browser 3.2.2

3.2.1 Infocounter Analysis via floppy and PC


Required Tools
• Floppy disk
• Service PC, required to analyze the infocounter.
(1) Insert the floppy disk in the printer
floppy drive
(2) In the service menu select
"3 Save data - 1 Infocounter file"
(3) Follow the further instructions on
the display.
(4) Insert the floppy in the Service
PC.
(5) Unzip the file "di55_xxxx_icn.zip"
(xxxx stands for serial number,
date and time). 5364_Reg01_009_e.cdr

(6) Start an editor (e.g. Notepad or Figure 1


Wordpad).
(7) Open the file "infocounters.txt"
(8) Analyze the infocounter file as described in section 3.2.3,
How to evaluate the Infocounter

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 8 DRYSTAR 5500 Chapter 9 / 5
2005-09-12 (Type 5364/100) Agfa Company Confidential
Maintenance DD+DIS211.05E

3.2.2 Infocounter analysis via browser

Required Tools
• Service PC, required to analyze the infocounter.

BROWSER (1) Connect the Service PC to the Service PC


Printer
network
(2) Start the browser (Internet
Explorer ≥ 5)
(3) Enter the IP address of the printer
in the address window.
In case the printer homepage is
not displayed, refer to Technical
536432el.cdr
Documentation section 3.2
(4) Go to "Service Engineer Tools -
Infocounters Report"
(5) Analyze the infocounter file as
Figure 2
described in section 3.2.3,
How to evaluate the Infocounter

3.2.3 How to evaluate the Infocounter

Infocounters.txt evaluation
What to check in the Comment
infocounter
1.1 Serial number and Compare device serial number with chapter 8
Installation date ‘Manufacturing Standard Modification’ to
determine whether the device possibly requires a
modification.
Refer to 3.3 (infocounter.txt), ‘HW Modification
History’, to check already made modifications.
1.4 Software Info Compare the currently installed SW version with
the latest SW version described in chapter 3.8,
SW releases, patches. In case a recommended
SW patch is available, install this SW patch.
Refer to 5.2.

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 6 DRYSTAR 5500 Edition 1, Revision 8
Agfa Company Confidential (Type 5364/100) 2005-09-12
DD+DIS211.05E Maintenance

Infocounters.txt evaluation
What to check in the Comment
infocounter
2.3 Reliability Should normally be not higher than 0.05. Higher
fault rates need a thorough investigation.
3.3 Hardware Modification By comparing the status of the device with the
History available modifications (refer to Technical
Documentation chapter 8) the exact hardware
status can be determined.
3.4 Software modification By checking the software modification history it
history can be determined, whether a recent software
upgrade solved a problem, that occurred quite
often in the error list in the past
5.3 Retries Many retries (> 1%, based on relative prints)
have to be investigated: They may reduce the
throughput of the device.
• Retries film pickup units: Check rollers /
check mechanics / check vacuum system
• Retries communication MEC: Check cabling
between MEC board and CPU-Board.
• Retries communication ETH (Ethernet):
Check network cabling, hub.
5.7 Error History Check which errors occurred in the last days,
how often they appeared as well as the CBF
(cycles between failures) of these errors. This
gives an overview of the current status of the
machine.
5.8 Error list (sorted by Troubleshoot these errors with the help of the
relative) and technical documentation, chapter 3.3,
5.9 Error list (sorted by ‘Troubleshooting’.
total):

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 8 DRYSTAR 5500 Chapter 9 / 7
2005-09-12 (Type 5364/100) Agfa Company Confidential
Maintenance DD+DIS211.05E

3.3 Image quality check

(1) In the service menu select '8


Adjust & Calibrate - 4 Print
Testimage"
(2) Print following test images on the
largest format available in the film
magazines:
• STI_5500.TIF
• STI2_5500.TIF
• medical_image.tif

The test image as displayed


in Figure 3 to Figure 5 will be
printed.

(3) Mark the test films with a


waterproof pen as testfilm before
maintenance
Figure 3

Figure 4 Figure 5

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 8 DRYSTAR 5500 Edition 1, Revision 8
Agfa Company Confidential (Type 5364/100) 2005-09-12
DD+DIS211.05E Maintenance

3.4 Extended System check


Required Tools
• Loopback connectors (1 piece for network; 1 piece for RS232 interface)
• Technical Documentation chapter 4, Reference Diagrams

(1) In the service menu select "2 Maintenance - 1 Quality Checks - 2


Extended System Check"
(2) Follow the further instructions on the local display.

Note: The Extended System Check takes approx. 15 minutes.

