Hematology Analyzer
AV5
User’s Manual
1.0 Release
Abacus Vet5 software v1.1
2|Page
Abacus Vet5 User’s Manual
This user’s manual is intended to give detailed information for end users and operators of the Abacus Vet5
hematology analyzer.
Warranty
THE WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Diatron’s obligation or liability under this warranty does not include any transportation or other charges or
liability for direct, indirect or consequential damages or delay resulting from the improper use or application
of the product or the use of parts or accessories not approved by Diatron or repairs by people other than
Diatron authorized personnel.
This warranty shall not extend to:
any Diatron product which has been subjected to misuse, negligence or accident.
any Diatron product from which Diatron’s original serial number tag or product identification
markings have been altered or removed.
any product of any other manufacturer.
Diatron’s obligation or liability does not include any charges or liability for direct, indirect or consequential
damages resulting from any language translation of this manual which is taken as the reference document of
the Abacus Vet5 hematology analyzer.
Company contact
Manufacturer: DIATRON MI ZRT
Address: Táblás utca 39.
Budapest, HUNGARY
H-1097
Contacts: Tel: +36-1-436 9800
Fax: +36-1-436 9809
www.diatron.com
support@diatron.com
3|Page
Abacus Vet5 User’s Manual
Table of Contents
1 Indications for use ............................................................................................. 11
2 For Your Safety.................................................................................................... 12
Who Should Use This Manual .....................................................................................................................12
Special Symbols Used In This Manual and on the Analyzer ............................................................12
General Precautions........................................................................................................................................12
Environmental Requirements ....................................................................................................................14
Electrical Requirements ................................................................................................................................14
Space Requirements .......................................................................................................................................15
Weight Requirements ....................................................................................................................................15
Waste Disposal..................................................................................................................................................16
Known Limitations ..........................................................................................................................................16
Emergency Situations ................................................................................................................................16
3 Product Support ................................................................................................. 17
4 Installation ........................................................................................................... 18
Abacus Vet5 Analyzer Package...................................................................................................................18
Accessories .........................................................................................................................................................18
Installation check-list .....................................................................................................................................19
Select a suitable location ....................................................................................................................................... 19
Connecting reagents ................................................................................................................................................ 19
Connect power ........................................................................................................................................................... 20
Peripherals .................................................................................................................................................................. 20
4|Page
Abacus Vet5 User’s Manual
5|Page
Abacus Vet5 User’s Manual
10 Database Operations......................................................................................... 67
6|Page
Abacus Vet5 User’s Manual
7|Page
Abacus Vet5 User’s Manual
8|Page
Abacus Vet5 User’s Manual
9|Page
Abacus Vet5 User’s Manual
10 | P a g e
Abacus Vet5 User’s Manual
Depending on working mode the analyzer takes 25 or 50l of K2 or K3-EDTA anti-coagulated whole blood
from an open sample vial placed into an adapter in the sample rotor station. It can process 24-30 samples
per hour in 3-part mode, 16-20 samples per hour in 5-part mode.
Abacus Vet5 is indicated for use in identifying patients with hematologic parameters within and outside of
established reference (normal) ranges.
11 | P a g e
Abacus Vet5 User’s Manual
Sharp needle warning The sampling needle may be a hazard to the operator.
General Precautions
Please, read this section carefully before operating the analyzer.
The sampling needle and other components inside the analyzer may cause injury, or can
get damaged if handled incorrectly. Only certified personnel should open the covers.
Running samples with opened cover is not recommended due to the risk of possible
injury. Always wear safety gloves while performing maintenance actions.
Do not execute any actions with the device when the side door or the back door is open.
The moving inner parts can cause injury.
Make sure to retain the original packaging material for safe transportation and storage in
the future.
To prepare the analyzer for shipping, storage or extended periods of inactivity, please
drain the reagents (run Preparing for shipment procedure) and repackage it in its original
packaging. Do not expose the Abacus Vet5 to direct sunlight, extreme temperature or
humidity over 90%.
12 | P a g e
Abacus Vet5 User’s Manual
The analyzer operates with chemically and biologically active reagents. Physical contact
with these reagents should be avoided. Please read reagent descriptions and MSDS
carefully for possible emergency actions.
Only Diatron certified service personnel that have successfully completed the Abacus
Vet5 Service Training program are qualified to service the Abacus Vet5 analyzer.
Before operating the Abacus Vet5 analyzer, all operators should complete an Abacus Vet5
Operator Training program. This program is offered by certified service personnel of
Diatron.
Replacement materials or spare parts (tubes, valves, etc.) which might have been in
contact with animal blood or reagents should be handled as a potentially biologically
hazardous and chemically dangerous material. All applicable laws and regulations must
be observed in the handling and disposal of these materials.
Abacus Vet5 is designed for laboratory operation. Mobile operation is not supported.
Operate it within the ambient temperature range described in section 2.4.
This IVD equipment complies with the emission and immunity requirements described in
relevant part of the IEC 61326 series.
This equipment has been designed and tested to CISPR 11 Class A. In a domestic
environment it may cause radio interference, in which case, you may need to take
measures to mitigate the interference.
Waste contains poisonous substances (because of chemical content) and animal origin
substances meaning biohazard. These substances are representing potential danger to
environment. For this reason, safe handling of the waste liquid is very important.
Based on the performed risk analysis, it’s guaranteed that the trained user doesn’t need
further operator action in case of residual risk as long as the operator works according to
the user manual and takes care of the warning labels.
The safety status of the device will not change (will not be worse) after maintenance or service repair.
13 | P a g e
Abacus Vet5 User’s Manual
Environmental Requirements
Operate Abacus Vet5 analyzer within the ambient temperature range of 15-30°C and a relative humidity
range less than 85%. The optimum operating temperature is between 20-25°C .
Abacus Vet5 should be stored within the temperature range of 5-35°C. Avoid exposing it to direct sunlight or
to extreme high or low temperatures. If the analyzer was subjected to extreme temperatures during
shipment or storage, the analyzer must be placed for at least one hour in a room whose temperature is within
the operational range before installation or use.
Reagents should be stored at a temperature range of 15-30°C. Reagents may experience a temperature range
of at most 5-35°C for a maximum of 3 days.
Analyzer operation at an altitude above 2000 meters (6560 feet) is prohibited.
Electrical Requirements
The analyzer should only be operated from a wall outlet meeting these power input requirements:
The analyzer MUST be operated from its power supply module (GT-41069P9012-T3) packaged
into the accessory box
Input voltage: 100-240VAC; 50Hz to 60Hz, max. 78 VA
The analyzer must be earth grounded
Abacus Vet5 analyzer comes with a power cord appropriate for your power system. Use of the appropriate
power cord assures adequate grounding of the system. If the power is not reliable, contact your
representative for options such as installation of an external UPS (uninterruptible power supply) module.
Failure to properly ground Abacus Vet5 bypasses important safety features and may
result in electrical hazard.
The instrument should not be placed near potentially interfering devices capable of emitting radio
frequencies (e.g. radio or television transmitters, radars, centrifuges, X-ray devices, fans, etc.).
This analyzer is designed to be safe for transient voltages to INSTALLATION CATEGORY II and POLLUTION
DEGREE 2.
14 | P a g e
Abacus Vet5 User’s Manual
Space Requirements
It is important to install the instrument in a suitable location. A poor location can adversely affect its
performance. Select a well-ventilated location near a power source and close to a suitable drain.
Place the unit on a clean, level surface. Leave at least 0.5 m (18 inches) space on both sides and above the
instrument to access pneumatics and the back easily. A minimum of 0.2 m (8 inches) must be maintained
between the rear panel and the wall to allow for heat dissipation and tubing clearance.
2 L diluent
C R waste
Placing reagents above the analyzer could result in reagent overflow and spilling.
Weight Requirements
Please select a table, laboratory shelf, or other location, which can support the weight of Abacus Vet5 with
accessories, and is free from vibration from other devices.
To allow reliable operation and to provide a safe working environment, make sure that
the table supporting the unit is stable enough to carry the weight of the instrument and
accessories.
15 | P a g e
Abacus Vet5 User’s Manual
Waste Disposal
Abacus Vet5 analyzer waste contains animal blood and reagents that are chemically and biologically active,
and should be considered to be potential infection and biohazard threat. Safe laboratory practices must be
followed including the use of personal protective when operating the Abacus Vet5 and handling blood,
reagents, and waste.