4 Maintenance points - Power off


The maintenance with powered-off machine comprises following tasks:
Task Details: see...
Check of general condition 4.1
Check of condition of belts and gears 4.2
Check of Drum motor and replacement if necessary 4.3
Visual check of the cables 4.4
Visual check of discharge brushes 4.5
Replacement of vacuum filters 4.6
Cleaning of the suction cups 4.7
Cleaning of Air Circulation Paths and Ventilation 4.8
Openings
Cleaning of Electronics 4.9
Cleaning of Power supply 4.10
Check and Cleaning of Drum 4.11
Cleaning of the cleaning rollers 4.12
Cleaning of Thermal print head 4.13
Electrical check 4.14

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 8 DRYSTAR 5500 Chapter 9 / 9
2005-09-12 (Type 5364/100) Agfa Company Confidential
Maintenance DD+DIS211.05E

4.1 Check of general condition


(1) Perform a visual check of
• Power- and network cables.
• Covers and the Front door.
• The Output trays.
• The upper and lower Input tray.
(2) Disconnect the power and network cable
(3) Remove all covers to prepare the printer for the next steps
(4) Remove the films in the film trays including white plastic sheet

4.2 Check of condition of belts and gears


(1) Check all belts and gears for visible wear

4.3 Check of Drum motor and replacement if necessary


(1) Check the flatfield area in the
STI2 test image printed previously
under step "3.3 Image quality
check" for banding lines with a
distance of 3.46 mm.
(2) Check drum motor for black dust
and radial play
(3) Exchange the drum motor in
following cases:
• In case 3.46 mm banding is
visible, and this banding is also
visible in the medical_image
printed under step "3.3 Image
quality check" or
• In case black dust is visible (in
case it also has radial play this
will be visible as 3.46 mm
banding)
Figure 6

4.4 Visual check of the cables


(1) Check cables inside the printer: The insulation may not show any sign of
wear or damage.

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 10 DRYSTAR 5500 Edition 1, Revision 8
Agfa Company Confidential (Type 5364/100) 2005-09-12
DD+DIS211.05E Maintenance

4.5 Visual check of discharge brushes


(1) Check discharge brushes at pickup frame, at cleaning module and sorter
for wear. Replace if required.

4.6 Replacement of vacuum filters


(1) Replace both vacuum filters.
Observe correct mounting
direction.

Long side

536409fg.cdr

Short side

Figure 7

4.7 Cleaning of the suction cups


(1) Clean the suction cups of both
film pickup assemblies with a lint
free cloth lightly moistened with
Isopropyl alcohol: This cleaning
action can be made from the
printer right hand side.

536409cc.cdr

Figure 8

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 8 DRYSTAR 5500 Chapter 9 / 11
2005-09-12 (Type 5364/100) Agfa Company Confidential
Maintenance DD+DIS211.05E

4.8 Cleaning of Air Circulation Paths and Ventilation Openings


(1) Clean the ventilation openings at
the rear cover with a vacuum
cleaner
(2) Clean the air circulation paths
with a vacuum cleaner.

Figure 9

4.9 Cleaning of Electronics Compartment


(3) Remove the EMI shielding
(4) Clean the electronics
compartment with a vacuum
cleaner.

Figure 10

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 12 DRYSTAR 5500 Edition 1, Revision 8
Agfa Company Confidential (Type 5364/100) 2005-09-12
DD+DIS211.05E Maintenance

4.10 Cleaning of Power supply


(1) Remove the power supply
(2) Clean the power supply with a
vacuum cleaner.
(3) Remount the power supply

Figure 11

4.11 Check and Cleaning of Drum


(1) Open the drum compartment door
(2) Remove the flat belt
(3) Check the drum for dirt or
damage
(4) In case signs of damage are
visible, also check the STI2 test
image printed under step "3.3
Image quality check": A damage
will be visible in the flatfield area
of the STI2 test image in distance
of 111 mm.
(5) Clean the rubber of the drum with
a lint free cloth lightly moistened
with Isopropyl alcohol or Ethanol
(6) Remount the flat belt Figure 12

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 8 DRYSTAR 5500 Chapter 9 / 13
2005-09-12 (Type 5364/100) Agfa Company Confidential
Maintenance DD+DIS211.05E

4.12 Cleaning of the cleaning rollers


(1) Open the cleaning rollers module
(2) Clean both rollers with an Agfa
cleaning tissue or a lint free cloth
lightly soaked in Isopropyl
alcohol:
• Turn the rollers by hand
• Clean in length direction of
the rollers

Figure 13

4.13 Cleaning of Thermal print head


(1) Clean the TH resistor line with a
lint free cloth lightly moistened Resistor
with Isopropyl - alcohol or a Line
cleaning tissue.
(2) Remount all covers IC cover
• The last cleaning action must
be carried out in one way!
• For the Isopropyl alcohol
observe the rule: As less water
is in the alcohol as better. We
recommend using a 90% 536409ci.cdr

Isopropyl alcohol.
Figure 14

• Be carefull at TH cleaning.
• The electronic circuits under the IC cover can be damaged.
• During cleaning do not press against the IC cover.
Only use very little alcohol to moisten the cloth.