Known Limitations
Abacus Vet5 is intended for analysis of animal blood samples. Anti-coagulated and homogenized (well-
mixed) animal blood samples must be free from contamination and clots.
Blood samples should be analyzed within 7 hours of venipuncture, otherwise precision is not guaranteed,
especially on 5-part differential WBC parameters.
Emergency Situations
Always follow all applicable laws and regulations in regards to emergency situations.
The following list provides some examples when emergency shutdown is needed:
Personal injury or risk of personal injury
Fire or flood in the laboratory, or wrong reagent is connected to Abacus Vet5
Risk of high voltage on the electric grid so that Abacus Vet5 can sustain permanent damage
In emergency-situation, cut off power by disconnecting mains power cable, or turning the main switch off. In
case of electric risk, cut off power from external power supply unit, do not use the main switch.
Use a fire extinguisher if necessary.
In case of fire, do not use water to extinguish the fire unless the Abacus Vet5 is
disconnected from the electrical power network.
After emergency power off, be cautious when reaching into the unit through the side door
or the back door, since the sampling needle may be in a position that it can cause injury.
16 | P a g e
Abacus Vet5 User’s Manual
3 Product Support
Your sales representative is trained and qualified to answer questions about applications and the operation
of the Abacus Vet5 analyzer. If you have additional questions regarding the Abacus Vet5, please ask for
manufacturer support through your sales representative.
Diatron makes every effort to provide excellent support and up to date information and services regarding
the Abacus Vet5 analyzer. Software updates and application documents are available on the Diatron web site
at http://www.diatron.com.
17 | P a g e
Abacus Vet5 User’s Manual
4 Installation
Your Diatron certified service engineer will perform the initial installation of Abacus Vet5 analyzer and train
selected laboratory personnel on the proper use and maintenance of it.
The analyzer should be operated by properly trained laboratory personnel only.
Accessories
The accessories box of Abacus Vet5 contains the following items:
User’s Manual (this document)
Tube kit in a plastic bag, 6 caps with dropdown tubes and 6
color-coded tubes:
o Diluent tube (green)
o Lyse tube (yellow)
o Cleaner tube (blue)
o Lyse2 tube(orange)
o Rinse tube(white)
o Waste tube (red)
18 | P a g e
Abacus Vet5 User’s Manual
Installation check-list
Abacus Vet5 is a complex veterinary device. Follow the checklist below for important information. Always
handle the instrument with care. Refer to Chapter 2 for Safety and other requirements.
Connecting reagents
Abacus Vet5 is capable of aspirating reagents not more than 1m below the analyzer’s base. Follow
suggestions detailed at 2.6
Step1. Open the Diatro Vet5 reagent pack (DRP02HK) box. Open each reagent container (lyse, lyse 2, rinse,
cleaner, and diluent) by unscrewing the container caps and piercing the bottle top seals.
Step 2. Open the plastic bag located in the accessory box, and take
out the six color-coded reagent tubes and caps with drop-down
tubes.
Step 3. Re-cap each container using the enclosed caps with drop
down tubes.
Do not block the Vent hole
small air vent hole on the container caps to provide air flow.
Note: Be sure to save the original reagent caps so you can re-
cap the containers when the reagent pack is used up.
19 | P a g e
Abacus Vet5 User’s Manual
Green = Diluent
Yellow = Lyse
Blue = Cleaner
Red = Waste
Orange = Lyse 2
White = Rinse
Connect the reagent bottles one by one to analyzer with matching color pickup tubes to reagent inlets. The
pickup tube must fully push onto connector, both analyzer and bottle side.
Install the waste vacutainer. Place the waste container and reagent pack below the analyzer, or on the same
level.
Before starting the analyzer:
Be sure once again that the color on each tube, cap and connector in the back of the instrument
match.
Before putting the drop-down tubes into the reagent containers, wipe the dust off from them using
a lint-free wet paper wipe.
Connect power
Make sure that the main switch (next to the power connector at rear) is in ‘0’ position;
Connect the DC cable of the external power supply module to the Abacus Vet5 power input.
Tighten the connector by turning the securing screw (2 – 3 turns).
Connect the power cord to the external power supply;
Plug-in the power cord into the wall-outlet. See section 2.5 for electrical requirements.
Connect keyboard and LAN (Ethernet) cables to the appropriate connectors (optional);
Power up the Abacus Vet5 by flipping the power switch to on ‘I’ position;
Check the operational status of the peripherals and the operation of touch screen.
Peripherals
Abacus Vet5 is capable of connecting to the following peripherals:
Various HP printers directly to USB port or printers through optional PrinterBox® device
Keyboard with USB connector
Bar-code readers with USB connector
USB memory devices up to 16 GBytes in capacity (such as a USB pen drive)
20 | P a g e
Abacus Vet5 User’s Manual
Host computer via USB link connected to USB B port on the back. (To utilize this USB connection, you
need a USB cable, and a suitable driver should be installed on the computer. Refer to your service
personnel for instructions.)
Host computer or local area network (LAN) via Ethernet link.
As a general rule: if the peripheral you want to connect has its own power source / power
supply then turn off both the peripheral device to be connected and the Abacus Vet5
instrument as well before connecting the peripheral device.
The peripheral connectors on Abacus Vet5 are SELV (safety extra low voltage) connectors.
Only connect approved external devices that are SELV rated to the instrument to avoid
the risk of electrical shock.
21 | P a g e
Abacus Vet5 User’s Manual
General Overview
Abacus Vet5 is a fully automated, high quality hematology analyzer for use in veterinary laboratories. It
provides precise and accurate 5-part (or in certain cases 3-part) WBC differential measurement results using
impedance measurement technology.
Abacus Vet5 analyzer uses the volumetric impedance method for measuring of leukocytes (WBC),
erythrocytes (RBC) and platelet (PLT) concentrations. Measurement of the hemoglobin (HGB) content of red
blood cells is accomplished by photometric measurement technology. 5-part WBC differential accomplished
by using a differential lysing reagent.
A vivid color touch-screen display is featured with an intuitive, informative, and attractive user interface.
Abacus Vet5 analyzer’s unique software system supports the use of PCL5 or PCL 3.2 compatible external
printers of HP brand with USB connections. The internal database is capable of storing 5,000 patient results,
separate QC, and calibration result records including flags and graphical histograms. To ensure up to date
operation the analyzer software can be upgraded on the field.
Abacus Vet5 offers Ethernet LIS connectivity using the HL7 protocol beside the standard serial interface to
communicate the laboratory informatics system.
Measured Parameters
Abacus Vet5 determines the following blood parameters, including 5-part WBC differential.
22 | P a g e
Abacus Vet5 User’s Manual
23 | P a g e
Abacus Vet5 User’s Manual
Genuine Reagents
For operating Abacus Vet5 analyzer use the following reagent pack only. All these reagents are for in-vitro
use only. Reagent storage temperature: 15-30 °C.
If eyes or skin gets contact with any of the reagents, flush abundantly with water.
If a large quantity ingested, drink water immediately and contact a doctor.
MSDS (Material Safety Data Sheet) of these reagents can be downloaded from Diatron’s
website: www.diatron.com
Performance and lifespan of the Abacus Vet5 are guaranteed only if reagents supplied
by the manufacturer are used.
The reagent pack (DRP02HK) consists of bottles containing diluent, cleaner, lyse, lyse 2, and rinse
solutions. The following table lists the reagents volumes and bottle sizes in the reagent pack.
Color code
Reagent Description Volume
on cap
Diluent Green This is an isotonic, filtered saline solution, used to dilute whole 9 liter
blood samples and rinse the fluidic system between measuring
cycles.
It is connected to a dedicated diluent input at the rear of the
analyzer.
Rinse White Used with diluent to prevent salt built-up on aperture. 800 ml
Cleaner Dark Blue Cleaning solution of the fluidics, connected to the dedicated rear 300 ml
reagent input.
The analyzer uses this reagent from time to time, to clean the tubing,
counting chambers and apertures.
Lyse Yellow Lyse reagent is connected to the rear reagent input. 300 ml
It creates hemolysate for 3-part WBC differential, total WBC count
and HGB determination.
Lyse2 Orange Used to dilute whole blood and differentially hemolyse white blood 800 ml
cells to separate eosinophil granulocytes from other WBC by
volume. Suitable for determining EOS, EOS%, BAS, and BAS%
parameters.
24 | P a g e
Abacus Vet5 User’s Manual
Control material
Hematology controls are used to check precision, accuracy and to calibrate the system, if necessary.
Performing QC determinations regularly verifies continued optimal performance.
The Abacus Vet5 is pre-programmed to monitor DiatroCont 5 control. This should be matched to the types
of samples usually run on the instrument. Specification for this material (assay values and allowed tolerances
along with expiry date) is always enclosed with the approved control material.
DC55021 - DC5 Hematology Control 4,5 ml low
DC55022 - DC5 Hematology Control 4,5 ml normal
DC55023 - DC5 Hematology Control 4,5 ml high.
Control assay is refreshed by 90 days, please contact Diatron MI Plc for actual information.
Using Control and Quality control support see more at chapter 12.
Principles of Operation
Abacus Vet5 analyzer uses a combination of methods to determine measurement results:
Volumetric impedance method is used to determine the cellular concentrations and volume
distributions of erythrocytes (RBC), and platelets (PLT) from so-called RBC dilution.
Volumetric impedance method is used in combination with a lytic reagent to determine distribution
of leukocytes (WBC), and their three sub-populations (LYM, MON, NEU, EOS, BAS).
Photometric measurement of light absorbance of the WBC dilution determines hemoglobin (HGB)
concentration of erythrocytes (RBC).
25 | P a g e
Abacus Vet5 User’s Manual
A constant direct current flows between the electrodes on both sides of the aperture. Each cell passing
through the aperture causes a change in the electrical impedance of the conductive blood cell suspension
(diluted blood). This impedance change is detected by the Abacus Vet5 electronics and converted to an
electrical voltage pulse.
The number of pulses generated are proportional to the number of particles in the diluted sample.
Intensity (amplitude) of each voltage pulse is proportional to the volume of the particle. The volume
distribution diagrams of the particles are displayed as the WBC, EOS, RBC, and PLT histograms measured in
femtoliter (fl) units.
26 | P a g e
Abacus Vet5 User’s Manual
Electronic discrimination by size allows separation of platelets (PLT), erythrocytes (RBC), white blood cell
(WBC) and WBC differential populations. Discriminators are indicated by dotted vertical lines on the
histograms.
27 | P a g e
Abacus Vet5 User’s Manual
The membranes of the WBCs become selectively permeable, so that they begin to shrink down to their nuclei
in the slightly hypertonic lyse solution. Effectively hemolysed samples contain WBC particles in the 30–300
fl region (for veterinary species).
For the three-part counting a primary dilution of 1:160 is created by diluting 25 µl of whole blood with 4 ml
of diluent into the chamber.
After taking the RBC sample of 25 µl, the remaining diluted blood is treated with 0.5–0.7 ml of lyse reagent
— this depends on the animal profile selected — to destroy red blood cells. The remaining solution is suitable
for photometric measurements of HGB, and counting of WBC. After these measurements done, the software
is able to determine HGB, WBC, LYM, LYM%, MON, MON%, and total GRA, and GRA%. GRA contains all types
of granulocytes: GRA = NEU + EOS + BAS, and GRA% = NEU% + EOS% + BAS%. Volume distribution of three-
part measurement can be seen on the three-part histogram. The cells are separated in three regions
depending on their sizes: LYM, MON, and GRA, from left to right, separated by discriminators.
Five-part WBC differential results are determined during two separate dilutions, and two counting sessions
after each other. (See sequence numbers in Figure 6 , Figure 7) The first session is used to count EOS only,
while the second is the three-part differential and RBC counting as described in previous point. In three-
part only measurements, the EOS counting cycle is omitted.
28 | P a g e
Abacus Vet5 User’s Manual
To determine EOS, EOS% and BAS, BAS%, a second sample preparation and counting is required. 25 µl of
whole blood sample is diluted with a species-dependent volume of lyse 2 solution, a hemolytic reagent and
diluent, to form a 1:200 dilution. During the incubation time, white blood cells will be differentially
hemolysed, so that eosinophils will retain a higher cellular volume.
After counting and sizing the cells, the software will interpret all cells above the discriminator as eosinophils.
The user can observe the distribution of the cells on the EOS histogram.
The number and percentage of BAS cells will be calculated using EOS and other internal parameters using a
mathematical formula.
The EOS count is then followed with a three-part differential count (described above) to provide the rest of
the WBC count, the RBC and the PLT and related parameters.
29 | P a g e
Abacus Vet5 User’s Manual
Measurement Sequence
Stages of blood testing process during normal (5 part) mode:
1. The sampling needle moves into the sample vial, and takes 25 µl of whole blood sample.
This 25µl of whole blood sample is transported to the chamber through the sampling needle
2 with an addition of Diluent, and Lyse2 reagents to form dilution for the 5-part
differentiation measurement.
After the first measurement and washing process, the sampling needle moves again into the
3
sample vial, and takes again 25 µl of whole blood sample.
The 25µl of whole blood sample is transported to the chamber through the sampling needle
4
with an addition of 4ml Diluent, to form 1:160 MIX dilution.
25µl of this diluted sample is aspirated back into the needle and stored in the needle during
5
WBC measurement and the hemoglobin analysis.
Lyse reagent is added to the MIX dilution for WBC measurement. Abacus Vet5 uses bubbles
6 for mixing lyse with MIX dilution. This lytic dilution is suitable for photometric
measurement and WBC counting.
After the measurements the system drains and cleans the measuring chamber and the
8
related tubing system. The aperture gets cleaned with high voltage burning pulses.
The 25ul MIX sample (stored in the needle) is added with 5ml diluent to the chamber. This
9
forms a 1:32000 overall RBC dilution.
10 With this dilution the RBC and PLT measurements are performed.
30 | P a g e
Abacus Vet5 User’s Manual
Front view
Thermo printer
LCD
touchscreen Sample rotor
station
Reagent key
connector
Rear connectors
Power inlet
Device
identification
USB A ports
LAN port
Reagent
Connectors
USB B port
31 | P a g e
Abacus Vet5 User’s Manual
Diluent inlet
Lyse inlet
Lyse2 inlet
Cleaner inlet
Side views
Service door
with lock
Reagent key
slot
USB A port
32 | P a g e
Abacus Vet5 User’s Manual
Open the paper lid (pull the lid upwards by the handle)
Remove central plastic roller of old paper roll
Unwind new paper roll, so that the “starting edge” is
coming from down under towards you.
Gently drop the new roll into the holder of the printer, hold
“starting edge” with your hand, and make sure it comes out on the
front of the printer as indicated in the picture
Close the lid, making sure that the paper captured between
it and the printer mechanics, and it is securely closed
33 | P a g e
Abacus Vet5 User’s Manual
6 User Interface
The user interface of the Abacus Vet5 analyzer is designed to be intuitive and easy to use, and provide access
to all user level functions. You can interact with the Abacus Vet5 by using only the touch screen and the
virtual on-screen keypad, or by using an external keyboard.
Selectable menu
buttons
Title of Menu or
function
Background image
(Inactive area)
Context sensitive
function buttons
34 | P a g e
Abacus Vet5 User’s Manual
Button color
Active buttons are displayed in full color, while inactive or not selectable buttons remain faded.
Entering Information
Abacus Vet5 sometimes requires the operator to enter information such as sample ID or patient data. In
order to enter information into a data field simply tap or click on its box. The data field will change color
indicating that data entry is active for that field. Information can be entered using the virtual on-screen
keypad or with an optional external keyboard.
Data is not saved until user presses Enter or Save or Accept to proceed.
Menu System
Abacus Vet5 offers a menu tree to access all functions. Tree items can be activated by pressing/tapping on
them directly on the touchscreen or rolling through items by keyboard.
Main Menu
The Main menu will appear after start-up or when Home button is pressed in the menu tree.
35 | P a g e
Abacus Vet5 User’s Manual
Menu Tree
This chapter is a listing of the Abacus Vet5 analyzer software menu trees.
Measure menu
1.Measure opens result 1.Run new sample 1.1.Run
screen screen asks for:
-Type 1.2.Blank
-Patient ID 1.3.Worklist
-Name 1.4.Normal ranges
2.Re-Run -Age 1.5.Cancel
3.Tech. -Gender
4.Print -Doc
5.Discard -Predilu YES/NO
6.Exit -Lyse volume
-Samp depth
Database menu
36 | P a g e
Abacus Vet5 User’s Manual
Maintenance menu
2.Calibration 1. Factors
2.Measure
3.History
4. Prediluted Factor
History
37 | P a g e
Abacus Vet5 User’s Manual
38 | P a g e
Abacus Vet5 User’s Manual
Settings menu
1.2.Format 1. Limits
2.Warnings
3. Histogram
12.Tech info
13.Diagnostic Flags
14. Header 1,2,3,4,
39 | P a g e
Abacus Vet5 User’s Manual
Autosend
Disabled
PC only
EMR only
PC and EMR
USB serial
Exit menu
6. Exit 1.Shutdown
2. Prepare for shipment
40 | P a g e
Abacus Vet5 User’s Manual
Turn on Abacus Vet5 using the main power switch on the back, above
the DC power input connector (see section 5.6.2).