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 14 DRYSTAR 5500 Edition 1, Revision 8
Agfa Company Confidential (Type 5364/100) 2005-09-12
DD+DIS211.05E Maintenance

4.14 Electrical check


The electrical check is split up in two steps:

Step Task Details see..


1 Perform the basic protective earth test 4.14.1
2 Perform the electrical check according to 4.14.2
National Regulations.

4.14.1 Basic Protective Earth Test


As the power supply has been dismounted for cleaning, the protective earth
resistance has to be checked.
(1) Use a multimeter that is
capable to measure a
200
resistance < 1 Ohm
(2) Put one probe at the ground
connector of the mains input
(3) Put the other probe at a few
metallic parts, the customer
could touch.
The electrical resistance has to
be ≤ 0.2 Ohm. 536409ck.cdr
Refer to Figure 15.
Figure 15
(4) Insert the power cord at the
printer side; leave it
disconnected at the wall side.
(5) Put one probe at the ground
connector of the mains
connector.
(6) Put the other probe at a few
metallic parts, the customer
could touch.
536409cl.cdr
The electrical resistance in this
case has to be ≤ 0.3 Ohm.
Refer to Figure 16. Figure 16

In case the electrical resistance of one measurement exceeds the limits (0.2
Ohm without mains cable; 0.3 Ohm with mains cable) check the printer for
loose grounding connections!

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 8 DRYSTAR 5500 Chapter 9 / 15
2005-09-12 (Type 5364/100) Agfa Company Confidential
Maintenance DD+DIS211.05E

4.14.2 Electrical Check according to National Regulations.


(1) Perform the electrical check according to National Regulations.

The electrical checks that have to be performed in certain intervals (e.g. each
12 months) or after certain service actions (repair and/or maintenance)
depend on National Regulations.
Example for Germany: BGV A3 and DIN VDE 0751 or subsequent
regulations.

5 Maintenance points - Power on


The maintenance with powered-on machine comprises following tasks:
Task Details: see...
Check of safety switches 5.1
SW patch installation 5.2
Film calibration 5.3
Density Reproduction Check 5.4
Test image evaluation 5.5
Host system image quality check 5.6

5.1 Check of safety switches


(1) Open front door, drum
compartment door and top door to
check function of safety switches

Figure 17

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 16 DRYSTAR 5500 Edition 1, Revision 8
Agfa Company Confidential (Type 5364/100) 2005-09-12
DD+DIS211.05E Maintenance

5.2 SW patch installation


(1) Install the latest SW patch, in case a recommended SW patch for the
printer is available. Refer to chapter 3.8, Software Releases, Patches.

5.3 Film calibration


How? (1) Select in service menu '8 Adjust
& Calibrate - 3 Calibrate - 1 Film'
(2) Follow the instructions on the
display.
(3) Repeat film calibration for
second input tray

536336as.cdr

Figure 18

5.4 Density Reproduction Check


Background With the density reproduction check the difference for a specific film type
Information between optical density defined in the sensitometric calibration file and the
printed optical density, measured with an external densitometer is checked.
This is a plausibility check for the MDM: In case the measured density values
are out of the specified range, the MDM requires a re-calibration.

This procedure must be carried out with every used film type:
(1) Print out test image SMPTE.tif and measure the density of the eleven
100 % – 0% squares with a calibrated MacBeth TR-924, X-Rite 331,
341, X-Rite 301, X-Rite 310 or Gretag D200-2 densitometer.
(2) To check linearity fill in measured values in corresponding Excel file:
Ds5500 density reproduction check DT2B.xls or
Ds5500 density reproduction check DT2C.xls

These Excel files can be downloaded from MedNet, download area


'GSO library - Hardcopy - Dry Imagers - Drystar 5500 - Freeware'.

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 8 DRYSTAR 5500 Chapter 9 / 17
2005-09-12 (Type 5364/100) Agfa Company Confidential
Maintenance DD+DIS211.05E

(3) After the measured values are filled in, the density reproduction is
calculated automatically.
(4) At the right hand side of the table is shown, whether the Density
Reproduction is ok or not. In case it is not ok you are requested to
carry out a MDM calibration.