‘On’ position is indicated by the ‘I’ symbol.
During start-up, the following screen will appear.
The software loading finishes at the Main Menu.
If you wish to browse the database, then no further user action is
required.
If you wish to run a sample, then go to Measure menu. Starting a new
measurement will perform priming of reagents and a cleaning cycle
automatically.
After powering on, wait 5 minutes before initiating any measuring process to allow the
instrument to reach optimal working temperature.
In some cases, a priming cycle is necessary prior to sample measurement. The analyzer will perform prime
cycles automatically if the fluid sensors are turned on, and additional liquid in the fluidic system is required.
Run priming cycle in the following circumstances:
After installation;
After extended time out of use;
Replacement of any component related to the Fluidic System;
Replacement of reagents with the analyzer turned on.
See more about priming at 11.4.1
41 | P a g e
Abacus Vet5 User’s Manual
Do not turn off the instrument by simply flipping the power switch on the rear panel. Doing
so, may result in erroneous operation, loss of data or prolonged start up cycle during later
use.
The built in Shut down procedure will set the analyzer electronics, fluidics and mechanics in a definite state:
Abacus Vet5 saves data, settings, closes the databases and other files. If the power is simply cut before
all data items are securely saved, some data might be lost.
The SW will perform cleaning to remove debris and fill chambers and apertures to avoid clogging and
contamination. Without proper shut down, blood-sample can remain inside the instrument
increasing the risk of contamination or if dried in, blood can permanently block the sample-paths.
The analyzer will move the mechanical parts (sample needle, syringe pumps, valves) into their safe
home positions to reduce risk of injury, and avoid more time and higher amount of reagents during
the next start-up process.
Therefore always follow the instructions below when switching the instrument off.
To shut down Abacus Vet5, press EXIT key in the Main menu. The following screen appears.
Press Shutdown.
Abacus Vet5 will display a message and give a tone indicating that it is
safe to turn off power, after the shut down procedure is finished.
Turn off the instrument using the power switch on the rear panel. Off
position is marked by ‘0’ symbol.
42 | P a g e
Abacus Vet5 User’s Manual
Emergency shut-down
Please, refer to section 2.10 for Emergency shut-down procedures.
43 | P a g e
Abacus Vet5 User’s Manual
8 Sample Measurement
This chapter contains important information about running samples.
Sampling Depth
The operator of Abacus Vet5 analyzer can define the exact sample taking position inside the vial (see figure
below). This software option can be accessed at Settings -> Measure ->Needle height (see more at 14.3.3)
Sampling depth can be adjusted in the following steps: -2mm,’--’, +5mm, +10mm, +15mm.
Default sampling depth is: 0 mm, indicated by ‘––‘ in the selection list, and it means approximately 3 mm
clearance between the bottom of sample tube and the tip of the sampling needle.
Example to set up sampling depth correctly: if the bottom of the tube is lifted by 12 mm, select +15 mm
setting to make sure the needle will not intend to touch the bottom of the sample tube.
If sampling depth is not adjusted to the tube correctly, the mechanics will push the
bottom of the tube, which may result damage to the sampling needle, or pneumatic error
of Vertical motor.
44 | P a g e
Abacus Vet5 User’s Manual
Microtainer adapter
Below you can see how to use 0,5 ml microtainer tube in adapters.
45 | P a g e
Abacus Vet5 User’s Manual
46 | P a g e
Abacus Vet5 User’s Manual
All blood samples must be handled as potentially infectious material that represent
biohazard.
47 | P a g e
Abacus Vet5 User’s Manual
Sample Identification
Use caution and ensure accurate data entry when manually entering sample ID or
patient ID information into the user interface of the Abacus Vet5 analyzer to avoid
possible misidentification of sample results.
The following sample information items are used to uniquely identify samples:
Sample Mode (Blank, Control, Profile, pre-diluted, etc.)
Time of measurement: assigned automatically by the Abacus Vet5 analyzer SW
Sample ID: assigned by the operator – or automatically assigned by Abacus Vet5 SW
o Abacus Vet5 doesn’t require the sample ID to be unique
o Sample ID can submitted by:
Manual typing
Bar code reading (with a manual bar code reader, see more at14.2)
Automatically assigned by Abacus Vet5 SW
o Sample ID is not applicable on blank measurements
Patient ID: assigned by the operator by manually typing in
o Not applicable for blank and QC measurements
o Can be changed in the database later on
48 | P a g e
Abacus Vet5 User’s Manual
Running Samples
After powering up Abacus Vet5, you need to perform the following steps in order to run samples.
Ensure that:
All reagents and the waste are properly connected
All environmental conditions are met
Blank measurement has been run and accepted (see section 8.6.1
below)
Quality Control material have been tested and produced
acceptable results, so that calibration of the analyzer is valid (see
chapter 12)
Specimen is ready to use, mixed well and it is in the recommended
time for measurement (see section 8.3)
49 | P a g e
Abacus Vet5 User’s Manual
Blank values over high limits will not be displayed and “---” symbol appears instead. See more at 9.2.2
If the blank values are within the acceptable ranges, press Accept to
store the blank values.
If one or more parameters are out of range, repeat Blank measurement
2-3 times, by pressing Re-blank key.
If the Blank is stable high for certain parameters, run Cleaning or Hard
cleaning in Maintenance menu (refer to section 11.1).
Press Print to print blank results.
Press Back to return to measure menu.
50 | P a g e
Abacus Vet5 User’s Manual
Do not reach inside the instrument during sampling because the needle can cause injury.
51 | P a g e
Abacus Vet5 User’s Manual
After about 3,5 minute the analyzer will display the results. You
can remove sample from holder now
52 | P a g e
Abacus Vet5 User’s Manual
Select sample ID from list for actually processed sample, or scan the
bar code of a sample from the list to automatically set Sample ID and
go to run mode.
Press Run to measure EMR (remote) identified sample.
Notice: for more detailed information about messaging protocols and data transfers between analyzer and
EMR system see the Diatron Communication Protocol document.
53 | P a g e
Abacus Vet5 User’s Manual
Make sure to use physiological saline and not diluent to prepare diluted samples, otherwise
some parameters may change due to the stabilizing effect of diluent.
54 | P a g e
Abacus Vet5 User’s Manual
9 Interpretation of Results
This chapter provides description about interpreting and understanding results and flags.
Results Screen
Result screen of Abacus Vet5 is displayed in 2 separate screens (Parameter and Technical details), where the
header section is same.
Displayed information
Sample identification header (marked with red on the left )
consists of:
PID – patient ID, date and time of sample run, SID – sample ID
and sample Type, WARNING – flags, LYSE setting, Pre-diluted
and Needle settings will also appear here.
Values within the normal range are displayed white digits, values
outside the normal range are displayed with red and flagged:
“– “ if lower than the normal range, or
“+” if higher than the normal range specified by the measurement
type. (area marked white)
Histograms(marked with orange): Graphical information about
volume distribution of WBC, RBC ,EOS and PLT. Histograms
contain vertical discriminator lines that separate different cell
populations.
The context sensitive buttons provides the following further operations :
Run, initiate the next sample measure
Re-Run: does repeat the current sample, the current results will be saved.
Discard: results will be deleted from measure database.
Tech/Result: depending on what is current screen, it switches between the results and technical
details screen. .
Print, sending the current results to default printer. Exit menu drops back to main screen.
55 | P a g e
Abacus Vet5 User’s Manual
Context sensitive buttons – at the bottom of the screen – work same way as described in the previous
subchapter.
There are numerous diagnostic and warning flags, see more details at the next chapter, .
56 | P a g e
Abacus Vet5 User’s Manual
Measurement Flags
This section describes diagnostic and warning flags indicated by Abacus Vet5.
57 | P a g e
Abacus Vet5 User’s Manual
58 | P a g e
Abacus Vet5 User’s Manual
Important: in case a flagged blank measurement – beside the above ranges - is accepted, then all next
patient measurement will be flagged with same warning flags. In exterme case even the measured
values also will be masked with ‘*’ instead of displaying that parameter.
Tip: Perform cleaning,then run Blank measurement again to correct blank measurement flags. See more at
Troubleshooting guide at chapter 16.4
k RBC pulse peak too small Check aperture or get contact service
l PLT limit exceeded Run a cleaning, check reagents. Change diluent
Table 6 summary of RBC/PLT measure related result flags
59 | P a g e
Abacus Vet5 User’s Manual
Vortex sample for 5–10 seconds at high speed and re-run. Per-
form manual smear to examine cells.
Y EOS clogging Aperture clogging. Perform 3 to 5 cleaning cycles and repeat
the measurement. If the problem persists, contact Technical
Support.
Low-temperature reagents (mainly diluent) can cause this. Be
sure to allow reagents to reach room temperature
Z EOS cellcount too high, noise Possible lysing problem. Make sure there is adequate lyse 2
in EOS channel reagent in the container, and prime if needed
60 | P a g e
Abacus Vet5 User’s Manual
61 | P a g e
Abacus Vet5 User’s Manual
62 | P a g e
Potential sample induced interferences
65 | P a g e
Abacus Vet5 User’s Manual
66 | P a g e
Abacus Vet5 User’s Manual
10 Database Operations
Patient results are stored in the analyzer’s memory in chronological order, and can be viewed in the database
at any time.
Memory capacity is 5.000 patient records, including all parameters, histograms, flags, sample ID and
date/time of measurement.
If storage space is full, the latest (actual) record will overwrite the oldest record in the Database.
We suggest to use data backup feature to archive record regularly. (se more at 10.4.3)
67 | P a g e
Abacus Vet5 User’s Manual
The user can switch between Parameters and Tech detail views by pressing the Tech/Result button.
68 | P a g e
Abacus Vet5 User’s Manual
69 | P a g e
Abacus Vet5 User’s Manual
A numerical input pop-up screen will appear after pressing e.g. Date to
field.
Type the desired date and press Enter. ‘/’ character will appear
automatically. Use Delete to delete the last character.
Press Cancel to discard changes.
Data input is similar for Patient ID and Sample ID fields as well.
Selection on filtering can be applied on the whole database (All) or on
the already Selected records only. Push the Records field to toggle
between these two states.
Press Type field (see below) if you wish to filter/select specific breed
records of measurements. The following selection screen will appear.
70 | P a g e
Abacus Vet5 User’s Manual
Backup button will open back-up options menu. Selected records can
be saved in different file formats onto USB mass storage device
connected to the analyzer, such as a USB pen drive.
71 | P a g e
Abacus Vet5 User’s Manual
Press Abort to stop sending the data record, and return to Manage
menu or wait until transmission is over.
72 | P a g e
Abacus Vet5 User’s Manual
Before proceeding with backup function, insert a pen drive into any of USB A ports.
Select one of the Backup options below:
Save: to save selected records in .dat format. Data files that saved
in this format can be viewed later, using Backup view function.
Export: to export selected records as .txt file. In this ASCII format,
data can be easily imported into Excel and other programs.
(Refer to section 14.1 to set up format of text exported to file.)
Backup view: function will open .dat format data files previously
saved onto a pen drive. See section 10.4.4 for details.
Backup one day: function will allow you to backup data record of
a specific day in .dat format. See section 10.4.5.
Press Home to go to Main menu.
Press Back to return to Manage menu.
Wait until the save operation is completed (progress bar is over), check
the pen drive if there is further activity (LED flashing).
Remove pen drive from USB port if no activity detected to avoid data
loss.
73 | P a g e
Abacus Vet5 User’s Manual
The saved backup data is not merged with database which is stored on
the analyzer.
The data opened in this backup view mode can be selected for view, print
or transmission.
Insert the pen drive containing saved data into any USB A port
Tap Backup view in Backup menu.
The analyzer will list saved database (.dat) files found. Use the
arrows to select the specific file to view
press OK to view records, the software switches to database view
to browse records
After opening the record database from pen drive, the following menu appears. The header will show the
actually displayed backup file name.
Records loaded from the USB storage device can be displayed, printed,
filtered, managed, and statistic trends could calculated just like for
normal database records.
74 | P a g e
Abacus Vet5 User’s Manual
Tap Backup one day button in Backup menu, and follow the steps
below:
Press Cancel to return to Backup menu without saving data onto the
pen drive.
Exporting database
This feature allows to use mesured data in 3rd party systems, for
example to import in spreadsheet application.
Connect a pen drive into any USB A port.
Tap Export button on Backup screen.
Wait until progressbar is shown and action is finished.
Press Enter, then tap Accept. The SW will save data records in .txt
(Tab separated) format.
The ongoing export process is indicated by a processbar on screen.
Press Cancel to return to Backup menu without saving data onto the
pen drive.
75 | P a g e
Abacus Vet5 User’s Manual
Use the white arrow icons on screen to move through all L-J chart
screens .
76 | P a g e
Abacus Vet5 User’s Manual
77 | P a g e
Abacus Vet5 User’s Manual
11 Maintenance menu
Maintenance screen collects the followin analyzer features : Cleaning,
Calibration, Diagnostics, and Reagent Status menus.
These feature groups will be described through this chapter.
To open maintenace group tap on the Maintenance button on the Main
screen :
Cleaning menu
Cleaning: Starts one wash cycle using the cleaning reagent connected
to cleaner input. A cleaning cycle is automatically performed during a
start-up and a shutdown procedure.
Clean the wash head: It is recommended to use this option when
executing maintenance..
Hard Cleaning: It must be performed weekly as part of routine
maintenance. Clogging errors or high PLT blank indicates the need for
additional hard cleaning.
Bleaching: Used only on demand, during troubleshooting when the
Blank results are constantly high.
Drain Chamber: Used during troubleshooting or component
replacement only.
78 | P a g e
Abacus Vet5 User’s Manual
Hard cleaning
Analyzer asks for about 2ml cleaning solution (Diatron D801 Hypoclean CC ) in a sample vial. This chemical
will be breath through sampling needle, filled into chamber then diluted. After a few minutes waiting it is
flushed out regularly, then a clean cycle is executed automatically.
Hard cleaning is part of the weekly maintenance.
Bleaching
This feature allows an overall hypo solution cleansing of the whole fluidic system. Detailed step by step
procedure is available at 15.2.3.3, Maintenace procedures.
Drain chamber
Used during troubleshooting or component replacement only. Drain of the chamber should be run before
removal or replacement of parts related to the measuring chambers or apertures (Service personnel task).
79 | P a g e
Abacus Vet5 User’s Manual
Calibration Menu
In Calibration menu, the operator can initiate calibration of Abacus Vet5 or can review calibration history.
Prior to Calibration, a Self-Test and a Blank Measurement must be performed with acceptable results.
Calibration is required at new installation or when the control values are out of range. Also obligatory if any
of measure hardware component is replaced.
Press Factors to check or modify Calibration factors manually.
Tap Database to see calibration database.
Push Calibration with measurement to run calibrator.
Tap History to see history of calibration events by factors.
Pressing Prediluted enters to Prediluted calibration menu. Prediluted
calibration factors are used to calculate results measured in Pre-
diluted mode. Refer to section 8.6.5 for details on pre-diluted sample
mode.
80 | P a g e
Abacus Vet5 User’s Manual
Press Accept when all factors are recorded to save the new values.
This will generate a new event in the calibration history log, with the
current date and time, by storing the new factors. See calibration
history in section 11.2.3.
Press Print to print the factors.
Press Cancel to discard changes and return to Calibration menu.
After the new factors are recorded and accepted the software drops
back to Calibration screen.
Note: calibration factor can be modified from 0.7 to 1.3. All in all,
three factor is existing: factory setting, service factor, user factor.
81 | P a g e
Abacus Vet5 User’s Manual
Mix the calibrator thoroughly, insert it into the sample adapter and
press Run when ready to start-
Target values, means of runs, CV% and the new calculated factors will
appear (see screenshot to the left).
If CV% (coefficient of variance) values and the new factors are
satisfactory, press Accept to save the new calibration factors.
Calibration of the analyzer is ready.
82 | P a g e
Abacus Vet5 User’s Manual
Calibration History
This menu contains historical calibration information. The changes of the factors by time can be followed.
Use the arrow icons on screen to go through the set of factors
83 | P a g e
Abacus Vet5 User’s Manual
Prediluted calibration factors and history menus. See sections 11.2.1 and 11.2.3 for details on use.