Select densitometer: X-rite 341


4 Dmax = 314

Square Measured density Target density Density diff. Check


100% 20 20 0 0
90% 47 49,4 -2,4 0
80% 73 78,8 -5,8 0
70% 104 108,2 -4,2 0
60% 133 137,6 -4,6 0
Density Reproduction = OK
50%
40%
163
189
167
196,4
-4
-7,4
0
0
No MDM calibration needed for DT2B film
30% 218 225,8 -7,8 0
20% 246 255,2 -9,2 0
10% 274 284,6 -10,6 0
0% 306 314 -8 0
When entering the measured density values the decimal point appears in a different spot,
e.g. measured value = 0.23, value to enter = 23.0
Density Lower limit Upper limit
0 0 0 Density difference SMPTE image DT2B film in Ds 5500
30 -5 5
45 -7 7
20
60 -8 8
75 -9 9 15
90 -10 10 X-rite 310
105 -11 11 10 X-rite 331
Density Difference

120 -12 12 X-rite 341


135 -12 12 5
X-rite 301 Lower limit
150 -13 13 Mac Beth TR924
0 Upper limit
165 -13 13 Gretag D200-2
180 -14 14 Density difference
-5
195 -14 14
210 -14 14 -10
225 -15 15
240 -15 15 -15
255 -15 15
-20
270 -16 16
0 50 100 150 200 250 300 350
285 -16 16
300 -16 16 Target Density
315 -16 16
536409af.cdr

Figure 19

(1) The measured density values have to be entered without decimal point.
(e.g. measured value = .23, value to enter = 23)

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 18 DRYSTAR 5500 Edition 1, Revision 8
Agfa Company Confidential (Type 5364/100) 2005-09-12
DD+DIS211.05E Maintenance

5.5 Test image evaluation


(1) In the service menu select '8 Adjust & Calibrate - 4 Print Testimage"
(2) Print following test images on the largest format available in the film
trays:
• STI_5500.TIF
• STI2_5500.TIF
• medical_image.tif

(3) Mark the test films with a waterproof pen as testfilms after maintenance
(4) Check all three images at the lightbox. Following items must been
checked:
o Line artifacts in transport direction
o Line artifacts vertical to the transport direction
o Point artifacts
o Surface artifacts

(5) Evaluate the test images at follows:


1. If no artifacts are seen on the STI / STI2 image continue with next
step
2. If artifacts seen on STI / STI2 image aren’t visible on the
medical_image continue with next step.
3. In other case (i.e. artifacts which are visible in STI / STI2 are
visible on medical_image, too) refer to Technical Documentation
chapter "3.3 Troubleshooting”

(6) Compare the test images "after maintenance" with the test images made
"before maintenance": The image quality of the last printed
medical_image and STI/STI2 test image must be equal or better than the
images made before maintenance.

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 8 DRYSTAR 5500 Chapter 9 / 19
2005-09-12 (Type 5364/100) Agfa Company Confidential
Maintenance DD+DIS211.05E

5.6 Host system image quality check


(1) Print a diagnostic image from every host system.
(2) Check images of each host at the lightbox for artifacts in diagnostic
image.
(3) In case artifacts are visible, compare with artifacts in last STI /STI2
image.
o In case artifacts are also visible in STI / STI2 image, refer to
Technical Documentation chapter "3.3 Troubleshooting”.
o In case artifact is only visible in images of a specific host but not in
STI / STI2 image, initiate a service call of modality service engineer

6 Completion of the maintenance


The Completion of the maintenance comprises following tasks:
Task Details see..
Backup Data on Floppy 6.1
Reset of the relative counters in the Infocounter 6.2
Customer Conversation 6.3

6.1 Backup Data on Floppy


(1) In the Key-operator menu select "5 Save configuration".
(2) Follow the instructions on the printer display.

6.2 Reset of the relative counters in the Infocounter


(1) In the Service menu select "2 Maintenance - 2 Clear relative counter".
(2) When leaving the Service menu select "Preventive maintenance" as
service action.

6.3 Customer Conversation


(1) Fill out the Maintenance checklist and hand it out to the customer.
(2) Explain the results of the maintenance to the customer.