84 | P a g e
Abacus Vet5 User’s Manual
Diagnostic menu
The most important service functions are collected under this menu group:
85 | P a g e
Abacus Vet5 User’s Manual
Statistic menu
In current menu the user can follow the numbers of most important events as:
- Number of measured samples
- Number of reagent changes.
- Startups and shutdowns.
- Clog events
- Wash head cleanings initiated
- Pneumatic events
- Internal errors reported to software
These numbers gives a clean picture how operator used the machine,
if the necessary maintenances has completed, if the system was
shutdown regularly. Large difference between startups and
shutdowns might indicate unreliable power mains beside several
power-cuts.
It is also important that the logged measure numbers projects when
the preventive maintenance is necessary.
This screen is available for standard level users as well, but only
service account can clear the event log statistic.
86 | P a g e
Abacus Vet5 User’s Manual
The stress cycle results also can be printed or saved into external USB data device. IMPORTANT: Blank data
is not stored and will be deleted if leave this screen with Exit command.
After pressing the Start button, the results of test will appear in the middle column, between target values.
Result will be indicated in red or green color depending it has passed or failed.
Self test results can be printed or saved, using the Print or Save buttons accordingly. To be able to save list,
an external USB storage must be present, otherwise system drops a warning.
87 | P a g e
Abacus Vet5 User’s Manual
88 | P a g e
Abacus Vet5 User’s Manual
89 | P a g e
Abacus Vet5 User’s Manual
Red lines represent the least and largest measured values that has
occured in the record set. Trend line itself is drawn by black.
At the end of the diagram the number represents the number of
points the diagram calculated of. (In this example its 53)
The numerical values of individual parameters are written onto
the white box located at end of the chart.
90 | P a g e
Abacus Vet5 User’s Manual
Reagent Status
This menu helps in monitoring reagents, set up and check lot numbers,
expiration date, and actual level.
There are individual sensors for the all input reagents, which can be
turned On/Off by switching the sensor icon by pulling left/right.
For priming press Prime at the selected reagent, or Prime all to prime all
reagents. Reagent sensors will control priming.
Go to Replace menu, if you wish to replace the reagent pack (see later
more).
Waste and reagent level % calculation is based on the preset of the tank
capacity. (On this analyzer only waste tank volume can be adjusted)
Pressing Volume button, you can redefine the volume of the waste
container.
If you receive reagent sensor error messages, then run Calibrate sensors
to reset sensors (see section 11.4.4 below).
Tap Measurement credits to check measurement credit counter (see section)
Pressing Reset will restore Waste counter to 0%. Make sure to dispose of the contents of waste container.
91 | P a g e
Abacus Vet5 User’s Manual
Replacing Reagents
As first change reagent containers then connect them to analyzer according to reagent pack installation
instructions. Then follow the steps below to introduce new reagent set to software. All reagent replacement
event is logged.
Step 1. From the Main Menu screen select Maintenance -> Reagent
Status function.
Step 2. Insert the hardware key to the marked side connector of the
front plate. Key always enclosed to Lyse bottle.
When you select Replace the instrument will look for a reagent key
(HK).
Step 4. Confirm that pack has already changed and pickup tubes
are connected to inlets pressing the big Change pack button.
Hint: the empty diluent container of the previous pack can be used
further as new waste container.
92 | P a g e
Abacus Vet5 User’s Manual
The key (above) is the part of Lyse reagent and enclosed to reagent pack. If key is missing from reagent pack
then contact your sales representative. Connect it only to the special USB connector where it is indicated
(front side connector).
In case a used key is inserted into system the enclosed message
appears (see left). If this happens, insert an unused key and press
Retry.
All measurement cycles will decrease the number of measurement credits, except for
initialization, standby and wake up cycles.
93 | P a g e
Abacus Vet5 User’s Manual
As a result, the below status screen appears, both leftover credits and
HK current credit content is displayed:
94 | P a g e
Abacus Vet5 User’s Manual
Liquid sensors
Abacus Vet5 is equipped with four channel liquid sensor that can check all input reagents and can be turned
On/Off in Reagent status menu. (Notice: Diluent and Cleaner shares on the same optic sensor but using it
beside different scheduling)
You can perform measurements (animal sample, QC, calibration, blank) only if all the
liquid sensors are turned on (enabled), i.e. their check boxes are ‘On’.
Use Calibrate sensors function to start a procedure to readjust trigger/threshold setting of the liquid
sensors.
95 | P a g e
Abacus Vet5 User’s Manual
Target values of the control material should be set only once, when you start using the
given lot. Resetting parameters deletes previous QC results of that level.
Any change in the QC material setting deletes previous QC results. It is strongly
recommended to print QC results prior to changes.
In QC menu the operator can enter a new lot of controls (by scanning the barcode, loading from data file or
by entering the assay data from the assay sheet manually), review data of previous control lots, and print
Levy-Jennings graphs. For general instructions on running samples, refer to Chapter 8.6.
Always follow the manufacturer’s instructions for warming and mixing of the QC material
before use. Instructions are listed in the package insert instruction sheet.
Keep track of the expiration date and open-bottle stability of the QC material.
Please, remove the cap before running the control.
Performing QC measurement
The QC control screen divided to three identical groups, where the low,
normal and high level control measurements are handled individually.
Open the next screen tapping any of the three buttons to start work with
the Quality Control blood. (see next page)
96 | P a g e
Abacus Vet5 User’s Manual
Opening one of the QC levels, the new screen provides the following
functions for the operator:
References: opening this function the reference data sheet of the QC
material opens and values can be entered. (see next)
Measure: this button initiate the measuring process.
Diagram: opens the actual QC level Levey Jennings diagrams.
Database: opens the current QC level measurement history giving
access to data records.
Home button drops the Main screen, Back button guides one menu
level back.
If you have a target value file on a USB pen drive for this lot, then
connect the pen drive and press Load button. Contact your local
distributor for proper data file (e.g. QC_31611.ini’).
When finished, press Accept to save QC lot data, or Cancel to return
to previous menu without saving changes.
97 | P a g e
Abacus Vet5 User’s Manual
Important note: as the analyzer handles three and not more QC level all the time, when new LOT number is
recorded the following warning message will be shown :
98 | P a g e
Abacus Vet5 User’s Manual
Performing a QC measurement
After the target data saved , the software returns to QC main screen.
Tap on the Measure button, a standard measure screen opens .
Put the prepared control blood into sample holder, then tap on Run.
The measurement is performed, the record will be saved standard
measure database indicating the measure type is ‘QC Low’, ‘QC Normal’
or ‘QC High’
If QC measurement results are not satisfying, then you can repeat it
using the Discard button.
You can switch to technical screen with Tech. button or immediately
jump to LJ charts pressing the Diagram button.
Levy-Jennings QC chart
Levy-Jennings charts of the actual LOT are displayed after opening this
database from QC main screen with Diagram button or jumping here
from the QC measure screen.
The mean and the specified range are displayed on the right for each
parameter.
Press Print to print table or Levy-Jennings charts.
Use ⇦ and ⇨ to move between data points.
Use ⇧ and ⇩ to browse between parameters.
Press Database to see the QC measurement results in database view.
Refer to Chapter 9.4. on database options.
99 | P a g e
Abacus Vet5 User’s Manual
QC database
The former QC measurement records of current LOT can be overviewed
or managed opening the QC database screen.
Use Quality Control -> QC<type> -> Database button to open this list.
Handling the records here is same as described in 10.4, Managing
records chapter.
Deatails button switches to result screen to look over measured results.
Trends button switches to Levey Jennings chart, as described in the
previous chapter. Filter and Manage buttons working same way as it is
described in 10.4, Managing records chapter. You can Print results also.
The Exit button switches back to QC main screen.
Important note. This database flushed each time if reference data saved
on same LOT number or existing LOT reference value is modified.
100 | P a g e
Abacus Vet5 User’s Manual
13 Multi-User Mode
Multi-user mode provides the clinical laboratory manager the means to control access to the Abacus Vet5
analyzer by ensuring that only authorized users can operate the analyzer. With the multi-user mode, the
laboratory manager can assign some users the authority to perform critical system functions such as
maintenance and calibration, and restrict other users to routine activities such as running samples.
Types of Users
Multi-user mode provides three types of users:
Admin (administrator) users have access to all system functions.
Advanced user can access all functions, except for setting up users.
Basic users can run samples, but cannot perform calibration and system settings.
NOTICE: special functions and extended software functions are available for Service account. See service
manual for further details. Only trained service person is entitled to act as service user on the analyzer.
Logging in
The operator has to log in using its password in order to use the
analyzer. Users can be set up by the person having Admin access to the
analyzer.
In Exit menu, press Log in to start log in.
In order to manage users, Admin level of user has to log in.
101 | P a g e
Abacus Vet5 User’s Manual
User Management
After logging in as Admin user, the User management menu is
available to set up and modify users.
Tap Add new user to set up a new user account.
Press Remove user to delete an existing user account.
Tap Automatic login set, to set up the default user.
Tap Edit user to modify an existing user account.
102 | P a g e
Abacus Vet5 User’s Manual
103 | P a g e
Abacus Vet5 User’s Manual
Autologin setting
Press the press Automatic login set button on the User Management screen. A selection window pops up.
If automatic log in feature is set for any of the listed users, the analyzer will all the time log into that account
as default.
Step 1. Log in as Admin and press Automatic login set. The following menu
appears.
Step2. Select the user account for automatic log in.
By tapping Automatic login off the operator can turn off this feature, then
the analyzer will start with the Exit menu to start with Log in.
Be careful to set Admin as automatic log in user, because then all multi-user modes will
be available, including editing Admin password.
104 | P a g e
Abacus Vet5 User’s Manual
14 Settings menu
From the Settings menu five more submenus can be accessed with custom options.
105 | P a g e
Abacus Vet5 User’s Manual
Printer settings
Enter this menu to select printer used for printing and other settings related to printing.
There are the following options available:
Tap Device to select printer and related settings, or
Push Format to adjust individual fields printed
Home button takes you to Main menu
Back button will step back to Settings menu
Under Device menu tap the white fields to modify the followings:
Printer: USB or Builtin.
Mode: (only selectable if Printer is
USB). Normal b&w, Draft b&w,
Normal color, Draft color
Paper: A4 or rollpaper.
Unit: cm or inch
Top margin, Left Margin: value is
given by Unit setting (cm or inch).
Automatic printing: if Enabled, the
analyzer will print all measurement
results automatically after each
measurement.
In printer format menu, tap the
white field to Enable/Disable option.
Limits: if Enabled, normal ranges will be printed on the report
Warnings: if Enabled Warning flags will appear on printout
Technical information: controls printing of lyse volume and probe
voltages
Diagnostic flags: if Enabled diagnostic flags will be printed
Operator can set up text printed into the Laboratory headers.
Header lines 1..4: can contain customized text, name of lab, address,
telephone number, contact person, etc.
Tap the appropriate white field to enter or modify text.
Press Accept to save settings and exit, or press Cancel to exit without
saving changes.
106 | P a g e
Abacus Vet5 User’s Manual
General settings
In General settings screen the following options can be set up or modified:
107 | P a g e
Abacus Vet5 User’s Manual
Measurement settings
This menu collects the options related to measurement data formats and
ranges.
The below options are available:
Units submenu to define base units
Normal ranges, to define limits
Needle settings: to adjust sample taking position inside the vial
Press Home to return to Main menu.
Press Back to return to Settings menu.
Tap the white fields to change and Accept to save changes made.
108 | P a g e
Abacus Vet5 User’s Manual
Needle settings
This menu serves to adjust sample taking position of the needle
inside the vial. The needle tip can be shifted from -2mm to + 15 mm,
related to default position. (-2 mm means lower, + range means
higher position)
Tap on the required needle height , then accept setting tapping to
Accept button below.
109 | P a g e
Abacus Vet5 User’s Manual
110 | P a g e
Abacus Vet5 User’s Manual
Communication settings
Communication settings menu serves for setting up external computer connection for data transfer.
Tap onto Communication in Settings menu to enter Communication
settings menu.
There are two submenus available:
Device communication – this menu allows the operator to set up LAN
based communication option.
Device communication
Tap Device communication in Communication settings menu to enter
this menu. The following settings can be modified:
DHCP : Disabled or Enabled. If DHCP is enabled, the master device
on the local network will assign an IP address, subnet, gateway and
DNS for Abacus Vet5 automatically.
IP address: If DHCP is disabled, enter a valid IP address in order to use
Abacus Vet5 in the network. Contact the network administrator for
details on IP address in your laboratory network.
Gateway: typically set to 0.0.0.0 (when no DHCP)
DNS: typically set to 0.0.0.0 (when no DHCP)
Current settings are shown for user reference below in shaded area.
Press Accept to save settings and return to previous menu.
Press Cancel to exit without saving changes.
111 | P a g e
Abacus Vet5 User’s Manual
Peer communication
Tap Peer communication in Communication settings menu to enter this menu. The following settings can
be modified:
PC: communication with one PC can be Active or Inactive.
PC remote IP address: A valid network address should be set up here if
PC setting is Active, otherwise the analyzer cannot transmit data to the
target PC.
EMR: EMR (remote work list mode) can be Active or Inactive.
EMR remote IP address/EMR remote port: A valid network address of
the LIS server should be set up here if EMR setting is Active, otherwise
the analyzer cannot transmit/receive work list to/from LIS.
USB-Serial link: Can be Enabled or Disabled. If a PC is connected via the
USB-B port, this setting should be Enabled.
USB-Serial Baud Rate: To connect PC via USB B emulated serial link, set
up baud rate here identically to the setting of the connected PC serial
link baud rate.
Automatic send: This setting can be one or combination of links
enabled. After a measurement is completed, data record will be sent
automatically to the ports selected here.
Press Accept to save settings and exit or Cancel to exit without saving changes.
By tapping automatic send button, the active/enabled ports can be
configured for automatic data transmission after every measurement.
Select one of the options on the screen for automatic sending of data
records after measurement:
Disabled: no records will be transmitted automatically
PC only: records will be sent to PC connected via LAN port
EMR (LIS) only: records will be sent to LIS address via LAN
PC and EMR(LIS): records will be sent to both addresses via LAN
(Ethernet) port
USB-Serial: records will be sent on USB B port to PC
Contact service personnel or LIS administrator for more information on
network settings.
112 | P a g e
Abacus Vet5 User’s Manual
For LIS integration details please contact the device distributor or the manufacturer and refer to related
documentation: Diatron Communication Protocols. The Abacus Vet5 analyzer is HL7 v 2.5 compliant, able to
work on both Ethernet or serial interface.
After successful download, install the appropriate chip driver and make sure a virtual COM port is created
in operating system. To check this, open the Windows Device Manager Example:
113 | P a g e
Abacus Vet5 User’s Manual
114 | P a g e
Abacus Vet5 User’s Manual
15 Maintenance of analyzer
User-accessible parts
There is a side and a back service door to access analyzer’s serviceable parts. It allows the operator reach the
fluidic system and the mechanical parts for maintenance.
XY needle
movement Glass vacuum
mechanics chamber
Wash head
Sampling
micro dilutor
Measuring
chamber
HGB head
Valves
Measure probe
Valves
Reagent
sensors Dilutor block
Pump
115 | P a g e
Abacus Vet5 User’s Manual
The operator is allowed to access, maintain and check the following parts behind the side
door and the back door of Abacus Vet5:
Tubing and chambers
Wash head on the XY needle moving mechanics
The following parts must NOT be opened or serviced by the user:
power supply
instrument housing, and electronic boards
Maintenance protocols
The operator should carry out below user maintenance/check-up activities to the analyzer.
Weekly maintenance
The sharp tip of the sampling needle can cause injury or infection. Always use rubber
gloves and a cotton swab to clean the wash head to avoid injury to fingers.
The purpose of this procedure is to clean the lower surface of the washing head. During the regular operation
diluent might build up a salt layer when it evaporates.
The physical access of the washing head is supported from the software. (see below). This will force the device
to move the sampling probe to a position where it can be reach easily.
Steps:
A) Go to Maintenance ->Cleaning ->Clean the wash head menu in software
B) After activating the function the needle is moving to a safe position. Follow instruction appears on
display.
116 | P a g e
Abacus Vet5 User’s Manual
Do the cleansing operation. Use a soft wet cloth, immersed in warm tap water to remove salt build-up.
Finally confirm if manual operation is ready by tapping ‘Accept’.
Be extreme careful while hands inside the machine due to moving
parts. Remove hand before pressing “Accept”
D) Close the side door.
Hard cleaning
Go to Maintenance ->Cleaning ->Hard cleaning menu to start this procedure on analyzer.