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 20 DRYSTAR 5500 Edition 1, Revision 8
Agfa Company Confidential (Type 5364/100) 2005-09-12
AGFA and the Agfa-Rhombus are trademarks of Agfa-Gevaert AG, Germany

Herausgegeben von / Published by / Edité par /

Agfa-Gevaert AG
Fototechnik
Tegernseer Landstraße 161
D - 81539 München

Printed in Germany / Imprimé en Allemagne


HEALTHCARE Maintenance
Global Services Organisation Checklist
Order-No.: DD+DIS211.05E

DRYSTAR 5500
Type 5364/100

Work Instruction for order no. SN Film cycles

The maintenance has to be carried out:

Every 12 months or 25 000 printed films

Maintenance must be carried out according to the maintenance instructions DD+DIS211.05E.

Diagnostics not
OK OK
Questioning of the • Questioning of the customer about problems with
customer the printer.
Infocounter analysis • Read-out and analyzation of the infocounter
Image quality check • Printing of STI, STI2 and medical test image for
comparison with test images at final system check
Extended system • Execution of service menu "Extended system
check check".

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.

printed in Germany 2005-09-12

Agfa Company Confidential


Maintenance Checklist DD+DIS211.05E

Maintenance points - Power off not


OK OK
Check of general • Visual check of power and network cable.
condition • Visual check of covers and film magazines
Check of condition of • Visual check of all belts and gears in the printer.
belts and gears
Check of Drum motor • Replacement of drum drive motor in case 3.46 mm
and replacement if banding visible in STI2 and medical image
necessary
Visual check of the • Check of cables inside the printer for damage
cables
Visual check of • Check of three discharge brushes for wear and
discharge brushes replacement if necessary
Replacement of • Replacement of both vacuum filters
vacuum filters
Cleaning of the suction • Cleaning of the suction cups of both film pickup
cups assemblies with a lint free cloth lightly moistened
with Isopropyl alcohol
Cleaning of Air • Cleaning of Air Circulation Paths and Ventilation
Circulation Paths Openings with a vacuum cleaner.
Cleaning of Electronics • Cleaning of the Electronic rack with a vacuum
rack cleaner.
Cleaning of Power • Cleaning of the Power supply with a vacuum
supply cleaner
Check and Cleaning of • Check of the rubber of the drum for damage.
Drum • Cleaning of the rubber of the drum with a lint free
cloth lightly moistened with Isopropyl alcohol or
Ethanol.
Cleaning of the • Cleaning of the cleaning rollers with a lint free
cleaning rollers cloth lightly moistened with Isopropyl alcohol
Cleaning of Thermal • Cleaning of the Thermal print head with a lint free
print head cloth lightly moistened with Isopropyl alcohol
Electrical check • Check of the protective earth resistance

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Maintenence Checklist Page 2 / 4 DRYSTAR 5500 Edition 1, Revision 8
Agfa Company Confidential (Type 5364/100) 2005-09-12
DD+DIS211.05E Maintenance Checklist

Maintenance points - Power on not


OK OK
Check of safety • Opening of front door, drum compartment door
switches and top door to check function of safety switches
SW patch installation • Installation of SW patch, in case a recommended
SW patch for the printer is available
Film calibration • Calibration of the films in both input trays.
Density Reproduction • Density Reproduction Check to confirm proper
Check operating MDM. Execution of MDM calibration if
not okay.
Test image evaluation • Printing of STI , STI2 and medical test image
• Comparison with test images made during printer
diagnostics
• Check for artifacts on the film. In case of artifacts
in STI / STI2 and medical test image
troubleshooting as described in chapter 3.3
Troubleshooting.
Host system image • Printing of diagnostic image from every host
quality check system
• Check concerning artifacts in diagnostic image.
Comparison with artifacts in last STI /STI2 image.
o In case artifacts are also visible in STI / STI2
image, troubleshooting as described in chapter
3.3 Troubleshooting.
o In case artifact is only visible in images of a
specific host but not in STI / STI2 image,
initiation of service call of modality service
engineer

Completion not
OK OK
Backup • Backup of machine specific parameters
Reset of the relative • Reset of relative infocounter
infocounter • Confirmation of the maintenance in the printer
service menu.
Customer conversation • Explanation of results of preventive maintenance.

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 8 DRYSTAR 5500 Maintenence Checklist Page 3 / 4
2005-09-12 (Type 5364/100) Agfa Company Confidential
Maintenance Checklist DD+DIS211.05E

Remarks:

...................................................................................................................................................

...................................................................................................................................................

...................................................................................................................................................

...................................................................................................................................................

...................................................................................................................................................

...................................................................................................................................................

...................................................................................................................................................

................................................................... ...................................................................
Date / Signature Service Technician Customer

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Maintenence Checklist Page 4 / 4 DRYSTAR 5500 Edition 1, Revision 8
Agfa Company Confidential (Type 5364/100) 2005-09-12

Anda mungkin juga menyukai