Analyzer asks for about 2ml cleaning solution (Diatron D801 Hypoclean CC ) in a sample vial. This chemical
will be breath through sampling needle, filled into chamber then diluted. After a few minutes wait it is flushed
out regularly, then a clean cycle is executed automatically.
Hard cleaning is part of the weekly maintenance.
In case the hard cleaning was not performed for a week the software puts a warning message on user
interface.
117 | P a g e
Abacus Vet5 User’s Manual
Bleaching process
This is an overall cleansing process where the whole tube system is cleaned from residuals. Bleaching is not
part of weekly maintenance, should perform only on demand. For the process, prepare:
- Bleaching mixture at least 100ml beside concentration of 0,25% hypochlorite (5 ml 5-6%
concentrated hypo dissolved in 100ml distilled water)
- Clean distilled water 200ml
- The “Prepare for shipment” tube kit
Steps for the bleaching process:
A) Disconnect all reagents from analyzer, and connect the “Prepare for shipment” tube kit instead of
them. IMPORTANT: Leave waste line connected during the whole process!
B) Submerge the tube’s end into the bleaching mixture
C) Start process tapping on “Bleaching” button. Follow software instructions.
D) When prompted, change from bleaching solution to distilled water
E) Wait until rinse finish
F) Remove the tube kit, do not reconnect the regents still
G) Do a full priming WITHOUT any regent. This will pump out all remaining.
H) Reconnect reagents
I) Do a full priming (Maintenance -> Reagent status-> Prime all)
J) Run one clean cycle
Too long or too frequent bleaching could damage tubes and dissolved particles causing high PLT
problem.
Semi-annual Maintenance
Only certified service personnel can carry out periodic service maintenance to Abacus Vet5. The detailed
maintenance protocol is available in the Abacus Vet5 Service manual.
118 | P a g e
Abacus Vet5 User’s Manual
Disconnect Abacus Vet5 power supply from the electric network before cleaning the
chassis of the analyzer and front cover.
Clean the instrument and its power supply on the outside only, using a damp cloth with some soft
detergent.
DO NOT let liquids get inside the unit or to electrical connectors.
119 | P a g e
Abacus Vet5 User’s Manual
16 Troubleshooting
Abacus Vet5 provides all the necessary software maintenance functions required to keep the analyzer
operating in optimum working order. Despite excellent maintenance, the analyzer can still encounter
problems where operator intervention is required. Abacus Vet5 analyzer software informs the operator of
situations when further help is required. These are called error messages.
Error messages
Abacus Vet5 analyzer provides additional information in the form of
numerical codes that help the operator to identify the root cause of
the error. These error codes are important for service personnel.
They can interpret the messages and take necessary actions.
If you receive error messages with codes, please write them down. If
possible, describe what you were doing before the error appeared.
This important information helps your service engineer to help you
when necessary. The software will store and re-display every error
message, but it cannot describe the operation you were performing
when the error happened.
When an error comes up, the software usually attempt to offer a
solution, or will try to resolve the problem and retry the current
operation. If this recovery process fails, or the problem continues to
occur, please contact your service engineer.
120 | P a g e
Abacus Vet5 User’s Manual
Mechanical Problems
Most mechanical problems arise because of some physical blockage of the moving parts. Problems caused by
severe blocks are usually accompanied by a grinding noise. The noise is a result of a motor not being able to
move, but this does not usually result in mechanical damage to moving parts.
Dilutor Errors
Pinched, clogged tube around dilutor or at the rear reagent tubes – check for free flow of reagents.
Physical obstruction (foreign material) got into the tubing.
121 | P a g e
Abacus Vet5 User’s Manual
122 | P a g e
Abacus Vet5 User’s Manual
17 Performance Characteristics
Performance characteristics of Abacus Vet5 are described in below sections:
Linearity
Parameter Linearity Ranges Maximum
WBC (109/l) 0–100 150
EOS (109/l) 0.2–3.0 10.0
RBC (1012/l) 0–15 20
PLT (109/l) 0–700 1000
HGB (g/dl) 0–25 40
HCT (%) 0–100 —
MCV (fl) 30...150 —
MPV (fl) 3...30 —
Precision
123 | P a g e
Abacus Vet5 User’s Manual
Sample Stability
Long-term sample stability studies on blood specimens drawn in K2- and K3EDTA collection tubes at room
temperature showed no significant clinical variation for all parameters between 30 minutes and 7 hours post
phlebotomy. MPV results can show instability in the first two hours but are stable afterwards.
Measuring times
Interfering substances
The following substances can interfere with parameter measurement and alternate measurement
procedures may be required.
Parameter Interference
124 | P a g e
Abacus Vet5 User’s Manual
18 Print Formats
125 | P a g e
Abacus Vet5 User’s Manual
126 | P a g e
Abacus Vet5 User’s Manual
127 | P a g e
Abacus Vet5 User’s Manual
19 Specifications
Item Description
Operating conditions Temperature: 15–30 °C. Optimal temperature is 20-25 °C.
Relative humidity: less than 85% non-condensing
Storage and transport Temperature: 5–35 °C.
conditions Relative humidity: 15-90% non-condensing
Power supply unit External, auto-ranging power supply unit (GT-41069P9012-T3)
Input: 100-240 VAC, 50–60Hz
Maximum input current:1.5A.
Output: 12VDC, max. 6.5A, max. 78W
User interface Easy-to-use, menu driven user interface with a touch screen
Languages available English, French, German, Spanish, Italian, Czech, Polish, Russian
Data storage capacity 5,000 results with RBC, PLT, and 3/5-part diff. WBC histograms
Data export to external USB
Host computer USB-B port (direct connection to host computer via USB port)
interface PC, LAN: Ethernet port on the back
Data back-up method USB mass storage device up to 16 GByte size
Send data to LIS (Laboratory Information System)
Printer interface USB with support for most HP printer languages (DeskJet, LaserJet,
PCL3, PS, LIDIL)
Built-in printer 58 mm thermal printer, monochrome printout with histograms
Display 8” (173x104 mm), LED backlit, color graphics LCD
External keyboard USB connection
Software upgrade via USB-A port using USB mass storage device (max. 16 GBytes of flash
method drive)
Sampling volume 25 µl of whole blood in three-part mode, 50 µl of whole blood in five-
part mode. 50 µl of whole blood in five-part prediluted mode
of 1:6; 25 µl of prediluted blood in three-part prediluted mode of 1:6.
Chambers 1 chamber for diluting and counting
Reagent system Isotonic Diluent, Lyse, Lyse2, Rinse, Cleaner (see section 5.3)
Aperture diameter 80 μm
Throughput 20-30 tests/hour in 3 part mode, 16 – 20 tests/hour in 5 part mode
Sampling method Open tube system with automatic sample rotor
Specimen types Animal whole blood, Control blood
supported
Recommended sample K2- or K3-EDTA primary blood sample tubes
collection method
Sample identification Individual sample ID entered by user. Barcode with external manual
method barcode reader (optional accessory)
Sample types Control, Animal profiles
Clog prevention High-voltage pulse on apertures in each analysis cycle; chemical
cleaning; high-pressure back flush of apertures.
128 | P a g e
Abacus Vet5 User’s Manual
Item Description
Cleaning procedure High-voltage burst of the apertures, high-pressure back-flush, chemical
cleaning of the apertures.
Quality control Support for 5 part differential control blood. Measured QC parameters:
mean, ± range, SD and CV for all measured and calculated parameters,
Levy-Jennings charts, and separate QC database.
Calibration Automatic with 3 measurements, or manual (factorial) calibration of
WBC, HGB, RBC, RDWc, PLT, MCV, MPV and EOS absolute
Dimensions (WxDxH) 315 x 275 x 370 mm
Net weight 13 kg
129 | P a g e
Abacus Vet5 User’s Manual
130 | P a g e
Abacus Vet5 User’s Manual
20 Fluidic Schematics
131 | P a g e
Abacus Vet5 User’s Manual
21 Cabling schematic
132 | P a g e
Abacus Vet5 User’s Manual
2. 9.
3. 10.
4. 11.
5. 12.
6. 13.
7.
133 | P a g e
Abacus Vet5 User’s Manual
134 | P a g e
Abacus Vet5 User’s Manual
23 Document history
Abacus Vet5
User’s Manual
Revision 1.0
Revision history
Rev Author Changes Effective Date
Péter Zsámboki
1.0 Zoltán Képíró Initial revision 24-06-2016
Antal Uszkay
Approvals
135 | P a g e